You are on page 1of 19

System Specifications i2000 i2000SR c8x00 c16x00

The following system specifications are included in this section: Site Requirements for Operations Manual Use Approximate Physical Measurements Clearance Measurements for Operation and Service Electrical, Power, and Heat Output Specifications Water and Waste Specifications Seismic Anchoring Specifications Environmental Specifications Floor Loading Specifications Dimensional Layout Drawings

Operations Manual Use


Use Minimum System Requirements

A computer workstation should be available in the lab for installation and use of the ARCHITECT System Operations Manual CD-ROM. This workstation is in addition to the SCC. If an additional workstation is not available, a printed operations manual can be ordered.

Additional computer workstation for display of the operations manual

Windows 95 or Windows 2000 Operating System Pentium II Processor 640 Mb RAM 24x CD-ROM Drive 17" Color Monitor (800x600 resolution) Caution It is highly recommended that the customer perform a system backup prior to installation of the operations manual. Note This installation program installs Internet Explorer 5.0 over the current version of Internet Explorer. To ensure proper viewing, set the resolution of the monitor to 800x600.

Approximate Physical Measurements


System / Module Depth Width Height Weight

c16000 (with Retest Sample Handler not including SCC, or Waste Option) c8000 (with Retest Sample Handler not including SCC, or Waste Option) i2000SR (with Retest Sample Handler, not including SCC, ARCHITECT ARM, or Waste Option) ci16200 Integrated System (with Retest Sample Handler, not including SCC, ARCHITECT ARM, or Waste Option) ci8200 Integrated System (with Retest Sample Handler, not including SCC, ARCHITECT ARM, or Waste Option) SCC

49" (124.5 cm) 49" (124.5 cm) 49" (124.5 cm) 49" (124.5 cm) 49" (124.5 cm) 21" (53.3 cm) 17.8" (45.2 cm)

79" (200.6 cm) 79" (200.6 cm) 61" (154.9 cm) 127" (322.6 cm) 127" (322.6 cm) 15.5" (39.4 cm) 6.9" (17.5 cm)

48" (121.9 cm) 48" (121.9 cm) 48" (121.9 cm) 48" (121.9 cm) 48" (121.9 cm) 22" (55.9 cm) 9.1" (23.1 cm)

1545 lbs. (701 kg) 1425 lbs. (646.4 kg) 1081 lbs. (490.3 kg) 2679 lbs. (1215 kg) 2447 lbs. (1109.9 kg) 55 lbs. (25 kg) 43 lbs. (19.5 kg) 18 lbs. (8.2 kg) 69 lbs. (31.3 kg)

Uninterruptible Power Supply (UPS) (Optional - for the SCC) Waste Pump ARCHITECT ARM (Automatic Reconstitution Unit)

13" (33.0 13" (33.0 13" (33.0 cm) cm) cm) 17.8" (45.2 cm) 22.9" (58.2 cm) 28.3" (71.9 cm)

Clearance Measurements for Operation and Service


Device Location Clearance Description

Note See Dimensional Layout Drawings. Sides (All Modules) Rear (c8000/i2000SR /c16000) 24" (61.0 cm) 30" (76.2 cm) Left and right clearance (for access) Single Module System clearance between rear module covers and wall (for safety access to main circuit breakers, cooling, airflow, cable and tubing routing, and maintenance) Multi-Module System clearance between the rear module covers and wall (for safety access to main circuit breakers, traffic, and maintenance) Clearance above modules (to open top cover) Clearance in front of modules (for safety and to allow access for removal and replacement of bulk liquids/waste) The c8000 module requires 31" (78.7 cm) wide openings for doorways and entryways between the delivery area and the laboratory. Verify door width for each door from dock to lab is 31" (78.7 cm) or wider. Note For doors less than 31" (78.7 cm), removal of door and hinges may be required. For door openings less than 31" (78.7 cm) wide, an alternate method of getting the module to the lab needs to be found.

