You are on page 1of 50

Section 3 - System Parameters

Table of Contents

1. INTRODUCTION ...................................................................................................... 3

2. SET SYSTEM PARAMETERS ................................................................................. 5

2.1 SYSTEM CONFIGURATION .................................................................................... 5


2.2 APRS .............................................................................................................. 16
2.3 GENERATOR LINES ........................................................................................... 25
2.4 ARCHIVE DEVICE .............................................................................................. 31
2.5 PRINTER DEVICE CONFIGURATION ..................................................................... 37
2.6 QUERY/RETRIEVE DEVICES ............................................................................... 44
2.7 WORKLIST DEVICES .......................................................................................... 45
2.8 PATIENT DATA CONFIGURATION ........................................................................ 46
2.9 BACKUP ........................................................................................................... 49
2.10 RESTORE ......................................................................................................... 50

Document #726-602-G1 3-1


Section 3 - System Parameters

Document #726-602-G1 3-2


Section 3 - System Parameters

1. Introduction
The DIS-200 Configuration consists of a series of screens, each screen offering tabs for
access to menus specific to the parameters of the individual screen. Refer to the chart
below as a guideline. Screen specific instructions will follow.

Parameter Available Settings


System Configuration • Site Info – Allows user to enter site-specific information
including Site Name and address.
• Interfaces - Specify Detector and Iris type, set base
system frequency, Set Cine and Fluoro Reference
signals, Enable or Disable ABS Control and set
Maximum spot Acquisition Rate
• System Options – Specify Scan Converter and
Reference Image Display, Set Thumbnail Frame,
Enable or Disable Patient List print status, Set Patient
Delete Options.
APR • Use to set default Window, Level and Edge for new
APRs created within the clinical application.
• Set Printer Configuration, Layout format, Batch Print
Processing settings, Batch print Edge Level and
Batch Print Polarity in Printing Mode
• Set device Configuration, Auto Send, Batch Device
Edge Level and Batch Device Polarity in
Archive/transfer Mode
Generator Lines • Enable or Disable Two-State and Analog Inputs; Set
Analog, Pulsed and Relay outputs.

• Add Archive Device


• Select Archive Device
Archive Device • Set Storage Commit timeout
• Set Network Information

Printer Device • Set Printer Destination and Priority level.


• Set Film Orientation and Size
• Set Image width, height and pixel size
• Set Border and Empty Image Density
• Enable or disable trim and set level of polarity

Query/Retrieve Device • Configure DICOM Query and Retrieve Service Class


Providers
Worklist Device • Configure DICOM Worklist information Service Class
Providers
Patient Data Configuration • Set Required field for Patient, Study and Series data

Document #726-602-G1 3-3


Section 3 - System Parameters

Document #726-602-G1 3-4


Section 3 - System Parameters

2. Set System Parameters

2.1 System Configuration

Use instructions in Section 1 to access Service Applications screen.

Click on CONFIG to check system parameters.

Main Image Screen with Service Application User Interface

Configuration button defaults to System Configuration Screen

Document #726-602-G1 3-5


Section 3 - System Parameters

2.1.1 Site Info


This screen contains important system information that is entered when the system is
programmed. This information should not be changed unless ownership of the system
changes hands. At that point, the Site Name and address should be changed but
nothing else.

Configuration Screen – System Configuration – Site Info

Document #726-602-G1 3-6


Section 3 - System Parameters

This data will appear on all stored images


Site Name
For record-keeping purposes
Site Address
Phone number for Shimadzu
Technical Support Phone
Number
For customer reference
Manufacturer Name
For customer reference
Manufacturer Address
For customer reference
Model Name
The system serial is assigned by Shimadzu and carries forth through
System Serial Number the life of the system. It must NEVER be changed. The serial number
must be provided when contacting service personnel and is used to
track parts and service throughout the life of the system.

2.1.2 Interfaces
Use Interfaces to set system parameters necessary for running system calibration and
other functions.

Configuration Screen - System Configuration - Interfaces

Document #726-602-G1 3-7


Section 3 - System Parameters

Not used. Only the Small Format Camera is available for this system.
Detector Type

Not used. Only the Direct-Couple Iris is available for this system.
Iris Type

Click on down arrow to Enable or Disable ABS Control


ABS Control

Click within circle to Select either 30/60 or 25/50 Hz


Base System Frequency

Click on down arrow to enable or disable Injector Control


Injector Control

Click on down arrow to display pull-down menu.


