Basics of LC/MS

Primer

Contents
Why Liquid Chromatography/Mass Spectrometry? Instrumentation Ion Sources Electrospray ionization Atmospheric pressure chemical ionization Atmospheric pressure photoionization Mass Analyzers Quadrupole Time-of-flight Ion trap Fourier transform-ion cyclotron resonance (FT-ICR) Collision-Induced Dissociation and Multiple-Stage MS CID in single-stage MS CID and multiple-stage MS Adapting LC Methods Sample preparation Ionization chemistry Capillary LC/MS and CE/MS Applications Molecular Weight Determination Differentiation of similar octapeptides Determining the molecular weight of green fluorescent protein Structural Determination Structural determination of ginsenosides using MSn analysis Pharmaceutical Applications Rapid chromatography of benzodiazepines Detection of degradation products for salbutamol Identification of bile acid metabolites Biochemical Applications Rapid protein identification using capillary LC/MS/MS and database searching Clinical Applications High-sensitivity detection of trimipramine and thioridazine Food Applications Identification of aflatoxins in food Determination of vitamin D3 in poultry feed supplements using MS3 Environmental Applications Detection of phenylurea herbicides Detection of low levels of carbaryl in food CE/MS Applications Analysis of peptides using CE/MS/MS 4 6 6 7 8 9 10 10 11 11 12 13 14 14 16 17 17 19 20 20 20 21 22 22 24 24 24 25 26 26 28 28 28 28 30 32 32 32 34 34

Why Liquid Chromatography/ Mass Spectrometry?


Liquid chromatography is a fundamental separation technique in the life sciences and related fields of chemistry. Unlike gas chromatography, which is unsuitable for nonvolatile and thermally fragile molecules, liquid chromatography can safely separate a very wide range of organic compounds, from small-molecule drug metabolites to peptides and proteins. Traditional detectors for liquid chromatography include refractive index, electrochemical, fluorescence, and ultraviolet-visible (UV-Vis) detectors. Some of these generate twodimensional data; that is, data representing signal strength as a function of time. Others, including fluorescence and diodearray UV-Vis detectors, generate three-dimensional data. Three-dimensional data include not only signal strength but spectral data for each point in time.

Mass spectrometers also generate threedimensional data. In addition to signal strength, they generate mass spectral data that can provide valuable information about the molecular weight, structure, identity, quantity, and purity of a sample. Mass spectral data add specificity that increases confidence in the results of both qualitative and quantitative analyses.

CH2CH2CH2NHCOCH3 CH2CH2CH2OH CH2CH2CH(NH2)COOH

% Rel. Abundance

% Rel. Abundance

m/z

Figure 1. Two-dimensional abundance data and three-dimensional mass spectral data from a mass spectrometer

For most compounds, a mass spectrometer is more sensitive and far more specific than all other LC detectors. It can analyze compounds that lack a suitable chromophore. It can also identify components in unresolved chromatographic peaks, reducing the need for perfect chromatography. Mass spectral data complements data from other LC detectors. While two compounds may have similar UV spectra or similar mass spectra, it is uncommon for them to have both. The two orthogonal sets of data can be used to confidently identify, confirm, and quantify compounds.

Some mass spectrometers have the ability to perform multiple steps of mass spectrometry on a single sample. They can generate a mass spectrum, select a specific ion from that spectrum, fragment the ion, and generate another mass spectrum; repeating the entire cycle many times. Such mass spectrometers can literally deconstruct a complex molecule piece by piece until its structure is determined.

1200000 800000 400000

MS TIC

4000 2000

m/z 648

12000 8000 4000

m/z 376

300 200 100 5 10 15

m/z 1325
20 25 30 35 40

min.

Figure 2. Identification of three components in a chromatographically unresolved peak

Instrumentation
Mass spectrometers work by ionizing molecules and then sorting and identifying the ions according to their mass-to-charge (m/z) ratios. Two key components in this process are the ion source, which generates the ions, and the mass analyzer, which sorts the ions. Several different types of ion sources are commonly used for LC/MS. Each is suitable for different classes of compounds. Several different types of mass analyzers are also used. Each has advantages and disadvantages depending on the type of information needed.

molecules from the analyte molecules (direct liquid inlet, thermospray) or did so before ionization (particle beam). The analyte molecules were then ionized in the mass spectrometer under vacuum, often by traditional electron ionization. These approaches were successful only for a very limited number of compounds. The introduction of atmospheric pressure ionization (API) techniques greatly expanded the number of compounds that can be successfully analyzed by LC/MS. In atmospheric pressure ionization, the analyte molecules are ionized first, at atmospheric pressure. The analyte ions are then mechanically and electrostatically separated from neutral molecules. Common atmospheric pressure ionization techniques are: Electrospray ionization (ESI) Atmospheric pressure chemical ionization (APCI) Atmospheric pressure photoionization (APPI)

