Test my product using Sine or Random?

As you know, in the vibration world, there are quite a few test "types" to which you can
expose your product. The major choices are Sine, Random, Classical Shock, Transient
Shock, Field Recorded Time History, Sine-on-Random, Random-on-Random and Sine-
and-Random-on-Random. Frequently, our customers will request advice on which of
these types of test to run on their product, and in particular, how to choose between the
two most common test types: sine or random. Their desire is to know which test, sine or
random, is best to most quickly pinpoint flaws in their product. If they can only run one
test, either sine or random, which should it be? Recently, I received an even more specific
request from a customer. This customer (Don) presented both a sine test and a random
test and wanted to know, given both a sine test and a random test, how he could
determine which is the most severe? Let's take a look at the two tests and decide how to
answer to his question.
=============================================================
Here is Don’s question:

How would the following specifications compare with regard to amplitude/severity?


------------------------------------------------------------------------------------------------------
Sine test:
3.5 G at 5 to 50 Hz
1.5 G at 50 to 300 Hz
Limit vibration to 0.4 inches double amplitude
Test all axes at the same level

Figure 1: Sample Sine test profile

Acceleration Profile
8
Acceleration (G peak)

0.2
5 10 100 300
Frequency (Hz)
------------------------------------------------------------------------------------------------------
Random test:
0.01500 G2/Hz from 10 Hz to 0.01500 G2/Hz at 40 Hz
0.01500 G2/Hz at 40Hz to 0.00015 G2/Hz at 500Hz.
Results in 1.05 G rms
Test all axes at the same level

Figure 2: Sample Random test profile

Power Spectral Density


1x10 -1
Acceleration (G²/Hz)

1x10 -2

1x10 -3

1x10 -4

10 100 500
Frequency (Hz)
=============================================================

This is a very valid and interesting question. Why? Because the answer is not obvious.
The sine vibration is measured in G peak, while the random vibration is measured as G
rms, with the peak G levels typically left to a statistical assumption. A quick calculation
tells us that the random test, which can have peak values up to 4 or even 5 times the RMS
level, will apply 4 x 1.05 G rms, or 4.20 G peak to our product. Since the sine test is only
3.5 G peak, we would expect the random test to be more damaging, right?

Looking for support from some equations, let’s make some reasonable assumptions: 1)
failures due to vibration are caused at the peak G level seen by the product, 2) most
products have resonances at one or more frequencies, and 3) at these resonances the
vibration levels applied to the product are amplified by the Q factor of the resonance.
Following this train of logic we conclude that the failures will occur when the vibration is
at one of the resonant frequencies. Now when we use a sine vibration, the full vibration
levels are concentrated at the resonant frequency, and the vibration levels are simply
amplified by the amplification factor Q.

Aproduct,sine = Q x Acontrol,sine (1)

When we use a random vibration, it is not so simple because not all of the vibration is
amplified by the resonance. Let’s further assume the resonance has a Q factor of 5 or
more. In that case the resonance will act as an amplifying band-pass filter with
amplification equal to Q, and a bandwidth equal to ∆f, where
Q = fn / ∆f (2)
fn = resonant frequency
∆f = half-power bandwidth of the resonance

It will also be helpful to refer to the following relationship, which tells us that, for a given
RMS level, the PSD level is inversely proportional to the full bandwidth of the random
spectrum. In more general terms, a more concentrated random vibration will have a
higher PSD value.

PSDcontrol = Acontrol,rms2 / ∆F * (3)


∆F = full bandwidth of the random PSD
*
equation (3) applies to flat spectrum only

A random test is defined in terms of a PSD, which is an amplitude-squared measure, so at


the resonant frequency the PSD levels of the product will be amplified by Q2. Since the
resonance acts as an amplifying band-pass filter, we can approximate the vibration levels
at the product by looking at just the energy at the resonant frequency that passes through
and is amplified by the resonance (again assuming the random waveform has up to 4
sigma peaks):

Aproduct,peak = 4 x Aproduct,rms
= 4 x [ PSDproduct x ∆f ]1/2
= 4 x [ Q2 x PSDcontrol x ∆f ]1/2
= 4 x [ PSDcontrol x Q * fn ]1/2 (4)

From equation (4) we note three features of the peak amplitude for a random test:
1. It is proportional to only the square root of Q! As a result, a high-Q resonance
will result in a more severe test in Sine than it will in Random, if all other
parameters are equal.
2. It is proportional to the square root of the resonant frequency, fn, so the higher the
resonant frequency, the higher the peak values in the output.
3. Referring back to equation (3), we also note that the more concentrated the
random vibration are, the higher will be the peak vibration levels.

