Professional Documents
Culture Documents
IM O1OM/O1/e/06.96
IMA Limited Parkwell House Otley Road Guiseley West Yorkshire LS20 8BH England Tel: +44 (0)1943 878877 Fax:+44 (0)1943 879988 e-mail: sales@ima.co.uk Website: www.ima.co.uk
1
Endress+Hauser reserves the right to change or modify our product appearance and specifications at any time and without notice. Therefore, information in this document is subject to change without notice and does not represent a commitment on the part of Endress+Hauser No part of this manual may be reproduced or transmitted in any form or by any means, electronic or mechanical, including photocopying and recording, for any purpose without the express written permission of Endress+Hauser.
Contents
1.0 General System Information
1.1 Unpacking and Inspection............................................................5 1.2 Introduction......................................................................5 1.3 MMY170 Series Analyser................................................6 1.4 DY 75 Probe....................................................................7 1.5 Theory of Operation ........................................................8 1.6 System Specifications .....................................................9 1.6.1 Probe Specifications...............................................9 1.6.2 Analyser Specifications ..........................................9
2.0
2.1
Installation Guidelines................................................. 11
Mounting the Probe......................................................................11 2.1.1 Inline Mounting .......................................................11 2.1.2 Bypass Mounting....................................................11 2.1.3 Mounting in a Sample System................................12 2.2 Mounting the Analyser.....................................................13 2.2.1 Checking the Set-Up of the Monorack Power Supply 13 2.3 Analyser Connection .......................................................15 2.3.1 Analyser in Monorack Enclosure - Connection Detail 16 2.4 Description of the Hygrotec MMY 170 Programming Matrix17 2.4.1 Special Functions of the Push Buttons...................18 2.4.2 Locking/Unlocking/Factory Reset Functions...........19 2.4.3 Functions of the Matrix ...........................................20
3.0 Troubleshooting..................................................... 33
3.1 Calibration Table..........................................................................33 3.2 Error Codes ....................................................................34 3.2.1 Error Code E 201 ...................................................34 3.2.2 Error Code E 202 ...................................................34 3.2.3 Error Code E 401 ...................................................35 3.2.4 Error Code E 402 ...................................................35 3.2.5 Error Code E 400 ...................................................36 3.2.6 Error Codes E 602 to E 607 ...................................36 3.2.7 Error Codes E 608 and E 610 ................................36 3.2.8 Error Codes E 613 and E 614 ................................36 3.2.9 Error Codes E 106, E107, E 301, E 302, E 600,and E 601..............................................................36 3.3 Probe Replacement/Recalibration...................................37
3.3.1 Probe Removal/Replacement ........................................37 3.3.2 Repackaging for Shipment.............................................37 3.3.3 Reprogramming of the Analyser Upon Return of a Probe from the Factory .....................................................................38
1.2 Introduction
The Hygrotec MMY 170 is a microprocessor controlled dewpoint/ppmv analyser Introduction consisting of a Endress+Hauser gold/aluminum oxide probe (DY 75) and the MMY 170 electronics module (analyser). The Hygrotec measures moisture in dewpoint temperature in the range of --90C to +20C ppmv from 0.01 to 9999, and temperature for compensation. Its four digit LCD displays the measured value. Process control is obtained with two current outputs, two voltage outputs, two alarm relays, and one system alarm relay. The Hygrotecs direct user access provides simple setting and editing of the users parameters utilizing its six push buttons and matrix-style programming structure. Multi-channel configuration is obtained utilizing the E+H bus and the computer interface ZA 672. (Refer to Field Communication, ZA 672 Modbus Gateway.) To set parameters more conveniently, the Commulog VU 260 Z can be connected to each module. Any sensor is affected by temperature. In order to minimize these effects, temperature compensation is carried out using factors which are calculated during factory calibration. (See page 33.) The measuring system consists of: o up to 64 Commutec transmitters or 128 measuring points o ZA 672 gateway o personal computer at the command port o Commulog o PC communication Software
Fig. I
Hygrotec MMY 170 1.3 MMY 170 Series Analyser Monorack Enclosure
The Hygrotec MMY170 analyser in the following enclosures:
Fig 2
Dimensions in mm
1.5
Theory of Operation
The moisture sensor of each DY 75 probe changes capacitance with moisture changes
in the process stream. The capacitance changes are converted into an interference-free PFM (Pulse Frequency Modulation) signal by the probes electronics. This technology allows the use of up to 1000 metres of ordinary interconnecting cable between the probe and the analyser, since noise does not interfere with the (PFM) signal. Temperature is measured with platinum RTD and values are transmitted via PFM signals.
