Professional Documents
Culture Documents
Abstract-This paper reports on an investigation of feedback ity, FNS has been used in the laboratory for standing,
control of coronal plane posture in paraplegic subjects who
stand using functional neuromuscular stimulation (FNS). A walking and stair climbing. Although feasibility has been
feedback control system directed at regulating coronal plane demonstrated [20], [22], [24], [25], [32], current lower
hip angle in neutral position was designed, implemented, and extremity FNS technology is deficient in ways which have
evaluated in two paraplegic subjects. The control system in- delayed and limited its clinical use. In general, FNS sys-
cluded sensor mounting and signal processing techniques, a tems need to be easier to use, more reliable, and safer;
two-stage feedback controller, stimulation hardware, and a set they also must provide a greater level of function to the
of percutaneous intramuscular electrodes. The feedback con-
troller consisted of two-stages in cascade: a modified discrete- user. The use of feedback control in lower extremity FNS
time proportional-integral-derivative (PID) stage and a nonlin- systems has been proposed as a means to overcome these
ear single-input, multiple-output stage to determine the stim- limitations [9], [30].
ulation to be sent to several muscles. The focus of this work Various stimulation controllers have been developed
was on evaluating the performance of the feedback controller and tested either in computer simulation, in animal ex-
by comparing the response of the feedback-eontrolled system
to that of an open-loop stimulation system. In an evaluation periments, or in human experiments. These control sys-
based on temporal response characteristics the controlled sys- tems modulate stimulus parameters in order to regulate
tem exhibited a 41% reduction in root-meamsquared (rms) er- one or more system outputs (e.g., joint angle or angular
ror (where error is defined as the deviation from the desired velocity, joint torque, contact force). Control systems for
angle), a 52% reduction in steady-state error, and a 22% re- FNS act on sets of nonlinear actuators. Controllers con-
duction in hip compliance. In addition, the feedback-controlled
system exhibited significant reductions in variability of these sisting of two stages in cascade have often been used: one
measures on several days. These results demonstrate the ability stage to control the system dynamics and a second to ef-
of the feedback controller to improve the temporal response fect co-stimulation and compensate for nonlinearities [26].
characteristics of the FNS control system. Both fixed-parameter feedback controllers [lo], [29],
[30], [33] and adaptive controllers [3], [6] have been in-
INTRODUCTION
AND BACKGROUND
vestigated. Proportional [30], PID [ 131, pole-placement
[ 101, and adaptive [6] controllers have been used for the
I N recent years, researchers have investigated the use of
functional neuromuscular stimulation (FNS) as a tool
in the rehabilitation of patients with spinal cord injury,
first stage.
Discrete-time proportional-integral-derivative (PID)
controllers have been utilized during stance at the knee
stroke, and other neurological disorders. This technique [5], [12] and at the knee and ankle [ l l ] . These earlier
employs electrical pulses to activate paralyzed muscle for attempts demonstrated disturbance resistance and the abil-
the purpose of achieving functional movements. It has ity to stabilize a particular joint, but did not directly result
been used to restore or enhance lower and upper extremity in improved stance for the user. Upper body movements
functional movements as well as a variety of other phys- during stance caused unwanted motion at the hip and trunk
iological functions [9], [ 151, [22], [28]. For individuals in both the coronal and sagittal planes. This suggested
with neurological impairments affecting the lower extrem- that feedback control of the hip and trunk was also needed.
Initial attempts at using PID control of coronal plane
Manuscript received July 20, 1989; revised January 29, 1991. This work hip angle were made on subjects who wore braces which
was supported by the Rehabilitation Research and Development Service of
the Department of Veterans Affairs. restricted all motion at the ankle, knee, hip, and trunk
J . J. Abbas is with the Department of Biomedical Engineering, Case except for coronal plane motion at the hip [ 131. Feedback
Western Reserve University, Cleveland, OH 44106 and the Motion Study of coronal plane hip angle was provided by goniometers
Laboratory, Cleveland Veterans Affairs Medical Center, Cleveland, OH
44106. mounted on the braces. These preliminary experiments
H. J . Chizeck is with the Department of Systems Engineering, Case demonstrated the feasibility of the approach [ 131, but the
Westem Reserve University, CLeveland, OH 44106 and the Motion Study practicality of this implementation was limited by the ex-
Laboratory, Cleveland Veterans Affairs Medical Center, Cleveland, OH
44106. tensive bracing used. Comparative trials with and without
IEEE Log Number 9100509. feedback control were not performed.
