You are on page 1of 58

The Immortality is yours take it---Aquiles

INSTRUMENTATION AND MEASUREMENTS IN THE HUMAN BODY1


Bayoán R. Ortiz, Xavier Piñero and Joel Blanco 2

Abstract---One of the major contributions of biomedical the absorption of light does not trigger action potentials but
engineering to the life sciences and clinical medicine has modulates the membrane potential of the cones [9].
been in biomedical instrumentation. Advances in this
field have resulted in the development of new types of The Schirmer strip test
biomedical instruments and the development of
numerous clinical approaches, such as electronic patient
monitoring, an important aspect of critical care In Schirmer test, we measure the quantity of tears that are
medicine, as well as a variety of devices to assist produced by the eye. If the tears are collected for some
individuals with disabilities. time, lets say 5 minutes or so, then one can determine
whether the amount produced is sufficient for maintaining
Key Words---Classification of Biomedical Instruments, eye health or not. If not much tears are produced, then one
Nervous System, Cardiovascular System, Respiratory have a tear deficient dry eye. If one produce enough tears,
System, Endocrine System, Measurement Techniques, but still have symptoms of ocular discomfort, then one may
Medical Measurement Constraints. have evaporative dry eye for example due to blepharitis or
Rosacea.
INTRODUCTION
In Schirmer test a 35 mm x 5 mm size filter paper strip is
used to measures the amount of tears that are produced over
Fluids compose over 80 percent of the human body. There a period of 5 minutes. The strip is placed at the junction of
are many kinds of instruments to measure fluids in the the middle and lateral thirds of the lower eye lid. The test is
body. NASA and other care agencies around the United done under ambient light. The patient is instructed to look
States helped in the design of instruments. Each measuring forward and to blink normally during the course of the test.
instrument has the characteristics of capacity, accuracy, A negative test (more than 10 mm wetting of the filter
error, precision, linearity, and durability, simplicity of paper in 5 minutes) means you produce normal quantity of
operation, output, calibration, and cost. The purpose of tears. Patients with dry eyes have wetting values of less
medical instrumentation is the measurement of than 5 mm in 5 minutes. An important limitation of
physiological variables: electrical activity of the heart, Schirmer test is that there may be considerable variability
respiratory airflow, blood pressure, etc. Measurement in the results of tests done at different times and by
techniques can be studied separately for each organ system, different doctors. So whereas this is perhaps the most
such as the cardiovascular, pulmonary, nervous, and common dry eye test performed, its main utility may really
endocrine systems. This approach isolates all important be in diagnosing patients with severe dry eyes. Sequential
measurements for specialists who need to know only about tests to follow the course of a patient with mildm dry eye
a specific area, but it results in considerable overlap of may not be of value. There is one point of some debate in
quantities sensed and principles of transduction [2]. Schirmer I test. When an anesthetic eyedrop is not used
then this test is thought to measure the basal + reflex tear
THE HUMAN EYE secretion. When an anesthetic eyedrop is used then this test
is thought to measure only the basic tear secretion.
The response of differences in absorption. cones is not all-
or-none. Light of a given wavelength (color), say 500 nm
(green), stimulates all three types of cones, but the green- There is compelling reason to believe that the tears
absorbing cones will be stimulated most strongly. Like rods measured by these two different methods may not
sufficiently differentiate between basic and reflex tear
production. Most clinicians perform this test after using
anesthetic eyedrops to numb the eye. However, The
National Eye Institute workshop on dry eyes recommended
1
This review article was prepared on July 10, 2004 for the not to use anesthetic eyedrops before performing this test.
course on Fluid Mechanics, INGE 4015. Course The cutoff value is similar whether or not anesthetic is
Instructor: Dr Megh R., Goyal, Professor in Biomedical used. To measure the reflex tear secretion Schirmer II test
Engineering, General Engineering Department, PO Box may be performed. Schirmer II test is performed by
5984, Mayaguez, Puerto Rico 00681-5984. For details irritating the nasal mucosa with a cotton-tipped applicator
contact: Email: m_goyal@ece.uprm.edu or visit at: prior to measuring tear production. Many clinicians regard
http://www.ece.uprm.edu/~m_goyal/home.htm the Schirmer test as unduly invasive and of little value for
mild to moderate dry eyes. Other less invasive methods to
2
Authors are in the alphabetical order. assess the adequacy of tear production have been
3
The numbers in the brackets refer to the refernce in the developed. The Phenol red thread test is one such test and
bibliography with cited page. is commercially available (zone-quick).

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-1
Figure 3. Non-contact tonometer [33].
Figure 1. Red thread test [16].
The Applanation Tonometer
A cotton thread impregnated with phenol red dye is used.
The pressure within an elastic-walled sphere or cylinder
Phenol red is pH sensitive and changes from yellow to red
such as the eye, an artery, or the pregnant uterus can be
when wetted by tears (Figure 1). The crimped end of a
measured from outside the walls by the use of an
70mm long thread is placed in the lower conjunctival
applanation tonometer (Figure 4). This may be a
fornix. After 15 seconds, the length of the color change on
mechanical device consisting of a flat ring in which there is
the thread - indicating the length of the thread wetted by the
a concentric disc maintained flush with the surface of the
tears -is measured in millimeters. Wetting lengths should
outer ring but connected to a force transducer. When a
normally be between 9mm and 20mm. Patients with dry
section of the wall is flattened by the device the pressure
eyes have wetting values of less than 9 mm.
required on the central disc to keep the wall flat is equal to
the pressure within the sphere or cylinder. The pressure is
The Non-Contact Tonometer calculated as the force applied divided by the area of the
inner disc. Large devices of this type are commonly used to
The non-contact tonometer is an applanation tonometer and measure the intra-uterine pressure during labor. This device
works on the principle of a time interval. Measuring the is sometimes called a tocodynamometer or guard ring
time it takes from the initial generation of the puff of air to tocograph, and the measuring instrument called a
where the cornea is exactly flattened (in milliseconds) to cardiotocograph, which also records heart rate. Smaller
the point where the timing device stops. It takes less time devices have been used for measuring pressures within
for the puff of air to flatten a soft eye than it does a hard arteries and also within the eye by direct application to the
eye [35]. cornea [25].

Figure 2. How non-contact tonometer works [35]. Figure 4. Applanation tonometer [40].

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-2
PanOptic Ophthalmoscope [27]

1. Makes it easy to enter small, undilated pupils

2. Provides a 5X larger view of the fundus than ever


before achieved with a standard ophthalmoscope
in an undilated eye

3. Enables a 25º field-of-view vs. the standard 5º


field-of-view

4. Increases magnification by 26% over a standard


scope

5. Greater working distance between practitioner


and patient improves comfort for all

6. Affordable new head is compatible with all


existing Welch Allyn 3.5V battery handles and
wall transformers [27].

The Biomicroscopy

Ultrasound biomicroscopy (UBM) can provide high-


resolution images in the living human eye; this resolution
Figure 5. Schiotz tonometer [25]. has been compared to microscopic resolution. This system
is particularly useful in studying the anterior segment and
Indentation (Schiotz) tonometer analyzing the different types of glaucoma. Anatomical
aspects, physiopathological processes and surgical results
This type of tonometer uses a plunger to indent the cornea. in glaucoma could be evaluated with UBM. This
After placing a drop of anesthesia on the eye to numb its exploration is also important in determining differential
surface, the doctor carefully places the instrument directly diagnosis, risk factors, and prognostic elements. In
on the cornea. The pressure within the eye is determined by addition, using UBM after filtering surgery helps detect
measuring how much the cornea is indented by a given early complications and understand poor surgical outcome.
weight. This test is less accurate than applanation
tonometry and is less commonly used today by THE HUMAN EAR
ophthalmologists and optometrists. However, other health
professionals (such as family practice doctors or urgent The human ear serves as an astounding transducer,
care doctors) may still commonly use this test [11]. converting sound energy to mechanical energy to a nerve
impulse which is transmitted to the brain (Figure 7).

Figure 6. PanOptic Ophthalmoscope [27]. Figure 7. Human ear [23].

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-3
The ear's ability to do this allows us to perceive the pitch of RESPIRATORY SYSTEM
sounds by detection of the wave's frequencies, the loudness
of sound by detection of the wave's amplitude and the The primary function of the respiratory system is to supply
timbre of the sound by the detection of the various the blood with oxygen in order for the blood to deliver
frequencies which make up a complex sound wave. oxygen to all parts of the body. The respiratory system does
this through breathing. When we breathe, we inhale oxygen
The ophthalmoscope and otoscope and exhale carbon dioxide. This exchange of gases is the
respiratory system's means of getting oxygen to the blood.
Due to the halogen illumination, this ophthalmoscope
provides a clean white light for optimum examination of The diaphragm's job is to help pump the carbon dioxide out
the eyeball. As with the otoscope, the on/off switch has of the lungs and pull the oxygen into the lungs. The
been cleverly designed into the stainless steel pocket clip to diaphragm is a sheet of muscles that lies across the bottom
provide easy operation with either hand. of the chest cavity. As the diaphragm contracts and relaxes,
breathing takes place. When the diaphragm contracts,
The lens wheel has a range of 0 to +20 and 0 to -20 oxygen is pulled into the lungs. When the diaphragm
dioptres. A full set of apertures are also provided - main relaxes, carbon dioxide is pumped out of the lungs [7].
beam, small beam, semi circle, green filter and a fixation
star. There is also a built in dust cover made of repeated
unsealed glass to prevent dirt entering the head. The glass
protection is made of silicon.

As with the otoscope the smooth finish of the overall unit


affords easy cleaning and maintenance [31].

Figure 7.: Ophthalmoscope (left) and otoscope (right) [36]. Figure 8. The electronic spirometer [38].

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-4
Figure 10. Spirostar [29].
Figure 9. The vitalograph [42].

The simple, non-invasive way to measure breath carbon


monoxide levels.

• Hand held, battery operated.


• Easy-to-read display.
• Reliable calibration
• Inexpensive, hygienic operation.
• Excellent for therapy.

Smoking cessation can be a success with the motivation


provided by the BreathCO carbon monoxide monitor. The
BreathCO provides visual proof to the smoker of the
dangerous CO levels in their lungs
Figure 11. SV-3 Spirometer [24].
Showing smokers the poison they are carrying in their
blood and damage they are doing to their bodies as a result The bronchoscope (Figure12)
of their habit is an excellent motivator.
It is an instrument for inspecting the interior of the
The Spirostar DX (Figure 10 and 11) tracheobronchial tree and carrying out endobronchial
diagnostic and therapeutic maneuvers, such as taking
specimens for culture and biopsy and removing foreign
New universal SpiroStar spirometer represents the state of
bodies [33].
the art in dynamic spirometry. It is compact weighting only
22 g. SpiroStar can be connected into a standard (RS-232C)
serial port of a computer. SpiroStar is accurate and precise
and it complies with the ATS spirometry standards (1994)
giving one the assurance of reliable results. Heuristic
system algorithm monitors the stability of the system and
guarantees repeatable and reliable session results.The
application software is Windows (95 / 98 / NT 4.0/ 2000 /
XP) compatible Spiro2000, which provides easy operation,
flexible parameter selection and comprehensive choice of
report formats. One can modify the reports to suite the
requirements. Session results can be transferred into
hospital information systems via local area network or
emailed to consulting physician fast and
faultlessly.Calibration database provides a tool for
professional quality assurance and management system.
SpiroStar includes an automatic temperature sensor to
monitor and correct variations in operating environment.
SpiroStar uses disposable Medikro Flow Transducer, which
provides for better patient safety by eliminating potential
contamination. It increases the system throughput and saves
operating costs. The instrument and the system are
designed to assist one in developing and maintaining a
better and safer lung function testing system for the
patients. Figure 12. Bronchoscope [40].

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-5
The Incentive Spirometer This exercise should help the patient [13].

Using the incentive spirometer after surgery keeps the 1. Increase in your inhaled lung volume.
lungs clear. The incentive spirometer also keeps the lungs 2. Improve getting rid of mucus or spit.
active when the patient is recovering from surgery, as if the 3. Avoid serious lung infection.
patient is at home performing daily activities [20].
The Body Plethysmograph
How to use the incentive spirometer (Figure 13):
Spirometry is the standard method for measuring most
1. Sit on the edge of the bed if possible, or sit up as relative lung volumes; however, it is incapable of providing
far as one can in bed. information about absolute volumes of air in the lung. Thus
2. Hold the incentive spirometer in an upright a different approach is required to measure residual
position. volume, functional residual capacity, and total
3. Place the mouthpiece in the mouth and seal the lungcapacity. Two of the most common methods of
lips tightly around it. obtaining information about these volumes are gasdilution
4. Breathe in slowly and as deeply as possible, tests and body plethysmography.
raising the yellow piston toward the top of the
column. The yellow coach indicator should be in In body plethysmography, the patient sits inside an airtight
the blue outlined area. box, inhales or exhales to a particular volume (usually
5. Hold the breath as long as possible (at least for 5 FRC), and then a shutter drops across their breathing tube.
seconds). Allow the piston to fall to the bottom of The subject makes respiratory efforts against the closed
the column. shutter (this looks, and feels, like panting), causing their
6. Rest for a few seconds and repeat Steps 1 to 5 at chest volume to expand and decompressing the air in their
least 10 times every hour when awake. lungs. The increase in their chest volume slightly reduces
7. Position the yellow indicator on the left side of the box volume (the non-person volume of the box and thus
the spirometer to show the best effort. Use the slightly increases the pressure in the box.
indicator as a goal to work toward during each
repetition.
8. After each set of ten deep breaths, practice Using the data from the plethysmography requires use of
coughing to be sure the lungs are clear. If the Boyles Law.
patient has an incision, support the incision when
coughing by placing a pillow firmly against it. P1V1 = P2V2 /1/
9. Once the patient is able to get out of bed, walk in
the hallway and cough well, one may stop using Where: P1 ans V1 are initial pressure and volume, and P2
the incentive spirometer unless otherwise and V2 are final pressure and volume, measured at constant
instructed by the health care provider [ 20]. temperature.

To compute the original volume of air in the lungs, we first


compute the change in volume of the chest.

We solve for the volume of the box during the respiratory


effort. The difference between this volume and the initial
volume of the box, is the change in volume of the box,
which is the same as the change in volume of the chest.
Armed with this piece of information, we use Boyle's Law
again, this time on the fixed amount of gas in the chest
before and at the end of a respiratoryeffort. We set the
initial volume of the chest (unknown) times the initial
pressure at the mouth (known), equal to the inspiratory
volume of the chest (the same unknown volume plus the
change in the volume of the chest, which we have just
computed) times the pressure at the mouth during the
inspiratory effort (known). Now we solve for the unknown
volume, which will be the original volume of gas present in
the lungs when the shutter was closed. As mentioned
before, the shutter is usually closed at the end of a normal
exhalation, or at FRC.

Body plethysmography is particularly appropriate for


Figure 13. The incentive spirometer [13]. patients who have air spaces within the lung that do not

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-6
communicate with the bronchial tree. In these individules, can flow past the cuff only when the arterial pressure
gas dilution methods ofmeasurement would give an exceeds the pressure in the cuff. Also, when the cuff is
erroneously low volume reading [ 12]. inflated to a pressure that only partially occludes the
brachial artery, turbulence is generated in the blood as it
The Oxymeter spurts through the tiny arterial opening during each systole.
The Korotkoff sounds that can be heard through a
stethoscope placed over the artery downstream from the
The pulse Oxymiter allows the continous measurements of cuff are generated by this turbulence [43].
arterial oxygen saturation (%SpO2) as well as the pulse rate
(beats/min.). This non invasive monitor uses a sensor
placed on a finger, toe, or an ear lobe [31]. Blood pressure varies between individuals and even during
the normal course of a day in response to emotion, exertion,
sleep, and other physical and mental changes. The average
normal blood pressure is about 120/80 mm Hg. Higher
THE CIRCULATORY SYSTEM blood pressures that are sustained over a long period of
time may indicate hypertension, a damaging circulatory
The Circulatory System or cardiovascular system, in condition. Lower blood pressures could signal shock from
humans, is the combined function of the heart, blood, and heart failure, dehydration, internal bleeding, or blood loss
blood vessels to transport oxygen and nutrients to organs [43].
and tissues throughout the body and carry away waste
products. Among its vital functions, the circulatory system Blood pressure measurement with a sphygmo manometer
increases the flow of blood to meet increased energy and a stethoscope can be obtained can be obtained by
demands during exercise and regulates body temperature. inflating the pressure cuff, on the upper arm, to a pressure
In addition, when foreign substances or organisms invade well above systolic pressure. No sounds can be heard
the body, the circulatory system swiftly conveys disease- through the stethoscope that is placed over the brachial
fighting elements of the immune system, such as white artery, which has been collapsed by the pressure of the cuff.
blood cells and antibodies to regions under attack. Also, in This pressure is then gradually reduced. Korotkoff sound
the case of injury or bleeding, the circulatory system sends can be heard because through the stethoscope because
clotting cells and proteins to the affected site, which pressure is below systolic pressure and small amounts of
quickly stop bleeding and promote healing [43]. blood spurts past the cuff. When this fist Korotkoff sound
is heard it is recorded as the systolic blood pressure and is
In order to maintain an adequate flow of blood to all parts indicated on the manometer. As the pressure in the cuff
of the body, a certain level of blood pressure is needed. continuous to drop, the Korotkoff sounds continue until the
Blood pressure, for instance, enables a person to rise cuff pressure is no longer sufficient to occlude the vessel
quickly from a horizontal position without blood pooling in during any part of the cycle. When the Korotkoff sounds
the legs, which would cause fainting from deprivation of disappear bellow this pressure, the diastolic pressure value
blood to the brain. Normal blood pressure is regulated by a is marked. This method is called the auscultatory method
number of factors: the contraction of the heart, the elasticity for locating the systolic and diastolic pressure value [43].
of arterial walls, blood volume, and resistance of blood
vessels to the passage of blood [43]. Automatic indirect method

Measurement of blood pressure A number of automatic and semiautomatic systems have


been developed. Most devices that utilize a pressure
Blood pressure is considered a good indicator of the status transducer connected to the sphygmo manometer cuff, a
of the cardiovascular system, and as one of the microphone placed beneath the cuff, and a standard
physiological variables that can be quite readily measured. physiological system on which cuff pressure and korotkoff
Blood pressure is measured by means of direct and indirect sounds are recorded. The procedure is essentially parallel to
methods [43]. the manual method [43].

Blood pressure is measured at two points: the high point at This instrument is actually semiautomatic because the
which the heart contracts to empty its blood into the recording must still be interpreted by the observer. Fully
circulation, called systole; and the low point at which the automated devices use some type of signal detecting circuit
heart relaxes to fill with blood returned by the circulation, to determine the occurrence of the first and last korotkof
called diastole. Blood pressure is measured in millimeters sounds and retain and display the cuff pressure reading for
(mm) of mercury by an instrument called a sphygmo these points, either electronically or with mercury
manometer consisting of an inflatable rubber cuff manometer that are cut off by solenoid valves. An example
connected to a pressure-detecting device with a dial. The of automatic blood pressure method is the programmed
cuff is wrapped around the upper arm and inflated by electrophyg momanometer PE- 300. This instrument is
squeezing a rubber bulb connected to it by a tube. The designed for use in combination with an occluding cuff,
sphygmo manometer works on the principle that when the microphone, or pulse transducer, and a recorder for the
cuff is placed on the upper arm and inflated, arterial blood automatic measurements of indirect systolic and diastolic

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-7
blood pressure. The device called electronic sphygmo Direct measurements
manometer incorporates a large gage for easy reading. It is
also called a “semiautomatic,” because the cuff has to be Methods for direct blood pressure measurements provide a
inflated manually. The microphone and light flash at continuous read out and recoding of the blood pressure
systolic pressure detect automatically the Korotkoff sounds waveform and are considerably more accurate than the
and stops at the diastolic value [43]. indirect methods. The direct measurement of blood
pressure is usually obtained by one of three methods:
Most electronic monitors using arterial cannulae with percutaneous insertion, catheterization (vessel cut down) or
electronic transducers and displaysprovide both diagnostic implantation of a transducer in a vessel or in the heart [43].
systolic/ diastolic waveform information and the added
opinion of a single value MAP (mean arterial pressure) Typically, for percutaneous insertion, a local anesthetic is
indication. A continuously decreasing or increasing MAP injected near the site of invasion. The vessel is occluded
can ultimately result in a hypertensive or hypontesive and a hollow needle is inserted at a slight angle toward the
crisis. A simple formula for calculating MAP: vessel. When the needle is in place, a catheter is fed
through the hollow needle, usually with some sort of a
guide. When the catheter is in place in the vessel, the
MAP= 1/3 (systolic- diastolic) + diastolic /2/ needle and the guide are withdrawn [43].

Mean arterial pressure is a weighted average of systolic and Catheterization


diastolic pressure. Between the diastolic low and the
systolic peak, MAP falls about one- third of the way. Catheterization has become the major diagnosis technique
for analyzing the heart and other components of the
The mean arterial pressure is determined by the cardiovascular system. Essentially, a catheter is a long tube
oscillometric method used in the Dinamap (figure 14). A that is introduced into the heart and major vessel by way of
solid- state pressure transducer located within the enclosure a superficial vein or artery. Measurements of blood
senses the pressure pulsations introduced within a cuff pressure with a catheter can be achieved in two ways. The
bladder. Pressure oscillations increase in amplitude and first is to introduce a sterile saline solution into the catheter
reach a maximum as the cuff pressure is decrease and so that the fluid pressure is transmitted to a transducer
passes through the pressure equal to the mean arterial blood outside the body. In the second method, the transducer is
pressure [43]. introduced into the catheter and pushed into the point at
which the pressure is to be measured [43].

Measurements of the heart sounds

The stethoscope is an instrument used for auscultation—


that is, to detect and study sounds arising within organs
such as the heart, lung, and stomach prior to treatment. It is
simply a device that carries sound energy from the chest of
the patient to the ear of the doctor via a column of air. The
stethoscope consists of a bell and diaphragm, or receiving
head, connected by an Y-joint and rubber tubing to two ear
pieces. Since the system is strictly acoustical, there is no
amplification of sound, except for any that might occur
through resonance and other acoustical characteristic. The
sounds may also be amplified electronically. With the
electronic stethoscope, high- fidelity equipment would be
able to reproduce the entire frequency range, of which is
missed by the regular stethoscope [43].

Instruments for graphically recording heart sounds have


been more successful. The phonocardiogram is a graphic
record of the heart sounds. The basic transducer is a
microphone having the necessary frequency response,
generally ranging from bellow 5 Hz to above 100 Hz. An
amplifier with similar response characteristics is required,
which may offer a selective low pass filter to allow the high
frequency cutoff to be adjusted for noise and other
considerations [43].

The presence of higher frequencies or murmurs in the


phonocardiogram indicates a possible heart disorder. A
Figure 14. Dinamap [43]. spectral analysis of heart sounds can provide a useful

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-8
diagnostic tool for discriminating between normal and
abnormal hearts. For this type of analysis it is required a
digital computer with a high- speed analog to- digital
conversion capability and some form of Fourier- transform
software.

