Professional Documents
Culture Documents
&
REMstar®|Pro 2
www.respironics.com
Service Manual
This REMstar Auto with C-Flex and REMstar Pro 2 CPAP devices are the subjects of U.S. patents #5239995 and #6105575. Other patents pending.
REMstar, Encore, Encore Pro SmartCard, and Whisper Swivel are registered trademarks of Respironics, Inc. The C-Flex mark is used under license.
Respironics disclaims all liability for economic loss, loss of profits, overhead, or consequential damages which may be claimed to arise from any sale or use
of this product. Some states do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation or exclusion may not
apply to you.
This warranty is given in lieu of all other express warranties. In addition, any implied warranty, including any warranty of merchantability or fitness for the
particular purpose, is limited to two years. Some states do not allow limitations on how long an implied warranty lasts, so the above limitation may not
apply to you. This warranty gives you specific legal rights, and you may also have other rights which vary from state to state.
The warranty for repairs is ninety days for labor and one year on the part(s) that was replaced.
To exercise your rights under this warranty, contact your local authorized Respironics Inc. 1001 Murry Ridge Lane, Murrysville, Pennsylvania 15668,
1-800-345-6443 (USA and Canada only) or 1-724-387-4000.
Table of Contents
Chapter 1: Introduction ...................................................... 1-1 Chapter 4: System Setup Procedures ................................. 4-1
1.1 System Overview... .................................... .....................1-2 4.1 Overview ........................................................................ 4-2
1.2 Service Notice ................................................................. 1-2 4.2 Programming a SmartCard .............................................. 4-2
1.3 Technical Support Statement ............................................ 1-2 4.3 System Setup .................................................................. 4-3
1.4 Training ............................................................................ 1-2 4.3.1 Therapy Mode Menu Navigation .......................... 4-4
4.3.2 Therapy Mode Display Screens ............................ 4-5
Chapter 2: Warnings, Cautions and Notes ........................ 2-1 4.4 Heated Humidifier Setup ................................................. 4-7
2.1 Warnings .......................................................................... 2-2 4.4.1 Humidifier Temperature Setup ............................... 4-7
2.2 Cautions........................................................................... 2-2 4.5 Changing Settings Using a SmartCard ............................. 4-8
2.3 Notes ............................................................................... 2-2 4.6 DC Power Accessories ................................................... 4-9
Chapter 3: Specifications and Theory of Operation ......... 3-1 Chapter 5: Routine Maintenance ...................................... 5-1
3.1 Overview ......................................................................... 3-2 5.1 Overview ........................................................................ 5-2
3.2 Specifications ................................................................... 3-3 5.2 Cleaning the System ........................................................ 5-2
3.3 Symbols and Meanings .................................................... 3-5 5.3 Cleaning / Replacing the Intake Filter .............................. 5-2
3.4 Functional Description ...................................................... 3-6 5.3.1 Removing, Cleaning, and
3.5 Theory of Operation (Subassemblies) ............................... 3-6 Replacing the Filter ............................................... 5-2
3.5.1 Main PCA ............................................................. 3-6 5.4 Preventive Maintenance Schedule ................................... 5-5
3.5.2 Control Pad & Liquid Crystal Display (LCD) ........ 3-7
3.5.3 Power Supply ........................................................ 3-7
3.5.4 Encore Pro SmartCard® ....................................... 3-7
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual T-1
Table of Contents
Chapter 6: Troubleshooting and Diagnostics................... 6-1 8.3 Real-time Clock Calibration ............................................ 8-3
6.1 Overview ........................................................................ 6-2 8.4 Load Serial and Model Numbers .................................... 8-5
6.2 Common System Level Problems ................................... 6-3 8.5 Respironics Heated Humidifier Test ................................. 8-6
6.3 Troubleshooting Table ..................................................... 6-4 8.6 Current Draw Test ........................................................ 8-13
6.4 System Error Codes ..................................................... 6-10 8.7 Manual Heated Humidifier Test ..................................... 8-14
6.4.1 Extracting Error Codes from the Unit .................. 6-10 8.8 Performance verification ................................................ 8-15
8.9 Testing Data Sheets....................................................... 8-12
6.4.2 Extracting Error Codes from the Unit
Using the Utility Software ................................... 6-10
Appendix A: Tools and Equipment .................................... A-1
6.5 Humidifier LED Error Code Chart ................................ 6-22
A.1 Service Tools and Supplies ............................................ A-2
A.2 Acceptable Test Equipment............................................ A-3
Chapter 7: Repair & Replacement................................... 7-1 A.2.1 Digital Manometer .............................................. A-3
7.1 Overview ........................................................................ 7-2 A.2.2 Digital Multimeter ................................................ A-3
7.2 Repair Kits ..................................................................... 7-3 A.2.3 Flow Meter......................................................... A-4
7.3 Repair and Replacement ................................................. 7-8
7.3.1 Top Enclosure Replacement ................................ 7-9 Appendix B: Schematics .................................................... B-1
7.3.2 Key Pad Replacement ...................................... 7-11 B.1 Schematics Statement.............................................. . ..B-2
7.3.3 Main Printed Circuit Assembly Motor Driver ................................................................ B-3
(Main PCA) Replacement ................................ 7-13 Power Supply ............................................................... B-4
7.3.4 Outlet Port Replacement ................................... 7-15 Display ......................................................................... B-5
7.3.5 Flow Element Replacement ............................... 7-16 To CPU Board .............................................................. B-6
7.3.6 Dog Bone Coupler ........................................... 7-17 Analog Circuit.............................................................. B-7
7.3.7 Blower Assembly Replacement ......................... 7-18 Microprocessor ............................................................ B-8
7.4.8 Bottom Enclosure Replacement ........................ 7-20
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual T-2
Chapter 1: Introduction
Chapter 1: Introduction
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 1-1
Chapter 1: Introduction
The REMstar Auto with C-Flex and REMstar Pro 2 are Continuous Positive
Email: service.operations@respironics.com
Airway Pressure (CPAP), microprocessor-controlled devices used for the
treatment of adult Obstructive Sleep Apnea (OSA). Phone: (724) 755-8220
Fax: (724) 755-8230
1.2 Service Notice
This service manual is intended for technicians with prior training or
experience in servicing CPAP devices.
www.respironics.com
U.S.A. and Canada International
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 1-2
Chapter 2: Warnings, Cautions, and Notes
Chapter 2: Warnings,
Cautions, and
Notes
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 2-1
Chapter 2: Warnings, Cautions, and Notes
Chapter 2: Warnings, Cautions, • To assure the safety of the service technician and the specified
performance of the device, Respironics recommends that only
and Notes technicians having prior training or experience servicing CPAP
devices perform any repairs or adjustments to this device.
WARNING: Indicates the possibility of injury to the patient,
operator, or technician.
2.2 Cautions
CAUTION: Indicates the possibility of damage to the device.
• Federal Law (U.S.) to sale by or on the order of a physician.
NOTE: Places emphasis on an operating/servicing characteristic. • Care should be taken to avoid exposure of the devices to tempera-
tures at or near the extremes of those specified in Chapter 3. If
2.1 Warnings exposure to such temperatures has occurred, the device should be
allowed to return to room temperature before being turned on.
• Do not immerse this device in any fluids.
• The information in this manual is provided for service personnel
• If oxygen is used with this device, the oxygen flow must be reference and is not intended for system setup or use. System
turned off when the device is turned off. Accumulated oxygen in setup should be performed by appropriate personnel using the
the system enclosure increases the risk of fire. “Home Care Provider Setup Instructions”.
• Oxygen supports combustion. Do not use oxygen in the pres- • Tobacco smoke may cause tar buildup within the device that may
ence of open flames, cigarette smoke, electrical spark, or other result in damage.
sources of ignition.
• Discontinue use of this device if any parts or components are
• This device is not suitable for use in the presence of a flammable damaged. Repair or replace device or components as necessary
anesthetic mixture in combination with oxygen or air or in the prior to continuing use.
presence of nitrous oxide.
• Check power cord for damage or signs of wear.
• Disconnect the electrical supply before cleaning or performing
any repairs on this device.
• Electronic components used in this device are subject to damage 2.3 Notes
from static electricity. Repairs made to this device must be • Notes are provided throughout this manual as necessary.
performed only in an antistatic, Electrostatic Discharge (ESD)
protected environment.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 2-2
Chapter 3: Specifications and Theory of Operation
Chapter 3: Specifications
Features,
Descriptions,
and Theory of
Operation
3.1 Overview ..................................................... 3-2
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 3-1
Chapter 3: Specifications and Theory of Operation
3.1 Overview
This chapter describes the REMstar Auto with C-Flex and REMstar Pro 2
specifications and system features. A functional description and theory of
operation is also provided.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 3-2
Chapter 3: Specifications and Theory of Operation
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 3-3
Chapter 3: Specifications and Theory of Operation
Specifications (continued)
Static Pressure Accuracy +/- 1.0 cm H2O
STANDARDS COMPLIANCE: with no humidifier.
ALTITUDE COMPENSATION:
OPERATING CHARACTERISTICS:
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 3-4
Chapter 3: Specifications and Theory of Operation
3.3 Symbols and Meanings
Display Screen Label Meaning
Meaning
Symbol Symbol
Indicates that the device requires user attention as Attention, consult accompanying documents
indicated on the screen
DC Power
Indicates that a SmartCard is inserted and detected
Indicates that the alphanumeric digits are AC Power
cm H 2O
displaying a pressure value
Class II (Double Insulated)
Indicates that the user may clear the Therapy Session
Counter
European CE Declaration of Conformity
Indicates that a C-Flex comfort setting is being displayed
or is active
Notified Body Approval for Standards Compliance
Indicates that the humidifier is turned on and/or its setting
is being displayed Canadian/US Certification
Indicates that the Therapy Hour Meter is being displayed.
