You are on page 1of 9

HCT.CO.,LTD 105-1, Jangamri Majangmyeon Ichensi Gyeonggido, 467-811.

Republic of Korea
T. 82-31-6456365 Email. product@hctpd.com www.hct.co.kr
HCT - Technology Changing How We Live www.particle-science.com

PC-3106P (Optical Particle Counter)

PC-3106P is an instrument which can measure the particles suspended in the air above 0.3m in diameter. Its sampling flow rate is 1 CFM
( 28.3L/min ). It provides several analog input (4~20mA) terminals for external sensor or probe air velocity, temperature & humidity, differential
pressure, etc. PC-3106P complies with the requirements of ISO 21501-4 or JIS B9921 and is calibrated using a condensation particle counter with
NIST traceable PSL. PC-3106P provides measurement result reports for ISO 14644-1 and FS209E via the onboard thermal printer. Measurement data
can be stored in the internal memory up to 30,000 records.

PC-3106P can either be used as a stand-alone particle counter or as a part of in the Facility Monitoring System (FMS).

Feature and Benefits

Up to 6 channels of simultaneous particle count data


Data storage 30,000 samples records
4 to 20 mA input (T/H, Velocity, Diff. Pressure)
Touch screen operation
Reports for ISO14644-1, FS209E

Applications

Cleanroom Monitoring
Facility Monitoring
Hospital Surgical Suites
Pharmaceutical
Electronics (Semiconductor, Flat Panel, LED,Hard Disk) Monitoring

SPECIFICATIONS

Particle channel sizes 0.3, 0.5, 1.0, 5.0, 10, 25 or 0.3, 0.5, 1.0, 3.0, 5.0, 10 , 6ch

Counting efficiency 50% at 0.3 , 100% for particles > 0.45 (Meets ISO and JIS Standards)

Concentration limit < 1,000,000 / ft3 at 5% coincidence loss

Light source Laser diode

Zero count < 1 count per 5 minutes (per ISO21501-4 and JIS B9921)

Flowrate 28.3 L/min (1.0 CFM) 5% accuracy (meets ISO 21501-4 and JIS B9921)

Sampling time 1 sec ~ 1 hrs

Data storage 30,000 samples records (internal memory)

Communication Ethernet, USB Serial, RS 485

Printer Thermal Printer

Vacuum source Internal pump

Battery Li Ion battery

Battery runtime 3 hours (2 Battery)

Dimension (LWH) 30 x 24 x 23 cm(11.8 x 9.45 x 9.05 in)

Weight 5.2 kg(11.46 lbs)

Power 24V / 4.0 Adapter 100 ~240VAC, 50/60 Hz

Parts Number 6C06008 Power Cable


6C06007 USB Download cable, type A to B & LAN cable
6P02016 Li-Ion battery
7004022 Purge Filter Assy
6P02030 Adaptor
8007141 Software program CD
9005011 Printer paper
1401017 Isokinetic probe & Cap

HCT Instruments INC. 1229 38th Ave N MS 161, Myrtle Beach, So Carolina 29577 Tel: 970-481-5721, Email. info@hctinstruments.com www.hctinstruments.com
HCT.CO.,LTD 105-1, Jangamri Majangmyeon Ichensi Gyeonggido, 467-811. Republic of Korea
T. 82-31-6456365 Email. product@hctpd.com www.hct.co.kr
HCT - Technology Changing How We Live www.particle-science.com

HPS-0405W (Hybrid Particle Sensor)


HPS is a new product that combines the functions of the existing Optical Particle Counter and Virtual Impactor all into one package. It can count the
number of particles per particle size and also sample the particles at the same time.

Feature and Benefits


Real Time Insitu Particle Counting and Sizing With
Four Channels of Simultaneous Particle Count Data
Particle Capture For Off Line Analysis
Ease to install
Easily installed in the Process Equipment Eliminating Transport Tubing Losses
Ag Sample Collecting Plate for Off line Particle analysis Sensor inside
Meets or Exceeds current JIS (JIS : Japanese Industrial Standards)

Applications

Clean Room and Facility Monitoring


Particle Trend Analysis
Indoor Air Quality (IAQ) Analysis
Real Time InSitu Particle Monitoring in the Process Equipment.

