You are on page 1of 37

Calibration Fundamentals

Biometrix Corporation (800)-890-8909

Objectives

Understand why we calibrate Understand 21 CFR Part 820 Introduce the basis for the basic SI units Explain the standards used in the calibration chain. Explain uncertainty and the 4:1 ratio.

Biometrix Corporation (800)-890-8909

Metrology

What is Metrology? The science that deals with measurement What does it mean to calibrate? To check, adjust, or determine display accuracy by comparison with a standard (the graduations of a quantitative measuring instrument) To make corrections in; adjust a set of operations that establish, under specified conditions, the relationship between the values of quantities indicated by a measuring instrument or measuring system and the corresponding values realized by standards
Biometrix Corporation (800)-890-8909 3

Metrology

Why Calibrate?

The quality of drug product is dependent upon certain conditions of manufacture, including temperature, pressure, and pH. In order to ensure consistent and correct drug manufacture, instrumentation relating to the drug production must be calibrated. In order to accurately describe the processes by which drugs are created, and gauge the effect of any deviations on the drug process

Biometrix Corporation (800)-890-8909

Metrology

..and because the government requires it. 21 CFR Part 820.72 governs Inspection, Measuring and Test Equipment.

Biometrix Corporation (800)-890-8909

21 CFR Part 820

(a) Control of inspection, measuring, and test equipment. Each manufacturer shall ensure that all inspection, measuring, and test equipment, including mechanical, automated, or electronic inspection and test equipment, is suitable for its intended purposes and is capable of producing valid results. Each manufacturer shall establish and maintain procedures to ensure that equipment is routinely calibrated, inspected, checked, and maintained. The procedures shall include provisions for handling, preservation, and storage of equipment, so that its accuracy and fitness for use are maintained. These activities shall be documented.

Biometrix Corporation (800)-890-8909

21 CFR Part 820

Calibration. Calibration procedures shall include specific directions and limits for accuracy and precision. When accuracy and precision limits are not met, there shall be provisions for remedial action to reestablish the limits and to evaluate whether there was any adverse effect on the device`s quality. These activities shall be documented.

Biometrix Corporation (800)-890-8909

21 CFR Part 820

Calibration standards. Calibration standards used for inspection, measuring, and test equipment shall be traceable to national or international standards. If national or international standards are not practical or available, the manufacturer shall use an independent reproducible standard. If no applicable standard exists, the manufacturer shall establish and maintain an in-house standard.
Biometrix Corporation (800)-890-8909 8

21 CFR Part 820

Calibration records. The equipment identification, calibration dates, the individual performing each calibration, and the next calibration date shall be documented. These records shall be displayed on or near each piece of equipment or shall be readily available to the personnel using such equipment and to the individuals responsible for calibrating the equipment.

Biometrix Corporation (800)-890-8909

N.I.S.T.

As a non-regulatory agency of the United States Department of Commerces Technology Administration, the National Institute of Standards (NIST) develops and promotes measurement, standards, and technology to enhance productivity, facilitate trade, and improve the quality of life. As part of this mission, NIST scientists and engineers continually refine the science of measurement, making possible the ultraprecise engineering and manufacturing required for todays most advanced technologies. They also are directly involved in standards development and testing done by the private sector and government agencies. NIST was originally called the National Bureau of Standards (NBS), a name that it had from 1901 until 1988

Biometrix Corporation (800)-890-8909

10

Definitions

Unit (of measurement)

Particular quantity, defined and adopted by convention, with which other quantities of the same kind are compared in order to express their magnitudes relative to that quantity. Magnitude of a particular quantity generally expressed as a unit of measurement multiplied by a number.

Value (of a quantity)

Biometrix Corporation (800)-890-8909

11

Definitions

Measurement Set of operations having the object of determining a value of a quantity. Error (of measurement)

Result of a measurement minus a true value of the measured.