RAD Expose on Failure

Acquire black image

Click on down arrow to display pull-down menu


RAD Expose on Timeout

Document #726-602-G1 3-8


Section 3 - System Parameters

Left click box to activate Continuous Fluoro. To turn off this feature, click
Continuous Fluoro in box over the arrow.

Use check boxes to select pulse rate (pulse per second)


Pulse Rates

Click on down arrow to display pull-down menu. Rates are stated in


Maximum Spot Acquisition
fps (frames per second). Current selections:
Rate

Click on down arrow to display pull-down menu.


Spot Subtract Calibration

Document #726-602-G1 3-9


Section 3 - System Parameters

2.1.3 System Options

Use System Options to check which options are enabled for system use or to change
settings with the addition of optional equipment.

Configuration Screen - System Configuration – System Options

Click on down arrow to select from available options. Screen indicates


Scan Converter
no scan converter installed.

Document #726-602-G1 3 - 10
Section 3 - System Parameters

Not used
Reference Image
Display
Use Slider to set which frame within a sequence will be displayed as the
Thumbnail Frame thumbnail.

Click on down arrow to select from available options


DICOM Storage Commit

Click on down arrow to select from available options


Modality Performed
Proc. Step

Manually enter the maximum number of images to be saved to the Print


Print ShotSave Options ShotSave folder. There is no limit to the number of images that can be
saved in this folder.

NOTE: Before the new number can take precedence, the system must
be shut down and restarted.

Click on down arrow to select from either Performing or Referring


Display/Print Physician
physician.
Name

Provides warning prior to deletion of patient records.


Patient Delete
Operations

Document #726-602-G1 3 - 11
Section 3 - System Parameters

Click on down arrow to select from available options

Click on down arrow to select from available options


Digital Host Control

Opacification Delay
Use slider to set appropriate Opacification delay setting.

Click on down arrow to select from available options


Patient List Archive
Status

Click on down arrow to enable or disable


Patient List Print Status

Document #726-602-G1 3 - 12
Section 3 - System Parameters

2.1.4 Image Processing

Use Image Processing to turn on/off H-Reverse and V-Reverse settings for image
processing. Use down arrow to turn Horizontal and Vertical Reverse on or off.

Not available
at this time.

Click on down arrows to turn


these options on or off

Configuration Screen - System Configuration – Image Processing

Document #726-602-G1 3 - 13
Section 3 - System Parameters

2.1.5 Monitor LUTs (Lookup Tables)

Use Monitor LUTs to set specifications for in-room Primary and Reference Monitors.

Configuration Screen - System Configuration – Monitor LUTs

Set LUT for In-room Use down arrow to make selection


Primary Monitor

Set LUT for In-room Use down arrow to make selection


Reference Monitor

Document #726-602-G1 3 - 14
Section 3 - System Parameters

2.1.6 DICOM
Use DICOM tab to set specifications for DICOM functions. All fields can be altered
manually by positioning the cursor over them and typing in the information.

Configuration Screen - System Configuration – DICOM

Document #726-602-G1 3 - 15
Section 3 - System Parameters

2.2 APRs

The APR screen allows access to a series of screens from which specific settings can be
made for Fluoro, Spot, Roadmap, Printing and Archive Transfer. Changes made within
these screens will become the default settings for new APRs created within the clinical
application. While the edge level can be adjusted within the clinical application, this is the
only place where the Brightness and Contrast levels for acquisition can be changed.

2.2.1 Fluoro
Use Fluoro to set default system settings or to change settings as required.

Configuration Screen – APR – Fluoro

Document #726-602-G1 3 - 16
Section 3 - System Parameters

Use down arrow to select the desired level of Edge Enhancement: None,
Edge Level 1, 2, 3, or 4.

This setting can be overridden by using the APR settings within the
clinical application.

Use Slider to select default brightness level.


Brightness

Contrast
Use Slider to select default contrast level.

Document #726-602-G1 3 - 17
Section 3 - System Parameters

2.2.2 Spot

Use Spot to set default system settings or to change settings as required.

Configuration Screen – APR – Spot

Use down arrow to select the desired level of Edge Enhancement:


Edge Level
None, 1, 2, 3, or 4.

This setting can be overridden by using the APR settings within the
clinical application.

Document #726-602-G1 3 - 18
Section 3 - System Parameters

Auto-Image Optimization (AIO) is recommended for all unsubtracted


Brightness
image types. However, in the absence of AIO, the default Brightness
setting can be configured for each image type.
Use slider control to select the default Spot Brightness value. This will
only be applied if AIO for the Spot image type is OFF.