Processing Module (c8000, i2000SR, ci8200, c16000, ci16200 Integrated System)

Rear (ci8200/ci16000)

30" (76.2 cm)

Above (All Modules) Front (All Modules) Door and Entry Requirements (c8000) Doorways And Entryways

20" (50.8 cm) 34" (86.4 cm) 31" (78.7 cm)

Door and Entry Requirements (c16000)

Doorways And Entryways

34.5" (87.6 The c16000 module requires 34.5" cm) (87.6 cm) wide openings for doorways and entryways between the delivery area and the laboratory. Verify door width for each door from dock to lab is 34.5" (87.6 cm) or wider. Note For doors less than 34.5" (87.6 cm), removal of door and hinges may be required. For door openings less than 34.5" (87.6 cm) wide, an alternate method of getting the module to the lab needs to be found. 36" (91.4 The i2000SR module requires 36"

Door and Entry

Doorways And

Requirements (i2000SR)

Entryways

cm)

(91.4 cm) wide openings for doorways and entryways between the delivery area and the laboratory. Note For doors less than 36" (91.4 cm), but greater than 33.5" (85.1 cm), some disassembly of the module or removal of door & hinges may be required. For doors less than 33.5" (85.1 cm) wide, an alternate method of getting the module to the lab needs to be found. Clearance behind and on either side of SCC & Monitor (for cooling and airflow) Clearance above monitor (for cooling and airflow) Clearance in front of the SCC (for work area) Printer Parallel Cable Length Tubing and Fitting Clearance Access Panel Access Access/Airflow Tubing and Fittings Tubing and Fittings

Sides and Rear (All Modules) Above (All Modules) Front (All Modules) Printer Front/Rear Waste Pump Sides Front Rear ARCHITECT ARM Right Left

6" (15.2 cm) 24" (61.0 cm) 34" (86.4 cm) 6' (1.8 m) 9" (22.9 cm) 5" (12.7 cm) 20" (50.8 cm) 10" (25.4 cm) 12" (30.5 cm) 18" (45.7 cm)

SCC

Electrical, Power, and Heat Output Specifications


50Hz Outlet Specification 60Hz Outlet Specification Cabling / Power Outlets Estimated Heat Output

System

Outlet Type

1 - 220V Outlet (Not including SCC & Options)

c8000/c16000

US - A single 220 VAC 30 Amp Circuit is required when installing the UPS. For non-UPS installs, a 220 VAC, 20 Amp Circuit is required for each instrument. EUR - 220 VAC, 16 Amp Circuit (180-264 VAC @ 47 to 63 HZ) Resettable Circuit Breaker Note Requires dedicated/isolated line for each module.

US - NEMA L6-30R (30A 250V TwistLock) - This is required for the UPS being supplied. The instrument power cord is a NEMA L620R, which plugs into the UPS. EUR - IEC309 (220V - 16A 250V)

2,200 Watts

i2000/i2000SR

US - A single 220 VAC 30 Amp Circuit is required when installing the UPS. For non-UPS installs, a 220 VAC, 20 Amp Circuit is required for each instrument. EUR - 220 VAC, 16 Amp Circuit (176-264 VAC @ 47 to 63 HZ) Resettable Circuit Breaker Note Requires dedicated/isolated line for each module.

1 - 220V Outlet (Not including SCC & Options)

US - NEMA L6-30R (30A 250V TwistLock) - This is required for the UPS being supplied. The instrument power cord is a NEMA L620R, which plugs into the UPS. EUR - IEC309 (220V - 16A 250V)

2,500 Watts

ci8200/ci16200

1 - 220V Outlet (Not including SCC & Options) US - A single 30 Amp Circuit EUR - see c8000/c16000 and i2000/i2000SR

US - NEMA L6-30R (30A 250V TwistLock) This is required for the UPS being supplied. The instrument power cord is a NEMA L620R, which plugs into the UPS.

