Technical Information

Proline Promag 10L


Electromagnetic Flow Measuring System Flow measurement of liquids in water or wastewater applications

Application Electromagnetic flowmeter for bidirectional measurement of liquids with a minimum conductivity of 50 S/cm: Drinking water Wastewater Sewage sludge Flow measurement up to 40000 m/h (175000 gal/min) Fluid temperature up to +90 C (+194 F) Process pressures up to 16 bar (232 psi) Lengths in accordance with DVGW/ISO Application-specific lining of the measuring pipe from polyurethane, hard rubber or PTFE with the following drinking water permissions: KTW WRAS NSF ACS

Your benefits Promag measuring devices offer you cost-effective flow measurement with a high degree of accuracy for a wide range of process conditions. The uniform Proline transmitter concept comprises: High degree of reliability and measuring stability Uniform operating concept The tried-and-tested Promag sensors offer: No pressure loss Not sensitive to vibrations Simple installation and commissioning Flexible flange mounting

TI00100D/06/EN/14.11 71136637

Proline Promag 10L

Table of contents
Function and system design. . . . . . . . . . . . . . . . . . . . . 3
Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

Mechanical construction . . . . . . . . . . . . . . . . . . . . . . 18
Design, dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Measuring tube specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . Material . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Material load diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Fitted electrodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Process connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Surface roughness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 27 30 31 32 34 34 34

Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Measuring ranges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Operable flow range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Load . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Low flow cutoff . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Galvanic isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

Human interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
Display elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 Operating elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 Remote operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34

Certificates and approvals . . . . . . . . . . . . . . . . . . . . . 35 Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4


Electrical connection, measuring unit . . . . . . . . . . . . . . . . . . . . . . 4 Electrical connection, terminal assignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Electrical connection, remote version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Supply voltage (power supply) . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Cable entry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Remote version cable specifications . . . . . . . . . . . . . . . . . . . . . . . . 6 Power consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Power supply failure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Potential equalization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-tick mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Drinking water approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Other standards and guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . 35 35 35 35

Ordering information. . . . . . . . . . . . . . . . . . . . . . . . . 35 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35 Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35 Registered trademarks . . . . . . . . . . . . . . . . . . . . . . . . 35

Performance characteristics. . . . . . . . . . . . . . . . . . . . . 8
Reference operating conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Maximum measured error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Repeatability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

Operating conditions: Installations . . . . . . . . . . . . . . . 9


Installation instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Inlet and outlet run . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Adapters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Length of connecting cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

Operating conditions: Environment. . . . . . . . . . . . . . 15


Ambient temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Storage temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Degree of protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Shock and vibration resistance . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Electromagnetic compatibility (EMC) . . . . . . . . . . . . . . . . . . . . . 15

Operating conditions: Process . . . . . . . . . . . . . . . . . . 16


Medium temperature range . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 Conductivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 Medium pressure range (nominal pressure) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 Pressure tightness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 Limiting flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16 Pressure loss . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

Endress+Hauser

Proline Promag 10L

Function and system design


Measuring principle Following Faraday's law of magnetic induction, a voltage is induced in a conductor moving through a magnetic field. In the electromagnetic measuring principle, the flowing medium is the moving conductor. The voltage induced is proportional to the flow velocity and is supplied to the amplifier by means of two measuring electrodes. The flow volume is calculated by means of the pipe cross-sectional area. The DC magnetic field is created through a switched direct current of alternating polarity.

Ue I

V B I L
A0003191

Ue = B L v Q=Av Ue B L v Q A I Induced voltage Magnetic induction (magnetic field) Electrode spacing Flow velocity Volume flow Pipe cross-section Current strength

Measuring system

The measuring system consists of a transmitter and a sensor. Two versions are available: Compact version: Transmitter and sensor form a mechanical unit. Remote version: Sensor is mounted separate from the transmitter. Transmitter: Promag 10 (key operation, two-line, unilluminated display) Sensor: Promag L - DN 50 to 300 (2 to 12") - DN 350 to 1200 (14 to 48")

"

Caution! To avoid corrosion, the sensor and process connection material must be selected considering the environmental and process conditions.

Input
Measured variable Measuring ranges Flow velocity (proportional to induced voltage) Measuring ranges for liquids Typically v = 0.01 to 10 m/s (0.03 to 33 ft/s) with the specified accuracy Over 1000 : 1

Operable flow range

Endress+Hauser

Proline Promag 10L

Output
Output signal Current output Galvanically isolated Active: 4 to 20 mA, RL < 700 (for HART: RL 250 ) Full scale value adjustable Temperature coefficient: typ. 2 A/C, resolution: 1.5 A

Pulse/status output Galvanically isolated Passive: 30 V DC / 250 mA Open collector Can be configured as: Pulse output: Pulse value and pulse polarity can be selected, max. pulse width adjustable (5 to 2000 ms), pulse frequency max. 100 Hz Status output: for example, can be configured for error messages, empty pipe detection, flow recognition, limit value

Signal on alarm

Current output Failsafe mode can be selected (e.g. in accordance with NAMUR Recommendation NE 43) Pulse output Failsafe mode can be selected Status output "Not conductive" in the event of fault or power supply failure Section "output signal" Switch-on points for low flow are selectable. All circuits for inputs, outputs and power supply are galvanically isolated from each other.

Load Low flow cutoff Galvanic isolation

Power supply
Electrical connection, measuring unit

e
e b e b

a i

f g h
24 25 26 27 1 2 + + L1 N (L+) (L-)

A0003192

Connecting the transmitter (aluminum field housing), cable cross-section max. 2.5 mm2 (14 AWG) a b c d e f g h i Electronics compartment cover Power supply cable Ground terminal for power supply cable Terminal connector for power supply cable Signal cable Ground terminal for signal cable Terminal connector for signal cable Service connector Ground terminal for potential equalization

Endress+Hauser

Proline Promag 10L

Electrical connection, terminal assignment

Order version 24 (+) 10***-***********A Functional values 25 () 26 (+)

Terminal No. 27 () 1 (L1/L+) 2 (N/L)

Pulse/status output

HART current output

Power supply Section "Supply voltage"

4, Section "Output signal"

Electrical connection, remote version


S1 E1 E2

GND

S2

d c a

4 37 36

42 41

n.c. 5

n.c. 7 4 37

n.c. 42 41

E1

E2

GND

A0012461

Connecting the remote version a b c d n.c. Wall-mount housing connection compartment Sensor connection housing cover Signal cable Coil current cable Not connected, insulated cable shields

Terminal numbers and cable colors: 5/6 = brown, 7/8 = white, 4 = green, 37/36 = yellow

Supply voltage (power supply)

85 to 250 V AC, 45 to 65 Hz 20 to 28 V AC, 45 to 65 Hz 11 to 40 V DC Power supply and signal cables (inputs/ outputs): Cable entry M20 1.5 (8 to 12 mm / 0.31 to 0.47") Thread for cable entries, " NPT, G " Connecting cable for remote version: Cable entry M20 1.5 (8 to 12 mm / 0.31 to 0.47") Thread for cable entries, " NPT, G "