Use slider control to select the default Spot Contrast value. This will
Contrast
only be used if AIO for the Fluoro image type is OFF.

Use slider to set default level of subtracted brightness


Subtract Brightness

Use slider to set default level of subtracted contrast


Subtracted Contrast

Document #726-602-G1 3 - 19
Section 3 - System Parameters

2.2.3 Roadmap

Configuration Screen – APR – Roadmap

Use down arrow to select the desired level of Edge Enhancement:


Edge Level
None, 1, 2, 3, or 4.

This setting can be overridden by using the APR settings within the
clinical application.

Brightness
Use slider control to select the default Brightness value.

Document #726-602-G1 3 - 20
Section 3 - System Parameters

Use slider control to select the default Contrast value.


Contrast

2.2.4 Printing

Use Printing to make default system settings or to change settings as required.

Configuration Screen – APR – Printing

Document #726-602-G1 3 - 21
Section 3 - System Parameters

Printer Configuration
Click on down arrow ( ) to select default printer from a list on
drop-down menu.

Print Layout Format


Click on down arrow ( ) to select default print layout from a list
on drop-down menu.

9 on 1
12 on 1
Batch Print Processing Settings
Click on down arrow ( ) to select default batch print
processing settings from a list on drop-down menu.

Batch print Edge Level Settings


Click on down arrow ( ) to select default edge level setting
from drop-down menu.

Batch Print Polarity Select the default polarity to be used for the Print File and Print
Sequence operations. Click on down arrow ( ) for drop-down
menu.

Document #726-602-G1 3 - 22
Section 3 - System Parameters

2.2.8 Archive Transfer


Use Archive/Transfer to make default system settings or to change settings as required.

Configuration Screen – ARP – Archive/Transfer

Archive Configuration Not currently functional

Auto Send
Click on down arrow ( ) for drop-down menu.

Batch Device Processing


Click on down arrow ( ) for drop-down menu.
Settings

Document #726-602-G1 3 - 23
Section 3 - System Parameters

Batch Settings
As Acquired

Batch Device Edge Level Select the level of edge enhancement to be used for the Print File
and Print Sequence operations. Click on down arrow ( ) for
drop-down menu.

This setting can be overridden within the clinical application.


Batch Device Polarity
Select the polarity to be used. Click on down arrow ( ) for drop-
down menu.

Process/Annotations Select processing features by clicking on down arrow for drop


down menu:

Document #726-602-G1 3 - 24
Section 3 - System Parameters

2.3 Generator Lines

Generator settings are made initially when the generator is connected to the system.
Use the Configuration function to check settings or to change them if necessary. Refer
to Section 2 – Generator Interface for more information.

2.3.1 Two State Inputs


Use Two State Inputs to enable or disable settings as required.

Configuration Screen - Generator Lines – Two State Inputs

Click on down arrow to enable or disable individual input settings


Two-State Inputs

Document #726-602-G1 3 - 25
Section 3 - System Parameters

Signal Definitions: The signals between the x-ray generator and Imaging system are defined
as follows:

Signal Description
Fluoro On This signal is active when fluoro radiation is ON. It causes the system to
display the live fluoro image on the image monitor in “Fluoro” mode. When
inactive, the system will maintain the last full frame of fluoro on the image
monitor. This signal is overridden when the Prepare for photospot signal is
active.
P Fluoro ON (Pulsed Fluoro ON) This signal indicates to the system that the generator
is currently generating an x-ray pulse. It is used to synchronize the
acquisition of images with the x-ray exposure.
Prep Rad Exp This signal is active when the operator is preparing for Radiographic
Exposure. It signals when the generator is switching from “Fluoro” to
“Radiographic” mode. This signal is used for both cine and spot
acquisition.
Req Rad Exp This signal is active when the operator has requested a Radiographic
exposure and is used for both cine and spot acquisition.
Rad Exp On This signal indicates to the system that the generator is producing
radiographic radiation for photospot imaging. It is considered an optional
signal. However, the signal line must be installed when the Spot in/Out
relay open condition does not terminate an exposure. Additionally, this
line must be installed for Tomography procedures.

NOTE: Failure to install the RAD EXPOSE ON line will limit the usable
radiographic exposure time to 200 milliseconds.