Now we can also compare the peak vibration levels on the product for both Sine and
Random tests by comparing equations (1) and (3) with Aproduct,sine = Aproduct,peak,random.

Acontrol,sine = Aproduct,sine / Q
= Aproduct,peak,random / Q
= 4 x [ fn x PSDcontrol / Q ]1/2 (5)

With this we have an equation with which, given a value for the Q factor, we can
compare a sine test with a random test. From this result, shown in figure 3, we see that
the Sine test from figure 1 is much more severe than the random test from figure 2. Only
in the case of a resonance at 50 Hz, where the sine test steps down in amplitude, with
Q=5 does the random test level come close the sine test level. So the equations are telling
us that the Sine test will be more severe than the Random test!

Figure 3: Comparison of the Sine equivalents of the Random profile, with various Q.
Note that for higher Q, the Sine equivalent of the Random profile has a lower amplitude.
5
Acceleration (G peak)

Sine

Random
equivalent
Q=5

Q=10

Q=20

Q=50

0.1 Q=100
5 10 100 500
Frequency (Hz)
===========================================================

Application

At the time Don asked his question, John was at Sperry Marine in Charlottesville, VA,
setting up for some equipment installation training with Dave Maxwell and Joe
Reisinger. This was a great time to demonstrate and test for the differences! As a single
test is worth a thousand opinions, we set up to run both the random test and the sine test
on a slip plate. The slip plate had two elements mounted on it, each with different
resonant frequencies. The elements were aluminum masses attached by threaded rods of
different lengths and thicknesses, with accelerometers mounted on the mass at the top of
each rod connected to channels 4 and 6.

What we needed to do for the comparison was run the tests, and then look at the G levels
seen. We knew they would be 3.5 G peak for the sine test at the Control point, and,
making a 4 sigma peak assumption, we expected to find 4.2 G peak at the Control point
for the random test also. The two plots in figures 4 and 5 show the controlled test along
with the response data for the two vertical rods with masses attached.
Figure 4: Sine test results
Acceleration Profile
100
Freq: 26 Hz Freq: 62 Hz
Ch4: 42 G Ch6: 97.04 G
Acceleration (G peak)

10

Ch6
Ch4
Control 0.15 10 100 300
Demand Frequency (Hz)

Figure 5: Random test results


Power Spectral Density
1x10 2
Freq: 27.2 Hz Freq: 62.3 Hz
Ch4: 44 G²/Hz Ch6: 74 G²/Hz
1x10 1 Q ≈ 48 Q ≈ 110

1x10 0
Acceleration (G²/Hz)

1x10-1

1x10-2

1x10-3

1x10-4

1x10-5

Ch6 1x10-6
Ch4
1x10-710 100 500
Control
Frequency (Hz)
Demand
While running the tests, we also simultaneously streamed the accelerometer data to the
hard disk drive for later analysis. This allowed us to be able to make a direct comparison
of the peak G levels for each of the tests. As expected, at the Control point, the peak G
levels for the Sine test were of course 3.5 G peak and 1.5 G peak. The Random test
levels at the Control point were 1.05 G rms, as expected, and 4.8 G peak, which is a little
bit higher than the 4 sigma peaks we predicted, but not unusual for a Gaussian random
vibration (figure 6).

Figure 6: The random test vibration levels measured at the Control point.

Now, as we were running the test, we observed the resonant elements mounted to the slip
table going berserk! This also showed up on the controller plots (figures 4 and 5), which
showed much larger accelerations on channels 4 and 6 than measured at the Control
point. What could be going on here? Recall from equations (1) and (4) above, that when
we have resonances in the product, they will amplify the vibration levels. The
accelerometers monitoring our two resonant elements show the resonance frequencies are
27 Hz and 62 Hz, right in the middle of our test range!

