You are on page 1of 6

MOONRAY COMPACT

TECHNICAL DATA SHEET – ALL TYPE DETECTORS


DETECTOR
BASE CONFIGURATION - MOONRAY COMPACT

FD 30 FD 22 I.I 13” I.I. 9”

Dimensions Dimensions Dimensions Dimensions


SID: ......................................................1034 mm SID: .....................................................
..................... 1060 mm SID: ................................................... 966 mm SID: ..................................................... 1019 mm
Free space: ..........................................834 mm Free space: ........................................
................................ 860 mm Free space: ...................................... 760 mm Free space: ......................................... 819 mm
Depth of the c-arm: ..............................670 mm Depth of the c-arm: ............................ 670 mm Depth of the c-arm: ........................... 670 mm Depth of the c-arm: ............................. 670 mm
Motorized vertical movement: ...............420 mm Motorized vertical movement: ............. 420 mm Motorized vertical movement: ........... 420 mm Motorized vertical movement: ............. 420 mm
Horizontal movement: ...........................230 mm Horizontal movement: ......................... 230 mm Horizontal movement: ....................... 230 mm Horizontal movement: ......................... 230 mm
Wig-wag: ..............................................20° (± 10°) Wig-wag: ............................................
................................ 20° (± 10°) Wig-wag: ........................................... 20° (± 10°) Wig-wag: ............................................. 20° (± 10°)
Orbital movement: .................................138° (+ 90° / - 48°) Orbital movement: ............................... 138° (+ 90° / - 48°) Orbital movement: ............................. 138° (+ 90° / - 48°) Orbital movement: ............................... 138° (+ 90° / - 48°)
Rotational movement: ...........................720° (± 360°) Rotational movement: ......................... 720° (± 360°) Rotational movement: ....................... 720° (± 360°) Rotational movement: ......................... 720° (± 360°)
Main Body dimensions: .........................1956x784x1686 mm Main Body dimensions: ....................... 1956x784x1686 mm Main Body dimensions: ..................... 1943x784x1879 mm Main Body dimensions: ....................... 1943x784x1859 mm
Main Body Weight: ................................340 kg Main Body Weight: .............................. 340 kg Main Body Weight: ............................ 340 kg Main Body Weight: .............................. 340 kg

Video Chain FD 30 Video Chain FD 22 Video Chain II 13” Video Chain II 9”


Flat digital detector 30x30cm CMOS technology Flat digital detector 22x22cm CMOS technology II of 13”, with 4 input fields: II of 9”, with 3 input fields:
Scintillator: ........................................ Caesium Iodide Scintillator: .......................................
................................ Caesium Iodide Nominal dimensions......................... 30.3 – 21.5 -16 - 12 cm Nominal dimensions .......................... 23 -16 - 12 cm
Dimension: ........................................ 29,6X29,6 cm Dimension: ......................................
................................ 21.5 x 21.5 cm Resolution: ..................................... 40 – 50 - 56 – 60 pl/cm Resolution: ........................................ 52 – 58 – 68 pl/cm
Pixel resolution: ................................ 1952X1952 Pixel resolution: ............................... 1416x1420 DQE %: .......................................... 65% DQE %: ............................................. 65%
DQE(0) @ 1u Gy, RQA5 .................. 72% DQE(0) @ 1u Gy, RQA5 ................. 75%
Image Bit Depth ................................ 16 bit Image Bit Depth............................... 16 bit Cmos camera Gigabit-Ethernet: Cmos camera Gigabit-Ethernet:
Fields of view .................................... 4 Fields of view ...................................
................................ 3 Matrix: ............................................ 1k x 1k pixels Matrix: ............................................... 1k x 1k pixels
Dot pitch ........................................... 151,8 µm Dot pitch ..........................................
................................ .151,8 µm Scan................................................ Progressive Scan................................................... Progressive
Digital Pre-view of the image rotation, without rx exposure Digital Pre-view of the image rotation, without rx exposure

