You are on page 1of 1

Manufacturer Endress + Hauser

General Model No Prosonic Flow 93WB1-KE5D20ACBBAA


Type ULTRA SONIC MICROPROCESSOR BASED INSERTION TYPE
Flow Sensor Ultrasonic Insertion Type
Sensor Range DN200 - DN4000
Sensor Housing SS 316L
Sensor Protection IP 68
Sensors Sensor Holder DN200 - DN1800
No. of sensors 2 Nos.
Mounting Weldable Cup of SS304
Sensor Cable Integral PVC Sheathed Cable 30 Mtrs.
Cable Connector Plug in type Nickled Brass
INPUT From Sensor
Measured Variable Flow Velocity
Electrical Measuring Range V = 0….15m/s at the specified measuring accuracy
Operable Flow Range Over 150 :1
Input Signal 3 - 30 VDC, Ri = 5 kOhms isolated
Transmitter & Indicator Prosonic Flow 93
Housing Powder Coated Die-Cast Aluminium for field mounting
Protection IP 67
Mounting Wall Mounted.
Device Accuracy +/- 0.5% of Reading plus +/- 0.02% of FSD
The dry calibration represents additional uncertainty due to mounting and actual pipe
Additional uncertainty properties. This additional uncertainty is better than ±1.5% typically. Manufacturer
issues Compliance Test Certificate as standard.
Temperature Compnsation Integral
Amb. Temp. Limit -40 to 80 Deg. C.
Process Temp. Limit -40 to 80 Deg. C.

Transmitter 4 Line ( 16 Characters each line) Illuminated LCD Display for Flow rate & Totalizer
Display
with Touch Control Function
Process Pressure PN16, 232 psi
Pressure Loss No pressure loss
Cable Entry 1/2" NPT(F)
Output 4 - 20 mA DC HART Isolated + Frequency
Power Supply 24 VDC ( 16 - 62 VDC )/230VAC
Power Consumption AC : < 18 VA , DC :< 10W
Load Impedence less than 250 ohms ( HART 4 - 20 mA)
Repeatablity ±0.3% for flow velocities >0.3m/s
Calibration Dry
Stability 0.01 % for One Year