Heart sounds are sometimes measured from other vantage


points rather than the chest or over the heart where
microphones and phonocardiograms are designed to be
placed. Special microphone transducers are placed at the
tips of the catheters to pick up heart sound from within the
chambers of the heart or from the major blood vessels near
the heart. The difference in acoustic paths makes these
heart- sound patterns appear somewhat different from the
usual phonocardiogrram patterns as might be expected. The
vibrocardiograph and apex cardiograph, use microphones Figure 15. Cardiac catheterization [43].
as transducers. Since the measurements involve the low
frequency vibrations of the heart against the chest wall, the The blood density increases as the capillaries fill with the
measurement is normally one of the displacements of the amount of light reaching the photocell. The result cause
force rather than the sound. The microphone must be a resistance changes in the photocell that can be measured on
good force transducer with suitable low- frequency a Wheatstone bridge and recorded. Because the photocell
coupling from the chest wall to the microphone element. device cannot be calibrated to reflect absolute or even
The microphone for the apex cardiogram must be coupled relative volumes, pulsations recorded in this manner are
to a point between the ribs. For this purpose, a soft rubber somewhat similar to those obtained by a true
or plastic cone attached to the element of the microphone plethysmograph [43].
gives good results. The volume changes in any part of the
The impedance plethysmograph is more reliable. In this
body are due to the pulsation of blood occurring with each
device, the volume changes in a segment of a limb or digit
heart beat. Such measurements are useful in the diagnosis
are reflected as impedance changes. These impedance
of arterial obstruction as well as for pulse-wave velocity
changes are due primarily to changes in the conductivity of
measurements. Instruments measuring volume changes or
the current path with each pulsation of the blood. The
providing outputs that can be related to them are called
measurements can be made using a two- electrode or a
plethysmographs. Such an instrument consists of a rigid
four- electrode system. The electrodes are either simple
cup or chamber placed over the limb or digit in which the
conductive strips of tape attached to the skin or conductive
volume changes are to be measured. The cup is tightly
bands wrapped around the limb or digit to be measured. In
sealed to the member to be measured so that any changes of
either case, the electrodes contact the skin through a
volume in the limb or digit reflect as pressure changes
suitableelectrolyte jelly or paste to form an electrode
inside the chamber. Either fluid or air can be used to fill the
interface and to remove the effect of skin resistance. In a
chamber.
two-electrode system, a constant current is forced through
the tissue between the two electrodes, and the resulting
Plethysmographs may be designed for constant pressure or
voltage changes are measured. In the four- electrode
constant volume within the chamber. In either case, some
system, the constant current is forced through two outer, or
form of pressure or displacement transducer must be
currents electrodes, and the voltage between the two inner,
including to respond to pressure changes within the
or measurement, electrodes is measured. The internal body
chamber and to provide a signal that can calibrate to
resistance between the electrodes forms the physiological
represent the volume of the limb or digit. Several devices,
divider. The advantage of the four-electrode system is a
called plethysmographs, actually measure some variable
much smaller amount of current through the measuring
related to volume rather than volume itself. One class of
electrodes, thus reducing the possibility of error due to
these “pseudo- plethysmographs” measures changes in
changes in electrode resistance [43].
diameter at certain cross section of a finger, toe, arm, leg,
or other segment of the body. Since volume is related to The reoencephalography is a special form of impedance
diameter, this type of devices is sufficiently accurate for plethysmograph, the measurement of impedance changes
many purposes [43]. between electrodes positioned on the scalp. This technique
provides information related to cerebral blood flow and is
The photoelectric plethysmograph operates on the principle sometimes used to detect circulatory differences between
that volume changes in a limb or digit result in changes in the two sides of the head. Such information might help in
the optical density through and just beneath the skin over a locating blockages in the internal carotid system, which
vascular region. A light source in a opaque chamber supplies blood to the brain [43].
illuminates a small area of the fingerprint or other region to
which the transducer is applied. Light scattered and The oculo pneumo plethysmograph measures every minute
transmitted through the capillaries of the region is picked volume changes that occur in the eye with each arterial
up by the photocell, which is shield from all other light. blood pulsation. A small eye cup is placed over the sclera
of each eye and is connected to a transducer positioned

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-9
over the patient’s head by a short section of flexible tubing. Transmission of this signal can be detected by the ECG
A vacuum, which can be varied from zero to –300 mm Hg, electrodes and graphically displayed as the ECG waveform.
is applied to hold the eye cups in place. Pulsations are The typical waveform is plotted and each major impulse
recorded on two channels of a three- channel pen recorder, identified by letters of the alphabet. The distance between
one for each eye. The third channel is used to record the the impulses is also important, and these are known as
vacuum. By periodically allowing the vacuum to build up segments. The P wave is associated with the initial impulse
to –300 mm Hg and deplete to zero, the instrument can also of the sinoatrial node. The QRS complex is a depiction of
be used as a recording suction ophtalmodynamometer an the ventricular contraction. The QRS signal can appear
instrument for measuring arterial blood pressure within the differently for a variety of reasons. The first downward
eye [43]. deflection is the Q wave, thefirst upward deflection is the R
wave, and the first negative deflection after the R wave is
Electrocardiography known as the S wave. Interestingly, even if the Q wave is
absent, the resulting waveform is still considered a QRS.
Electrocardiography (figure 16) is a medical procedure by The T wave represents relaxation of the ventricles. In some
which a physician obtains a tracing of the electrical activity instances, a U wave may be seen, signaling the final phase
of the heart. The rhythmic beating of the heart is of ventricular relaxation. The distances between these
maintained by an orderly series of discharges originating in waves represent the speed of signal conduction, and
the sinus node of the right atrium and proceeding through intervals and determined for which there are normal values
the atrioventricular node and the bundle of neuromuscular [43].
fibers to the ventricles. By attaching electrodes to various
parts of the body, a record of this current can be obtained. The use of 12 leads for an ECG allows the rhythm of the
This record is called an electrocardiogram, or ECG for heart to be examined through different planes. A complete
short. Prominent parts of the ECG are the P wave, a or 12-lead ECG uses electrodes attached to all four limbs
deflection caused by the current originating in the atrium; and six positions on the chest wall. By using these different
the QRS ( figure 17) complex, showing the passage of the attachment points, the ECG can examine the heart rhythm
electrical activity into the ventricles; and the T wave, as the from a variety of viewpoints, providing a more complete
ventricles reset themselves. The ECG is often helpful in assessment of cardiac rhythm Quite simply, one lead
showing the cause of an abnormal heart rhythm or an provides a one-dimensional view of heart rhythm, and
evolving heart attack [43]. multiple leads allow a three-dimensional depiction. Leads
are simply electrode pads consisting of a nonconductive
An ECG is the graphic display of the electrical impulses material, adhesive, conducting jelly, and metal connection.
that control muscular contraction and relaxation of the The individual electrodes are placed in the extremities
heart. Monitoring the ECG allows the nurse or doctor to (usually the wrists and ankles) and at six sites on the chest
determine the cause of abnormal cardiac rhythms and wall. The leads are divided into six limb leads and six
design the appropriate treatment. The ECG is frequently chest leads. The six limb leads from four electrodes are
referred to as an EKG, and one portion of the ECG is accomplished by varying the signal orientation between the
typically continuously monitored in the intensive care unit four extremities. The six chest leads result from the six
[43]. electrodes placed on the chest wall [43].
Electrical conduction in the heart is accomplished by a
deceptively simple conduction system. Under normal The six limb leads are known as leads I, II, III, aVR, aVL,
conditions, the sinoatrial node is the pacemaker of the and aVF. The first three leads I, II, and III are bipolar leads.
heart, controlling heart rate. The sinoatrial node is Bipolar refers to the fact that these leads compare the
controlled by the autonomic nervous system. Once the difference in voltage between any two of the limb leads.
sinoatrial node initiates the electrical impulse, the signal is Lead I result from designating the left arm electrode as a
carried through a system of conducting fibers known as the positive electrode and the right arm electrode as a negative
atrial conduction tracts. During passage of this signal, the electrode. Lead II results from designating the left leg
atria contract, filling the ventricles with blood. This portion electrode as positive and the right arm electrode as
of the cardiac cycle is the systole. The signal is delivered negative. Lead III results from designating the left arm as
to the atrioventricular node, which carries the signal from negative and the left leg as a positive electrode. The
the atria to the ventricles. The bundle of His lies just difference in voltage between the two electrodes is
beyond the atrioventricular node and is responsible for a compared to create the ECG signal. Leads aVR, aVL, and
slight delay before transmission of the signal to the aVF are known as augmented leads. In each of these cases,
ventricles. This delay is generally believed to last one limb electrode is designated as positive and all others
approximately 0.1 seconds and allows the ventricles to are designated as negative. The aVR, lead designates the
completely fill with blood before contraction. The signal is right arm as the positive electrode, the aVL lead designates
then passed through the left and right bundle branches to the left arm as the positive electrode, and the aVF lead
the Purkinje fibers and Purkinje network embedded in the designates the left leg as the positive electrode. The term
ventricular walls. This coordinated signal results in augmented lead refers to the doubling of the amplification
uniform contraction of the ventricles and movement of of the signal required to create the ECG signal in these
blood to the pulmonary and systemic circulation [43]. leads. Leads I, II, and III are bipolar leads, but the
augmented leads are unipolar leads [43].

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-10
Figure16. Electrocardiography [43].
The limb leads provide a view of the electrical conduction
of the heart in a frontal plane. That is, the limb leads can
detect electrical activity directed within this plane. The
plane is shown in the form of a circle with the patient’s
head at the top and the patient’s feet at the bottom. The
relationship between the positive and negative electrodes
for each lead (I, II, III, aVR, aVL, and aVF) can be
determined by tracing the angle from the designated
positive and negative sides. This angle is known as the
angle of orientation. As an example, the angle of
orientation for lead I is 0 (straight line from left to right arm
electrode). The angles of orientation of leads II and III are
60 and 90 degrees, respectively. The augmented leads are
somewhat more complicated because three of the leads are
negative. In this case, the average of the three negative
leads is used in comparison with the positive lead. This
provides an angle of orientation for aVR of –150; for aVL of
–30; and for aVF +90. By tracing the angle from one
electrode positive to another, you can see how the ECG Figure17. PQRST waves [43].
machine gets a unique and distinctive view of the electrical
activity of the heart [43].
The chest leads are unipolar leads designated V1’, V2’, “rhythm strips,” which typically display the signal from a
V3’, V4’, V5’, and V6’. These leads are frequently single lead, generally have “tic” marks at the top of the
referred to as the precordial leads and are placed across the paper, outside the grid representing 3 seconds. The vertical
chest in a horizontal plane. The chest leads detect electrical axis uses the same boxes, with the scale being the signal
moving anteriorly or posteriorly (front to back). As such, amplitude. Standard ECG recordings use 10-mm to
the chest leads view the heart through an angle of represent 1 millivolt. The determination of time and
orientation different from that of the limb leads. The voltage is important for detecting abnormalities in heart
positioning of the chest electrodes is important for rate and rhythm [43].
providing the proper use of the heart. When properly Pressure transducer
placed, V1 and V2 are above the right ventricle, V3 and V4
are over the intraventricular septum, and V5 and V6 are This device consists of a catheter, a small open flexible
called lateral leads as a consequence of placement and left tube, filled with a saline solution, and a pressure sensor.
lateral view of the heart. Early ECG technology required One end of the catheter is inserted into the lumen of an
that the operator move the electrode from each chest lead artery and the other end is connected to the pressure sensor.
position to the next to record the chest leads. Today, the The top part of the transducer, known as the dome, is a
six chest electrodes are manufactured in a continuous strip. closed chamber that is connected to the catheter. On one of
The relative position of the electrode is predetermined. The the walls of the chamber is a thin diaphragm that can be
proper positions of these require that the clinician deflected by the force resulting from the difference in
appreciate body size differences among patients and place pressure of the fluid in the chamber and the atmospheric
the electrodes as required [43]. pressure. Strain gauge displacement sensors measure the
deflection of the diaphragm. This deflection is related to the
ECG paper resembles graph paper, with grid like boxes pressure difference with the following equation:
created from horizontal and vertical lines. The lines are
different widths and intensity to allow differentiation of d = [3 *(1 - µ2 )*R4 (∆P)] / (16*E*t3) /3/
time (horizontal) and signal amplitude (vertical). The
individual boxes created by the thin lines are 1 mm2, and Where: d = Displacement of diaphragm center.
the heavy lines are 5 mm2. The horizontal axis of time is µ = Poisson’s ratio.
usually based on a paper speed of 25 mm/sec, although R = Diaphragm radius.
other speeds are commonly available. At this rate, each 1- t = Diaphragm thickness.
mm box represents 0.04 seconds and each 5-mm box E = Young’s modulus for the diaphragm.
represents 0.2 seconds. By simple extrapolation, five of the
∆P = Pressure difference across the diaphragm.
5-mm boxes are equivalent to 1 second. ECG papers, or

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-11
The diaphragm is connected to four wire strain gauges ultrasound beam are produced by a rapidly oscillating
located in the transducer. These wires measure the crystal, and are inaudible to humans. A device called a
displacement of the diaphragm. These strain gauges are transducer is used to transmit the sound waves and receive
arranged in such a way that when the pressure increases the echoes. The transducer must be in close contact with
two of the wires elongate while the other two decrease in the skin, and a jellylike substance is smeared on the skin to
length. They are connected as a Wheatstone bridge, which improve the transmission of sound. Ultrasound can be used
causes the output voltage to be proportional to the pressure to examine the arterial system, heart, pancreas, peritoneal
in the chamber [43]. cavity, urinary system, ovaries, venous system, brain, and
spinal cord. Fluid conducts the sound waves well, making
We have described this device in terms of hydrostatic ultrasound a useful technique for diagnosing cysts (which
pressures. The arterial blood pressure, however, is a time- are filled with liquid), examining fluid-filled structures
varying pressure that follows a complex pattern between such as the bladder or biliary system, and viewing the fetus
systolic and diastolic pressure over the course of time in the amniotic sac. Air, bone, and other calcified tissues
between two heartbeats. Accurate reproduction of this absorb nearly all the ultrasound beam, so this technique
waveform at the pressure transducer entails several cannot be used to examine the bones or lungs [43].
considerations. The fluid in the catheter and the pressure
transducer dome has a mass associated with it. The 1. Ultrasound properties: Ultrasound exists as a sequence
catheter, if it is flexible, and the dome and the diaphragm of alternate compressions and refractions of a suitable
have an associated compliance. Thus the fluid column and medium and is propagated through that medium at some
pressure transducer behave very much as a mass on a velocity. Its behavior also depends on the frequency of the
spring. The system is a resonant frequency, which is often sonic energy and the density and mechanical compliance of
sufficiently low to cause some distortion of the arterial the medium through which it travels. The propagation path
blood pressure waveform. The system can also generate for ultrasound into or through the body must not include
artifact in responding to sharp changes in pressure. For bone or any gaseous medium, such as air because almost all
example, a step function in pressure often produces ringing the energy will be reflected and practically none will
at the resonant frequency of the system, further distorting continue through the second medium. An airless contact is
the waveform. It is, therefore, important when using such a usually produced through use of aqueous gel or water bag
system to consider not only static behavior of the between the transducer and the skin consecutively to
transducer, but also the dynamic behavior of the applying ultrasound to the body. The density and other
transducer-catheter system [43]. properties of various materials, including several of
biological interest. The temperature and ultrasonic
A pressure sensor that does not require the fluid coupling frequency are given for most of the measurements. The
components eliminates problem associated with the velocity of sound propagation through a medium varies
catheter and pressure transducer dome by miniaturizing the with the density of the medium, elastic properties and
pressure sensor so that it can be located in the artery at the temperature. The characteristic impedance of a material is
site of the pressure measurement. Such a sensor, needless the product of its density and the velocity of sound through
to say, must be much smaller that the one presented before. it. Some energy is absorbed and the wave is attenuated a
This requires that the diaphragm also is much smaller; certain amount for each centimeter, as ultrasound travels
therefore, if unbounded wire strain gauges are used to through the material. The amount of attenuation is a
measure the diaphragm deflection, the sensitivity is function of both the frequency of ultrasound and
considerably lower that it would be for the external type of characteristics of the material. The attenuation constant, α,
pressure sensor. Thus, a miniature pressure transducer for is defined by the equation:
inter-arterial application must employ displacement sensor
with the highest possible sensitivity. The overall sensor is α = c*ƒ /4/
1.25 by 0.5 by 3.75 mm and occupies a very small volume.
Since the diaphragm is only a fraction of this size, its Where: α = Attenuation constant.
deflection will be very small, and this sensor can be c = Proportionality constant.
considered to be noncompliant. It will therefore have a very ƒ = Ultrasound frequency.
high resonant frequency that should well outside the
frequency of normal physiological signals from various In equation /4/ with a given amount of ultrasonic energy,
biological sources [43]. the attenuation increases with some power of the
frequency, which means that the higher the frequency, the
Ultrasound less the distance it can penetrate into the body.

Ultrasound is a medical diagnostic technique in which very 2. Ultrasonic imaging: The invasive imaging of internal
high frequency sound is directed into the desired body part. organs of the body is the most widely used applications of
The tissue interfaces reflect the sound, and a computer ultrasound in diagnostic medicine. The valuable
produces a photograph or a moving image on a television information can be provided without the necessity of
processes the resulting pattern of sound reflection. surgery or the use of harmful radiation. In general, the
Ultrasound can be used to examine many parts of the body, imaging system utilizes the pulsed ultrasound or pulsed
but its best-known application is the examination of the Doppler mode. The received information is displayed in
fetus during pregnancy. The sound waves used in the one of the following display modes after the amplification:

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-12
a. M- scan display: Here, the received pulse are used to Fd = (fr – f0 )= (2*f0*v) / c /6/
brighten the trace rather than control the vertical deflection.
Each transmitted pulse triggers the oscilloscope sweep. The Where: f0 = The initial frequency.
echoes appear as dots with brightness proportional to the f = The frequency reflected.
intensity of each echo, this because the brightness level is v = The velocity.
set below the visibility threshold. The transducer is held c = Constant.
stationary, thus, the movement of the dots along the sweep Fd = Dopler frequency.
represent movement of the received targets.
Doppler equipment: The simplest Doppler instruments
b. B- scan display: It presents a two dimensional image of use continuous wave and also are the least expensive. In
a stationary organ or body structure. His transducer is this, the transmitter continuously excites the ultrasonic
moved with respect to the body while the vertical transducer with a sinusoidal electrical signal of frequency.
deflection of the oscilloscope or movement of the chart The transducer converts this electrical energy into
paper is made to correspond to the movement of the ultrasonic energy, which propagates into the medium. Echo
transducer. signals resulting from reflection and scattering return to the
c. A- scan display: The simplest form of display. Here, transducer creating an electrical signal that is applied to the
each transmitted pulse triggers the sweep of an receiver amplifier. Within the amplifier is boosted in
oscilloscope. The vertical deflections on the trace are that strength and applied to a demodulator and low-pass filter,
pulse and the returning echoes displayed. In some cases the with the result that only l0 w-frequency Doppler shift
transducer is kept stationary so that any movement of signals emerge in the output [43].
echoes along the transducer will be the result of moving
targets. The display of an echoencephalogram is an Pulsed Doppler
example of an A- scan [43].
Pulsed Doppler provides for discrimination of Doppler
Doppler effect signal from different depths, allowing for the detection of
Doppler is the apparent variation in frequency of any moving interfaces and scatters only from within a well-
emitted wave, such as a wave of light or sound, as the defined sample of volume. The sample volume can be
source of the wave approaches or moves away, relative to positioned anywhere along the axis of the ultrasound beam.
an observer. The effect takes its name from the Austrian The Pulsed Doppler consists of: master oscillator, gated
physicist Christian Johann Doppler, who first stated the transmitter, receiver, demodulator and transducer. In such
physical principle in 1842. Doppler's principle explains devices, the ultrasonic transducer is exited with a short-
why, if a source of sound of a constant pitch is moving duration burst from a gated transmitter. Scattered and
toward an observer, the sound seems higher in pitch, reflected echo signals are deflected by the same transducer,
whereas if the source is moving away it seems lower. An amplified in the receiver, and applied in the demodulator
observer listening to the whistle of an express train from a [43].
station platform or another train can hear this change in
pitch. The lines in the spectrum of a luminous body such as The Doppler signal is produced by a pulsed Doppler
a star are similarly shifted toward the violet if the distance instrument is generated from the pulse-to-pulse changes in
between the star and the earth is decreasing and toward the phase of the echo signal returning from a target that is
red if the distance is increasing. By measuring this shift, the either receding from or moving toward the ultrasonic
relative motion of the earth and the star can be calculated. transducer. A transmit pulse is emitted by the transducer,
Doppler ultrasound is used to measure the flow of a liquid, and subsequent echoes are detected from reflectors in the
for example blood flow. The amount of frequency shift can medium as a single pulse echo sequence [43].
be expressed in the formula:
Duplex ultrasound
f = (2*V)/ 8 /5/
A pulsed echo scanner and a Doppler instrument provide
Where: f = Shift in frequency of the reflected wave. complementary information so that the scanner can best
V= Velocity of the interface. outline anatomic structures, whereas a Doppler instrument
yields information regarding flow and movement patterns.
8= Wavelength of the transmitter ultrasound.
Duplex ultrasound instrument are real time B-mode scanner
with built-in Doppler capabilities.
Ultrasonic Doppler equipment can be employed for
detecting the motion of relatively large structures, such as
In typical applications, the pulse echo B-mode image
heart valves and fetal heart walls. This equipment is also
obtained with a duplex scanner is used to localize areas
commonly used for detecting and evaluating the
where flow is to be examined using Doppler. Many duplex
characteristics of blood flow in arteries and veins. The
instruments allow the operator to indicate the direction of
sound source or ultrasonic transducer is placed in contact
blood flow with respect to the ultrasound beam. The angle
with the external skin surface and the ultrasound beam is
formed by the ultrasound beam and the flow vector must be
directed toward the vessel of interest. The sources of echo
known to estimate flow velocity from the frequency of the
signals are the red blood cells following in the vessel. The
Doppler signal [43].
Doppler shift is given by the equation:

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-13
Ultrasonic methods are used in cardiology, in eye analysis, Ultrasonic devices
brain studies, obstetric and gynecology studies. In
Most of the devices in the ultrasound techniques employ
cardiology they study ultrasound examination of the heart
the same principles. These principles can be Doppler or
known as echocardiography. Echocardiography is used to
Duplex. Some of those devices are:
study congenital heart diseases, coronary artery diseases,
tumors of the heart and other cardiac disorders. They use a ACUSON 128XP/10 95’: Uses Doppler to detect the pulse
device that do a studied called echocardiogram, it is the of the patient in the required area. Also use Duplex to
output obtained by a machine that use essentially duplex to obtain an image of this area. With a colored scale the
measure the blood flow, the direction of the flow, the technician can determine the direction of the blood flow,
volume of the flow, and its velocity. the speed and his volume.
ADVANCED ACUSON ASPEN: Uses same principle as
To determine the direction of the flow they colored a
ACUSON 128XP, but has more versality. Here, all the
picture in a monitor device according with the reflection of
studies (images or videos) can be stored in a memory up to
the ultrasound wave and by a legend they know if the blood
one month [43].
flow is normal or not (see acuson output). Also, they can
use a blind Doppler ultrasound, which is printed in a graph Echocardiogram
as a function of the heart beating from where they can
analyze the length of the wave, the amplitude and the An echocardiogram uses ultrasound, to diagnose a variety
rhythm, to record the heart beating and by they knowledge of heart problems. To perform the test a transducer is held
determine the patient condition as hypertension or any against the chest. The transducer sends ultrasound waves
other condition [43]. that reflect (echo) off the various parts of the heart.

Echocardiology has been extremely useful in diagnosis A computer processes the information coming from the
many cardiac abnormalities, among them are calcific aortic transducer to construct an image of the heart. The image is
stenosis, pulmonary valve stenosis, mitral valve stenosis, displayed on a television screen, and it can also be recorded
left atrial myxoma, and other heart diseases. In selecting a on videotape or printed on paper [43].
transducer for an echocardiography, following factors
should be considered for optimum results: The echocardiogram study usually combines three
1. The type of tissue to be encountered. different techniques. M-mode echo, produces an image that
2. The anatomic area involved. looks more like a tracing than an actual heart. The M-mode
3. Physical size of the patient. echo is especially useful for measuring the exact size of the
4. The type of investigation to be performed. heart chambers.
5. The depth of the structure to be studied.
The two dimensional (2D) echo shows the actual shape and
It is possible to distinguish between different soft tissues motion of the different heart structures. In a way, the
and is possible to measure the motion of structures of the images represent "slices" of the heart in motion [43].
heart with ultrasound techniques. There is no interference
by echoes from other body structures, since the lungs, Doppler echo allows doctors to evaluate the flow of blood
which are literally air bags, surround the heart. through the heart. The signals representing the flow of
blood are displayed as a series of black and white tracings
The position and movements of each interface of the heart or as color images on the television screen [43].
such as the atrial and ventricular walls, the septum, and
various valves, can be measured by the reflected The echocardiogram provides following information about
ultrasound. The echoes from walls and valves are the heart, such as:
predictable since the components of the heart move in a 1. The heart chambers and thickness of the heart
known manner [43]. muscle.
Echocardiography is also used in the detection of fluids. 2. Whether the heart is pumping at full strength or is
When the pericardium, is inflamed, which is known as weakened. It can also help determine whether the
pericarditis, there is sometimes an escape of fluid. The various parts of the heart pump equally.
echocardiogram can detect the presence of this fluid [43]. 3. The shape and motion of the heart valves. It can
help determine if a valve is narrowed or leaking
To prevent blurring of the image due to heart movements and show how severe the problem is.
fast scanning speeds are required. Many simultaneous
recordings can be taken in different positions to give a 4. Detect the presence of fluid around the heart,
thorough dynamic analysis, as to give cross-sectional blood clots or masses inside the heart, and
images at various points along the length of the heart. abnormal holes between heart chambers.
5. An echocardiogram is sometimes performed
The medical profession believes that the potential of this during or immediately after an exercise test. It
diagnostic method will continue to increase as the helps doctors determine how well the heart
techniques and interpretations continue to advance [43]. pumps when it is made to work harder.

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-14
DIALYSIS placed in the great vein of the neck. This catheter can be
used for the external dialysis during several weeks up to
Dialysis is a process by which free blood toxins and fluid months [43].
excess are removed from the blood in the kidney. Two
main methods are used in dialysis: Hemodialysis and The catheters are covered with clean and dry bandages. The
Peritoneal dialysis [43]. most common permanent entrances are fistulas and grafts.
A fistula is formed under the skin sewing a vein to an
The hemodialysis uses a filter to remove toxins and fluids artery. These increase the amount of blood that flows
from the blood. The filter known as dialyzer acts as an through the vein and causes that the vein enlarges [43].
artificial kidney. The amount of blood that circulates at any
time around the artificial kidney is smaller than half liter. The process by which the fistula acquires the strength and
The dialyzer has two compartments by a semi permeable the size sufficient to allow the insertion of the needle is
membrane. Only particles of determined size can cross denominated as mature. Normally this process entails of
through this membrane. for up to twelve weeks. because the cross- sectional area at
the place of the velocity measurement is well defined with
A solution called dialysate circulates to one side of a either type of transducer.
membrane and the blood flows through the other side.
Dialysate is composed of water, dextrose (sugar) and This technique requires that a transducer surrounds an
elements similar to the normal fluids of the body without excised blood vessel, these are mainly used during surgery.
toxins [43]. The indicator dilution methods are the only methods of
blood flow measurement that really measure the blood flow
In the diffusion process across a semi permeable and not the blood velocity [43].
membrane, particles of high concentration tends to move to
those of lower concentration until the two solutions reach A graft is formed under the skin connecting an artery and a
equal concentration. In the dialysis, fresh dialysate flows vein with synthetically material. After surgery intervention,
through the dializer and the waste of the blood turns one may need three to four weeks for the swelling sensation
towards it until a vast proportion of waste products have of the graft to disappear. This graft can be used during the
been evacuated from the blood. dialysis [43].

The osmosis is similar to the diffusion, but instead of Peritoneal dialysis


evacuating waste products, evacuates the excess in blood.
The ultra filtration is the process by which pressure is The peritoneal dialysis is based on the same process of
applied in order to eliminate fluid of the blood. The filtration as in the hemodialisys. But instead of using an
pressure on the side of dialysated is applied to extract the artificial kidney as filter, it uses peritoneal membrane.
fluids toward outside.
The peritoneal membrane, also call peritoneum, is a thin
The dialysis personal can regulate the amount of fluids to membrane that covers the abdominal cavity [43].
be extracted during treatment of the dialysis. After a few
hours of dialysis, the blood is relatively purified and the
excessive fluids and waste are removed [43].

Entrance and care: Hemodialysis

In order to make possible that the blood flows from the


body to the dializer and vice versa an ample opening it is
needed to allow a considerable flow of blood. In order to
create this opening, it is necessary to perform a small
surgery, an entrance to the channels of the patient’s blood,
opening which is usually made in an arm or leg [43].

There are several types of entrances: temporally and


permanent. Temporally entrances include the following
types of fistulas (hollow flexible tubes that are partially out
of the body): the intraclavicula, the Femur, and the internal
Jugular. The intraclavicula is a temporally catheter placed
in the intraclavicular vein, exactly underneath the bone of
the neck. It shouldn’t be used during a period of time that
surpasses two weeks [43].

In the Femur the catheter is placed temporally in the great


vein in the ingle. It is only used during the stay of the
patient in the hospital. In the internal Jugular the catheter is Figure 18. Dialysis machine [43].

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-15
In order to obtain the entrance in the peritoneum, a flexible The orifice of the probe must fit tightly around the vessel
sounding or tube is surgically implanted in the low for a proper orientation. The probes can be implanted for
abdomen. It measures 30.5 cm in length but only 10 or 12 chronic use or during surgery, depending on the model of
cm of this is outside the body. the probe [43].