Fig. 5
Calibration Curve
Ambient and storage temperature: -60 C to +50 C Nominal operating temperature: -20 C to +40 C Maximum operating pressure: 350 bar Helium leak-rate: < 10-7 mbar I/s Maximum gas velocity: 50.0 m/s at 1 bar 5.0 m/s at 10 bar 0.5 m/s at 100 bar Signal transmission: Pulse Frequency Modulation (PFM) via three-wire cable; Moisture signal, I-signal, and common return Probe Mounting adapter: Stainless steel 1.4571 with G 1/2 thread (DIN/ISO 228) and Oring seal out of Fluor Caoutchouc (SMK) Rating: I P 65 Torque when tightening: 50 Nm Wrench width: 30 mm Optional: Weight: Protective cap with 10 m filter 830 grams
1.62 Analyser Specifications Electronics: Microprocessor controlled Temperature Unit: C Moisture units. Dewpoint temperature in C and concentration in ppmv (concentration in ppmw optional) Instrument range: -100 C to +20 C standard, 0.01 to 9999 ppmv and ppmw Display: Four character LCD with bar graph 010 100% and two control fields
10
Inline Mounting
Fig.7
Mount the probe vertically such that the sensor tip points down. The probe is typically mounted in a G pipe thread connection. Insure that the tip of the probe does not touch the inside wall of the pipe.
Bypass Mounting
Fig. 8 Caution! Close both bypass valves before servicing the probe.
2.12 Bypass Mounting Utilizing a bypass line assures no interruption of the process stream during servicing of the probe. To mount the probe in a bypass, the above procedure for in-line mounting should be followed. Caution should be exercised when reopening the bypass after a probe has been serviced as this may contaminate the main stream. Stream Caution !
11
If the gas velocity or temperature of an application exceeds the probe specifications on page 9 or if particulates are entrained in the stream, a sample system should be used. In a sample system with low dewpoints, all wetted parts should be of stainless steel. (For further details, refer to the General Hints for Trace Moisture Measurement manual.) While the probes measured value is not flow sensitive, the flow rate in a sample system influences the response time of the system. A flow rate of 30 to 150 I/h through the flow cell is ideal. In addition, the sample gas temperature in the flow cell must remain above the dewpoint temperature to prevent condensation. (If dewpoint is 0 C, maintaining a +10 C difference is recommended.) Figure 9 illustrates a basic sample system. PRIOR TO REMOVING THE PROBE, BE CERTAIN THAT THE FLOW CELL IS NOT U
(a) If measurement is desired under system pressure: Valve 1 is fully open and the sample flow is regulated by Valve 2.
(b) If measurement is desired under atmospheric pressure: Valve 2 is fully open and the sample flow is regulated by Valve 1.
NOTE: 1,5 m of 6 mm tubing, if vented to atmosphere, will insure accuracy of sample in the
CAUTION! PRIOR TO REMOVING THE PROBE< BE CERTAIN THAT THE FLOW CELL IS NOT UNDER PRESSURE!
12
13
Fig. 10
Hygrotec electronics module straight out of the Monorack enclosure (refer to Figure 10). (c) Upon removal of the electronics module, locate the two flat head screws and loosen them. (d) Grasping the Monorack enclosure with both hands carefully pull the black terminal connection section away from the front section of the enclosure and set it aside (refer to Figure11).
Fig. 11
14
Fig. 12
NOTE: THERE ARE TWO OTHER VERSIONS AVAILABLE, IF SPECIFIED ON THE ORDER FOR THE ENCLOSURE: 1. 100/200VAC,or 2. 24/48VAC
15
Fig. 13 Fig.14
Looking onto the contact pins of the Hygrotec electronics module or the connection side of the female connector in the mounting rail.
Fig. 15
16
This section is relative only to the Monorack enclosure. The contact pins of the Hygrotec electronics module plug into the black terminal connection section of the Monorack enclosure. Actual hard wiring of the analyser to its probe, outputs and power cord is performed at the terminal connection section (refer to Figure 16). When making these connections, all wiring should conform with local electrical codes and standards. Recheck all connections for security and correct placement of wires. (Severe damage can result from incorrect wiring.)
Fig. 16
17
Fig. 17
18
Fig. 18
Note that seven (7) matrix locations are for display only and may not be changed by the user (refer to Figure 17). For convenience, input (or programming) fields have a flashing digit in the display, whereas display only fields do not. The display only fields are as follows:
o V H 00 = normal display (in dewpoint 0C) o V H 01 = normal display (concentration in ppmv) o V H 08 = displays probe frequency o V H 40 = displays temperature o V H 48 = displays temperature frequency o V H 90 = during a system alarm, displays the error code for the fault encountered o V H 91 = during normal operation, the most recent error code is displayed for reference o V H 93 = displays the factory issued reference number designating the device type and software version
(e)
Default Values A default value is assigned to each programmable matrix field. The values are present after a reset to factory programmed data has been executed (see V H 95).