Ion Sources
Much of the advancement in LC/MS over the last ten years has been in the development of ion sources and techniques that ionize the analyte molecules and separate the resulting ions from the mobile phase. Earlier LC/MS systems used interfaces that either did not separate the mobile phase
100,000 Electrospray Ionization 10,000 Molecular Weight

Figure 3. Applications of various LC/MS ionization techniques

1000

APPI APCI

100

10 Nonpolar Polarity Very Polar

Electrospray ionization Electrospray relies in part on chemistry to generate analyte ions in solution before the analyte reaches the mass spectrometer. The LC eluent is sprayed (nebulized) into a chamber at atmospheric pressure in the presence of a strong electrostatic field and heated drying gas.
Ions Nebulizer gas

Some gas-phase reactions, mostly proton transfer and charge exchange, can also occur between the time ions are ejected from the droplets and the time they reach the mass analyzer. Electrospray is especially useful for analyzing large biomolecules such as proteins, peptides, and oligonucleotides, but can also analyze smaller molecules like benzodiazepines and sulfated conjugates. Heated nitrogen drying gas Large molecules often acquire more + + than one charge. Thanks to this multiple charging, electrospray can be used to analyze molecules as large as 150,000 u even though the mass range (or more accurately mass-to-charge range) for a typical LC/MS instruments is around 3000 m/z. For example: 100,000 u / 10 z = 1,000 m/z When a large molecule acquires many charges, a mathematical process called deconvolution is often used to determine the actual molecular weight of the analyte.

Solvent spray
+ + + + + +

+ + + + + + + +

Dielectric capillary entrance

Figure 4. Electrospray ion source

The electrostatic field causes further dissociation of the analyte molecules. The heated drying gas causes the solvent in the droplets to evaporate. As the droplets shrink, the charge concentration in the droplets increases. Eventually, the repulsive force between ions with like charges exceeds the cohesive forces and ions are ejected (desorbed) into the gas phase. These ions are attracted to and pass through a capillary sampling orifice into the mass analyzer.

Evaporation

Analyte ion ejected

Figure 5. Desorption of ions from solution

+-+ + + - + + - -+ + - + ++

++ ++ -+ + + - -++ ++

+ ++- + + + -- -+++++ ++ +-+ ++

Atmospheric pressure chemical ionization In APCI, the LC eluent is sprayed through a heated (typically 250C 400C) vaporizer at atmospheric pressure. The heat vaporizes the liquid. The resulting gas-phase solvent molecules are ionized by electrons discharged from a corona needle. The solvent ions then transfer charge to the analyte molecules through chemical reactions (chemical ionization). The analyte ions pass through a capillary sampling orifice into the mass analyzer.

APCI is applicable to a wide range of polar and nonpolar molecules. It rarely results in multiple charging so it is typically used for molecules less than 1,500 u. Due to this, and because it involves high temperatures, APCI is less well-suited than electrospray for analysis of large biomolecules that may be thermally unstable. APCI is used with normal-phase chromatography more often than electrospray is because the analytes are usually nonpolar.

HPLC inlet Nebulizer (sprayer) Vaporizer (heater)

Figure 6. APCI ion source

Nebulizing gas

Drying gas

+ + +

+ + + + + + + + + + +

Corona discharge needle

Capillary

Atmospheric pressure photoionization Atmospheric pressure photoionization (APPI) for LC/MS is a relatively new technique. As in APCI, a vaporizer converts the LC eluent to the gas phase. A discharge lamp generates photons in a narrow range of ionization energies. The range of energies is carefully chosen to ionize as many analyte molecules as possible while minimizing the ionization of solvent molecules. The resulting ions pass through a capillary sampling orifice into the mass analyzer.
Figure 7. APPI ion source

APPI is applicable to many of the same compounds that are typically analyzed by APCI. It shows particular promise in two applications, highly nonpolar compounds and low flow rates (<100 l/min), where APCI sensitivity is sometimes reduced. In all cases, the nature of the analyte(s) and the separation conditions have a strong influence on which ionization technique: electrospray, APCI, or APPI, will generate the best results. The most effective technique is not always easy to predict.
HPLC inlet Nebulizer (sprayer) Vaporizer (heater)

Nebulizing gas

UV lamp

Drying gas

hv
+

+ + +

+ + + + + + + + + + +

Capillary

Mass Analyzers
Although in theory any type of mass analyzer could be used for LC/MS, four types: Quadrupole Time-of-flight Ion trap Fourier transform-ion cyclotron resonance (FT-ICR or FT-MS)

through the filter at a given time. Quadrupoles tend to be the simplest and least expensive mass analyzers. Quadrupole mass analyzers can operate in two modes: Scanning (scan) mode Selected ion monitoring (SIM) mode In scan mode, the mass analyzer monitors a range of mass-to-charge ratios. In SIM mode, the mass analyzer monitors only a few massto-charge ratios. SIM mode is significantly more sensitive than scan mode but provides information about fewer ions. Scan mode is typically used for qualitative analyses or for quantitation when all analyte masses are not known in advance. SIM mode is used for quantitation and monitoring of target compounds.