TABLE I
MODIFIED
DIGITAL PID CONTROLLER EQUATIONS
Wmaa
Fig. 3. Feedback control system implementation. This diagram illustrates the various components o f the feedback control sys-
tem and their interaction; functional units are enclosed in solid line boxes and physical devices are indicated by the shaded
boxes. The feedback controller calculations are performed by a MicroVAX I1 laboratory minicomputer (DEC). The stimulator
and command processor components of the system are contained in a V40 microprocessor (NEC) based unit. Communication
between the feedback controller and the stimulator/command processor is executed via a dual-port RAM (DPR). Note that the
open-loop control system utilizes only the V40 and the EPROM components of the system.
the muscles. The system-to-patient interface displays sys- A mounting and signal processing procedure was de-
tem status information and menu selections for the user. veloped [Fig. 4(a)] to measure the coronal plane hip angle
In the closed-loop system, the stimulus phase parame- while the subject is standing with both feet touching the
ters are read from tables on the DPR, rather than from ground. A sensor pair was mounted on each side of the
RAM. The VAX-based feedback controller generates the frontal aspect of the pelvis such that the line formed by
stimulation parameters based on sensor measurements. It the sensor was parallel to the femur and passes just lateral
then sends stimulation parameters to the stimulator via the to the anterior superior iliac spine (ASIS). This mounting
DPR in real-time. Also, real-time communication be- location was adequate, but not necessarily optimal; the
tween the V40 and the VAX via the message board of the effect of different mounting locations was not exhaus-
DPR is enabled to allow for upper level control actions, tively explored. Using circuitry and processing for the set
such as switching between control modes. of sensor signals [Fig. 4(b)], the accuracy of the mea-
surement was improved to within 3.2". This noisy mea-
Sensors and Signal Processing surement was adequate, however, for contro1,purposes and
The liquid-metal strain gage (Whitney Gage, Park for discerning the differences between the two control
Medical Electronics) consists of a silastic tube (outer di- strategies compared in this study, as is demonstrated be-
ameter = 0.04 in) filled with liquid mercury. As the tube low.
is elongated, the diameter of the liquid-metal capillary is
reduced and its length increased, resulting in an increase Subject Description-Electrodes and Muscles
in electrical resistance. These gages can be used in a stan- The evaluation of the coronal plane hip controller was
dard modification of a Wheatstone bridge circuit (as typ- performed on two adult male subjects with complete tho-
ically used with low resistance strain gages) to obtain a racic-level lesions of the spinal cord. The subjects are
voltage output. In static tests the gages have been shown long-term participants in the lower extremity FNS re-
to generate outputs linearly proportional to strain up to search and development project at the Motion Study Lab-
levels of 40%; in dynamic tests they have been shown to oratory at the CVAMC [24]. One subject (CG), level T7,
be frequency-independent with no phase-shift up to 50 Hz was 32-years-old, three years post-injury , and had been
[8]. The sensors can be easily mounted on the skin using in the program for seven months. The other subject (RL),
Tegaderm" (3M), a stretchable medical adhesive dress- level T9, was 42-years-old, nine years post-injury , and
ing. The size, shape, and elasticity of the sensors are such had been in the program for four years.
that they are nonrestricting and cosmetically acceptable. As a part of their participation in the research program,
There are several examples of biomedical applications of both subjects were equipped with a set of functioning per-
these gages [31]. cutaneous, intramuscular (IM) electrodes implanted in
The accuracy of these sensors is compromised by sev- several muscles of the lower extremities. The electrodes
eral factors. When mounted on the skin, they directly were made from 76 pm 10-strand, stainless-steel, Teflon-
measure skin stretch, not joint angle. Also, as motion at coated wire; with each strand 25 pm in diameter. Details
the joint is transduced through the skin, spatially-distrib- of the electrode design and implantation procedure was
uted noise is added and the signal is filtered. Comparison given in [24]. In this study, a subset of those electrodes
of measurements of skin stretch at locations about the joint was used for muscular activation. In particular, IM elec-
center of rotation with measurements of joint angle ob- trodes were used bilaterally for activation of the vasti of
tained using a strobe photographic technique exhibited a quadriceps for knee extension, gluteus maximus for hip
velocity-dependent hystersis [ 11. The viscoelastic prop- extension, gluteus medius for hip abduction, and the pos-
erties bf skin which could be responsible for this observed terior portion of adductor magnus for hip adduction-hip
hysteresis have been documented elsewhere [23]. extension. To supplement the set of percutaneous elec-
ABBAS A N D CHIZECK: FEEDBACK CONTROL OF HIP ANGLE IN PARAPLEGIC SUBJECTS 69 I
I I
Fig. 4. (a) Four stages of sensor signal processing. First, a coordinate transformation converts measurements of skin stretch to
joint angles. This was done by differentially mounting pairs of sensors about the joint center. Second, spatially distributed noise
was reduced by averaging the outputs of several pairs of sensors. Third. high-frequency noise was reduced by smoothing.