Cable entry

Endress+Hauser

Proline Promag 10L

Remote version cable specifications

Coil cable 2 0.75 mm2 (18 AWG) PVC cable with common, braided copper shield ( 7 mm / 0.28") Conductor resistance: 37 /km (0.011 /ft) Capacitance core/core, shield grounded: 120 pF/m (37 pF/ft) Operating temperature: 20 to +80 C (68 to +176 F) Cable cross-section: max. 2.5 mm2 (14 AWG) Test voltage for cable insulation: 1433 AC r.m.s. 50/60 Hz or 2026 V DC Signal cable 3 0.38 mm2 (20 AWG) PVC cable with common, braided copper shield ( 7 mm / 0.28") and individual shielded cores With empty pipe detection (EPD): 4 0.38 mm2 (20 AWG) PVC cable with common, braided copper shield ( 7 mm / 0.28") and individual shielded cores Conductor resistance: 50 /km (0.015 /ft) Capacitance core/shield: 420 pF/m (128 pF/ft) Operating temperature: 20 to +80 C (68 to +176 F) Cable cross-section: max. 2.5 mm2 (14 AWG)

1 2 3 4 5 6 7 a b
A0003194

a b 1 2 3 4 5 6 7

Signal cable Coil current cable Core Core insulation Core shield Core jacket Core reinforcement Cable shield Outer jacket

Operation in zones of severe electrical interference The measuring device complies with the general safety requirements in accordance with EN 61010 and the EMC requirements of IEC/EN 61326.

"
Power consumption

Caution! Grounding is by means of the ground terminals provided for the purpose inside the connection housing. Ensure that the stripped and twisted lengths of cable shield to the ground terminal are as short as possible. 85 to 250 V AC: < 12 VA (incl. sensor) 20 to 28 V AC: < 8 VA (incl. sensor) 11 to 40 V DC: < 6 W (incl. sensor) Switch-on current: Max. 16 A (< 5 ms) for 250 V AC Max. 5.5 A (< 5 ms) for 28 V AC Max. 3.3 A (< 5 ms) for 24 V DC

Power supply failure

Lasting min. cycle frequency: EEPROM saves measuring system data

Endress+Hauser

Proline Promag 10L

Potential equalization

Warning! The measuring system must be included in the potential equalization. Perfect measurement is only ensured when the fluid and the sensor have the same electrical potential. This is ensured by the reference electrode integrated in the sensor as standard. The following should also be taken into consideration for potential equalization: Internal grounding concepts in the company Operating conditions, such as the material/ grounding of the pipes (see table) Standard situation
Operating conditions When using the measuring device in a: Metal, grounded pipe Potential equalization takes place via the ground terminal of the transmitter. Note! ! When installing in metal pipes, we recommend you connect the ground terminal of the transmitter housing with the piping.
A0010831

Potential equalization

Via the ground terminal of the transmitter

Special situations
Operating conditions When using the measuring device in a: Metal pipe that is not grounded This connection method also applies in situations where: Customary potential equalization cannot be ensured. Excessively high equalizing currents can be expected. Both sensor flanges are connected to the pipe flange by means of a ground cable (copper wire, at least 6 mm / 0.0093 in) and grounded. Connect the transmitter or sensor connection housing, as applicable, to ground potential by means of the ground terminal provided for the purpose. The ground cable is mounted directly on the conductive flange coating with the flange screws. Note! ! The ground cable for flange-to-flange connections can be ordered separately as an accessory from Endress+Hauser. Potential equalization

A0011567

Via the ground terminal of the transmitter and the flanges of the pipe When using the measuring device in a: Plastic pipe Pipe with insulating lining This connection method also applies in situations where: Customary potential equalization cannot be ensured. Excessively high equalizing currents can be expected. Potential equalization takes place using additional ground disks, which are connected to the ground terminal via a ground cable (copper wire, at least 6 mm / 0.0093 in). When installing the ground disks, please comply with the enclosed Installation Instructions.

A0010833

Via the ground terminal of the transmitter and the optionally available ground disks

Endress+Hauser

Proline Promag 10L

Operating conditions When using the measuring device in a: Pipe with a cathodic protection unit The device is installed potential-free in the pipe. Only the two flanges of the pipe are connected with a ground cable (copper wire, at least 6 mm / 0.0093 in). Here, the ground cable is mounted directly on the conductive flange coating with flange screws. Note the following when installing: The applicable regulations regarding potential-free installation must be observed. There should be no electrically conductive connection between the pipe and the device. The mounting material must withstand the applicable torques.

Potential equalization

1 2 2

A0010834

Potential equalization and cathodic protection 1 2 Power supply isolation transformer Electrically isolated

Performance characteristics
Reference operating conditions As per DIN EN 29104 and VDI/VDE 2641: Fluid temperature: +28 C 2 K (+82 F 2 K) Ambient temperature: +22 C 2 K (+72 F 2 K) Warm-up period: 30 minutes Installation conditions: Inlet run > 10 DN Outlet run > 5 DN Sensor and transmitter grounded. The sensor is centered in the pipe. Maximum measured error Current output: also typically 5 A Pulse output: 0.5% o.r. 2 mm/s (0.5% o.r. 0.08 in/s) (o.r. = of reading) Fluctuations in the supply voltage do not have any effect within the specified range.
[%] 2.5 2.0 0.5 % 1.5 1.0 0.5 0 0 1 2 4 6 8 10 [m/s] v [ft]
A0003200

10

15

20

25

30

32

Max. measured error in % of reading

Repeatability

Max. 0.2% o.r. 2 mm/s (0.2% o.r. 0.08 in/s) (o.r. = of reading)

Endress+Hauser

Proline Promag 10L

Operating conditions: Installations


Installation instructions Mounting location Entrained air or gas bubble formation in the measuring tube can result in an increase in measuring errors. Avoid the following installation locations in the pipe: Highest point of a pipeline. Risk of air accumulating! Directly upstream from a free pipe outlet in a vertical pipeline.

h 2 x DN

A0003202

Mounting location

Installation of pumps Sensors may not be installed on the pump suction side. This precaution is to avoid low pressure and the consequent risk of damage to the lining of the measuring tube. Information on the pressure tightness of the measuring tube lining 16, Section "Pressure tightness". Pulsation dampers may be needed when using piston pumps, piston diaphragm pumps or hose pumps. Information on the shock and vibration resistance of the measuring system 15, Section "Shock and vibration resistance".

A0003203

Installation of pumps

Endress+Hauser

Proline Promag 10L

Partially filled pipes Partially filled pipes with gradients necessitate a drain-type configuration. The empty pipe detection function (EPD) provides additional security in detecting empty or partially filled pipes.

"

Caution! Risk of solids accumulating. Do not install the sensor at the lowest point in the drain. It is advisable to install a cleaning valve.