are used most often. Each has advantages and disadvantages depending on the requirements of a particular analysis. Quadrupole A quadrupole mass analyzer consists of four parallel rods arranged in a square. The analyte ions are directed down the center of the square. Voltages applied to the rods generate electromagnetic fields. These fields determine which mass-to-charge ratio of ions can pass

Figure 8. Quadrupole mass analyzer


To detector

From ion source


1 Scan Mass Range

Scan Abundance

m/z

time SIM
1 Scan

m/z
Discrete Masses

Figure 9. The quadrupole mass analyzer can scan over a range of mass-to-charge ratios or alternate between just a few

Abundance

m/z

time

m/z

10

Time-of-flight In a time-of-flight (TOF) mass analyzer, a uniform electromagnetic force is applied to all ions at the same time, causing them to accelerate down a flight tube. Lighter ions

travel faster and arrive at the detector first, so the mass-to-charge ratios of the ions are determined by their arrival times. Time-offlight mass analyzers have a wide mass range and can be very accurate in their mass measurements. Ion trap An ion trap mass analyzer consists of a circular ring electrode plus two end caps that together form a chamber. Ions entering the chamber are trapped there by electromagnetic fields. Another field can be applied to selectively eject ions from the trap. Ion traps have the advantage of being able to perform multiple stages of mass spectrometry without additional mass analyzers.

Flight tube (field-free region)

+
Repeller

Detector

Accumulation From ion source Flight tube (field-free region)

Repeller

Detector

Ejection + + + + Abundance

+ +

Accumulation

Ejection

m/z
Abundance

Figure 10. Time-of-flight mass analyzer

m/z

Figure 11. Ion trap mass analyzer

11

Fourier transform-ion cyclotron resonance (FT-ICR) An FT-ICR mass analyzer (also called FT-MS) is another type of trapping analyzer. Ions entering a chamber are trapped in circular orbits by powerful electrical and magnetic fields. When excited by a radio-frequency (RF) electrical field, the ions generate a timedependent current. This current is converted by Fourier transform into orbital frequencies of the ions which correspond to their mass-tocharge ratios.

Like ion traps, FT-ICR mass analyzers can perform multiple stages of mass spectrometry without additional mass analyzers. They also have a wide mass range and excellent mass resolution. They are, however, the most expensive of the mass analyzers.

Figure 12. FT-ICR mass analyzer

12

The atmospheric pressure ionization techniques discussed are all relatively soft techniques. They generate primarily: Molecular ions M+ or M Protonated molecules [M + H]+ Simple adduct ions [M + Na]+ Ions representing simple losses such as the loss of a water [M + H H2O]+

CH3 350000 O 300000 O S NH N N CH3 [M + H]

279.1

Collision-Induced Dissociation and Multiple-Stage MS

250000 H2N

200000

150000

280.0

100000

301.0

50000

The resulting molecular weight information is very valuable, but complementary structural information is often needed. To obtain structural information, analyte ions are fragmented by colliding them with neutral molecules in a process known as collisioninduced dissociation (CID) or collisionally activated dissociation (CAD). Voltages are applied to the analyte ions to add energy to the collisions and create more fragmentation.
124.1

0 100 200 300 m/z

Figure 13. Mass spectrum of sulfamethazine acquired without collision-induced dissociation exhibits little fragmentation

m/z 156 m/z 186


O 186.0 N NH N CH3 O S 279.1 CH3

80000

Figure 14. Mass spectrum of sulfamethazine acquired with collision-induced dissociation exhibits more fragmentation and thus more structural information

m/z 108 m/z 213


H2N

156.1

60000

[M + H]

m/z 124

40000

108.2

301.0

[M + Na] 323.0

20000

107.1

213.2

125.1

0 100 200 300

157.1

187.0

280.1

m/z

281.0

[M + Na]

13

CID in single-stage MS CID is most often associated with multistage mass spectrometers where it takes place between each stage of MS filtering, but CID can also be accomplished in single-stage quadrupole or time-of-flight mass spectrometers. In single-stage mass spectrometers, CID takes place in the ion source and is thus sometimes called source CID or in-source CID. Analyte (precursor) ions are accelerated and collide with residual neutral molecules to yield fragments called product ions. The advantage of performing CID in singlestage instruments is their simplicity and relatively low cost. The disadvantage is that ALL ions present are fragmented. There is no way to select a specific precursor ion so there is no sure way to determine which

product ions came from which precursor ion. The resulting spectra may include mass peaks from background ions or coeluting compounds as well as those from the analyte of interest. This tradeoff may be acceptable when analyzing relatively pure samples, but does not give good results if chromatographic peaks are not well resolved or background levels are high. CID and multiple-stage MS Multiple-stage MS (also called tandem MS or MS/MS or MSn) is a powerful way to obtain structural information. In triple-quadrupole or quadrupole/quadrupole/time-of-flight instruments (see Figure 16), the first quadrupole is used to select the precursor ion. CID takes place in the second stage (quadrupole or octopole), which is called the collision cell.