Finally, the velocity-dependent hysteresis was reduced by the addition of a derivative-dependent term. (b) Sensor circuitry and
processing for estimating joint angle. The two sensors on the left side (R1 and R2) were placed in series in one arm of a
Wheatstone bridge (also in series with a fixed resistance), and the two on the right (R3 and R4) were placed in series in the other
arm. The coordinate transformation and spatial averaging were done in the circuitry. The outputs of the bridge were differentially
amplified (IOOOX). low-pass filtered ( A = 25 Hz).and sampled at a rate of 50 Hz.The signal was then smoothed using a five-
point rectangular window ( J = 6 Hz),the difference approximation to the derivative was calculated, scaled, and added to the
smoothed signal.
ztL*2anqt
cases, surface electrodes placed on the posterior aspect of
the thigh were used to activate hamstrings for additional
diitubonce
hip extension. a
-fix!
Systems to be Evaluated
PW mop
The objective of the coronal plane hip controller was to
maintain neutral upright hip angle position in the coronal Fig. 5 . Diagrams of the OPen-looP (OL)and closed-loop (CL)control SYs-
tem compared in this study.
plane. A set of experiments was designed to compare its
performance with the response of an open-loop stimula-
tion system. Block diagram representations of the systems
tested are given in Fig. 5 . The first system (OL) uses open-
loop stimulation of extensors at the knee, hip, and trunk,
and of hip abductors and adductors. The second system
(CL) adds the two-stage .subsystem controller to control
coronal plane hip angle. Note that the OL system was
equivalent to the CL system with the feedback loop cut.
That is, the OL system utilized stimulation levels for all
muscles that were the same as those of the CL system
under a condition of zero error. When using either the OL
or CL system, the subject stood with his hands on an in-
strumented horizontal bar for auxiliary support.
Application of the Disturbance Fig. 6. Sketch of experimental setup illustrating the subject and instru-
The two controllers were evaluated by comparing their .mentation used in the experiments.
performance in maintaining upright coronal plane posture
as the subject stood wiFh both feet on the ground and with that would not corrupt the skin stretch sensor measure-
both hands on a horizontal support bar. Disturbances were ments. A rectangular frame (24 X 15 in) made of light-
applied in the form of ap impulse-like force which moved weight aluminum was placed around the subject's waist
the pelvis laterally, thus causing the subject to bend in the (Fig. 6 ) . The frame was held in place by two bolts, each
coronal plane at the -hip. Application of the disturbance of which passed through the frame on either side and ter-
was done such that the applied force was measurable and minated at a pad which was placed on the lateral aspect
the region of application was restricted to areas of &e skin of the pelvis just below the iliac spine. With proper place-
692 IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING. VOL. 38, NO. 7. JULY 1991
ment of the pads, the disturbance force was applied in the TABLE I1
DESCRIPTION
OF AUXILIARY
SUPPORT BARMEASUREMENTS
region of the bony landmark, thus limiting the deforma-
tion of the skin to a small area. The disturbance was ap- A static, planar double inverted pendulum model with arms added as
plied by tugging on a rope attached to one side of the rigid segments was used to study the moments acting at the hip due to the
aluminum frame. Although this type of disturbance is not hands. Assuming that any deviation from upright does not affect the
moment arm significantly, the following equations describe the moment
perfectly repeatable, the applied disturbance force was at the hip due to the hands (Mhh), the sum of the forces acting on the
measured using a load cell (Model 3167, Lebow Assoc. horizontal bar in the coronal and vertical directions (Fhh.,.
Fhh.;),and the
sum of the moments about right support point of the horizontal bar (MI):
Inc.) placed in series with the rope (Fig. 6).