The sounding in an operating theater with the help of the The canula type transducer is a plastic canula around which
local anesthesia is inserted frequently. During the dialysis the magnet is arranged for blood flow. For this transducer it
the abdominal cavity with a solution, dialisate fills. is required that the blood vessel be cut and its ends slipped
over the canula and secured with a suture.
Since the peritoneum is rich in tiny glasses of blood, it
continuously provides a provision with blood that is filtered A transducer very similar to this one, extracorporeal blood
by osmosis and diffusion. The flowed excess of toxins in flow probe is also used to measure blood flow in
the blood happens to the dialisate one, which is evacuated extracorporeal devices, such as dialyzers [43].
and replaced periodically.
Magnetic blood flow transducers are also manufactured as
An interchange is the process to evacuate dialisate of the catheter tip transducers. The normal transducer design is
abdomen and to introduce dialisate fresh in the abdomen. essentially turned “inside out,” with the electromagnetic
The time space is known with the name of time of stay that being located inside the catheter, which has the electrodes
the dialysis solution remains in the peritoneal cavity during at the outside.
the peritoneal dialysis (Figure 17).
Since the cross section of the blood vessel at the place of
Measurements of blood flow and cardiac output measurement is not defined, the catheter transducer cannot
be calibrated in flow units [43].
The variation in blood supply may be due to of various
diseases. It would be of great help in diagnosing such Indicator dilution
diseases by measuring the blood flow in the vessel which
supplies the blood to a particular organ. The blood flow, is Measurement of blood temperature and injection of cool
a rather elusive variable that cannot be measured easily saline are performed with a specially designed single
[43]. catheter. The volume injected of cold saline is normally 10
ml, and the circulating blood is rewarmed rapidly and no
The rate of flow of a liquid or gas in a pipe is expressed as measurable recirculation occurs.
the volume of the substance that passes through the vessel
in a given unit of time. The device used for this technique is called the Swan- Ganz
catheter, which contains four separate lumens.
Usually, flow rates are expressed in liters per minute or
millimeters per minute. Almost all blood flow meters To inject the cooled saline, is used one of the lumens is
employ following physical principles: used. The second lumen contains two thin wires whith lead
to a tiny electrical temperature sensor close to the top of the
1. Electromagnetic induction. catheter.
2. Indicator dilution.
The third lumen ends at the catheter tip and can be used to
Magnetic blood flow meter measure the blood pressure. To inflate a small rubber
balloon at the tip of the catheter, the fourth lumen of the
Magnetic blood flow meters measure the mean velocity of catheter itself is used. The balloon is inflated and the
the blood stream. These transducers can be calibrated returning venous blood carries the catheter until its tip is
directly in units of flow column of conductive fluid that positioned in the pulmonary artery once the catheter has
flows through a tube located in the magnetic field the same been inserted into the vein [43].
principles applies.
The thermistor, which is an electronic temperature- sensing
A permanent magnet or electromagnet positioned around device, is connected into a bridge circuit for measuring and
the blood vessel generates a magnetic field perpendicular to recording blood temperature during injection. The analog
the direction of the blood flow. computer that consists of electronic integrator and
necessary controls makes possible the reading of the
The voltage induced in the moving blood column is cardiac output directly [43].
measured with stationary electrodes located on opposite
sides of the blood vessel and perpendicular to the direction Noninvasive blood glucose monitoring
of the magnetic field [43].
Noninvasive blood glucose monitoring involves either
The slip on or C type is one of the most commonly used folowing promising technologies:
types of implantable magnetic blood flow probes.
Squeezing an excised blood vessel together and slipping it 1. Near Infrared light (NIR) spectroscopy
through the slot of the probe, uses the C type. 2. Far Infrared Radiation (FIR) spectroscopy

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-16
3. Radio Wave Impedance With spectroscopy, light absorption at selected wavelengths
4. Optical Rotation of Polarized Light is analyzed for each blood glucose level. A polynomial
5. Fluid extraction from skin formula is generated converting the sum of the relative
6. Interstitial fluid harvesting. contributions of absorption at the selected wavelengths to
the blood glucose concentration. This technology is used in
Near infrared radiation (NIR) spectroscopy oximetry to measure the oxygen saturation of blood [43].

The term NIR refers to the use of an external light source The major problem with using NIR spectroscopy for blood
with wavelengths in the infrared spectrum near the glucose monitoring requires frequent recalibration under
wavelengths of visible light. following situations :

An NIR source can pass through or be reflected by a body 1. Use of medications that absorb NIR.
part. Glucose and other body constituents absorb a small 2. Alterations in the blood hemoglobin level, body
amount of the light at each wavelength [43]. temperature, or state of hydration.

Spectroscopy, an established technology used to measure Far infrared radiation (FIR) spectroscopy
energy containing many wavelengths, detects the amount of
NIR absorbed at each wavelength by comparing a reference Thee FIR spectroscopy measures absorption of far infrared
radiation contained in natural thermal emissions or body
heat. FIR it does not require an external energy source
[43].

The term thermal emissions refers to deep layers of the


human body emitting thermal radiation or body heat with
wavelengths in the far infrared radiation spectrum far from
the wavelengths of visible light. When FIR passes out of
the body, glucose in the blood absorbs part of the radiation
[43].

Thermal energy absorption of FIR by blood glucose can be


spectroscopically determined by comparing measured and
predicted amounts of thermal energy at the skin surface.
The predicted amount of thermal energy radiated can be
calculated by the Planck distribution function.
Simultaneous measurement of thermal energy absorption
outside the FIR "glucose band" determines the reference
intensity that is a necessary variable for calculating the
blood glucose concentration. The percentage of thermal
energy absorption can be arithmetically converted to a
blood glucose concentration. This technology is used in
tympanic thermometry to measure body temperature [43].

FIR spectroscopy for blood glucose monitoring has two


problems: First, the signal size of human thermal emissions
is very small. Second, the prototype device incorporates
cryogenically cooled infrared detectors. Replenishment of
the cryogenic fluid, currently liquid nitrogen, is
inconvenient [43].

Radio wave impedance

Radio wave impedance technology for blood glucose


monitoring also faces two problems: first, impedance is
also affected by factors other than glucose, which must be
accounted for to determine the relationship between
impedance and blood glucose concentration. These factors
Figure 19. Peritoneal Dialysis [43]. include the concentration of electrolytes in the blood, finger
width, and body temperature. Second, an inexpensive
beam with the detection beam that has passed through or is disposable finger clip may be necessary to conduct the
reflected by the body. radio waves [43].

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-17
called the GlucoWatch, will extract fluid from the skin
using reverse iontophoresis. A blood glucose level could be
reported as often as three times per hour. A programmable
alarm will sound if high or low panic values are exceeded
[43].

Interstitial fluid harvesting

This technology involves transcutaneous harvesting and


measurement of interstitial fluid from skin.

While the aforementioned technologies produce neither


skin trauma nor pain, transcutaneous harvesting of
interstitial fluid can be accomplished with nearly no skin
trauma and minimal sensation. Therefore, this technology
is classified not as "noninvasive," but rather as "nearly
noninvasive [43].

Transcutaneous harvesting of interstitial fluid produces no


Figure 20. Sensitiving glucose sensing [43] significant breaks in the skin surface, in contrast to
"minimally invasive" technology, which involves inserting
Optical rotation of polarized light an indwelling subcutaneous glucose sensor. Transcutaneous
interstitial fluid harvesting technology is so similar to the
The polarimetry is a optical rotation of polarized light. noninvasive technologies that this "nearly noninvasive"
When polarized light passes through a fluid that contains technology is included in this discussion of noninvasive
glucose, the plane of polarization rotates proportionate to monitoring [43].
the glucose concentration. A beam of infrared polarized
light can be used to pass through a body compartment, and Collecting interstitial fluid is less inconvenient than
the amount of optical rotation can then be measured [43]. collecting blood in terms of:

In polarimetry, a beam splitter divides a polarized light 1. Pain


beam into a reference beam and a detection beam, which 2. Skin trauma
passes through the body. The beams are then compared to 3. Site restriction (sampling is not limited to the
determine the amount of phase shift produced by passage fingertips), and risk of contamination
through the body. A blood glucose level is calculated by
applying a conversion factor to the phase shift. This Interstitial fluid harvesting involves extraction of fluid from
technology is used industrially to measure the the skin includes the following methods:
concentration of sugar in foods and dextrose in intravenous
solutions [43]. 1. Penetration with a 29-gauge needle
2. Disruption with chemicals, and
Fluid extraction from skin 3. Microporation with a laser.

This technology extracts and measures tissue fluid from The concentration of interstitial fluid glucose has been
skin. Also known as reverse iontophoresis, this method is shown to be equivalent to that of serum glucose when the
accurate and produces multiple measurements over a 24- serum glucose level is stable. Recent studies have also
hour period. A device using this technology will measure demonstrated such equivalence when the serum glucose
trends in blood glucose concentrations and will be level is rapidly rising or falling. The finding that dermal
programmed to control an insulin delivery system that will interstitial glucose is an indicator of ambient glycemia
create an artificial pancreas. Fluid extraction from skin is helps validate the proposed substitution of interstitial fluid
the only noninvasive blood glucose monitoring technology for blood as an accurate measurement of diabetic control
capable of measuring blood glucose levels continuously [43].
without patient effort [43].
HUMAN REPRODUCTIVE SYSTEM
Reverse iontophoresis involves application of an electrical
current to the skin followed by extraction of fluid. The Ultrasound
current pulls out salt, which carries water, which in turn
carries glucose. The glucose concentration of this extracted Ultrasound is a scan used to study internal body structures.
fluid can be measured and is proportionate to that of blood. It works by sending out high frequency sound waves,
Iontophoresis, from which reverse iontophoresis derives, is directed at the internal body part being examined,
an effective drug delivery technology [43]. andrecording the reflected sound or 'echoes' to create an
A noninvasive blood glucose monitoring system that will image.
resemble a wristwatch is being developed. This system,

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-18
HUMAN DIGESTIVE SYSTEM

Digestion includes mechanical and chemical processes. The


mechanical processes include chewing to reduce food to
small particles, the churning action of the stomach and
intestinal peristaltic action. These forces move the food
through the digestive tract and mix it with various
secretions. Three chemical reactions take place in this
process: conversion of carbohydrates into such simple
sugars as glucose, breaking down of protein into such
amino acids as alanine, and conversion of fats into fatty
acid and glycerol. Specific enzymes accomplish these
processes [9].

When food is chewed and swallowed the six salivary


glands produce secretions that are mixed with the food. The
saliva breaks down starches into dextrin and maltose,
dissolves solid food to make it susceptible to the action of
later intestinal secretions, stimulates secretion of digestive
enzymes, and lubricates the mouth and esophagus for the
Figure 21. Sperm Count Test [30]. passage of solids [9].

The sound waves are emitted from a small, vibrating crystal The digestive system is responsible for the digestion,
in a hand-held scanner. The high frequency of the sound absorption, and assimilation of fluids, macro- and micro-
means it can't be heard by the human ear. An ultrasound nutrients (vitamins and minerals) and the elimination of
scan doesn't produce high quality images, but it is non- waste from the gastrointestinal tract. Food and drink that
invasive and doesn't involve radiation. It is commonly used we ingest is broken down by our digestive system into
to study the abdominal organs, pelvic organs and breast smaller simple particles (molecules) before it is absorbed
tissue. Other names for ultrasound scan include sonogram by the small intestine and transported into the blood stream
and ultrasonography [6]. that carries the nutrients, vitamins, and minerals to cells
throughout the body. In the cells the molecules provide
Pregnancy scan is used to check for fetal abnormalities energy and nourishment to the body [9].
(such as spinal bifida), check the age and sex of the fetus,
and monitor fetal growth and development. An ultrasound Colonoscopy
scan during pregnancy may need to be performed every
three months (each trimester), according to the doctor's The term "colonoscopy" means looking inside the colon. It
recommendations. is a procedure performed by a gastroenterologist.The
colonoscopy (figure 21) is used to look inside the colon.It
Sperm count is a long, thin, flexible tube with a tiny video camera and a
light on the end. By adjusting the various controls on the
It is number of sperms in an ejaculate. The sperm count is a colonoscope, the gastroenterologist can carefully guide the
indicator of male fertility. The test can look at many instrument in any direction to look at the inside of the
aspects of sperm production and release. In 1 in 20 samples colon. The high quality picture from the colonoscope is
no sperm at all are seen. This is due to either obstruction or shown on a TV monitor, and gives a clear, detailed view.
Colonoscopy is more precise than an X-ray. This procedure
a failure of production. Although, in theory, only 1 sperm is also allows other instruments to be passed through the
needed to conceive, there is a direct relationship between colonoscope. These may be used to remove a suspicious-
the number of sperm and the chances of pregnancy. The looking growth or to take a biopsy.A shorter version of the
cycle of sperm production in the testis takes about 70 days colonoscope is called a sigmoidoscope, which is used to
so the test shouldn't be repeated too close as it takes a while screen the lower part of the large bowel only [9].
for the cycle to recover [6].
Colonoscopy can also check colon cancer and treat colon
Rigid Scan polyps-abnormal growths on the inside lining of the
intestine. The colonoscopy process includes:
The Rigid Scan. is a pressure device that consists of a
collector for data and two rings. The patient should put the 1. The colonocscope is inserted in the anus and
rings in the upper and lower part of his penis. The scan will gently guided to the cecum, the first part of your
read the blood pressure of the patient during the night and digestive tract.
the ring will contract or expand depending on the penis 2. Upon reaching the colon, the specialist can guide
activity during the night. This helps the specialist to the colonoscope to view the entire area.
determine: If the circulation in the penis is a normal one or 3. Upon finding something suspicious, the specialist
if the patient is impotent [6]. can take a biopsy.

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-19
Endoscopic Retrograde Cholangio Pancreatography
(ERCP)

Retrograde refers to the direction in which the endoscope is


used to inject a liquid enabling X-rays to be taken of the
parts of the GI tract called the bile duct system and
pancreas. Cholangio refers to the bile duct system and
pancreas. The process of taking these X-rays is known as
cholangiopancreatography. ERCP may be useful in
diagnosing and treating problems causing jaundice (a
yellowing of the whites of the eyes) or pain in the
abdomen.

Other special exams that take pictures using X-rays or


sound waves may provide complement important
information from ERCP. The ERCP process includes:

1. The gastroeneterologist may determine that a


stone is blocking a common duct.
1. 2. An endoscope will be lowered down the
esophagus, through the stomach, and into the
duodenum. A small tube will be threaded down
into the duct.
Figure 22. Colonoscopy [43]. 2. A dye will be injected backwards through the
ducts, allowing X-rays to be taken.
Upper GI Endoscopy 3. The specialist will be able to remove the stone
that is blocking the duct with the endoscope.
The term "endoscopy" refers to a special technique for
looking inside a part of the body. "Upper GI" is the portion
of the gastrointestinal tract, the digestive system, which
includes the esophagus, the swallowing tube leading to the
stomach [9].

Upper GI endoscopy (figure 24) is a procedure performed


by a gastroenterologist, who uses the endoscope to
diagnose and, in some cases, treat problems of the upper
digestive system. The endoscope (figure 54) is a long, thin,
flexible tube with a tiny video camera and light on the end.
By adjusting the various controls on the endoscope, the
gastroenterologist can safely guide the instrument to
carefully examine the inside lining of the upper digestive
system.The high quality picture from the endoscope is
shown on a TV monitor; it gives a clear, detailed view. In 1 2
many cases, upper GI endoscopy is a more precise
examination than X-ray studies.

Upper GI endoscopy can be helpful in the evaluation or


diagnosis of various problems: difficulty in swallowing,
pain in the stomach or abdomen, and bleeding, ulcers, and
tumors.

1. The endoscopy process

1. The endoscopy is inserted in the mouth and


gently edged down the esophagus.
2. The endoscope will be inserted until it reaches
the stomach.
3. Once in the stomach, the specialist will look 3 4
closely for any problem areas. Figure 23. Sequence of a colonoscopy [43].
4. If anything suspicious is found, the doctor will
take a sample for biopsy.

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-20
DIGESTIVE SYSTEM TESTS

Cholangiography (Intravenous)

Intravenous cholangiography is a contrast x-ray used to


detect gallstones, obstructions, or other abnormalities of the
gallbladder and bile ducts. The same procedure may be
done using a CT scan rather than conventional x-ray [43].

The patient will have a high-fat meal the evening before the
test, followed by nothing but water or fat-free liquids. A
laxative may be given, since a full bowel can create
shadows on the x-ray [43].

The test usually lasts about 2 to 2 1/2 hours, if the bile


ducts are not obstructed. If they are, it may take 4 to 24
hours for the contrast medium to concentrate in the
gallbladder [43].

The test is done in a hospital. After preliminary x-rays are


taken with the patient lying down on a tilting x-ray table,
the contrast medium is administered intravenously and
more pictures are taken at different angles [43].

The only risk is possible allergic reaction to the contrast


medium. To help minimize this, a small amount is
administered first and reaction checked before the full dose
is given. There may, however, be side effects including
nausea, vomiting, hives, or flushing [43].

Cholangiography (Percutaneous Transhepatic)


Figure 24. Endoscope [43].
This is a contrast x-ray study of the bile duct, often used to
diagnose causes of jaundice. It is especially useful with
patients who have had their gallbladders removed. No food
is given for about 8 hours before the test [43].

This procedure lasts about 30 minutes. It always performed


in the hospital. The patient lies on a tilting x-ray table and a
local anesthetic is injected into the skin covering the liver
and the capsule surrounding it. A long, flexible needle
guided by a fluoro-scope is inserted into the liver in an
attempt to find a dilated bile duct, and the contrast medium
administered through it. Then a series of x-rays is taken as
the table is rotated to different angles [43].

A number of complications can result from this procedure,


including bleeding, septicemia (a bacterial infection of the
blood), and bile peritonitis caused by bile leaking from the
punctured duct, but these are rare. Although the radiation
exposure from a single cholangiography is small, there is a
risk if the procedure is repeated several times. The patient
will feel some sting from the needle used for the anesthetic,
but will not feel the needle being inserted into the liver.
When the dye is injected, there may be a feeling of fullness
or pressure. Side effects of the contrast dye sometimes
include nausea, vomiting, excessive salivation, hives, and
sweating [43].

Cholecystography

Figure 25. Upper GI [43]. Cholecystography is a contrast x- ray study of the

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-21
gallbladder used to detect gallstones or to diagnose long, flexible tube called a catheter is passed through the
inflammatory disease and tumors. At noon on the day nose into the stomach. The patient then lies down and the
before the test, the patient eats a meal that contains normal doctor, guided by a fluoroscope screen, continues to pass
amounts of fat, which stimulates release of bile from the the catheter into the duodenum. Injection or IV gives a
gallbladder, emptying it of normal bile and preparing it to hormone or drug to relax the duodenum, and then the
accept bile containing a contrast medium. This is followed contrast medium is administered through the catheter and
by a fat-free dinner, which causes the contrast-containing x-rays are taken from a number of angles. Some of the
bile to collect in the gallbladder. After dinner, at 5-minute barium is then withdrawn and air injected through the
intervals, the patient takes a series of pills containing the catheter, which may cause the patient to have cramps. More
contrast dye. The test takes about 30 to 45 minutes. This x-rays are taken before the catheter is removed [43].
outpatient procedure is simply a series of x-ray pictures
taken with the patient in different positions. After the initial Risks, which include allergy to the contrast medium or
pictures, the patient may be given a high-fat meal or a pill possible internal damage from the introduction of the
to produce the same effect, and asked to wait 15 to 20 catheter, are rare. More common are side effects such:
minutes, after which another set of pictures is taken [43]. nausea, vomiting, hives, flushing, blurred vision, dry
mouth, thirst, irregular heart rhythms [43].
Except for possible allergic reaction to the contrast
medium, there is no risk involved in this test. Some people Endoscopy
suffer temporary side effects caused by the pills, most
commonly diarrhea and occasionally nausea, vomiting, or Endoscopy is a diagnostic test that enables the doctor to
difficulty in urinating [43]. look inside the upper digestive system. The instrument used
to perform this simple test is the endoscope, a long thin
Colonoscopy flexible tube that has at one end a miniaturized color TV
camera with a wide angle lens. By passing this "scope" into
Colonoscopy is the direct examination of the large intestine your stomach, the doctor can directly examine the lining of
using a flexible fiber-optic tube inserted through the anus. the upper digestive system on a television monitor. The
It is used to detect inflammatory or ulcerative bowel technical name for this test is Esophago-Gastro-
disease, to check for polyps or tumors, or to locate the site Duodenoscopy. To simplify things the shortened form of
of gastrointestinal bleeding [43]. the name endoscopy is usually used, or the initials EGD.
The examination is quick and painless. There is no incision
The patient is given a liquid diet for 1 to 2 days before the [43].
exam, followed by a laxative and one or more enemas to
completely empty the bowels. A sedative or tranquilizer Alternate testing
may also be given [43].
As part of the evaluation, the doctor may have already
This procedure takes 30 to 60 minutes. It can be done as an ordered a barium x-ray of upper digestive system. This x-
outpatient procedure, but is usually done in the hospital. ray exam may have been helpful in directing attention to an
With the patient lying on his or her side and knees flexed, a area of possible abnormality. But x-rays often do not fully
colonoscope is passed through the intestine. Air may be reveal all [43].
injected to expand the folds of the colon, enabling the
doctor to see the entire surface area. When the tube has Fortunately, endoscopy permits a much more accurate view
reached the entire length of the colon, it is slowly of the upper digestive system. Rather than studying an
withdrawn as the doctor examines the lining for any lesions indirect x-ray "shadow picture" of the stomach, endoscopy
or polyps. Special attachments allow the removal of some allows the doctor to directly view the upper digestive tract
small polyps or the collection of a biopsy sample [43]. in "living color" and to examine the lining in remarkable
detail. Thus, it is much more accurate [43].
There is very little risk of bowel perforation, but the patient
should report any bleeding, dark stool, or abdominal pain. Understandably, most patients are a bit apprehensive about
Many patients feel embarrassment and discomfort during the idea of "swallowing a scope." However, with the help
this procedure, especially when air is injected and expelled of modern medical instruments and sedation, patients
[43]. usually tolerate this test with relative ease. Endoscopy is
useful in diagnosing the following:
Duodenography
1 Abdominal pain.
Duodenography (also called hypotonic duodenography) 2. Bleeding from the digestive tract.
provides x-ray pictures of the duodenum and the pancreas 3. Cancers of the stomach or esophagus.
and is used to diagnose tumors or lesions that may be 4. Chronic heartburn and indigestion.
causes of upper abdominal pain. The patient must fast for 6 5. Detection and treatment of esophageal varices.
to 12 hours before the test [43]. 6. Diagnosis and removal of stomach polyps.
7. Dilatation of esophageal strictures.
This test takes about 30 minutes. It can be done on an 8. Gastritis, or stomach inflammation.
inpatient or outpatient basis. The patient is seated, and a 9. Hiatal hernia.

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-22
10. Removal of swallowed objects. depending on the objective of the test. The longer test
11. Trouble swallowing. requires hospitalization, but the shorter ones may be done
12. Treatment of the "Ulcer Bacteria". on an outpatient basis. The doctor passes a long, flexible
13. Ulcers of the esophagus. tube through a nostril down the patient's throat into the
14. Ulcers of the stomach u Ulcers of the duodenum. stomach. Once the tube has been started, the patient is
15. Unexplained chest pain. asked to swallow until the tube enters the stomach. The
gastric juices are then aspirated and collected through the
Endoscopy Retrograde Cholangiopancreatography tube over a period of hours [43].
(ERCP)
The test is uncomfortable and may cause gagging when the
ERCP is x-ray visualization of the ducts leading from the tube is initially passed and some irritation in the nostril or
pancreas and the gallbladder is used to diagnose the throat afterward. Other risks, which are rare, include lung
presence of stones or tumors in these ducts. collapse or ulceration of the larynx if the tube is passed into
the trachea instead of the esophagus [43].
The patient is asked to fast for 12 hours before the test and
is usually given a sedative or tranquilizer. Local anesthetics Gastroscopy
(generally unpleasant tasting) is used in either spray or
gargle form to suppress gagging, and it causes the patient to Gastroscopy (also called esophagogastroduo-denoscopy)
lose some control of saliva. A mouth guard may be inserted allows a direct view of the lining of the esophagus,
to protect the teeth [43]. stomach, and duodenum through an endoscopic tube. It is
used to determine the cause of bleeding to perform a
ERCP takes about 60 minutes. While it can be done on an biopsy, or to diagnose inflammatory disease, tumors,
outpatient basis, it is usually done on an inpatient basis. ulcers, and structural abnormalities [43].
The endoscope is inserted into the mouth and down the
throat and the patient is instructed to swallow to help pass it The patient is asked to fast for 6 to 12 hours beforehand
down into the esophagus. Guided by a fluoroscope, the and is given a sedative or tranquilizer. A local anesthetic is
doctor then continues to pass it down into the stomach and sprayed into the mouth and throat. The procedure takes
duodenum, and then injects a drug into the duodenum to about 30 minutes. It may be done on an inpatient or
relax it. Next the contrast medium is injected through the outpatient basis. A flexible fiber-optic tube is passed
endoscope and a series of x-rays is taken. Another set may through the mouth down the esophagus and into the
be taken in different positions after the scope is removed. stomach and duodenum [43].
The patient may feel side effects from the drug or hormone
used to relax the duodenum and from the contrast medium. The tube, which transmits light and allows the doctor to see
These include nausea, hives, blurred vision, dry mouth, the internal structure of various organs, also has
urinary retention, and a feeling of burning or flushing. The attachments that allow the suctioning off of secretions,
throat may be sore for 3 or 4 days afterward [43]. taking of biopsy samples, and introduction of air, which
expands the organs and makes examination easier. As with
Besides the slight risk of infection or perforation of internal all invasive procedures, there is some risk of internal
organs that accompanies any invasive procedure, there is damage to organs or structures, such as perforation, but this
also the problem of urine retention, especially in men with is relatively rare. The patient may feel some gagging,
prostate problems. This can usually be avoided by voiding cramps, or fullness as the tube is passed through various
completely before the exam. organs or when air is introduced, and will probably have a
sore throat afterward [43].
Esophagography
Liver scan
Although this test is occasionally done alone, when the
doctor is reasonably sure that the problem is confined to the This isotope study provides information on the structure
esophagus, it is usually done as part of the upper GI series. and functioning of the liver and is used to detect cirrhosis,
abscesses, or growths. There is no risk, other than the
Gastric secretion studies remote possibility of isotope overdose. No special
preparation is necessary. The procedure takes about 45 to
This procedure allows the physician to collect and study 60 minutes for a structure scan and 30 minutes for each
samples of the gastric juices secreted by the stomach to function scan, which are done in a series with the final one
diagnose causes of epigastric pain or learn more about after 24 hours. The procedure may be done on an
suspected ulcers before surgery. The patient is usually fed a outpatient basis. The isotope is injected and a blocking
liquid meal and then asked to fast for 12 hours and to agent may be used [43].
restrict liquids and smoking for 8 hours prior to the test.
Part of the test may involve eating, drinking, or chewing to Lower GI Series
stimulate production of gastric juices. [43]
Like the upper GI Series, this x-ray exam (also called
Although the secretions are all collected in the same barium enema) uses barium as a contrast medium to
manner, the test can last anywhere from 2 to 12 hours, provide fluoroscopic views, in this case of the large

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-23
intestine. The procedure is used to detect polyps, tumors, The patient follows a high-protein diet for several days
gastroenteritis, telescopic bowel, irritable colon or other before the test, then fasts for 8 to 12 hours, with the
causes of lower abdominal pain, or blood, mucus, or pus in exception of a glass of skim milk a few hours before the
the stool [43]. test. Another high-protein liquid is given at the time of the
test.
Before the examination, the patient may be asked to fast or
go on a liquid diet, and then laxatives or one or more A pancreas scan takes 1 to 2 hours. It may be done on an
enemas will be used to empty the bowels. inpatient or outpatient basis. The isotope sometimes causes
nausea or vomiting [43].
The test takes 30 to 45 minutes. It can be done in a doctor's
office, testing facility, or hospital outpatient clinic. The Proctosigmoidiscopy
barium enema is administered through a tube inserted into
the rectum while the patient lies on his side on a tilting x- A proctoscope exam (also called proctoscopy) allows the
ray table. The table will be rotated to allow the barium to doctor to see the rectum and the lower part of the large
completely coat the large intestine and x-ray pictures will intestine. It may be used to detect hemorrhoids, polyps, and
be taken. The patient will then be helped to a bathroom or abscesses or to determine the cause of bleeding. It is
given a bedpan and told to expel as much of the barium as routinely used to screen for cancer after age 40, even if
possible. More pictures will then be taken of the thin film there are no symptoms [43].
that remains on the mucous lining of the intestine. In some
cases, this will be followed by a double contrast study, The patient may be instructed to have a liquid diet or to fast
meaning air will be carefully injected into the colon to for a few hours, followed by an enema or suppository.
provide further contrast on the film in order to detect This procedure takes 10 to 20 minutes. It may be carried
possible polyps [43]. out in a doctor's office or in a clinic. With the patient in the
knee-chest position, the doctor will do a digital rectal exam,
Although this procedure is unpleasant and often using a gloved finger, and then insert the proctoscope,
uncomfortable, there are no serious complications. Risks, which may be a rigid or a flexible tube. The procedure is
such as perforation of the bowel, are rare. The procedure similar to the colonoscopy, but not as extensive [43].
causes some discomfort, including a strong urge to defecate
and mild or, especially with the air contrast study, Although the patient may feel some irritation around the
moderates to severe cramps. The stool will be white or light anus afterward, this is a routine, safe procedure, with the
colored for a few days following the test, and the barium risk of injury to the large intestine being very rare. The
may be constipating. For this reason, the test is sometimes patient may feel some discomfort and the urge to defecate
followed by a laxative or another enema [43]. during the procedure [43].