LC/MS Abundance

Figure 15. In-source CID with a singlequadrupole mass analyzer

Capillary

CID

Q1

m/z

Capillary

Q1

Q2 (CID)

Q3

Abundance

LC/MS/MS

m/z

Figure 16. MS/MS in a triple-quadrupole mass spectrometer

14

The third stage (quadrupole or TOF) then generates a spectrum of the resulting product ions. It can also perform selected ion monitoring of only a few product ions when quantitating target compounds.

In ion trap and FT-ICR mass spectrometers, all ions except the desired precursor ion are ejected from the trap. The precursor ion is then energized and collided to generate product ions. The product ions can be ejected to generate a mass A. spectrum, or a particular product ion can be retained and collided to obtain another set of product ions. This process can be sequentially + automated so that the most + + abundant ion(s) from each stage of MS are retained and collided. This is a very powerful technique for determining the structure of Accumulation molecules. It is also a powerful technique for obtaining peptide C. mass information that relates to the sequence of amino acids in a peptide. + +
+ + +

A major advantage of multiple-stage MS is its ability to use the first stage of MS to discard nonanalyte ions. Sample cleanup and chromatographic separation become much less critical. With relatively pure samples, it is quite common to do away with chromatographic separation altogether and infuse samples directly into the mass spectrometer to obtain product mass spectra for characterization or confirmation.

B.

Nonprecursor ions ejected D.

++

Collision

Product ions ejected and detected

Abundance

Figure 17. MS/MS in an ion trap mass spectrometer

m/z

15

Adapting LC Methods
Early LC/MS systems were limited by fundamental issues like the amount of LC eluent the mass spectrometer could accept. Significant changes to LC methods were often required to adapt them to MS detectors. Modern LC/MS systems are more versatile. Many mass spectrometers can accept flow rates of up to 2 ml/min. With minor modifications, the same instruments can also generate good results at microliter and nanoliter flow rates. Ion sources with orthogonal (off-axis) nebulizers are more tolerant of nonvolatile buffers and require little or no adjustment, even with differing solvent compositions and flow rates.

Changes to LC methods required for modern LC/MS systems generally involve changes in sample preparation and solution chemistry to: Ensure adequate analyte concentration Maximize ionization through careful selection of solvents and buffers Minimize the presence of compounds that compete for ionization or suppress signal through gas-phase reactions

Figure 18. Salt deposits in this Agilent APCI ion source had little effect on performance thanks to orthogonal spray orientation and robust ion source design

16

Sample preparation Sample preparation generally consists of concentrating the analyte and removing compounds that can cause background ions or suppress ionization. Examples of sample preparation include: On-column concentration to increase analyte concentration Desalting to reduce the sodium and potassium adduct formation that commonly occurs in electrospray Filtration to separate a lowmolecular-weight drug from proteins in plasma, milk, or tissue Ionization chemistry Because formation of analyte ions in solution is essential to achieving good electrospray results, careful attention must be paid to proper solution chemistry. For electrospray: Select more volatile buffers to reduce the buildup of salts in the ion source Adjust solvent pH according to the polarity of ions desired and the pH of the sample

Use solvents that have low heats of vaporization and low surface tensions to enhance ion desorption Make sure that gas-phase reactions do not neutralize ions through proton transfer or ion pair reactions If pH adjustments interfere with proper chromatography, postcolumn modification of the solvent may be a good solution. This can improve MS response without compromising chromatography.
+13

pH 2.5
+12 +14 +11 +15 +10 +9

+12 +11 +10 +13 +9 +14 +15 +8

pH 6

pH 12

Figure 19. Effect of solvent pH on the abundance of multiply charged ions of the protein Lysozyme in electrospray mode

17

Solution chemistry is less critical for APCI operation because ionization occurs in the gas phase, not the liquid phase, but solvent selection can still have a significant effect on APCI analyte signal response. Select more volatile solvents Select solvents with a lower charge affinity than the analyte Protic solvents generally work better than nonprotic solvents for positive ion mode For negative ionization, solvents that readily capture an electron must be used Ammonium salts in the mobile phase can cause ammonium adduct formation

Vaporizer temperature also affects APCI ionization results. The temperature must be hot enough to vaporize the solvent but not so hot as to cause thermal degradation of the analyte molecules.

200C

1 400C

7 8

9 10

11

12 13 14 15

16

10

11

12 13 14 15 16

Figure 20. APCI analysis with an inadequate vaporizer temperature (200C) yields poor results for some compounds compared to a more typical vaporizer temperature (400C)

18

Capillary LC/MS and CE/MS


Capillary LC and capillary electrophoresis (CE) are commonly used alongside HPLC. Capillary LC at microliter to nanoliter flow rates often provides better sensitivity than conventional flow rates for extremely small sample quantities. It is commonly used for protein and peptide analysis but has also proven very useful for the analysis of small quantities of drugs.