SPECIFICATION

Size range 0.5 ~ 10.0


Particle channel sizes 0.5, 1.0, 2.0, 5.0 (4ch) or 0.5, 1.0, 5.0, 10.0 (4ch)
Counting efficiency 50 % at 0.5 , 100 % > 1.0
Flow rate 28.3 L/min (1.0 CFM) ) 5% accuracy
Max. concentration 1,000,000 particles/CFM @ 0.5 um * System components excluded : Stand
Sampling time 1 sec ~ 60 sec
Light source Laser Diode (35 mW)
Zero count level < 1/sec @ 5 min.
Vacuum requirement External Vacuum Pump ( < 160 mmHg )
90 Kpa 10% (Vacuum)
Communication Ethernet
Enclosure Steel
Surface of sampling plate Ag coated
Dimension (LWH) 7 x 14 12.5 cm (2.8 x 5.5 x 5 in)
Weight 0.89 kg (1.96 lbs)
Power (main) 100 ~ 230 VAC, 50 ~ 65 HZ
Parts Number
(Adapter) 5V, 4A
1001027 Main inlet
1003004 Inlet tube connector
1003003 Stand (for bottom)
1003002 Bracket (for wall)
Before Sampling

HPS Monitoring Software


300

1.0 mm
2.0 mm
250
5.0 mm
10.0 mm
Particle Counts, #.sec

200

After Sampling
150

100

50

0
0 100 200 300 400 500 600 700 800 900 1000

Time, sec

HCT Instruments INC. PO Box 3532 Myrtle Beach, SC 29578 T. 843-808-9224, Email. info@hctinstruments.com www.hctinstruments.com
HCT.CO.,LTD 105-1, Jangamri Majangmyeon Ichensi Gyeonggido, 467-811. Republic of Korea
T. 82-31-6456365 Email. product@hctpd.com www.hct.co.kr
HCT - Technology Changing How We Live www.particle-science.com

AG-01 (Aerosol Atomizer) CAS-01 (Clean Air Supply)

Aerosol generation is very important and necessary in the field of Clean Air Supply, CAS-01 can provide clean and dry compressed air for
aerosol science. aerosol experiments.
AG-01 is a nebulizer type aerosol generator. Aerosol jets generated by The Water & Oil trap removes large water & oil drops included in the
compressed air throughout the critical orifice collide with the wall inside compressed air. Compressed air is then dried in the desiccant dryer,
AG-01. Small droplets (including particle droplets) divided by impaction then exhausted throughout the HEPA filter.
is exhausted outside AG-01. .

Application: SPECIFICATIONS

Filter Media Test Maximum pressure : 10 kg/cm3


Particle generation in wind tunnel Flow rate : 50 LPM (Max.)
Particle deposition in wafer deposition system Dimension (LWH): 450 x 240 x 400 mm (17.7 x 9.45 x 15.7 in)
Weight : 10 kg (22.05 lb)
SPECIFICATIONS
Part Number
Air Flow Rate : 4.5 L/min at 2.5 kg/cm3 2501004 Oil trap
Particle Concentration : < 106 particles/cm3 7004004 HEPA Filter
Geometric Standard Deviation : 1.3 ~2.0
Pressure Source : Air compressor
Number of Jets : 1
Dimensions (height) : 130 mm (5.12 in)
Weight : 0.5 kg (1.1 lbs)

Part Number
2601003 Orifice 0.3 mm
2601004 Orifice 0.5 mm
2601001 Lid (tempered glass)
2601006 Lid (Quartz)

XRC-05 (Soft X-Ray Charger)