This symbol is used in conjunction with the symbol Type BF Applied Part
to indicate that the session counter is being displayed
IPX0 Drip Proof Equipment
Indicates that the control panel LED backlight setting is
being displayed or is active
Indicates that a Patient Disconnect alert is active or its
setting is being displayed
Indicates that the ramp function is in progress, or during
provider mode, the ramp length is being displayed
Indicates that the session counter is being displayed;
This symbol is used in conjunction with the symbol.
Indicates that the ramp starting pressure is being displayed
Indicates that the device is in provider mode and not
user mode
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 3-5
Chapter 3: Specifications and Theory of Operation
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 3-6
Chapter 3: Specifications and Theory of Operation
3.5.2 Control Pad & Liquid Crystal Display 3.5.4 Encore SmartCard
(LCD) The REMstar Auto with C-Flex and REMstar Pro 2 are delivered with the
The control pad consists of five user buttons (Ramp Button, On / Off Encore SmartCard installed. Information is stored in a small silicon chip
button, Heated Humidifier Button, and Right and Left User Buttons for embedded into the card. When installed into the into the devices, the
other functions). The On / Off button activates and deactivates the SmartCard records the date, time at pressure, pressure levels, and other
pressure. The range of pressure is 4.0 to 20.0 cm H2O with AC and DC information on the sleep session. Using the Encore Pro Software, the
operation (adjustable in increments of 0.5 cm H2O). usage data can be automatically downloaded and viewed.
NOTE: Domestic and International versions of both the REMstar Auto The SmartCard can also be programmed with the patient’s prescription by
with C-Flex and REMstar Pro 2 are available. The LCD display using the Encore Pro Software and SmartCard reader/writer. When the
screens and control buttons are represented by English text on programmed SmartCard is inserted into the device, the prescription settings
the domestic variation and international symbols are used on are automatically transferred into the device and erased from the SmartCard.
the second variation of the unit. The SmartCard will then start to collect patient data.
FOSQ: The REMstar Auto with C-Flex and REMstar Pro 2 are programmed
to allow the users to take a “Functional Outcomes of Sleep Questionnaire”
3.5.3 Power Supply (FOSQ). FOSQ is a quality of life questionnaire designed specifically for
The unit has a switching power supply which is built on the Main PCA and people with sleep disorders. The results allow health care professionals to
mounted within the unit’s top enclosure. The switching power supply see how therapy has improved the quality of their patient’s life. Refer to
operates properly on any AC voltage between 100 - 240 VAC; 47 - 63 HZ the “FOSQ” brochure (from Respironics) for further information. The FOSQ
and produces an output of approximately 12 VDC. The switching power information is also stored on the SmartCard.
supply automatically adjusts for varying input voltages. No voltage
selector switch or fuse changing is required, which allows for worldwide
operation.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 3-7
Chapter 3: Specifications and Theory of Operation
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 3-8
Chapter 4: System Setup Procedures
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 4-1
Chapter 4: System Setup Procedures
NOTE: This section provides directions for accessing the therapy Step 3 Remove the SmartCard from the Reader/Writer. If desired,
menu. Prescribed therapy can be set by using the SmartCard or write the patient’s name on the front of the card.
the Therapy menu. (The patient’s User’s Manual provides
directions to access only the Patient menu, which limits access
to settings. To prevent patients from tampering with the
settings, the directions to access the Therapy menu should not
be revealed to the patient).
NOTE: If the SmartCard is not installed, the device usage will not be
recorded.
NOTE: If the system is not turned off properly, data will not be recorded
to the SmartCard.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 4-2
Chapter 4: System Setup Procedures
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 4-3
Chapter 4: System Setup Procedures
4.3.1 Therapy Mode Menu Navigation
When in Therapy Mode, press the left and right user Note: The menu screens shown below are displayed
buttons to scroll through the display screens. Use the on Domestic units. Please note that Interna-
Humidifier and Ramp buttons to change the settings on tional units’ push-buttons and menu screens
each screen. To exit Therapy Mode, push the Start/Stop are represented with graphics, symbols, and
button. icons. Refer to Chapter 3 for additional
To/from the LED Backlight
Setting screen information.
Right User Left User
Button Button
RAMP
SETUP START
Ramp Start Pressure Setting screen
Right User Left User
Button Button
NOTE: This screen appears only
Minimum Pressure Setting Screen if the Ramp Length is set to a
NOTE: This screen appears only MIN
SETUP
value between five and 45 minutes.
cm Right User
on the REMstar Auto with C-Flex H2O Button
Left User
Button
when either APAP or AFLE
Mode is Selected. SETUP
Right User Left User Patient Disconnect
Button Button
Alert Setting screen
Maximum Pressure Setting Screen ALERT
MAX
NOTE: This screen appears only SETUP
Right User Left User
cm
on the REMstar Auto with C-Flex H2O Button Button
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 4-4
Chapter 4: System Setup Procedures
4.3.2 Therapy Mode Display Screens
AFLE Indicates that the Unit is
Reset the Number of Nights the Unit Was Used for More
Than Four Hours in Auto-CPAP Mode with C-
Flex Comfort Setting
Push the Humidifier or Ramp button to erase the number of nights the unit
was used for more than four hours. Press the right user button to continue
to the Mode Setting screen.
Once you have selected the operating mode, press the right user button to
Select the Operating Mode continue to the next screen.
Select from one of two (REMstar Pro 2) or four (REMstar Auto with C-Flex) Set the Minimum Pressure (REMstar Auto with C-Flex
operating modes as follows: only)
• CPAP-Maintains a constant level of pressure throughout the
If you have selected APAP or AFLE mode, you must set minimum and
breathing cycle.
maximum pressure settings. To change the minimum pressure setting, press
• CFLE-CPAP with pressure relief upon exhalation the ramp or humidifier button. The range for the minimum pressure setting
• APAP-Automatically adjusted CPAP therapy (REMstar Auto with C- is 4 to 20 cm H2O, adjustable in increments of 0.5 cm H2O.
Flex only)
• AFLE-Automatically adjusted CPAP therapy with pressure relief NOTE: The minimum pressure cannot be set higher than the maximum
upon exhalation (REMstar Auto with C-Flex only) pressure. Example: If the current settings are 4 cm min. and 6 cm
max., and you wish to change them to 8 cm min. and 12 cm max.,
you must set the maximum pressure first. Press the right user
APAP Indicates that the Unit is button to go to the maximum pressure setting, and then press
in Auto-CPAP Mode without the left user button to go back to the minimum pressure setting.
C-Flex comfort setting
Once you have set the minimum pressure setting, press the right user
button to continue to the next screen.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 4-5
Chapter 4: System Setup Procedures
Set the Maximum Pressure Set the Ramp Length
If you have selected APAP or AFLE mode, you must set minimum and If the operating mode is set to CPAP or CFLE, the range for the Ramp
maximum pressure settings. To change the maximum pressure setting, press Length is 5 to 45 minutes. If the operating mode is set to APAP or AFLE
the ramp or humidifier button. The range for the maximum pressure setting (REMstar Auto with C-FLEX only), the range for Ramp Length is 0, 2, 3, or 4
is 4 (or the value of the minimum pressure setting) to 20 cm H2O, adjustable hours.
in increments of 0.5 cm H2O.
Once you have set the patient disconnect alert to on or off, press the right
NOTE: The maximum pressure cannot be set lower than the minimum user button to go to the next setting.
pressure setting. Example: If the current settings are 8 cm min.
and 12 cm max., and you wish to change them to 6 cm min. and 9 Set the Patient Disconnect Alert to On or Off
cm max., you must set the minimum pressure first. Press the left
user button to return to the Minimum Pressure setting screen,
This patient disconnect alert is an audible tone that sounds when the unit
and then press the right user button to go back to the Maximum
detects a large continuous air leak (such as mask removal) in the patient
Pressure setting screen.
circuit. This alert can be turned on (1) or off (0) using the ramp or humidi-
fier button.
Once you have set the maximum pressure, press the right user button to
continue to the next setup display screen. If you selected APAP as the
Once you have set the patient disconnect alert to on or off, press the right
operating mode, the patient disconnect alert setup screen will be displayed
user button to go to the next setting.
next. If you selected AFLE as the operating mode, the C-Flex level setting
screen will be displayed next.
Set the Button Lights On or Off
Set the C-Flex Level
The button lights setting allows you to have the lights behind the buttons
If you have selected the AFLE (REMstar Auto with C-Flex) or CFLE turned on or off while the airflow is turned on. (The lights will always be on
operating mode, you must select a C-Flex setting. Use the ramp or humidi- when the airflow is off).
fier button to set the C-Flex to either 1, 2, or 3. Once you have selected the
C-Flex setting, press the right user button to continue to the next setup
display screen.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 4-6
Chapter 4: System Setup Procedures
Temperature Setting
Preheat Mode
To enter preheat mode, press and hold the humidifier button on the
REMstar Auto with C-Flex or REMstar Pro 2 until the humidifier icon
appears on the display screen. The green LED on the front of the Heated
Humidifier will illuminate. Preheat mode will last 20 minutes if the airflow is
not turned on.
Heating Mode
To enter heating mode, press the On/Off button on the unit to start the
blower. Once the blower has started, press and hold the humidifier button
until the humidifier icon appears on the display screen. The green LED will
illuminate on the front of the Heated Humidifier.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 4-7
Chapter 4: System Setup Procedures
Turn the airflow on to verify the new prescription setting. You may now
Step 1 Inserting the SmartCard
either remove the SmartCard or leave it in the unit to record device usage.
Once the prescription settings have been transferred to the unit, they will
Make sure that the unit is plugged into an electrical outlet. Insert the
be deleted from the SmartCard.
programmed SmartCard into the slot on the right side of the unit (logo side
facing up). When the “card detected” screen appears on the unit’s LCD,
the card is inserted correctly.
Indicates a SmartCard
MIN is Inserted
cm
H2O
CARD
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 4-8
Chapter 4: System Setup Procedures
4.6 DC Power Accessories CAUTION: Only use the Respironics DC Power Accessories.
Use of any other DC Power accessories may result in
The Respironics Shielded DC Power Cord (RI P/N 1001956) can be used damage.
to operate the REMstar Auto with C-Flex or REMstar Pro 2 in a stationary CAUTION: DC Power is not intended to be used as battery backup.
recreational vehicle, boat, or motor home. DO NOT connect the DC Power while the REMstar Auto
with C-Flex or REMstar Pro 2 is operating on AC Power.
