Professional Documents
Culture Documents
II. Cleaning and Calibration Guidelines for Automatic Pipettes (Rainin, Gilson,
MLA and Pipetman).
New pipets or pipets which have just been returned from renovation at the factory
do not need lubricated but must be calibrated prior to being used. All
maintenance and calibrations must be recorded. Each pipet has a serial no. etched
on the barrel of the pipet to be used for identification. If a pipet is cleaned, its
calibration must be checked before it is placed in use.
III. Frequency:
A. Before newly purchased pipettes are placed into service
B. Every six months at a minimum.
C. Anytime a pipette becomes dismantled and repaired
D. As a means of troubleshooting assay problems
IV. Equipment
1. Pipette
2. Pipette QC log sheet
3. Pipette tips for appropriate volumes and pipettes
4. Weigh boats
5. Water, reagent grade, room temperature
6. Centrigrade thermometer
7. A temperature vs. density diagram
8. Mettler-Toledo Analytical Balance
9. Calculator
10. Kim Wipes
11. Silicone Grease
Pipette Calibration Check________________________________________________2
2. Clean outer pipette body and nose piece with an alcohol wipe and
kimwipe. Use a small piece of cotton on applicator stick to clean
out barrel. Be very careful not to lose the cotton plug.
3. Record the pipette serial number on the pipette QC log sheet for
the pipet that was cleaned.
Note: The manufacturer does not permit the Pipetman Pipet to be taken
apart and cleaned. We may only wipe it down with alcohol or send
it back to the Rainin Company for repairs.
c. If the above steps fail, take the pipette out of service until it
can be repaired and rechecked & may need returned to the
company for repairs
For Plastic Disposable S/P Pipettes used to prepare reagents or solutions where the
amount added needs to be precise and accurate, pipettes are checked semi-annually to
ensure they deliver +/- 3% of the stated volume.
Follow above steps 2 – 10. If the percent tolerance for each of the three designated
volumes is +/-3%, the disposable pipette is considered to be within acceptable calibration.
If the % tolerance is outside the range for any of the three measurements, a new set of
weighings should be made and repeat steps 2-9 above.
If the above steps fail, please contact a Lead or Supervisor for assistance to determine
appropriate corrective action.
A. Delivery Techniques
Pipette Calibration Check________________________________________________4
There are different techniques for filling and discharging automatic pipets
(i.e. MLA).
Note: If the plunger on an automatic pipet does not move smoothly, the accuracy
and precision may suffer. Bring this to the attention of the Lead
Technologist. The pipet should be cleaned, lubricated and recalibrated. If
the pipet is not repairable, send for service.
1. Operating Procedure
a. Set the desired volume by holding the PIPETMAN body in
one hand and turning the volume adjustment knob until the
correct volume shows on the digital indicator. The friction
o-ring of the volume adjustment knob “cold flows” in order
to lock the volume securely. For best volume setting
precision, always approach the desired volume by dialing
downward (at least one third revolution) from a larger
setting volume.
If an air bubble is noted within the tip during intake, dispense the sample
to original vessel, check tip immersion depth, and pipet more slowly. If an
air bubble appears a second time, discard the tip and use a new one to
repeat.
Prevent liquids from being drawn into the PIPETMAN shaft to prevent
possible resultant damage by observing the following precautions:
a. Pipet very slowly, holding the PIPETMAN not greater than
20o from the vertical axis.
b. Never invert or lay the PIPETMAN on its side with liquid
in the tip.
c. Use the special Safety Filters supplied with the P5000 &
P1000 instruments and C-5000 & C-1000 tips.
D. Service Information
1. Pipetman Pipet
References:
1. Eppendorf Pipette Instruction Manual
2. FACService Tech Notes from Becton Dickinson
3. Instructions for Pipetman Continuously Adjustable Digital Microliter Pipets, Rainin
Instruments Company, Inc., 1982.
4. Instructions for MLA pipets; MLA Pipettes, 270 Marble Ave. Pleasantville, NY. 10570
Pipette Calibration Check________________________________________________7.