19
Location in Matrix VH 89
Description and Function By selecting any value <670 or >679, programming of the matrix is prevented. This system locking feature prevents either unauthorized or accidental modification of setpoints or process control parameters. Values in the fields can be reviewed but not changed. Entering a value from 670 to 679 unlocks the matrix for programming. Default: 670
2. Reset to Factory Programmed Data Location in Matrix VH 95 Description of Function This field allows resetting of all matrix parameters to their factory (default) settings by entering a number from 670 to 679. This field should only be programmed during service. Default: 0
Caution! Resetting to default values clears the calibration table (programmed in V H 22, V H 23 and V H 24) and compensation table (programmed in V H 62, V H 63 and V H 64).
20
Hygrotec MMY 170 2.4.3 Functions of the Matrix (refer to Figure 17)
The following describes specifically each field of the matrix grouped by common function:
DISPLAY AND OUTPUT MODE 1. Dewpoint Display Location in Matrix VH 00 2. ppmv Display Location in Matrix VH 01 Description and Function The ppmv is shown. To operate the analyser in ppmv select2 in V H 80. This enables all settings (alarm, Output) to refer to ppmv. Description and Function This is the normal display of the analyser when in operation. The dewpoint is shown in C.
3. Selecting the Analogue Output 1 Range Location in Matrix VH 03 Description and Function This field provides selection of the analogue output range 0 3 corresponding to the dewpoint. For 0-20 mA current output, select 0w; for 4-20 mA, select 1 Default: 0
4. Damping Location in Matrix VH 04 Description and Function In this field, the response of the transmitter to changes can be adjusted in the order of 0 to 100 seconds. Default: 4 5. Selecting the Analogue Output Offset (0/4 mA) Location in Matrix VH 05 Description and Function The dewpoint value corresponding to the analogue output offset (0/4 mA) is entered here. Default: -80 C
21
Hygrotec MMY 170 6. Selecting the Analogue Output 1 Span (20 mA) Location in Matrix VH 06 Description and Function The dewpoint value corresponding to the analogue output span (20 mA) is entered here. Default: +20 C
7. Selecting the Status of the Current Output During System Alarm Location in Matrix VH 07 Description and Function This field determines the status of the current output during a system fault. That is, the current output will hold (freeze) either at -10% of range, at 110% of range, or at the last measured value when 0, 1 or 2, respectively are selected for this value.
,
Default: 2
8. Displaying the Probe Frequency Location in Matrix VH 08 Description and Function This field continuously displays the moisture sensor frequency
ALARM SET POINTS 9. Selecting Alarm Setpoint 1 Location in Matrix Description and Function The first process value (dewpoint/temperature) alarm VH setpoint (C) is displayed and programmed in this location. 10
Default: -40C
22
Hygrotec MMY 170 10. Selecting the Status of Alarm Setpoint 1 (Min or Max; Fail-Safe Mode) Location in Matrix Description and Function The value in this field determines whether the first VH alarm relay is a minimum or maximum alarm. That is, 11 when 0 is selected, the alarm relay will trip if the dewpoint temperature is less than the setpoint value (minimum). When 1 is selected, the alarm condition exists if the dewpoint is greater than the setpoint value (maximum). Default: 1 11. Selecting the Hysteresis for Alarm Relay 1 Location in Matrix Description and Function To prevent chattering of the relay, this value is the VH amount above a minimum setpoint or below a 12 maximum setpoint that the dewpoint/temperature must be before the first alarm relay changes status. Default: 2 C 12. Selecting the Status of Alarm Relay 1 During System Alarm Location in Matrix Description and Function This field indicates the state of the first process alarm VH during a system fault. Selecting 0 will cause the 13 process alarm to be de-energized (alarm state) when the system alarm is de-energized. Selecting 1 will put the process alarm relay in the state that corresponds to the analogue output (-10% of range, 110% of range, or holding the last measured value; see V H 0 7, programming step 7, page 21 or V H 4 7, programming step 26 on page 26). Default: 0 13. Relay 1 Assignment to Outputs Location in Matrix Description and Function Assignment of Relay 1: VH 14 If 1, relay connected to current output 1, moisture. If 2, relay connected to current output 2, temperature. Default: 1
23
Hygrotec MMY 170 LINEARIZATION 14. Confirming or Clearing Probe Calibration Data Location in Matrix VH 20 Description and Function This field of the matrix controls the calibration data for the DY 75 probe. This matrix location either confirms the existing calibration data (0) or clears the calibration data table ( 1). Default: 1 Caution I Caution should be exercised since entering 1 (followed by pressing E) will erase all calibration values currently stored. 15. Selecting the Calibration Point Location in Matrix VH 22 Description and Function This location indicates the number of the particular calibration points that is to be examined or changed (NOTE: 13 calibration points are typically utilized. However, if the user desires a wider calibration range, up to 15 calibration points are available. See example table on page 33.) Default: 1
16. Verifying or Changing the Dewpoint of Selected Calibration Point Location in Matrix VH 23 Description and Function The value in this field is the dewpoint (in C) of the calibration point chosen in V H 2 2, programming step 15, above.