XRC-05 is a bipolar charger that uses soft X-ray instead of radioactive
source such as Am241 (Americium-241), Kr85 (Krypton-85), Po210
(Polonium-210). It changes irregularly charged particles to equilibrium
DD-01 (Diffusion Dryer)
charged state. Particles introduced into XRC-05 is collided with ions (+)
or electrons (-) generated by soft X-ray radiation. Ions or electrons DD-01 removes water vapors contained in the aerosol and dries-off
being attached onto the particle surfaces, neutralizing the particles. water on the surface of the droplet.
XRC-05 does not generate Ozone due to its air ionization effect. The Water & Oil trap mounted at the inlet of DD-01 removes large
water & oil drops. Aerosols are dried by the diffusion of water vapor
in the desiccant dryer filled with molecular sieve or silica gel.

SPECIFICATIONS

Ion Generation Method: Soft X-ray SPECIFICATIONS


Source: Soft X-ray tube
Power Consumptions: 7.2W Flow rate : 0 ~ 5 L/min
Cooling Method: Natural cooling Maximum pressure : 420kPa
Weight: 1.11 kg /2.45 (lb) Relative humidity ; 20 % when incoming R.H. is 60 %
Input Power: AC 100~240V, Dimension (LWH): 570 x 160 x 210 mm ( 224.4 x 6.3 x 82.7 in)
50/60Hz (Adaptor Input) DC 12V, 3.33A
Life Time: 1 year Part Number
Indicating: Ion generation indicating 2401001 Lid
Dimension (LWH):273 X 72 X191 mm (10.8 x 2.8 x 7.5 in) 2401002 Stand
2401003 Water trap
Part Number 2401004 Outer Cylinder
2302002 DC 12V Adapter & power cord 2401005 Inner Diffusion Screen
2302004 Safety Key 9005009 Silica gel (Optional)

HCT Instruments INC. 1229 38th Ave N MS 161, Myrtle Beach, So Carolina 29577 Tel: 970-481-5721, Email. info@hctinstruments.com www.hctinstruments.com
HCT.CO.,LTD 105-1, Jangamri Majangmyeon Ichensi Gyeonggido, 467-811. Republic of Korea
T. 82-31-6456365 Email. product@hctpd.com www.hct.co.kr
HCT - Technology Changing How We Live www.particle-science.com

AS-03 (Air-Sampler) VI-104 (4-Stage Impactor)

Due to strict environmental rules, it is highly necessary to sample According to environmental regulations, it is needed to analyze
aerosols or gas-phase materials. Especially, the product protection airborne particles to identify their components or composition. Such
against fine particles or gas-phase materials is essential in particles should be collected to be analyzed. VI-104 is the instrument to
semiconductor, LCD or any precision manufacturing fields. The Model be used to collect them.
AS-03 is the instrument to sample these gas-phase materials.
Application:
Portable sampler
Maximum of 12 hours (Using a built-in battery for operating with 2 Aerosol Research : Particle sampling, collection, generation and
LPM) transport
Particle analysis in clean-room and associated controlled environments
User-friendliness Indoor and outdoor air quality research
The touch screen enables a convenient control. Bio-aerosol research
The sampling start or shutdown can be scheduled using the built-in
timer.
Sampling can be done by setting either the sampling time or the
total flow rate.

Optimization for clean-room environments


A built-in HEPA filter is used to prevent contamination by a pump,
optimal for sampling in a clean-room

Accuracy of flow rates


The differential pressure sensor assures a higher accuracy for the air SPECIFICATIONS
sampler.
Guarantee of the accuracy of flow rates (5% full scale) throughout Sampling flow rate : 20 L/min
flow rate correction. Particle Cut-off Diameter: 1.0, 2.5, 5.0 and 10 m
Front panel display : Touch screen LCD Panel
(Set-up reserved time / operation time / Display present time)
Application: Dimension (LWH):
Controller : 320 x 235 x 240 mm (12.6 x 9.3 x 9.5 in)
Sampling in hospital or bioengineering research center Impactor : 78 x 265 H mm (78 x 10.4 in)
Sampling in clean-room Weight : 6.5 kg (14.3lb)
Power : 100 ~ 240 VAC, 50/60 Hz
Sampling in factories or industrial field
Part Number
6C06001 Power Cable
7004004 HEPA Filter