2 x DN 5 x DN
A0003204

Installation with partially filled pipes

Down pipes Install a siphon or a vent valve downstream of the sensor in down pipes h 5 m (16.4 ft). This precaution is to avoid low pressure and the consequent risk of damage to the lining of the measuring tube. This measure also prevents the liquid current stopping in the pipe which could cause air locks. Information on the pressure tightness of the measuring tube lining 16, Section "Pressure tightness".

A0008157

Installation measures for vertical pipes 1 2 h Vent valve Pipe siphon Length of the down pipe

10

Endress+Hauser

Proline Promag 10L

Orientation An optimum orientation helps avoid gas and air accumulations and deposits in the measuring tube. However, the measuring device also offers the additional function of empty pipe detection (EPD) for detecting partially filled measuring tubes or if outgassing fluids or fluctuating operating pressures are present. Vertical orientation This is the ideal orientation for self-emptying piping systems and for use in conjunction with empty pipe detection.

A0008158

Vertical orientation

Horizontal orientation The measuring electrode axis should be horizontal. This prevents brief insulation of the two measuring electrodes by entrained air bubbles.

"

Caution! Empty pipe detection only works correctly with horizontal orientation if the transmitter housing is facing upwards. Otherwise there is no guarantee that empty pipe detection will respond if the measuring tube is only partially filled or empty.

A
1

A
3

A0003207

Horizontal orientation 1 2 3 EPD electrode for empty pipe detection Measuring electrodes for signal detection Reference electrode for potential equalization

Endress+Hauser

11

Proline Promag 10L

Vibrations Secure the piping and the sensor if vibration is severe.

"

Caution! If vibrations are too severe, we recommend the sensor and transmitter be mounted separately. Information on the permitted shock and vibration resistance 15, Section "Shock and vibration resistance".
L

A0003208

Measures to prevent vibration of the measuring device L > 10 m (33 ft)

Inlet and outlet run

If possible, install the sensor well clear of assemblies such as valves, T-pieces, elbows etc. Note the following inlet and outlet runs to comply with measuring accuracy specifications: Inlet run: 5 DN Outlet run: 2 DN

5 x DN

2 x DN

A0003210

Inlet and outlet run

12

Endress+Hauser

Proline Promag 10L

Adapters

Suitable adapters to DIN EN 545 (double-flange reducers) can be used to install the sensor in larger-diameter pipes. The resultant increase in the rate of flow improves measuring accuracy with very slow-moving fluids. The nomogram shown here can be used to calculate the pressure loss caused by reducers and expanders.

Note! The nomogram only applies to liquids of viscosity similar to water. 1. 2. Calculate the ratio of the diameters d/D. From the nomogram read off the pressure loss as a function of flow velocity (downstream from the reduction) and the d/D ratio.

[mbar] 100
8 m/s 7 m/s 6 m/s

10
5 m/s

max. 8 d D

4 m/s 3 m/s 2 m/s

1
1 m/s

d / D 0.5

0.6

0.7

0.8

0.9

A0003213

Pressure loss due to adapters

Endress+Hauser

13

Proline Promag 10L

Length of connecting cable

When mounting the remote version, please note the following to achieve correct measuring results: Fix cable run or lay in armored conduit. Cable movements can falsify the measuring signal especially in the case of low fluid conductivities. Route the cable well clear of electrical machines and switching elements. If necessary, ensure potential equalization between sensor and transmitter. The permitted cable length Lmax is determined by the fluid conductivity. A minimum conductivity of 50 S/cm is needed for all fluids. When the empty pipe detection function is switched on (EPD), the maximum connecting cable length is 10 m (33 ft).

[S/cm] 200 L max

100

50

[m] 50 0 200 100 400 200 [ft] 600


A0003214

L max

Permitted length of connecting cable for remote version Area marked in gray = permitted range; Lmax = length of connecting cable in [m] ([ft]); fluid conductivity in [S/cm]

14

Endress+Hauser

Proline Promag 10L

Operating conditions: Environment


Ambient temperature range Transmitter 20 to +60 C (4 to +140 F) Sensor Flange material carbon steel: 10 to +60 C (14 to +140 F) Flange material stainless steel (DN 300): 40 to +60 C (40 to +140 F) Caution! The permitted temperature range of the measuring tube lining may not be undershot or overshot ( 16, Section "Medium temperature range"). Please note the following points: Install the device in a shady location. Avoid direct sunlight, particularly in warm climatic regions. The transmitter must be mounted separate from the sensor if both the ambient and fluid temperatures are high. Storage temperature The storage temperature corresponds to the operating temperature range of the measuring transmitter and the appropriate measuring sensors.

"

"
Degree of protection

Caution! The measuring device must be protected against direct sunlight during storage in order to avoid unacceptably high surface temperatures. A storage location must be selected where moisture does not collect in the measuring device. This will help prevent fungus and bacteria infestation which can damage the liner. Standard: IP 67 (NEMA 4X) for transmitter and sensor. Optional: IP 68 (NEMA 6P) for sensor for remote version (DN 300 only with stainless steel). For information regarding applications where the device is buried directly in the soil or is installed in a flooded wastewater basin please contact your local Endress+Hauser Sales Center. Acceleration up to 2 g following IEC 600 68-2-6 As per IEC/EN 61326 as well as NAMUR Recommendation NE 21 Emission: to limit value for industry EN 55011

Shock and vibration resistance Electromagnetic compatibility (EMC)

Endress+Hauser

15

Proline Promag 10L

Operating conditions: Process


Medium temperature range

0 to +80 C (+32 to +176 F) for hard rubber (DN 350 to 1200) 20 to +50 C (4 to +122 F) for polyurethane (DN 50 to 1200) 20 to +90 C (4 to +194 F) for PTFE (DN 50 to 300)
The minimum conductivity is: 50 S/cm Note! In the remote version, the necessary minimum conductivity also depends on the cable length ( 14, Section "Length of connecting cable"). EN 1092-1 (DIN 2501) PN 6 (DN 350 to 1200) PN 10 (DN 50 to 1200) PN 16 (DN 50 to 150) EN 1092-1, lap joint flange, stampel plate PN 10 (DN 50 to 300) ANSI B 16.5 Class 150 (2" to 24") AWWA Class D (32" to 48") AS2129 Table E (350 to 1200) AS4087 PN 16 (350 to 1200)

Conductivity

!
Medium pressure range (nominal pressure)

Pressure tightness

Measuring tube lining: Polyurethane, hard rubber


Promag L Nominal diameter [mm] 50 to 1200 [inch] 2 to 48" Polyurethane Measuring tube lining Resistance of measuring tube lining to partial vacuum Limit values for abs. pressure [mbar] ([psi]) at various fluid temperatures 25 C 77 F 0 0 50 C 122 F 0 0 80 C 176 F 0

350 to 1200 14 to 48" Hard rubber

Measuring tube lining: PTFE


Nominal diameter Limit values for abs. pressure [mbar] ([psi]) at fluid temperatures: 25 C (77 F) [mm] 50 65 80 100 125 150 200 250 300 [inch] 2" 3" 4" 6" 8" 10" 12" [mbar] 0 0 0 0 135 135 200 330 400 [psi] 0 0 0 0 1.96 1.96 2.90 4.79 5.80 90 C (194 F) [mbar] 0 40 40 135 240 240 290 400 500 [psi] 0 0.58 0.58 1.96 3.48 3.48 4.21 5.80 7.25

Limiting flow

The diameter of the pipe and the flow rate determine the nominal diameter of the sensor. The optimum flow velocity is between 2 to 3 m/s (6.5 to 9.8 ft/s). The velocity of flow (v), moreover, has to be matched to the physical properties of the fluid: v < 2 m/s (6.5 ft/s): for abrasive fluids such as potter's clay, lime milk, ore slurry etc. v > 2 m/s (6.5 ft/s): for fluids causing build-up such as wastewater sludges etc.

