You are on page 1of 1

ICAL-Standardization Report

Instrument-Type : SPECTROTEST CCD Date : 6/25/2018


Instrument-Number : 122218/07 Time : 12:56

Company/Owner : SPECTROTEST

Sample(s) : RH18
Measurements : 6
Excitation-Mode : SPARK

CCD-Nr. RSD[%] RSD-Limit[%] Correc.[%] Correc.-Limit[%] Remark

1 5.09 2.50 3.80 6.70


2 2.57 2.50 3.90 7.20
3 3.27 2.50 5.20 8.40
4 1.99 2.50 2.70 6.70
5 2.16 2.50 2.80 6.50
6 2.05 2.50 2.80 6.90
7 1.94 2.50 3.40 7.40
8 1.82 2.50 3.50 8.00
9 1.91 2.50 4.30 7.70
10 0.26 2.50 71.10 6.50 Not OK
11 2.25 2.50 6.00 7.00
12 2.27 2.50 7.20 6.50 Not OK
13 2.81 2.50 7.50 8.00
14 0.00 2.50 --- 0.00 No Monitoring
15 0.00 2.50 --- 0.00 No Monitoring

Average of RSD : 2.34 (Limit: 2.50)

ICAL-Standardization not successfull