DMA-20/05 (Differential Mobility Analyzer)


SPECIFICATIONS

Particle Size Range :


- DMA-05 : 2 ~ 160 nm
- DMA-20 : 7 ~ 830 nm
Maximum input concentration : 107 particles/cm3 @ 10 nm
Aerosol pressure range : 1 0.1 ATM
Flow rate :
- Aerosol : 0.1~1.5 L/min
- Sheath Air : 1 ~ 15 L./min
- By pass : None
SPECIFICATIONS Voltage : 10 ~ 10,000 VDC
Dimensions : DMA-05 : 210 x 36 mm (8.3 x 1.4 in)
Sampling flow rate : 1~3L/min DMA-20 : 420 x 76 mm (16.5 x 3.0 in)
Flow rate accuracy : 5 % of full scale Weight : DMA-05 : 2 kg (4.41 lb) / DMA-20 : 7 kg (15.43lb)
Gas temperature range : 10~50C
Gas pressure range : 1 0.2 ATM
Operating hours : 12 hrs @ 2 L/min Application:
(in full charging battery)
Battery charging time : 4~5 hrs Aerosol research : particle generation,
LCD display : Graphic color LCD (400 x 240 pixel) nucleation, condensation, coagulation and
Dimensions (LWH) : 160 x 190 x 190 mm (7.5 x 6.3 x 7.5 in) transport
Impinger : 240 mm x 45 (9.5 lb x 45) Particle charging and electrical mobility
Weight : 2.6 kg (5.7 lb) studies
Power : 100~240 VAC, 50/60 Hz Filter media test
Mono-disperse particle generation for
Part Number instrument calibration
6C06001 Power Cable
1901001 Impinger

HCT Instruments INC. 1229 38th Ave N MS 161, Myrtle Beach, So Carolina 29577 Tel: 970-481-5721, Email. info@hctinstruments.com www.hctinstruments.com
HCT.CO.,LTD 105-1, Jangamri Majangmyeon Ichensi Gyeonggido, 467-811. Republic of Korea
T. 82-31-6456365 Email. product@hctpd.com www.hct.co.kr
HCT - Technology Changing How We Live www.particle-science.com

WCPC-1003S(Water-based Condensation Particle Counter)


A new approached is applied using particle condensation technique. When a particle is place in a super-saturated vapor, the super-saturated vapor
molecules will start condense onto the particle. The particle with condensed vapor will continuously grow and become a large droplet. This large size
droplet can easily be detected by optical electronic system.

Benefits and Applications

Single Particle Counting with real-time for accurate measurement (0 ~ 10,000 particles/ cm3)
With water as the working fluid and HEPA-filtering of the exhaust use the clean room environment
With High Pressure Diffuser(HPD-75), Possible to Monitoring in high pressure gas lines

Feature

10 nm Detection Limit
2.83 L/min of aerosol Flow rate
Built-in Color Touch Screen Display (7)
Adopted in Internal Thermal Printer
Contaminant particle detection in cleanroom and high purity gas
systems
Ethernet, USB Serial Communication

SPECIFICATIONS

Particle Size Range : Min. Detectable Particle(D50): 10 nm, verified with DMA-classified
Particle Concentration Range : Single Particle Counting: 0 to 10,000 particles/
Particle Concentration and Accuracy : 10 % at 10,000 particles/
Flow rate : 2.83 0.28 L/min
Response Time : ~ 0.8 sec., (95% Response to Concentration Step Change)
False Background Counts : ~5x10-6 /, 12 hrs average
Aerosol Medium : Inert Gas for Air, N2, Ar
Environmental Operating Conditions : Ambient Temp. & Humidity: 10 to 35
and 0 to 90% RH, Non condensing
The efficiency curve
Inlet Pressure : 75 to 110 kPa
Working Fluid : Distilled Water only 100