Let us examine the time data for those accelerometers, which we had conveniently
streamed to the hard disk drive, and see what is happening at the resonant elements. This
may shed more light on the test equivalences. First looking at the frequency response for
the Ch4 accelerometer (figure 5), we estimate a Q factor of 48 and a resonant frequency
of 27 Hz. However we note that the bandwidth, ∆f = fn / Q = 0.6 Hz, which is less than
the frequency resolution of the PSD plot (using 800 lines). This is a good indication that
the Q factor estimated from this PSD will be underestimated. Since we had recorded the
time waveforms to disk, we are able to load these waveforms into MatlabTM and use high
resolution spectral post-processing (13000 lines) to zoom in on the peaks. With this more
accurate estimate, we find the Q level to be 140. Computing using equation (4), we
expect a peak level of 4 x [ 0.015 x 140 x 27 ]1/2, or 30 G peak . Figure 7 shows the
actual acceleration levels on Ch4, where we find the peak G level was 28 G.
Next looking at the frequency response for the Ch6 accelerometer (figure 5) we estimate
a Q factor of 110 and a resonant frequency of 62 Hz. Again we find the bandwidth, 0.6
Hz, to be less than the frequency resolution of the PSD plot, so we need to post-process
this recorded waveform to obtain a more accurate estimate of the Q factor. Using 13000
lines of resolution we estimate the Q factor to be 200. Computing using equation (4), we
expect to see peak levels of 4 x [ 0.00673 x 200 x 62 ]1/2, or 37 G peak. Figure 8 shows
the actual acceleration levels on Ch6, where we find the peak G level was 42 G.

Figure 7: The random test vibration levels measured on Ch4 (27 Hz resonance)

Figure 8: The random test vibration levels measured on Ch6 (63Hz resonance)
Observations

Let’s make a table of the results, so we can get the whole picture:
Q Sine Random Random Predicted (Eqn. 4)
(G peak) (G rms) (G peak) (G peak)
Control point 3.5 G 1.05 G 4.8 G 4.2 G
Ch4 (27 Hz resonance) 140 42 G 8.35 G 28 G 30 G
Ch6 (62 Hz resonance) 200 97 G 10.9 G 42 G 37 G

From these results we find that, at the control point, the random test has higher peak G
levels than the sine test. This would lead us to believe that the random test was the more
severe test. However, at the resonant points the sine test has higher peak G levels than
the random test, so from the product’s perspective the sine test was the more severe test.

You may also notice that the sine resonance peak values were not as high as would be
predicted by the Q factor. By reviewing the recorded time data for Ch4 we are able to
determine that, at the 27 Hz resonance, the sine sweep passed through the resonance too
quickly to fully excite the resonance. Repeating the test using a slower sweep rate, or a
resonance search-and-dwell feature, would allow us to more accurately find the
resonance frequency and Q factor. From this we can conclude that for high Q values, to
get the full Q amplification effects of a sine test, you must either sweep slowly or do a
resonance dwell at each resonance.

By examining the recorded data for Ch6 we are also able to determine that at the 62Hz
resonance the acceleration levels were so high that they exceeded the accelerometer’s
measurement capacity, and therefore the measured waveform was saturated well below
the actual acceleration levels. This is an important point, because quite often
accelerometers are sized based on the test profile’s acceleration level, while the
resonances may see 10x or 100x the profile acceleration level. Repeating the test using a
higher capacity accelerometer is required to accurately measure the acceleration levels at
that resonance.

The actual random peak values are a little different from the predicted values, but this is
to be expected due to the random nature of the waveform. Each random test you run will
be different from any other random test, so the peak values will vary from test to test, and
even over different time intervals within a single test.

Conclusion

The relative severity of a sine test and a random test will vary depending on your
product’s resonant frequencies and Qs. In general, when sine and random tests have the
same peak vibration levels at the control point, the product will see higher vibration
levels with a sine test than with a random test due to the resonances in the product.
Equation (5) gives you way to convert a random PSD into an approximately equivalent
sine peak acceleration, as long as you know the resonant frequencies and Q factors of the
product.
However, you must also consider that sine tests only excite a single product resonance at
a time, so a sine test will not test the interaction between two resonances in your product.
Since a random test excites the full frequency range all at the same time, it can be used to
find problems resulting from the interaction between two resonances.