2 MOONRAY COMPACT | REV 20/12/2019


SOFTWARE CONFIGURATION - MOONRAY COMPACT

Standard Software Optional Software

STATIC X-RAY IMAGING SOFTWARE DYNAMIC X-RAY IMAGING SOFTWARE FULL ENDOVASCULAR IMAGING SOFTWARE

Digital Static Storage Module Digital Dynamic Storage Module 25 fps Digital Dynamic Storage Module 25 fps
Digital matrix of dynamic images:.................... 1k x 1k Digital matrix of dynamic images: .................... 1k x 1k Digital matrix of dynamic images: ................... 1k x 1k
Storage on HD: ................................................ Automatic/Manual Storage on HD: ................................................ Automatic/Manual Storage on HD: ................................................
................................ Automatic/Manual
Storage images are visualized through icons with identifying marker Selectable storage frequency: ......................... 1,2,4,6,8,12 and 25 fps Selectable storage frequency:......................... 1,2,4,6,8,12 and 25 fps
Possibility to delete sequences or complete sessions. Storage images are visualized through icons with identifying marker Storage images are visualized through icons with identifying marker
Possibility to revise, simultaneously, two different static sequences on both “Cine Loop” function to revise sequences in continuous mode or frame by “Cine Loop” function to revise sequences in continuous mode or frame by
monitor screens. frame, this function is selectable according to user’s needs. frame, this function is selectable according to user’s needs.
Possibility to delete sequences or complete sessions. Possibility to delete sequences or complete sessions.
Possibility to revise, simultaneously, two different differ dynamic sequences on both Possibility to revise, simultaneously, two different dynamic sequences on both
monitor screens. monitor screens.

Cine Pulse DHD (Dynamic High Definition) digital Cine Pulse DHD (Dynamic High Definition) digital
module module
Digital dynamic images storage synchronized with generator pulses: Digital dynamic images storage synchronized with generator pulses:
Pulses rate: ......................................................12-25 pps Pulses rate: .....................................................
................................ 12-25 pps
range kV: ..........................................................40 – 120 kV range kV: .........................................................
................................ 40 – 120 kV
range mA: ........................................................from 30 up to 80 mA range mA: ........................................................
................................ from 30 up to 80 mA

Digital Vascular and Endovascular module


DSA (Digital Subtraction Angiography) modality, with:
ROAD MAPPING modality
Re-mask - Max Op - Land-mark - Pixel shift

MOONRAY COMPACT | REV 20/12/2019 3


GENERATOR & X-RAY CONFIGURATION - MOONRAY COMPACT

4 kW – Fixed Anode 12 kW – Rotating Anode 18 kW - Rotating Anode


Type: high frequency monotank Type: high frequency monotank Type: high frequency monotank
Nominal Power: ........................................... 4 kW Nominal Power: ........................................... 12 kW Nominal Power: ........................................... 18 kW
Power available in Radiography: ................ 3.2 kW. Power available in Radiography: ................ 8 kW. Power available in Radiography: ................ 8 kW.
DDC (Dynamic Density Compensation) DDC (Dynamic Density Compensation) DDC (Dynamic Density Compensation)
LDC (Low Dose Control) LDC (Low Dose Control) LDC (Low Dose Control)
Voltage range: ............................................. 40-120 kV Voltage range: ............................................. 40-120 kV Voltage range: ............................................. 40-120 kV
Current range: Current range: Current range:
Fluoroscopy pulsed: .................................... up to 20 mA Fluoroscopy pulsed: .................................... up to 20 mA Fluoroscopy pulsed: .................................... up to 20 mA
Boost pulsed …………………………………… 6 or 10 mA Boost pulsed ……………………………………6 or 10 mA Boost pulsed …………………………………… 6 or 10 mA
CINE PULSE DHD 12 pps and 25 pps: ...... up to 80 mA CINE PULSE DHD 12 pps and 25 pps:....... up to 80 mA CINE PULSE DHD 12 pps and 25 pps:....... up to 80 mA
Radiography not direct: ............................... 25 – 100 mA; 0,02 – 3 sec Radiography not direct: ............................... 25 – 100 mA; 0,02 – 3 sec Radiography not direct: ............................... 25 – 100 mA; 0,02 – 3 sec
Pulse rate in all pulsed modalities: .............. 0,5-1-2-4-6-12 – 25 pps Pulse rate in all pulsed modalities: .............. 0,5-1-2-4-6-12 – 25 pps Pulse rate in all pulsed modalities: .............. 0,5-1-2-4-6-12 – 25 pps
Fixed anode: ................................................ YES Rotating anode: ........................................... up to 3000 rpm Rotating anode: ........................................... up to 3000 rpm
Anode thermal capacity: .............................. 225 KJoule (300 KHU) Anode thermal capacity: .............................. 150 KJoule (200 KHU) Anode thermal capacity: .............................. 225 KJoule (300 KHU)
Anode thermal dissipation: .......................... 750 W (60 KHU/min) Anode thermal dissipation: .......................... 300 W (24 KHU/min) Anode thermal dissipation: .......................... 750 W (60 KHU/min)
Housing thermal capacity: ........................... 1.600 KJoule (2.150 KHU) Housing thermal capacity: ........................... 1.600 KJoule (2.150 KHU) Housing thermal capacity: ........................... 1.600 KJoule (2.150 KHU)
Housing thermal dissipation max with iLX:.. 300 W (24 KHU/min) Housing thermal dissipation with iLX plus: .. 420 (35 KHU/min) Housing thermal dissipation with iLX plus: .. 420 W (35 KHU/min)
Inherent filtration: ......................................... 0,8 mmAl @ 80 kV Inherent filtration: ......................................... 1,4 mmAl @ 80 kV Inherent filtration: ......................................... 0,7 mmAl @ 80 kV
Double focal spot: ........................................ 0,6 mm & 1,5 mm Double focal spot: ........................................ 0,3 mm & 0,6 mm Double focal spot: ........................................ 0,3 mm & 0,6 mm
Power on focal spot ..................................... 1,3 kW (spot 0.6); 4 kW (spot 1.5) Power on focal spot ..................................... 5 kW (spot 0.3); 17 kW (spot 0.6) Power on focal spot ..................................... 6 kW (spot 0.3); 25 kW (spot 0.6)

