III. Mg concentration Standard Mg++ concentration is 2 mM, but sometimes you may find that the concentration needs to be raised (rarely lowered) to get a PCR to work. Raising Mg lowers specificity, and is
1 of 3
1/30/2013 7:29 PM
http://irc.igd.cornell.edu/Protocols/PCR_principles.htm
roughly comparable to lowering the annealing temperature. It may cause multiple bands to appear (or, occasionally, disappear). In my experience, most troubleshooting is more easily accomplished by playing with other variables (temperature, DNA quality or amount). IV. Length of expected product The length of the extension step (72 C) should be a function of the length of the product you are trying to amplify. A general guideline is 1 minute/kb of product length, but in fact this is more than is needed, particularly if you are doing 3-step (i.e., conventional) PCR, as extension will take place during the annealing step and during the ramp time. Taq polymerase is a very fast and very processive enzyme. Optimizing your reaction, saving time The original PCR programs typically used 1 minute denaturation and annealing steps. This is way more time than is needed. Also, if your product will be much less than 1 kb, a 1 minute extension is also too long. When possible, use 2-step PCR: 1. Initial denaturation, 2-3 min at 94 C 2. Denaturation, 15 sec at 94 C 3. Simultaneous annealing and extension, x minutes at 68 C x depends on product length, 1 min/kb 4. Return to step 2 for 29-39 additional cycles. When the Tm of your primers is below 65 C, or when 2-step PCR does not yield a product, use 3-step PCR: 1. Initial denaturation, 2-3 min at 94 C 2. Denaturation, 15 sec at 94 C 3. Annealing, 15 sec at x C x depends on Tm, should be as high as possible, often Tm 4. Extension, x sec x depends on product length, 1 min/kb 5. Return to step 2 for 29-39 additional cycles. Most PCR programs set up in this way should be complete in well under 2 hours. This saves you time and frees up the machine for other users. Shorter times also prolong the activity of your Taq in the later cycles. I usually let the thermocycler block heat up to at least 80 C before I put my tubes in. This is not a true hot-start but I think it may improve specificity. When your program is finished, remove tubes as promptly as possible and CANCEL YOUR PROGRAM. Do not leave an open block set on 4 C!! Gradient PCR When you are trying out heterologous primers, you can run a set of PCRs with different annealing temperatures all in the same block. This saves time and does not tie up multiple blocks. Long PCR Long PCR requires special enzymes and concentrations of reaction components. Often it is advisable to, after ten cycles, add 10 seconds to the extension time of every consecutive cycle, as the Taq polymerase loses activity and takes more time to make the full length product. Nested PCR This is a very reliable way to get a better PCR product when results are marginal for a number of reasons. Design a new F or
2 of 3
1/30/2013 7:29 PM
http://irc.igd.cornell.edu/Protocols/PCR_principles.htm
R primer (F or R) internal to the original, compatible with the partner, and use a tiny amount of the primary PCR product as template in the PCR with F + R (or F + R, as the case may be). Frequently this will -eliminate extra bands if the first PCR was messy -produce a robust band where the first PCR was weak or even invisible This method also saves on genomic DNA. For example, you could do a small-scale primary PCR of a 2 kb product using a modest amount of DNA. It might not work very well, but from that primary PCR you can amplify two overlapping 1.2 kb products for sequencing (using primers F + R and F + R). Remember that you need only a very little of the primary product in the nested PCR because it is very low complexity (see above). I often dilute the primary PCR 1:100 and add 1 uL to the nested PCR. Sometimes this is too much. Also, you often can cut your cycles back to 25-30 for the nested PCR. You have to determine this empirically. Finally, a few words about enzymes Enzymes are stored at -20oC in a non-frost free freezer, typically in 50% glycerol. The tubes should never be allowed to reach room temperature and gloves should be worn when handling or you must handle the tube very carefully and keep your fingers away from the opening. Always use a new, clean pipet tip every time you use a stock tube of enzyme. Before you open a new tube of enzyme, spin it briefly first as there is often enzyme in the cap. This is particularly true for temperature-sensitive enzymes (i.e., all except Taq), that may be put in ice: enzyme in the cap does not stay cold. Spin tubes as necessary to keep enzyme at bottom. Avoid trying to measure out minute quantities of enzyme, as the 50% glycerol storage buffer makes this impossible. When pipetting enzyme from a stock tube, place the end of the tip just far enough into the enzyme to get what you need, do not plunge the tip way down into the solution, as the outside of the tip will become covered with enzyme and your measurement will be off. Whenever possible, make a cocktail of enzyme, buffer, water, etc., and aliquot this as appropriate. (DO NOT add enzyme to unbuffered water, you can denature it. Mix water and buffer first, place on ice, then add enzyme.) The volume of the enzyme should be less than 1/10 of the final volume of the reaction mixture, as too much glycerol can interfere with enzyme activity. Enzymes are expensive and perishable. Use them carefully!!
3 of 3
1/30/2013 7:29 PM