17. Verifying or Changing the Frequency of Selected Calibration Point Location in Matrix VH 24 Description and Function This value is the probe frequency corresponding to the dewpoint given in V H 2 3, programming step 16, above.
24
Location in Matrix VH 25
20. Verifying or Changing the Dewpoint of Selected Calibration Point Location in Matrix VH 38 Description and Function
When the value 6 has been selected at matrix location V H 8 0 programming step 40, page 29), selection of this field generates a 0 mA signal for calibration. Connecting a mA-meter to the output the correct current (0 mA) can be adjusted by increasing or decreasing the displayed digits.
25
Hygrotec MMY 170 21. Adjusting the Current Output 1 (20 mA) During Calibration Location in Matrix VH 39 Description and Function Selection of this field assists during calibration 39 generating a 20 mA signal. Connecting a mA-meter to the output the correct current (20 mA) can be adjusted by increasing or decreasing the displayed digits. This function is activated when 6 in V H 8 0 (programming step 40, page 29) is selected.
DISPLAY TEMPERATURE
22. Display Temperature Location in Matrix Description and Function Temperature is shown in C VH 40
23. Selecting the Analogue Output 2 Range Location in Matrix VH 43 Description and Function This field provides selection of the analogue output range corresponding to the temperature. For 0-20 mA current output, select 0; for 4-20 mA, select 1. Default: 0
24. Selecting the Analogue Output 2 Offset (0/4 mA) Location in Matrix VH 45 Description and Function The temperature value corresponding to the analogue output offset (0/4 mA) is entered here. Default: 0
25. Selecting the Analogue Output 2 Span (20 mA) Location in Matrix VH 46 Description and Function The temperature value corresponding to the analogue output span (20 mA) is entered here. Default: 100 0C
26
Hygrotec MMY 170 26. Selecting the Status of the Current Output During System Alarm Location in Matrix VH 47 Description and Function This field determines the status of the current output during a system fault. That is, the current output will hold (freeze) either at -10% of range, at 110% of range, or at the last measured value when 0, 1, or 2 are selected for this value.
respectively
Default: 2
27. Displaying the Temperature Frequency Location in Matrix Description and Function This field continuously displays the temperature VH frequency. 48
ALARM SETPOINTS 28. Selecting Alarm Setpoint 2 Location in Matrix VH 50 Description and Function The second process (moisture/temperature) alarm setpoint (C dewpoint) is displayed and programmed in this location. Default: 25 0C
29. Selecting the Status of Alarm Setpoint 2 (Mm or Max; Fail-Safe Mode) Location in Matrix VH 51 Description and Function The value in this field determines whether the second alarm relay is a minimum or maximum alarm. That is, when 0 is selected, the alarm relay will trip if the dewpoint/temperature is less than the setpoint value (minimum). When 1 is selected, the alarm condition exists if the dewpoint is greater than the setpoint value (maximum). Default: 1
27
Hygrotec MMY 170 30. Selecting the Hysteresis for Alarm Relay 2 Location in Matrix VH 52 Description and Function To prevent chattering of the relay, this value is the amount above a minimum setpoint or below a maximum setpoint that the dewpoint/temperature must be before the first alarm relay changes status. Default: 2C
31. Selecting the Status of Alarm Relay 2 During System Alarm Location in Matrix VH 53 Description and Function This field indicates the state of the second process alarm during a system fault. Selecting 0 will cause the process alarm to be de-energized (alarm state) when the system alarm is de-energized. Selecting 1 will put the process alarm relay in the state that corresponds to the analogue output (-10% of range, 110% of range, or holding at the last measured value; see V H 4 7, programming step 26, page 26 or V H 0 7, programming step 7, page 21). Default: 0
32. Relay 2 Assignment During System Alarm Location in Matrix VH 54 Description and Function Assignment of Relay 2: If 1, relay connected to current output 1, moisture. If 2, relay connected to current output 2, temperature. Default: 2
28
Hygrotec MMY 170 TEMPERATURE COMPENSATION 33. Confirming Probe Calibration Data (refer also to table on page 33) Location in Matrix VH 60 Description and Function This field of the matrix controls the compensation data for the DY 75 probe. This matrix location confirms the existing calibration data ( 0). Default: 0 34. Selecting the Compensation Point Location in Matrix VH 62 Description and Function This location indicates the number of the particular compensation factors that are to be examined or changed. NOTE: 7 calibration rows are utilized. See example table on page 33.