Biometrix Corporation (800)-890-8909

12

Definitions

Accuracy
Conformity to fact. Precision; exactness. The ability of a measurement to match the actual value of the quantity being measured

Biometrix Corporation (800)-890-8909

13

Definitions

Traceability

Property of the result of a measurement or the value of a standard whereby it can be related to stated references, usually national or international standards, through an unbroken chain of comparisons all having stated uncertainties. Includes conductivity solutions, pH solutions, viscosity solutions, and turbidity solutions.

Biometrix Corporation (800)-890-8909

14

Definitions

Standard Reference material (SRM)

Reference material, accompanied by a certificate, one or more of whose property values are certified by a procedure which establishes traceability to an accurate realization of the unit in which the property values are expressed, and for which each certified value is accompanied by an uncertainty at a stated level of confidence.

Biometrix Corporation (800)-890-8909

15

Definitions

Reference Material (RM)

Material or substance one or more of whose property values are sufficiently homogeneous and well established to be used for the calibration of an apparatus, the assessment of a measurement method, or for assigning values to materials. A reference material may be in the form of a pure or mixed gas, liquid or solid. Examples are water for the calibration of viscometers, sapphire as a heat-capacity calibrant in calorimetry, and solutions used for calibration in chemical analyses.

Biometrix Corporation (800)-890-8909

16

Definitions

Working Standard

Standard that is used routinely to calibrate or check material measures, measuring instruments or reference materials. A working standard is usually calibrated against a reference standard.

Biometrix Corporation (800)-890-8909

17

Definitions

Reference Standard

Standard, generally having the highest metrological quality at a given location or in a given organization, from which measurements made there are derived.

Biometrix Corporation (800)-890-8909

18

Definitions

Primary Standard

Standard that is designated or widely acknowledged as having the highest metrological qualities and whose value is accepted without reference to other standards of the same quantity.

Biometrix Corporation (800)-890-8909

19

Definitions

Resolution

A measure of the smallest portion of the signal that can be observed. For example, a thermometer with a display that reads to three decimal places would have a resolution of 0.001C. In general, the resolution of an instrument has a better rating than its accuracy.

Biometrix Corporation (800)-890-8909

20

Definitions

Repeatability

closeness of the agreement between the results of successive measurements of the measurand (eg mass value) carried out under the same conditions of measurement. Conditions include: same procedure, observer, instrument, conditions, location; and carried out over a short period of time.

Biometrix Corporation (800)-890-8909

21

Definitions

Reproducibility

closeness of the agreement between the results of measurements of the measurand (eg mass value) carried out under changed conditions of measurement. Includes changing some of those conditions which are held constant for repeatability, and may refer to measurements carried out over a long period of time.

Biometrix Corporation (800)-890-8909

22

Definitions

Uncertainty

parameter, associated with the result of a measurement, that characterizes the dispersion of values that could reasonably be attributed to the measurand

Biometrix Corporation (800)-890-8909

23

Definitions

Standard

Material measure, measuring instrument, reference material or measuring system intended to define, realize, conserve or reproduce a unit or one or more values of a quantity to serve as a reference. Examples

kg mass standard 100 standard resistor standard ammeter cesium frequency standard

Biometrix Corporation (800)-890-8909

24

International System of Units (SI)

The SI is founded on seven SI base units for seven base quantities assumed to be mutually independent. Length Mass Time Temperature Quantity of Substance Luminous Intensity Electric Current

Biometrix Corporation (800)-890-8909

25

Length

Meter

The meter is the length of the path traveled by light in vacuum during a time interval of 1/299 792 458 of a second. Symbol = m It used to be one ten millionth of the distance of the meridian through Paris.

Biometrix Corporation (800)-890-8909

26

Mass

Kilogram

The kilogram is the unit of mass; it is equal to the mass of the international prototype of the kilogram. The international prototype is a large chunk of platinum-iridium in Paris. The international prototype will soon be a sphere of pure Silicon with an exactly known number of atoms Symbol = kg
Biometrix Corporation (800)-890-8909 27

Time

Second

The second is the duration of 9 192 631 770 periods of the radiation corresponding to the transition between the two hyperfine levels of the ground state of the cesium 133 atom. Symbol = s It used to be 1/86400 of the astronomical day.