16

Endress+Hauser

Proline Promag 10L

Flow characteristic values (SI units)


Diameter Recommended flow Min./max. full scale value [mm] 50 65 80 100 125 150 200 250 300 350 375 400 450 500 600 700 800 900 1000 1200 [inch] 2" 3" 4" 6" 8" 10" 12" 14" 15" 16" 18" 20" 24" 32" 36" 40" 42" 48" (v ~ 0.3 or 10 m/s) 35 to 1100 dm/min 60 to 2000 dm/min 90 to 3000 dm /min 145 to 4700 dm/min

Factory settings Full scale value Current output (v ~ 2.5 m/s) 300 dm/min 500 dm/min 750 dm /min 1200 dm/min

Pulse value (~ 2 pulses/s) 2.50 dm 5.00 dm 5.00 dm 10.00 15.00 dm 0.025 m


Low flow cut off (v ~ 0.04 m/s) 5 dm/min 8 dm/min 12 dm/min 20 dm/min 30 dm/min 2.5 m/h 5.0 m/h 7.5 m/h 10 m/h 15 m/h 20 m/h 20 m/h 25 m/h 30 m/h 40 m/h 50 m/h 75 m/h 100 m/h 125 m/h 150 m/h

dm

220 to 7500 dm /min 20 to 600 m /h 35 to 1100 m/h 55 to 1700 m/h 80 to 2400 m /h 110 to 3300 m/h

1850 dm /min 150 m /h 300 m/h 500 m/h 750 m /h 1000 m/h

0.05 m 0.05 m 0.10 m 0.10 0.15 m 0.15 m

140 to 4200 m /h 140 to 4200 m /h 180 to 5400 m/h 220 to 6600 m/h 310 to 9600 m /h 420 to 13500 m/h

1200 m /h 1200 m /h 1500 m/h 2000 m/h 2500 m /h 3500 m/h


0.25 m 0.25 m 0.30 m 0.50 0.75 m 0.75 m 1.50

550 to 18000 m /h 690 to 22500 m /h 850 to 28000 m/h 1250 to 40000 m/h

4000 m /h 4500 m /h 6000 m/h 8000 m/h

1.00 m m

Flow characteristic values (US units)


Diameter Recommended flow rate Min./max. full scale value [inch] 2" 3" 4" 6" 8" 10" 12" 14" 15" 16" 18" 20" 24" 28" 30" 32" 36" 40" 42" 48" [mm] 50 65 80 100 125 150 200 250 300 350 375 400 450 500 600 700 800 900 1000 1200 (v ~ 0.3 or 10 m/s) 10 to 300 gal/min 16 to 500 gal/min 24 to 800 gal/min 40 to 1250 gal/min 60 to 1950 gal/min 90 to 2650 gal/min 155 to 4850 gal/min 250 to 7500 gal/min 350 to 10600 gal/min 500 to 15000 gal/min 600 to 19000 gal/min 600 to 19000 gal/min 800 to 24000 gal/min 1000 to 30000 gal/min 1400 to 44000 gal/min 1900 to 60000 2150 to 67000 2450 to 80000 gal/min gal/min gal/min Full scale value Current output (v ~ 2.5 m/s) 75 gal/min 130 gal/min 200 gal/min 300 gal/min 450 gal/min 600 gal/min 1200 gal/min 1500 gal/min 2400 gal/min 3600 gal/min 4800 gal/min 4800 gal/min 6000 gal/min 7500 gal/min 10500 gal/min 13500 gal/min 16500 gal/min 19500 gal/min 24000 gal/min 30000 gal/min 33000 gal/min 42000 gal/min Factory settings Pulse value (~ 2 pulses/s) 0.50 gal 1 gal 2 gal 2 gal 5 gal 5 gal 10 gal 15 gal 25 gal 30 gal 50 gal 50 gal 50 gal 75 gal 100 gal 125 gal 150 gal 200 gal 225 gal 250 gal 250 gal 400 gal Low flow cut off (v ~ 0.04 m/s) 1.25 gal/min 2.0 gal/min 2.5 gal/min 4.0 gal/min 7.0 gal/min 12 gal/min 15 gal/min 30 gal/min 45 gal/min 60 gal/min 60 gal/min 60 gal/min 90 gal/min 120 gal/min 180 gal/min 210 gal/min 270 gal/min 300 gal/min 360 gal/min 480 gal/min 600 gal/min 600 gal/min

3100 to 100000 gal/min 3800 to 125000 gal/min 4200 to 135000 gal/min 5500 to 175000 gal/min

Pressure loss

No pressure loss if the sensor is installed in a pipe with the same nominal diameter. Pressure losses for configurations incorporating adapters according to DIN EN 545 ( 13, Section "Adapters").

Endress+Hauser

17

Proline Promag 10L

Mechanical construction
Design, dimensions Transmitter, remote version
E F

ANSCHLUSSKLEMMEN - FIELD TERMINALS

J
A0010718

Transmitter dimensions, remote version

Dimensions in SI units
A 178 J 123 B 113 K 150 C 135 L 100 D 20 to 30 M 25 E 161 to 181 N 133 F 113 O 177.5 G 100 P 327.5 H 8.6 (M8)

All dimensions in [mm]

Dimensions in US units
A 7.00 J 4.84 B 4.45 K 5.90 C 5.31 L 3.94 D 0.79 to 1.81 M 0.98 E 6.34 to 7.13 N 5.24 F 4.44 O 6.99 G 3.94 P 12.89 H 0.34 (M8)

All dimensions in [inch]

A
248 2 (9.76 0.08)

B
238 (9.37)

ANSCHLUSSKLEMMEN - FIELD TERMINALS

ANSCHLUSSKLEMMEN - FIELD TERMINALS

mm (inch)
A0010719

Transmitter mounting, remote version A B Direct wall mounting Pipe mounting

18

Endress+Hauser

Proline Promag 10L

Sensor, remote version, DN 50 to 300

B C

D F
J H L
A0012462

Dimensions in SI units
DN 50 65 80 100 125 150 200 250 300 L1) 200 200 200 250 250 300 350 450 500 129 163 143 102 A B C D E 286 336 336 336 417 417 472 522 572 F 202 227 227 227 267 267 292 317 342 G 84 109 109 109 150 150 180 205 230 H 120 180 180 180 260 260 324 400 460 J 94 94 94 94 140 140 156 156 166