Document #726-602-G1 3 - 26
Section 3 - System Parameters

2.3.2 Analog Outputs

Configuration Screen - Generator Lines Screen – Analog Outputs

Click on down arrow to enable or disable function


ABS Control

Document #726-602-G1 3 - 27
Section 3 - System Parameters

2.3.3 Digital Outputs

Configuration Screen - Generator Lines Screen – Analog Outputs

Fluoro V- Drive
Click on down arrow to enable or disable function

Document #726-602-G1 3 - 28
Section 3 - System Parameters

Show Timing Diagram

Click to display graph representing fluoro pulse statistics:

Pulse Width

Click on down arrow to select pulse width from list on drop-down menu.

Polarity

Click on down arrow to select polarity level from list of drop-down menu.

Phase Delay

Enter phase delay in seconds manually.

Document #726-602-G1 3 - 29
Section 3 - System Parameters

2.3.4 Relay Outputs

Configuration Screen - Generator Lines Screen – Relay Outputs

Spot Expose
Click on down arrow enable or disable function

Injector Sync
Click on down arrow enable or disable function

Document #726-602-G1 3 - 30
Section 3 - System Parameters

2.4 Archive Device


Use this screen to make adjustments for a specific archive device. This screen is also
used to add a new Archive Device or to delete an existing one.

Device Specified under Archive Device

Configuration Screen – Archive Devices

Document #726-602-G1 3 - 31
Section 3 - System Parameters

2.4.1 DICOMNetwork

1. Position cursor over device name in left column and left click.
2. The Archive Device Configuration screen will be displayed.

Archive Device Configuration Screen


3. Make changes as required.

Document #726-602-G1 3 - 32
Section 3 - System Parameters

To add additional devices:

1. Left click the Add device button in the lower left of the
screen.
2. The “Select Device to Add” dialog box will be displayed

“Select Device to Add” dialog box

3. Enter name of the device in the Configuration window.


4. Select device type.
5. Click “Create” to add or “Cancel” stop the process. The Add Storage AE window
will be displayed.

Document #726-602-G1 3 - 33
Section 3 - System Parameters

6. Complete information in screens provided for either DICOM Network or DICOM


Media.
7. Click SAVE to complete setup process or Exit to abort the setup and exit the
system. You will be asked to confirm changes.

8. Click YES to save changes; NO to abort the process.

2.4.2 DICOMMedia

1. Position cursor over device name in left column and left click.
2. The Archive Device Configuration screen will be displayed.

Archive Device Configuration Screen


3. Make changes as required.

Document #726-602-G1 3 - 34
Section 3 - System Parameters

To add additional devices:


1. Left click the Add device button in the lower left of the
screen.
2. The “Select Device to Add” dialog box will be displayed.

“Select Device to Add” dialog box

3. Enter name of the device in the Configuration window..


4. Select device type.
5. Click “Create” to add or “Cancel” stop the process.
6. Complete information in screens provided for either DICOM Network or DICOM
Media.

Document #726-602-G1 3 - 35
Section 3 - System Parameters

7. Click SAVE to complete setup process or Exit to abort the setup and exit the
system. You will be asked to confirm changes.

8. Click YES to save changes; NO to abort the process.

Document #726-602-G1 3 - 36
Section 3 - System Parameters

2.5 Printer Device Configuration

Customize print functions for specific archive device

Installed print devices will be


listed in this column

Configuration Screen – Printer Device Configuration

To view or change settings for a specific printer:

1. Position cursor over device name in left column and left click.
2. The Printer Device Configuration screen will be displayed.
3. View selections or make changes using the available options.

NOTE: Different types of print devices have specific options available.

Document #726-602-G1 3 - 37
Section 3 - System Parameters

DICOM Print Device Settings

Configuration Screen – Print Device Configuration –DICOM Printer

Specific printer will be listed in this box

Use arrow to access drop down box to select to high, low or medium priority.

Use arrow to access drop down box to select either portrait or landscape film
orientation.

Document #726-602-G1 3 - 38
Section 3 - System Parameters

Use arrow to access drop down box to select film size from those available.

Use arrow to access drop down box to select magnification type.

Use arrow to access drop down box to select specific border density

Use arrow to access drop down box to select empty film density

Use arrow to access drop down box to select to either to apply trim or not
(Yes/No)

Document #726-602-G1 3 - 39
Section 3 - System Parameters

Use arrow to access drop down box to select to either normal or inverted as
polarity setting

Use sliders to set Minimum and Maximum density

When Selections are complete, click either Save to retain information or exit to leave the
configuration screen without saving information.

