ares 901-ZX:TR MULTITESTER OPERATION SUGGESTIONS 1 Unknown quantity. When unknown voltage or current is measured, always start with the highest range. After the first reading, the range switch can be reset to a lower range for a more accurate reading. 2 Selection of a range. For accurate reading, use the range next higher to the approximate value known. For instance, the voltage of a dry cell should be checked on the DC 2.5V range; on the 10V or 50V range, no better result is obtained. For resistance, reading is most accurate around in the middle of the scale. 3 Protective front cover, The metacrylic-resin front cover is treated with anti-electrifi- cation coating. Do not wipe it hard, or the coating will come off to cause erroneous reading. Volatile solvent should not be used to clean it; use a soft brush to remove dust. Metacrylic resin is softer than glass quite liable to get bruised. 4 Maintenance cares. Avoid placing the instrument where severe shock or vibra- tion is encountered, or where there is high temperature or moisture. Do not leave it in the direct sun or oily atmosphere. _—a 5 Measurement of small-size transistors. While testing a small-size RF germanium transistor, IcEo gradually increases heated by the current applied. It may cause the pointer to creep, and must be checked quickly. 6 Resistor burned. Not seldom some internal resistor is burnt on account of AC line voltage erroneously applied to the 250mA DC or the X1 ohm range. A resistor burnt can be replaced referring to Fig. 9 and the table. The protection device is to safe- guard the indicator from accidental damage, while some resistor may burn on account of the overload, but its replacement readily restores the instrument to normal performance. 6 NOT ; Noss Range Resistor | N GO0G 250mA DC} Nat - 0.970 100mA 7” No2 - 1.59 10mA ” No3 - 22.59 Oo imA Nod - 2250 2 x100 No5 - 3.72k2 corr ” x10 No6 - 3440 1 oN ‘N ok -340 wot Nez No3 Noa xl No7 | Fig. 9 — 2 - SUPPLEMENTARY DATA 1 Parts list. Part No. Description ZXRO1 | Resistor (1.3 -1.8k 1) for mV calibration RI ZXRO2Z | Resistor (8kQ)0.5V DC multiplier R2 ZXRO03 | Resistor (40k) 2.5V DC multiplier R3 ZXRO4 | Resistor (150k) 10V DC multiplier R4 ZXRO05 | Resistor (800k(2)50V DC multiplier RS ZXRO06 | Resistor (4MQ) 250V DC multiplier R6 ZXRO7 | Resistor (5MQ) 500V DC multiplier R7 ZXRO08 | Resistor (10MQ) 1kV DC multiplier R8 ZXRO9 | Resistor (2.75kQ) series RO ZXR10 | Resistor (2250) tmA DC shunt R10 ZXRI11 | Resistor (22.52) 10mA DC shunt Ril ZXRI1Z | Resistor (1.5) 100mA DC shunt Riz ZXR13 | Resistor (0.971) 250mA DC shunt R13 ZXRI4 | Resistor (0.0250) 10A DC shunt Ri4 2XRI15 | Resistor (11.9k1) 2.5V AC multiplier RIS ZXRI6 | Resistor (30kQ) 10V AC multiplier RIG ZXRIT | Resistor (160k) 50V AC multiplier RI7 ZXRI8 | Resistor (800k) 250V AC multiplier R18 2XR19 | Resistor (3MQ)1kV AC multiplier R19 ZXR20 | Resistor (31.8k) for AC sensitivity calibration R20 ZXR21_| Resistor (2k) for rectifying circuit R21 ZXR22 | Resistor (2k) for rectifying circuit R22 Resistor (340) Q X1 shunt = BH Resistor (344.2) Q X10 shunt Resistor (3.72k) 2100 shunt Resistor (220k) 2 X1k shunt Resistor (34. 2k) Q series Resistor (334k) 0 X10k series Potentiometer (10kQ)0 Q, adjuster Resistor (6.7k) shunt Germanium diode Germanium diode Capacitor (0.1 F) Capacitor (0.054F) Meter movement (442A) Range selector switch Front panel (501-ZXR type) Dry cell (UM-2/1.5V), 2 required Dry cell (015/22.5V) Varister Rear case (501- ZX TR type) Carrying handle Pin-type jack, 6 required Range selector knob Meter movement case Meter movement front cover Test leads, pair {——— Alligator clip cord, 2 required Polarity switch R.S_ Reference symbol

You might also like