Pancreas Scan Sialography

This radioisotope scan provides information on the This procedure (also called ptyalography) allows x-ray
structure and functioning of the pancreas to help detect visualization of the salivary glands to diagnose stones in
cancer, cysts, or infection. There is a rare risk of the salivary ducts or other causes of an enlarged or painful
radioisotope overdose. salivary gland [43].

No special preparation is required, except possibly fasting


for a few hours beforehand. The patient may be given a
sedative. This procedure takes about 45 to 60 minutes. It is
usually performed on an outpatient basis. A catheter is
inserted through the mouth into the duct of the salivary
gland to be studied and a contrast medium is injected
through the catheter. The doctor examines the gland using a
fluoroscope and may take x-rays. There is some discomfort
when the catheter is inserted and pain when the dye reaches
the salivary gland [43].

As with all invasive procedures, there is a rare risk of


internal damage or infection. The patient will continue to
taste the contrast dye and experience some soreness in the
mouth after the procedure [43].

Small Bowel Examination

This exam is usually done in conjunction with an upper GI


series. This fluoroscopic examination (also called barium
Figure 26. Gastroscope [43]. milk shake or barium swallow) of the esophagus, stomach,

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-24
and small intestine is used to diagnose cases of hiatal Measurements for the nervous system
hernia, ulcers, tumors, obstruction, or enteritis, and when
there are symptoms such as difficulty in swallowing,
regurgitation, burning or gnawing abdominal pain, We can measure direct electrical activity of the nervous
diarrhea, weight loss, or bleeding [43]. system by many ways. The effects of the nervous system
on the other system of the body are manifested in most
The patient may be asked to eat a low-residue diet for 2 or physiological measurements. The Gate Control Theory
3 days before the test and will be asked not to eat or smoke suggests that by electrically stimulating sensory nerve
for 6 to 12 hours prior to the exam. receptors, a gate mechanism is closed in a segment of the
spinal cord, preventing pain-carrying messages from
It may take anywhere from 30 minutes (for just the reaching the brain and blocking the perception of pain. We
esophagus) to 60 minutes (for the esophagus and stomach) can stimulate sensor neurons with the specific type of
to 6 or more hours (if the small intestine is also included). stimulus and measure the responses in various nerves, in
This test may be done in a testing facility or in a hospital on some cases, in individual neurons either in the peripheral or
an inpatient or outpatient basis. It begins with the patient central nervous system measurement. The Endorphin
swallowing a "barium milk shake," a sweetened, flavored, Release Theory suggests that electrical impulses stimulate
but nonetheless chalky-tasting substance containing barium the production of endorphins and enkaphalins in the body.
sulfate, a contrast medium that outlines the upper digestive These natural, morphine-like substances block pain
tract. Sometimes a more liquid radiopaque material is used. messages from reaching the brain [15].
The patient is strapped securely to a tilting x-ray table, Open MRI equipment
which starts in a vertical position with the patient standing,
and is tilted at various angles throughout the test to help The MRI is a medical diagnostic technique that creates
spread the contrast medium and to get different views on images of the body using the principle of nuclear magnetic
the fluoroscope. Pressure may be applied to the patient's resonance. This machine can generate images of thin
abdomen to spread the medium further. If the test is to sections of any part of the body, including the heart,
include a small bowel exam, the patient will have to wait arteries and veins from any angle and direction, in a
several hours until the barium filters down into the relative short period of time.
intestine, and may be able to leave and come back, or to MRI is possible in the human body because the body is
spend the time reading [43]. filled with small biological magnets, the most abundant and
responsive of which the proton, the nucleus of the hydrogen
Although the barium is unpleasant tasting and will make atom. The MRI stimulates the body with ratio waves to
the patient's stool white for several days (unless a laxative change the steady-state orientation of protons. It then stops
is given to speed its passing), there is no pain, little the ratio waves and listens to the body’s electromagnetic
discomfort, and virtually no risk with this test, unless it is transmissions at a select frequency. The transmitted signal
repeated several times, when radiation then becomes a risk is used to construct internal images of the body. MRI is
[43]. preferred for diagnosing most diseases of the brain and the
central nervous system. It can distinguish soft tissue in both
THE NERVOUS SYSTEM normal and diseased states [15].
The nervous system of the human being is responsible for
sending, receiving, and processing nerve impulses
throughout the body. All the organs and muscles inside our
body rely upon these nerve impulses to function. It could
be considered as the master control unit inside the body.
Sense organs provide the nervous system with information
about the environment by means of such senses as sight,
hearing, smell, taste, tough, pressure, and pain. Nerves are
connected throughout the whole body to the brain. These
carry the information throughout the body in the form of
electrochemical signals called impulses. These impulses
travel from the brain and spinal cord to the nerves located
throughout the body. It is largely made up of specialized
cells called neurons, they are the basic unit of the nervous
system, a neuron is a single cell with the cell body,
sometimes is called the Sonoma, one more imput fibers
called dendrites, a long transmitting fiber called the axon.
Each of this neuron has a cell body, or cyton, which
contains the nucleus and organelles. It takes the corporation
of three system to carry out the mission of the nervous
system. These are the central, the peripheral, and the
autonomic nervous systems [15].
Figure 27. Open MRI [43].

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-25
This machine allows clinicians to conduct advanced technology and one-half second scanning enable faster and
procedures, including, high-resolution routine imaging, all more precise CT imaging. It provides high-performance
forms of vascular imaging techniques including 2D/3D CT angiographies with multi-slice, half-second acquisition.
TOF, PC and CE. MRA, high-resolution (RF refocused) Precision in CT scanning and intervention are attained by
EPI, breath-hold imaging (abdomen and chest); a wide use of Aspire CI technology. Cardiac gating offers
range of pulse sequences usually associated with high-field capabilities to provide cost-effective, fast and simple
systems; Bolus tracking (optional on Excelart). coronary artery screening. The advanced CAT Aquilion is
Superconducting speed and performance are combined with engineered for 1/2 second, multi-slice scanning [15].
the patient access advantages of open MRI design for
unmatched clinical versatility. OPART incorporates the Patients can be scanned faster clinical exploration.
latest advances in gradient, RF and computer technology to Furthermore, the Aquilion platform can be enhanced at a
deliver advanced clinical applications, such as MR future date, without replacing the gantry, by the addition of
Cholangio-Pancreatography, MR Myelography and MR multi-slice scanning capability. Electroencephalogram
Urography, and Images of the highest diagnostic clarity. (EEG)with much more anatomical detail, since the
The system's superconducting magnet provides all of these Aquilion is redesigned to withstand the g-forces created by
studies with OPART's four sided open design. This implies the incredible speed of 1/2 second scanning. The Aquilion
that a broader patient population can be accommodated. It’s can be combined with Aspire CI for improved CT
all around patient access enables the scanning of large, interventional applications, and shatter the limits of
claustrophobic or pediatric patients. And also allows for Electroencephalogram is the measurement of the electrical
examinations which are not possible in many closed activity of the brain. The measurements are obtained from
systems OPART features a sophisticated icon based user electrodes placed on the surface of the scalp. For clinical
interface that allows intuitive operation, enabling high measurements, surface or subdermal needle electrodes are
patient tohroughput and allowing routine examinations to used. The ground reference electrode is often a metal clip
be conducted quickly and efficiently [15]. on the earlobe. A suitable electrolyte paste or jelly is used
CAT Scan in conjunction with the electrodes to enhance coupling of
the ionic potentials to the input of the measuring device. To
reduce interference and minimize the effect of electrode
This system uses X rays and computers to produce three- movement, the resistance of the path through the scalp
dimensional images of the human body( figure 27). It between electrodes must be kept as low as possible [15].
provides detailed views of the body’s soft tissues, including
blood vessels, muscle tissue, and organs, such as the brain.
A CAT of the brain can show if an accident victim has Normal patterns on the electroencephalogram, or EEG,
sustained a brain injury, or if bleeding is occurring in the include the alpha wave, indicative of a relaxed state, and
brain. It also may be used to determine the location and size the theta wave, usually found in children and thought by
of a cancerous tumor and whether the cancer has spread to some to reflect creative activity in adults. Medically the
other parts of the body. A new application for the CT most important use of electroencephalography is to aid in
include CT angiography, which provides detailed images of the diagnosis of epilepsy. The EEG tracing can identify a
the interior of small blood vessels and images of the specific site at which damage to the brain has occurred.
interior of hollow structures, such as the colon and the Study of EEGs has contributed to research into the nature
airways of the lungs [15]. of sleep, helping to identify four stages of sleep. Scientists
have also analyzed brain waves stimulated by sensory
input, such as a flash of light or a sound, as a means of
To enhance image, patients may be given an injection of a determining which parts of the brain carry out various
substance that will increase the contrast between different functions [15].
tissues. The patient can be also asked to drink a liquid that
makes internal organs more clearly and visible in the CT
scan. The CT scan contains an X-ray source, which emits
beams of the X-rays that pass through the patient and are
recorded on the other side by the detector. As the source
and the detector rotate in a 360- angle circle around the
patient, X-ray emissions are recorded from many angles.
Then the computer interprets the data, it translates the
images that appear as cross-sections on a television monitor
[15].

A system so powerful its potential once seemed


unattainable. With more elements and thinner slices, CAT
offers images with remarkable clarity and detail, as well as
more rapidly scans a larger coverage area. Toshiba CT
systems feature expanded clinical applications and system
utility to meet the advanced needs of cardiology, trauma
care, surgery, pediatrics and oncology. Toshiba’s
technology offerings address these needs with: multi-slice Figure 28. CT Scan [43].

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-26
Most clinical electroencephalographs provide the capability
of simultaneously recording EEC signals from several
regions of the brain. The readout in a clinical
electroencephalograph is a multichannel pen recorder with
a pen for each channel. The standard chart speed is 30
mm/sec, but most electroencephalographs also provide a
speed of 60 mm/sec for improved detail of higher-
frequency signals. Many electroencephalographs also have
provisions for interfacing with an analog tape recorder to
permit recording and playback of the EEG signal.
Dedicated cardiology imaging system Features improved
patient throughput with 180°, 90° or 76° myocardial Spect
Acquisition. Real-time infrared body contouring reduces
setup time [15].

The electrode sock

Patients with diabetic foot pain, sprains and fractures in the


feet, ankles and other pediatric ailments now have an easy,
at-home, treatment option: the Electrode Sock. The sock
provides stimulation to the entire foot and ankle to increase
blood flow, thereby reducing swelling and associated pain
[43].

Transcutaneous electrical nerve stimulator (TENS)

TENS is an acronym for Transcutaneous Electrical Nerve


Stimulation. Developed in the late 1960’s, the TENS
stimulator is a battery-powered device which transmits an
electrical impulse through lead wires and surface electrodes
to underlying nerves. The stimulator converts the direct Figure 29. Electroencephalogram [43].
current of the battery into pulses of stimulation. Most
stimulators feature adjustable settings to control amplitude The lack of ATP due to trauma of the tissue results in the
(intensity) of stimulation by controlling voltage, current, decreased production of sodium and an increase in
and pulse width (duration) of each pulse [43]. metabolic wastes, which is perceived as pain. The use of
Microcurrent at an injured area helps to realign the body’s
Electrodes are placed at specific sites on the body for electrical current, increase the production of ATP, resulting
treatment of pain. The current travels through electrodes in increased healing and recovery, as well as blocking the
and into the skin stimulating specific nerve pathways to pain that is perceived [43].
produce a tingling or massaging sensation that reduces the
perception of pain. When used as directed, TENS is a safe, Electrical muscle stimulator
non-invasive, drug-free method of pain management [43].
EMS is the use of electrical stimulation on muscle groups
Microcurrent electrical neuromuscular stimulator to contract and reeducate muscles. Some of the uses of
EMS are as follows: (1) Maintaining and increasing range
MENS use a very low voltage current, usually between 1 of motion in conditions where the reduction of
µA and 1000µA. A microamp (µA) is 1/1000 of a milliamp physiological range of motion is due to or the result of
(mA), so 1000 µA equals 1 mA. Most TENS devices have fractures with consequent immobilization, operative
a milliamplitude of 1-80 mA. Where TENS is used to hide intervention, or arthroscopy, in shoulders, knees, and backs.
pain, microcurrent, is thought to work on a more cellular (2) The Prevention or Retardation of muscle Disuse
level. It has been theorized that healthy tissue is the result Atrophy: Muscle disuse atrophy is a reduction 'in muscle
of the direct flow of electrical current throughout the body. contraction and size due to prolonged impairment or joint
Electrical balance is disrupted when the body is injured at a immobility from surgery, injury or disease. The use of
particular site, causing the electrical current to change electrical stimulation to contract the muscles builds and
course. The use of Microcurrent over the injured site is strengthens the muscles, assisting in prevention of disuse
thought to realign this flow, thus aid in tissue repair. It's atrophy. (3) Relaxation of muscle Spasms: Muscle
been found that ATP (Adenosine Triphosphate) in the cell spasms and cramping often occur in areas of localized pain
helps to promote protein synthesis and healing [43]. and tenderness. Stimulation is used to fatigue the "spastic"
muscle. (4) Muscle Reeducation: Evidence has shown that
a combination of both exercise and electrical stimulation is
far superior in strengthening atrophied muscles. Increased
Local Blood Circulation: Rhythmic muscle contraction

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-27
helps improve blood circulation, thereby aiding in the Interferential generator
reduction of localized swelling and tenderness. (5)
Immediate Post-surgical Stimulation of Calf Muscles to Interferential electrical stimulation's is a unique way of
Prevent Venous Thrombosis: The use of EMS to increase effectively delivering therapeutic frequencies to tissue.
blood circulation assists in the prevention of venous Conventional TENS and Neuromuscular stimulators use
thrombosis [43]. discrete electrical pulses delivered at low frequencies of 2-
160 Hz per second. However, Interferential stimulators use
High voltage pulsed galvanic stimulation a fixed carrier frequency of 4,000 Hz per second and also a
second adjustable frequency of 4,001-4,400 Hz per second.
High-voltage pulsed galvanic stimulation (HVPGS) is When the fixed and adjustable frequencies combine
gaining widespread use for wound healing, edema (heterodyne), these produce the desired signal frequency
reduction and pain relief Carpal Tunnel Syndrome and (Interference frequency). Interferential stimulation is
Diabetic Foot are two major areas of use. Devices in this concentrated at the point of intersection between the
class are characterized by a unique twin- peak monophasic electrodes. This concentration occurs deep in the tissues as
waveform with very short pulse duration (microseconds) well as at the surface of the skin. Conventional TENS and
and a therapeutic voltage greater than 100 volts. The Neuromuscular stimulators deliver most of the stimulation
combination of very short pulse duration and high peak directly under the electrodes. Thus, with Interferential
current, yet low total current per second (microcurrent) Stimulators, current perfuses to greater depths and over a
allows relatively comfortable stimulation. Furthermore, this larger volume of tissue than other forms of electrical
combination provides an efficient means of exciting therapy. When current is applied to the skin, capacitive skin
sensory, motor and pain-conducting nerve fibers. resistance decreases as pulse frequency increases.'
Perceptual discrimination of those responses is relatively
easy to achieve and thus its clinical versatility. Background For example, at a frequency of 4,000 Hz (Interferential
skin offers a great amount of resistance to the flow of unit) capacitive skin resistance is eighty (80) times lower
electrical current. When current is passed through a circuit than with a frequency of 50 Hz (in the TENS range). Thus,
that contains resistors, voltage drop occurs and energy is Interferential current crosses the skin with greater ease and
lost. This phenomenon occurs when traditional low voltage with less stimulation of cutaneous nociceptors allowing
units are used in treatment A high voltage device produces greater patient comfort during electrical stimulation. In
a spontaneous breakdown in skin resistance and HVPGS addition, because medium-frequency (Interferential)
current passes through the skin with negligible thermal and current is tolerated better by the skin, the dosage can be
electrochemical effects. Bell Telephone Laboratories increased, thus improving the ability of the Interferential
developed the first high voltage stimulator in 1945. By current to permeate tissues and allowing easier access to
decreasing the pulse duration and increasing the voltage, deep structures. This explains why Interferential current
the developers noted that deep tissues could be stimulated may be most suitable for treating patients with deep pain,
without producing tissue damage. The first published report for promoting osteogenesis in delayed and nonunion
(I 966) described its effectiveness in wound healing of fractures and in pseudothrosis. Also for stimulating deep
animal limbs [43]. skeletal muscle to augment the muscle pump mechanism in
venous insufficiency, and for depressing the activity of
certain cervical and lumbosacral sympathetic ganglia in
patients with increased arterial constrictor tone [43].

Clinical electromyography (EMG)

Clinical electromyography consists of nerve conduction


studies (NCS) and needle electromyography (needle EMG).
In the strict sense of the word, EMG only refers to the
needle electrode examination of muscles, however is has
traditionally been used to refer to both NCS and needle
EMG. It is a diagnostic modality for the evaluating the
peripheral nervous system which, can provide invaluable
information that may not be obtainable with any other
diagnostic test. It can also complement the information
obtained from other sources, such as imaging studies, for
optimal localization of a lesion [43].

There are dozens of muscles and nerves that can be tested,


and it is the electromyographer's job to tailor the
examination to the patient's particular clinical problem.
What is tested is as important as how competently it is
done.

Figure 30. The Electrode sock [43].

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-28
Peripheral nerves transmit electrical impulses and can be Carpal tunnel syndrome (CTS)
simplistically thought of as extremely efficient conductive
wires. For NCS studies, a noninvasive stimulator applies CTS is the single most common referring diagnosis in the
brief electrical impulses to a peripheral nerve EMG laboratory and by far the most common entrapment
transcutaneously, the nerve then transmits the impulse and neuropathy. Electrophysiological features include:
a response is recorded by electrodes at some distance away.
The time it takes for the stimulus to reach the recording
electrodes (latency) can be accurately measured and a 1. Prolongation of the median nerve motor latency
velocity of transmission calculated. Both motor and sensory upon stimulation at the wrist ("terminal" or
nerves can be examined. Healthy nerves will transmit the "distal" latency).
electrical impulse faster than diseased ones [43]. 2. Normal median motor conduction velocity in the
forearm.
3. Slow median sensory conduction velocity across
Needle EMG does not introduce any electrical stimulation,
the wrist.
instead it records the intrinsic electrical activity of skeletal
muscle fibers. The needle is quite slim (about a 25 gauge)
and produces minor discomfort which most patients can Normal motor and sensory NCS of the ulnar nerve (not
tolerate. Needle EMG findings suggestive of denervation involved in CTS). NCS are positive in 91-98% of patients
include: with clinical CTS8 [43].
1. Fibrillations,
2. Positive sharp waves, and Although false negatives do occur, those are so infrequent
3. Giant motor unit potentials (MUP). that the diagnosis should be reconsidered. An
asymptomatic contralateral CTS can be seen in 32% of
Abnormalities pointing to a myopathy include small short- patients [43].
duration MUP. Since the root and peripheral nerve supply
to the muscles is known, needle EMG can be used to Ulnar neuropathy
localize peripheral nerve or root lesions by noting which
muscles show denervation changes.
The second most frequent entrapment neuropathy is ulnar
neuropathy at the elbow. Entrapment at the wrist is much
For most conditions, NCS and needle EMG complement less frequent. An useful clinical tip in making a distinction
each other and are performed in the same session. is that when the lesion is at the wrist, only the sensation in
the palm is affected, whereas with a lesion at the elbow
both the palm and the dorsum of the hand are numb. This is
so because the ulnar nerve branch which supplies sensation
for the dorsum of the hand (dorsal ulnar cutaneous nerve)
leaves the main trunk of the ulnar nerve at the forearm so it
would not be involved in lesions at the wrist [43].

The cardinal EMG findings in elbow involvement are:

1. Sslowing of ulnar motor conduction velocity


across the elbow.
2. Conduction block across the elbow segment.
3. Needle EMG evidence of denervation in ulnar-
innevated muscles.

Peripheral neuropathy

In this group of conditions the peripheral nerves are


affected in a diffuse fashion and more or less
symmetrically. This is distinguished from entrapment
neuropathies where only one nerve is affected. In the USA
the most common etiologies are diabetes mellitus (by far
the most frequent), alcohol abuse, and uremia. In third
world countries, leprosy is one of the leading causes. EMG
features of peripherial neuropathy include:

1. 1.Diffuse slowing of nerve conduction velocity in


sensory and/or motor nerves, usually worse in the
Figure 31. Electromyography machine [43]. lower limbs,

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-29
2. 2.Sensory nerves are usually affected earlier than Table 1. Conditions referred for EMG.
motor ones,
Yield of EMG
3. 3.Needle EMG evidence of denervation in distal Condition Comment
EMG Features
muscles is not always present and is not essential
for the diagnosis. NCS are diagnostic in 80 % of + median
cases8. n. TL, Most common
CTS 91-98% + median entrapment
SCV, nl neuropathy
Apart from establishing the diagnosis, NCS can also be ulnar NCS
helpful in suggesting the etiology of the neuropathy.
Peripheral neuropathies are classified as "axonal" or Ulnar slow MCV Wrist
"demyelinating", depending on which part of the nerve is Neuropathy 73-91% across entrapment is
most affected by the disease process. The great majority of (elbow) elbow, CB less frequent
polyneuropathies are of the axonal type. Only a handful are Diabetes most
demyelinating, so finding NCS features of demyelination diffusely common
really narrows down the differential diagnosis6,8. Examples Polyneuropathy >80%
slow etiology in
of demyelinating peripheral neuropathies include: Guillain- USA
Barre syndrome, chronic demyelinating neuropathy, and
polyneuropathy associated with gammopathy [43]. denervation
in EMC not
Radiculopathy 80-100% paraspinals positive before
Cervical and lumber radiculopathy
and 7-20 days
myotome
Needle EMG is essential to establish the diagnosis. NCS
are usually normal in radiculopathies, but are required to small-short Polymyositis is
Myopathy varies*
exclude concurrent peripheral neuropathy or MUP most common
mononeuropathy which may affect the needle EMG
examination. EMG features include:
The yield of EMC varies with the etiology and severity of
1. Needle EMG evidence of denervation involving the myopathy ABBREVIATIONS: n. = nerve, TL =
paraspinal muscles and, terminal latency, NCS = nerve conduction studies, SCV =
2. Denervation involving limb muscles of the same sensory conduction velocity, MCV = motor conduction
root innervation (myotome). The yield is at least velocity CB = conduction block, MUP = motor unit
80%. potentials

Myopathy

In adults, polymyositis is the most common cause of Both NCS and needle EMG are only mildly uncomfortable
acquired myopathy. Needle EMG is essential for diagnosis. to most patients. No special preparation is required for the
NCS are performed for the same reasons as in test, except to avoid applying oils or lotions on the skin as
radiculopathies [43]. it interferes with electrode adherence.

In cases of radiculopathy, if a specific root level is As with any other ancillary medical test, clinical correlation
suspected, mention it. For entrapment neuropathies, include is crucial. It should be ask if the EMG diagnosis makes
the nerve and the suspected site of involvement (wrist, sense with the clinical picture? An inadequately performed
elbow, etc.). A short pertinent clinical history is also very or interpreted EMG can become a nightmare and lead to
helpful. Important conditions to mention are diabetes unnecessary procedures and even surgery1. If the EMG
mellitus, alcoholism, uremia, cancer, and collagen vascular diagnosis is confusing, discuss it with the
diseases. In general, any severe systemic disease is worth electromyographer or request a formal neurological or
mentioning. It is best to request both NCS and needle EMG neurosurgical consultation. As with everything else in
since these usually complement each other [43]. medicine, common sense is the key [43].

