Quick Reference
Instructions for Updating the i-STAT 1
with Serial Downloaders and Jammlite,
using www.pointofcare.abbott

Please refer to Technical Bulletin: Updating i‐STAT 1 Handheld Software Using 
www.pointofcare.abbott ART: 731335 for detailed instructions

Art: 732159‐00B Rev. Date: 11‐Oct‐17
i‐STAT is a trademark of the Abbott Group of Companies in various jurisdictions 1
[A.1] Required Materials & Setup:

Cables ‐ Abbott Supplied Computer ‐ Customer Supplied  i‐STAT 1 Analyzer


 USB Cable (a)  Windows XP or Windows 7 (c)  Electronic Simulator
 Power Cable (for DRC) (b)  Internet Access (c)
D *Installed DRC Drivers Required
c
R
C b

Cables ‐ Abbott Supplied Computer ‐ Customer Supplied  i‐STAT 1 Analyzer


 9‐Pin Null Modem Cable (a)  Windows XP or Windows 7 (c)  Electronic Simulator
 Power Cable (for DS/DRS) (b)  Internet Access (c)
D  Com port (c)
R c

S a

Cables ‐ Abbott Supplied Computer ‐ Customer Supplied  i‐STAT 1 Analyzer


 9‐Pin Null Modem Cable (a)  Windows XP or Windows 7 (c)  Electronic Simulator
 Power Cable (for DS/DRS) (b)  Internet Access (c)
 Com port (c)
D c
S a

b
Art: 732159‐00B Rev. Date: 11‐Oct‐17 2
[A.2] Download the Software:

Go to www.pointofcare.abbott/us/en/offerings/support/software-license-keys/istat-
system-software-update

Do you use i-STAT data management with the i-STAT System?


Select the appropriate use model

Select the Software Update File

Locate the Software Update File on your computer

Note: If a black window appears asking to overwrite files, 
answer (Y) Yes to all questions until the window disappears

Locate the JAMMLITE Utility on your computer

Art: 732159‐00B Rev. Date: 11‐Oct‐17 3
[A.3] Performing the Update via JAMMLITE:

Art: 732159‐00B Rev. Date: 11‐Oct‐17 4
[A.4] Transmitting Software to the i-STAT Analyzer:

Art: 732159‐00B Rev. Date: 11‐Oct‐17 5
[A.5] Troubleshooting:

Check Connections to the Downloader


 Power and serial cables are connected
 Driver for DRC model downloader is installed
o Go to www.pointofcare.abbott/us/en/offersings/support/software-
license-keys/istat-system-software-update
 Full Administrative Rights may be required to install

Choose a different Communication Port


 Trial of all available COM ports may be necessary

Click the Update Button


 Proceed as per the instructions on Step A.3

If message persists, contact Abbott Point of Care Technical Services:


1‐800‐366‐8020, Option 1, Option 3
Art: 732159‐00B Rev. Date: 11‐Oct‐17 6
[A.6] Finishing the Software Update Process:

Run the External Electronic Simulator


Analyzer Process Pathway:
• Power Button
• Menu Button  Administration Menu
• 3 – Quality Tests
• 4 – Simulator
• Follow the On Screen Instructions

Update is Complete

To Update Additional i-STAT Analyzers:


 JAMMLITE: Repeat these instructions from Step A.3
 Handheld-to-Handheld: Refer to “Quick Reference Instructions for Handheld-to-Handheld Transfer”

To Update Central Data Station:


 Refer to “Quick Reference Instructions for Updating Central Data Station”