4,700 Watts

EUR - IEC309 (220V - 16A 250V) c8000/c16000 Power Cord Length i2000SR Power Cord Length Power Cord Length EUR - 220 US - 110 VAC, VAC, 12 Amp 14 Amp Circuit Circuit (90 - 132 VAC (180 - 284 VAC @ 47 - 63 HZ) @ 47 - 63 HZ) Resettable Resettable Circuit Breaker Circuit Breaker Country Specific 110 VAC, 14 Amp Circuit (90 - 132 VAC @ 47 - 63 HZ) Resettable Circuit Breaker 1 Outlet 15' (4.6 m) 6' (1.8 m) US - 110V - Included in above NEMA 5-15 3estimates prong EUR - 220V IEC 320 M grounded US only Included in above 110V - NEMA 5-15 3-prong estimates 13' (4.0 m)

SCC The E platform has an autosensing power supply and needs no switching. Modem

1 Power Outlet, 1 direct, analog, dedicated modem line (voice mail cannot be enabled)

Analog telephone line available at SCC location. Direct line (no switchboard interface) is required. Dedicated line is recommended, but is not required. Modem Phone Line Modem line does not route through an internal facility switchboard. Modem line does not have voice mail enabled. Options Power Cord Length Printer 110 VAC, 14 220 VAC, 12 110/220 Amp Circuit Amp Circuit Switchable (Do not plug the (180 - 264 VAC (90 - 132 VAC @ 47 - 63 HZ) @ 47 - 63 HZ) printer into the Resettable Resettable UPS.) Circuit Breaker Circuit Breaker UPS (110) (SCC Only) N/A 110 VAC, 14 Amp Circuit (90 - 132 VAC @ 47 - 63 HZ) Resettable Circuit Breaker 6' (1.8 m) 1 Outlet 110V - NEMA 5-15 3-prong 220V - IEC 320 M grounded 110V - NEMA 5-15 3-prong N/A

1 Outlet

N/A

UPS (220) (SCC Only) Waste Pump Switch must be manually set. (Plug the waste pump into the UPS, if one is provided.) ARCHITECT ARM Appropriate fuse (shipped with ARCHITECT ARM) must be installed

Contact ACS

Contact ACS

1 Outlet 1 Outlet

Contact ACS 110V - NEMA 5-15 3-prong 220V - IEC 320 M grounded

Contact ACS N/A

115 VAC, 1.6 220 VAC, 0.8 Amp Circuit Amp Circuit (207 - 253 VAC (104 - 126 VAC @ 47 - 63 HZ) @ 47 - 63 HZ) Resettable Resettable Circuit Breaker Circuit Breaker

1 Outlet (90 - 132 VAC (180 - 264 VAC @ 50/60 HZ)

US: NEMA 515 P EUR: CE E 7/7 or equivalent

N/A

Water and Waste Specifications


Minimum resistivity

Water Quality

1 MegOhm-cm @ 25 C (77 F) Note NCCLS Type II Reagent Grade water is acceptable. Caution ARCHITECT ARM shuts off if minimum resistivity is < 1 MegOhm.

Maximum Bacterial Content

1000 colony forming units per mL (cfu/mL)

Regulator and pressure gauge located prior to c8000 and c16000 water input. No brass fittings in the water input line. Note The fitting on the back of the c8000 and c16000 is not brass, it is a different alloy. Water pressure set to 15-25 psi at 15 C (59 F) to 37 C (98.6 F) Water Connection c8000/c16000 Water Flow Rate The flow rate is determined by the water pressure and the tubing size. If the regulator pressure, tubing and connectors meet specifications, the flow rate is acceptable. Average Consumption Pressure 8-100 psi Minimum Flow Rate Water (ARCHITECT ARM Installed) Input Temperature Water Connection 102L/hour (1.7 L/minute) at pressure > 15 psi, or 132L/hour (2.2 L/minute) at pressure 8-15 psi. 15 C (59 F) to 37 C (98.6 F) Provide a female connector to support a male connector 3/8" (9.5mm) OD Tube x 1/4"-1/8" (6.4-3.2 mm) NPT. 54 liters per hour (c16000) 25 liters per hour (c8000) Provide tubing and a female connector to support a male connector at the water input fitting of the analyzer. Tubing to be 3/8" (9.5 mm) OD hard tubing. Fitting to be 3/8" 1/4" NPT (9.5 - 6.35 mm) male thread (both ends) compression fitting.