The time of onset of symptoms is very important since THE URINARY SYSTEM
needle EMC evidence of denervation takes at least 7-10
days to develop following nerve injury 7. Thus, a negative The human body has several systems to remove waste. The
needle EMG examination done within 10 days of the onset urinary system produces and discharges urine. The kidneys
of symptoms should be repeated later. A two weeks interval play a vital role within the urinary system by filtering out
is recommended to be certain that enough time has passed water and waste from the blood. These waste materials are
for needle EMG findings to be detectable [43]. excreted (eliminated) by the kidneys in the form of urine
[22].

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-30
An adult bladder holds about 1 pint (475ml) of urine. You
excrete anywhere from 1.5 - 3.5 pints (.7 to 2 liters) of
urine a day. Did you know that sweat is made up of the
same substances as urine? That's right! Although they
contain different concentrations of water, salt, and urea,
sweat and urine are made up of the same stuff [11].

The cystoscope [22] Figure 32. Endoscope [43].

Cystoscopy is a procedure that enables your health care 3. Normal Values


provider to view the inside of your bladder and urethra in
great detail using a specialized endoscope (a tube with a The bladder wall should appear smooth. The bladder
small camera used to perform tests and surgeries) called a should be normal size, shape, and position. There should be
cystoscope [22]. no obstructions, growths, or stones.

1. How is the test iperformed? 4. Special considerations

There are two types of cystoscopes, the standard rigid One may notice a small amount of blood in the urine
cystoscope and the flexible cystoscope. The method for following this procedure. If the bleeding continues after
insertion of the cystoscope varies, but the test is the same. three voids (urinations), contact the health care provider.
The choice of which scope to use depends on the purpose
of the exam. There is also a chance of infection. If one develops pain,
chills, fever, or reduced urine output, contact ther health
If the standard rigid cystoscope is used, you lie in the care provider.
lithotomy position (on back with your knees up and apart).
The flexible cystoscope may be easier to insert than the Endoscope [22]
standard rigid model. It does not require the lithotomy
position for insertion.
What is Endoscope and Why is it Performed?
The procedure usually takes between 5 and 20 minutes. The
urethra is cleansed and a local anesthetic is applied. The Endoscope allows physicians to peer through the body's
scope is then inserted through the urethra into the bladder. passageways. Endoscope is the examination and inspection
of the interior of body organs, joints or cavities through an
Water or saline is inserted through the cystoscope and fills endoscope.
the bladder. As this occurs, you are asked to describe the
feeling to your health care provider, which provides An endoscope is a device using fiber optics and powerful
information about your condition [22]. lens systems to provide lighting and visualization of the
interior of a joint. The portion of the endoscope inserted
As the fluid fills the bladder, it stretches the bladder wall, into the body may be rigid or flexible, depending upon the
enabling the physician to view the entire bladder wall. You medical procedure [22].
will feel the sensation of a need to urinate when the bladder
is full; however, it needs to remain full until the An endoscope uses two fiber optic lines. A "light fiber"
examination is complete [22]. carries light into the body cavity and an "image fiber"
carries the image of the body cavity back to the physician's
If any tissue appears abnormal, a small specimen can be viewing lens. There is also a separate port to allow for
taken (biopsy) through the cystoscope to be analyzed. administration of drugs, suction, and irrigation. This port
may also be used to introduce small folding instruments
such as forceps, scissors, brushes, snares and baskets for
2. Why the test is performed ? tissue excision (removal), sampling, or other diagnostic and
therapeutic work. Endoscopes may be used in conjunction
1. Diagnose and evaluate urinary tract disorders with a camera or video recorder to document images of the
inside of the joint or chronicle an endoscopic procedure.
2. Check for cancer of the bladder or urethra New endoscopes have digital capabilities for manipulating
and enhancing the video images [22].
3. Diagnose an enlarged prostate (men)

4. Help determine the cause of pain during urination A rigid endoscope (Figure 30) is used for arthroscopy. The
"image fiber" leads from the ocular (eye piece) to the
5. Diagnose recurrent bladder infections inserted end of the scope. The "light fiber" is below and

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-31
leads from the light source to the working end of the small volume doses - less than 10 cc. Use a 16 gauge for
endoscope. large volume doses of 10 cc or greater [7].

Why Is Endoscopy Performed? Needles of course come in different lengths. When drawing
blood from pigs you will be using a needle that is 6 inches
Endoscope can be used to diagnose various conditions by long! Drawing blood from the neck of a sheep or the tail
close examination of internal organ and body structures. vein of a cow is usually done with a 1 inch needle. For
Endoscope can also guide therapy and repair, such as the intramuscular injections a 1 inch needle is commonly used
removal of torn cartilage from the bearing surfaces of a with young animals while a 1.5 inch needle would be used
joint. Biopsy (tissue sampling for pathologic testing) may in adults. For subcutaneous injections a .5 to 1 inch needle
also be performed under endoscope guidance. Local or is favored [22].
general anesthetic may be used during endoscope,
depending upon the type of procedure being performed The viscosity of the product also plays a part in gauge
[22]. selection. If the product is "thick" and pushes through the
needle hard, opt for the 16 gauge needle even for the small
Internal abnormalities revealed through endoscope include: dose. If selecting one needle for all-around performance, a
abscesses, biliary (liver) cirrhosis, bleeding, bronchitis, 1-inch, 16 gauge needle works well. Its a good all around
cancer, cysts, degenerative disease, gallbladder stones, needle that meets the requirements of intramuscular and
hernia, inflammation, metastatic cancer, polyps, tumors, subcutaneous injections. When using needles for injections
ulcers, and other diseases and conditions [22]. it is best to change them every 10 to 15 injections, or each
time you stop to refill an automatic syringe. Dull needles
cause excess issue damage. which increases the chances of
Endoscope is a minimally invasive procedure and carries reactions and abscesses forming. Also, over used needles
with it certain minor risks depending upon the type of can become dirty and spread germs. Never just throw away
procedure being performed. However, these risks are used needles. They need to be stored in a "sharps"
typically far outweighed by the diagnostic and therapeutic container. This is a plastic box used to store needles until
potential of the procedure [22]. they can be properly disposed of. Often, the veterinarian
can or a local hospital or the local public health office can
Prior to the widespread use of endoscope and diagnostic remove and replace the container [22].
imaging, most internal conditions could only be diagnosed
or treated with open surgery. Until the last several decades,
exploratory surgery was routinely performed when a
patient was critically ill and the source of illness was not
known. For example, in certain dire cases the patient's
thorax or abdomen were surgically opened and examined to
try to determine the source of illness [22].

Endoscope can often be done on an outpatient basis.


"Outpatient" means that the procedure does not require
hospital admission and acute care and observation and may
be performed outside the premises of a hospital. Outpatient
procedures performed at hospitals or ambulatory centers
allow the patient to go home or return to work within a
short while after their procedure [10].
Figure 33. Needles [30].
Needles and syringes

Needles have evolved over the years so that today they


come in many different sizes and diameters, or gauges.
Needles also have different bevels on the tip to better do
the jobs they are used for. Gauges run from 4 to 30 , with
the smaller number being the largest around. A four gauge
needle would be used to insert an t6rocar with a 10 gauge
used to place a catheter in a jugular vein of a cow. On the
other end of the scale a 30 gauge needle would be used by a
dentist to give you Novocain before drilling your brains
out. The most commonly used gauges of needles in
domestic animal production would be in the 16 to 22 range
with 18 the most common. Use an 18-gauge needle for
Figure 34. Syringes [31].

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-32
Syringes also have evolved over the years. Today there are dialyser, which is connected to the dialysis machine. A
many sizes and shapes of syringes. Many are specialized, pump on the machine, assisted by the body's blood
or used for just one thing. Basically there are just two types pressure, moves the blood through the tubing and dialyser.
though. Syringes that are made to be disposed of after one Only about 300ml (approximately 2 cups) is out of the
use, and those that are made to be used many times [7]. body at any one time. The blood is pumped at around 2-300
mls per minute and over a four hour dialysis session the
Disposable syringes are in theory used only one time and whole 5 litres of blood in the average person’s body will
then discarded. The syringe used to draw blood from a flow through the artificial kidney approximately six times.
human probably falls under this description. However, Haemodialysis is usually performed three times a week at
plastic, disposable syringes are the backbone of many regularly spaced intervals. Dialysis sessions usually last 4-6
livestock producers health programs [7]. hours [7].

Automatic or re-usable Syringes. are made to be used for 3. What are dialysers (artificial kidney) and dialysate?
many years . They must be maintained with proper cleaning
and storing if they are to do so however [7]. The dialyser is a plastic cylinder that contains thousands of
very fine tubes or fibres. Each of these has tiny pores in its
Haemodialysis treatment walls, only small enough for waste and extra fluid to pass
through. Blood cells and protein cannot fit through the
holes. Blood is pumped to the dialyser and flows through
Dialysis means the replacement of the excretory function of the fibres. A specially formulated fluid called ‘dialysate’
the kidneys by artificial means. Dialysis is only necessary washes around the outside of the fibres. Dialysate has a
when the kidneys have failed. There are two principal similar chemical composition to healthy blood, so the
forms of dialysis - haemodialysis and peritoneal dialysis. electrolytes (which are unbalanced in renal failure) can
This Fact Sheet addresses haemodialysis. Haemodialysis is equilibrate and be returned to the normal range. The used
the process whereby blood is pumped from the patient dialysate goes down the drain into ordinary sewage waste
through an artificial kidney (called a dialyser) and back to [7].
the patient [7].
4. Is haemodialysis painful?
1. What type of dialysis should be used?
Haemodialysis is not painful. The patient does not feel the
The choice between haemodialysis or peritoneal dialysis blood being moved through the dialyser. A local
depends upon many factors including the availability of anaesthetic is often used prior to the insertion of the
resources, age, overall health and lifestyle. The advantages needles. The fistula arm with the needles in it must be kept
and disadvantages of each type of dialysis need to be reasonably still. People on haemodialysis can read, talk,
weighed up in consultation with the clinical team caring for play board or card games, watch television, write, use a
the person with kidney failure and their family. It is computer or sleep. They cannot get up and move around
possible to change from haemodialysis to peritoneal [7].
dialysis or vice versa if it is found that the initial treatment
option does not suit the person or their lifestyle. Discussion 5. For how long must a person continue dialysis treatment?
with your nephrologist is essential if a treatment change is
desired [7]. Dialysis must be performed for the rest of a persons life or
until they receive a kidney transplant.
2. What happens during haemodialysis?
6. Extracorporeal blood circuit

As with peritoneal dialysis, haemodialysis replaces lost


kidney function. During haemodialysis, an artificial kidney
machine (or "dialysis machine") replaces the excretory
function of the failed kidneys. Blood from the patient is
pumped from the body through special tubing to the
dialysis machine, where it travels through the dialyser and
back to the patient. The dialyser replaces some of the
balancing functions of the native kidney. To do this access
to the patient’s blood supply is needed. This is achieved by
the creation of a fistula or graft and is usually placed in the
forearm [7].

For each treatment, the patient sits next to the machine.


Two needles are inserted into the fistula or graft - one takes
blood to the machine, the other returns blood to the patient.
During treatment, the blood in the body flows through the Figure 35. Flow Traces in a Centrifugal pump [18]

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-33
Optical, acoustic, or electronic means, and direct imaging,
electronic processing may form the image, or a
combination of these methods receives it. The microscope
may be static, in which the object is viewed directly, or
dynamic, in which the image is built up by successive scans
of the object [5].
The magnifying power of a microscope is an expression of
the number of times the object being examined appears to
be enlarged. Magnifying power is a dimensionless ratio.
The resolution Microscopeof a microscope is a measure of
the size of the smallest detail on the object that can be
observed. Resolution is expressed in linear units, usually
millimeters [5].

The most familiar type of microscope is the optical, or


light, microscope, in which lenses are used to form the
image. Optical microscopes can be simple, consisting of a
Figure 36. Pressure contours in a Axial Blood flow pump
single lens, or compound, consisting of several optical
[18].
components in tandem. Simple optical magnifiers may have
magnifying powers of from 1 to 10 magnitudes, with
Several bloodflow pumps have been successfully
resolutions to about 0.01 millimeter (mm) possible.
characterized using figure 18, including axial and
Compound optical microscopes have magnifying powers
centrifugal pumps. CFD analysis has been used to quantify
ranging from about 2.5 to 1,000 magnitudes and provide
stress and hemolysis (using the built-in hemolysis model)
resolutions of 0.01 to 0.0002 mm [5].
levels in several blood pumps.. Multiple designs of blood
flow pumps have been virtually prototyped for performance
Other types of microscopes utilize the wave nature of
optimization [18].
various physical processes, the most important being the
electron microscope, which uses a beam of electrons in its
THE ENDOCRINE SYSTEM image formation. Electron microscopes have magnifying
powers of 1,000 to 1,000,000 magnitudes possible, with
The endocrine system is a group of organs and body tissues resolutions ranging from 0.001 to 0.000 000 01 mm.
that produce, store, and secrete chemical substances known Special microscopes, such as acoustic microscopes or
as hormones. Hormones are the body’s chemical scanning tunneling microscopes, use other physical effects,
messengers and transfer information and instructions from further extending the range of objects that can be
one set of cells to another. Endocrine organs have a great viewed.Indeed, with a scanning tunneling microscope, even
deal of influence over body thanks to the hormones. Their individual molecules and atoms can be seen [5].
job is to regulate the body’s growth and development, the
control of the function of various tissues, the support of PHYSICAL THERAPY
pregnancy and other reproductive functions, and the
regulation of metabolism [5]. High voltage pulsed galvanic stimulation

HVPG stimulation is a treatment modality to decrease pain


To study the system, the observer performs a blood analysis and swelling. Galvanic stimulation is direct current (DC),
that indicates any endocrine alteration. A technique used moving fluid from the affected area back into circulation. It
for diagnosis of the endocrine diseases is the can be used to treat and provide therapeutic relieve from
radioinmunoanalysis. It is based on the usage of specific the following indications: relaxation of muscle spasm,
antibodies for a determined hormone. A urine analysis is arthritic pain, increasing local blood circulation, and
also used to quantify parameters in the diagnostic of maintaining or increasing range of motion. It is also very
endocrine diseases such as glucosuria or concentration of good in edema reduction, reducing motor spasticity,
glucose in urine, present in diabetes. It can also determine preventing or retarding disuse atrophy and managing post
the quantity of hormones excreted by their urine that gives traumatic or post surgical pain conditions, and it is ideal
an idea of the functioning of the endocrine glands. The X- when used in therapy for carpal tunnel syndrome [43].
rays are used to observe the endocrine glands; to see their
shape and size to determine any disorder on them. Some The Thera Tech Equipment has two types of portable
visual tests are also performed for diagnosis of some galvanic stimulators designed for home use and offers
complications [5]. conventional electrodes, as well as, the complete line of
electromesh garments. The sterling impulse is powered by
MICROSCOPE a 9-volt battery system, single alkaline work best. It records
from 6 minutes to 999.9 hours in 6-minute increments [43].
The instrument that produces enlarged images of small
objects, allowing them to be viewed at a scale convenient In order to prevent the patient from wearing the stimulator
for examination and analysis, is called the microscope. to long, an automatic shut-off feature is incorporated. The

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-34
glove also supplies electrostimulation from a Pulsed- it passes through tissue cells, resulting in a temperature
Galvanic Stimulator and other stimulation modalities [43]. increase [43].

Ectrica1 stimulating currents The electrotherapeutic shortwave and microwave devices


have preset frequencies and wavelengths that cannot be
The electrical stimulating currents that affect nerve and altered. Shortwave diathermy units are set at either 13.56
muscle tissue have the longest wavelengths and the lowest MHz with a corresponding wavelength of 22 m or 27.12
frequencies of any of the modalities. The wavelengths of MHz with a wavelength of 11 rn.
electrical stimulating units are extremely long, ranging
somewhere around 15,000 km (clinically used frequencies Microwave units have shorter wavelengths than do
range from 1 to 2000 Hz). Most stimulators have the shortwave diathermy units and are set at wavelengths of 33
flexibility to alter the frequency output of the device to or 12 cm with respective frequencies of 915 or 2450 MHz.
elicit a desired physiologic response. The nerve and muscle The depth of penetration with microwave is bit deeper than
stimulating currents are capable of pain modulations either with shortwave because the amount of energy when using
through stimulation of cutaneous sensory nerves at high microwave is concentrated in one spot rather than spread
frequencies or through production of β-endorphin at lower out over a large area [43].
frequencies, producing muscle contraction and relaxation
or tetany depending on the type of current (alternating or Ultraviolet therapy
direct) and frequency, and producing a net movement of
ions and thus eliciting a chemical change (direct current The physiologic effects of ultraviolet are mainly chemical
only) [43]. in nature and occur entirely in the cutaneous layers of skin.
The maximum depth of penetration with ultraviolet is about
Infrared modalities 1 mm. Ultraviolet wavelength ranges between 1850 and
4000 Å. The ultraviolet region is subdivided into two
The infrared modalities are used to produce a local and different areas; near ultraviolet (2900 to 4000 Å) and far
occasionally a generalized heating or cooling of the ultraviolet (1850 to 2900 Å). Clinically used frequencies
superficial tissues. It is generally accepted that the infrared with ultraviolet range between 7 X 1013 and 7 x 1014 Hz.
modalities have a maximum depth of penetration of 1 cm or Ultraviolet therapy is used rarely by the sports therapist to
less. The infrared modalities can elicit either increases or treat cutaneous lesions [43].
decreases in circulation depending on whether heat or cold
is used. They are also known to have some analgesic The acoustic spectrum and ultrasound
effects as a result of stimulation of sensory cutaneous nerve
endings [43]. Ultrasound devices produce a type of energy that must be
classified as a portion of the acoustic spectrum rather than
The wavelengths of the infrared modalities are much the electromagnetic spectrum. Ultrasound is frequently
shorter than are those of the electrical stimulating cur rents classified along with shortwave and microwave diathermy
and the diathermies and are expressed in Angstrom (Å) as a deep-heating conversion type modality, and it is
units; 1 Å is equal to l0 meters (m). certainly true that all of these are capable of producing a
temperature increase in human tissue to a considerable
The infrared and ultraviolet wavelengths are temperature depth. However, ultrasound is a mechanical vibration, a
dependent. Those modalities with the lower temperature sound wave, produced and transformed from high-
have the longer wavelength. Temperatures used with the frequency electrical energy [43].
infrared modalities range from 0° C with ice to more than
3000° C with the infrared lamps. The wavelengths in this Acoustic and electromagnetic radiations have very different
temperature range fall between 10,000 and 105,000 Å with physical characteristics. When acoustic vibrations are
corresponding frequencies ranging between 2 x 1012 and 4 produced, they travel at a velocity that is significantly
x 1013 Hz [43]. lower than electromagnetic radiations. Electromagnetic
waves travel at approximately 300 million meters per
The critical factor is the superficial increase or decrease in second while sound waves travel at speeds from hundreds
tissue temperature that elicits the same physiologic to several thousand meters per second [43].
response regardless of wavelength [43].
Frequencies of ultrasound wave production are between
Shortwave and microwave diathermy 700,000 and 1,000,000 cycles per second. The depth of
penetration with ultrasound is much greater than with any
The diathermies are considered to be high-frequency of the electromagnetic radiations. At a frequency of 1
currents because they have more than a million cycles per megacycle, 50% of the energy produced will penetrate to a
second. When impulses of such a short duration come in depth of about 5 cm. The reason for this great depth of
contact with human tissue there is not sufficient time for penetration is that ultrasound travels very well through
ion movement to take place. Consequently there is no homogeneous tissue while electromagnetic radiations are
stimulation of either motor or sensory nerves. The energy almost entirely absorbed. Thus when therapeutic
of this rapidly vibrating electrical current produces heat as penetration to deeper tissues is desired, ultrasound is the
modality of choice [43].

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-35
Clinical ultrasound is capable of producing a temperature nerve and muscle cell membranes to develop and maintain
increase in the deeper tissues and thus to some extent pain the resting potential, the membranes are excitable [43].
relief as a result of both mechanical and chemical effects.
To create transmission of an impulse in the nerve tissue,
Electrical treatments there must be a reduction in this resting membrane
potential below a threshold level. Changes in the
Electrical treatments can provide dramatic results at times. membrane’s permeability then occur. These changes create
The use of electricity in treating an injury can be beneficial, an active potential that will propagate the impulse down the
but the therapist must base the use of electricity on factual nerve in both directions from the location of the stimulus
information about the effects of electricity on biologic [43].
tissues. Electrical therapy should not be used in a shotgun
approach if we are to maximize the effectiveness of this An action potential is created by a stimulus from chemical,
modality [43]. electrical, thermal, or mechanical means. The stimulus
always creates the same result, membrane depolarization.
All biologic tissue will have some response to having an
electrical current passed through it. The type and extent of Not all stimuli are effective in causing this depolarization.
this response are dependent on the type of tissue and its To be an effective agent, the stimulus must have an
response characteristics, and the nature of the current adequate intensity and last long enough to equal or exceed
applied. The tissue should respond to electrical energy in a the membrane’s basic threshold for excitation. The stimulus
manner similar to the manner in which it normally must alter the membrane so that a number of ions are
functions or grows. This is essential in the use of any pushed across the membrane, exceeding the ability of the
therapeutic modality. Clinically, therapists use electrical active transport pumps to maintain the resting potentials. A
currents for: stimulus of this magnitude forces the membrane to
depolarize and results in an action potential [43].
• Creating muscle contraction through nerve or
muscle stimulations. Clinical response to electrical currents
• Stimulating sensory nerves to help in treating
pain. When an electrical system is applied to the soft tissues of
• Creating an electrical field in biologic tissues to an extremity, the system provides the force to induce ion
stimulate or alter the healing process. movement in the tissues and create an action potential. At
• Creating an electrical field on the skin surface to the positive electrode (also called the anode), positive ions
drive ions beneficial to the healing process into or are repelled and negative ions are attracted. At the negative
through the skin. electrode (also called the cathode), the negative ions are
repelled and the positive ions are attracted [43].
Muscle and nerve responses to electrical currents
Depolarization
The major therapeutic use of electricity is concentrated
around muscle contraction or sensory stimulation or both. As the charged ions move across the nerve fiber
membranes beneath the anode and cathode, membrane
Nerve and muscle are both excitable tissues. This depolarization occurs. The cathode is usually the site of
excitability is dependent on the cell membrane’s voltage depolarization. As the concentration of negatively charged
sensitive permeability. Nerve or muscle cell membrane ions increases, the membrane’s voltage potential becomes
regulates the interchange of substances between the inside low and is brought toward its threshold for depolarization.
of the cell and the environment outside the cell. The cell The anode makes the nerve cell membrane potential more
membrane allows certain ions to move freely into or out of positive, increasing the threshold necessary for
the cell while almost totally keeping other ions in place depolarization [43].
outside or inside the cell. This voltage sensitive
permeability creates an unequal distribution of charged ions Depolarization effects
on each side of the membrane, which in turn creates a
potential difference between the charge of the interior of As the nerve impulse reaches its effect on organ or another
the cell and the exterior of the cell. The membrane is then nerve cell, the impulse is transferred between the two at a
considered polarized. The potential difference between motor end plate or at a synapse. At this junction, a
inside and outside is known as the resting potential because transmitter substance is released from the nerve rather than
the cell tries to maintain a homeostatic, or balanced, having the impulse jump from one to another. This
electrical relationship between the inside and outside of the transmitter substance causes the other excitable tissue to
tissue. Both electrical and chemical equilibrium must be discharge.In terms of muscle excitation, a twitch muscle
established along the membrane; that is, the net movement contraction would result. This contraction, initiated by an
of ions into the cell is equaled by the net movement of ions electrical stimulus, would be the same as a twitch
out of the cell. The cell continually moves Na ions out of contraction coming from voluntary activity. Voluntary
the cell and K to the inside of the cell by active transport to muscular activity is different only in the rate and synchrony
maintain this equilibrium. In addition to the ability of the (simultaneous response) of the muscle fiber contractions
[43].