Capillary electrophoresis (CE) is a technique with a high separation efficiency and the added benefit of being able to handle very complex matrices. It has proven useful for such diverse samples as: peptides, drugs of abuse, drugs in natural products, flavanoids, and aromatic amines. With specialized nebulizers and method adjustments, electrospray ionization can be used for both capillary LC/MS and CE/MS. The mass spectrometry benefits of selectivity and sensitivity are available to both of these separation techniques.

2.0

Base Peak 50 mM NaPO3


Terbutaline 5 g/mL
HO OH HO NH NH HO HO OH

% Relative Abundance

1.5

Salbutamol 5 g/mL

1.0

0.5

0.0 2 4 6 8 10 12

Time (min)

Figure 21. CE/MS/MS analysis of terbutaline and salbutamol shows good chromatographic separation and good signal even in the presence of a high concentration of sodium phosphate salt

19

Applications
LC/MS is suitable for many applications, from pharmaceutical development to environmental analysis. Its ability to detect a wide range of compounds with great sensitivity and specificity has made it popular in a variety of fields.

Differentiation of similar octapeptides Figure 22 shows the spectra of two peptides whose mass-to-charge ratios differ by only 1 m/z. The only difference in the sequence is at the C-terminus where one peptide has threonine and the other has threonine amide. The smaller fragments are identical in the two spectra, indicating that large portions of the two peptides are very similar. The larger fragments contain the differentiating peptides.

Molecular Weight Determination


One fundamental application of LC/MS is the determination of molecular weights. This information is key to determining identity.

100 80 60

Ala-Ser-Thr-Thr-Thr-Asn-Tyr-Thr

739.3

444.2

498.2

576.3

343.1

20 0

397.2

841.3

100 80 60

Ala-Ser-Thr-Thr-Thr-Asn-Tyr-Thr amide

857.5

739.3

20 0

379.1

400

444.2

576.3

600

800

840.3

40

879.5

880.3

40

858.5

Figure 22. Mass spectra differentiating two very similar octapeptides

m/z

20

Determining the molecular weight of green fluorescent protein Green fluorescent protein (GFP) is a 27,000Dalton protein with 238 amino acids. It emits a green light when excited by ultraviolet light. During electrospray ionization, GFP acquires multiple charges. This allows it to be analyzed by a mass spectrometer with a relatively limited mass (mass-to-charge) range. Mass deconvolution is then used to determine the molecular weight of the protein.
Figure 23. Molecular weight determination of green fluorescent protein by electrospray LC/MS

The upper part of the display in Figure 23 shows the full scan mass spectrum of GFP. The pattern of mass spectral peaks is characteristic of a multiply charged analyte. Each peak represents the molecule with a different number of charges. The lower display is a deconvoluted mass spectrum generated by the data system for the singly charged analyte.

789.95 100000 80000 60000 40000 20000 726.25 746.15 767.55 120000

839.45 895.45 866.45 926.15 959.15 994.85 1032.95 1074.15 1118.85 1167.55

1220.45

1278.75

1342.65

1413.25

1000 1000000 800000 600000 400000 200000

Deconvoluted spectrum MW = 26828.84

26700

26800

26900

27000

1579.65 1500

21

Structural Determination
Another fundamental application of LC/MS is the determination of information about molecular structure. This can be in addition to molecular weight information or instead of molecular weight information if the identity of the analyte is already known. Structural determination of ginsenosides using MSn analysis Ginseng root, a traditional Chinese herbal remedy, contains more than a dozen biologically active saponins called ginsenosides. Since most ginsenosides contain multiple oligosaccharide chains at different positions in the molecule, structural elucidation of these compounds can be quite complicated.

MSn analysis in an ion trap mass spectrometer permits multiple stages of precursor ion isolation and fragmentation. This stepwise fragmentation permits individual fragmentation pathways to be followed and provides a great deal of structural information. Figure 24 shows the full scan mass spectrum from a direct infusion of the ginsenoside Rb1. The most prominent feature is the sodium adduct ion [M + Na]+ at m/z 1131.7. MS/MS of m/z 1131.7 yields a product ion at m/z 789.7 corresponding to cleavage of a single glycosidic bond (Figure 25). Subsequent isolation and fragmentation of m/z 789.7 (Figure 26) yields two products: a more abundant ion at m/z 365.1 corresponding to loss of the oligosaccharide chain (Glc 2 Glc), and a less abundant ion at m/z 627.5 representing the loss of a deoxyhexose sugar.
Figure 24. Full scan mass spectrum of ginsenoside Rb1 showing primarily sodium adduct ions

100

HO OH OH

1131.7
OO OH O O OH OH OH

80

OH

% Relative Abundance

[M + Na]+ m/z 1131.7 60


HO OH OO

40

HO

HO OH HO OH OO

20

0 200 400 600 800 1000 1200

m/z

22

100

HO

OO OH OH OH

789.7
O OH OH OH OH O

MS/MS of 1131.7

Figure 25. Full scan product ion (MS/MS) spectrum from the sodium adduct at m/z 1131.7