The Respironics DC Battery Adapter Cable (RI P/N 532209), when used System damage may occur.
with the Respironics Shielded DC Power Cord, enables the Respironics
CPAP system to be operated from a 12 VDC freestanding battery.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 4-9
Chapter 4: System Setup Procedures
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 4-10
Chapter 5: Routine Maintenance
Chapter 5: Routine
Maintenance
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 5-1
Chapter 5: Routine Maintenance
Wipe the outside of the unit with a cloth slightly dampened with water and
5.3.1 Removing, Cleaning, and Replacing the
a mild detergent. Let the unit dry before reconnecting the electrical supply. Filter
Step 1 With the airflow turned off, disconnect the AC or DC power
cord from the back of the device.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 5-2
Chapter 5: Routine Maintenance
Step 2 Remove the filter cap by gently lifting up on the bottom of the Step 3 Remove the pollen filter by gently pulling the edges of the filter.
filter cap, then pulling it away from the back of the device. Rinse the filter in a steady stream of running water. Squeeze out
the water, and repeat. Allow the pollen filter to air dry for eight to
12 hours, or in a clothes dryer for 15 to 20 minutes.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 5-3
Chapter 5: Routine Maintenance
Step 5 Install the filter cap. The opening in the unit’s filter cap should
face down and away from the patient.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 5-4
Chapter 5: Routine Maintenance
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 5-5
Chapter 5: Routine Maintenance
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 5-6
Chapter 6: Troubleshooting and Diagnostics
Chapter 6: Trouble-
Shooting and
Diagnostics
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 6-1
Chapter 6: Troubleshooting and Diagnostics
6.1 Overview
WARNING: Electrical Shock Hazard: Disconnect the electrical
supply before repairing the unit.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 6-2
Chapter 6: Troubleshooting and Diagnostics
__ _ A large circuit leak, such as mask • No airflow in approximately 90 • Check the patient circuit for
removal, has been detected. seconds. any leaks and adjust the mask
and headgear.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 6-3
Chapter 6: Troubleshooting and Diagnostics
6.3 Troubleshooting Table
Symptom Cause Verification Corrective Action
Audible Alarm not Sounding - Main PCA Perform power up sequence. If audible alarm does not work,
During audible alarm condition, Visually verify the illumination replace Main PCA.
the alarm is not audible. of the LED’s and display. Reestab-
lish original alert conditions. Listen If visual indicators are not
for audible alarm. present at power up, perform
user interface checkout proce-
dure.
Device does not operate from the Faulty DC plug, low DC voltage, Ensure the DC voltage supply is a If DC source is not at 11.5
12 VDC power source- fuse in DC cord is blown minimum of 11.5 VDC. Ensure the VDC or higher, replace or
DC voltage is stable. Monitor the recharge power source as
Main PCA voltage at the back of the DC required.
connector mounted on the Main
PCA. If DC voltage is below 11.5 If 12 VDC supply and DC cords
VDC, replace or recharge DC are ok, replace Main PCA.
source. If there is no DC measured
at the Main PCA, then check the
DC adapter cord for continuity.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 6-4
Chapter 6: Troubleshooting and Diagnostics
Display Indicator - Main printed circuit assembly Perform user interface checkout If device fails user interface
LED or other indicator not working. procedure. checkout procedure, replace
Main PCA.
Display - Display assembly Inspect solder joints between Replace Main PCA.
Display is blank or contains display and Main PCA.
erroneous information. Main PCA
Intermittent Power Supply Power cord, AC inlet, Main PCA, Inspect power cord for fraying at Replace Main PCA.
Problem - loose connections cable ends.
An intermittent on / off condition
exists, unit alarms randomly, or Inspect all connections on the
indicator lights blink sporadically. Main PCA.
Noise - Missing or damaged rubber feet Inspect the bottom of the unit If rubber feet are missing,
on bottom of unit, blower for missing or damaged rubber feet. replace rubber feet. Tighten any
malfunctioning loose screws.
Blower impeller imbalance Place hand on device, if vibration is Resolve any air leaks.
detected, replace blower S/A.
Replace Blower.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 6-5
Chapter 6: Troubleshooting and Diagnostics
Odor - Tubing smells new, device smells Visually inspect patient tubing for Run device in a clean environ-
new, airborne residue buildup contamination. ment for a few hours to eliminate
new smell. Wash tubing with
soap and water. To clear residue
buildup, replace all subassem-
blies in the patient air stream
(blower filters, patient circuit).
Outlet Air Temperature - Dirty filters, defective Blower, Main Ensure the filter is clean and not Replace in order until solved:
The outlet air temperature is too PCA restricting air flow. Monitor the
warm. outlet temperature at the end of the Filters
six foot patient tubing. A rise in
temperature can be expected. Blower
Main PCA
Pressure Related Problems - Inspect pressure tubing for secure Secure pressure tubing connections.
The outlet pressure does not Pressure tubing has been blocked, connections and kinks. Check
change or properly adjust. disconnected, or kinked. Blower for leaks. Replace blower, pressure tubing, or
Main PCA.
Main PCA or blower malfunction
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 6-6
Chapter 6: Troubleshooting and Diagnostics
Pressure Offset - The delivered pressure is higher or Perform power up process. Check internal tubing for kinks.
Static pressure fluctuates from lower than the set value by more
-1.0 cm to +1.0 cm H2O of set than -1.0 cm to +1.0 cm H2O. Replace the Main PCA.
pressure.
No Stored Therapy Data SmartCard was not installed in No therapy data after usage Depress the blower on / off
device prior to use button before disconnecting the
Main AC power.
Not a Respironics SmartCard
(error code displayed) Reformat SmartCard.
Device on, displays pres- Loose connection Check motor plug / PCA Ensure connections are secure.
sure setpoint, No airflow connection.
Blower / motor failure Replace the Blower Assembly.
Troubleshoot by substitution
Main PCA failure with a known good Blower S/A. Replace Main PCA.
Occluded or restricted pressure tube Check pressure with calibrated Replace pressure tubing.
Pressure check out of range
manometer.
Defective pressure sensor Calibrate Main PCA.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 6-7
Chapter 6: Troubleshooting and Diagnostics
Pressure Variation - The Internal air leak Replace filter. Replace Blower.
pressure varies around the set
value, pressure fluctuates greater Filters dirty Perform Testing Process (Refer to If pressure still drifts, then
than 1.0 cm H2O. Chapter 8). replace Main PCA.
Main PCA
Check internal tubing.
Blower
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 6-8
Chapter 6: Troubleshooting and Diagnostics
Humidifier will not turn on Humidifier not turned back on Inspect main power cord for fraying Replace power cord.
after CPAP was turned off at cable ends.
Faulty power cord Inspect power cord between CPAP Replace Main PCA in
and humidifier base. humidifier base.
Broken or damaged light pipe
Inspect light pipe inside CPAP for Realign or replace light tube.
No infrared connection misalignment or kinking.
Damaged heater plate Make sure receiver on Humidifier Replace Humidifier Main PCA.
base is not damaged.
Humidifier Main PCA
Inspect Main PCA to see if there is Replace Main PCA.
CPAP Main PCA damage to the transmitter on the
Main PCA.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 6-9
Chapter 6: Troubleshooting and Diagnostics
6.4 System Error codes 6.4.2 Extracting Error Codes Using the Utility Software
This section contains information on extracting error codes that have Step 1 Verify utility software is installed on the computer.
occurred during use of the REMstar Auto with C-Flex or REMstar Pro 2.
There are two ways to extract the information. Step 2 Connect the communication cable between the SleepLink
card and the serial port COM1 on your computer.
6.4.1 Extracting Error Codes from the Unit
Step 3 Insert the SleepLink card into the device and plug the device into
Step 1 Hold the humidifier button and ramp button down while the proper AC voltage source.
plugging in the power cord. Continue holding the buttons
down until the unit beeps. Step 4 With your mouse, click on the Start Button of your
computer.
Step 2 The most recent error code will be displayed. By pressing
the humidifier key, the unit will show the previous error code, Step 5 Select:
allowing the user to view the entire error log in a circular buffer Programs
fashion. Only the ten most recent errors will be shown. An E-0 Respironics
indicates no error. Tools
Read Error Log
Step 3 By pressing the ramp key, the unit will display the blower hours
when the error occurred. Pressing the humidifier key will allow the Step 6 The software will load automatically.
user to view the hours that each error occurred in a circular buffer
fashion.
Step 7 With your mouse, click on the Run Arrow located
Step 4 To erase the error logs, press the right user button. An X will in the upper left corner of the computer screen.
appear in the lower left corner of the LCD. Then press the
Pressure On / Off key. This will erase all errors and the blower Step 8 The software will automatically read and display any error
hours at which the errors occurred, and return the unit to the codes stored in the unit along with the date and time the errors
system self test screen. occurred.
NOTE: The ten most recent error codes will be displayed. The oldest
error code will be overwritten when a new error occurs.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 6-10
Chapter 6: Troubleshooting and Diagnostics
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 6-11
Chapter 6: Troubleshooting and Diagnostics
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 6-12
Chapter 6: Troubleshooting and Diagnostics
E-30 ERR_INVALID_NVDATA_SCHEMA •Normal for first time power up •Remove & restore power
•Defective EEPROM •Replace Main PCA
E-32 ERR_INVALID_NVDATA_CHECKSUM •Normal for first time power up •Remove & restore power
•Defective EEPROM •Replace Main PCA
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 6-13
Chapter 6: Troubleshooting and Diagnostics
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 6-14
Chapter 6: Troubleshooting and Diagnostics
E-57 ERR_KEYPAD_STUCK_KEY •Verify all keys protrude through top •Align keys with top
•Defective Keypad •Replace Keypad
•Defective pad on PCA •Replace Main PCA
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 6-15
Chapter 6: Troubleshooting and Diagnostics
1018658 REMstar Auto with C-Flex and REM star Pro 2 Service Manual 6-16
Chapter 6: Troubleshooting and Diagnostics
Error Code Description Probable Cause Corrective Action
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 6-17
Chapter 6: Troubleshooting and Diagnostics
6.5 SmartCard Error Codes
C-1 SMARTCARD_12C_WRITE_FAIL Card inserted upside-down. Remove card and reinsert properly.
