You are on page 1of 28

Gas Chromatography – Harris

Chapter 24
Gaseous Mobile Phase, Solid or
Liquid Stationary Phases

11/22/2005 Chem 253 - Chapter 24 1

Schematic & Major Components (fig


24-1)

11/22/2005 Chem 253 - Chapter 24 2

1
Carrier
Gas
Reservoir
Gas Flow
Make-up Gas
(optional)

Injection Port

Sample
Introduction

Separation
column

Injection Port Oven Detector Oven


Temperature T1 Temperature T3
Column Oven
Temperature T2

11/22/2005 Chem 253 - Chapter 24 3

Summary of GC components.
z Injection port – Syringe containing sample is introduced through
a septum, injection port oven temperature heated to temperatures
that ensures fast volatilization of sample, i.e. above the b.p. of all
sample components, usually 275 0C. Keep in mind these terms:
split and splitess injection

z Carrier Gas – Sweeps analyte to separation column. Usually, He,


or N2, sometimes H2.

z Column – Either packed (old) or capillary (newer) columns. Each


type requires its own set of plumbing components (threads, etc.)
Interestingly all GC and HPLC plumbing components in the
English system of measurements.

11/22/2005 Chem 253 - Chapter 24 4

2
Summary of GC components.
z Column T – May be below b.p. of sample components, but high
enough to keep a significant quantity in the gas phase.
Temperature may be ramped up to get separation based on b.p.
differences of components along with chromatographic
separation.

z Makeup Gas – Sometimes required since detector volume is too


large for carrier gas flow. Remember that makeup gas is usually
required for capillary columns.

z Detector – Ideally, want universal detector, sensitive to anything


that passes by. Wide dynamic range, low D.L. the usual wants.
We want to keep all sample components volatilized, detector
oven temperature usually 300 0C.

11/22/2005 Chem 253 - Chapter 24 5

Effects of the Column and Mobile


Phase Flow in GC.
z Why are modern GC based on a capillary column?

z Back to the van Deemter Eqn. H = A + B/u + Cu

z Remember, u = flow rate,


z A = multiple paths
z B/u = longitudinal diffusion effects
z Cu = MT effects

z We want to minimize H as much as possible.

11/22/2005 Chem 253 - Chapter 24 6

3
11/22/2005 Chem 253 - Chapter 24 7

11/22/2005 Chem 253 - Chapter 24 8

4
11/22/2005 Chem 253 - Chapter 24 9

“Older” packed columns

z usually 1/8” (3.2 mm OD, 2.2 mm I.D.)


diameter, 1 – 2 m length
z design impedes gas flow
z max flow rate about 1 mL/min or 8
cm/min.

11/22/2005 Chem 253 - Chapter 24 10

5
“Newer’ capillary columns
z 0.25 mm inner diameter.
z major point, less restricted flow path.
z flow rate about 20 ml/min or 20 cm/min

11/22/2005 Chem 253 - Chapter 24 11

We want to minimize H as much as


possible.
z H = A + B/u + Cu

z Which of the above affects H the most in GC?


• B/u effects!
z Diffusion coefficients are large in the gas phase.
σ = 2 D uL
z Simply increasing the flow rate partially
addresses the B/u effects in GC

11/22/2005 Chem 253 - Chapter 24 12

6
Other major factors that influence the performance of
capillary columns relative to packed ones are evident
in the table below.
Typical 1/8 “ Typical Capillary Comments
packed
I.D. 2.2 mm 0.25 mm No packing material less
0.1-0.53 mm restricted flow
df 5 µm 0.25 µm MT for s.p. part of Cu
f (k ' )d 2f
Cs α
Ds

L 1-2 m 10-60 m HN = L
N 4,000 180,000

Advantages •Lower cost •Higher sep.


•Easy to make efficiency
•Larger •Faster sep.
samples •Better for complex
mixtures

11/22/2005 Chem 253 - Chapter 24 13

Other advantages of tubular over packed GC


columns:

z No “A” term, straight path.

