INDICATIONS FOR USE

The InterPlate IFD is indicated for intervertebral body fusion of the spine in skeletally

mature patients. The device system is designed for use with autograft to facilitate fusion. One device is used per intervertebral space. The InterPlate C-P Cervical System made with PEEK OPTIMA polymer is intended for use at one level in the cervical spine, from C3 to T1, for the treatment of cervical disc disease (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies). The cervical device is to be used in patients who have had six weeks of non-operative treatment. Refer to the package insert for a complete list of warnings, precautions, indications, and contraindications.

Part #
RSB0712P RSB0812P RSB0912P RSB1012P RSB0715P RSB0815P RSB0915P RSB1015P RSB0007P RSB0008P RSB0009P RSB0010P RSB3111P RSB3113P RSB402P RSB403P

Description
7mm PLATE (12mm WIDE) 8mm PLATE (12mm WIDE) 9mm PLATE (12mm WIDE) 10mm PLATE (12mm WIDE) 7mm PLATE (15mm WIDE) 8mm PLATE (15mm WIDE) 9mm PLATE (15mm WIDE) 10mm PLATE (15mm WIDE) 7mm COVER ASSEMBLY 8mm COVER ASSEMBLY 9mm COVER ASSEMBLY 10mm COVER ASSEMBLY BONE SCREW STANDARD BONE SCREW RESCUE SIZER 7 x 12 SIZER 8 x 12

Part #
RSB404P RSB405P RSB407P RSB408P RSB409P RSB410P RSB411P RSB412P RSB413P RSB414P RSB415 RSB416P RSB418 RSB420 RSB421 RSB422P RSB500P

Description
SIZER 9 x 12 SIZER 10 x 12 SIZER 7 x 15 SIZER 8 x 15 SIZER 9 x 15 SIZER 10 x 15 INSERTER CENTER PUNCH BONE SCREW DRIVER COVER DRIVER HANDLE COVER DRIVER SHAFT ROUND HOLE GUIDE TUBE FLEX DRIVER COVER SCREW EXTRACTOR BONE SCREW EXTRACTOR PLATE EXTRACTOR STERILIZATION TRAY

INDICATIONS FOR USE

The height of the ridge is matched to the length of screw travel in the slot. When the ridge is fully embedded and the screw has reached the end of the slot, the device rests on flats on the caudal surface.

SIZES
13 and 15 mm inner widths 17.5 and 19.5 mm overall widths 7, 8, 9, 10 mm heights

Provides the benefits of an interbody spacer and an anterior cervical plate, without any of the drawbacks Combined fixation plate and spacer saves time with one step implantation Lower profile than anterior cervical plates and reduces gardening Can be implanted next to existing plates or artificial discs Accommodates a large amount of graft material Translating caudal screw and unique wedge reduce stress shielding of graft material Secure anti-backout screw cover accommodates variable screw angles Self-drilling, self-tapping screws with guide tube or freehand placement Multiple sizes Radiopaque markers Manufactured from PEEK-OPTIMA polymer to facilitate fusion visualization

CRITICAL DIMENSIONS

InterPlate C-P Dimensions Implant dimensions are shown in the following figures and Table 1.

Table 1
Dimension (mm)
A) Nominal Height B) Posterior Height C) Overall width D) Width E) Depth

7x12
7.0 6.0 17.5 13.0 14.2

8x12
8.0 7.0 17.5 13.0 14.2

9x12
9.0 8.0 17.5 13.0 14.2

10x12
10.0 9.0 17.5 13.0 14.2

7x15
7.0 5.8 19.5 15.0 15.8

8x15
8.0 6.8 19.5 15.0 15.8

9x15
9.0 7.8 19.5 15.0 15.8

10x15
10.0 8.8 19.5 15.0 15.8

Standard Bone Screws: 3.1mm thread diameter x 11.0mm thread length Rescue Bone Screws: 3.1mm thread diameter x 13.0mm thread length

SCREW ROM
Round Hole Sagittal: 33 total included angle Slot Sagittal: 43 total included angle Slot Transverse: 9 total included angle