1) The length (L) is regardless of the pressure rating selected. Fitting length to DVGW. All dimensions in [mm]

Dimensions in US units
DN 2" 3" 4" 6" 8" 10" 12"
1)

L1) 7.87 7.87 9.84 11.8 13.8 17.7 19.7

E 11.3 13.2 13.2

F 7.95 8.94 8.94 10.5 11.5 12.5 13.5

G
G 3.32 4.30 4.30 5.91 7.10 8.08 9.06

H 4.72 7.10 7.10 10.2 12.8 15.8 18.1

J 3.70 3.70 3.70 5.51 6.14 6.14 6.54

5.08

6.42

5.63

4.02

16.4 18.6 20.6 22.5

The length (L) is regardless of the pressure rating selected. Fitting length to DVGW All dimensions in [inch]

Endress+Hauser

19

Proline Promag 10L

Sensor, remote version, DN 350 to 1200


A B

F
G H L
A0014987

Dimensions in SI units
DN 350 375 400 450 500 600 700 750 800 900 1000 1050 1200 L 550 600 600 600 600 600 700 750 800 900 1000 1050 1200 129 163 102 A B C F 353 379 379 407 432 473 543 580 599 649 699 735 713 G 290 290 290 290 290 290 424 454 500 580 660 755 828

All dimensions in [mm]

DN PN 6 350 375 400 450 500 600 700 750 800 900 1000 1050 1200 598 649 704 754 850 973 1087 1187 1287 1516

E with pressure rating PN 10 605 661 714 767 863 990 1107 1207 1314 1541 ANSI AWWA 620 677 724 781 879 1006 1072 1129 1233 1344 1408 1569 AS 615 654 669 727 784 885 998 1078 1129 1237 1227 1558 PN 6 490 540 595 645 755 860 975 1075 1175 1405

H with pressure rating PN 10 505 565 615 670 780 895 1015 1115 1230 1455 ANSI AWWA 533 597 635 699 813 927 984 1060 1168 1289 1346 1511 AS 525 550 580 640 705 825 910 995 1060 1175 1255 1490

All dimensions in [mm]

20

Endress+Hauser

Proline Promag 10L

Dimensions in US units
DN 14" 15" 16" 18" 20" 24" 28" 30" 32" 36" 40" 42" 48" L 21.6 23.6 23.6 23.6 23.6 23.6 27.6 29.5 31.5 35.4 39.4 41.3 47.2 5.08 6.42 4.02 A B C F 13.9 14.9 14.9 16.0 17.0 18.6 21.4 22.8 23.6 25.6 27.5 28.9 32.0 G 11.4 11.4 11.4 11.4 11.4 11.4 16.7 17.9 19.7 22.8 26.0 29.7 32.6

All dimensions in [inch]

DN PN 6 14" 15" 16" 18" 20" 24" 28" 30" 32" 36" 40" 42" 48" 23.5 25.6 27.7 29.7 33.5 38.3 42.8 46.7 50.6 59.7

E with pressure rating PN 10 23.8 26.0 28.1 30.2 34.0 39.0 43.6 47.5 51.7 60.6 ANSI AWWA 24.4 26.7 28.5 30.7 34.6 39.6 42.2 44.4 48.5 52.9 55.4 61.8 AS 24.2 25.7 26.3 28.6 30.9 34.8 39.3 42.4 44.4 48.7 52.2 61.3 PN 6 19.3 21.3 23.4 25,4 29.7 33.9 38.4 42.3 46.3 55.3

H with pressure rating PN 10 19.9 22.2 24.2 26,4 30.7 35.2 40.0 43.9 48.4 57.3 ANSI AWWA 21.0 23.5 25.0 27.5 32.0 36.5 38.7 41.7 46.0 50.7 53.0 59.5 AS 20.7 21.7 22.8 25.2 27.8 32.5 35.8 39.2 41.7 46.3 49.4 58.7

All dimensions in [inch]

Endress+Hauser

21

Proline Promag 10L

Compact version, DN 50 to 300

C A B D

E G

Esc

J K L

A0012464

Dimensions in SI units
DN 50 65 80 100 125 150 200 250 300
1)

L1) 200 200 200 250 250 300 350 450 500

F 341 391 391 391

G 257 282 282 282 322 322 347 372 397

H 84 109 109 109 150 150 180 205 230

J 94 94 94 94 140 140 156 156 166

K 120 180 180 180 260 260 324 400 460

178

20 to 30 153 to 168

121

150

472 472 527 577 627

The length (L) is regardless of the pressure rating selected. Fitting length to DVGW. All dimensions in [mm]

Dimensions in US units
DN 2" 3" 4" 6" 8" 10" 12" L1) 7.87 7.87 7.87 9.84 9.84 11.8 13.8 17.7 19.7 7.01 0.79 to 1.81 6.02 to 6.61 4.76 5.91 A B C D E F 13.4 15.4 15.4 15.4 18.6 18.6 20.8 22.7 24.7 G 10.1 11.1 11.1 11.1 12.7 12.7 13.7 14.7 15.6 H 3.32 4.30 4.30 4.30 5.91 5.91 7.10 8.08 9.06 J 3.70 3.70 3.70 3.70 5.51 5.51 6.14 6.14 6.54 K 4.72 7.10 7.10 7.10 10.2 10.2 12.8 15.8 18.1

1) The length (L) is regardless of the pressure rating selected. Fitting length to DVGW. All dimensions in [inch]

22

Endress+Hauser

Proline Promag 10L

Compact version DN 350 to 600

Esc

F
G H L
A0014993

Dimensions in SI units
DN 350 375 400 450 500 600 700 750 800 900 1000 1050 1200 L 550 600 600 600 600 600 700 750 800 900 1000 1050 1200 178 161 to 181 150 A B C F 401 427 427 455 480 521 591 628 647 697 747 783 861 G 290 290 290 290 290 290 424 454 500 580 660 755 828

Alle Abmessungen in [mm]

DN PN 6 350 375 400 450 500 600 700 646 697 752 802 898 1021

E at pressure rating: PN 10 653 709 762 815 911 1038 ANSI AWWA 668 725 772 829 927 1054 AS 663 702 717 775 832 933 1046 PN 6 490 540 595 645 755 860

H at pressure rating: PN 10 505 565 615 670 780 895 ANSI AWWA 533 597 635 699 813 927 AS 525 550 580 640 705 825 910

Endress+Hauser

23

Proline Promag 10L

DN PN 6 750 800 900 1000 1050 1200 1135 1235 1335 1564

E at pressure rating: PN 10 1155 1255 1362 1588 ANSI AWWA 1120 1177 1281 1392 1456 1617 AS 1126 1177 1284 1374 1606 PN 6 975 1075 1175 1405

H at pressure rating: PN 10 1015 1115 1230 1455 ANSI AWWA 984 1060 1168 1289 1346 1511 AS 995 1060 1175 1255 1490

All dimensions in [mm]