Otherwise, card is corrupt and should be reprogrammed.
C-2 SMARTCARD_BAD_CARD_TYPE Card is not supported by this device. Otherwise, card is corrupt and
should be reprogrammed.
C-4 SMARTCARD_SCRIPT_DURING_PROVIDER_MODE Remove card. Exit Provider Mode screens. Then reinsert card.
C-5 SMARTCARD_PRESENT_IN_CAL_MODE Remove card. Exit Calibration Mode. Then reinsert card.
C-8 SMARTCARD_12C_READ_FAIL Card inserted upside down. Remove card and reinsert properly.
Otherwise, card is corrupt and should be reprogrammed.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 6-18
Chapter 6: Troubleshooting and Diagnostics
C-103 SMARTCARD_UNKNOWN_CPB_FORMAT Card contains a prescription format that is not supported. Perhaps
card is intended for another device.
C-104 SMARTCARD_UNKNOWN_CPB_PRESCRIPTION Card contains a prescription type that is not supported. Perhaps
card is intended for another device.
C-105 SMARTCARD_SCRIPT_PARM_RANGE Card contains prescription settings that are outside of this device’s
range. Perhaps card is intended for another device.
C-201 SMARTCARD_BAD_CDB_VERSION Card contains a storage format that is not supported. Perhaps card
is intended for another device.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 6-19
Chapter 6: Troubleshooting and Diagnostics
Code Label User Action
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 6-20
Chapter 6: Troubleshooting and Diagnostics
Code Label User Action
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 6-21
Chapter 6: Troubleshooting and Diagnostics
6.6 Humidifier LED Error Code Chart
LED Blinks Problem / Cause Corrective Action
Conditions: When the Heated Humidifier is enabled, the LED will be on constantly.
When the Heated Humidifier is disabled, the LED will be off.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 7-1
Chapter 7: Repair and Replacement
7.1 Overview For technical assistance or replacement part ordering information contact
Respironics Corporate Technical Service.
This chapter illustrates the replaceable components in the REMstar Auto
with C-Flex and REMstar Pro 2. Procedures for replacing the components
are also provided in this chapter. U.S. and Canada
Phone: 1-800-345-6443
Fax: 1-888-886-0245
WARNING: Disconnect the electrical supply before attempting to
make any repairs to the device.
International
CAUTION: Electronic components used in this device are
subject to damage from static electricity. Repairs Phone: 724-387-4000
made to this device must be performed only in an
Fax: 724-387-5012
antistatic, ESD-protected environment.
Email: service@respironics.com
www.respironics.com
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 7-2
Chapter 7: Repair and Replacement
Bottom Enclosure RP# 1011929* (Refer to note* page 7-7) Blower Assembly RP# 1022381
Outlet Port RP# 1008477 Main PCA RP# ** (Refer to note** page 7-7)
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 7-3
Chapter 7: Repair and Replacement
AC Power Cord RP# 1005894*** (Refer to note*** page 7-7) Sleep Link Communication Cable RP# 1007492
Mini DIN Plug RP# 1011220 Flow Element Kit RP# 1008490
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 7-4
Chapter 7: Repair and Replacement
Blower Top Foam RP# 1011219 Top Enclosure RP# 1022379 (REMstar Auto w/ C-Flex)
1022378 (REMstar Pro 2)
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 7-7
Chapter 7: Repair and Replacement
7.3 Repair and Replacement
Bottom
Top Filter Enclosure
Cover Filters
Enclosure
pg. 5-2 pg. 5-2 pg. 7-20
pg. 7-9
6
Blower Assembly
pg. 7-18
6
Main Printed 6
Circuit Assembly (PCA) Blower Top Foam
6 6
Dog Bone Outlet Port /
Coupler Flow Element
Key
pg. 7-17 pg. 7-15 Pad
6 pg. 7-11
Blower Top Foam 6
Main Printed
pg. 7-18
Circuit Assembly (PCA)
6
pg. 7-13
Blower Assembly
6
pg. 7-18 Top
Filter
Filters Enclosure
6 Cover
Bottom pg. 5-2 pg. 5-2 pg. 7-9
Enclosure
pg. 7-20
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 7-8
Chapter 7: Repair and Replacement
Procedure
7.3.1 Top Enclosure Replacement
Step 1 Remove the Top Enclosure:
Replacement Part Number 1022379 (REMstar Auto with C-Flex) a. Place the device on a protected work surface and
1022378 (REMstar Pro 2) carefully turn it over, exposing the bottom.
3
c. While securely holding the top and the bottom
enclosure together, carefully return the device to its
upright position.
Top Enclosure (REMstar d. Lift the top enclosure up and away from the bottom
Auto with C-Flex Shown) enclosure.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 7-9
Chapter 7: Repair and Replacement
Top Enclosure Replacement (Continued)
c. Insert and tighten the three Phillips screws to secure the Top
Enclosure to the Bottom Enclosure.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 7-10
Chapter 7: Repair and Replacement
Procedure
7.3.2 Keypad Replacement
Replacement Part Number 1020294 (International) Step 1 Remove the Top Enclosure (Refer to section 7.3.1)
1020325 (Domestic)
Step 2 Remove the Keypad:
Included in Kit: Tools Required:
Keypad Phillips screwdriver (No. 2 medium) a. Carefully lift the Keypad from the Main PCA.
Flat-blade screwdriver (small) There are three rubber locking pins that hold the keypad to
the Main PCA.
Keypad
Locking Pins
(International
Keypad Shown)
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 7-11
Chapter 7: Repair and Replacement
Keypad Replacement (Continued)
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 7-12
Chapter 7: Repair and Replacement
Procedure
7.3.3 Main Printed Circuit Assembly
Step 1 Remove the Top Enclosure (Refer to section 7.3.1)
(Main PCA) Replacement
Step 2 Remove the Keypad (Refer to section 7.3.2)
Replacement Part Numbers:
Step 3 Remove the Main PCA:
• 1021760 REMstar Auto with C-Flex (International)
• 1021761 REMstar Auto with C-Flex (Domestic) a. Lift the Main PCA and hold it slightly above the
Blower Assembly to provide access to the Blower
• 1021764 REMstar Pro 2 (International) Wiring Harnesses and tubing connected to the Main
• 1021765 REMstar Pro 2 (Domestic) PCA.
NOTE: The correct part number of the main PCA for your unit is linked
to the model number.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 7-13
Chapter 7: Repair and Replacement
Main PCA Replacement (Continued)
PCA Stands
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 7-14
Chapter 7: Repair and Replacement
Procedure
7.3.4 Outlet Port Replacement
Step 1 Remove the Top Enclosure (Refer to section 7.3.1)
Step 2 Remove the Main PCA (Refer to section 7.3.3)
Replacement Part Number 1008477
Step 3 Remove the Outlet Port:
Included in Kit: Tools Required:
Outlet port Phillips screwdriver (No. 2 medium) a. Remove the pressure tubing from the Outlet Port.
Outlet Port
Step 4 Install the Outlet Port:
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 7-15
Chapter 7: Repair and Replacement
7.3.5 Flow Element Replacement Procedure
Replacement Part Number 1008490 Step 1 Remove the Top Enclosure (Refer to section 7.3.1)
Step 2 Remove the Main PCA (Refer to section 7.3.3)
Included in Kit: Tools Required:
Flow Element Phillips screwdriver (No. 2 medium) Step 3 Remove the Outlet Port (Refer to section 7.3.4)
Tubing
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 7-16
Chapter 7: Repair and Replacement
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 7-17
Chapter 7: Repair and Replacement
Procedure
7.3.7 Blower Assembly Replacement
Step 1 Remove the Top Enclosure (Refer to section 7.3.1)
Replacement Part Number 1022381
Step 2 Remove the Main PCA (Refer to section 7.3.3)
Included in Kit: Tools Required:
Blower Assembly Phillips screwdriver (No. 2 medium) Step 3 Remove the Outlet Port (Refer to section 7.3.4)
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 7-18
Chapter 7: Repair and Replacement
Blower Assembly Replacement (Continued)
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 7-19
Chapter 7: Repair and Replacement
Included in Kit: Tools Required: Step 2 Remove the Main PCA (Refer to section 7.3.3)
Bottom Enclosure (w / label) Phillips screwdriver (No. 2 medium)
8 Hi - Lo x 1/2″ (x3) Step 3 Remove the Outlet Port (Refer to section 7.3.4)
Bottom Enclosure
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 7-20
Chapter 7: Repair and Replacement
Bottom Enclosure Replacement (Continued)
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 7-21
Chapter 7: Repair and Replacement
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 7-22
Chapter 8: Testing, Calibration, and Verification
Chapter 8:
Testing, Calibration, and
Verification
8.1 Overview ..................................................... 8-2
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 8-1
Chapter 8: Testing, Calibration, and Verification
Chapter 8: Testing 8.2 Run-in
8.1 Overview Before the REMstar Auto with C-Flex or REMstar Pro 2 can be calibrated,
the device must be run-in for at least 2 hours to avoid possible calibration
This chapter provides run-in, calibration, AC current draw testing, and failures.
performance verification procedures for the REMstar Auto with C-Flex and
REMstar Pro 2. Calibration and AC current draw testing are necessary only Equipment
when a repair has been made to the unit. However, these procedures may
• 0.25” Test Orifice (p/n 332353)
be used in conjunction with the performance verification to determine
whether or not the unit is functioning properly. Verification shall be
performed at periodic intervals commensurate with hospital or Homecare Procedure
Provider guidelines for preventative maintenance, between rentals, or
during normal patient usage. Step 1 Connect the test orifice to the Outlet Port of the unit.