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-36
Muscular responses to electrical current stimulated muscles. Normal firing in voluntary
muscle contraction varies from one moment to
Stimulation of the motor nerve is the method used in most the next because some motor units are
clinical applications of electrical muscular contraction. In contracting while others are inactive. Voluntary
the absence of nerve supply, muscle contraction can be contractions do not lead to muscular fatigue as
stimulated by an electrical current by causing the muscle early in the exercise period as electrical
membrane to depolarize [43]. contractions. This synchrony of contraction may
also be important in using electrical muscle
The all or none response is another important concept in contractions to improve muscular strength [43].
applying electrical current to nerve or muscle tissue. Once a
stimulus reaches a depolarizing threshold, the nerve or Therapeutic uses of electrically induced muscle
muscle membrane depolarizes and propagation of the contraction
impulse or muscle contraction occurs. This reaction
remains the same regardless of increases in the strength of A variety of therapeutic gains can be made by electrically
the stimulus used. Either the stimulus will cause stimulating a muscle contraction:
depolarization, the all, or the stimulus will not cause
depolarization, the none. There is no gradation of response • Muscle reeducation
other than by recruiting more nerve fibers by stimulation. • Muscle pump contractions
The response of the single nerve or muscle is maximal or • Retardation of atrophy
none at all [43]. • Muscle strengthening
• Increasing range of motion
This all or none phenomenon does not mean that muscle
fiber contraction and overall muscle activity cannot be Any electrical stimulator, high voltage, low voltage,
influenced by changing the intensity, pulses per second, or alternating current, hybrid current, or transcutaneous
duration of the stimulating current. Adjustments in current electrical nerve stimulating (TENS) units, may be used to
parameters can cause muscle fiber contraction [43]. cause muscle contraction. The efficiency and effectiveness
of the treatment can be increased by following the
Effects of changes in current parameters protocols as closely as possible with available equipment
[43].
1. The motor nerves are not stimulated by a steady
flow of direct current. The nerve re-polarizes Muscle reeducation
under the influence of the current and will not
depolarize until a sudden change in current Muscular inhibition after surgery or injury is the primary
intensity occurs. indication for muscle reeducation. If the neuromuscular
mechanisms of a muscle have not been damaged, then
2. If continuous direct current were the only current central nervous system inhibition of this muscle is usually a
mode available, we would get a muscle factor in loss of control. Muscle contraction can usually be
contraction only when the current intensity rose caused by electrically stimulating the muscle. Forcing the
to a stimulus threshold. Once the membrane re- muscle to contract causes an increase in the sensory input
polarized, another change in the current intensity from that muscle. The patient feels the muscle contract,
would be needed to force another depolarization sees the muscle contract, and can attempt to duplicate this
and contraction [43]. muscular response.
3. The biggest difference in the effect of alternating Muscle pump contractions
and direct current is the ability of direct current to
cause chemical changes. Chemical effects from Electrically induced muscle contraction can be used to
using direct current will only occur when the duplicate the regular muscle contractions that help
stimulus is continuous and it is applied over a stimulate circulation by pumping fluid and blood through
period of time. These chemical changes become venous and lymphatic channels back into the heart. One of
measurable after about 1 minute, but the effect is the major problems in most traumatic injuries and surgical
cumulative over the total treatment time. This interventions is the excessive accumulation of fluid. This
type of current is available in most low-voltage edema is the result of damage to the vascular structure, loss
equipment [43]. of normal muscular activity, and dependency of the
extremity. Electrical stimulation of muscle contractions in
4. The difference between electrically induced the affected extremity can help in reestablishing the proper
muscle contraction and voluntary muscle circulatory pattern while keeping the injured part protected
contraction is the asynchrony of firing of motor [43].
units under voluntary control versus the
synchronous firing of electrically stimulated Retardation of atrophy
motor units. Each tune the electrical stimulus is
applied, the same motor units should respond. Prevention or retardation of atrophy has traditionally been
This may lead to greater fatigue in the electrically the reason for treating patients with electrically stimulated

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-37
muscle contraction. The maintenance of muscle tissue, after • Electrodes may be placed close to the spinal cord
an injury that eliminates normal muscular exercise, can be segment that innervates an area that is painful.
accomplished by substituting an electrically stimulated • Peripheral nerves that innervate the painful area
muscle contraction. The electrical stimulation reproduces may be stimulated by placing electrodes over
many of the physical and chemical events associated with sites where the nerve becomes superficial and can
normal voluntary muscle contraction and helps to maintain be easily stimulated.
normal muscle function [43]. • Both acupuncture and trigger points have been
conveniently mapped out and illustrated. A
Muscle strengthening reference on acupuncture or trigger areas should
be consulted and the sports therapist should
Muscle strengthening from electrical muscle stimulation systematically attempt to stimulate the points
has been used with some good results in patients with listed as successful for certain areas and types of
weakness of a muscle group. Several studies also indicate path, If they are effective the patient will have de
that strength gain can be achieved [43]. creased pain. These can also be identified using
an ohm meter point locator to determine areas of
Increasing range of motion decreased skin resistance.
• Combinations of any of the above systems and
Increasing the range of motion in contracted joints is also a bilateral electrode placement can also be
possible and documented use of electrical muscle successful.
stimulation. Electrically stimulating a muscle pulls the joint • Crossing patterns, also referred to as an
through the limited range. The continued contraction of this interferential technique, involve electrode
muscle group over an extended time appears to make the application so that the electrical signals from
contracted joint and muscle tissues modify and lengthen. each set of electrodes add together at some point
Reduction of contractures in patients with hemiplegia has in the body and the intensity summates. The
been reported, although no studies have reported this type pulses must arrive at the point to be stimulated at
of use in contracted joints from injuries or surgery [43]. the same time and with the same wave form and
frequency. The electrodes are usually arranged in
Therapeutic uses of electrical stimulation of sensory a crisscross pat tern around the point to be
nerves stimulated.
Clinically efforts are made to stimulate the sensory nerves The practitioner should not be limited to any one system
to change the patient’s perception of a painful stimulus but should evaluate electrode placement for each patient.
coming from an injured area [43]. The effectiveness of this treatment is closely tied in with
proper electrode placement. As in all trial and error
Electrical stimulation of sensory nerves may stimulate the treatment approaches a systematic organized search is
release of enkephalin from local sites throughout the central always better than a shotgun hit and miss approach [43].
nervous system and the release of β-endorphin from the
pituitary gland into the cerebral spinal fluid. The Protocol hyper stimulation and β-endorphin
mechanism that causes the release and then the binding of
enkephalin and β -endorphin to certain nerve cells is still To use the influence of hyper stimulation analgesia and
unclear. It is certain that a diminution or elimination of pain endorphin release, a point stimulation setup must be used.
perception is caused by applying an electrical current to A large dispersive pad and a small pad or hand-held probe
areas close to the site of pain or to acupuncture or trigger point electrode are used in this approach. Electrode
points, both local and distant to the pain area [43]. placement is decided. The point electrode is applied and the
intensity is increased until it is perceived by the patient.
Electrode placement The probe is then moved around the area and the patent is
asked to report relative changes in perception of intensity.
When using any of the treatment protocols aimed at the When a location of maximum intensity perception is found,
electrical stimulation of sensory nerves for pain the current intensity is increased to maximum tolerable
suppression, there are several guidelines that will help the levels [43].
practitioner select the appropriate sites for electrode
placement. Transcutaneous electrical nerve stimulation A combination of intense point stimulation and
(TENS) uses similar sized electrodes placed according to a transcutaneous electrical nerve stimulation may be used.
pattern and moved in a trial and error pattern until pain is The transcutaneous electrical nerve stimulation applications
decreased. The following patterns may be used: should be used as much as is needed to make the patient
comfortable and the intense point stimulation should be
• Electrodes may be placed on or around the used on a periodic basis. Periodic use of intense point
painful area. stimulation gives maximal pain relief for a period of time
• Electrodes may be placed over specific and allows some gains in overall pain suppression. Daily
dermatomes, myotomes, or sclerotomes that intense point stimulation may eventually bias the central
correspond to the painful area. nervous system and decrease the effectiveness of this type
of stimulation [43].

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-38
Clinical uses of low-voltage uninterrupted direct occurs by three mechanisms: conduction, convection, and
current radiation [43].

The application of uninterrupted low-voltage direct current Conduction occurs when the body is in direct contact with
causes several physiologic changes that can be used the heat or cold source. Convection occurs when particles
therapeutically. The therapeutic benefits are related to the (air or water) move across the body, thus creating a
polar and vasomotor effects and to the acid reaction around temperature variation. The final mechanism is radiation or
the positive pole and the alkaline reaction at the negative the transfer of heat from a warmer source to a cooler source
pole [43]. through a conduct medium such as air. It must also be
pointed out that the body may either gain or lose heat
There is also a vasomotor effect on the skin, increasing through any of these three processes of heat transfer. The
blood flow between the electrodes. The benefits from this infrared modalities use these three methods of heat transfer
type of direct current are usually attributed to the increased to affect a tissue temperature increase or decrease.
blood flow through the treatment area [43].
Heat modalities are referred to as thermotherapy.
Promotion of wound healing Thermotherapy is used when a tissue temperature rise is the
goal of treatment. The use of cold, or cryotherapy, is most
Similar direct current has been used to treat skin ulcers that effective in the acute stages of the healing process the
have poor blood flow. The treated ulcers show accelerated following injury when tissue temperature loss is the goal of
healing rates when compared to untreated skin ulcers. The therapy. Cold applications can be continued into the
wound areas should be treated for the first 3 days with the reconditioning stage of injury management. The term
negative electrode at the wound site and the positive hydrotherapy can be applied to any cryotherapy or
electrode positioned 25 cm from the wound. The negative thermotherapy technique that uses water as the medium for
electrode and its alkaline reaction seem to inhibit the heat transfer [43].
growth of bacteria without destroying it. After the initial
negative pole treatment, the electrodes were reversed and Physiologic effects of heat
the positive pole was placed in the wound. The positive
pole promotes the migration of skin cells toward the center Local superficial heating (infrared heat) is recommended in
of the wound and decreases healing time in these ulcers sub acute conditions for reducing pain and inflammation
[43]. through analgesic effects. Superficial heating produces
lower tissue temperatures at the site of the pathology
Promotion of fracture healing (injury) relative to the higher temperatures in the superficial
tissues, resulting in analgesia. During the later stages of
The use of uninterrupted low-voltage direct current may be injury healing, a deeper heating effect is usually desirable;
an adjunctive modality in the treatment of fractures, it can be achieved by using the diathermies or ultrasound
especially fractures prone to non-union. Fracture healing [43].
may be accelerated by passing a direct current through the
fracture site. Getting the current into the bony area without Heat dilates blood vessels, causing the resting capillaries
an invasive technique is difficult [43]. (patent small blood vessels) to open up and increase
circulation. The skin is supplied with sympathetic
Iontophoresis vasoconstrictor fibers that secrete norepinephrine at their
endings (especially evident in feet, hands, lips, nose, and
Direct current has been used for many years to drive ions ears). At normal body temperature the sympathetic
from the heavy metals into and through the skin for vasoconstrictor nerves keep vascular anastomoses (blood
treatment of skin infections or for a counter-irritating vessel junctions) almost totally closed, but when the
effect. There are two techniques of application: (1) an superficial tissue is heated, the number of sympathetic
active pad is applied over gauze that is saturated with a impulses is greatly reduced so that the anastomoses dilate
solution containing the ions, and (2) an immersion and allow large quantities of warm blood to flow into the
technique in which the active electrode is suspended in a venous plexuses (a group of veins). This promotes heat loss
container of solution, then the part to be treated is from the body, which can increase blood flow about
immersed in the container. In both cases, a large dispersive twofold [43].
pad is applied to the patient and the proper polarity of the
active electrode is selected based on the polarity of the ions The hyperemia (increased blood flow) created by heat has a
in the solution [43]. beneficial effect on injuries. This is based on increases of
blood flow and pooling of blood during the metabolic
Mechanisms of heat transfer processes. Recent hematomas (blood clots) should never be
treated with heat until resolution of bleeding in completed
Easy application and convenience of use of hot and cold [43].
modalities provide the therapist with the necessary tools for
primary injury care. Heat is defined as the internal vibration The rate of metabolism of tissues is dependent in part on
of the molecules within a body. The transmission of heat temperature. The metabolic rate has been shown to increase
approximately 13% for each 1 C increase in temperature. A

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-39
similar decrease in metabolism has been demonstrated modulation. With ice treatments, the patient usually reports
when temperatures are lowered [43]. an Uncomfortable sensation of cold followed by burning,
then an aching sensation, and finally complete numbness
A primary effect of local heating is an increase in the local [43].
metabolic rate with a resulting increase in the production of
metabolites and additional heat. These two factors lead to Cold does depress the excitability of free nerve endings and
an increased intravascular hydrostatic pressure, causing peripheral oral nerve fibers, and this increases the pain
arteriolar vasodilatation and increased capillary blood flow. threshold. This is of great value in short-term treatment.
However, with increased hydrostatic pressure, there is a Cold applications can also enhance voluntary control in
tendency toward formation of edema, which may increase spastic conditions, and in acute traumatic conditions they
the time required for rehabilitation of a particular injury. may decrease painful spasms that result from local muscle
Increased capillary blood flow is important with many irritability. Reduction in muscle spasm relative to acute
types of injury in which there is mild or moderate athletic trauma has been observed by all active sports
inflammation, because it causes an increase in the supply of therapists [43].
oxygen, antibodies, leukocytes, and other necessary
nutrients and enzymes, along with an increased clearing of The initial reaction to cold is local vasoconstriction of all
metabolites. With higher heat intensities vasodilatation and smooth muscle by the central nervous system in order to
increased blood flow will spread to remote areas, causing conserve heat. Localized vasoconstriction is responsible for
increased metabolism in the unheated area. This is known the decrease in the tendency toward formation and
as consensual heat vasodilatation and may be useful in accumulation of edema probably as a result of a decrease in
many conditions where local heating is contraindicated local hydrostatic pressure- There is also a decrease in the
[43]. amount of nutrients and phagocytes (cells that eliminate
debris) delivered to the area, thus reducing phagocyte
The application of heat can produce an analgesic effect, activity [43].
resulting in a reduction in the intensity of pain. The
analgesic effect is the most frequent indication for the use The following physiologic effects of cold have been
of heat. Although the mechanisms under this phenomenon documented:
are not well understood current thinking is that it is in some
way related to the gate control theory of pain modulation • Decreased local temperature, in some cases to a
[43]. considerable depth
• Decreased metabolism
Heat is applied in musculoskeletal and neuromuscular • Vasoconstriction of arterioles and capillaries
disorders, such as sprains, strains, articular (joint-related) • Decreased blood flow
problems, and muscle spasms, which all describe various • Decreased nerve conduction velocity
types of muscle pain [43]. • Decreased delivery of leucocytes and phagocytes
• Decreased lymphatic and venous drainage
Effects of cooling
• Decreased muscle excitability
• Decreased muscle spindle depolarization
The physiologic effects of cold are for the most part
opposite those of heat, the primary effect being a local • Decreased viscosity of muscle
decrease in temperature. Local cooling is strongly indicated • Decreased formation and accumulation of edema
in the acute stage of the healing process to decrease pain, • Extreme anesthetic effects
muscle spasm, and inflammation. Cold is also known to
reduce the metabolic rate, with a corresponding decrease in The application of cold causes decreased cell permeability,
production of metabolites and metabolic heat. decreased cellular metabolism, and decreased accumulation
of edema and should be continued in 5 to 45 minute
In acute injury there is increasing agreement that the use of applications for tip to 72 hours following initial trauma
cold is the initial treatment for most conditions involving [43].
strains, sprains, and contusions. It is most commonly used
immediately after injury to decrease pain and promote local Effects of tissue temperature change on circulation
vasoconstriction thus controlling hemorrhage and edema
(swelling). It is also used in the acute phase of Local application of heat or cold is indicated for thermal
inflammatory conditions such as bursitis, tenosynovitis, and physiologic effects. The main physiologic effect is on
tendonitis, in which heat may cause additional pain and superficial circulation because of the response of the
swelling. Cold is also used to reduce pain and the reflex temperature receptors in the skin and the sympathetic
muscle spasm and spastic conditions that accompany it. Its nervous system [43].
analgesic effect is probably one of its greatest benefits. One
explanation of the analgesic effect is that cold decreases the The circulatory apparatus is composed of two major vessel
velocity of nerve conduction, although it does not entirely types: nutritive arteries, capillaries, and veins; and vascular
eliminate it. It is also possible that cold bombards central structures for heating the skin. Two types of vascular
pain receptor areas with so many cold impulses that pain structures are the subcutaneous venous plexus, which holds
impulses are lost through the gate control therapy of pain large quantities of blood that heat the surface of the skin

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-40
and the arteriovenous anastomosis, which provides vascular primarily superficial and directly affect the cutaneous blood
communication between arteries and venous plexuses. The vessels and the cutaneous nerve receptors.
walls of the plexuses have strong muscular coats innervated Absorption of infrared energy cutaneously increases and
by sympathetic vasoconstrictor nerve fibers that secrete decreases circulation subcutaneously in both the muscle
norepinephrine. When constricted blood flow is reduced in and fat layers. If the energy is absorbed cutaneously over a
the venous plexus to almost nothing. When maximally long enough period of time to raise the temperature of the
dilated, there is an extremely rapid flow of warm blood into circulating blood, the hypothalamus will reflexly increase
the plexuses. The arteriovenous anastomoses are found blood flow to the underlying tissue. Likewise, absorption of
principally in the volar or palmar surfaces of the hands and cold cutaneously can decrease blood flow via a similar
feet, the lips, the nose, and the ears [43]. mechanism in the area of treatment [43].

When cold is applied directly to the skin, the skin vessels Thus if the primary treatment goal is a tissue temperature
progressively constrict to a temperature of about 15 °C increase with a corresponding increase in blood flow to the
(590 °F), at which point they reach their maximum deeper tissues, it is perhaps wiser to choose a modality,
constriction. This constriction results primarily from such as diathermy or ultrasound that produces energy that
increased sensitivity of the vessels to nerve stimulation, but can penetrate the cutaneous tissues and be directly absorbed
it probably also results at least partly from a reflex that by the deep tissues [43].
passes to the cord and then back to the vessels. At
temperatures below l5 °C (59 °F) the vessels begin to If the primary treatment goal is to reduce tissue temperature
dilate. This dilation is caused by a direct local effect of the and decrease blood flow to an injured area, it should be
cold on the vessels themselves, producing paralysis of the clear that the superficial application of ice or cold is the
contractile mechanism of the vessel wall or block of the only modality capable of producing such a response [43].
nerve impulses coming to the vessels. At temperatures
approaching 0 °C (32 °F), the skin vessels frequently reach Perhaps the most effective use of the infrared modalities
maximum vasodilatation. Vasodilatation associated with should be for the purpose of analgesia or reducing the
intense cold plays a protective role in preventing freezing sensation of pain associated with injury. The infrared
of the exposed portions of the body, and it is commonly modalities stimulate primarily the cutaneous nerve
referred to as hunting reaction [43]. receptors. Within the philosophy of an aggressive program
of rehabilitation, as is standard in most sports medicine
Three types of sensory receptors are involved in the sub- settings, the reduction of pain as a means of facilitating
epithelial tissue: cold, warm, and pain. The pain receptors therapeutic exercise is a common practice [43].
are free nerve endings. Temperature and pain are
transmitted to the brain via the lateral spinothalamic tract. Cryotherapy techniques
The nerve fibers respond differently at different
temperatures. Above 45 °C (113 °F) cold receptors begin Cryotherapy is the use of cold in the treatment of acute
to discharge again and pain receptors are stimulated. Nerve trauma and sub acute injury and for the decrease of
fibers transmitting sensations of pain respond to the discomfort following athletic reconditioning and
extremes of temperature. Both warm and cold receptors rehabilitation. Tools of cryotherapy include ice packs, cold
adapt rapidly to temperature change; the more rapid the whirlpool, ice whirlpool, ice massage, commercial
temperature change, the more rapid the receptor adaptation chemical cold spray, and contrast baths. Application of
[43]. cryotherapy produces a three to four stage sensation. First
there is a sensation of cold followed by a stinging, then a
General exposure to cold elicits cutaneous vasoconstriction, burning or aching feeling and finally numbness. Each stage
shivering, piloerection, and an increase in epinephrine is related to the nerve endings as they temporarily cease to
secretion, so vascular contraction occurs. Simultaneously, function as a result of de creased blood flow [43].
metabolism and heat production are increased to maintain
the body temperature [43]. Depth of penetration is dependent on the amount of cold
and the length of the treatment time because the body is
Increased blood flow supplies additional oxygen to the well-equipped to maintain skin and subcutaneous tissue
area, thus explaining the analgesic and relaxation effects on viability through the capillary bed by reflex vasodilatation
muscle spasm. An increased proprioceptive reflex of up to 4 times normal blood flow.
mechanism may explain these effects. Receptor end organs
located in the muscle spindle are reflexly inhibited by heat The body has the ability to decrease blood flow to the body
temporarily, while sudden cooling tends to excite the segment that is supposedly losing too much body heat by
receptor end organ [43]. shunting the blood flow. Depth of penetration is also
related to intensity and duration of cold application and the
circulatory response to the body segment exposed. If the
Clinical use of the infrared modalities person has normal circulatory responses, frostbite should
not be a concern. Even so, caution should be exercised
Infrared energy can have a depth of penetration greater than when applying intense cold directly to the skin. If deeper
1 cm, thus the effects of the infrared modalities are penetration is desired, ice therapy is most effective using
ice towels, ice packs, ice massage, and ice whirlpools [43].

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-41
Ice massage • The patient must simply remain still during the
treatment to maintain appropriate positioning of
Ice massage can be applied by the therapist or by the the cold pack.
person itself if the person can reach the area of application • Cold pack must be molded onto the skirt
to administer self-treatment. It is best for the first three • The pack should be covered with a towel to limit
treatments to be administered by the therapist to give the loss of cold.
person the full benefit of the treatment. When positioning • A timer should be set or time should otherwise be
the person’s body segment to be treated, it should be noted.
relaxed and the he should be made comfortable. • Treatment time should be 20 minutes.
Appropriate seating and positioning should be taken into
consideration with the application of ice. Administration Physiologic response:
must be thorough to get the most out of the treatment. Ice
massage is perhaps best indicated in conditions in which • Cold progression proceeds through the four
some type of stretching activity is to be used [43]. stages.
• Erythema occurs.
Treatment
1. Ice packs
• Athlete position. Sidelying, prone, supine,
hooklying, or sitting, depending on the area to be Like cold hydrocollator packs, ice packs are indicated in
treated. acute stages of in as well as for prevention of additional
• Self-treatment. Used when athletes can swelling after exercise of the injured part. Treatment
comfortably reach the area to be treated by includes:
themselves.
• Circular motion. Application of ice massage in a • Athlete position. Position depends on the part to
circular pattern with each succeeding stroke be treated.
covering half the previous stroke. • The patient must simply remain still during
• Longitudinal strokes. Application of ice massage treatment.
in a longitudinal motion with each stroke • Pack must be placed on skin.
overlapping half the previous stroke.
• Pack should be secured in place with toweling or
• Peripheral coverage. Ice should be applied for 15 elastic bandage.
to 20 minutes, consistent patterning of circular • Pack should be covered with towel to limit cold
and longitudinal strokes include the sequence
loss.
described in the clinical uses section.
• A timer should be set or time should otherwise be
noted.
Physiologic responses
• Treatment time should be 20 minutes.
• Cold progression proceeds through the four
Physiologic response:
stages: cold, sting ing, burning, and numbness.
• Reddening of the skin (erythema) occurs as a
• Cold progression proceeds through the four
result of blanching or lack of blood in the
stages.
capillary bed. A common example occurs when
• Erythema occurs.
one works outside in the cold without gloves or
appropriate footwear and returns inside to find
the toes beet-red. The body is attempting to pool 2. Cold whirlpool
blood in the area to prevent further tempera ture
The cold whirlpool is indicated in acute and subacute
loss.
conditions in which exercise of the injured pan during a
• Ice applications of 5 to 15 minutes and greater
cold treatment is desired.
than 10 C (50 F) will not stimulate the hunting
reaction and do not Stimulate the reflex
vasodilation that creates the bodys own
Treatment includes:
physically induced heat or increased blood flow.
• Temperature should be set at 5O to 6O °F.
Commercial (cold) hydrocollator packs
• Body segment must be immersed.
Cold hydrocoilator packs are indicated in any acute injury • For total body immersion the water temperature
to a muscle skeletal structure. Treatment includes: should be set at 65 to 80° F.
• Treatment tune should be 5 to 15 minutes.
• Athlete position. Side lying, prone, supine,
hookiying, or sitting, depending on the area to be
treated Physiological responses

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-42
• Cold progression proceeds through the four Ice immersion
stages.
• Erythema occurs. Ice buckets allow ease of application for the therapist.
Again, a wet area should be selected (where spilled water is
3. Cold spray not a concern), with the patient positioned for comfort. The
immersion, like the contrast bath, should be maintained
Cold sprays, such as Fluori-Methane or ethyl chloride, do until desired results ale reached. If cryokinetics are part of
not provide adequate deep penetration, but they do provide the treatment, then the container should be large enough to
adjunctive therapy for acupressure techniques to reduce allow for the movement of the body segment. Ice
muscle spasm. Physiologically this is accomplished by immersion is similar to cold whirlpool in that the body
stimulating the A fibers involved in the gate control theory. segment may be subject to gravity-dependent positions
The primary action of a cold spray is reduction of the pain [43].
spasm sequence secondary to direct trauma. It will,
however, not reduce hemorrhage be cause it works on the Cryoldnetics is a technique that combines the use of
superficial nerve endings to reduce the spasm via the various cryotherapy treatments with exercise. The goal of
stimulation of A fibers to reduce the so-called painful arc. cryokinetics is to numb the injured part to the point of
Precautions concerning the use of cold spray include analgesia and then work toward achieving normal range of
protecting the face of the patient from the fumes and motion through controlled exercises. The exercises are
spraying the skin at an acute angle rather than at a continued until a sensation of pain returns, at which point
perpendicular angle. Cold spray is indicated when the injured part is iced again for several minutes, until it is
stretching of an injured part is desired along with cold numb, and exercise continues. This sequence is repeated
treatment. Treatment includes: four or five times. Cryokinetics is thought to inhibit painful
impulses, allowing the injured tissues to return to early
• The area to be treated should be sprayed and then activity, while simultaneously stimulating circulation,
stretched. through exercise [43].
• Spasm should be reduced.
• Treatment should be distal to proximal. Thermotherapy techniques
• A quick jetstream spray or stroking motion
should be used. Heat is still used as a universal treatment for pain and
• Coaling should be superficial, no frosting should discomfort. Much of the benefit is derived from the
treatment simply feeling good. However, in the early stages
occur
following injury, heat causes increased capillary blood
• Cold sprays may be used in conjunction with
pressure and increased cellular permeability; this results in
acupressure
additional swelling or edema accumulation. No person with
• Treatment time should be set according to body
edema should be treated with any heat modality until the
segment reasons for the edema are determined. It is in the best
interests of the therapist to use cryotherapy techniques or
Physiologic response contrast baths to reduce the edema prior to heat
applications. Superficial heat applications seem to feel
• Muscle spasm is reduced. more comfortable for complaints of the neck, back, low
• Golgi tendon organ response is facilitated. back, and pelvic areas and may be most appropriate for the
• Muscle spindle response is inhibited. athlete who exhibits some allergic response to scold
• Ligament and other musculoskeletal structures applications. However, it must be remembered that the
may be stimulated. tissues in these areas are absolutely no different from those
4. Contrast bath in the extremities. Thus, the same physiologic responses to
the use of heat or cold will be elicited all areas of the body
Contrast baths are used to treat subacute swelling, gravity- [43].
dependent Contras swelling, and vasodilatation -
vasoconstriction response. Treatment includes:

• Hot and cold immersions are alternated.


• Treatment time should heat least 20 minutes.
Treatments should consist of five 1-minute cold
immersions and five 3 warm immersions.
Although the exact ratio of cold to hot treatment
is highly variable.

Physiologic response:
• Vasoconstriction and vasodilatation occur.
• There is a reduction of necrotic cells at the
cellular level. Figure 37. Whirlpool [43].
• Edema is decreased.