Dimensions in US units
DN 14" 15" 16" 18" 20" 24" 28" 30" 32" 36" 40" 42" 48" L 21.6 23.6 23.6 23.6 23.6 23.6 27.6 29.5 31.5 35.4 39.4 41.3 47.2 7.00 6.347.13 5.91 A B C F 15.8 16.8 16.8 17.9 18.9 20.5 23.2 24.7 25.5 27.4 29.4 30.8 33.9 G 11.4 11.4 11.4 11.4 11.4 11.4 16.7 17.9 19.7 22.8 26.0 29.7 32.6

All dimensions in [inch]

DN PN 6 14" 15" 16" 18" 20" 24" 28" 30" 32" 36" 40" 42" 48" 25.4 27.4 29.8 31.6 35.4 40.2 44.7 48.6 52.5 61.6

E at pressure rating: PN 10 25.7 27.9 30.0 32.0 35.9 40.93 45.5 49.4 53.6 62.5 ANSI AWWA 26.3 28.5 30.4 32.6 36.5 41.5 44.1 46.3 50.4 54.8 57.3 63.6 AS 26.1 27.6 28.2 30.5 32.8 36.7 41.2 44.3 46.3 50.6 54.1 63.2 PN 6 19.3 21.3 23.4 25.4 29.7 33.9 38.4 42.3 46.3 55.3

H at pressure rating: PN 10 19.9 22.2 24.2 26.4 30.7 35.2 40.0 43.9 48.4 57.3 ANSI AWWA 21.0 23.5 25.0 27.5 32.0 36.5 38.7 41.7 46.0 50.7 53.0 59.5 AS 20.7 21.7 22.8 25.2 27.8 32.5 35.8 39.2 41.7 46.3 49.4 58.7

All dimensions in [inch]

24

Endress+Hauser

Proline Promag 10L

Ground disk

DN 300 (12") 6.5 (0.26)

DN 350 (14") DN 600 (24") 9 (0.35)

DN 700 (28")
6.4 (0.25)

B B
A

A D

t = 2 (0.08)

t = 2 (0.08) t = 2 (0.08)

mm (inch)

A0015442

Dimensions in SI and US units


DN [mm] 50 65 80 100 125 150 200 250 300 350 [inch] 2" 2 " 3" 4" 5" 6" 8" 10" 12" 14"
1) 1) 1) 1) 1) 1) 1) 1) 1)

Pressure rating [mm] 52 68 80 104 130 158 206 260 312 343

A [inch] 2.05 2.68 3.15 4.09 5.12 6.22 8.11 10.24 12.28 13.50 420 470 393 15.47 480 525 439 17.28 538 575 493 19.41 592 676 593 23.35 693 [mm] 101 121 131 156 187 217 267 328 375 433

B [inch] 3.98 4.76 5.16 6.14 7.36 8.54 10.51 12.91 14.76 16.54 479 17.05 18.50 542 18.90 20.67 583 21.18 23.31 650 22.64 27.28 766 26.61 [mm] 115.5 131.5 154.5 186.5 206.5 256 288 359 413

D [inch] 4.55 5.18 6.08 7.34 8.13 10.08 11.34 14.13 16.26 18.86 [mm] 108 118 135 153 160 184 205 240 273 365

H [inch] 4.25 4.65 5.31 6.02 6.30 7.24 8.07 9.45 10.75 14.37

DIN, PN 6 DIN, PN 10 ANSI, Cl.150 DIN, PN 6 400 16" DIN, PN 10 ANSI, Cl.150 DIN, PN 6 450 18" DIN, PN 10 ANSI, Cl.150 DIN, PN 6 500 20" DIN, PN 10 ANSI, Cl.150 DIN, PN 6 600 24" DIN, PN 10 ANSI, Cl.150

21.34

395

15.55

22.95

417

16.42

25.59

460

18.11

30.16

522

20.55

Endress+Hauser

25

Proline Promag 10L

DN [mm] [inch]

Pressure rating [mm] DIN, PN 6 DIN, PN 10 697 693 687 693

A [inch] 27.44 27.28 27.05 27.28 [mm]

B [inch] [mm] 786 813 807 832

D [inch] 30.94 32.01 31.77 32.76 [mm] 460 480 490 494

H [inch] 18.11 18.9 19.29 19.45

700

28" AS, PN 16 AWWA, Class D AS, PN 16

750

30" AWWA, Class D DIN, PN 6 DIN, PN 10 743 799 795 789 795 897 893 886 893 999 995 988 995 1044 1203 1196 1196 1188 29.25 31.46 31.30 31.06 31.30 35.31 35.16 34.88 35.16 39.33 39.17 38.90 39.17 41.10 47.36 47.09 47.09 46.77 833 893 920 914 940 993 1020 1014 1048 1093 1127 1131 1163 1220 1310 1344 1385 1345 32.8 35.16 36.22 35.98 37.01 39.09 40.16 39.92 41.26 43.03 44.37 44.53 45.79 48.03 51.57 52.91 54.53 52.95 523 520 540 550 561 570 590 595 615 620 650 660 675 704 733 760 786 775 20.59 20.47 21.26 21.65 22.09 22.44 23.23 23.43 24.21 24.41 25.59 25.98 26.57 27.72 28.86 29.92 30.94 30.51

800

32" AS, PN 16 AWWA, Class D DIN, PN 6 DIN, PN 10

900

36" AS, PN 16 AWWA, Class D DIN, PN 6 DIN, PN 10

1000

40" AS, PN 16 AWWA, Class D

1050

42"

AWWA, Class D DIN, PN 6 DIN, PN 10

1200

48" AS, PN 16 AWWA, Class D

1)

Ground disks can be used for all flange norms/ pressure ratings.

26

Endress+Hauser

Proline Promag 10L

Weight

SI units Compact version (lap joint flanges / welded flanges DN >350)


Weight data in kg Nominal diameter [mm] 50 65 80 100 125 150 200 250 300 350 375 400 450 500 600 700 750 800 900 1000 1200 [inch] 2" 2 " 4" 5" 6" 8" 10" 12" 14" 15" 16" 18" 24" 28" 30" 32" 36" 40" 42" 48" PN 10 20" PN 16 3" EN (DIN) 9.0 10.4 12.4 14.4 19.9 23.9 43.4 63.4 68.4 88.4 104. 4 112.4 132.4 155. 4 246. 4 320.4 400.4 473.4 722.4 PN 6 Compact version (including transmitter) EN (DIN) 77.4 89.4 99.4 114.4 155.4 198.4 246.4 314.4 364.4 535.4 AWWA / Class D ANSI / Class 150 ANSI / AWWA 9.00 12.4 14.4 23.9 43.4 63.4 68.4 137.4 168.4 191.4 228.4 302.4 275.4 327.4 394.4 480.4 599.4 682,4 912.4 PN 16. Tabelle E AS 99.4 105.4 120.4 133.4/143.4* 182.4 260.4 339.4 439.4 499.4 696.4 767.4 1225.4

Transmitter Promag (compact version): 1.8 kg (Weight data valid without packaging material) *DN450 AS Tab E

Endress+Hauser

27

Proline Promag 10L

Remote version (lap joint flanges / welded flanges DN >350)


