2007 SERVICE NOTE Page 1 of 2

FILE TITLE VIVID 7 SN 76167B


TAB 1000
Loss of Doppler Spectrum

APPLIES TO:
Vivid 7 models with BT-02 Software v2.x.x and Vivid 7 RFT models with Software v3.x.x and 3.x.x.

PROBLEM:
Intermittantly no Doppler spectrum, just a constant signal in the spectrum.

CAUSE:
Marginal timing on Doppler board

SOLUTION:
Exchange Faulty Doppler board with another Doppler board that has Part Number: FB200865

PROCEDURE:

WARNING DO NOT TOUCH ANY BOARDS WITH INTEGRATED CIRCUITS PRIOR TO TAKING THE
NECESSARY ESD PRECAUTIONS:
1. ALWAYS CONNECT YOURSELF, VIA AN ARM-WRIST STRAP, TO THE ADVISED ESD
CONNECTION POINT LOCATED ON THE REAR OF THE SCANNER (TO THE RIGHT OF
THE POWER CONNECTOR).
2. FOLLOW GENERAL GUIDELINES FOR HANDLING OF ELECTROSTATIC SENSITIVE
EQUIPMENT.
1.) Shut down Scanner.
2.) Remove the Scanner Side Panel that gives access to the Card Rack.
3.) Remove the Card Rack Front.
4.) Exchange Doppler Boards.
5.) Install Card Rack Front, Side Panel and Boot Up.

Submitted by:Oddmund Fjordbakk

Copyright© 2005 and 2007 by General Electric Company Inc. All Rights Reserved
05. Sept. 2007 SERVICE NOTE Page 2 of 2

FILE TITLE VIVID 7 SN 76167B


For any further information, please contact:

Technical Support
GEHC - Americas GEHC - Europe GEHC - Asia

OLC - Europe Asia Support Center - Singapore


Phone: Toll Free Numbers for Phone: 65-277-3512
Online Services Ultrasound
(English/German): Fax: 65-272-3997
Phone: 1-800-321-7937 (USA/Canada)
Phone: +49 (0)212-2802-207 (General Imaging)
1-262-524-5300 (Latin America)
+49 (0)212-2802-208 (Cardiac) Japan Support Center
Fax: 1-414-647-4125
+43 (0)7682-3800-26 (Kretz Products) Phone: 81-426-48-2950
Fax: +49 (0)212-2802-431 Fax: 81-426-48-2902

REVISION HISTORY:

Revision Date Reason for change

B Sept. 2007 Removed Faulty Doppler Board part number

Copyright© 2005 and 2007 by General Electric Company Inc. All Rights Reserved

You might also like