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-43
Primary goals of thermotherapy include increased blood secretions in the respiratory system, opens the airways,
flow and increased muscle temperature to stimulate reduces resistance to respiration, regulates the vegetative
analgesia, increased nutrition to the cellular level, reduction system, relaxes, and can improve mental outlook. Children
of edema, and removal of metabolites and other products of can start to take saunas at two or three years of age [43].
the inflammatory process [43].
Full and partial immersion baths
Hydrotherapy
Various substances can be added to warm and rising
Hydrotherapy is the use of water in the treatment of temperature baths. The following are the different kinds of
disease. Hydrothermal therapy additionally uses its bath used:
temperature effects, as in hot baths, saunas, wraps, etc.
Rising temperature hip bath.
The recuperative and healing properties of hydrotherapy
are based on its mechanical and/or thermal effects. It This is taken in a tub filled with a hand's breadth of tepid
exploits the body's reaction to hot and cold stimuli, to the water. Hot water is then gradually added until the level
protracted application of heat, to pressure exerted by the reaches the navel. The final temperature should be 103-
water and to the sensation it gives. The nerves carry 104'F. Following this procedure, the patient is wrapped
impulses felt at the skin deeper into the body, where they warm and proceeds to bed. It should last 15-30 minutes, not
are instrumental in stimulating the immune system, more than three times per week [43].
influencing the production of stress hormones, invigorating
the circulation and digestion, encouraging blood flow, and Cold footbath.
lessening pain sensitivity. Generally, heat quiets and
soothes the body, slowing down the activity of internal The feet are placed into a footbath filled to calf depth with
organs. Cold, in contrast, stimulates and invigorates, cold water. Stop when a cold stimulus is felt or when the
increasing internal activity. If you are experiencing tense water is no longer perceived as being particularly cold.
muscles and anxiety from your stress, a hot shower or bath Stroke off excess water, dress, and walk or run until dry. A
is in order. If you are feeling tired and stressed out, you special form of this treatment is "walking in water," which
might want to try taking a warm shower or bath followed involves walking stork-like on a non-slip mat placed under
by a short, invigorating cold shower to help stimulate your the water. This type of treatment is best avoided by people
body and mind [43]. who suffer from cold feet, very high blood pressure, an
irritable bladder, urinary tract infection, diabetes, or
When you submerge yourself in a bath, a pool, or a vascular occlusion [43].
whirlpool, you experience a kind of weightlessness. Your
body is relieved from the constant pull of gravity. Water Rising temperature footbath, warm footbath.
also has a hydrostatic effect. It has a massage-like feeling
as the water gently kneads your body. Water, in motion, The feet are immersed in a footbath filled with water at
stimulates touch receptors on the skin, boosting blood body temperature. Hot water is gradually added to give a
circulation and releasing tight muscles [43]. final temperature of 103-104'F. In warm foot baths water of
this temperature is added straight away. Keep warm
Hydrotherapy and hydrothermal therapy are chiefly used to afterwards. The procedure should last 10-15 minutes and
tone up the body, to stimulate digestion, the circulation, and can be done daily. Best avoided by people with varicose
the immune system, and to bring relief from pain. veins, lymphostasis, or edema [43].
Description of indications are given under individual
method used. Water seems to have special powers in Cold arm bath.
getting rid of stress and rejuvenating our body. It affects the
skin and muscles. It calms the lungs, heart, stomach, and A basin is filled with cold water until it reaches a depth
endocrine system by stimulating nerve reflexes on the several inches above the immersed elbow. If the treatment
spinal cord [43]. becomes intolerable, stop and repeat as desired. Good of
headaches and sleeplessness. Best avoided by people with
Sauna and steam baths heart or circulatory problems [43].

Saunas and steam baths are similar in effect; the decision to Rising temperature arm bath.
take one rather than the other will be guided by personal
preference. In a sauna the heat acts more quickly to In principle, this is the same as the rising temperature foot
eliminate toxins through the skin, though some consider the bath. It should be followed by a cold arm douche, then by
moist air of a steam bath to have a more satisfying effect on half an hour's rest. Used for the treatment of bronchitis,
the respiratory system. Saunas are deeply relaxing and are a asthma, incipient respiratory infection, circulatory
great way to melt away stress. problems, angina pectoris [43].
Sitz bath.
A sauna is an eliminative procedure; it stimulates blood
flow, increases the heart rate, has an immune-modulating This is generally taken in a hip bath as a cold, rising
effect, promotes hormone production, encourages mucosal temperature, or warm sitz bath. Prior to a sitz bath, warm

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-44
the feet, e.g. through a warm foot bath. Parts of the body • Fill a thin cloth bag with 1/2 cup of herbs, either
not immersed in water should be covered. Cold sitz bath for placing it in the bath water or tying it to the
hemorrhoids or inflammation of the anus; warm or rising spigot so that the hot water runs through it as it
temperature sitz bath for difficulty in voiding the bladder, fills the tub. Again, soak for twenty to thirty
an irritable bladder, inflammation or infection of the minutes.
prostate, preparation for pregnancy. Do not use warm or
rising temperature sitz baths for hemorrhoids. A wrap is Certain herbs are quite effective for creating soothing
primarily used as a supportive measure for treating fever baths. Combine a handful each of valerian, lavender,
and local inflammation. The person reiving treatment linden, chamomile, hops, and burdock root, and add it to
should first adopt a relaxed position. Then a linen cloth is your bath according to one of the preceding methods. Soak
moistened with cold water (warm water for respiratory for thirty minutes in the tub. Another soothing herbal bath
diseases), well wrung out, and then wrapped tightly around calls for a handful each of hops, linden, valerian,
the appropriate part of the body, but not so tightly as to chamomile, yarrow, and passionflower. Prepare this bath
cause constriction. The moist linen cloth is in turn wrapped according to one of the preceding methods, or simmer the
with a dry cotton or linen cloth. The patient is then usually herbs in a quart of water, then drink ½ cup of the liquid
wrapped in a blanket or another cloth, and should rest for (with lemon and honey added, if you wish) and pour the
45-60 minutes or, if the intention is to induce sweating, for rest in the tub. While soaking in an herbal bath, you can
up to three hours. If the wrap is not felt to be warm after a read, meditate, listen to peaceful music, or just sit quietly,
quarter of an hour, heat should be applied in the form of a concentrating on relaxing yourself [43].
hot water bottle or by giving warm tea. The wrap should be
removed immediately if the person complains of feeling Importance of drinking sufficient water
unwell [43].
It is very important that we drink sufficient amount of
Warm packs. water in a day to make up for the water lost. The benefits of
drinking water are widely recognized. Drinking pure, fresh
A wrapping cloth is soaked in a hot infusion or decoction water is essential to our health and well-being. Steps for an
of herbs, then wrung out and applied to the patient's body. effective hydrotherapy:
Alternatively, the wrap may receive a coating of hot mud
mustard flour, or fango. As a further alternative, hayseed • For overall tension reduction, use a neutral bath
may be placed in a sack and steamed. Indications; painful (temperature between 92 to 94 °F) that is close to
chronic diseases such as arthrosis, renal disease, or cystitis, the skin temperature.
and for stimulating blood flow. Always check that the
temperature is tolerable before applying a wrap [43]. • Use water temperature between 102 to 106 °F for
loosening tight, tense muscles and reducing the
Cold packs. pain of stress-related conditions such as
backache. (Using temperatures higher than 106
Cooled cataplasm is spread onto the wrapping cloth and degrees is not recommended as it can raise your
placed on the part of the body. Crushed ice in a plastic bag body temperature very fast, inducing an artificial
may also be repeatedly applied for one minute, then fever.)
removed for four. Used for the treatment of various
inflammatory arthropathies, sprains and strains, pleurisy. • Take a cold shower after you step out of the bath.
Ice packs can also be used for headaches [43]. This brings and immediate rush of blood through
your system, as well as a rush of energy. (Try
Different forms of preparing a herbal bath alternating cold and hot shower to get a similar
effect. 3 minutes of hot water followed by 30
• Simmer 1/2 cup of herbs in 1 quart of water in a seconds of cold water and the 3 minutes of hot
covered pot for fifteen minutes. While the herbs water, etc.) Stay in the bath no more than 15 to
are simmering, take a short shower to cleanse 20 minutes. If you have high blood pressure or
your body, then fill the tub with hot or warm cardiovascular problems, don't stay long enough
water. Strain the liquid from the decoction into to raise your body temperature.
the bath water, and wrap the herbs in a
washcloth. Soak in the tub for at least twenty Paraffin baths
minutes, using the "herbal washcloth" to rub over
your body. - Paraffin bath is a simple and efficient, though somewhat
messy, technique for applying a fairly high degree of
• Add 1/2 cup of herbs to running bath water, localized heat. Paraffin treatments provide 6 times the
preferably hot. You might want to cover the drain amount of heat available in water because the mineral oil in
with a thin mesh screen to prevent the herbs from the paraffin lowers the melting point of the paraffin.
clogging the pipes. Soak in the tub for twenty to
thirty minutes. The risk of a burn with paraffin is substantial. The sports
therapist should weigh heavily the considerations between
a paraffin bath and warm whirlpool bath in the athletic

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-45
setting. The majority of paraffin baths are used for chronic increase in metabolic rates in a normal adult. These
arthritis patients’ hands and feet. If the patient has a chronic properties will increase blood flow sedate, decrease blood
hand or foot problem, the use of paraffin instead of water pressure, and promote healing by accelerating bio chemical
usually gives longer lasting pain relief [43]. reactions [43].

Physiologic responses Counter irritation, through mechanoreceptor and


thermoreceptor stimulation, reduces pain sensitivity, thus
• There is an increase in tissue temperature. permitting high temperatures with out painful heat
• Pain relief occurs. sensations. Pronounced hyperthermia accelerates the
• Thermal hyperthermia occurs. chemical metabolic processes and stimulates the normal
healing process. The high temperatures enhance tissue
Infrared lamps elasticity and reduce tissue viscosity, which improves
musculoskeletal mobility. Vascular responses are
When talking about infrared modalities, the therapist most stimulated as a result of the long-lasting hyperthermia and
typically thinks of the infrared lamp. The biggest advantage pressure fluctuations, resulting in increased blood flow to
of an infrared lamp is the superficial tissue temperature rise the injured area. Treatment includes:
and the fact that the unit does not touch the patient.
However, radiant heat is seldom used because it is limited • The patient must be positioned for comfort.
in depth of skin penetration to less than 1 mm. It is • The patient should place the body segment to be
recommended that a warm moist towel be placed over the treated (hand or foot) in the fluidotherapy unit.
body segment to be treated to enhance the heating effects. • Protective toweling must be placed at the umt
Dry towels should cover the remainder of the body not interface arid body segment.
being treated. These will allow a greater blood/tissue • Treatment time should be 15 to 20 minutes.
exchange by trapping the heat build-up in the moist towel
and reducing the stagnant air over the body segment [43]. Physiologic responses

Infrared generators may be divided into two categories: • Tissue temperature increases.
luminous and nonluminous. Nonluminous generators • Pain relief occurs.
consist of a spiral coil of resistant metal wire wound around • Thermal hyperthermia occurs.
a cone shaped piece of nonconducting material. The
resistance of the wire to the electric flow produces heat and Ultrasound
a dull red glow. A properly shaped reflector then radiates
the heat to the body. All incandescent bodies and tungsten Ultrasound has become an accepted modality in the
and carbon filament lamps are in the category of luminous treatment of injuries during the last three to four decades.
generators [43]. Its use continues to become more judicious as we learn
more about the healing process and the effect of ultrasound
Physiologic response on that process.

• A superficial rise in tissue temperature occurs. Ultrasound is truly a therapeutic modality that aids
• There is some decrease in pain. therapists in their attempts to return athletes to pre-injury
• Moisture and sweat appear on the skin surface. status. Before being ad ministered, however, it must be
preceded by an orthopedic evaluation to determine the
tissues involved the degree of injury, and the status of the
Fluid therapy healing process [43].

Fluid therapy is a unique, multifunctional physical Ultrasound must be respected for the significant effects it
medicine modality. The fluidotherapy unit is a dry heat produces on a variety of tissues and for its depth of
modality that uses a suspended air stream which has the penetration. In the hands of individuals who do not
properties of a liquid. Its therapeutic effectiveness in understand its effects and those who would ad minister it
rehabilitation and healing is based on its ability to before an appropriate musculoskeletal examination, it is a
simultaneously apply heat, massage, sensory stimulation modality that could be more of a liability than a benefit
for desensitization, levitation, and pressure oscillations. [43].
Unlike water, the dry, natural medium does not irritate the
skin or produce thermal shocks. This allows for much Ultrasound falls under the classification of the acoustic
higher treatment temperatures than with aqueous or spectrum rather than the electromagnetic spectrum. Sound
paraffin heat transfer. The pressure oscillations may waves that are at a higher frequency than are detectable by
actually minimize edema, even at very high treatment the human ear are termed ultrasound. Therefore frequencies
temperatures. Outstanding clinical success has been above 16,000 to 20000 Hz are defined as ultrasound.
reported in treatment of pain, range of motion, wounds, Standard therapeutic units used on the musculoskeletal
acute injuries, swelling, and blood flow insufficiency. system operate at a frequency between 800000 and
Fluidotherapy treatment of the hand at 115 °F (46.2 °C) 1,000,000 Hz. The major purpose for using ultrasound is in
results in a sixfold increase in blood flow and a fourfold the treatment of soft tissue. The propagation of ultrasonic

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-46
energy in biologic tissue depends on the absorption the extensibility of tight capsular tissue and in the mobility
characteristics of the tissue, as well as on the density of the of mature scar tissue have been proven with ultrasound and
tissue [43]. exercise [43].

1. Anatomy of the equipment Ultrasound produces the desirable therapeutic effects of


any deep heat modality. The effect of ultrasound that may
Ultrasound machines are not as mechanically and be the most distinguish able is its ability to selectively
electrically complex as many other therapeutic modalities. increase temperature in local, well circum scribed areas
A high frequency generator provides electricity via a [43].
coaxial cable to the hot electrode in the transducer. The
transducer contains a crystal made of quartz or a synthetic 3. Under water
ceramic. The sound head also includes an insulator and a
metal front plate that serves as the surface of the applicator. Ultrasound can be used effectively under water to treat
The unit also contains a timer directly inked to the on-off irregular surfaces such as a hand. The sound head need not
switch and either a digital display or meter that shows the be in contact with the body part and is moved in a circular
output of the total number of watts and the watts per square fashion. The applicator should be approximately ½ inch
centimeter of the surface of the sound head from the body part and the intensity setting is the same as
for regular application. Bubble formation sometimes occurs
2. Effects over the head as a result of gaseous separation and needs to
be wiped of by the person administering the treatment. This
Physiologic effects that occur because of heating of tissues degassing process is done by simply rubbing the bubbles
are referred to as thermal. Many of the effects of ultrasound off with the fingers and proceeding with the treatment [43].
are attributed to heat. It is generally accepted that heat
produces the following results: 4. Pulsed ultrasound

• Increases the extensibility of collagen tissue The theory behind the development and use of pulsed
• Decreases joint stiffness ultrasound is that the mechanical effects can be produced
• Reduces pain without thermal effects. The pulsed beams, as compared to
• Reduces muscle spasm continuous beams, allow for a rest period for cooling
• Assists in mobilizing inflammatory infiltrates, purposes [43].
edema, and exudates
• Increases blood flow 5. Simultaneous ultrasound and electrical stimulation
• Increases local metabolism
Machines are available to incorporate the use of ultrasound
with either low-voltage alternating current or high-voltage
Effects at the cellular level that are the result of the
direct current. In these units, the transducer becomes an
vibratory motion are referred to as mechanical effects.
Acceleration in the rate of diffusion across membranes has, active electrode that causes muscle stimulation as it is
passed over an area- This technique is used most
however, been shown to be a result of mechanical
commonly on spasmodic areas that would benefit from
vibration. This results in increased blood flow and
both modalities [43].
consequently absorption of the by-products of the healing
process, thus facilitating that process. This in itself is
significant to consider when an injury is in an acute or 6. Coupling agents
subacute phase [43].
Because sound waves at the frequencies emanating from
When energy is added to the body in the form of sound the machine do not readily travel through au, specific
coupling mediums are necessary. The coupling medium
waves, it is absorbed transmitted, reflected, or refracted.
must have good lubricating qualities so that the applicator
Tissues of high water content have a low rate of absorption
can glide against the skin on even very hairy
(thereby a high rate of penetration) while tissues of high
protein content have a high rate of absorption. Fat has a
relatively low absorption rate and muscle absorbs
considerably more. Bone that is rather superficial absorbs
more ultrasonic energy than any of the other tissues. This is
thought to be because of the high protein content and
density of bone compared to other tissue. Peripheral nerve
absorbs at a rate approximately twice that of muscle. It has
been concluded that soft tissue in close proximity to bone
receives more of the effects of ultrasound because of a
rebound phenomenon. Therefore, joint structures such as
the periosteum, capsule, tendon, and extra capsular
ligament are significantly affected by ultrasound treatment.
It has been demonstrated that ultrasound raises the
Figure 38. Ultrasound machine [43].
temperature in the collagen tissues of tendon. Increases in

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-47
individuals with out significant friction. There are many
commercial gels on the market to serve as this medium.
Studies have been done that showed no significant
difference between the common commercial gels, mineral
oil, and de gassed water [43].

Interferential therapy

The basic principle of Interferential Therapy is to utilizes


the strong physiological effects of low frequency electrical
stimulation of muscle and nerve tissues without the
associated painful and somewhat unpleasant side effects of
such stimulation. To produce low frequency effects at
sufficient intensity at depth, most patients experience
considerable discomfort in the superficial tissues. This is
due to the resistance (impedance) of the skin being
inversely proportional to the frequency of the stimulation.
The skin impedance at 50Hz is approximately 3200W with
4000Hz it is reduced to approximately 40W. The result of
applying this latter frequency is that it will pass more easily
through the skin, requiring less electrical energy input to
reach the deeper tissues & giving rise to less discomfort
[43].

Interferential therapy utilizes two of these medium


Figure 39. Ultrasound doppler machine [43] frequency currents, passed through the tissues
simultaneously, where they are set up so that their paths
cross. This interference gives rise to an interference or beat
frequency which has the characteristics of low frequency
stimulation [43].

The exact frequency of the resultant beat frequency can be


controlled by the input frequencies. By careful
manipulation of the input currents it is possible to achieve
any beat frequency that you might wish to use clinically.
Modern machines usually offer frequencies of 1-150Hz,
though some offer a choice of up to 250Hz or more.

The use of 2 pole I/F stimulation, where there is clearly no


interference within the body, it is made possible by
electronic manipulation of the currents - the interference
occurs within the machine. The magnitude of the low
frequency interference current is approximately equivalent
to the sum of the input amplitudes [43].

1. Physiological effects:

Excitable tissues can be stimulated by low frequency


alternating currents. Although to some extent, all tissues in
this category will be affected by a broad range of
stimulations, it is thought (Savage 1984) that different
tissues will have an optimal stimulation band, which can be
estimated by the conduction velocity of the tissue, its
latency and refractory period. These are detailed below:

• Sympathetic Nerve 1-5Hz.


• Parasympathetic Nerve 10-150Hz.
• Motor Nerve 10-50Hz.
• Sensory Nerve 90-100Hz.
Figure 40. Abdominal ultrasound machine [43]

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-48
• Nociceptive fibers 90-150Hz (130Hz There is no primary nervous control of oedema
specific). reabsorption and the direct electrical stimulation of blood
• Smooth Muscle 0-10Hz. flow is limited in its effectiveness. It is suggested therefore
that in order to achieve these effects, suitable combinations
The clinical application of I/F therapy can be based of muscle stimulation can be made [43].
logically on this data together with knowledge of
physiological behavior of stimulated tissue. Selection of a 4. Treatment:
wide treatment band can be considered less efficient than a
smaller selective band in that by treating with a frequency Electrode positioning should ensure adequate coverage of
range of say 1-100Hz, the appropriate treatment the area for stimulation. In some circumstances, a bipolar
frequencies can be covered, but only for a relatively small method is preferable if a longitudinal zone requires
percentage of the total treatment time. Additionally, some stimulation rather than an isolated tissue area. Placement of
parts of the range might be counterproductive for the the electrodes should be such that a crossover effect is
primary aims of the treatment [43]. achieved in the desired area. If the electrodes are not placed
so that a crossover is achieved, the physiological effects of
They are 4 main clinical applications for which I/F appears I/F can not be achieved [43].
to be used:
Nerves will accommodate to a constant signal and a sweep
• Pain relief. (or gradually changing frequency) is often used to
• Muscle stimulation. overcome this problem (as well as generating a range of
• Increased blood flow. effects). The sweep (range) should be appropriate to the
• Reduction of oedema. desired physiological effects, though again it is suggested
that an excessive range may minimize the clinical effect
In addition, claims are made for its role in stimulating [43].
healing and repair.
The mode of delivery of the selected sweep varies with
As I/F acts primarily on the excitable tissues, the strongest machines. The most common application is the 6 second
effects are likely to be those which are a direct result of rise and fall between the pre-set frequencies. For example,
such stimulation (i.e. pain relief and muscle stimulation). if a 10 - 25Hz range has been selected, the machine will
The other effects are more likely to be secondary deliver a changing frequency, starting at 10Hz, rising to
consequences of these [43]. 25Hz over a 6 second period. Once this upper limit has
been achieved, the frequency will once again fall, over a 6
2. Pain relief: second period to its starting point at 10Hz. This pattern is
repeated throughout the treatment session [43].
Electrical stimulation for pain relief has widespread clinical
use, thought the direct research evidence for the use of I/F Other patterns of sweep can be produced on many
in this role is limited. Logically one could use the higher machines as illustrated:
frequencies (90-150Hz) to stimulate the pain gate
mechanisms & thereby mask the pain symptoms. Treatment times vary widely according to the usual clinical
Alternatively, stimulation with lower frequencies (1-5Hz) parameters of acute/chronic conditions and the type of
can be used to activate the opoid mechanisms, again physiological effect desired. In acute conditions, shorter
providing a degree of relief. These two different modes of treatment times of 5-10 minutes may be sufficient to
action can be explained physiologically & will have achieve the effect. In other circumstances, it may be
different latent periods and varying duration of effect. It necessary to stimulate the tissues for 20-30 minutes. It is
remains possible that relief of pain may be achieved by suggested that short treatment times are initially adopted
stimulation of the reticular formation at frequencies of 10- especially with the acute case in case of symptom
25Hz or by blocking C fiber transmission at >50Hz [43]. exacerbation. These can be progressed if the aim has not
been achieved and no untoward side effects have been
3. Muscle stimulation: produced. There is no research evidence to support the
continuous progression of a treatment dose in order to
Stimulation of the motor nerves can be achieved with a increase or maintain its effect [43].
wide range of frequencies. Clearly, stimulation at low
frequency will result in a series of twitches, whist INTENSIVE CARE UNIT
stimulation at 50Hz will result in a tetanic contraction. The
choice of treatment parameters will depend on the desired Although many promising objective methods (measuring
effect, but to combine muscle stimulation with an increase systems) are available, there are no truly validated
in blood flow and a possible reduction in oedema, there is instruments for monitoring intensive care unit (ICU)
some logic in selecting a range which does not involve sedation. Auditory evoked potentials can be used only for
strong sustained tetanic contraction and a sweep of 10- research in patients with a deep level of sedation. Other
25Hz is often used [43]. measuring systems require further development and
validation to be useful in the ICU. Continuing research will
provide an objective system to improve the monitoring and

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-49
controlling of this essential treatment for ICU patients. 10. http://imaginis . com / endoscopy/
Subjective methods (scoring systems) that are based on
clinical observation have proven their usefulness in guiding 11. http:// my. webmd. com/ hw/ health_guide_atoz/
sedative therapy. The Glasgow Coma Score modified by Web MD Health
Cook and Palma (GCSC) achieves good face validity and
reliability, which assures its clinical utility for routine 12. http:// oac.med.jh mi.edu/res_phys /E ncyclopedia/
practice and research. Other scales, in particular the Body Plethysmography
Ramsay Scale, can be recommended preferably for clinical
use. An accurate use of available instruments can improve 13. http:// umjmh .org/ Health Library/ Respiratory/
the sedative treatment that we deliver to our patients [35]. Incentive Spirometer

CONCLUSIONS 14. http://users.rcn.com /jkimball.ma.ultranet/


The Human Eye. 13 June 2004
We had the opportunity to increase our knowledge about
the role of fluids in all different system of the human body 15. http://www4 .tpgi.com.au/ users/ amc gann/body/
and what instruments were used to measure it. Now we can The Nervous System. 14 Jun 2004
mention instruments for measuring some fluids of the
human body easily. For example, blood pressure, with the 16. http://www .aafp.org/afp/ 20020915 /991. html
sphygmomanometer and the Schirmer strip, which American Family Physician
measures the quantity of total tear secretion. Also, we learn
about how system in human body function and it parts. 17 .http://www.agingeye.net/dryeyes/dryeeyeseyeexam.pp
This is good for us because now we understand better how Agind Eye Times
the human body function and we can prevent sickness or
just now what the doctor is doing when he is taking our 18. http://www.cfdrc.com/applications/biomedical/pumps.
pressure. After study all this we will check our self Blood Pumps
frequently to see how our system are working.
19. http://www. cncmagazine. com/v6i22/v6i22i-.htm.
ACKNOWLEDGEMENTS CNC Machine Magazine

This study was aided in part by the teachings of Dr. Megh 20. http://www .clevelandclinic org/ health/ health-
R. Goyal, University of Puerto Rico at Mayaguez. How To Use An Incentive Spirometer

REFERENCES 21 http://www .dnacounter. com/ dnacounter/dnac_site/


Endoscope Products
1. Baker, Geddes. 1968, Principles of Applied Biomedical
Instrumentation. 2nd ed. John Wiley & Sons. 22. http://www .eyecareamerica. org/eyecare /museum/
Eye care america.
2. Bronzion, Joseph D. 1986, Biomedical Engineering and
Instrumentation Basic Concepts and Applications. PWS 23. http:// www. Futuremedamerica .com/ pdf/sv3
Publishers.
24 .http://www. glenbrook.k12.il.us /gbssci/ phys/
3. Cobbold, Richard S. C. 1974, Transducers for Sound Properties and their Perception.1996 : The
Biomedical Measurements. New York: Wiley Human Ear. 13 June 2004.
Interscience.
25 http://www. grogans. com/research.jsp?documentid
4. Webster, John G. 1998, Medical Instrumentation. 3rd Introducing the Welch Allyn PanOptic
ed. NY: John Wiley & Sons. Ophthalmoscope

5. http://biology.clc.uc.edu/courses/bio105/ endocrin.htm. 26 .http://www .guldenindustries .com/


Burps, Farts and Snot." 2004. 13 Jun 2004 Gulden Ophthalmics

6. http://biology.clc.uc.edu/courses/bio105/ reproduc.htm. 27 .http://www. hamiltoncomp. com/ product/syringe/


Carter, Stein . "Reproductive System." 1996. 13 June
2004 28 http://www. kidney.org.au/ renal_resources/ html
Haemodialysis: A Treatment Option
7. http:// classes.aces. uiuc.edu/ AnSci103/needles. html
29. http://www.lsc.org/online_science/gross/gross_urinary
8. http:// ct.essortment.com / humancirculat_rdgf.htm The Urinary System. 13 June 2004.
How the Human Circulatory System Works. 2002.
30 http://www. medikro. com/ medikro/ pulmonary/
9. http:// digestive.niddk.nih.go v/ddiseases /pubs/yrdd/.
Your Digestive System and How it Works. 2004. 31. http://www. medisave. co. uk/datasheets/diagnostic

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-50
32. http://www. medused. com/ Product Page.asp Adrenal glands- Paired endocrine glands, one located just
superior to each kidney; secretes
33. http://www. mercksource. com/
Afferent- Conduction of a nerve impulse toward an organ.
34 http://www .msf .org
Afferent- Leading toward some point of reference.
35 http://www. ncbi. nlm. nih. gov/ entrez /query.fcgi Compare with efferent.