Water Supply System : External 1 Liter Bottle used (5 days of continuous operation) 90

Vacuum : Internal Vacuum Pump 80

Outputs : USB Serial, Ethernet, SD Card, Internal Thermal Printer 70


Efficiency , %

Digital Display: 7 Color touch screen 60

Power : Adapter for 24 VDC, 6.0A 50

Communication: USB 40

Dimensions (LWH): 310 x 200 x 270 mm (12.2 x 7.9 x 10.6 in) 30

Weight: 9.4 kg (20.7 lb) 20

Parts Number 10

6C06001 Power Cable 0


7004014 CPC Extra Filter 1 10 100
7008001 Drain Bottle / Fill Bottle Particle diameter , nm
9005011 Printer paper
6C06007 USB Download cable, type A to B & LAN cable
9003005 Sampling Tube (1 m)
2602022 Monitoring software Program

HCT Instruments INC. 1229 38th Ave N MS 161, Myrtle Beach, So Carolina 29577 Tel: 970-481-5721, Email. info@hctinstruments.com www.hctinstruments.com
HCT.CO.,LTD 105-1, Jangamri Majangmyeon Ichensi Gyeonggido, 467-811. Republic of Korea
T. 82-31-6456365 Email. product@hctpd.com www.hct.co.kr
HCT - Technology Changing How We Live www.particle-science.com

DI-120 (Aerosol Dilutor)

HCT Model DI-120 reduces particle concentrations of high-concentration aerosols.

Feature & Benefits

Highly repeatable performance


Dual pressure gauges (indicate correct operation or signal need for maintenance)
Require NO power or NO compressed gas.
Simple maintenance
Trouble-free cleaning.

DI-120 test using the 0.3mm PSL

140

1.2 LPM
SPECIFICATIONS 120
1.5 LPM

100
Sample flow : 1 LPM (1.2 / 1.5 optional) Dilution ratio

Dilution ratio : 1:100 80

Dimensions (LWH) : 85 x 420 x 107 mm ( 3.35 x 16.54 x 4.21 in)


60
Weight : 9.82 kg ( 21.65 lbs)
40

20

0
0 5 10 15 20 25 30

Clean Air Gauge (mmH2O)

HPD-75 (High Pressure Diffuser)


Connecting pressurized air or gas to a particle counter is able to damage the instrument. HPD-75 can reduce the pressure of these gases and allow
the particle counters flow metering system to function as designed.

No external flow meters needed

SPECIFICATIONS

Air Flow Rate: 0.1 CFM


Material: Stainless steel body with orifice, exhaust filter
Pressure: 35 ~ 75 psi for 0.1 CFM
Inlet fitting: VCR Male Connector
Exhaust fitting: tube type
Sample gas: Dry, inert, non-toxic, non-flammable gases
(CDA, Nitrogen, Argon)
Dimensions (HWD): 182 x 155 x 77 mm (7.17 x 6.10 x 3.03 in)
Weight : 782 g (1.72 lbs)
Included : Mounting brackets and exhaust tube (1 m)

HCT Instruments INC. PO Box 3532 Myrtle Beach, SC 29578 T. 843-808-9224, Email. info@hctinstruments.com www.hctinstruments.com
HCT.CO.,LTD 105-1, Jangamri Majangmyeon Ichensi Gyeonggido, 467-811. Republic of Korea
T. 82-31-6456365 Email. product@hctpd.com www.hct.co.kr
HCT - Technology Changing How We Live www.particle-science.com

EPS-n (Electrical Particle Sizer with XRC-05)


EPS-n is an instrument which classifies a particle according to its electrical mobility. Electrical mobility of a particle is decided by fluid resistance for
ce and electrostatic force and can be converted to an equivalent particle diameter. The biggest advantage of EPS-n is that Soft X ray is used as the
neutralizer instead of radioactive sources. The Soft X ray charger is installed inside the EPS-n making it smaller and more portable than its previous
models.