11/22/2005 Chem 253 - Chapter 24 14

7
Example of the separation efficiency of capillary vs.
packed column.

11/22/2005 Chem 253 - Chapter 24 15

11/22/2005 Chem 253 - Chapter 24 16

8
GC stationary phases.
z Older packed columns – uniform silica particles (150-250 µm)
required to ensure uniform path lengths (the “A” term in the van
Deemter eqn.) Surfaces are chemically modified (see below). The
columns themselves were either glass or stainless steel.

z Capillary columns – fused silica which like the particles in the


packed column require chemical modification (below).The
stationary phase surface (silica) is a hydroylated surface. This
caused problems with nonpolar stationary phases.

z When polar or mildly polar species partition into the s.p. they
stand a good chance of being trapped, causing a excessive band-
broadening beyond what is expected from van Deemter
considerations.

11/22/2005 Chem 253 - Chapter 24 17

Trapped polar solute


Mobile Phase

Stationary Phase

OH OH OH OH OH OH OH OH

Silica Support

11/22/2005 Chem 253 - Chapter 24 18

9
Surface modification with trimethylchlorosilane
reduces surface polarity.

CH3
H3C CH3
Si
Cl
CH3
+ H3C CH3
Si
OH OH OH O + HCl
O Si O Si O Si O Si O
O O O O

11/22/2005 Chem 253 - Chapter 24 19

Effects can be dramatic:


z A = EtOH
z B = methylethyl
ketone
z C = benzene
z D = cyclohexane

z Chromatogram 1
= unmodified

z Chromatogram 2
= silinazied

11/22/2005 Chem 253 - Chapter 24 20

10
Liquid stationary phases

11/22/2005 Chem 253 - Chapter 24 21

Important Notes:

• All columns have a Tmax, never exceed this!


Thermal degradation, or s.p. simply boils
off.

• Most s.p. are air sensitive and easily oxidize


at high temperatures.

• Capillary columns cost $150-600 (2002


dollars)

11/22/2005 Chem 253 - Chapter 24 22

11
Effect of Stationary Phase Polarity on
Retention Times.

z #1 n-heptane
z #2 tetrahydrofuran
z #3 2-butanone
z #4 n-propanol

z We can reason this order by


considering the polarity of the
analytes and of the s.p.

z In GC the s.p. characteristics


have the largest effect on the
order of elution.

11/22/2005 Chem 253 - Chapter 24 23

Column Temperature and


Temperature Programming.
z The oven T must be controlled to within ±0.5
0C from RT to 400 0C. A fan circulates air
throughout the oven ensuring uniform heating.

z Constant temperature isothermal separations


are good for simple mixtures.

z Typically T programming is required. Slowly


ramping T throughout the separation provides
a basis for the separation of sample
components based on BP.

11/22/2005 Chem 253 - Chapter 24 24

12
Isothermal vs. Temp Programming
z The more
volatile cpds
elute very
close together
under
isothermal
conditions
(upper).

z Ramping T
from 50 to 250
oC at 8 oC/min
separates out
those volatile
cpds (lower).

11/22/2005 Chem 253 - Chapter 24 25

How does T programming work?

• T programming is an example of the control of the capacity


t − tm
factor: k A = t
' r
which in turn controls the resolution term:
m

⎛ α − 1 ⎞⎛ k B ⎞
'
N
Rs = ⎜ ⎟⎜⎜ ⎟
' ⎟
4 ⎝ α ⎠⎝ 1 + k B ⎠

• Note that T programming does not change the order of


elution.

11/22/2005 Chem 253 - Chapter 24 26

13
Carrier Gases
• O2 is usually avoided since it will oxidize the s.p.

• 3 most common gases N2, H2, He.

• Each have very different Hmin characteristics.

• The lighter gases He and H2 require faster analysis flow rates


20-50 cm/min.

• Diffusion in He and H2 causes more band-broadening in m.p.


• Greater MT effects decrease band-broadening at faster flow
rates
• H = A + B/u + Cu

• N2 Hmin occurs at about 10 cm/min.