35. Verifying or Changing of Selected Compensation Point Location in Matrix VH 63 Description and Function
The value on this field is the first compensation point of the row chosen in V H 6 2 programming step 34 above
36. Verifying or Changing of Selected Calibration Point Location in Matrix VH 64 Description and Function
The value on this field is the second compensation point of the row chosen in V H 6 2 programming step 34 above
37. Moving Through the Calibration Table (Verifying or Changing the Values) Location in Matrix VH 64 Description and Function By using the 1 or buttons (followed by pressing E), the user may step sequentially through the table of compensation rows to verify or change each one. In this case, the user is moved automatically to 1) V H 6 3 to program the desired compensation value. After programming all the compensation values, the user must move to matrix location V H 6 0 to activate the programmed compensation table.
29
Hygrotec MMY 170 38. Adjusting the Current Output 2 (0 mA) During Calibration Location in Matrix VH 78 Description and Function When the value7 has been selected at matrix location V H 8 0 (programming step 40, below), selection of this field generates a 0 mA signal for calibration. Connecting a mA-meter to the output, the correct current (0 mA) can be adjusted by increasing or decreasing the displayed digits. 39. Adjusting the Current Output 2 (20 mA) During Calibration Location in Matrix VH 79 Description and Function Selection of this field assists during calibration generating a 20 mA signal. Connecting a mA-meter to the output, the correct current (20 mA) can be adjusted by increasing or decreasing the displayed digits. This function is activated when 7 in V H 8 0 (programming step 40, below) is selected.
MODE OF OPERATION 40. Changing from Normal Mode to Simulation/Calibration Mode Location in Matrix VH 80 Description and Function This field provides the capability for the user to select either dewpoint (1) or ppmv (2) as the normal operation of the analyser and perform various simulations (see V H 9 6, V H 9 8, and V H 9 9, programming steps 47, 48, and 49, respectively, page 31 and 32) and electronic calibrations (V H 3 8, programming step 20, page 24,and V H 3 9, programming step 21, page 25. If 7 is selected, V H 7 8, programming step 38, page 29, and V H 7 9, programming step 39, page 29). NOTE: When6 or 7 is selected, the system alarm LED (:) flashes, indicating that the analyser is not in a normal operation mode. The values1 or2 must be selected for normal moisture measurement and analyser operation.
~
30
Hygrotec MMY 170 41. Locking/Unlocking the Matrix for Programming Location in Matrix Description and Function By selecting any value <670 or >679, programming of VH the matrix is prevented. This system locking feature 89 prevents either prevents either unauthorized or accidental modification of setpoints or process control parameters. Values in the fields can be reviewed but not changed. Entering a value from 670 to 679 unlocks the matrix for programming. Default: 6 10 SERVICE AND SIMULATION 42. Displaying the Error Code During System Alarm Location in Matrix Description and Function In the event of a system fault, this field displays the VH diagnostic error code for the fault encountered 90
43. Displaying the Most Recent Error Code Location in Matrix Description and Function When a system fault condition is cleared, the value of VH 9 1 the error code is stored in this location. That is, 91 during normal operation, the most recent error code is displayed for reference. This field may be cleared by pressing the E button.
44. Identification Field Location in Matrix VH 93 45. E+H Bus Address Location in Matrix VH 94 Description and Function
This field indicates a factory issued reference number that designates the device type and software version
31
Hygrotec MMY 170 46. Reset to Factory Programmed Data Location in Matrix Description and Function This field allows resetting of all matrix parameters to VH their factory (default) settings by entering a number 95 from 670 to 679. This field should only be programmed during service. Default: 0 Caution! Resetting to default values clears the calibration tables (programmed in V H 2 2 programming step 15, page 23, V H 2 3, programming step 16, page 23, and V H 2 4, programming step 17, page 23) and the compensation tables (programmed in V H 6 2, programming step 34, page 28, V H 6 3, programming step 35, page 28, and V H 6 4 programming step 36, page 28).
