Operation Instructions

Infrared Thermometer

Article-No. 64 0314
1. Safety instructions 2. Please note 3.2 Display 5. Operation
5.1 Measure
Please prepare the measuring very When the ambient temperature varies A K-type temperature sensor is switched
carefully. The device should regularly be greatly, you must wait 30 minutes until the on To measure the temperature of an object,
cleaned with a dry cotton-cloth. Do not use temperature of the device has adjusted to B Laser is switched on (during the mea- press the trigger (7) briefly to switch the
white gas or solvent. Do not expose the the ambient temperature. surement shows a laser beam at ap- device on. Aim the Thermometer on the
meter to direct sunlight, ultraviolet rays or proximately the center of the measuring object which has to be measured and press
extreme temperatures. Protect your device spot) the trigger (7) again. As long as the trigger
from dust and keep the meter dry, moisture 3. Functions C Trigger is pressed (7) is pressed, the temperature is measured
may damage the device. Prevent it from 3.1 Description D Warning when reaching the maximum continuously. After releasing the trigger (7),
any shock. (H) and minimum (L) measured value the display will show the last read tempera-
1. Button for K-type temperature sensor E Indicates the unit (° C / ° F) ture for 7 seconds („HOLD“). After that, the
Repairs to the device must be carried out 2. Backlight / UP function F Holding the measurement result unit will switch off.
only by the Vogel Germany GmbH & Co. 3. LCD („HOLD“)
KG. 4. Button for emissivity G Measurement mode
5. Laser pointer and H Measurement mode value 5.2 Backlight
The battery should be taken out, if not in °C/°F toggle switch / DN function I Current measurement result
use for a long time. 6. MODE button J Battery indicator (low battery) To pinpoint the display even in dark places,
7. Trigger the backlight can be switched on. Press and
Regard to environmental protection: 8. Battery case hold the trigger (7) and press the button
Do not dispose of batteries or B C D EF (2) after it. The backlight will switched on.
device in domestic waste. Please Repeat the process to switch the light off.
give it to a special collecting SCAN
J SCAN
depot. K H L °C°F A K H L °C°F
5.3 Laser pointer
3 I
LAL MAX
1.1 Safety symbols HAL AVG LAL MAX The Laser pointer can be switched by pres-
Є MIN G HAL AVG H
Є MIN sing and holding the trigger (7). Press now
Important Safety Information, the button (5). The laser pointer will be
which points to the instruction 1 K MODE SET 4 switched on and the symbol for the laser (B)
K MODE SET
manual. will be shown in the display. To switch the
2 /UP T/ /DN
5 /UP T/ /DN laser pointer off, press the trigger (7) and
According the European CE the button (5) again.
safety standards, this unit
is compatible with the fol- 6
lowing standards: EN61326-1 4. Quick start 5.4 Switching °C/°F
EN61010-1 EN60825-1
Open Battery compartment by folding down The unity is changed after pressing the
Do not aim with the laser beam the battery cover (8) and connect the batte- button (5). The display will show, depen-
direct or indirect by a reflec- ry with the correct polarity. Press the trigger ding on the unity, °C (for Celsius) or °F (for
7
ting surface to an eye. (7) and in the display flashes „SCAN“ (C) Fahrenheit).
8 und the current measurement result (I) will
be displayed. The last reading will be shown
for 7 seconds in the display when the trigger
is released. Then the device turns off.

2 3
5.5 Change the default values 5.6 Measuring results 6. Battery replacement 8. Ratio of distance to spot