Mhh == (FLX + FRx) * /I + (FLZ + FRZ) * w/2
I/
0 0.5 1
..it
1.5
bod
2
,J
2.5
Fig. 7. Example traces showing the applied disturbance, the resulting hip angle trajectories, and the two components of the
Support force measurements. The top plot displays example disturbance force tracings for an OL (solid line) and a CL (dashed
line) trial. The observed difference in the traces was due to the trial-to-trial variation in the application of the disturbances; in
general, there was no significant difference between the OL and CL disturbances. The second plot displays the resulting coronal
plane hip angle trajectories for each system. While the observed reduction in peak angular excursion and increase in natural
frequency were typical, the improvement in steady-state error was, in general, greater than that which was observed in the
+
example. The third and fourth plots display the coronal plane support forces ( F , x F z x ) and the difference in the vertical plane
support forces (Flz - Fzz) measured on the horizontal support bar. Also note here that the observed differences in the traces
were due to trial-to-trial variations; there were no consistently significant differences between the support force measurements
of the OL and CL trials.
gain, which was then used throughout that day. To ac- CL trial are given in Fig. 7. Two different types of quan-
count for any changes in sensor offset, the sensor offset tities were calculated for each trial. First, three pei$or-
was recalibrated at the beginning of each run with the sub- mance measures were used to compare the responses of
ject standing. For the subjects used in this study, 100% the two control systems: rms error, steady-state error and
abduction corresponded to approximately 20". compliance. The error was defined as the deviation from
Each day of experimentatiod consisted of four to eight the desired hip angle (0'). The root-mean-squared ( h S )
runs. In a run, either the OL or the CL system was tested; error provided an overall measure of system output track-
the order in which the systems were tested was varied and ing. This has often been used as the primary measure of
the subject was not aware of which system was being performance for FNS control systems, as well as other
tested. Each run lasted approximately three minutes with types of control systems. Steady-state error was calcu-
six to eight minutes of rest between runs. In each run the lated as the mean value of hip angle over the last half-
subject first stood up using prespecified stimulation se- second of the trial. The ratio of initial peak angular ex-
quences. The frame used for application of the distur- cursion of the hip to the peak disturbance force was used
bance was then secured about his waist and his feet placed as a measure of compliance.
on markers Axed on the force plates. Upon a trigger from Secondly, four other response attributes were calcu-
the subject, either CL or OL operation was initiated. If lated, each with respect to the steady-state value for that
the CL system was being tested, the real-time controller trial (Fig. 8). Rms-wrt-ss was calculated as the square
was then activated. The subject was instructed to look root of the mean squared deviation from the steady-state
straight ahead and to use his arms as little as possible for value. Overshoot, return time, and setting time were cal-
support. Disturbances were then applied repeatedly (every culated as described in Fig. 8. Together, they characterize
5-10 s) for the duration of the run. A trial was defined as the transient response of the system.
a 3-s segment of data which was triggered by the appli-
cation of the disturbance. The results reported below were RESU LTS
based on 448 such trials over five days. When the subject
signalled the termination of a run the feedback controller, Pei$ormance Measures
if active, was deactivated. A fixed level of stimulation to The purpose of the evaluation was to compare the re-
the extensors was maintained while the frame was re- sponse of the CL system with that of the OL system. The
moved; the subject then sat down using pre-specified first objective was to determine if, on any given day, the
stimulation sequences. action of the controller resulted in significant changes in
the performance measures. For each trial, the values of
Temporal Response Measures each of the measures were calculated; then the values ob-
Sample tracings of input disturbance force, hip angle tained on each day were grouped by controller type. The
trajectory, and auxiliary support forces for an OL and a data for each of the measures for the OL and the CL
694 IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, VOL. 38. NO. 7. JULY 1991
TABLE 111
COMPARISON OF OL AND CL SYSTEM PERFORMANCE A N D ATTRIBUTES
experiment day 1 2 3 4 5
subject # I 1 2 2 2
# OL trials 12 26 21 23 69
# CL trials 38 34 60 56 43
Performance measures
r m s error X X X X X
S . S . error X X X X X
compliance X X X X
Fig. 8. Diagrammatic representation of the temporal response character- System attributes (with respect to steady-state value)
istics used for the control system evaluation. The quantities shown are de- rms-wrt-ss X X X
fined as follows: overshoot = l(p2 - s ~ ) \ return
, time = irlO% - rpl. return time X X
settling time = is 10%- t p l . where pl = initial peak of the angular tra- settling time 0 0
jectory (positive or negative), p? = second peak of the angular trajectory overshoot 0 0
(positive or negative), fpl = time at which p l occurred, ss = steady-state ~~ ~ ~
value of hip angle, dh = 10%* I(pl - ss)J trlO% = time at which the The top four rows indicate the experiment day, subject, and the number
angle had returned to within ss f dh, is 10% = time at which the angle of OL and CL trials performed on that day. The middle three rows denote
had settled to within ss f dh. statistically significant differences ( P < 0.05 using the Kurskal-Wallis
nonparametric test) between the OL and CL performance measyes on each
of the five experiment days. X indicates that the CL values for that day
groups were tested for normality of distribution using the were significantly less than the OL values, 0indicates that the OL val-
ues were significantly less than the CL. An empty slot in the table indicates
chi-square test. For some df the measures the normality that there was no significant difference between the two groups. Similarly,
of the distribution could not be confirmed on some days. the bottom four rows indicate statistically significant differences for the
Therefore, statistical significance of the difference be- system attributes calculated with respect to the steady-state value.