~,
47. Simulating Probe Frequency (for test purposes) Location in Matrix Description and Function VH This field (when V H 8 0 is set to 6 programming step 40, page 29) allows user entry of a particular 96 frequency value (i.e. simulating a probe). The user may then observe both the dewpoint calculated (and displayed in V H 9 8, programming step 48, below) and the analogue output signal generated by the analyser (and displayed in V H 9 9, programming step 49, page 32). NOTE: The frequency simulation is only valid for moisture. 48. Simulating Probe Dewpoint (for test purposes) Location in Matrix VH 98 Description and Function As with V H 9 6 (programming step 47, above), this field allows user entry of a dewpoint value (simulated analyser software result) for testing the analogue output, alarms or other analyser functions. NOTE: Selecting 7 in V H 8 0 allows simulation of temperature value.
32
Hygrotec MMY 170 49. Simulating Current Output (for test purposes) Location in Matrix VH 99 Description and Function
This field allows simulation of a particular current output by programming a value (in mA) that is then transmitted to the analogue output terminals and verified by the corresponding frequency and dewpoint values in V H 9 6 and V H 9 8, respectively (programming steps 47 and 48 page 31 NOTE: Selecting 7 in V H 8 0 allows simulation of output 2
33
3.0 Troubleshooting Standardised Calibration Table 3.1 Calibration Table Typical table. Correct entries may have to be checked while troubleshooting.
Calibration Point Number 1 2 3 4 5 6 7 8 9 10 11 12 13 Frequency/Hz 188.9 187.3 186.2 184.3 180.1 169.1 160.7 152.6 134.1 110.7 98.0 87.3 76.1 Dewpoint/C -100.0 -90.0 -80.0 -70.0 -60.0 -50.0 -40.0 -30.0 -20.0 -10.0 0.0 +10.0 +20.0
34
The first two error codes are relative to the Hygrotec's calibration table. To aid the user in visualizing the structure of a typical calibration/compensation table, one is provided above.
This error code indicates that the measured frequency is out of calibration table range at the high dewpoint or <20 Hz.
(a) Following programming step 8, page 21, V H 0 8, read the probe frequency (b) If the probe frequency displayed is <20 Hz but >zero, the probe should be returned to the factory. (C) If the frequency displayed is >20 Hz, perform programming step 15, page 23, V H 2 2, and select the lowest frequency calibration point (factory program = point #12). Follow programming step 17, page 23, V H 2 4, and read the table value. If the table value is higher than the actual frequency V H 0 8, the probe is out of calibration table range and should be returned to the factory for recalibration. (d) If the frequency displayed is in the range of 15 to 17 Hz, the sensor is shorted. Replace the probe.
3.2.2 Error Code E 202
This error code indicates that the measured frequency is out of calibration table range at the lowest dewpoint or >3000 Hz.
(a) Following programming step 8, page 21, V H 0 8, read the probe frequency. (b) If the probe frequency displayed is >3000 Hz, the probe should be returned to the factory (c) If the probe frequency displayed is <3000 Hz, perform programming step 15, page 23, V H 2 2, and select the highest frequency calibration point (factory program = point #1). Follow programming step 17, page 23, V H 2 4, read the table value. If the table reading is lower than the actual frequency V H 0 8. the probe is out of calibration table range and should be returned to the factory for recalibration.
35
(a) Following programming step 8, page 21 V H 0 8, read the probes moisture frequency. If the frequency is 0.5 Hz, the analyser is not receiving a frequency signal from the probe. (b) Check all connections for security and correct placement, and all cables for breaks or severe bends. Read the probe frequency again. If the frequency is still 0.5 Hz, the probe is probably at fault and should be returned to the factory for repair. (c) If you have another probe that is in good working order, disconnect power from the analyser, disconnect the suspect probe, and connect the working probe to the analyser. Power up the analyser. If the error code E 4 0 1 appears in the display and the frequency is 0.5 Hz when reading V H 0 8, the analyser might also require repair. Therefore, the suspect probe and the analyser should be returned to the factory. However, upon power up with the working probe, if no error code appears and the probe and analyser function satisfactorily, only the suspect probe should be returned to the factory for repair.
3.2.4 Error Code E 402
This error code indicates that input d2 is not receiving a signal to the analyser.
(a) Following programming step 27, page 26, V H 4 8, the display should read three different frequency values in sequence. If the display does not show three different frequencies in sequence, the analyser is not receiving a frequency signal from the probe. The last value read remains on the display. (b) Check all connections for security and correct placement, and all cables for breaks or severe bends. Read the probe frequency again, If the frequency is still 0.5 Hz, the probe is probably at fault and should be returned to the factory for repair. (c) If you have another probe that is in good working order, disconnect power from the analyser, disconnect the suspect probe, and connect the working probe to the analyser. Power up the analyser. If error code E 4 0 2 appears in the display or the frequency does not change when reading V H 4 8, the analyser might also require repair. Therefore, the suspect probe and the analyser should be returned to the factory. However, upon power up with the working probe, if no error code appears and the probe and analyser function satisfactorily, only the suspect probe should be returned to the factory for repair.