Calibration and testing of the REMstar Auto with C-Flex and REMstar Pro 2 Step 2 Enter the therapy setup menu (refer to Chap. 4) and set the
requires the use of Respironics Utility Software. Before proceeding, log on pressure to 20 cm H2O.
to (http://servicesoftware.respironics.com) and download the Utility
Software on to your PC. Step 3 Run the unit for at least 2 hours before performing
the calibration.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 8-2
Chapter 8: Testing, Calibration, and Verification
8.3 Real-time Clock Calibration
The Unit Under Test (UUT) will record and track several different param-
eters (e.g. patient pressures, apnea events, when and how the device is
used, error codes, etc.) To do this in a logical manner and for a technician or
therapist to know exactly when a certain event has occurred, the REMstar
Auto with C-Flex and REMstar Pro 2 runs and records events at real-time. Click here
This time is checked and compared to your computer’s time when running
the calibration software. If the times do not match, the calibration will fail.
Therefore, you must perform the real-time clock calibration before you
proceed with the calibration.
Equipment
Procedure
Step 1 Ensure AC power is connected to the UUT.
Step 4 Click on the red “Set Clock / Verify Clock” button to begin
calibration (the button will turn green.)
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 8-3
Chapter 8: Testing, Calibration, and Verification
Real-time Clock Calibration (continued)
Step 5 In the upper lefthand corner, click on the white runtime arrow. Step 6 When the program has finished, a green “PASS” indicator
The program will now set the appropriate date and time on will appear.
the UUT.
“PASS”
Indicator
Runtime
Arrow
Figure 8-3
Calibration Complete
Figure 8-2
Setting Time and Date
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 8-4
Chapter 8: Testing, Calibration, and Verification
8.4 Main Printed Circuit Assembly (Main Step 1 Connect the UUT to a PC:
PCA) / Load Serial and Model Number a. Connect the communication cable between the SleepLink card
and the serial port COM1 PC.
NOTE: During the manufacturing process, the Serial Number is
programmed into the memory of the Main PCA. This process b. Connect the UUT to the proper AC voltage source.
gives Homecare Providers the ability to link the information
from the Encore SmartCard to a specific device for a more
complete patient record. Although this process is important for Step 2 Load the Serial and Model Number
tracking patient information, it is not necessary for the perfor-
mance of the unit. a. Click on the Start Button of your computer.
b. Select:
Programs
Respironics
Parts Needed: Tools Required:
Tools
Sleep Link Card (1007492)
Load Serial and Model Number
Communication Cable None
Utility Software (download from servicesoftware.respironics.com
c. With your mouse, click on the RUN arrow to
start the software program.
Minimum System Requirements
• Windows 95/98/2000/NT d. When prompted, enter the Serial Number of the unit. Press
enter, or click on “OK” with your mouse.
• Color Monitor with 256 colors, 800 x 600 resolution
• Windows-compatible pointing device (mouse) e. Type in the Model Number of the unit. Press
• Windows-supported serial port (COM1 - COM4) enter, or click on “OK” with your mouse.
• 2.5 MB free hard disk space
f. The software will automatically transfer the information into
the memory of the new PCA. A green PASS box will appear.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 8-5
Chapter 8: Testing, Calibration, and Verification
Step 3 Set the UUT on the REMstar Heated Humidifier and connect
the AC power cord to the Humidifier.
Equipment
• Utility Software Step 4 Connect the O2 adapter to the outlet of the UUT and occlude
• Digital Manometer (see Appendix A) the adapter.
• Windows®-compatible personal computer running
• Windows 95 or higher version software NOTE: Remove the humidifier’s reservoir if connected. All connections
Printer must be made to the outlet port of UUT.
• SmartLink card with cable. (P/N 1007492)
• Flowmeter (range: +180 to -180 LPM, 3% accuracy, 1 LPM resolu-
tion)
• Flow control valve
• Pressure pick-off adapter (O2 Enrichment Adapter)
• Pressure tubing
• Negative flow source (CPAP capable of 60 LPM)
• Smooth Bore Patient Tubing
• REMstar Heated Humidifier
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 8-6
Chapter 8: Testing, Calibration, and Verification
System Calibration (continued)
Step 7 Start the calibration by clicking on the white runtime arrow in
the upper lefthand corner of the window.
Step 5 Connect the pressure tubing from the O2 port of the O2
adapter to the digital manometer. (Refer to figure 8-4 for
setup.) Runtime Arrow
SleepLink Card
Manometer
Figure 8-5
System Setup
Step 8 The program will then prompt the user to enter the serial
number. After typing in the serial number, click (OK). Next,
type in the model number and click (OK). Enter the user name
and hit (ENTER) on the keyboard.
O2 Adapter
Figure 8-4
w/ End Cap Step 9 The program will then run through a series of humidifier,
Pressure Setup
lighting, alarm, and display tests. Follow the instructions on
Step 6 Open the Calibration Software. the screen.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 8-7
Chapter 8: Testing, Calibration, and Verification
System Calibration (continued)
Step 10 When the prompt in Figure 8-6 appears, the program is ready
to start the pressure sensor calibration. Click on the (OK)
button.
Figure 8-7
Increase/Decrease Blower Speed
Step 12 Type in the pressure reading from the manometer and press
(ENTER) on the keyboard.
Figure 8-6
System Calibration
Step 11 Observe the manometer. If the pressure is higher than 4.00cm
H2O, click on the red button to decrease the blower speed. If
the pressure on the manometer is lower than 4.00cm H2O,
click on the green button to increase blower speed. Each time
a button is clicked, the pressure will change by approximately
0.2cm H2O. When 4.00cm H2O is achieved, click on
(DONE). Figure 8-8
NOTE: Let the pressures stabilize for at least one minute before Entering Manometer reading
making any adjustments during the calibration.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 8-8
Chapter 8: Testing, Calibration, and Verification
System Calibration (continued)
Step 16 Observe the flow meter. If the flow is greater than 160 LPM,
click on the red button to decrease the flow. If the flow is less
than 150 LPM, click on the green button to increase the flow.
When the UUT is delivering between 150 - 160 LPM, click on
the (DONE) button.
Figure 8-9
Flow Element Calibration
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 8-9
Chapter 8: Testing, Calibration, and Verification
System Calibration (continued)
Step 19 The program will now request several different flow settings
Step 17 Observe the flow meter and adjust the flow control valve until decreasing in value down to and including zero flow. For
150 LPM of flow is achieved. Click on the green (OK) each setting, use the flow control valve to set the flow and
button. then type in each value. After zero flow is confirmed, the
following prompt will appear.
Figure 8-12
Flow Element Calibration Figure 8-14
Step 18 Enter in the flow reading from the flow meter and press Zero Flow Confirmation
(ENTER) on the keyboard.
Step 20 The program will now calibrate negative flow. Connect
another section of patient tubing between the outlet of the
flow control valve and the outlet of the negative flow source
(refer to Figure 8-15). Turn on the negative flow source and
click the green (OK) button.
NOTE: For the negative flow source, another REMstar device may be
used with the pressure set to 20cm H2O.
Figure 8-13
Entering the Flow Meter Reading
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 8-10
Chapter 8: Testing, Calibration, and Verification
Figure 8-16
Negative Flow Value Display
Figure 8-15
System Calibration with Negative Flow Source
Step 21 The program will now calibrate the negative portion of the
flow table. Although the flow meter may not display a
negative flow, pressure against the normal direction of flow of
the device is considered a negative flow. Adjust the flow
control valve until 10 LPM is displayed on the flow meter and
click (OK). Refer to figure 8-16.
Step 22 Before entering the flow value, ensure that the negative or
minus symbol is typed in first. The software will reject a
positive value here. Figure 8-17
Entering the Negative Flow Value
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 8-11
Chapter 8: Testing, Calibration, and Verification
System Calibration (continued) Step 28 When the pressure verification is complete, the program will
request a button test for the UUT. Press the buttons on the
Step 23 Once again, use the flow control valve to set the negative device in the order that they are numbered on the program
flows that the program requests and enter the values. screen.
Step 24 The program has now completed calibrating the flow and Step 29 When the button test has been completed and the UUT has
pressure sensor. Pressure and flow accuracy of the device successfully passed all tests, a green (PASS) prompt will
must now be verified. The program will now read the pressure appear on the computer screen. Click the (OK) button to view
and flow parameters from the device and compare these the test data sheet. Click (OK) on the print results button to
readings to the pressure and flow values that the user will print the test data sheet. Sign the sheet and retain as part of a
enter. device history record. This completes the calibration proce-
dure. The Current Draw Test must now be performed (proceed
Step 25 The program will first verify positive flows. Turn off the to section 8.5).
negative flow source and disconnect the patient tubing from
between the negative flow source and the flow valve. Com-
pletely open the flow valve. Follow the prompts on the screen
for the flow settings.
Step 26 After zero flow is confirmed, the program will verify negative
flow. Connect the patient tubing between the outlet of the
flow valve and the negative flow source. Turn on the negative
flow source and follow the prompts.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 8-12
Chapter 8: Testing, Calibration, and Verification
8.6 Current Draw Test
Step 4 Record the current reading from the multimeter in the
This procedure provides instructions for testing the current draw on the space provided on the Testing Data Sheet. Sign and date the
test sheet and retain for a device history record. This con-
REMstar Auto with C-Flex.
cludes the REMstar Auto with C-Flex Current Draw Test
Equipment
• 0.25” Test Adapter (p/n 332353) (see Appendix A)
• Multimeter
• Test Leads
NOTE: Test leads can be made by opening one lead of the power cord
and installing the proper cable ends to connect to the multi-
meter.
Step 1 Place the test adapter on the outlet of the UUT. Enter the
Setup menu and set the pressure to 20 cm H2O.
Step 3 Turn on the UUT and allow the device to stabilize for at least
one minute.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 8-13
Chapter 8: Testing, Calibration, and Verification
This procedure provides instructions for the Heated Humidifier System to c. Start the preheat mode and observe the following:
be tested without the test station set-up. This test is optional and may be 1. The icon appears on the CPAP display screen.
performed if the automated heated humidifier test is not performed.
