Publication No.

SM3360-16
2002DEC10
Supersedes SM3360-16
2001OCT15

SERVICE MANUAL
for the
Vitros 950 Chemistry System
Vitros 950AT Chemistry System
IMMUNO-WASH FLUID METERING ASSEMBLY

E015_4301HA

Ortho-Clinical Diagnostics, Inc.

PLEASE NOTE

The information contained herein is based on the experience and knowledge relating to the subject matter
gained by Ortho-Clinical Diagnostics, Inc. (OCD) prior to publication.
No patent license is granted by this information.
OCD reserves the right to change this information without notice, and makes no warranty, express or
implied, with respect to this information. OCD shall not be liable for any loss or damage, including
consequential or special damages, resulting from any use of this information, even if loss or damage is
caused by the negligence or other fault of OCD.

This equipment includes parts and assemblies sensitive to damage from electrostatic
discharge. Use caution to prevent damage during all service procedures.

Table of Contents
Description

Page

Adjustments. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
TIP to Metering Position - Alignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
TIP to RESERVOIR - Alignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3
3
5
7

2002DEC10 SM3360-16

Adjustments

Section 1: Adjustments
TIP to Metering Position - Alignment
Purpose:
Specification:
When to Do:

To align the MOUNTING PLATE of the IMMUNO-RATE WASH METERING


ASSEMBLY.
The TIP on the IMMUNO-RATE ARM must not touch the RESERVOIR COVER when
the TIP enters or leaves the RESERVOIR.
After replacement of the IMMUNO-RATE WASH METERING ASSEMBLY, PUMP
ASSEMBLY, or PUMP CARRIAGE.

To Adjust:
[1] Check that a TIP is on the IMMUNO-RATE ARM.
[2] Touch:
DIAGNOSTICS
MECHANISM EXERCISE DIAGNOSTICS
IMMUNO RATE SYSTEM
METERING ARM
CYCLE ARM
MULTIPLE
10
ENT
START
Detach 1

IMMUNO-WASH FLUID
METERING ASSEMBLY

IMMUNO-RATE
ARM

TIP

SCREW (3)
MOUNTING PLATE

E015_4303BCA
E015_4303BA

[3] Observe the TIP as it enters the Metering hole. Does the TIP enter the hole without touching the side?

Yes
Advance to step 9.

2002DEC10 SM3360-16

No
Advance to step 4.

SERVICE MANUAL

[4] Loosen the 3 SCREWS on the MOUNTING PLATE.


IMMUNO-RATE ARM

RESERVOIR

TIP

O4308DCA

[5] Touch START and move the IMMUNO-RATE WASH METERING ASSEMBLY so the TIP does not touch
the side of the hole.
[6] Tighten the 3 SCREWS evenly.
[7] Touch START and check that the adjustment is correct.
[8] Do:
Adjustment
TIP to RESERVOIR - Alignment
[9] Return the System to normal operation.

2002DEC10 SM3360-16

Adjustments

TIP to RESERVOIR - Alignment


Purpose:

To align the RESERVOIR of the IMMUNO-RATE WASH METERING


ASSEMBLY.
The TIP on the IMMUNO-RATE ARM must be in the center or
1.0 mm (< 0.040 in.) from the center of the hole in the RESERVOIR COVER.
After replacement of the IMMUNO-RATE WASH METERING ASSEMBLY,
PUMP ASSEMBLY, or PUMP CARRIAGE.
None.
None.

Specification:
When To Do:
Special Tools:
Special Requirements:
To Adjust:

[1] Check that a TIP is on the IMMUNO-RATE ARM.


[2] Touch:
DIAGNOSTICS
MECHANISM EXERCISE DIAGNOSTICS
IMMUNO RATE SYSTEM
METERING ARM
ARM TO RESERVOIR POSITION
[3] Observe the TIP as it enters the hole in the RESERVOIR COVER. Does the TIP enter the hole without touching
the side?

Yes

No

Advance to step 8.

Advance to step 4.

Detach 2

IMMUNO-RATE ARM

TIP

RESERVOIR
NEST
SCREW (3)

E015_4302BCA
E015_4302BA

[4] Loosen the 3 SCREWS on the RESERVOIR NEST.


[5] Touch START and move the RESERVOIR NEST so the TIP does not touch the side of the hole.
[6] Evenly tighten the 3 SCREWS.
[7] Touch START and check that the adjustment is correct.
[8] Return the System to normal operation.
2002DEC10 SM3360-16

SERVICE MANUAL

2002DEC10 SM3360-16

Parts List

Section 2: Parts List


List of Figures
Description

Page

IMMUNO-WASH FLUID METERING ASSEMBLY . . . . . . . . . . . . . . . . . . . . . . . . . . .

RESERVOIR ASSEMBLY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

10

2002DEC10 SM3360-16

SERVICE MANUAL

Figure 1

IMMUNO-WASH FLUID METERING ASSEMBLY


14
13
12
11
15

16
19

10

17

18

5
1
4

3
O4307ECA

2002DEC10 SM3360-16

Parts List

Figure 1
Item
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19

Part No.
2B3103
999347
273442
903756
748353
999611
356192
354984
354626
356933
1C3350
356936
104459
748360
1C2191
1H0126
2B3854
337918
1C2169

IMMUNO-WASH FLUID METERING ASSEMBLY

Description
Sensor - Gap
Clamp - Sensor
Screw
Washer
Screw - Mounting
Motor and Pinion Assembly
Shaft - Guide
Bearing - Cam
Cam - Lift
Cam - Washer
Spring
Bushing
Washer
Screw
Sensor
Pump Assembly
Flag - Sensor
Screw
Spring

2002DEC10 SM3360-16

Qty.
1
1
1
1
2
1
1
1
1
1
1
1
1
1
1
1
1
1
1

Notes

SERVICE MANUAL

Figure 2

RESERVOIR ASSEMBLY
4

7
6
8

2
9

1
E012_5025DCA
E012_5025DA

Figure 2
Item
1
2
3
4
5
6
7
8
9

10

Part No.
356496
356668
356582
356531
356497
356864
748360
066556
356669

RESERVOIR ASSEMBLY

Description
Housing - Reservoir
Spring
Pin
Holder - Cover
Cover - Reservoir
Seal - Reservoir
Screw - SEMS, M4 x 12
Washer
Reservoir Assembly

Qty.
1
1
1
1
1
1
3
3
1

Notes

2002DEC10 SM3360-16

Parts List

This page is intentionally blank.

2002DEC10 SM3360-16

11

Vitros is a trademark.

Sm336016.FM+
Printed In USA

Ortho-Clinical Diagnostics, Inc.

100 Indigo Creek Drive

Rochester, N.Y. 14626