Adding a DICOM Printer

1. Left click to access “Select Printer to Add” dialog box in


the lower left corner of the screen.
2. The Select Printer to Add screen will be displayed.

“Select Printer to Add” dialog box

3. Enter the name of the printer in Configuration window.


4. Select printer type.
5. Click “Create” to add or “Cancel” stop the process. The Add Printer AE window
will be displayed.

Document #726-602-G1 3 - 40
Section 3 - System Parameters

Use drop down arrow to select from options


available

Use drop down arrow to select either Normal


or Inverted Polarity
Use drop down arrow to select format from
available options.

Document #726-602-G1 3 - 41
Section 3 - System Parameters

Enter the name of the Windows Printer in the


widow provided

Use drop down arrow to select either Portrait


or Landscape paper orientation.

Use drop down arrow to select from available


options.

Enter information manually in the windows


provided.

Document #726-602-G1 3 - 42

Use drop down arrow to select from available


Section 3 - System Parameters

Use drop down arrow to select from available


options.

Use drop down arrow to select from either


Duplex or Simplex.

Enter information manually in the windows


provided.
Enter information manually in the windows
provided.
Enter information manually in the windows
provided
Enter information manually in the windows
provided.

Document #726-602-G1 3 - 43
Section 3 - System Parameters

2.6 Query/Retrieve Devices

The Query / Retrieve Menus will allow configuration of DICOM Query and Retrieve
Service Class Providers.

Click to display Query/Retrieve devices dialog box.

Complete information and then click Next.

Document #726-602-G1 3 - 44
Section 3 - System Parameters

2.7 Worklist Devices

The Worklist Devices Menus will allow configuration of DICOM Worklist information
Service Class Providers.

Click to display Query/Retrieve devices dialog box.

Complete information and then click Next.

Document #726-602-G1 3 - 45
Section 3 - System Parameters

2.8 Patient Data Configuration

2.8.1 Patient Data


Specify required and non-required fields for entry of patient data. Patient data cannot be
processed if a field marked “required” is left blank. Information cannot be entered if field
is marked not used; the box is “grayed” out. If a field is marked used, information can be
entered but is not required for processing patient data entry.

Configuration Screen – Patient Data Configuration – Patient Data

Other than the patient name, which is preset, all fields can be changed by using the drop-down menu.

Document #726-602-G1 3 - 46
Section 3 - System Parameters

2.8.2 Study Data


Specify required and non-required fields for entry of study data. Study data cannot be
processed if a field marked “required” is left blank. Information cannot be entered if field
is marked not used; the box is “grayed” out. If a field is marked used, information can
be entered but is not required for processing study data entry.

Configuration Screen – Patient Data Configuration - Study Data

Other than the date and time, which are preset, all fields can be changed by using the drop-down
menu.

Document #726-602-G1 3 - 47
Section 3 - System Parameters

2.8.3 Series Data


Specify required and non-required fields for entry of series data. Series data cannot be
processed if a field marked “required” is left blank. Information cannot be entered if field
is marked not used; the box is “grayed” out. If a field is marked used, information can
be entered but is not required for processing series data entry.

Configuration Screen – Patient Data Configuration – Series Data

Other than the Series Number and Date, which are preset, all fields can be changed by
using the drop-down menu.

Document #726-602-G1 3 - 48
Section 3 - System Parameters

2.9 Backup

Click to access back-up function. A confirmation screen will be


displayed

Backup Configuration Dialog Box

Click yes to continue backup procedure


The Backup Configuration window will be displayed.
Repeat as necessary for each Config file to be backed up
(Sysconfig.xml, LUTS.xml, APRConfig.xml and Userconfig.xml)

Select file name and type and then click save. The files will be saved in the Backup folder (or other
location as specified).

If “NO” is selected, the system will default to System Configuration –


Site Info Screen

Document #726-602-G1 3 - 49
Section 3 - System Parameters

2.10 Restore
This function allows the restoration of a previous configuration that was saved using
Backup function.
Click to access restore configuration. A confirmation screen will
be displayed

Restore Previous Configuration Dialog Box

Click yes to continue restore procedure


The Windows “Open” window will be displayed.

Select file name and type and then click Open. The selected file will be restored.
Repeat as necessary for each Config. File to be restored (SysConfig.xml and UserConfig.xml)

If “NO” is selected, the system will default to System Configuration –


Site Info Screen

Document #726-602-G1 3 - 50

You might also like