4 MOONRAY COMPACT | REV 20/12/2019


STANDARD FEATURES - MOONRAY COMPACT

Safety and Thermal Indicators Dose measurement and report:


Digital dynamic indicator of x-ray tube thermal level. ADR (Automatic Dose Report)
Digital dynamic indicator of monotank temperature. DAP (Calculation Software).
Fluoroscopy Timer with acoustic alarm and manual reset. Visualization of DAP for each image.
“X-ray on”: light and acoustic signal Total Exposure Report, visualized according to the different modalities of work
(pulsed fluoroscopy, radiography).
The report can be printed by the local medical printer (optional)
Collimators Digital dose report exportation available as TXT – PDF - MDB file
Iris diaphragm, motorized handling and digital pre-view.
Digital semi-transparent.
Environments conditions and power supply
Temperature: ...................................................+10°C / +40°C
Humidity: ..........................................................10% / 90%
Digital Memory Single phase, neutral ground, 220-240V ± 10%; 50/60Hz;
LIH (Last Image Hold) Single phase, neutral ground, 110-140V ± 10%; 50/60Hz
Storage on HD: ................................................ Automatic/Manual EN60309 current plug 16 A; out of isolation transformer.
Memory capacity on HD: ................................. 250.000 images
User-friendly database, divided into operative sessions Certifications
Several key words to search images The equipment is CE certified (CE0051), is in conformance with European
Storage images are visualized through icons. Directive 93/42 regarding medical devices and is in compliance with
Clipboard, to compare any stored image with live images D.L. n°187/00.
Possibility to erase single images or complete sessions
Others
Connection cable between c-arm and workstation trolley with double connector
Powder coating for easy cleaning and disinfection

MOONRAY COMPACT | REV 20/12/2019 5


Recommended Software
Additional Configuration
(To be requested when placing order)
CLINICAL APPLICATION
 DAP hardware
 Dissipation system iLX / iLX Plus Cardiology
 Integrated laser on Flat Detector (Electrophysiology)
 Monoblock integrated laser
 Interface for external “X-Ray ON” lamp Complementary Radiology
 Monitor B/W
Gastroenterology

General Surgery

Neuro
Neurosurgery

Ort
Orthopedics/Traumatology

Pain Therapy
Accessories Pneumology

Trauma Center
 DAP DATA Collection Software
 DICOM connection module:(STORAGE – PRINT – Urology
WORKLIST - QUERY & RETRIEVE - Master – DOSE
STRUCTURED REPORT) Vascular Surgery /
Vasc
 Digital medical thermal printer Endovascular
 Infrared remote control

STANDARD RECOMMENDED

SIMAD s.r.l. Via Benedetto Zallone 25 Tel: (+39) 051 686 08 11 C.F./P.I./Reg.Imp.di Bologna
Società Unipersonale 40066 Pieve di Cento Fax: (+39) 051 686 12 46 01854280367
Direzione e Bologna PEC:simad@legalmail.it REA di Bologna 433497
coordinamento Italy www.simad.net Iscr. RAEE IT09110000006357
di Sago Medica s.r.l. info@simad.net Cap. Soc. € 76.500,00 I.V.

SIMAD SRL reserves the right to make changes for the continuous improvement of all products without notice.
All brands, product names and logos appearing in this document are the property of their respective owners. © 2020 SIMAD S.r.l. All rights reserved..
reserved.

You might also like