36
This error code indicates that input 1 and input 2 are not receiving a signal to the analyser. (See error codes on page 35.)
These error codes indicate faulty calibration data entry Re-enter calibration data.
3.2.9 Error Codes E 106, E 107, E 301. E 302, E 600, and E 601
Contact factory.
37
38
(g) Continue through all the calibration table entries utilizing the values on the calibration data sheet provided with the recalibrated probe and matrix locations V H 2 3, V H 2 4, and V H 2 5. When you have entered all the values, perform programming step 14, page 23, V H 2 0, pressing 0 (zero) to activate the programmed calibration table values.
(a) Perform programming step 34, page 28, V H 6 2 and enter 1 to modify compensation factors A and B of the first row. (b) Perform programming step 35, page 28, V H 6 3 and enter compensation factor A (listed on the compensation table data sheet). (c) Perform programming step 36, page 28, V H 6 4 and enter compensation factor B (listed on the compensation table data sheet). (d) Perform programming step 37, page 28, V H 6 5 to access the next compensation point (using the + button followed by pressing F). Modify the compensation factor A and the compensation factor B values for the next compensation row (point #2) using V H 6 3 and V H 6 4 as described above.
39
40
Fig. 19
41
Location in Matrix VH 02
Location in Matrix VH 31
Description and Function In addition to the Cs values of 10 hydrocarbons stored in the memory of the MMY 170, the user can program Cs values of 5 hydrocarbon liquids on their own by selecting Cs table no. 0 to 4. NOTE: The Cs values are not visible in fields V H 3 2 to V H 3 7. These fields show the last manually entered Cs values.
4.1.3 Changing from Normal Mode to Simulation/Calibration Mode Location in Matrix VH 80 Description and Function This field provides the capability for the user to select either dewpoint (1), ppmv (2), or ppmw (3) as the normal operation of the analyser and perform various simulations (see V H 9 6, V H 9 8, and V H 9 9, programming steps 47, 48, and 49, respectively, page 31 and 32) and electronic calibrations (V H 3 8, programming step 20, page 24, and V H 3 9 programming step 21, page 25). If 7is selected, V H 7 8, programming step 38, page 29, and V H 7 9, programming step 39, page 29). NOTE: When 6 or 7 is selected, the system alarm LED (:) flashes, indicating the analyser is not in normal operation mode. The value 3 must be selected for moisture measurement in ppmw.
42
Hygrotec MMY 170 4.2 Preprogrammed (10) Saturation Concentration (Cs) Values
VH 31 Position 5 6 7 8 9 10 11 12 13 14
Hydrocarbon Liquid
Cs value in PPMw (at each given temperature 0C 10 C 20 C 454 639 67 101 19 28 53 45 140 94 416 40 500 730 100
50 C 1570
Benzol Butane Hexane Transformer Oil Propane Pentane Propylene Gasoline Butadiene Kerosene
160
43
10C
310 500
20C
450 295 640 61 13700 331 317 282 95 1000 9000 40 600 303
30C
675 383 510
40C
960 1005 110
50C
1350 670 1200 190
60C
1825
305
570
Carbon disulfide CS2 Chlorobenzene Chloroform (Trichloromethane) Cumene 156 Cycloheptatriene Cyolohexane 36 Cyclohexene Cyclopentane 46 1 ,4-Bis(cyclopentyl)butane 2-Cyclopentyloctane Decalin Diethylether Dichloropropane Diethylbenzene Di-isobutylene 2,3-Dimethylbutane 2,3-Dimethyl-1 -butene 2,6-Dimethylheptane 2,4-Dimethylhexane 2,7-Dimethyloctane
219
67 252 86 88 48
159
226 145 58 53 53 48
431 274 192 459 160 180 152 134 431 186 273
574
29
323 301
516 465
219 574
344 756
44
0 0C
Heptane 1-Heptene Hexadecane Hexane 1-Hexene n-Hexylcyclopentane lsobutylacetate lsoprene Isopropylcyclopentane p-Isopropyltoluene Kerosene 2-Methylbutane 2-Methyl-2-butene Methylcyclohexane Methylcyclopentane 27
100C
54 186
200C
96 249 69 101
300C