80

m/z 789.7 [M + Na] m/z 1131.7


+

% Relative Abundance

60

HO O OH HO HO OH HO O

m/z 365.2

40

OH

20

365.2

0 200 400 600 800 1000

1131.7 1200

m/z

Figure 26. Subsequent full scan product ion spectrum (MS3) from the ion at m/z 789.7

365.1 100

1131.7
OH

789.7

80

Na+ m/z 627.5

% Relative Abundance

m/z 365.1
HO O O

60
HO

OH OH O OH O

40

OH OH

20

627.5

0 200 300 400 500 600 700 800

m/z

23

Pharmaceutical Applications
Rapid chromatography of benzodiazepines The information available in a mass spectrum allows some compounds to be separated even though they are chromatographically unresolved. In this example, a series of benzodiazepines was analyzed using both UV and MS detectors. The UV trace could not be used for quantitation, but the extracted ion chromatograms from the MS could be used. The mass spectral information provides additional confirmation of identity. Chlorine has a characteristic pattern because of the relative abundance of the two most abundant isotopes. In Figure 27, the triazolam spectrum shows that triazolam has two chlorines and the diazepam spectrum shows that diazepam has only one.

Identification of bile acid metabolites The MSn capabilities of the ion trap mass spectrometer make it a powerful tool for the structural analysis of complex mixtures. Intelligent, data-dependent acquisition techniques can increase ion trap effectiveness and productivity. They permit the identification of minor metabolites at very low abundances from a single analysis. One application is the identification of metabolic products of drug candidates.

Rapid chromatography and isotopic information


mAU 2 1.5 1 0.5 0
1 2 3 4 min

UV
343.1

Figure 27. MS identification and quantification of individual benzodiazepines from an incompletely resolved mixture

Triazolam

345.1

2 Cl
350 300 250 200
Diazepam

MS
Triazolam Clobazam

285.2

Diazepam
287.2

150 100 50
oxazepam

Alprazolam Estrazolam Lorazepam Diazepam

1 Cl

250

300

350

0 1 2 3 4 min

24

This example uses the in vitro incubation of the bile acid deoxycholic acid with rat liver microsomes to simulate metabolism of a drug candidate. Intelligent, data-dependent acquisition was used to select the two most abundant, relevant ions in each MS scan. These precursor ions were automatically fragmented and full scan product ion spectra collected. Figure 28A shows the base peak chromatogram. Figure 28B shows the extracted ion chromatogram of the [M-H] ion at m/z 407

corresponding to a predicted minor metabolite (cholic acid) that eluted at 9.41 minutes. The full scan MS/MS product spectrum (Figure 28C) from the ion at m/z 407 confirms the identity. Significant time was saved because the confirming MS/MS product ion spectra were acquired automatically in the same run as the full scan MS data.

11.64 5 4 3 2 1 0 2 4 3 2 1 0 2 4 6 289.5
H O OH H OH H H OH H H OH

Base Peak: m/z 150500 Minor Metabolite


10.24 0.77

Figure 28. Identification of a minor metabolite of deoxycholic acid through MS/MS

12.01 12.28 13.49

8 9.41

10

12

14

B
m/z 407

8 343.4

10

12

14

100 80 60 40 20 0 100 200

C
MS/MS of 407 at 9.41 min

325.3

389.5

Cholic Acid 300 400 500 m/z

25

Biochemical Applications
Rapid protein identification using capillary LC/MS/MS and database searching Traditional methods of protein identification generally require the isolation of individual proteins by two-dimensional gel electrophoresis. The combination of capillary LC/MS/MS with intelligent, data-dependent acquisition and probability-based database searching makes it possible to rapidly identify as many as 100 proteins in a single analysis.

In this example, a capillary LC and ion trap mass spectrometer were used to acquire data from a mixture of five tryptically digested proteins at a concentration of 1 pmol/l each (Figure 29). Using intelligent, automated datadependent acquisition, a full scan product ion (MS/MS) spectrum was acquire from the most abundant relevant ion in each mass scan throughout the entire run. All MS and MS/MS data were acquire from a single analysis.

Intens. x107 0.8

18.39 17.55

31.37 0.6 0.4 0.2 0.0 5 10 15 20 25 30 35 Time [min] 4.43 11.75 13.43 14.43 20.59 21.09 22.35 30.37 Base Peak: m/z 150 2000

Figure 29. Base peak chromatogram generated from 1 pmol total material injected on column

26

Protein identification was accomplished using MASCOT software that correlated the uninterpreted MS/MS data with sequences in a database. Figure 30 demonstrates the excellent match between the observed MS/MS spectrum from the most abundant ion (m/z 807.2) in the chromatographic peak at 17.55 minutes and the theoretical y-ion series predicted for a tryptic peptide from human apolipoprotein, one of the proteins in the sample mixture.