Liquid Waste

Instrument drains are gravity flow, with barb fittings at approximately 8 inches (20.3 cm) above the floor. Average Output When Running

c16000

< 54L/hour c16000 Maximum Flow Capacity = 300 L/hour

c8000

25 L/hour c8000 Maximum Flow Capacity = 300 L/hour

i2000SR ARCHITECT ARM

5.5 L/hour 18.4 L/hour Maximum Flow Capacity with ARCHITECT ARM 132 L/hour (2.2L/minute)

c8000/c16000 Liquid Waste

Floor level drain (no more than 4" (10cm) high) for gravity fed waste tubes (2) is required. Drain must be located within 9.5 ft. (2.9 m) from the rear of the system and be able to handle maximum flow rate of 300 liters per hour. High concentration waste lines can pump to an uphill drain if needed. If drain is not available or not permitted, the optional high concentration waste container can be utilized. (High concentration waste collection/handling as per local requirements/regulations.)

Waste Pump Required

If floor drain is not available and elevated drain (sink) is used, a external Waste Pump is required. A Waste Pump is available from Abbott that has four (4) input lines and one output line. A waste pump is required for each module. Waste pump maximum outlet tubing guidelines: Length of 30 feet (9.1m) and height of up to 5 feet (1.5m) Length of 10 feet (3m) and height of up to 8 feet (2.4m)

Solid Waste (i2000SR)

Operational Output of 200 reaction vessels per hour. Waste Container Capacity sized for five hours of operation (over 1000 reaction vessels). Container can be changed during operation. With container removed, the Solid Waste Chute holds 50 reaction vessels (15 minutes of run time).

Seismic Anchoring Specifications

SCC - per site specific requirements Seismic Anchoring (optional, information provided for site specific needs) No tie down points are provided for the SCC. To secure the SCC, site personnel should follow internal procedures for securing business equipment.

Waste Pump - per site specific requirements No tie down points are provided for the Waste Pump. To secure the Waste Pump, site personnel should follow internal procedures for securing business equipment.

ARCHITECT ARM - per site specific requirements No tie down points are provided for the ARCHITECT ARM. To secure the ARCHITECT ARM, site personnel should follow internal procedures for securing business equipment.

c8000/c16000 - per site specific requirements No tie down points are provided for the c8000 or c16000. To secure the c8000/c16000, site personnel should follow internal procedures for securing business equipment.

i2000SR Module - Four (4) threaded holes suitable for bolts are available Four (4) anchoring points are used during shipping. Based on site approval, these points can be used to secure the assay module to the floor after the shipping pallet is removed. The four (4) holes are threaded for bolts and are located on the bottom of the module, approximately 2" (5.1 cm) inside the frame, at the midpoint of each side of the module. Site specific anchoring should be performed by site personnel and coordinated with the Abbott installation. Length of bolts used is site specific. The table below is provided to assist in determining hardware requirements.

Diameter 1/2 "

Threads per Inch Length

13 Determine the 2 site specific lengths and add the three lengths to obtain a total length:

Add +

1" 3"

2.5 cm 7.6 cm

Thread engagement into bottom of module. Approximate distance from floor to bottom of module at attachment point. (May be site specific and should be confirmed during installation.) (Site specific) Additional length if needed to go through anchoring platform Total length of bolt needed for anchoring.

+ =

__"

___ cm

___" ___ cm

Environmental Specifications

Operational Temperature and Humidity Storage Temperature Approved Operating Altitude

15 C to 30 C (59 F to 86 F) noncondensing relative humidity of 10% to 85%, at 25 C (77 F).

5 C to 50 C (41 F to 122 F)

6500 ft. (1981.2 m) or lower.