tween the OL and the CL data was evaluated using the
Kmskal-Wallis nonparametric test (see Table 111) [27].
Differences were considered significant for P < 0.05. days, reduced return time on two days, increased settling
Mean values and standard error bars for each of the tem- time on two days, and increased overshoot on two days.
poral measures on each day are shown in Fig. 9. Note that
on all five days the CL controller resulted in a statistically Disturbance Forces
significant reductiorl in fms error. Although this improve- As indicated earlier, the disturbances for each trial were
ment was primarily a reflection of the reduction in steady- not precisely identical, but they were measured for each
state error, it was also influenced by the transient char- trial. The possibility that differences in disturbance rms
acteristics of the response. The CL controller resulted in value, peak value, energy, or rate of application might
a significantly reducedLcompliance on four days. have influenced the system response was tested via cor-
In addition to comparing daily performance of the two relation analysis. On all days, all of the performance
controllers, the overall performance with regard to each measures were found to be poorly correlated with the dis-
measure was evaluated. The daily mean values of each turbance measures ( p << 0.5); none of the correlation
performance measure were normalized with respect to the coefficients were consistently different from zero. Thus,
OL mean value for that day. For each day, the percent the disturbances for each trial were treated as being iden-
reduction for each measure due to the action of the con- tical, i.e., the data were not normalized.
troller was calculated as 100 * (1 - (CL measure/OL
measure)). A paired t-test was used to determine the sig- Support Forces
nificance of the changes in the absolute measures over the The measurements of auxiliary support force were used
set of all days. Results indicate that the CL controller gen- to determine whether or not the observed differences in
erated significant reductions in each of the performance performance were due to differences in support force used
measures: a 41 % reduction in rms error, a 52% reduction by the subjects. Recall that the hip moment due to the
in steady-state error, and a 22% reduction in compliance. action of the hands is the weighted sum of two terms: the
sum of the coronal plane forces measured on the support
System Attributes (Calculated with Respect to Steady- posts (Mhh,c),and the difference between the vertical plane
State) forces measured on the support posts (Mhhqt,).The rms
The quantities of rms-wrt-ss, return time, settling time, value, rms kalculated with respect to the initial value, and
and overshoot were all calculated with respect to the the peak value of the force component of each of these
steady-state value for each trial. The influence of the two terms, and of their weighted sum were calculated for
steady-state error must be considered in interpreting these each trial. These data were then analyzed to determine if
results. For example, a fast return to an incorrect steady- there were significant differences between the OL and CL
state position may result in a faster return or settling time trials. Results indicated that, for any given measure, there
or a smaller Overshoot than the same controller would was no consistent difference across the five days. The rms
achieve if it did not have the steady-state error. This said, value of cpronal plane force was lower for the CL trials
we note that the CL controller reduced rms-wrt-ss on three on three days, lower for the OL trials on one day, and
ABBAS A N D CHIZECK: FEEDBACK CONTROL OF HIP ANGLE IN PARAPLEGIC SUBJECTS 695
240
0
0
D
I
m 4
%a 4
Y * m
0
*
1 2 3 4 5 1 2 3 4 5 1 2 7 4 5
mxperiment day experiment day experiment day
(b) (C)
rma-m-t-ss return ti-
o h
::::p]
1 2 3 4 5 1 2 3 4 5
experimnt day experiment day
+
;;11;.1-
0.5
- 0
0
1
1 2 3
( 4
r e t t l i n g time
4 5
10
ot *. . . . .
0
1 1
(e)
overshoot
3 4 5
experimnt day experiment day
(f) ( g)
Fig. 9. Mean values of the performance measures and system attributes versus experiment day for both the OL and CL systems.
