You are on page 1of 5

UN-III oximeter module users manual

Address: Rm1,10,14F,Wan'anguojiqiyedaodao,Gexin Road #568,


Wujiashanjingjikaifaqu, Dongxihu District,Wuhan,China
Postcode: 430040
Phone: (86) 027-83083133
Fax: (86)027-83080833
Email: info@unmedtech.com
Website:www.unmedteh.com

Page 1 of 5

UN-III oximeter module users manual

Advanced DSP algorithm minishes the influence of motion


artifact
All digital-signal processing improves accuracy of low
perfusion
Small size, easy to be assembled

Page 2 of 5

UN-III oximeter module users manual

Range and accuracy


Range
Pulse rate: 30 250 bpm
SpO2: 0 100%
Resolution1SD
Pulse rate 30250BPM
2bpm Normal
3bpm Motion or low perfusion
SpO2 70100%
2% Normal
3% Motion or low perfusion
Below 70% Not specified

Environmental Requirements
Operating temperature: 045 C
Storage temperature: -2060 C
Operating Humidity3095%
Storage Humidity10100%
Operating Altitude-5005000m
Storage Altitude-50010000m

Power Specifications
Par.

Min.

Typ.

Max.

5V

3.6V

5V

5.25V

Ripple

---

---

100mVp-p

Communication Protocol
QSZ-MN-III Module communicates with the host via a asynchronous serial port at
TTL levels, the data format is 4800baud, 8 data bits, odd parity, 1stop bit 4800
8-Odd-1.

Page 3 of 5

UN-III oximeter module users manual


Byte

Bit

Description

Byte

Bit

Description

Signal strength 0

Pulse rate 0

Signal strength 1

Pulse rate 1

Signal strength 2

Pulse rate 2

Signal strength 3

Pulse rate 3

Pulse rate 4

1 = Searching too long

1 = Sensor unplugged

Pulse rate 5

1 = Pulse beep

Pulse rate 6

1 (Syn. bit)

0 (Syn. bit)

Plethysmogram 0

SpO2 0

Plethysmogram 1

SpO2 1

Plethysmogram 2

SpO2 2

Plethysmogram 3

SpO2 3

4
4

Plethysmogram 4

SpO2 4

Plethysmogram 5

SpO2 5

Plethysmogram 6

SpO2 6

0 (Syn. bit)

0 (Syn. bit)

Bargraph 0

Bargraph 1

Bargraph 2

Bargraph 3

1 = Sensor error

2
5

1 = Searching for pulse

Pulse rate 7

0 (Syn. bit)

Page 4 of 5

UN-III oximeter module users manual

Connections
P1-Probe connector
P1 Pin
1
2
3
4
5
6

Description

Probe output
3.3V
RED+
REDGND
NC

P2-Host connector
P2 Pin
1
2
3
4

Description

TX Output
RX Input
GND
3.6V-5V

Size
Unitmm

Probe
Only digital probes can be used for this module.
Page 5 of 5