2. The green LED on the front of the Heated Humidifier
is illuminated.
CAUTION: Electronic components used in this device are subject
d. Record the results (pass / fail) on the Test Data Sheet.
to damage from static electricity. Use and follow
appropriate Electro-Static Discharge (ESD) procedures. Step 2 Humidifier Heat Test
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 8-14
Chapter 8: Testing, Calibration, and Verification
8.8 Performance Verification Procedure
Step 1 Place the O2 port on the outlet of the device and connect the
The performance verification can be used as a way to determine if the pressure tubing between the test orifice and the manometer.
REMstar Auto with C-Flex and REMstar Pro 2 is functioning properly. This Occlude the O2 port on the device.
verification may be performed between rentals, during normal patient
usage, as routine maintenance, or as a screening procedure. Perform this Step 2 Enter the Setup menu and set the pressure to 4 cm H2O. Turn
procedure after the Run-in (refer to section 8.2) on the UUT and allow the unit to stabilize for at least one
minute.
Equipment
Step 3 Record the pressure reading of the manometer in the space
provided on the Testing Data Sheet.
• O2 Enrichment Attachment (p/n 312710) (see Appendix A)
• Pressure Tubing Step 4 Set the pressure on the UUT to 20 cm H2O. Allow the
• Manometer pressure to stabilize for at least one minute.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 8-15
Chapter 8: Testing, Calibration, and Verification
Performance Verification
Hours____________________
Tolerance Tolerance
CPAP Performance Verification 4cm H2O (_____) 3-5 cm H2O 20cmH2O (_____) 19-21 cm H2O
Hours____________________
Tolerance
Current Draw @ 20cm H2O (_____) <1.0 A
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual 8-16
Appendix A
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual A-1
Appendix A
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual A-2
Appendix A
A.2 Acceptable Test Equipment A.2.2 Digital Multimeter
Specifications:
A.2.1 Digital Manometer*
2.5 digit readout minimum
Specifications: 0.0 - 20.0 VDC
0 - 25 cm H2O (or better) 0.0 - 25.0 VAC
±0.3 cm H2O accuracy
±0.1 cm H2O resolution Acceptable Options:
• Fluke 83 or better model
Acceptable Options: • Any commercially available multimeter
• Respironics Digital Manometer that meets the above specifications.
• TSI Certifier FA Test System, High Flow Module
(RI P/N 1012598)
• Merical DP2001 A.2.3 Flow Meter
Specifications:
• Sensym PDM 200CD
Range: +180 to -180 LPM
• Any commercially available digital manometer
± 2% accuracy
that meets the above specifications.
± 1 LPM resolution
* A water column manometer may also be used.
Acceptable Options:
• TSI Certifier FA Test System, High Flow Module
(RI P/N 1012598)
• TSI 4000 Series Flow Meter
• Any commercially available flowmeter that meets the
above specifications.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual A-3
Appendix A
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual A-4
Appendix B
Appendix B: Schematics
Schematic Statement ............................................. B-2
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual B-1
Appendix B
Appendix B: Schematics
B.1 Schematic Statement
Schematics are supplied with this manual in direct support of the sale and
purchase of this product.
The schematics are proprietary and confidential. Do not copy the schemat-
ics or disclose them to third parties beyond the purpose for which they are
intended. Patents are pending.
The schematics are of the revision level in effect at the time this manual was
last revised. New revisions may or may not be distributed in the future.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual B-2
Appendix B
8 7 6 5 4 3 2 1
R EV C /N D ESC R IPTIO N D ATE D R AW N BY: C H EC KED BY: APPR O VED BY:
499
R6
10.0K
R 16
U 19:B
10.0K
10.0K
10.0K
R 19
R 87
R 88
C R 12 RAM P
R 105 164 114
G ND JTAG SEL W AKEU P H SM G -H 730
10.0K 169 115
N TR ST N TR ST SH D N
165 SPV_W D T
D TM S TM S R 106
D
167 163
TD O TD O PB18/BM S C R 13 HUM ID
166 55 10.0K G ND
TD I TD I PB19/TC LK0 H SM G -H 730
168 58
TC K TC K PB22/TC LK1
77 63
SM AR T_D ETEC T PA9/IR Q 0 PB25/TC LK2 D C _O N
78
M O TO R _C U R R EN T_TR IP PA10/IR Q 1
79 56 CR1 O N /O FF
PA11/IR Q 2 PB20/TIO A0
80 57 H SM G -H 730
PA12/IR Q 3 PB21/TIO B0 FLO W _O S
142
PB3/IR Q 4
10.0K
R 112
+5V
.1uF
C 89
143 61
PB4/IR Q 5 PB23/TIO A1 BU ZZER
144 62 +5V Q 14
PB5/IR Q 6 PB24/TIO B1
81
10.0K
BSS123LT1
R 99
PA13/F1Q
10.0K
R 93
64
PB26/TIO A2 SC L
82 65
PA14/SC K0 PB27/TIO B2 SD A
R 30
G ND 83
PA15/TXD 0 G ND
84 66 G ND +3.3V
PA16/R XD 0 PA0/TC LK3 SD A_D ISPLAY
*
TXD 0
PA3/TC LK4
69
SC L_D ISPLAY *
JP2
10.0K
R 29
68 90.9K BAV99L
TXD 1 PA2/TIO B3
G ND R 155 92
PA20/SC K2
.01uF
10.0K
C 76
R 75
+5V 93 70 1
+5V 10.0K PA21/TXD 2 PA4/TIO A4 BEM F +5V +5V
94 71
R XD 1 PA22/R XD 2 PA5/TIO B4
R 13 95 75 G ND
SC K
R 14 PA23/SPC K PA7/TIO A5 R AW _PW M
96 76
R 154
10.0K
R 103
R 101
10.0K
10.0K
10.0K
10.0K
10.0K
33.2 G ND +5V
R 37
R 18
R 21
R 27
R 34
M O SI
N PC S0
33.2 98
99
PA25/M O SI
PA26/N PC S0/N SS PB0
139
140
C TL_H I_A * * SW 4 RAM P
N PC S1 PA27/N PC S1 PB1 C TL_LO _A
100 141 1 2
PA28/N PC S2 PB2 C TL_H I_B
101 149 SW ITCH CO ND RUBBER 3M M SPSTNO
+3.3V PA29/N PC S3 PB8 C TL_LO _B
.1uF
C 16
150 SW 5
PB9 C TL_H I_C N PC S1
117 151 1 2
VD D A PB10 C TL_LO _C
M ISO
U3
152 SW ITCH CO ND RUBBER 3M M SPSTNO
PB11 1 8
126 153 SW 6 CS VC C
+5V
1000pF
AD VR EF PB12 2 7 G ND
.1uF
C 30
C 29
154 1 2 SO H O LD
PB13 3 6
127 155 SW ITCH CO ND RUBBER 3M M SPSTNO WP SC K SC K
D AVR EF PB14 4 5
156 VSS SI M O SI
PB15
157
PB16 N M 25C 040 +5V
G ND 130 158 SW 7
G NDA PB17 O N /O FF G ND
145
PB6/AD 0TR G N W D O VF
159 * C 80
1 2
R 104
10.0K
R 102
10.0K
G ND 146 160
SW ITCH CO ND RUBBER 3M M SPSTNO
PB7/AD 1TR G M CKO C 87 Y1 SW 8
105 G ND 15pF
.1uF
C 52
XIN C 77 1 2
118 106 R 81 4.9152 M H Z
PR ESS AD 0 XO U T SW ITCH CO ND RUBBER 3M M SPSTNO
119 108 1.00K C 28 15pF
FLO W AD 1 PLLR C
120 1200pF C 23 U 14
C AD 2 R 10 G ND +3.3V
121 SW 9 1 8 G ND
AD 3 N PC S0 C S /SH D N VD D
122 170 499 R IG H T
AD 4 N R ST 120pF 1 2
123 G ND 2 7
AD 5 PR ESS CH0 C LK SC K
I_LO _SID E
BATTER Y_BU
124
125
AD 6
AD 7 N R STBU
112 R 22
CR3
* SW ITCH CO ND RUBBER 3M M SPSTNO
SW 10
FLO W
3
CH1 DO UT
6
M ISO
B XIN 32
110 1.00M 1 2 B
128 111 + C 35 4 5
.1uF
C 38
32.768 KH Z
BAT54
CR5
AT91M 55800-33AI 3 G ND
4 SW 11 LEFT
ZER O
.1uF
C 24
R 23
39.2K
39.2K
G ND G ND
R 31
R 24
1 2
1
Q 21 VBU LK
SW ITCH CO ND RUBBER 3M M SPSTNO +5V
BSS84LT1
C 82
C 81
* * SW 12
R 130
2.00K
1 2
G ND G ND 2 3 LS1
10.0K
10.0K
.1uF
.1uF
R 33
C 40
C 37
R 26
1.00K
R4
47.5
1
R 73
2
SW 1 HUM ID
G ND 1 2 TXD 0
+ B1 BU ZZER
G ND G ND SW ITCH CO ND RUBBER 3M M SPSTNO IR
+3.3V BR 2032-1H E C R 9 LIN K
SW 2 R1
BU ZZER Q4
1 2 ZTN 155W -03 1.00K M M BT3904LT1
+5V +5V SW 3
.1uF
C 93
+3.3V U2 1 2 G ND
R 131 +5V
1 8 SW ITCH CO ND RUBBER 3M M SPSTNO
SENSE1 VD D G ND
ZER O R 132 2 7
R 124
10.0K
SENSE2 MR PP2 G ND
ZER O 3 6 2
PP1 W DI R ESET PP3 G ND
CR4
C 99
C 98
4 5 3
G ND R ESET N R ST
A TPS3305-33D BAV99L A
* * G ND 1 PR O D U C T LIN E:
.1uF
C 94
C -FLEX
G ND G ND
G ND
R 156 PR O C ESS: TITLE:
SPV_W D T N O TES: AU TO C -FLEX,D O M ESTIC
ZER O SCHEM ATIC
G ND 1. R EFER TO SEPAR ATE BILL O F M ATER IALS FO R AD D ITIO N AL PAR T IN FO R M ATIO N . M ICRO & DIG ITAL
2. -IN D IC ATES ITEM S N O T IN STALLED .