The EPS-n is a rudimentary instrument that generates mono-disperse aerosols and measures particle size distributions with high durability and repe
atability.

HCT provides two type of DMA, depending on your preferred particle size range.

Benefits and Applications

Aerosol Generation System


Fundamentals of aerosol research
Monitoring outdoor and indoor air quality
Filter Media Test for filter efficiency
Particle formation or growth research
Use for standard calibration tool of particle counter

SNPS
Fundamentals of aerosol research
Atmospheric environment research
Research on semiconductor fabrication and material synthesis

SPECIFICATIONS

Maximum particle concentration : 107 particles/


Particle size range
- Using DMA-20 : 7~ 830 nm
- Using DMA-05 : 2 ~ 160 nm
Flow rate :
- Aerosol : 0.1 ~ 1.5 L/min
- Sheath air : 1 ~ 15 L./min
Aerosol temperature range : 10 ~ 35C
TDMA Test
Aerosol pressure range : 1 0.1 ATM
Front panel display : 800 x 480 pixel
60000
Power : 100 ~ 240 VAC, 50/60 Hz, 1.0A Dp = 20.28 s = 1.07
Dimensions (LWH): 260x375x350 mm ( 14.8 x 10.2 x 17.8 in) 50000
Dp = 29.98 s = 1.06
Dp = 50.92 s = 1.06
Dp = 80.79 s = 1.06
Parts Number 40000 Dp = 101.33 s = 1.06
dType/dLogDp

6C06001 Power Cable


6C05061 High voltage Cable 30000

6C05063 BNC Cable


1801013 Impactor 20000

1804000 DMA-05 (Optional)


1801000 DMA-20 (Optional) 10000

0
10 100

Particle diameter(nm)

HCT Instruments INC. 1229 38th Ave N MS 161, Myrtle Beach, So Carolina 29577 Tel: 970-481-5721, Email. info@hctinstruments.com www.hctinstruments.com
HCT.CO.,LTD 105-1, Jangamri Majangmyeon Ichensi Gyeonggido, 467-811. Republic of Korea
T. 82-31-6456365 Email. product@hctpd.com www.hct.co.kr
HCT - Technology Changing How We Live www.particle-science.com

SNPS-n20 (Scanning Nano Particle Spectrometer with XRC-05)

SNPS-n20 is an instrument which can measure aerosol particles in the range from 7 to 830 nm in high resolution. SNPS-n20 consists of CPC and
EPS, and these two devices are connected through a tube with each other. Particles classified corresponding to its electrical mobility in the EPS are
introduced into the CPC, and then the particles are measured by the optical system in the CPC after condensational growth. Measured data is
transferred in real time to a computer using the Scanning Mobility Particle Spectrometer software. Raw data is converted to particle size distribution
through the inversion algorithm.
SNPS-n20 is suitable to measure airborne particle size distributions, so SNPS is highly recommended monitoring atmospheric aerosol in the
Meteorology. SNPS can be also used for measuring the size distribution of test aerosols in Inhalation toxicology test system or filter media test
system.

Applications Benefits
Indoor/outdoor air quality research Real-time nanoparticle size measurements : 75 sec (scan + retrace)
Occupational hygiene/workplace exposure monitoring High resolution : 64 channels per decade
Filter Media Test for filter efficiency Nano particle size up to submicron particle size : 7 nm ~830 nm
Inhalation toxicology Meet ISO 15900 : 2009
Fundamentals of aerosol research No radioactive materials
Nanotechnology applications Built-in data logging
Atmospheric environment research Size distributions down to 10 nanometers

SPECIFICATIONS

Maximum particle concentration : 107 particles/


Maximum particle size resolution : 64 channel
Particle size range
- Using DMA-20 : 7~ 830 nm
- Using DMA-05 : 2 ~ 160 nm
Flowrate :
- Aerosol : 0.1 ~ 1.5 L/min
- Sheath air : 1 ~ 15 L./min
CPC Type : Water based CPC or N-Butyl CPC
Aerosol temperature range : 10 ~ 35C
Aerosol pressure range : 1 0.1 ATM
Communication : RS-232
Power : 100 ~ 240 VAC, 50/60 Hz, 1.0A