LLDNWDSVTSTFSK
Y(9)

Theoretical
381.2
Y(3) Y(6) Y(8)

971.5 856.4 1386.6 1157.5


Y(10) Y(12)

670.3 569.3
Y(5)

100 670.4 % Relative Abundance 80 856.5

971.6 400 600 800 1000 1200 1400

1386.7 60 381.3 569.3 769.5 1157.5 1271.5

40

Observed MS/MS of m/z 807.2

20

0 200 600 1000 1400 1800 m/z

Figure 30. Full scan MS/MS spectra from the doubly charged parent ion m/z 807.2 and the matching theoretical sequence identified by database searching

Y(11)

769.4
Y(7)

1271.6

27

Clinical Applications
High-sensitivity detection of trimipramine and thioridazine For most compounds, MS is more sensitive than other LC detectors. Trimipramine is a tricyclic antidepressant with sedative properties. Thioridazine is a tranquilizer. Figure 31 shows these compounds in a urine extract at a level that could not be detected by UV. To get the maximum sensitivity from a single-quadrupole mass spectrometer, the analysis was done by selected ion monitoring.

Food Applications
Identification of aflatoxins in food Aflatoxins are toxic metabolites produced in foods by certain fungi. Figure 32 shows the total ion chromatogram from a mixture of four aflatoxins. Even though they are structurally very similar, each aflatoxin can be uniquely identified by its mass spectrum (Figure 33).

Figure 31. Trimipramine and thioridazine in a urine extract


mAU 200 0 80000 0 40000 0 20000 0 8000

UV 280 nm Trimipramine EIC m/z 295 EIC m/z 100

Trimipramine

EIC m/z 371 EIC m/z 126


1 2

Thioridazine Thioridazine

0 3 4 5 6 min

28

Scan of 2 ng

Figure 32. Total ion chromatogram of a mixture of aflatoxins

4
200000

TIC

160000

1) Aflatoxin G2
120000

2) Aflatoxin G1 3) Aflatoxin B2

80000

4) Aflatoxin B1
40000

6.00

8.00

10.00

12.00

14.00

16.00

18.00

Figure 33. Unique mass spectra allow positive identification of structurally similar aflatoxins

331 [M+H]

90

Aflatoxin G2
O O O O

315 + [M+H]

90 Aflatoxin B2
313 + [M-OH] 353 + [M+Na]
O O O

50
O O OCH 3

50
O O

OCH3

287

337 + [M+Na]

120

160

200

240

280

320

360

120

160

200

240

280

320

329 [M+H] +

313 [M+H] +

90 Aflatoxin G1
O O O O

311 + [M-OH] 351 [M+Na]+

90

Aflatoxin B1
O O

50
O O

50
243 283
O O

285
OCH3

OCH3

335 + [M+Na]

120

160

200

240

280

320

360

120

160

200

240

280

320

29

Determination of vitamin D3 in poultry feed supplements using MS3 Vitamin D is an essential constituent in human and animal nutrition. Livestock diets deficient in vitamin D can cause growth abnormalities. Traditional GC/MS analysis methods for vitamin D3 in feed extracts require extensive and time-consuming sample preparation and derivatization prior to analysis. Atmospheric pressure chemical ionization with ion trap detection provides a sensitive analytical method without the need for extensive sample

preparation and derivatization. Further, the multiple-stage MS capability of the ion trap eliminates the need for chromatographic separation, greatly speeding analyses. Flow injection analysis of a poultry feed extract yields a peak at m/z 385 suggesting the presence of vitamin D3. Isolation and fragmentation of the precursor ion at m/z 385 is inconclusive. The full scan product ion spectrum shows a prominent peak at m/z 367 representing the loss of a single water molecule but little other fragmentation (Figure 34).

m/z 367
100 367.4

[M+H]+ H2O

80

% Relative Abundance

H 60

Poultry Feed Extract MS2 m/z 385

40

HO

20 199.3 0 150 200 250 300 350 400 450 500 219.0 255.2 287.3 325.4

m/z

Figure 34. Full scan MS/MS product ion spectrum from the precursor ion at m/z 385 showing primarily the nonspecific loss of a water molecule

30

Isolation and fragmentation of the ion at m/z 367 yields a full scan MS3 spectrum (Figure 35A) rich in structurally specific product ions. This spectrum is an excellent

match with a similar analysis of a pure standard (Figure 35B) and conclusively confirms the presence of vitamin D3.