* DRN BY: J.KAU C IC 9-25-03 SIZE D R AW IN G N O .
S1017737
R EV.
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual B-3
Appendix B
8 7 6 5 4 3 2 1
R EV C /N D ESC R IPTIO N D ATE D R AW N BY: C H EC KED BY: APPR O VED BY:
SEE SH EET 1
D 0:15
U 19:C
A1:A19
D 35 7 D
D0 D0 AO /N LB A0
+5V
36 8
D1 D1 A1 A1
+5V
37 9
D2 D2 A2 A2
.1uF
C3
38 10
D3 D3 A3 A3
.1uF
.1uF
C 14
C 13
39 11 U 22
D4 D4 A4 A4 U 20
11 23 G ND
A1 A0 Vcc 41 11
40 12 10 OE OE VC C
D5 D5 A5 A5 A2 A1 6 33 G ND G ND
9 15 N C S2 CS VC C
A3 A2 DQ 0 D 15 40
41 13 8 17 17 UB NW R1
D6 D6 A6 A6 A4 A3 DQ 1 D 13 WE WE 39
7 19 LB A0 D 0:15
A5 A4 DQ 2 D 11
42 14 6 21 A1:A19 1 7
D7 D7 A7 A7 A6 A5 DQ 3 D9 A14 A0 I/O 1 D0
5 24 2 8
A7 A6 DQ 4 D7 A13 A1 I/O 2 D2
47 17 4 26 3 9
D8 D8 A8 A8 A8 A7 DQ 5 D5 A12 A2 I/O 3 D4
42 28 4 10
A9 A8 DQ 6 D3 A11 A3 I/O 4 D6
48 18 41 30 5 13
D9 D9 A9 A9 A10 A9 DQ 7 D1 A10 A4 I/O 5 D7
40 16 18 14
A11 A10 DQ 8 D 14 A5 A5 I/O 6 D5
49 19 39 18 19 15
D 10 D 10 A10 A10 A12 A11 DQ 9 D 12 A6 A6 I/O 7 D3
38 20 20 16
A13 A12 D Q 10 D 10 A1:A19 A7 A7 I/O 8 D1
50 20 37 22 21 29
D 11 D 11 A11 A11 A14 A13 D Q 11 D8 A8 A8 I/O 9 D8
36 25 24 30
A15 A14 D Q 12 D6 A9 A9 I/O 10 D 10
51 21 35 27 25 31
D 12 D 12 A12 A12 A16 A15 D Q 13 D4 A2 A10 I/O 11 D 12
34 29 R2 26 32
+3.3V A17 A16 D Q 14 D2 A19 A3 A11 I/O 12 D 14
*
52 22 3 31 27 35
D 13 D 13 A13 A13 A18 A17 D Q 15 D0 A4 A12 I/O 13 D9
C A15
42
A13 I/O 14
36
D 11 C
53 23 33 R3 43 37
D 14 D 14 A14 A14 +5V BYTE 2 A16 A14 I/O 15 D 13
RY/BY 44 38
12 A1 A15 I/O 16 D 15
R 113
10.0K
10.0
54 24 N C SO CE 22
D 15 D 15 A15 A15 14 1 A17 NC
OE OE NC
43
25 WE WE 13
A16 A16 Vss 12 23
100K
R5
44 32 G ND NC A18
R 123
N R ST R ESET Vss 34 28
172 26 G ND NC PP54
OE N R D /N O E A17 A17 AM 29F400AB-90SC
R 122 33.2 173 K6R 1016C 1C -J
WE N W R 0/N W E
R 25 33.2 174 27 G ND
NW R1 N W R 1/N U B A18 A18 G ND
33.2 171 G ND
N W AIT
28
A19 A19
3
N C SO N C S0
4 31
N C S1 A20
5
N C S2 N C S2
6 32
N C S3 A21
135
C 97
C 92
N C S4
136 33
N C S5 A22
137
* * 138
N C S6
N C S7 A23
34
15 175
VDDIO VD D C O R E
176
C 10 VDDIO
+3.3V 16 C 11
.1uF
C 12
G ND 1
B .1uF G ND B
G N D 29 2 .1uF
G ND J4:A
VDDIO +5V 1
G ND
C9
30 131 J4:B
G ND VD D C O R E TD O 2
.1uF 132
G N D 43 VDDIO
J4:C
VD D C O R E C 25 3
.1uF
C 31
44 133
VDDIO G ND
C7 134 G ND
.1uF
.1uF
C8
G ND J4:D
45 TC K 4
.1uF G ND G ND
46
G ND J4:E
147 N R ST 5
VDDIO
G ND 59
VDDIO C 20 J4:F
148 TM S 6
C 85 G ND
60 .1uF
G ND J4:G
.1uF 161 G N D 7
G N D 73 VDDIO
VDDIO C 15 J4:H
162 JTAG SEL 8
C 18 G ND
74 .1uF
G ND G ND J4:I
N TR ST 9
.1uF G N D 87
VD D C O R E J4:J
88 109 TD I 10
VDDIO VD D BU BATTER Y_BU
C 26
.1uF
C 21
89 113
G ND G N D BU
.1uF 90 116
G ND G N D BU +3.3V
102 104 R 15
G ND VDDIO VD D PLL
100
A C 27
103 107
A
1000pF
G ND G N D PLL
.1uF
C 19
C 22
.1uF PR O D U C T LIN E:
AT91M 55800-33AI C -FLEX
G ND
PR O C ESS: TITLE:
AU TO C -FLEX,D O M ESTIC
G ND SCHEM ATIC
M IC R O ,FLASH & D EBU G C O N N EC TO R
8 7 6 5 4 3 2 1
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual B-4
Appendix B
8 7 6 5 4 3 2 1
R EV C /N D ESC R IPTIO N D ATE D R AW N BY: C H EC KED BY: APPR O VED BY:
SEE SH EET 1
+5V +5V
D D
10.0K
10.0K
R 11
R 12
+5V
+5v_M O DEM
U7
*
1 40
*
SD A_D ISPLAY SD A S23 PP10 .1uF
2 39 J5:A E3
SC L_D ISPLAY SC L S22 PP9 1
3 38 C 17
PP6 SYNC S21 PP11
4 37 J5:B +5v_M O DEM
C LK S20 PP12 2 G ND
5 36
***
VD D S19 PP13
6 35 J5:C E15
+5V O SC S18 3 2
.1uF
C 32
7 34
A0 S17 C R 21
.1uF
C 33
8 33 J5:D E2 3
A1 S16 4 R XD 1
9 32 BAV99L
A2 S15
R 17
* G ND
10
11
SA0
VSS
S14
S13
31
30
5 J5:E E4
1 +5v_M O DEM
G ND 12 29 J5:F
VLC D S12 6 G ND
13 28 G ND
**
BP0 S11
14 27
ZER O
J5:G E6
R 20
BP2 S10 7 2
15 26
BP1 S9 C R 22
16 25 J5:H E5 3
BP3 S8 8 TXD 1
17 24 BAV99L
S0 S7
G ND 18 23
S1 S6
19 22 1
S2 S5
20 21
S3 S4
PC F8566T G ND
C C
D S1
S18 1
S19 23 S17
S9 2 S10
S18 22 BP0
S8 3 C1
S17 21 S16
S7 4 S9
S16 20 S15 SM AR T C AR D
S6 5 S8 +5v_M O DEM
S15 19 S14
S5 6 S7
S14 18 S3
S4 7 S6
S13 17 S13
S2 8 S5 J6:A E14
S12 16 S1 C1
S0 9 S4
S11 15 S12 1000 O H M S @ 100 M H Z
BP3 10 S3 +5v_M O DEM
.1uF
C 78
C4 14 S11 J6:B
BP2 11 S2 C2
C3 13 S10
BP1 12 S1
C2 +5v_M O DEM
U TS-J511JV
10.0K
R 74
G ND
2
C R 24
J6:C E10 3
C3 SC L
1000 O H M S @ 100 M H Z BAV99L
+5v_M O DEM
C4 J6:D E11
1
1000 O H M S @ 100 M H Z +5v_M O DEM
J6:E E16
C5
10.0K
1000 O H M S @ 100 M H Z G ND
R7
J6:F
C6 2
B G ND B
C R 23 3
+3.3V C7 J6:G E17 SD A
1
3 4 1000 O H M S @ 100 M H Z BAV99L
U 21:A J6:H E18
2 6 9 C8
U 21:B 1
1000 O H M S @ 100 M H Z
2.49K
5 8
R 148
+3.3V
2.49K
R 150
R 147
2.49K
R AW _PW M
12
R 151
1.30K
.22uF
LM 339M
C1
11
U 21:D 6
13
U 1:B 1
PW M
74LV00 +3.3V
7
(7=G N D ,14=+3.3V)
G ND G ND (3=VBU LK,12=G N D )
2.49K
R 152
A A
PR O D U C T LIN E:
R 149
2.49K
C -FLEX
PR O C ESS: TITLE:
AU TO C -FLEX,D O M ESTIC
SCHEM ATIC
G ND DIG ITAL
8 7 6 5 4 3 2 1
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual B-5
Appendix B
8 7 6 5 4 3 2 1
R EV C /N D ESC R IPTIO N D ATE D R AW N BY: C H EC KED BY: APPR O VED BY:
SEE SH EET 1
+8V
D D
1 J7:A
CO NN M 3 PIN .078
.01uF
C 48
+8V +8V
+8V
J7:B +5V
2
24.9K
R 49
CO NN M 3 PIN .078
10.0K
R 54
G ND LM 324A
3.16K
1.00K
R 64
R 53
12 G ND
1.00K
R 55
14 R 61 R 62 +5V
U 13:D
M T1 13 2.00K 30.1K
1 J7:C 11
.1uF
C 51
3 CO NN M 3 PIN .078
2 (4=+8V,11=G N D ) 13
TP8 2 U 12:D H ALL
3
4
R 65