11/22/2005 Chem 253 - Chapter 24 27

11/22/2005 Chem 253 - Chapter 24 28

14
Sample Injection

11/22/2005 Chem 253 - Chapter 24 29

Injection Port for Modern GC’s

11/22/2005 Chem 253 - Chapter 24 30

15
GC Injection Syringe
z It is important to rapidly vaporize the sample.

z Slow vaporization increases band broadening, by


increasing the sample “plug”.
z
Injection port temperature is usually held 50 0C higher
than the BP of the least volatile cpd.

11/22/2005 Chem 253 - Chapter 24 31

GC injection and band broadening


and anomalies.
z Manual injections – takes practice and
patience. Extremely slow injections will cause
band-broadening, wide sample “plug”. Jerky
injections may cause double peaks for the
same analyte species.

z Auto-injectors – robotized carousels of sample


vials. No practice needed. Can be set to make
measurements all night long. Rare in
academe, frequent in industry.

11/22/2005 Chem 253 - Chapter 24 32

16
11/22/2005 Chem 253 - Chapter 24 33

11/22/2005 Chem 253 - Chapter 24 34

17
11/22/2005 Chem 253 - Chapter 24 35

Split vs. Splitless Injection


z Sample injection is done by a syringe – 1 to 5 µL or ng’s of analyte for
the average capillary column.

z Capillary columns usually require split in injections, a sample reduction


method.

z Depending on the spilt ratio (adjustable) only 0.2 to 2 % of the sample


injection makes its way to the column. The rest is discarded.

11/22/2005 Chem 253 - Chapter 24 36

18
Effect of split versus splitless
injections on separation quality
z The chromatogram
(upper left)
demonstrates ideal
separation
characteristics. Upper
right illustrates what
happens when too
much analyte solute is
injected onto a
capillary column.

11/22/2005 Chem 253 - Chapter 24 37

Solid-Phase Microextraction
z Analyte pre-
concentration method,
remember stripping
voltammetry.

z Typical derivatizing
agents:
C18COO-Si-(O)3-

11/22/2005 Chem 253 - Chapter 24 38

19
11/22/2005 Chem 253 - Chapter 24 39

SPME of Nerve Agents


z Fig 24-21.

Advantages of SPME

“Green” method compared


to solvent extraction

Fibers are reusable 20-30x

102-103 LOD enhancement

Disadvantage – lot’s of
technique and method
development.

11/22/2005 Chem 253 - Chapter 24 40

20
Purge and Trap
Major Goal: Remove all analyte from sample.

Volatile components are removed from


sample by gentle heating and a stream of
purge gas, e.g. He.

Adsorbents trap targeted analytes, e.g.


molecular sieves.

After complete purge of sample, contents of


trap tube are injected into GC.

A purge and trap module is on the Cassini


probe to Saturn and moons looking for
hydrocarbons, H2, He, NH3.
http://saturn.jpl.nasa.gov/home/index.cfm

11/22/2005 Chem 253 - Chapter 24 41

GC Detectors
z Ideal detector characteristics, for flowing systems (e.g. GC)

 large dynamic/linear range


 response independent of flow rate, i.e. fast response times.
 Universal detection, responds to all species.

z Keep this in mind when we discuss HPLC and CE.

z Additional requirements for GC

 operation from RT to 400 oC


 detector response independent of detector oven T

11/22/2005 Chem 253 - Chapter 24 42

21
Thermal Conductivity Detector-TCD
z Measures heat loss from a hot filament – nearly universal

z filament heated to const T

 when only carrier gas flows heat loss to metal block is constant, filament T remains
constant

 when an analyte species flows past the filament generally thermal conductivity goes
down, T of filament will rise. (resistance of the filament will rise).

11/22/2005 Chem 253 - Chapter 24 43

Flame Ionization Detector (FID).


z Sensitive towards organics

z Analyte is burned in H2/air, which produces CH and CHO+,


radicals, remember our discussion regarding the blue cone in AA.

 CHO+ radicals are reduced at a cathode which produces a current


proportional to the radical quantity. About 10-12 A

 Specific for organic carbon, insensitive to inorganics, CO2, SO2 etc.

 Generally DL 100x less than TCD about pg/s (flow rate dependent)

 Dynamic range of 107

 Response to specific organic depends on the number of organic


carbons.