36. http:www .nidek .com /nt.html Alveolus- An air sac of the lung through which the gas
Ophthalmic & Diagnostic Equipment exchange with the blood takes place.

37. .http://www .nlm.nih.gov/ medlineplus/ency/article/ Amplitude- The intensity of current flow as indicated by
Medline Plus the height of the waveform from baseline.

38. .http://www .opt .indiana .edu/ riley/ Analgesia- Loss of sensibility to pain.
HomePage/NonContact/TEXT_Non_Contact.html
Anesthesia- Loss of sensation.
39. .http://www .optomed. com/ pd_ha2. cfm HA-2
Anode- Positively charged electrode in a direct cur rent
Hand Held Applanation Tonometer with Prism
system.
40. http://www. thebrowns23.f reeserve. co. uk/
Anus: The distal end and outlet of the digestive tract.
entries/APPLANATION_TONOMETER.html
Applanation Tanometer
Aorta. The largest and main systemic artery of the body;
arises from the left ventricle and branches to distribute
41. http://www.v eatchinstruments .com/ perkins.htm
blood to all parts of the body except the lungs.
Clement Clarke Perkins MK2 Applanation Tonometer
Aortic valve: Outlet valve from left ventricle to aorta.
42 .http://www .vitalograph.com /default.html?site_files/
Main Frame
Applicator- The electrode used to transfer energy in
microwave diathermy.
43 Goyal, M.R. 2001, First Congress on Biofluids at
UPR-RUM.
Arrhythmia: An alteration in rhythm of the heartbeat
either in time or force.
GLOSSARY
Artery: A thick walled blood vessel that carries away from
Ablution- Act of washing or bathing. heart chamber and forward the body organs. Compared
with veins.
Absorption- Energy that stimulates a particular tissue to
perform its normal function. Atherosclerosis: A form of arteriosclerosis, is the
reduction in blood flow through the arteries caused by
Absorption- The movement of nutrients and other greasy deposits called plaque that form on the insides of
substances through the wall of the digestive tract and into arteries and partially restrict the flow of blood.
the blood.
Atonic- Without tone.
Accommodation- Adaptation by the sensory receptors to
various stimuli over an extended period of time. Atrioventicular valve: A valve between each atrium and
its ventricle that prevents back flow of blood. The right
Acoustic-spectrum- The range of frequencies and Atrioventricular valve is the tricuspid valve; the left
wavelengths of sound waves. atrioventricular valve is the mitral valve.

Action potential- A recorded change in electrical potential Atrioventricular node: Mass of specialized cardiac tissues
between the inside and outside of a nerve cell, resulting in that receive and impulse from sinoatrial node (peacemaker)
muscular contraction. and conducts it to the ventricles.

Active electrode- Electrode at which greatest current Atrium: An anatomical cavity or passage; especially a
density occurs. main chamber of the heart into which blood returns from
circulation.
Adrenal cortex- The outer region of each adrenal gland;
secretes steroid hormones. Bile: The fluid secreted by the liver; emulsifies fats.

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-51
Blood pressure: The force exerted by blood against the
inner walls of the blood vessel. Cerebellum: A convoluted subdivision of the brain
concerned with the coordination of muscular movements,
Blood: A fluid, circulating connective tissue that transport muscle tone, and balance.
nutrients and others materials through the body.
Cerebrospinal Fluid (CFS): The fluid that bathes the
Bowman’s capsule: Doubled wall sac of cells that central nervous system.
surrounds the glomerulus of each nephron.
Brachial: relating to the arm or a comparable process. Cerebrum: Large, convoluted subdivision of the brain; it
functions as the center for learning, voluntary movement,
Brain stem: The part of the brain that includes the and interpretation of sensation.
medulla, pons, midbrain, thalamus and hypothalamus.
Circuit- The path of current from a generating source
Brain: A concentration of nervous tissue that control through the various components back to the generating
neural function. Enlarge portion of the central nervous source.
system.
Circulatory system: The body system that functions in
Bronchiole: Air duct in the lung that branches from a internal transport and protects the body from disease.
bronchus; divides to form air sacs Alveoli).
Clitoris: A small, erectile structure at the anterior part of
Bronchus: One of the branches of the trachea and its the vulva in females; homologous to the male penis.
immediate branches within the lung.
Coaxial cable- Heavy, well-insulated electrical wire.
Bundle of his: A small band of cardiac muscle fibers
transmitting the waves of depolarization from the atria to Cochlea: The structure of the inner ear that contains the
the ventricles during cardiac contraction. auditory receptors (organ of Corti).

Cable electrodes- An inductance type electrode in which Cold-induced vasodilatation- Vasodilatation following
the electrodes are coiled around a body part creating an cold application.
electromagnetic field.
Collagen tissue- Fibrous insoluble protein found in
Cannula: A small tube for insertion into a body cavity or connective tissue, bone, ligaments, and cartilage.
blood vessel.
Collecting duct: A duct in the kidney that receives filtrate
Capillaries: Microscope blood vessels in the tissues the from several nephrones and conducts it to the renal pelvis.
permit exchange of materials between cells and blood.
Colloid- The fluid suspension of the body’s intercellular
Cardiac arrest: Standstill of the heart, its action and fluid.
diseases.
Colon: Portion of the large intestine between the cecum
Cardiac cycle: One complete heart beat. and the rectum.

Cardiac output: The volume of blood pumped by the heart Condenser electrodes- An electrical current is con ducted
per unit of time. back and forth between the two electrodes. Highest
concentration is under the electrodes, which may be pads or
Cardiac: Pertaining to the heart. space plates. Highest concentration is also in fat tissue

Cardiology: The study of the heart, its action and diseases. Conduction- Heat loss or gain through direct contact.

Cardiovascular disease: Diseases of the heart or blood Cone: One of the conical photoreceptive cells of the retina
vessels. that is particularly sensitive to bright light and, by
distinguishing light of various wavelengths, mediates color
Carpal tunnel syndrome: A painful and disabling disorder vision. Compare with rod.
characterized by inflammation and swelling in the tendons
that run through the narrow carpal tunnel in the wrist, and Congestion- Presence of an abnormal amount of blood in
one of the most common of repetitive motion injuries. the vessels as a result of an increase in blood flow or
obstructed venous return.
Cathode- Negatively charged electrode in a direct current
system. Consensual heat vasodilatation- Increased blood flow that
spreads to a remote area of the body as a result of localized
Central Nervous System (CNS): The brain and spinal heating.
cord.

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-52
Contraindication- Special symptom or circumstance that
renders the use of a remedy or procedure in advisable. Direct current- Galvanic current that always flows in the
same direction and may flow in either a positive or a
Contrast bath- Hot (106° F) and cold (50° F) treatments in negative direction.
a combined sequence to stimulate superficial capillary
vasodilatation or vasoconstriction. Distal convoluted tube: The part of the renal tubule that
extends from the loop of Henle to the collecting duct.
Convection- Heat loss or gain through the movement of air
or water molecules across the skin. Douche- A current of water directed against the skin
surface (e.g., scotch douche alternating hot and cold water).
Cornea: Transparent covering of the eye.
Duodenum: The portion of the small intestine into which
Coronary artery and sinus: Vessels carrying blood to and the contents of the stomach first enter.
from the walls of the heart itself.
ECG: Abbreviation for electrocardiogram.
Corpus collosum: A large bundle of nerve fiber
interconnecting the two cerebral hemispheres. Echocardiogram: An ultrasonic record of the dimension
and movement of the hearts and his valves.
Corpus luteum: The temporary endocrine tissue in the
ovary that develops from the ruptured follicle after Edema- Excessive fluid in cells.
ovulation.
EEG: Abbreviation for electroencephalogram.
Cortical: Relating to, or consisting of the cortex.
Efferent: Leading away from point of reference.
Coupling agent- A substance used as a medium for the
transfer of sound waves. Ejaculation: A sudden expulsion, as in the ejection of
semen from the penis.
Cranial nerves: Ten to twelve pairs of nerves that emerge
directly from the brain. Electrical field- A technique of heating the tissues in
shortwave diathermy in which the patient is part of the
Cranium: The body framework that protects the brain. electrical circuit.

Cryokinetics- The use of cold and exercise in the treatment Electrical potential- The difference between charged
of pathology or disease. particles at a higher and lower potential.

Cryotherapy- The use of cold in the treatment of Electrical stimulation- The use of four or more electrodes.
pathology or disease. AA’ and BB’ arranged so the path of electric current A to
A’ is crossed by the path of the electric current B to B’.
Crystal- The part of the ultrasound head that vibrates and
changes shape. Electrocardiogram: A record of the electric activity of the
heart.
Dendrite: A branch of a neuron that receives and conducts
nerve impulses toward the cell body. Electrocardiograph: An instrument used for the
measurement of the electrical activity of the heart.
Depolarization- Process or act of neutralizing the cell
membrane’s resting potential. Electroencephalograph: An instrument for measuring and
recording electrical activities from the brain (brain waves).
Diastole: Phase of the cardiac cycle in which the heart is Electromagnetic spectrum- The range of frequencies and
relaxed. wavelengths associated with radiant energy.

Diastolic: Of or pertaining to the diastole. Electromyography- The pick-up and amplification of


electrical signals generated by the muscle as it contracts.
Diathermy- The application of high-frequency electrical
energy that is used to generate heat in body tissues as a Electrostatic or condenser field- The patient is placed
result of the resistance of the tissue to the passage of between electrodes and becomes a part of a series circuit.
energy.
Elimination: Ejection of undigested food from the body.
Diffusion: Net movement of particles from region of
higher concentration to a region of lower concentration, Endocrine gland: Gland that secretes hormones directly
resulting from random motion. into the blood or tissue fluid instead of into a duct.

Digestion: The breakdown of food into smaller molecules.

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-53
Endocrine system: Body system that helps regulate between 2E-4 and 0.0012 mm3 (formerly called maximum
metabolic activities; consists of ductless glands and tissues expiratory flow rate).
that secretes hormones
Forced expiratory volume: The volume of gas exhaled
Endorphins- Endogenous opiates whose actions have over a given time interval during the performance of a
analgesic properties. They are neurophor mones and not forced vital capacity.
neurotransrnitters (i.e., β-endor- phins).
Forced midexpiratory flow: The average rate of flow
Enkephalin- Neurotransmitter proteins that are pain- during the middle half of the forced expiratory volume.
relieving molecules. They block the passage of noxious
stimuli by servicing descending neurons to counter Forced vital capacity: The maximum volume of gas that
ascending signals. They inhibit the release of substance P can be expelled as forcefully and rapidly as possible after
and are produced by enkephalinergic neurons. maximum inspiration.

Enkephalinergic neurons- Neurons with short axons that Frequency- The number of cycles or pulses per second.
release enkephalin. They act as interneUrofls (internuncial
neurons) and are found in the substantia gelatinosa, nucleus Functional residual capacity: The volume of gas
raphae magnus, and periaqueductal gray matter. remaining in the lungs at the resisting expiratory level. The
resting end-expiratory level is used as the baseline because
Enzyme: An organic catalyst that accelerates a specific it varies less than the end- inspiratory level.
chemical reaction by lowering the activation energy
required for the reaction. Gallbladder: A small sac that stores bile.

Epididymis: A coiled tube that receives sperm from the Gamma system- Nerve fibers that reset the muscle spindle
testis and conveys it to the vas deferent. to its adjusted length.

Epiglottis: A thin, flexible structure that guards the Ganglion: A mass of neuron cell bodies.
entrance to the larynx, preventing food from entering the
airway during swallowing. Gland: See endocrine gland and exocrine gland.

Epinephrine: Hormone produced by the adrenal gland. Glomerulus: The cluster of capillaries at the proximal end
of the nephron; the glomerulus is surrounded by the
Eustachian tube: The auditory tube passing between the Bowman’s capsule.
middle ear cavity and the pharynx; permits the equalization
of pressure on the tympanic membrane. Gray matter: Nervous tissue in the brain and spinal cord
that contains the cell bodies, dendrite, and unmyelinated
Excretion: The discharge from the body from a waste axons.
product of metabolism (Do not confuse with the
elimination of undigested food materials). Ground- A wire that makes an electrical connection with
the earth.
Exocrine gland: Gland that excretes its product through a
duct that opens onto a free surface. Heart block: A delay or interference of the conduction
mechanism whereby impulses do not go through all or a
Expiratory reserve volume: The volume capable of being major part of the myocardium.
expired at the end-expiratory level of a quiet expiration.
Hemoglobin: The red iron- containing protein pigment in
External respiration: Movement of gases in and out of the blood that transport oxygen and carbon dioxide and aids
lungs. in regulation of pH.
Fallopian tube: Tube that carries ova from the ovary to the Heparin: An acid occurring in tissues, mostly in the liver.
utter. It can be produced chemically and can make the blood
incoaguable if injected into the blood stream intravenously.
Fluidotherapy- A modality of dry heat using a finely
divided solid suspended in an air stream with the properties Hepatic: Pertaining to the liver.
of liquid.
Hertz- A unit of frequency equal to one cycle per second.
Fluoroscopy: Process of using an instrument to observe the
internal structure of an opaque object (as the living body) High-voltage current- Current in which the wave form has
by means of X rays. an amplitude of greater than 150 volts with a relatively
short pulse duration of less than 100 sec.
Forced expiratory flow: The average rate of flow for a
specified portion of the forced expiratory volume, usually Hormone: An organic chemical messenger that is
produced in one part of the body and transported to another

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-54
part where it signals cells to alter some aspect of Iuxtaglomerular apparatus: A structure in the kidney that
metabolism. secretes rennin in response to a decrease in blood pressure.

Hydrocollator- A synthetic hot (1700 F) or cold (00 F) gel Kidney: The paired organ important in excretion of
used as an adjunctive modality to stimulate tissue metabolic waste and in osmoregulation.
temperature rise or tissue temperature lowering.
Korotkoff sounds: Sounds produced by sudden pulsation
Hydrocortisone- An antinflamatory steroid. of blood being forced through a partially occluded artery
and heard during auscultatory.
Hydrogymnastics- Exercises using the buoyant properties blood pressure determination.
of immersion in water.
Large intestine: The portion of the digestive tract
Hydrostatic pressure- The pressure exerted by a liquid at consisting of the cecum, colon, rectum and anus.
rest.
Larynx: The organ at the upper end of the trachea that
Hydrotherapy- Cryotherapy and thermotherapy techniques contains the vocal cords.
that use water as the medium for heat transfer.
Lens: The oval, transparent structure located behind the iris
Hyperstimulation analgesia- See stimulus-produced of the eye; bends incoming light rays and brings them to a
analgesia. focus on the retina.

Hypertension: The elevated blood pressure develops when Liver: A large, complex organ that secretes bile, helps
the body’s blood vessels narrow, causing the heart to pump maintain homeostasis by removing or adding nutrients to
harder than normal to push blood through the narrowed the blood, and performs many other metabolic functions.
openings.
Low-voltage current- Current in which the waveform has
Hyperventilation: Abnormally prolonged, rapid deep an amplitude of less than 150 volts.
breathing or over breathing.
Lumen: The cavity of a tubular organ or instrument.
Hypothalamus: Part of the brain that regulates the
pituitary gland, the autonomic system, emotional responses, Lung capacity: The amount of gas contained n the lung at
body temperature, water balance, and appetite; located the end of maximal inspiration.
below the thalamus.
Lung: An internal respiratory organ that functions in gas
Impedance- Resistance of the tissue to the passage of exchange; enables a person to breathe air.
electrical current.
Magnetic field- A technique of heating the tissues in
Indifferent or dispersive electrode- Large electrode used shortwave diathermy in which the patient is not part of the
to spread out electrical charge. electrical circuit.

Induction electrodes- Electrical current is passed through Maximal voluntary ventilation: The volume of air that a
a coil that in turn gives off eddy currents of electromagnetic subject can breathes with maximal effort over a given time
energy. This energy is absorbed by the tissues and heating interval.
occurs as a result of the resistance of the tissues.
Mechanical effects- Ultrasonic effects that involve
Inferior vena cava: Main vein feeding back to the heart movement as a result of vibratory motion.
from systemic circulation below the heart.
Infrared- The portion of the electromagnetic spectrum Membrane: A thin layer of tissue that covers a surface or
associated with thermal changes; located adjacent to the red divides a space or organ.
portion of the visible light spectrum. That part of the
electromagnetic spectrum dealing with infrared Meninges: The three membranes that envelop the brain and
wavelengths. spinal cord: the dura madre, arachnoid, and pia madre.

Inspiratory capacity: The maximal volume of gas that can Microshock- An electrical shock that is imperceptible
be inspired from the resting expiratory level. because of a leakage of current of less than 1 mamp.

Inspiratory reserve volume: Maximal volume of gas that Mitral valve: See atrioventricular valve.
can be inspired from the end-inspiratory position.
Modulation- Refers to any alteration in the magnitude or
Iris: The pigmented portion of the eye. any variation in the duration of an electrical current.

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-55
Myelin sheath: The white fatty material that forms a
sheath around the axons of certain nerve cells, which are Pacemaker (of the heart): See sinoatrial node.
then called myelinated nerve fibers.
Pad electrodes- Capacitor type electrode used with
Nephron: The functional, microscopic unit of the kidney. shortwave diathermy.

Nerve: A bundle of axons or dendrites wrapped in Pancreas: Large gland located in the abdominal cavity.
connective tissue that conveys impulses between the central The pancreas produces pancreatic juice containing
nervous system and some other part of the body. digestive enzymes; also serves as an endocrine gland;
secreting the hormone insulin and glucagons.
Nervous tissue: a type of tissue specialized for conducting
electrochemical impulses. Paraffin bath- A combined paraffin and mineral oil
immersion technique in which the paraffin sub stance is
Neuron: A nerve cell; a conducting cell of the nervous heated to 126° F for conductive heat gains; commonly used
system that typically consists of a cell body, dendrites, and on the hands and feet for distal temperature gains in blood
an axon. flow and temperature.

Neurotransmitter: A chemical messenger used by neurons Parathyroid gland: Small, pea- sized glands closely
to transmit impulses across the synapse. adjacent to the thyroid gland; their secretion regulates
calcium and phosphate metabolism.
Norepinephrine- A neurotransmitter that may enhance
pain. When it is inhibited, analgesia is in creased. Increased Parathyroid hormone: Hormone secreted by the
levels in the central nervous system decrease analgesia. parathyroid glands; regulates calcium and phosphate
metabolism.
Norepinephrine: A neurotransmitter that is also a hormone
secreted by the adrenal medulla. Pharynx: Part of the digestive tract. It is bounded
anteriorly by the mouth and nasal cavities and posteriorly
Occipital lobes: Posterior areas of the cerebrum; interpret by the esophagus and larynx; the throat region.
visual stimuli from the retina of the eye.
Plethysmography: The recording of the changes in the
Ohm’s law- The current in an electrical circuit is directly volume of a body part as modified by the circulation of the
proportional to the voltage and inversely proportional to the blood in it.
resistance.
Pulmonary circulation: The part of the circulation system
Organ of Corti: The structure within the inner ear that that delivers blood to and from the lungs for oxygenation.
contains receptor cells that sense sound vibrations.
Pulmonary minute volume (pulmonary ventilation):
Organ system: Body system; an organized group of tissues Volume of air respired per minute= tidal volume*
and organs that work together to perform a specialized set breathes/min.
of functions.
Pulsed ultrasound- Method of administering ultra-
Osmotic pressure- The pressure for movement of a solvent
across a membrane that is permeable to that solvent. Radiation- The process of emitting energy from some
source in the form of waves. A method of heat transfer
Otoliths: Small calcium carbonate crystals in the saccule through which heat can either be gained or lost.
and utricule of the inner ear; sense gravity and are
important in static equilibrium. Radiograph- Record produced on a photographic plate,
film, or paper by the action of roentgen rays or radium;
Ovary: One of the paired female gonads responsible for specifically x-rays.
producing eggs and sex hormones.
Repolarize- To return to a polarized state after a de
Ovulation: The release of a ovum from the ovary. polarizing event.

Ovum: female gamete. Residual volume: Air left in the lungs after deep exhale
(about 0.0012m3).
Oxygen dissociation curve: A curve depicting the
percentage saturation of hemoglobin with oxygen, as a Retina: The innermost of the three layers of the eyeball,
function of certain variables such as oxygen concentration, which is continuous with the optic nerve and contains the
carbon dioxide concentration, or pH. light- sensitive rod and cone cells.

Oxyhemoglobin: Hemoglobin that has combined with Rod: light sensitive cells of the retina that are particularly
oxygen. sensitive to dim light and mediate white and black vision.

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-56
Thalamus: The part of the brain that serves as a main relay
Scrotum: The external sac of skin in males that contains center, transmitting information between the spinal cord
the testes and their accessory organs. and the cerebrum.

Secondary vasodilatation- Dilation following exposure to Thermal- Pertaining to heat.


cold to sustain viable tissues.
Thermotherapy- The use of heat in the treatment of
Semen: Fluid composed of sperm suspended in various pathology or disease.
glandular secretions that is ejaculated from the penis during
orgasm. Thyroid gland: An endocrine gland that lies anterior to the
trachea and releases hormones that regulate the rate of
Semilunar valves: Valves between the ventricles of the metabolism.
heart and the arteries that carry blood away from the heart;
aortic and pulmonary valves. Thyroid hormones: Hormones, including thyroxin,
secreted by the thyroid gland; stimulate rate of metabolism.
Seminal vesicles: Glandular sacs that secrete a component
of seminal fluid. Tidal volume: Volume of gas inspired or expired during
each quiet respiration cycle.
Shawnn cell: Supporting cells found in nervous tissue
outside the central nervous system. Trachea: The main trunk of the system of tubes by which
air passes to and from the lungs.
Sinoatrial node: Mass of specialized cardiac muscle in
which the impulse triggering the heartbeat originates; the Transducer- A device that changes energy from one type
peacemaker of the heart. to another.

Small intestine: Portion of the digestive tract that extends Transmission- To pass through some medium.
from the stomach to the large intestine.
Tricuspid valve: The valve connecting the right atrium to
Sphygmomanometer: An instrument for measuring blood the right ventricle.
pressure, especially arterial blood pressure.
Ultrasound- A portion of the acoustic spectrum located
Spinal cord: The dorsal, tubular nerve cord. above audible sound. Sound waves higher than the 16,000
to 20,000 Hz detectable by the human ear.
Spinal nerves: The nerves that emerge from the spinal
cord. Ultrasound imaging: Technique in which high frequency
sound waves are used to provide an image (sonogram) of
Spirometer: An instrument for measuring the air entering an internal structure.
and leaving the lungs.
Ultraviolet- The portion of the electromagnetic spectrum
Spleen: Abdominal organ located just below the diaphragm associated with chemical changes, located adjacent to the
that removes worn- out blood cells and bacteria from the violet portion of the visible light spectrum.
blood and plays a role in immunity.
Ureter: One of the paired tubular structures that conducts
Stomach: Muscular region of the digestive tract extending urine from the kidney to the bladder.
from the esophagus to the small intestine.
Urethra: The tube that conducts urine from the bladder to
Superior vena cava: Main vein feeding back to the heart the outside of the body.
from systemic circulation above the heart.
Urinary bladder: An organ that receives urine from the
Systole: The contraction, or period of contraction, of the ureters and temporarily stores it.
heart, especially that of the ventricles. It coincides with the
interval between the first and second heart sound, during Urinary system: Body system that consists of kidneys,
which blood is forced into the aorta and the pulmonary urinary bladder, and associated ducts.
trunk.
Uterus: The hollow, muscular organ of the female
Systolic: Of or pertaining to systole. reproductive tract in which the fetus undergoes
development.
Testis: The male gonad that produces sperm. The testes are
located in the scrotum. Vagina: The elastic, muscular tube, extending from the
cervix to its orifice that receives the penis during sexual
intercourse and serves as the birth canal.

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-57
Vas deferens: One of the paired sperm ducts that connect
the epididymis of the testis to the ejaculatory duct.

Vasoconstrictions- Narrowing of the blood vessels.

Vasodilatation- Dilatation of the blood vessels.

Vasodilatation: Expansion of the diameter of blood


vessels.

Vein: A blood vessel that carries blood from the tissue


toward a chamber of the heart.

Ventricle: One of the chambers of the heart that receives


blood from an atrium.

Venule: A small vein; especially one of the minute veins


connecting the capillary bed with the larger systemic veins.

Vibration- A shaking massage technique; a fine tremulous


movement made by the hand or fingers placed firmly
against a body part and causing that part to vibrate. Often
used for a soothing effect; may be stimulating when more
energy is applied.

Viscoelastic properties- The property of a material to


show sensitivity to rate of loading.

Vital capacity: Volume of air that can be exhaled after the


deepest possible inhalation.

Volt- The electromotive force that must be applied to


produce a movement of electrons.

Voltage sensitive permeability- The quality of some cell


membranes that makes them permeable to different ions
based on the electric charge of the ions. Nerve and muscle
cell membranes allow negatively charged ions into the cell
while actively transporting some positively charged ions
outside the cell membrane.

Vulva: The external genital structure of the female.

Waveform- The shape of an electrical current as displayed


on an oscilloscope.

Wavelength- The distance from one point in a propagating


wave to the same point in the next wave.
Wavelength: The distance from one peak to the next;
energy from electromagnetic radiation is inversely
proportional to its wavelength.

White matter: Nervous tissue in the brain and spinal cord


that contains myelinated axons. Compare with gray matter.

β-endorphin- A neurohormone derived from β-lipo- tropin


and containing enkephalin. It is similar in structure and
properties to morphine. β-endor- phin has a half-life of 4
hours.

β-lipotropin- A pituitary hormone containing β-en-


dorphin and enkephalin and having opiate activity. [13]

July 10, 2004 Congress on Biofluid Dynamics of Human Body Systems at University of Puerto Rico Mayaguez BB-58

You might also like