The emission value, the maximum and mi- Besides the display of current value (I) the The battery symbol (J) appears, when The thermometer measures an area that
nimum measured values are preset. These highest reading, the average and the lowest the battery is getting low. The battery has to depends on the distance to the object (see
can be changed as follows. The button value are also shown in the display. be changed now. figure) and of the impact angle.
(2) increases the value; the button (5)
reduces the value. Take the device with one hand at the back- The measured object or area should always
5.6.1 Highest reading side and grab with the thumb and forefinger be greater than the spot of the infrared ther-
of the other hand to the Notches of the mometer. This thermometer is the ratio of
5.5.1 Set the emission value Press the button (5) until the display battery cover. These are located below the the distance to spot diameter:
shows MAX (G). The value (H) next to it is trigger (7). Pull carefully the battery cover
Not every body has the same emissivity, the highest value in the measured series. forward and fold it downwards. Remove the D: S = 12:1
the thermometer consider this. Press the used battery and insert a new battery (9V,
button (4). On the display a Є appears type 6LR61, article-no. 60 9281). Please The diameter of the measuring spot is
and behind the set emission rate (I) two 5.6.2 Average value take care of the polarity and ensure that calculated: S = D/12.
bars ( | | ) are shown. Press the button there are no damages at the cable and
(2) or the button (5) to change the Press the button (5) until the display connections.
emissivity. To save the value, press the shows AVG (G). The value (H) next to it is Insert the battery in the housing with the
button (4) again and the bars disappear. the average value in the measured series. connection to the top and shut the cover.
For more information, please read chapter Take care that the cover shuts audible.
10. Emissivity.
5.6.3 Lowest reading All previous values are lost
when changing the battery!
5.5.2 Set the maximum value Press the button (5) until the display
shows MIN (G). The value (H) next to it is 38mm =
300mm
75mm =
900mm
132mm =
1500mm
Press the button (5) till the Display the lowest value in the measured series.
show „HAL“ (G). The preset of this value is
550°C. To change this value, press the but- Change of the diameter of
tons (2) or (5). The display will show the measuring spot
„H“ (D) as index for exceeding the limit.

5.5.3 Set the minimum value


7. Using the temperature sensor
Press the button (5) till the Display
show „LAL“ (G). The preset of this value is The thermometer has underneath the infra-
-50°C. To change this value, press the but- red sensor a plug for connecting the k-type
tons (2) or (5). The display will show temperature sensor. Please take care of
„L“ (D) as index for undercutting the limit. the right polarity. The delivered tempera-
ture sensor can be plugged in here. Press
the button (1) and the K-symbol will be
shown on the display. Press the trigger (7)
to get the temperature from the connected Positive Minus pole (-)
pole (+)
temperature sensor.
More K-type temperature sensor is availab- Connections for the k-type temperature
le in our product range. sensor are below the infrared sensor

4 5
9. Measurement principle 10. Emission ratio 11. Technical data

Every device emits infrared rays, whose The emission ratio of an object indicates Emission ratio: 0.1 ~ 1.0 adjustable
intensity is denpendant on its temperature. how much radiation the object emits in
These rays are detected and evaluated comparison to an ideal heat radiator (a so Spectral range: 8 – 14 µm
by the infrared thermomter. The basis for called black body). The higher this value is
this principle is the Stefan-Boltzmann-Law, the more infrared energy is emitted by this Measurement range: -30°C ~ +550°C (-22°F ~ 1022°F)
which says that the total ray capacity P for object.
a black body is dependant on the absolute Most organic materials have an emission Surrounding temperature range: 0°C ~ 40°C (32°F ~ 104°F)
temperature T (in K). Together with the ratio of somewhere between 0,85 and 0,98.
Stefan-Boltzmann-Constant this is: This thermometer calculates with a adjus- Humidity: 10 ~ 95% RH not condensing
table emission ratio.
P = σ • A • T4 Objects with a lower emission ratio will Storage: -20 °C ~ +60 °C (-4°F ~ 140°F), without battery
show a lower temperature during measu-
Real bodies radiate with an intensity which rement than the actual temperature. Be Accuracy: -30 °C ~ 0 °C: ± 3 °C or ± 3 %
is ε - times less than others: aware of necessary re-calculation when
measuring objects with a lower emission 0 °C ~ +100 °C: ± 2 °C or ± 2 %
P = ε σ • A • T4 ratio (e.g. reflecting metal objects).
> +100 °C: ± 3 °C or ± 3 %
For a contact free temperature measure-
ment you have to know the emission ratio ε, Repeatability: 1% from measuring value or 1°C
the heat radiation capability of the object to
be measured. Measurement time: 5 msec, 95% response time

Infrared thermometers determine the heat- Laser Energy: < 1 mW


energy that is radiated by an object. The
device focusses the energy through its len- Automatic switch off: approx. 7 seconds
se onto a special material, which releases
an electric signal that is dependant on the Battery: 9V, type 6LR61, Article No. 60 9281)
temperature. A micro-computer processes
this signal and delivers the measurement Dimensions: 165 x 122 x 50 mm
value to the display.
Weight: approx. 215 g incl. battery
The laserbeam is used exclusively for
the positioning and has no influence on Accessories: Batteriy (9V, type 6LR61, Article No. 60 9281),
the temperature measurment.. Manual, K-type Temperature feeler,
Workshop calibration certificate, Bag