Subject no. 1 was used on days 1. 2 and subject no. 2 was used on days 3, 4, and 5 . (a) Displays the mean value of rms error
for the OL system (0) and for the CL system (X) versus experiment day. Vertical bars indicate f one standard error;
the units of the y-axis are given in percent-maximum-abduction (pma) as described in the text. Similar plots for steady-state
error, compliance, rms-wrt-ss, return time, settling time and overshoot are shown in (b)-( g), respectively.
showed no difference on one day. This was the most con- System Response Variability
sistent of the measures; no other.measure was higher or The variance of the performance measures and system
lower for one group on more than two days. The lack of attributes was used to evaluate the effect of the controller
observed consistent differences in the measures of the mo- on system variability. Results of this analysis are pre-
ments exerted by the hands during the OL and CL trials sented in Table IV. The chi-square test was used to eval-
suggests that the differences observed in the OL and CL uate the normality of the distribution of each group. If the
system responses cannot be attributed to the action of the distribution of the data for either the OL or CL group was
hands. Tracings of the time course of the forces contrib- determined not to be normal for a given day, comparison
uting to each component of the moment exerted by the of the variances would not be valid. These cases are in-
hands for the sample trials are given in Fig. 7(c) and (d). diated in the table with a + . For the remaining entries
Although in these trials the CL system required less upper in the table, Bartletts test [27] was used to test the equal-
body support this was not a general trend. ity of the variances. As indicated in Table IV, in several
In the course of the experiments, an additional differ- instances the CL controller brought about significant re-
ence between the OL and CL systems was observed. Fol- ductions of the variance of the performance measures. The
lowing the disturbances, the CL system often resulted in reduction in variance brought about by the CL controller
larger weight shifts between the two legs than the OL sys- was seen most clearly on days 1, 2, and 5.
tem. To quantify this observation, the rms values of the
difference in the vertical component of the two ground
DISCUSSION
reaction force vectors were calculated. On three days this
measure was lower for the OL system, which reflects the Performance Measures
higher compliance of that system. In some situations, this Lower extremity FNS systems which provide stance
result may be undesirable, thus weight distribution on the currently have two primary deficiencies: the high level of
feet may be a useful feedback signal for the controller. energy expenditure and the limited ability to respond to
-
696 IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING. VOL. 38. NO. 7. JULY 1991
amount of weight supported by the arms varied from day trically stimulated muscle, IEEE Trans. Biomed. Eng., vol. BME-
to day. The difference in system responses of the two sub- 34, pp. 140-147, Feb. 1987.
[7] G. A. Borges, K. R. Ferguson, and R. Kobetic, Development and
jects was most apparent in the measure of compliance operation of portable and laboratory electrical stimulation systems for
(Fig. 9). Subject no. 1joined the program three years post- walking in paraplegic subjects, IEEE Trans. Biomed. Eng., vol. 36.
injury; at the time of the first day of trials he had been in pp. 798-801, July 1989.
[8] T. D. Brown, L. Sigal, G. 0. Njus, J. M. Njus, R. J. Singerman,
the program for only a few months. The strength of the and R. A. Brand, Dynamic performance characteristics of the liquid
implanted muscles had not yet fully redeveloped, as was metal strain gage,.!. Biomechan., vol. 19. no. 2, pp. 165-173, 1986.
evidenced by a high compliance and minimal ability to [9] H. J. Chizeck, R. Kobetic, E. B. Marsolais, J. J. Abbas, E. Simon,
and I. Donner, Control of Functional Neuromuscular Stimulation
withstand disturbances. The second day of experimenta- systems for standing and locomotion in paraplegics, Proc. IEEE,
tion with this subject took place one month later. An in- vol. 76, no. 9, pp. 1155-1 165, 1988.
crease in muscle strength during this time was apparent, [IO] H. J. Chizeck, P. E. Crago, and L. S. Kofman, Robust closed-loop
control of isometric muscle force using pulsewidth modulation, IEEE
as indicated by reductions in compliance, return time and Trans. Biomed. Eng., vol. BME-35, July 1988.
settling time. Subject no. 2 had been in the program for [ I l l H. J. Chizeck, R. Lalonde, C. W. Chang, J. A. Rosenthal, and E.
several years and thus had greater muscle development B. Marsolais, Performance of a closed-loop controller for electri-
cally-stimulated standing in paralyzed patients, in Proc. 8th Ann.
than subject no. 1. The effect of the greater muscle RESNA Con$, 1985, pp. 231-233.
strength and passive stiffness was seen in lower values of [I21 P. Crago, H. J. Chizeck, M. Neuman, and F. T. Hambrecht, Sen-
hip compliance and return time, and a greater ability to sors for use with functional neuromuscular stimulation, IEEE Trans.
