You are on page 1of 5

Product Data

Intra-oral X-ray unit

Endos ACP

Tubehead features
High voltage circuit type Rated output voltage Tubehead current Tubehead power Total filtration Maximum deviation of output current Pre-heating time Interval between exposures (duty cycle) Minimum focus to skin distance Single phase, self-rectifying, grid controlled 70 kVp 6% @ 230V 70 kVp 6% @ 120V 8 mA 430 VA 2 mm Al eq. @ 70 kV 13% 100 ms 1:32

20 cm 30 cm with optional collimator cone X-ray field (at collimator tip) Circular, diameter 6 cm Rectangular 35x45 mm with optional diaphragm Leakage radiation @ 1m 0.1 mGy/h Technical factors for leakage radia- 70 kV, 8 mA, 1 s, 1:32 duty cycle tion Tube insert type OCX/70-G Focal spot 0.8 (IEC 336) Anode material Tungsten Anode angle 19 Inherent filtration 0.5 mm Al eq. @ 70 kV Anode thermal capacity 6 kJ

Scissors and extension arm features


Material Type of painting Mounting type Aluminum Glossy - Wall mount with local timer - Wall mount with remote X-ray pushbutton - Ceiling mount - Mobile unit on trolley - 30 cm - 60 cm - 80 cm

Choice of extension arm length (only for wall mounted units)

Product data Endos ACP Rev.0

Page 1

Timer features
Exposure times Exposure times control Timer accuracy Exposure times visualization Exposure button Stand-by function Number of anatomic programs Choice of patient size Choice of anatomic region From 0.02 to 3.2s in 33 steps Microprocessor controlled exposure times, with automatic compensation of line voltage fluctuations 10% or 20 ms (whatever is greater) @ 230V 10% or 32.2 ms (whatever is greater) @ 120V Digital display With extensible cable (3m) The system enters a low-consumption state after 5 minutes of inactivity 30 pre-set times 3 choices (small, medium, large) - Incisors/canines (upper an lower) - Premolars (upper an lower) - Upper molars - Lower molars - Bite wing Yes, with pushbutton on keyboard Yes Yes Yes Provisions for connection of: - Remote x-ray pushbutton - Unit powered signal lamp - X-ray active signal lamp

Automatic dose reduction for videoradiography systems Film speed setting CCD/ phosphor-plates setting Visualization of number of exposures External connections

Electrical characteristics
Line voltage Line frequency Absorbed current Absorbed power Line resistance Max line voltage regulation 230V 10% 50 Hz 4A rms @ 230V 920 VA @ 230V 0.8 Ohm N/A 120V 10% 60 Hz 8A rms @ 120V 1.1 kVA @ 120V N/A 3% @ 132V

Environmental conditions
Operating conditions Conditions for transport and storage Temperature: Humidity: Temperature: Humidity: Pressure: +10 to +40 Celsius 30% to 75% -20 to +70 Celsius up to 95 % non condensing up to 630 hPa

Product data Endos ACP Rev.0

Page 2

Standard and regulations


This product is designed and produced to meet the following standards: EN 60601-1, EN 60601-1-1, EN 60601-1-2, EN 60601-1-3, EN 60601-1-4, EN 60601-2-7, EN 60601-2-28, EN 60601-2-32, EN 1441 CE symbol grants the product compliance to the European Directive for Medi0434 cal Devices 93/42 as a class IIB device.

Keyboard functions

1 DISPLAY 2 7 8 9 13 17 14 15 16 10 11 12 3

5 6

1. 2. 3. 4. 5. 6. 7. 8. 9.

Ready for exposure signal lamp Manual decrease of exposure times Manual increase of exposure times X-ray emission warning lamp Video radiography signal lamp Video radiography selection Bite wing signal lamp Upper molars signal lamp Lower molars signal lamp

10. Premolars (upper and lower) signal lamp 11. Incisors/canines (upper and lower) signal lamp 12. Anatomic selection 13. Large patient size signal lamp 14. Medium patient size signal lamp 15. Small patient size signal lamp 16. Patient size selection 17. Exposure enable button Page 3

Product data Endos ACP Rev.0

Site planning (quotes in mm and inches)


WALL MOUNT

CEILING MOUNT

Product data Endos ACP Rev.0

Page 4

MOBILE UNIT ON TROLLEY

Note: Products are continuously under review in the light of technical advancement. The actual specification may therefore be subject to improvement or modification without notice.

VILLA SISTEMI MEDICALI s.p.a. 20090 BUCCINASCO (MI) ITALY, Via delle Azalee, 3 Tel. +39 02 48859.1, Fax +39 02 4881844 Sito web: www.villasm.com

Product data Endos ACP Rev.0

Page 5