Lab Environment

Instrument is for Indoor Use Only. Avoid direct sunlight, drafts from heating and cooling vents, and excessive dust. Instrument should not be placed next to a heat or vibration generating device. c8000/c16000 requires that the floor be level within 0.5 cm/m. Acoustical noise output from the system: 70 dB peak output for a 10 second period, and less than 48 dB during normal operation.

Acoustical Noise

Floor Loading Specifications


Instrument Weight at each Foot Force at each Foot

Note See Dimensional Layout Drawings. Left Front = 482 lbs. (219 kg) [1] c16000 Right Front = 385 lbs. (175 kg) [2] Left Rear = 310.5 lbs. (140.8 kg) [3] Right Rear = 367.5 lbs. (166.7 kg) [4] c16000 total weight = 1545 lbs. (701 kg) Note See Dimensional Layout Drawings. Left Front = 320 lbs. (145.15 kg) [1] c8000 Right Front = 456 lbs. (206.84 kg) [2] Left Rear = 339 lbs. (153.77 kg) [3] Right Rear = 310 lbs. (140.61 kg) [4] 69.57 lbs/in2 (4.89 kg/cm2) 99.13lbs/in2 (6.97 kg/cm2) 73.70 lbs/in2 (5.18 kg/cm2) 67.39 lbs/in2 (4.74 kg/cm2) 105 lbs/in2 (7.38 kg/cm2) 84 lbs/in2 (5.91 kg/cm2) 68 lbs/in2 (4.78 kg/cm2) 80 lbs/in2 (5.62 kg/cm2)

c8000 total weight = 1425 lbs. (646.4 kg) Note See Dimensional Layout Drawings. Left Front = 331 lbs. (150.14 kg) [1] i2000SR Right Front = 265 lbs. (120.20 kg) [2] Left Rear = 195 lbs. (88.45 kg) [3] Right Rear = 290 lbs. (131.54 kg) [4] 105.36 lbs/in2 (7.41 kg/cm2) 84.35lbs/in2 (5.93 kg/cm2) 62.07 lbs/in2 (4.36 kg/cm2) 92.31 lbs/in2 (6.49 kg/cm2)

i2000SR total weight = 1081 lbs. (490.3 kg) Note See Dimensional Layout Drawings. Left Front = 407 lbs. (184.61 kg) [1] Center Front = 552 lbs. (250.38 kg) [2] Right Front = 418 lbs. (189.60 kg) [3] Left Rear = 305 lbs. (138.35 kg) [4] Center Rear = 535 lbs. (242.67 kg) [5] Right Rear = 230 lbs. (104.33 kg) [6] 88.48 lbs/in2 (6.22 kg/cm2) 71.32 lbs/in2 (5.01 kg/cm2) 133.12lbs/in2 (9.36 kg/cm2) 66.30 lbs/in2 (4.66 kg/cm2) 69.12 lbs/in2 (4.86 kg/cm2) 73.25 lbs/in2 (5.15 kg/cm2)

ci8200

ci 8200 total weight = 2447 lbs. (1109.9 kg) Note See Dimensional Layout Drawings. Left Front = 534 lbs. (242 kg) [1] Center Front = 700 lbs. (318 kg) [2] ci16200 Right Front = 245 lbs. (111 kg) [3] Left Rear = 331 lbs. (150 kg) [4] Center Rear = 519 lbs. (235 kg) [5] Right Rear = 350 lbs. (159 kg) [6] 116.1 lbs/in2 (8.16 kg/cm2) 158.4 lbs/in2 (11.14 kg/cm2) 78 lbs/in2 (5.48 kg/cm2) 72 lbs/in2 (5.06 kg/cm2) 117.4 lbs/in2 (8.25 kg/cm2) 111.4 lbs/in2 (7.83 kg/cm2)

ci16200 total weight = 2679lbs. (1215 kg)

Dimensional Layout Drawings


c8000/c16000 with Retest Sample Handler (RSH)

i2000SR

c8000 or c16000 and i2000SR