You are on page 1of 3

SPECIFICATIONS OF DIGITAL STORAGE OSCILLOSCOPE WITH ISOLATED CHANNELS

There is a requirement of a set of items that consists of:

A. Digital storage oscilloscope with four isolated channels – 1 No;

B. High voltage probe compatible with the above DSO – 4 Nos.;

C. Accessories and documents; and

D. Installation at user’s site.

The details of every individual items are given below.

A. Specifications for Isolated Channels Digital Storage Oscilloscope - 1 No.

State Compliance
Sr.
Parameters Required Specification (Complied / Not
No.
Complied)
1 Isolated Channels 4
2 a) Sampling / Channel 2.0 G samples/s minimum
b) Bandwidth 200 MHz or more
3 Isolation Float Voltage 600 Vrms Min (reference to earth) CAT-II

4 Isolation Float Voltage 1200 Vrms Min (channel reference to channel


reference) CAT-II
5 Maximum Input Voltage 300 Vrms (tip to reference of standard passive probes),
1000 Vrms (tip to reference of high voltage passive
probe) CAT-II
6 Vertical Resolution At Least 8 Bits

7 Record Length 2500 points

8 DC Gain Accuracy ±3%

9 Horizontal Accuracy ± 50 ppm

10 Linear dynamic range ± 5V

11 Sensitivity 2 mV to 200 mV per div. ± 2V; >200mV to 5V/div. ±


50V
12 Time/Division Range 2.5 ns to 50s/division

13 Horizontal Trigger modes Auto, Normal, Single shot


Zoom
Source All four channels and External
Trigger
Edge (rising, level, pos/neg slope on any channel
falling)

1/3
Coupling AC, DC, Noise Reject, HF reject, LF Reject

Video Trigger on all lines or individual line, odd/even line


(NTSC, PAL, and SECAM)
Pulse Width – Trigger on a pulse width less than, greater than, equal
to, or not equal to a selectable time limit ranging from
33ns to 10s
Trigger View LCD colour
Display

14 Waveform Processing Add/Subtract, Multiply, FFT

15 Measurement System Automatic Waveform Measurement – Period,


Frequency, +width, -width, Rise time, Fall time, Max,
Min, Peak to peak, Mean, and RMS

16 Battery Capacity for two hot-swappable battery packs (Each


providing 4 hours of battery operation)

17 Power Measurements Instantaneous power waveform analysis, Waveform


analysis, harmonic analysis, switching loss, phase
angles, dv/dt and di/dt cursors

18 Operating Temperature 0 - 50°C

19 Humidity 60 – 90 % RH

20 Warranty Minimum 3 Years

21 Accessories  200 MHz, 10X passive probes (one per


channel).
 One Lithium-Ion battery with fuel gauge
For 4-hour battery life
 PC Connectivity Software – A collection of
programs that enable fast and easy
communication between MS Windows PCs and
Oscilloscope.
 User manual
 AC Adapter with Power Cord suitable for
Indian system.
 NIM/NIST-Traceable Certificate of
Calibration.
 Front Protective Cover.
 USB to RS-232 Cable for Scope-PC
Communication
 Soft Carrying Case
 Extra Battery shall be quoted separately
 Battery charger
 Storage media
 Card reader

B. Specifications for High Voltage Probe compatible with above mentioned power scope - 4 Nos.

Sr. Parameters Required Specification State Compliance


(Complied / Not Complied)
2/3
No.
1 Bandwidth DC to 200 MHz atleast
2 Voltage 1000 VRMS CATII
3 Floatable upto 600 VRMS CATIII
4 Attenuation 1:100
5 Rise Time ≤ 2.5 ns
6 Loading 4.0 pF at 100 MΩ
7 Compensation range in pF 10 to 22
8 Length 1 meter or more

C. Accessories and documents


Sr. No. Parameters State Compliance (Complied / Not
Complied)
1 Accessories: As required for full operation.
2 User manual and calibration certificate must be provided
Refer Sr. No. 6 of Note below before complying Sr. 1 of Table C.

D. Installation
Sr. Parameter State Compliance (Complied / Not Complied)
No.
1 Installation to be done by vendor at the site
Note :-
Sr. Clauses State Compliance
No. (Complied / Not
Complied)
2 If the vendor is not an OEM, he shall provide the authorization certificate from his
manufacturer showing his willingness to extend the maintenance and repair
support during Warranty. Without Authorization certificate, quotation will be
rejected.
3 Full data sheet of all products must be provided with quote
4 Vendor shall provide details of facilities they have in India to support warranty and
repair with quote
5 Detailed compliance table shall be attached with quote. Detailed application notes may be attached with
quote
6 Standard accessories and optional accessories must be clearly mentioned in the quotation.
7 Only required optional accessories will be purchased, this decision will remain with the indenter.
8 Quotations supported with the suitable catalogues and specification sheets will only be accepted
9 Please mention recommended accessories, if any

 The offers will be evaluated on the basis of the total cost of one set.

3/3