1. Turn
on
the
Instron
test
stand
(a
small
black
switch
facing
the
wall
on
the
lower
right
hand
side
of
the
machine)
2. After
the
Instron
has
warmed
up,
push
the
IEEE
button
on
the
control
panel
to
enable
it
3. Turn
on
and
log
on
to
the
desktop
computer
adjacent
to
the
test
stand
if
needed
4. Select
the
Instron
Series
IX
icon
on
the
desktop
5. If
a
video
error
pops
up
select
`ok
6. Select
the
Test
icon
7. Type
in
me218
for
the
data
file
and
push
`ok
8. Select
the
99
ME218Tensile
test
method
and
push
ok
9. Type
me218
for
the
ASCII
file
name
and
push
`ok
10. Open
the
bottom
vise
by
turning
the
hands
CW
NOTE:
DO
NOT
shear
the
small
$15
screw
in
the
vise.
If
you
do
you
owe
me
$15!
11. Obtain
a
sample
12. Measure
the
samples
diameter
and
record
it
13. Slide
the
test
sample
into
the
bottom
vise,
slide
the
sample
all
the
way
down
14. Lower
the
top
vise
until
in
nearly
touches
the
top
of
the
sample
15. Open
the
top
vise
by
turning
the
handles
CCW
NOTE:
DO
NOT
shear
the
small
screw
in
the
vise
16. Raise
the
sample
into
the
top
vise.
Make
sure
there
is
enough
surface
area
on
the
sample
in
each
vise
to
grip
properly
17. Clamp
the
bottom
vise
by
turning
the
handles
CCW.
Use
both
hands
to
tighten.
18. Put
the
extensometer
on
the
sample.
Make
sure
it
is
not
touching
the
top
or
bottom
vise
19. Push
the
IEEE
button
20. Reset
the
ground
level
by
pressing
GL
Reset
on
the
control
panel
21. Zero
the
load
sensor
by
pushing
BAL
in
the
yellow
Load
box
on
the
control
panel,
then
press
enter
on
the
keypad
22. Zero
the
strain
sensor
by
pushing
BAL
in
the
yellow
Strain
box
on
the
control
panel,
then
press
enter
on
the
keypad
23. Clamp
the
top
vise
by
turning
the
handles
CW,
use
both
hands
to
tighten
24. Push
the
IEEE
button
to
enable
communication
with
the
computer
25. Select
Start
Test
in
the
computer
program
26. Enter
the
diameter
of
the
sample
and
push
ok
twice
27. The
test
stand
will
begin
to
tension
the
sample
and
run
until
the
Yield
Point
is
reached
28. When
indicated,
remove
the
extensometer
29. Place
the
plastic
shield
in
front
of
the
test
piece
30. Select
ok
on
the
computer
31. The
test
stand
will
continue
to
tension
the
sample
until
it
breaks
32. After
the
test
is
finished,
push
ok
33. Observe
the
graph
and
record
the
following
on
the
worksheet:
a. Modulus
b. Yield
Strength
c. Ultimate
Strength
d. Ultimate
Strain
34. Remove
the
plastic
shield
35. Raise
the
top
vise
a
couple
inches
36. Unclamp
the
top
vise
by
turning
the
handles
CCW
and
remove
the
top
piece
NOTE:
DO
NOT
shear
the
$15
pin
37. Unclamp
the
bottom
vise
by
turning
the
handles
CW
and
remove
the
bottom
piece
NOTE:
DO
NOT
shear
the
$15
pin
38. To
continue
with
another
test
and
sample,
start
over
at
line
#11
39. To
stop
testing
exit
any
programs,
turn
off
the
test
stand,
and
log
off
of
the
computer