172 375 123 179 477 141 850 159 415
400C
308 209 317
500C
480 332
600C
52 520 59 223
1075
1295
32
55 61 73
260
400
1 -Methyl-2phenylcyclopentane 105 Methylenechloride (Dichloromettiane) 2-Methylhexane 56 2-Methyloctane 52 3-Methyloctane 50 Nitrobenzene Octane 51 n-.Pentane 1 -Phenyl-5-methyl- 1 -cyclopentene 2-Phenyl-2 ,4 6trEmethylheptane 96 Propane Propylene Propylidenecyclopentane 53
315 350
507
258
Styrene
Tetrachloroethylene Tetrachloromethane (carbon tetrachloride) Tetralin Toluene 316 Transformer Oil 19 28 Trichloroethane
615 70
750 100
965 160
45
0C
Trichloroethylene 1 ,3,5-Trimethyl-2-allylbenzene 1 ,3,5-Trimethylbenzene 2,2,3-Trimethylbutane 27 1 ,3,5-Trimethyl-2-ethylbenzene
10C
200C
20 246 291 160
300C
331 393 184 259 455 201 536
400C
438 519 315 350
500C
60C
57
507 461
332
538
46
p(H2O) mbar
.000014 .000017 .000021 .000026 .000031 .000038 .000046 .000055 .000061 .000080 .000097 .000116 .000139 .000166 .000198 .000235 .000279 .000331 .000392 .000464 .000547 .000644 .000757 .000889 .001042 .001219 .001424 .001661 .001935 .002251 .002614 .003031 .003509 .004058 .004686 .005403 .006222 .007156 .008219 .009428 .010800
Temperature C
-59 -58 -57 -56 -55 -54 -53 -52 -51 -50 -49 -48 -41 -46 -45 -44 -43 -42 -41 -40 -39 -38 -37 -36 -35 -34 -33 -32 -31 -30 -29 -28 -27 -26 -25 -24 -23 -22 -21 -20 -19
p(H2O) mbar
.012357 .014120 .016116 .018311 .020917 .023787 .027020 .030657 .034745 .039334 .044480 .050244 .056695 .063905 .071958 .080942 .090955 .102104 .114505 .128281 .143588 .160561 .179371 .200197 .223236 .248700 .276819 .307843 .342045 .379711 .421118 .466771 .516866 .571865 .63220 1 .698338 .770780 .850066 .936778 1.03154 1.13503
47
p(H2O) mbar
1.24795 1.37106 1.50528 1.65140 1.81040 1.98330 2.17118 2.37522 2.59664 2.83678 3.09704 3.37893 3.68406 4.01413 4.37096 4.75649 5.17276 5.62197 6.10642 6.56533 7.05388 7.57434 8.12849 8.71820 9.34540 10.012 10.720 11.413 12.271 13.118 14.016 14.967 15.975 17.042 18.171 19.366 20.628 21.962 23.371 24.858 26.428
Temperature C
23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60
p(H2O) mbar
28.084 29.829 31.669 33.606
35.646
37.793 40.052 42.428 44.925 47.548 50.304 53.197 56.233 59.419 62.759 66.261 69.930 73.774 77.794 82.011 86.419 91.030 95.851 100.890 106.155 111.655 117.397 123.391 129.646 136.170 142.973 150.067 157.457 165.156 173.176 181.525 190.214 199.256
48
49
50
Reference
To further understand the bus communication option, the following list of literature contains information that might be helpful to the user:
o Field Communication modbus gateway Z4~ 672 With command and data ports for digital control of intrinsically safe field instrumentation (TI 1481001e) o Field Communication ZA 673 PROF/BUS gateway Integrates intrinsically sate Rackbus instrumentation into PROFIBUS networks o Field Communication commutec Operating Program Software package for PC-supported configuration of commutee measuring instruments (TI 1131001e) o Field Communication commugraph presentation software Software for measured value presentation on personal computer and printer (TI 158/OO/e) o Field Communication commutool Commissioning and servicing software for Commutec transmitters (TI 163F100/e) o Field Communication commutec Interfaces Siemens Simatic 85. software package (TI 115/02.92/e) o Field Communication ZA 371 interface adapter Interfaces 5V TTL ~ 20 mA current-loop and RS-232C c~ 20 mA current loop (TI 1531001e) o Field Communication ZA 371 interface adapter Interfaces 5V TIE .~ RS-232C, 5V TTL ~ RS-422 or RS-232C ~ RS-422 (TI 152100/e)
51
52
APPENDIX
53
54
55
IMA Limited Parkwell House Otley Road Guiseley West Yorkshire LS20 8BH England Tel: +44 (0)1943 878877 Fax:+44 (0)1943 879988 e-mail: sales@ima.co.uk Website: www.ima.co.uk
56