11/22/2005 Chem 253 - Chapter 24 44

22
11/22/2005 Chem 253 - Chapter 24 45

Electron Capture Detector (ECD)


z Sensitive to electron withdrawing groups especially towards
organics conatining –F, -Cl, -Br, -I also, -CN, NO2

z Nickel-63 source emits energetic electrons collides with N2


(introduced as make-up gas or can be used as carrier gas)
producing more electrons:

Ni-63 => e-
e- + N2 => 2e- + N2+

z The result is a constant current that is detected by the electron


collector (anode).

11/22/2005 Chem 253 - Chapter 24 46

23
z As an analyte flows through
past the Ni-63 source
electron capture is possible
by electron-withdrawing
species:

z A + e- => A-

z Current decreases as a
result of e- capture by
analyte. This is one of the
few instances in which a
signal is produced by a
decrease in detectable
phenomenon.

11/22/2005 Chem 253 - Chapter 24 47

ECD Characrtersitics (the good)


 VERY low DL for detected species 10-15g/s for many halogenated substances (PCB,
DDT etc).

 OK dynamic range of 104.

z (The bad)

 Radioactive Ni-63 source

 EASILY contaminated with O2, H2O, sample overloading. High maintenance device.

 Highly variable response to halogenated substances, see table below:

 Can be a real headache when method developing a specific analysis, e.g. CH2Cl2 in the
presence of CCl4.

z Sometimes complementary information from FID helps.

11/22/2005 Chem 253 - Chapter 24 48

24
11/22/2005 Chem 253 - Chapter 24 49

Other GC detectors
z Nitrogen-Phosphorous Detector (NPD)

Also know as the thermionic detector (TID) or alkali flame detector. It


is an FID tweaked for N-P cpds, and organics.

z Flame Photometric Detector (FPD)

FID tweaked for S containing cpds.

z Photoionization Detector (PID)

UV ionization of organic analyte, coupled with high voltage cathode


and analode results in current proportional to ionized organics.

11/22/2005 Chem 253 - Chapter 24 50

25
z FT-IR we’ve discussed before. (because of generally low ε in A
=εbc, most analytes have a modest DL with FT-IR)

z Mass Spec

11/22/2005 Chem 253 - Chapter 24 51

Qualitative Analysis in GC.

z GC-MS offers structural determinations


(remember organic?)

z With other detectors identification is


possible with retention times of analyte
and standard, however it’s best if
another method is used as a
confirmation.

11/22/2005 Chem 253 - Chapter 24 52

26
Quantitative Analysis in GC.

z Calibration curve

z Standard Addition – generally matrix effects


are not as pronounced as in GC as I have
personally noticed in LC.

z Internal Standard – best used with CC and


SA methods. Almost always required since
sample injection is not always a reproducible
phenomenon, even with auto-samplers.

11/22/2005 Chem 253 - Chapter 24 53

Method of Internal Standards


z Internal standards are
often used in
chromatographic
methods, but useful in Analyte, X
other techniques where
sample sizes may vary IS, S
from rune to run

z Known quantities of both Signal


non-interfering species
and analyte species are
subjected to the same
analytical procedure and
the signal is measured:

[X] = 0.0837 mM Independent variable


[S] = 0.0666 mM

11/22/2005 Chem 253 - Chapter 24 54

27
F-ratio
The signal of the IS is 347, and that of the analyte is 423.

We now can define a response factor F, defined as

AX A 423 347
=F S = =F
[X ] [S ] 0.0837 0.0666

Solving for F = 0.9700

We should realize that F will not vary even if sample sizes may
do so.

11/22/2005 Chem 253 - Chapter 24 55

Example - Now consider an unknown, if 10.0 ml of 0.146 M IS


is added to 10.0 ml of unknown then diluted to 25.0 ml, what is
the unknown analyte concentration is As = 553 and Ax is 582?

[S] = 0.146 M (10.0 ml/25.0 ml) = 0.0584 M

553 582
= 0.9700
[X ] 0.0584

[X] = 0.05721 M

[X]sample = 0.05721 M (25.0 ml/10.0 ml) = 0.143 M

11/22/2005 Chem 253 - Chapter 24 56

28

You might also like