Instrument or Artifact
Instrument Type or Method
Measurand Level or Range
Number of measurements
Time interval
Stopwatch
Time interval measurement
0.01 to 1E5 s
10
Instruments
Measurement configuration
10 MHz Reference
Universal Counter
start
Frequency Distribution
Amplifier
stop
Stopwatch Calibrator
start
Cesium Clock
Stopwatch
List of uncertainties
1. UTC uncertainty
2. UTC-UTC(NMI) Link Uncertainty
3. UTC(NMI) Uncertainty
4. Cesium Uncertainty
5. Cable Uncertainty
6. Frequency Distribution Amplifier Uncertainty
1/8
stop
GPIB
2) This uncertainty is multiply by 2 , considering that the uncertainty of TAI link both start and
end points of the measurement. In this case the uncertainty is calculated as follows:
uA
5 2
86400 5
2 10 9
5 2
86400 5
1.5 10 14
ulink
3) From the data given in Circular T, as shown in the Appendix as Figure 1, the calculated standard
deviation and average of UTC-UTC(MNFS) for MJD55214 to 55439 are 516.6E-15 and
-23.9E-15, respectively. The uncertainty is calculated as follows:
u predict
(516.6 10 15 ) 2 5 ( 23.9 10 15 ) 2
1.2 10 12
4V
4V
47.68s
8.4 s
4.77 10 7 8.40 10 8
5.61 10 7 s
2/8
6.05 10 7 s
2
5.96 10 -7 s
2
t res 750 ps
Note : The signal slew rate of the calibrator output pulse were measured using a digital oscilloscope.
Type
Value
s
Distribution
Factor
Sensitivity
Coefficient
Standard
Uncertainty s
5.2E-15
5.2E-10
3/8
1.5E-14
x
100000
1.5E-9
3) UTC(NMI) Uncertainty
1.2E-12
x
100000
1.2E-7
4) Cesium Uncertainty
5E-13
x
100000
5E-8
5) Cable Uncertainty
(Worse Case : 7 0C
temperature change over
100000 s)
4E-15
x
100000
4E-10
6) Frequency Distribution
Amplifier
( 2 0C temperature
change over 100000 s)
4E-15
x
100000
4E-10
7) Universal Counter
Uncertainty
1.78E-6
1E-6
8) Stopwatch Calibrator
Uncertainty
2E-7
2E-7
9) Rresolution of
Stopwatch
5E-4
2.9E-4
1
3
4/8
APPENDIX
Cable Uncertainty
This is an economical solution on creating a significant temperature changes, we used a
25 watt lighting bulb and a timer to ON and OFF the bulb. The bulb was place inside the
polystyrene box with the cable under test (CUT). A thermocouple was attached to the
cables skin and the data acquisition records its temperature at 1 s interval. The
measurement set-up diagram, temperature records and measurement results
(Overlapping Allan deviation) are shown as in Figure 1, 2 and 3:
10 MHz from Caesium
Frequency Standard
External Reference In
Agilent 53131A
Records phase
comparison
measurement at
1 s gate time
Lighting Bulb
Polystyrene Box
Figure 1 : Measurement set-up for estimating the uncertainty of BNC cable in phase
comparison in the worst case event of 7C change over 1 day.
5/8
Figure 4: By using extrapolation of slope -1, we found that the uncertainty of cable in the worst
case event of 7C change over 100000 s is 4 x 10-15.
Frequency Distribution Amplifier Uncertainty
Figure 2 : Temperature measurement at 1 second interval
Figure 3: By using extrapolation of slope -1, we found that the uncertainty of cable in
worst case event of 7C change over 100000 s is 4 x 10-15.
6/8
10 MHz and
5 MHz
Agilent 53131A
Records phase
comparison
measurement at 1 s
gate time
7/8
History
2012-06-01 first version (created by SIRIM).