1.00K
+8V
6
U 13:B
LM 324A
R 63 * C R 11
3
FLO W
10
LM 339M
5 G ND 7 1.00K (3=VBU LK,12=G N D )
BAV99L
6 5
(4=+8V,11=G N D )
10.0K
R 56
1
.22uF
7
10uF
C 46
C 47
24.9K
R 59
8 LM 324A
2 R 60 R 69
AIR FLO W SEN SO R U 13:A 1 2.00K 30.1K G ND
3 G ND
(4=+8V,11=G N D ) G ND G ND
G ND
.1uF
C 49
4.22K
R 50
3 10.0K
LM 324A
9
2.2uF
C 75
U 13:C 8
10
(4=+8V,11=G N D )
G ND
5
FLO W _O S
7
U 11:B 5
6
LM 358A 2
U 12:A PP15
4
(4=G N D ,8=+8V)
LM 339M
(3=VBU LK,12=G N D )
+5V
R 58
7
4.75K 1
U 12:B PP14
6
ZER O
R 47
2.49K
R 57
LM 339M
(3=VBU LK,12=G N D )
B C 66 B
9
.1uF G ND 14
M T2 U 12:C PP17
1 8
NC PP19 C 67
2 +5V LM 339M
VS
3 (3=VBU LK,12=G N D )
G ND .01uF
4 G ND
VO U T
M PXV5004G C 6U G ND R 42 2
3
R 67 C R 25 G ND
2.00K 1 3
U 10:A PR ESS
2 ZER O
.01uF
BAV99L
C 62
C 63
.1uF
LM 358A
(4=G N D ,8=+8V) 1
G ND G ND R 36 G ND
3.01K
10.0K
R 35
+5V
G ND
* M T3
+VS
1
2
VO
4
NC
3 5
G ND
ASD X001 7
U 10:B
A 6
LM 358A
A
G ND PR O D U C T LIN E:
G ND (4=G N D ,8=+8V)
C -FLEX
PR O C ESS: TITLE:
AU TO C -FLEX,D O M ESTIC
SCHEM ATIC
AN ALO G C IR C U IT
8 7 6 5 4 3 2 1
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual B-6
Appendix B
8 7 6 5 4 3 2 1
R EV C /N D ESC R IPTIO N D ATE D R AW N BY: C H EC KED BY: APPR O VED BY:
R 97 SEE SH EET 1
VBU LK
10.0K VBU LK
R 84
G ND 3 1
10.0K
1 3
C TL_H I_C
4 U 8:A U 5:A PP4
R 85 2 2
6 Q 17 Q 16 LM 358A
U 5:B G ND
5 M M BT3904LT1 M M BT3904LT1 74LV08
.22uF
2.49K
C 79
PW M (4=G N D ,8=+5V) (7=G N D ,14=+3.3V)
74LV08
2.49K
(7=G N D ,14=+3.3V) G ND
R 86
90.9K
G ND +3.3V
R 82
Q 15 G ND
D 9 D
M M BT3906LT1 8
3 Q 23 U 4:C PP5
10
.1uF
C 84
R 83 4
G ND VO U T 5
232 74LV00
G ND VO U T 6
(7=G N D ,14=+3.3V)
VO U T 7
8 G ND
G ND VBU LK R 77 2 VO U T
R 78 232
1 N D S8858H
10.0K
90.9K
R 76
1 Q 19
C TL_LO _C
3 R 80 M M BT3904LT1
U 4:A Q 20
2 2.49K M M BT3904LT1
PW M
74LV00
(7=G N D ,14=+3.3V)
2.49K
R 79
Q 18
10.0K
10.0K
G ND
R 94
R 95
M M BT3906LT1
G ND
G ND G ND G ND
C 86
G ND
C C TL_H I_B
9
10.0K
* C
8 R 117
U 5:C Q 10 Q9
.22uF
C 91
10 2.49K M M BT3904LT1 M M BT3904LT1 G ND
PW M
PHASE_B E1:B CO NN M 3 PIN 100
74LV08 2
J1:B
(7=G N D ,14=+3.3V) 165 O H M S @ 100 M H Z
R 118
2.49K
R 114
90.9K
G ND
C 90
G ND
Q8
M M BT3906LT1 3 Q 24
*
R 115 4
G ND VO U T 5 G ND
232
VO U T 6
G ND PHASE_A E1:A CO NN M 3 PIN 100
3
VO U T 7 J1:C
165 O H M S @ 100 M H Z
VO U T 8
C 95
R 108 2
VBU LK
232 * +3.3V
*
R 109 1 N D S8858H
TP6 TP4
R 107
90.9K
10.0K
R 100
G ND
*
4 Q 11
*
C TL_LO _B
R 111 R 128 R 140 TP1
*
6 M M BT3904LT1
U 4:B Q 13
5 2.49K M M BT3904LT1 10
PW M U 1:D 13
BEM F
R 158
ZER O
C 96
74LV00
11
*
(7=G N D ,14=+3.3V) LM 339M
R 110
2.49K
G ND (3=VBU LK,12=G N D )
10.0K
Q 12
R 96
*
M M BT3906LT1 R 139 R 138
G ND G ND
*
B R 121 R 127 ZER O B
G ND
*
G ND
G ND TP2
C4
*
VBU LK
R 91 VBU LK TP7
G ND
10.0K R 144 TP5
* G ND
*
*
10.0K
12 R 120 R 126
C TL_H I_A
11 R 145
U 5:D Q3 Q2
.22uF
C 100
C6
PW M
*
74LV08 M TR _C TR _EST
(7=G N D ,14=+3.3V)
R 146
2.49K
R 142
90.9K
G ND
G ND
Q1 TP3 G ND
3 Q 25
*
*
M M BT3906LT1 R 143 4 R 119 R 125
G ND VO U T 5
232
VO U T 6
G ND
R 157
ZER O
7
C5
VO U T
*
VO U T 8 C R 10
VBU LK R 134 2
232 B160
R 135 1 N D S8858H
FAST_C U R R EN T G ND G ND
G ND
90.9K
R 133
10.0K
12
A
C TL_LO _A
13
U 4:D
11 R 137
2.49K
Q7
M M BT3904LT1
Q5
M M BT3904LT1
R 70 5
TP9
* A
PW M 1.00K 7 R 66 PR O D U C T LIN E:
0.12
R 28
1.00K
.22uF
10.0K
G ND
R 90
C 74
8 7 6 5 4 3 2 1
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual B-7
Appendix B
8 7 6 5 4 3 2 1
R EV C /N D ESC R IPTIO N D ATE D R AW N BY: C H EC KED BY: APPR O VED BY:
SEE SH EET 1
S14K320
L1 C 50
4 1
750K
R V1
.1uF
1/2 W
R 41
C 54
AC IN
2
L2 BR 86D L .001uF
+ C 55
P6SM B150AT3
SC H U R TER 4300.0099 2 3 4 3
F1 IC A-0885 35 m h .6A PW R G N D 3 68uF G ND
C R 17
1.00M
400V
R 44
R 40
68K
2W
2.5A
998 O H M S @ 100 M H Z
998 O H M S @ 100 M H Z
PW R G N D
1.00M
R 43
E9
E7
VBU LK VBU LK VBU LK VBU LK
M U R S160
C R 18
SM BJ10C A
U 16
CR7
1
T1 C 59 + C 41
C 104
.1uF
.1uF
C 44
C O N TR O L 7 1 1500uF
DRAIN
F2
25V
5A
4700pF
2
L-SEN SE
.01uF
4
SM BJ10C A
.1uF
C 69
C 68
SO U R C E C R 16
3 10
CR6
5 X-LIM IT E8
FR EQ 2
SB560
TO P249Y G ND G ND
998 O H M S @ 100 M H Z
PW R G N D PW R G N D PW R G N D PW R G N D C R 19 C 53 +
.1uF
C 70
C 5
1000uF
25V
G ND C
B160
6
.1uF
C 64
4 G ND G ND
G ND
IC A-0638
PW R G N D
C 71 C 72 C 73
*
U 15
.01uF
C R 14
C 65
6 1 R 51 R 52
PP18
301 M M SZ5240BT1
5 2
4 3 G ND
PP16
VBU LK
C N Y17F-2
VBU LK
R 48
6.2
+8V F3 CR2 1 L
U 17
KA78M 05R 5A M SP845 3 SS J3
1 3 2 SC
B IN O UT + B
47uF
G ND +
10V
C 56
.22uF
C 58
R 45
499
4 C 57 G ND
2.2uF
+5V
PW R G N D
G ND
G ND
R 46
232
+5V
R 141
4.22K
10.0K
R 98
+5v_M O DEM +3.3V +3.3V
+5V
U 18 U9 (3=VBU LK,12=G N D )
G ND LM 339M
KA78M 05R LT1521C ST-3.3
1 3 8
IN O UT IN O UT
U 1:C D C _O N
.22uF
.22uF
.22uF
.22uF
C 102
C 105
14
C 83
C 88
G ND G ND
.22uF
.22uF
9
C 60
C 61
2.2uF
2.2uF
C 43
C 45
4 TAB
R 129
10.0K
C 103
.1uF
G ND G ND G ND
G ND G ND G ND G ND
+5v_M O DEM
U6
KA78M 05R G ND G ND
1 3
IN O UT
G ND
.22uF
.22uF
C 34
C 39
4
+5V +3.3V +8V VBU LK
A A
G ND G ND G ND *
TP14
PR O D U C T LIN E:
TP12 TP13 TP11 TP10 C -FLEX
* G ND * * * PR O C ESS: TITLE:
AU TO C -FLEX,D O M ESTIC
SCHEM ATIC
PO W ER SU PPLY
8 7 6 5 4 3 2 1
1018658 REMstar Auto with C-Flex and REMstar Pro 2 Service Manual B-8
www.respironics.com