Biomed. Eng., vol. BME-33, pp. 256-268, Feb. 1986.
withstand disturbances. [I31 I. Donner, Electrical stabilization of the hip in the lateral plane via
a feedback controller, M.S. Thesis, Case Western Reserve Univ.,
Cleveland, Ohio, 1986.
CONCLUSIONS 1141 G. F. Franklin, I . D. Powell, and A. Emami-Naeimi, Feedback Con-
A feedback controller of coronal plane hip angle has trol of Dynamic Systems. Reading, MA: Addison-Wesley, 1986,
pp. 103-106.
been implemented in paraplegic subjects using FNS. The 151 T. F. Hambrecht and J. B. Reswick, Eds., Funcrional Electrical
performance of the feedback-controlled system has been Stimulation: Applications in Neural Prostheses, Vol. 3. New York:
quantitatively compared to that of an open-loop control Marcel Dekker, 1977, vol. 3, 1977, pp. 119-140.
161 H. Hemami and B. F. Wyman, Modeling and control of constrained
system. In the comparison, the feedback controller dem- dynamic systems with application to biped locomotion in the frontal
onstrated a 41 % reduction in rms error, a 52 % reduction plane, IEEE Trans. Automat. Contr., vol. AC-24, pp. 526-535,
in steady-state error, and a 22 % reduction in compliance. Aug. 1979.
171 R. T. Jaeger, Design and simulation of closed-loop electrical stim-
The observed changes in these performance measures ulation orthoses for restoration of quiet standing in paraplegics, J.
suggest that the feedback-controlled system provided a Biomech., vol. 19, pp. 825-835, 1986.
greater level of disturbance rejection than the open-loop [I81 2. Jin, R. Kobetic, and H. J. Chizeck, Parallel bars for three di-
mensional force measurement, in Proc. ICAART, Montreal, Que-
system. This improvement may allow the user of the FNS bec, 1988, pp. 148-149.
system to have greater use of the hands during stance. The 1191 G. Khang and F. E. Zajac, A Planar musculoskeletal model for
feedback controller also resulted in a reduction in the studying posture induced by functional neuromuscular stimulation,
in Proc. 9th Annual Con$ Rehab. Eng. (RESNA), Minneapolis, MN,
variance of these performance measures on some of the 1 9 8 6 . p ~ 445-447.
.
experiment days. This type of improvement in the repeat- [20] A. Kralj, T. Bajd, R. Turk, J. Krajnik, and H. Benko, Gait resto-
ability of the system response may help to make the FNS ration in paraplegic patients: A feasibility demonstration using mul-
tichannel surface electrode FES, J . Rehab. RD, vol. 20, p. 3, 1983.
system easier to learn to use and easier to use. [21] A. Kralj, T. Bajd, R. Turk, and H. Benko, Posture switching for
prolonging functional electrical stimulation standing in paraplegic pa-
tients, Paraplegia, vol. 24, pp. 221-230, 1986.
ACKNOWLEDGMENT [22] A. Kralj and T. Bajd, Functional Electrical Stimulation: Standing and
The authors thank E. B. Marsolais and R. Kobetic for Walking afier Spinal Cord Injury. Boca Raton, FL: CRC Press,
1989.
their assistance and collaboration and R. Shepheard for 1231 J. F. M. Manschot and A. J. M. Brakkee, The measurement and
the illustration presented in Fig. 6. modelling of the mechanical properties of the human skin in vivo-
11. The model, J. Biomechan., vol. 19, no. 7, pp. 517-521, 1986.
1241 E. B. Marsolais and R. Kobetic, Functional electrical stimulation
REFERENCES for walking in paraplegia, J. Bone Joint Surg., vol. 69, pp. 728-
733, 1987.
111 I . I . Abbas and H. I . Chizeck, Processing skin stretch sensor mea- [25] -, Development of a practical electrical stimulation system for
surements of hip and trunk angles, in Proc. 40th ACEMB, Niagara restoring gait in the paralyzed patient, Clin. Orrhop. Rei. Res., vol.
Falls, NY, Sept. 1987, p. 222. 233, pp. 64-74, Aug. 1988.