Parts Number
6C06001 Power Cable
6C05061 High voltage Cable
6C05063 BNC Cable
1801013 Impactor
1804000 DMA-05 (Optional)
1801000 DMA-20 (Optional)
7004014 CPC Extra Filter
7008001 Drain Bottle / Fill Bottle
1702002 HCT SNPS Software program

HCT Instruments INC. 1229 38th Ave N MS 161, Myrtle Beach, So Carolina 29577 Tel: 970-481-5721, Email. info@hctinstruments.com www.hctinstruments.com
HCT.CO.,LTD 105-1, Jangamri Majangmyeon Ichensi Gyeonggido, 467-811. Republic of Korea
T. 82-31-6456365 Email. product@hctpd.com www.hct.co.kr
HCT - Technology Changing How We Live www.particle-science.com

HCT-3710 (Mask Tester)


HCT has developed a mask tester ideal for respirator testing and perfect for complying with standards such as NIOSH and EN 143.
The multi-jet atomizer in the HCT-3710 is capable of generating a large amount of aerosol(120mg or 200mg) that can be easily adjusted by changing the
nozzle part having a different number of nozzles.

Feature & Benefits

The dehumidifier in the HCT-3710 maintains a less than 60% relative humidity for the
It can test various type of masks
generated aerosol.
by changing mask adaptor.
HCT-3710 also features the HCT Soft X-ray which acts as a neutralizer that does not
need special authorization to be used. It offers high performance efficiency for respirator
tests.
The flow rate for the test can be automatically adjusted.
The automatic flow control can maintain the set flow rate for the duration of the test
period providing a highly reliable mask performance review.
An anti-vibration and soundproof chamber is provided with the HCT-3710 to be used
during high flow pump operations.
The user can monitor the process in real time. Its dedicated software has a functional * Mask adaptor must be ordered separately. When you order
graph tool suitable for data analysis and was designed to process large amounts of data. this product, we provide a basic type of mask adaptor.
The test parameters can be adjusted by varying the level of Relative
Humidity and amount of aerosol challenge.

HCT-3710 has been enhanced to insure your final products meet global standards.
Complied with En 143, NIOSH 42 CFR part 84.
Generate a large amount of aerosol.
Easy to use smart software
No radioactive source
Low maintenance

Display Chart & Data Display Environment data

Features Specifications
Main Display at Window
Particle Generation Particle Generator Atomizer
Particle measurement system Photometer
Particle detection Sampling Flow rate Bypass & 1.5 L/min
Sampling particle Dynamic range 0 ~ 400 g/s
Aerosol type : NaCl, DOP, Paraffin
DOP Diameter (count median) 0.18 m, 0.2 m
DOP Mass 17 mg/min Display : Pressure, Temperature, Relative humidity, Flow rate
DOP Standard deviation <1.7 Chamber On/Off
Paraffin diameter (count median) 0.18 m, 0.2 m Control start and stop
Particle type Paraffin mass 17 mg/min
Paraffin Standard deviation <1.7
NaCl diameter (count median) 0.085 m, 10 nm
NaCl Mass 3.6 mg/min
NaCl Standard deviation <1.86
Detection rate Efficiency rate 30 ~ 99.999 %
Neutralizer Soft X-ray charger
Flow rate 0 ~ 120 L/min
Differential pressure 0 ~ 620 Pa
Temperature 10 ~ 50
Environment data Humidity 20 ~ 60 %
Communication USB to Serial
Power 100 ~ 240 VAC,
Dimension 1400 * 800 * 1080 (mm)
Weight 300 kg

HCT Instruments INC. PO Box 3532 Myrtle Beach, SC 29578 T. 843-808-9224, Email. info@hctinstruments.com www.hctinstruments.com