Weight data in kg Nominal diameter [mm] 50 65 80 100 125 150 200 250 300 350 375 400 450 500 600 700 750 800 900 1000 1200 [inch] 2" 2 " 4" 5" 6" 8" 10" 12" 14" 15" 16" 18" 24" 28" 30" 32" 36" 40" 42" 48" PN 10 20" PN 16 3" EN (DIN) 5.7 7.1 9.1 11.1 16.6 20.6 40.1 60.1 65.1 84.1 100.1 108.1 128.1 151.1 317.1 397.1 470.1 719.1 PN 6 Remote version (sensor plus sensor housing without cable) EN (DIN) 73.1 85.1 95.1 110.1 151.1 195.1 243.1 311.1 361.1 532.1 AWWA / Class D ANSI / Class 150 ANSI / AWWA 5.7 9.1 11.1 20.6 40.1 60.1 65.1 133.1 164.1 187.1 224.1 298.1 272.1 324.1 391.1 477.1 596.1 679.1 909.1 PN 16. Tabelle E AS 95.1 101.1 116.1 129.1/139.1* 178.1 256.1 349.1 436.1 496.1 693.1 764.1 1222.1

Transmitter Promag (remote version): 3.1 kg (Weight data valid without packaging material) *DN450 AS Tab E

Lap joint flanges, stamped plate


Weight data in kg Nominal diameter [mm] 50 65 80 100 125 150 200 250 300 [inch] 2" 2 " 3" PN 10 4" 5" 6" 8" 10" 12" Compact version EN (DIN) 5.4 6.2 7.2 13.2 17.2 35.7 54.2 55.2 PN 10 9.7 Remote version (without cable) Sensor EN (DIN) 3.4 4.2 5.2 7.7 11.2 15.2 33.7 52.2 53.2 Transmitter 4.2 4.2 4.2 4.2 4.2 4.2 4.2 4.2 4.2

Transmitter Promag (compact version): 1.8 kg (Weight data valid for standard pressure ratings and without packaging material)

28

Endress+Hauser

Proline Promag 10L

US units ANSI / AWWA (lap joint flanges / welded flanges DN >350)


Weight data in lbs Nominal diameter [mm] 50 65 80 100 125 150 200 250 300 350 375 400 450 500 600 700 750 800 900 1000 1200 [inch] 2" 2 " 3" 4" 5" ANSI / Class 150 6" 8" 10" 12" 14" 15" 16" 18" 20" 24" 28" AWWA / Class D 30" 32" 36" 40" 42" 48" Compact version ANSI / AWWA 19.4 27.4 31.4 95.4 139.4 239.4 302.4 371.4 421.4 503.4 666.4 607.4 869.4 1059.4 1320.4 1504.4 2011.4 AWWA / Class D 721.4 ANSI / Class 150 52.4 Remote version (without cable) ANSI / AWWA 19 26 31 52 95 139 238 294.8 262.8 413.8 494.8 657.8 599.8 714.8 862.8 1051.8 1313.8 1497.8 2004.8

Transmitter Promag (compact version): 3.9 lbs Transmitter Promag (remote version): 6.8 lbs (Weight data valid without packaging material)

Endress+Hauser

29

Proline Promag 10L

Measuring tube specifications

Internal diameter for pressure rating EN (DIN), AS 2129, AS 4087, ANSI and AWWA
Nominal diameter EN (DIN) [mm] [inch] 50 65* 80 100 125 150 200 250 300 2" 2" 3" 4" 5" 6" 8" 10" 12" PN 10/16 PN 10/16 PN 10/16 PN 10/16 PN 10/16 PN 10/16 PN 10/16 PN 10/16 PN 10/16 Class 150 PN 6 350 14" PN 10 PN 16 Table E Class 150 PN 10 375 15" PN 6 400 16" PN 10 PN 16 Table E Class 150 PN 6 450 18" PN 10 PN 16 Table E Class 150 PN 6 500 20" PN 10 PN 16 Table E Class 150 PN 6 600 24" PN 10 PN 16 Table E Class 150 PN 6 PN 10 700 28" PN 16 Table E Class D PN 6 PN 10 750 30" PN 16 Table E Class D PN 16 Class 150 Class 150 Class 150 Class 150 Class 150 Class 150 Class 150 Class 150 Pressure rating AS 2129 AS 4087 ANSI AWWA Internal diameter measuring tube Hard rubber [mm] 341 341 339 339 391 389 391 442 389 389 442 493 440 438 493 595 489 489 595 590 591 589 696 694 690 694 741 743 [inch] 13.4 13.4 13.3 13.3 15.4 15.3 15.4 17.4 15.3 15.3 17.4 19.4 17.3 17.2 19.4 23.4 19.2 19.2 23.4 23.2 23.2 23.1 27.4 27.3 27.2 27.3 29.2 29.3 Polyurethane [mm] 50.3 66.1 78.9 104.3 129.7 158.3 206.7 260.6 311.5 309.9 344 344 342 342 392 394 394 392 392 445 445 443 441 496 496 492 492 598 598 594 592 699 697 693 697 699 697 744 746 [inch] 2.0 2.6 3.1 4.1 5.1 6.2 8.1 10.3 12.3 12.2 13.5 13.5 13.4 13.4 15.4 13.5 13.5 13.4 13.4 17.5 17.5 17.4 17.3 19.5 19.5 19.3 19.3 23.5 23.5 23.4 23.3 27.5 27.4 27.3 27.4 27.5 27.4 29.3 29.4 PTFE [mm] 51.7 67.7 79.9 103.8 129.1 156.3 202.1 256.2 305.5 303.9 [inch] 2.0 2.7 3.1 4.1 5.1 6.2 8.0 10.1 12.0 12.0

30

Endress+Hauser

Proline Promag 10L

Nominal diameter EN (DIN) [mm] [inch] PN 6 PN 10 800 32"

Pressure rating AS 2129 AS 4087 ANSI AWWA

Internal diameter measuring tube Hard rubber [mm] 798 796 [inch] 31.4 31.3 31.2 31.3 35.3 35.2 35.0 35.2 39.3 39.3 39.0 39.1 41.2 47.4 47.2 46.9 47.0 Polyurethane [mm] 801 799 795 797 900 898 892 898 1002 1000 994 998 1049 1206 1202 1194 1198 [inch] 31.5 31.5 31.3 31.4 35.4 35.4 35.1 35.4 39.4 39.4 39.1 39.3 41.3 47.5 47.3 47.0 47.2 PTFE [mm] [inch]

PN 16 Table E Class D PN 6 PN 10

792 794 897 895

900

36"

PN 16 Table E Class D PN 6 PN 10

889 895 999 997

1000

40"

PN 16 Table E Class D PN 6 PN 10

991 995

1050

42"

PN 16 Table E Class D PN 6 PN 10

1046 1203 1199

1200

48"

PN 16 Table E Class D

1191 1195

* Specification according EN 1092-1 (not according DIN 2501)