121 J. J. Abbas, H. J. Chizeck, G. Borges, P. Chow, P. Lambert, and [26] R. J. Nakai, P. E. Crago, and H. J. Chizeck, Development of a
M. Moynahan. A software structure for implementing multi-state closed-loop control system for hand function restoration using func-
feedback controllers in FNS systems, in Proc. IEEE-EMBS, New tional electrical stimulation, in Proc. 2nd Int. Con& Rehab. Eng.,
Orleans, LA, 1988, pp. 1653-1654. 1984, pp. 543-544.
131 J. Allin and G. F. Inbar, FNS control schemes for the upper limb, [27] J. Neter, W. Wasserman, and M. H. Kutner. Applied Linear Statis-
IEEE Trans. Biomed. Eng., vol. BME-33, pp. 818-828, 1986. tical Models. Chicago, IL: Illinois, Richard D. Irwin. Inc., 1985.
[4] B. I . Andrews and R. M. Baxendale, A hybrid orthosis incorporat- [28] P. H. Peckham, Functional electrical stimulation: Current status and
ing artificial reflexes to spinal cord damaged patients, J. Physiol., future prospects of applications to the neuromuscular system in spinal
vol. 380, p. 19, 1986. cord injury, Puraplegia. vol. 25, pp. 279-288, 1987.
[5] T. J. Bajzek and R. J. Jaeger, Characterization and control of mus- [29] J. S. Petrofsky and C. A. Phillips, Computer controlled walking in
cle response to electrical stimulation, Ann. Biomed. Eng., vol. 15, the paralyzed individual, J. Neurol. Ortho. Surg., vol. 4, pp. 153-
pp. 485-501, 1987. 164, 1983.
[6] L. Bemotas, P. Crago, and H. J. Chizeck, Adaptive control of elec- [30] U. Stanic and A. Tmkoczy, Closed-loop positioning of hemiplegic
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING. VOL. 38. NO. 7, JULY 1991
patients joint by means of functional electrical stimulation. IEEE research interests include neuromotor prostheses, control and analysis of
Trans. Biomed. E n g . , vol. BME-21, pp. 365-370, 1974. biomechanical systems, and the design of neural network controllers for
J. E. Stone, M. H. Madsen, J. L. Milton, W. F. Swinson, and J. L. dynamic systems.
Turner, Developments in the design and use of liquid metal strain
gages, Experiment. Mechan., pp. 129-139, June 1983.
R. L. Waters, D. R. McNeal. W. Faloon. and B. Clifford, Func-
tional electrical stimulation of the peroneal nerve for hemiplegia, J.
Bone Joint Surg., vol. 67-A, pp. 792-793, 1985.
Howard Jay Chizeck (S74-M79) was born on
G. F. Wilhere. P. E. Crago, and H. J. Chizeck, Design and eval-
March 27, 1952 in Columbus, OH. He received
uation of a digital closed-loop controller for the regulation of muscle
the B.S. and M.S. degrees in systems and control
force by recruitment modulation, IEEE Trans. Biomed. Eng., vol.
Engineering from Case Western Reserve Univer-
BME-32, pp. 668-676, Sept. 1985.
sity, Cleveland, OH. in 1974 and 1976, respec-
tively, and the Sc.D. degree in Electrical Engi-
neering and Computer Science from the
Massachusetts Institute of Technology, Cam-
bridge, in 1982.
James J. Abbas (S88) was born in Lawrence, He is an associate professor in the Systems En-
MA in 1961. He received the Sc.B. degree in gineering and Biomedical Engineering Depart-
bioelectrical engineering from Brown University, ments at Case Western Reserve University. His research interests involve
Providence, RI, in 1982 and the M.S. degree in applications of control engineering to biomedical problems. Current proj-
biomedical engineering from Case Western Re- ects include the design and analysis of controllers for the electrical stimu-
serve University, Cleveland, OH in 1989. He is lation of paralyzed muscle, adaptive control of drug delivery, identification
currently working towards the Ph.D. degree in and control of cerebral ventricle volume dynamics for hydrocephalus man-
Biomedical Engineering at Case Westem Reserve agement, algebraic modelling of DNA sequence-geometry relationships and
University. applications of algebraic systems theory to coding.
From 1982 to 1984 he taught secondary school Dr. Chizeck is a member of the AAAS, Sigma Xi, the Rehabilitation
mathematics in Mzuzu, Malawi. Since 1986 he has Society of North America (RESNA), and the International Federation of
been employed as a Biomedical Engineer at the Motion Study Laboratory Automatic Control (IFAC) Technical Committee on Biomedical Engineer-
at the Cleveland Veterans Affairs Medical Center in Cleveland, OH. His ing.