Biometrix Corporation (800)-890-8909

28

Temperature

Kelvin

The Kelvin, unit of thermodynamic temperature, is the fraction 1/273.16 of the thermodynamic temperature of the triple point of water. What is the triple point?

The temperature and pressure at which solid, liquid, and gaseous water coexist in equilibrium.

Symbol = K

Biometrix Corporation (800)-890-8909

29

Electric Current

Ampere

The ampere is that constant current which, if maintained in two straight parallel conductors of infinite length, of negligible circular cross-section, and placed 1 meter apart in vacuum, would produce between these conductors a force equal to 2 x 10-7 Newton per meter of length.

Biometrix Corporation (800)-890-8909

30

SI Derived Units

Examples of SI derived Area square meter Volume cubic meter Speed velocity meter per second Acceleration meter per second squared Wave number reciprocal meter Mass density kilogram per cubic meter Specific volume cubic meter per kilogram Current density ampere per square meter Magnetic field strength ampere per meter Concentration mole per cubic meter Luminance candela per square meter Mass fraction kilogram per kilogram
Biometrix Corporation (800)-890-8909

m2 m3 m/s m/s2 m-1 kg/m3 m3/kg A/m2 A/m mol/m3 cd/m2 kg/km
31

CGMP Calibrations

CGMP Calibration must be traceable through all vendors and paperwork to the original NIST calibration

Example: NIST calibrates primary standard for primary standards lab Primary Standards Lab calibrates Intelligent RTD for Biotech Manufacturing Facility Biotech Manufacturing Facility calibrates HART temperature blocks with IRTD Calibration Group calibrates bioreactor instrumentation using the HART temperature blocks.

Biometrix Corporation (800)-890-8909

32

4:1 Ratio

When using one instrument or standard to calibrate another, the ratio of accuracies should be maintained at no lower than a 4:1 ratio. Example

Using an IRTD with an accuracy of +/- 0.01 K to calibrate a HART Block with an accuracy of +/0.05 K.

Biometrix Corporation (800)-890-8909

33

Uncertainty

To calculate the uncertainty of the overall test system, the uncertainties of the individual components are combined using a statistical approach. Each component is represented by a standard deviation (standard uncertainty). The deviations are combined under the rule of Propagation of Uncertainty. This rule is mathematically expressed as the root sum squared. The square root of the sum of the squares. When combined they are known as the

Combined Standard Uncertainty

Biometrix Corporation (800)-890-8909

34

Combined Standard Uncertainty

Example

A pressure measurement system contains a pressure transducer, ma transmitter, and PLC Analog Input. The uncertainty of the pressure transducer is +/- 0.5% The uncertainty of the ma transmitter is +/-0.2% The uncertainty of the PLC Analog Input is +/- 0.1% sqroot((0.5%)2 + (0.2%)2 + (0.1%)2 ) Combined Standard Uncertainty = 0.547%

Biometrix Corporation (800)-890-8909

35

Expanded Uncertainty

The expanded uncertainty, suggested symbol U, is obtained by multiplying the combined standard uncertainty by a coverage factor, suggested symbol k, which typically has a value between 2 and 3 (i.e., U = kuc) For a normal distribution and k = 2 or 3, the expanded uncertainty defines an interval having a level of confidence of 95.45% or 99.73%, respectively.

Biometrix Corporation (800)-890-8909

36

Expanded Uncertainty

If we apply the 4:1 ratio on top of the expanded uncertainty, the maximum specification for accuracy with a 99.73% confidence level would be..

0.547% Combined Standard Uncertainty K Factor of 3 for 99.73& confidence 4:1 ratio 4 x 3 x 0.547% = +/- 6.564 % Luckily most pressure transducers are significantly more accurate.

Biometrix Corporation (800)-890-8909

37

You might also like