100

255.1

A Poultry Feed Extract MS3 m/z 385 367

% Relative Abundance

80 60 241.2 40 185.0 20 158.5 0 150 200 250 300 213.1 227.2 285.4 271.1

311.2

325.5

353.2

350

400

500

m/z
100 255.2

B
271.1

% Relative Abundance

80 60 40 20 0 150 200 250 300 158.6 185.0

213.0 227.2

241.3 285.3 325.5

Pure Standard Vitamin D3 MS3 m/z 385 367

349.2 350

367.6 400 500

m/z

Figure 35. Full scan MS3 product ion spectra show much more structural information

31

Environmental Applications
Detection of phenylurea herbicides Many of the phenylurea herbicides are very similar and difficult to distinguish with a UV detector (Figure 36). Monuron and diuron have one benzene ring and differ by a single chlorine. Chloroxuron has two chlorines and a second benzene ring attached to the first by an oxygen. The UV-Vis spectra are similar for diuron and monuron, but different for chloroxuron. When analyzed using electrospray ionization on an LC/MS system, each compound has a uniquely identifiable mass spectrum.

Detection of low levels of carbaryl in food Pesticides in foods and beverages can be a significant route to human exposure. Analysis of the carbamate pesticide carbaryl in extracts of whole food by ion trap LC/MS/MS proved more specific than previous analyses by HPLC fluorescence and single-quadrupole mass spectrometry. The protonated carbaryl molecule (m/z 202) was detected in full scan mode using positive ion electrospray. A product ion

UV shows class; MS identifies species


233.1 Norm. 500 400 300 200 100 0 200 mAU 100 80 60 40 20 0 5 10 15 20 25 min 250 300 350 nm 199.2

Figure 36. Chromatogram of phenylurea herbicide with UV and MS spectra

Diuron

Monuron
291.2

Chloroxuron
100 200

300

400

500

* * *

32

at m/z 145 (Figure 37) generated by collisioninduced dissociation provided confirmation of carbaryl and was used for subsequent quantitative analysis.

Ion trap analysis was more sensitive than previous analysis using a single-quadrupole mass spectrometer operating in scanning mode and more sensitive than fluorescence detection. Ion trap LC/MS/MS also confirmed false positives in the HPLC fluorescence analysis caused by a coeluting compound. Based on the MS/MS spectrum of the coeluting compound, a possible structure was assigned as shown in Figure 38.

145 m/z
100

145

+H O O C N H CH3 H+ = 202 M

80
Carbaryl

Abundance

60

+ O

40

C2H4N O
+

214
NH C2H5

100
N H 86 H C O

20

80

145 O mw 213

0 100 125 150 175 200 225 250 275

Abundance

[M + H] + 202

60

+ N H H

m/z

86
40
O N H C2H4N O+

Figure 37. Full scan product ion spectrum generated by collision-induced dissociation of the [M + H]+ carbaryl ion at m/z 202

20

145

0 60 80 100 120 140 160 180 200 220

m/z

Figure 38. Full scan product ion spectrum of coeluting compound that produced false positives in HPLC fluorescence analysis

33

CE/MS Applications
Analysis of peptides using CE/MS/MS Capillary electrophoresis (CE) is a powerful complement to liquid chromatography. Different selectivity and higher chromatographic resolution are its biggest advantages when analyzing clean samples such as

synthetic peptides. When analyzing ppm-levels of analytes in complex matrices, minimal sample preparation and short analysis times enable high throughput.

CE/MS with an ion trap mass spectrometer is demonstrated using a standard peptide mix. The data-dependent acquisition capability of the ion trap triggers MS/MS on the most Intensity intense ion in each [x107] TIE, All MS scan. Figure 39 1.0 MS shows total ion electropherogram (TIE) and the 0.5 MS/MS base peak electrophero0.0 gram (BPE). Drops in 6 the TIE indicate where [x10 ] BPE 200-1350, All MS 3 2 MS/MS spectra were 4.0 4 acquired automatically. 1 2.0 The averaged mass 0.0 spectrum of peak 4 8.50 9.00 9.50 10.00 10.50 Time [min] (Figure 40) shows a singly charged molecule with m/z of 574.3. Figure 39. Total ion electropherogram (top) and base peak electropherogram This is confirmed (bottom) of a standard peptide mixture as Met-enkephalin (molecular weight 573.7) by the examination Intensity Intensity [x105] of the MS/MS mass 574.3 5 [x10 ] [M+H]+ spectrum. The MS/MS 4 4 574.3 575.2 spectrum shows all 576.2 239.1 0 of the b- and y-series 2 570 574 578 fragments within the 812.1 1147.2 493.1 scan range as well 0 as fragments from 200 400 600 800 1000 [m/z] other series.
Intensity [x104] 6 4 2 0 b2 221.0 232.9 225 b3 y2 278.1 297.1 262.1 275 a4 397.2 y3 354.2 b4 425.2

x2 323.1 325

x3 380.1 375

y4 411.1 [m/z]

Figure 40. Full scan mass spectrum (top) and MS/MS spectrum (bottom) of Met-enkephalin from peak 4 of the electropherogram

34

www.agilent.com/chem
Copyright 2001 Agilent Technologies Information, descriptions and specifications in this publication are subject to change without notice. All rights reserved. Reproduction, adaptation or translation without prior written permission is prohibited, except as allowed under the copyright laws. Printed in the U.S.A. February 15, 2001 5988-2045EN