Material

Transmitter housing: Compact housing: powder-coated die-cast aluminum Wall-mounted housing: powder-coated die-cast aluminum Sensor housing DN 50 to 300: powder-coated die-cast aluminum DN 350 to 1200: with protective lacquering Measuring tube: DN 300; stainless steel 1.4301/304 or 1.4306/304L DN 350; stainless steel 202 or 304 Electrodes: 1.4435/304L, Alloy C-22 Flange EN 1092-1 (DIN 2501): DN 300: 1.4306/304L; 1.4307/304L; 1.4301/304; 1.0038 (S235JRG2) EN 1092-1 (DIN 2501): DN 350: A105; 1.0038 (S235JRG2) ANSI: A105; 316L AWWA: A181/A105; 1.0425/316L (P265GH); 1.0044 (S275JR) AS 2129: A105; 1.0345 (P235GH); 1.0425/316L (P265GH); 1.0038 (S235JRG2); FE 410 WB AS 4087: A105; 1.0425/316L (P265GH); 1.0044 (S275JR) Seals: to DIN EN 1514-1

Ground disks: 1.4435/316L or Alloy C-22

Endress+Hauser

31

Proline Promag 10L

Material load diagram

"

Caution! The following diagrams contain material load diagrams (reference curves) for flange materials with regard to the medium temperature. However, the maximum medium temperatures permitted always depend on the lining material of the sensor and/or the sealing material ( 16). Flange connection to EN 1092-1 (DIN 2501); DN 300 Material: 1.4306/304L; 1.4307/304L
[psi]
300 200 100 0

[bar]

20 15 10 5 0 -60 -40 -20


-40 0
PN 16 PN 10

20 40 60 80 100 120 140 160 180 [C]


100 200 300 360 [F]
A0011571

Flange connection to EN 1092-1; DN 300 Material: 1.4301/304; lap joint flange, stamped plate
[psi]
300 200 100 0

[bar]

20 15 10 5 0 -60 -40 -20


-40 0
PN 10

20 40 60 80 100 120 140 160 180 [C]


100 200 300 360 [F]
A0011573

Flange connection to EN 1092-1 (DIN 2501) Material: A105, 1.0038 (S235JRG2); lap joint flange, stamped plate only in PN 10
[psi]
300 200 100 0

[bar]

20 15 10 5 0 -40 -20
-40 0
PN 16 PN 10
PN 6

20 40 60 80 100 120 140 160 180


100 200 300

[C]

360 [F]
A0011568

32

Endress+Hauser

Proline Promag 10L

Flange connection to ANSI B16.5, Class 150; DN 300 Material: 316L


[psi] [bar]
400 300 200 100 0

30 20 10 0 -40 -20
-40 0
DN 50250 (2"10") DN 300 (12")

20 40 60 80 100 120 140 160 180 [C]


100 200 300 360 [F]
A0011580

Flange connection to ANSI B16.5, Class 150; DN 300 Material: A105


[psi] [bar] 500 400 30 300 20 200 10 100 0 0 -40 -20
-40 0

DN 50250 (2"10") DN 300 (12")

20 40 60 80 100 120 140 160 180


100 200 300

[C]

360 [F]
A0011572

Endress+Hauser

33

Proline Promag 10L

Flange connection to AWWA C207, Class D; DN 350 Material: A181/A105; 1.0425/316L (P265GH); 1.0044 (S275JR)
[bar]
11 10 9 8 7
[psi]

159,5 145 130,5 116 101,5

0 -60 -40 -20 0 20 40 60 80 100 120 140 160

0
[C]
A0011572

Flange connection to AS 2129 Table E oder AS 4087 PN16; DN 350 Material: AS 2129 Table E: A105; 1.0345 (P235GH); 1.0425/316L (P265GH); 1.0038 (S235JRG2); FE 410 WB AS 4087 Class D: A105; 1.0425/316L (P265GH); 1.0044 (S275JR)
[psi] [bar]
300 200 100 0

20
PN 16

15
Table E

10 5 0

-40 -20
-40 0

20 40 60 80 100 120 140 160 [C]


100 200 300 [F]
A0011572

Fitted electrodes

Measuring electrodes, reference electrodes and empty pipe detection electrodes available as standard with: 1.4435/304L Alloy C-22 Flange connection: EN 1092-1 (DIN 2501) (Dimensions to DIN 2501, DN 65 PN 16 exclusively to EN 1092-1) ANSI B16.5 AWWA C207 AS Electrodes with 1.4435, Alloy C-22: 0.3 to 0.5 m (11.8 to 19.7 in) (All data refer to parts in contact with medium)

Process connections

Surface roughness

Human interface
Display elements Liquid crystal display: unilluminated, two-line, 16 characters per line Display (operating mode) preconfigured: volume flow and totalizer status 1 totalizer Local operation via three keys (S, O, F) Operation via HART protocol and FieldCare

Operating elements Remote operation

34

Endress+Hauser

Proline Promag 10L

Certificates and approvals


CE mark The measuring system is in conformity with the statutory requirements of the EC Directives. Endress+Hauser confirms successful testing of the device by affixing to it the CE mark. The measuring system meets the EMC requirements of the "Australian Communication and Media Authority (ACMA)". WRAS BS 6920 ACS NSF 61 KTW/W270

C-tick mark

Drinking water approval

Other standards and guidelines

EN 60529 Degrees of protection by housing (IP code). EN 61010 Protection Measures for Electrical Equipment for Measurement, Control, Regulation and Laboratory Procedures. IEC/EN 61326 "Emission in accordance with requirements for Class A". Electromagnetic compatibility (EMC requirements). ANSI/ISA-S82.01 Safety Standard for Electrical and Electronic Test, Measuring, Controlling and related Equipment - General Requirements. Pollution degree 2, Installation Category II. CAN/CSA-C22.2 No. 1010.1-92 Safety requirements for Electrical Equipment for Measurement and Control and Laboratory Use. Pollution degree 2, Installation Category II

Ordering information
Your Endress+Hauser service organization can provide detailed ordering information and information on the order codes on request.

Accessories
Various accessories, which can be ordered separately from Endress+Hauser, are available for the transmitter and the sensor. Your Endress+Hauser service organization can provide detailed information on the order codes in question.

Documentation
Flow measurement (FA005D/06) Operating Instructions Promag 10 (BA00082D/06)

Registered trademarks
KALREZ and VITON Registered trademarks of E.I. Du Pont de Nemours & Co., Wilmington, USA TRI-CLAMP Registered trademark of Ladish & Co., Inc., Kenosha, USA HART Registered trademark of the HART Communication Foundation, Austin, USA FieldCare, Fieldcheck, Field Xpert, Applicator Registered or registration-pending trademarks of Endress+Hauser Flowtec AG, Reinach, CH

Endress+Hauser

35

Instruments International
Endress+Hauser Instruments International AG Kaegenstrasse 2 4153 Reinach Switzerland Tel.+41 61 715 81 00 Fax+41 61 715 25 00 www.endress.com info@ii.endress.com

TI00100D/06/EN/14.11 71136637 FM+SGML6.0 ProMoDo