You are on page 1of 7

Wireless Heart Attack Detector with GPS

Project Proposal
September 15, 2004 Kuangwei Hwang Diane Limsui Lianhan Zhao

Introduction
Overview Coronary artery disease, the restriction of blood flow to the heart, is the leading cause of death in the United States. The chances of suffering a myocardial infarction are great and increase up to fifteen times after the first occurrence. During a heart attack, heart muscle is depraved of oxygen and will literally die if the artery remains blocked. The first few hours are critical in saving much of the dying heart muscle and preventing permanent heart damage. Unfortunately, the symptoms vary and the most common reason for critical delays in medical treatment is lack of early warning and patient unawareness. Objectives It is possible to detect the onset of a heart attack and eliminate patient error. Our device intends to benefit those at high risk and already receiving some form of surgical treatment. By implanting a chip into the body, the patient need not worry about device operation. The patient will only be required to carry a cell phone equipped with Bluetooth and GPS technology. When the implant detects a heart attack, it will alert the cell phone which in turn will automatically call for help and provide the patients location. The goal is to provide early heart attack detection so that the patient will be given medical attention within the first few critical hours, thus greatly improving his or her chances of survival. Product Features Continuous monitoring of hearts electrical activity Rapid detection of heart attacks Automatic call for medical assistance Identifies patient location to emergency personnel

Product Benefits Provides early detection of heart attacks Eliminates delays in receiving medical treatment Improves healthcare services to at risk population Saves lives and improves quality of living

Design
Block Diagram

Biosensors

Analog Circuitry

A/D Converter

Microcontroller

Bluetooth Module

GPS Unit

Cell Phone

Block Descriptions Biosensors - Disposable Ag-AgCl ECG round pad electrodes are to be placed on each wrist of the patient. The electrodes are imbedded in pre-soaked electrolyte foam with double-sided peel-off adhesive tape for attachment. The foam provides good electrical contact with the skin and reduces motion artifacts. The electrodes read the hearts electrical activity and outputs to the circuitry. Analog Circuitry The circuitry will consist of two buffers, a differential amplifier, and a band-pass filter. Each electrode will connect to a buffer which is needed to match the high impedance of skin to the low impedance differential amplifier. The differential amplifier then takes the difference between the data collected by the electrodes and provides a gain before outputting to the band-pass filter. The band-pass filter is needed to eliminate noise (other biological signals, environmental, motion, etc.) and provides additional gain. Finally, the ECG waveform is fed into the A/D Converter. A/D Converter The analog to digital converter will convert the analog data from the biosensors to digitally sampled data points, while allowing enough resolution and sampling rate for our purpose of detecting a heart attack. The data points will be sent to the microcontroller and be sampled at regular intervals.

Microcontroller The microcontroller is a BASIC Stamp 2 which will run a real time program to constantly monitor the output of the A/D Converter, comparing current data samples against stored samples. It will include an algorithm to process both the amplitude and frequency of the heart beat, to cover as many possible cases of a heart attack as reliably as possible. Once a heart attack is detected and confirmed, relevant data such as the time of occurrence will be collected, and a signal is sent to the Bluetooth Module to initiate the emergency dial up sequence via the cell phone. Bluetooth Module The Bluetooth module is an EmbeddedBlue Transceiver AppMod which conforms to v1.1 of the Bluetooth standard and provides connectivity for the BASIC Stamp. Bluetooth is chosen as the method of choice for wireless connectivity between the sensor package and the cell phone, because the connection can be treated as a low power wireless serial link, and Bluetooth is an emerging standard for personal area networks. Cell Phone A modified Motorola V600 will be the phone platform. This phone meets the requirements for having both Bluetooth and GPS built in. A Java applet application will be written to establish the link between the sensor package and the phone and to pass GPS and subscriber information to emergency personnel. GPS Unit GPS is becoming a standard feature of newer production model cell phones in order to become compliant with FCC regulations. Work may include enabling a GPS unit to feed data to the cell phone if time permits. Performance Requirements The system must be able to reliably detect heart attacks. The analog circuit requires a bandpass filter with a lower cutoff frequency of 0.5 Hz, an upper cutoff frequency of 150 Hz, and a mid-band frequency of 20 Hz. Emergency personnel receive enough calls as it is, and flooding the system with additional false alarms is undesirable. At the same time, the system cannot afford to miss detection of a heart attack. From an engineering standpoint, the algorithm must be both precise and accurate. An error rate of less than 1% is desired. As to hardware, it must be portable, low-power, and durable. In addition, it must be able to sample at a rate of 400 Hz, the resolution required the pick up the QRS peak. Finally, communications between the sensor package and phone should be dependable. Assuming the phone is in range, it should be capable of accepting a signal from the sensor at all times.

Verification
Testing Individual modules will be tested separately. The band-pass filter will be isolated and frequency response determined. An input signal from a function generator (sine wave of 0.1Vpp) will sweep over a range of frequencies while output magnitudes noted. To determine the common mode gain of the differential amplifier, both input signals are provided by the function generator (sine wave of 0.5Vpp at 20Hz) and output voltage recorded. To find the common mode rejection, one input signal will be grounded while the other once again provided by the same function generator signal. The overall gain of the ECG amplifier will be found by grounding one input to a buffer and providing a sine wave of 0.1Vpp at 20Hz into the other buffer while taking note of the output magnitude. Finally, the ECG amplifier will be tested on real ECG waveforms by placing an electrode on each wrist of a subject and connecting to the buffer inputs. The microcontroller should be able to run the algorithm using simulated data; the phone should be able to place a call. Put together, data simulating a heart attack will be fed into the system, and it should be able to detect the occurrence, and have the phone call for help. Tolerance Analysis Tolerance analysis will focus on the heart attack detection algorithm. Data on heart attacks will be collected from existing medical sources and analyzed. Many various forms of heart attacks exist. Furthermore, because of variations in heart rate due to age, condition, activity, and other factors, distinguishing between a heart attack and normal activity may be problematic. The goal of the data analysis is to have a solid grasp on the distinguishing characteristics of a heart attack and design the algorithm accordingly. Validation of the algorithm will be completed by feeding the data into the algorithm. The analog circuit should produce a standard ECG waveform with a QRS complex at least ten times greater in amplitude than baseline noise.

Cost & Schedule


Cost Estimates Part Ag-AgCl Electrodes AH 59-8438 1F Capacitor (25V Rating) 0.01F Capacitor (25V Rating) 1k Resistor (1/2W Rating) 6.2k Resistor (1/2W Rating) 10k Resistor (1/2W Rating) 33k Resistor (1/2W Rating) 100k Resistor (1/2W Rating) 3.9M Resistor (1/2W Rating) Op Amp National LM 741C A/D Converter TI ADS7800JP JavelinStamp Starter Kit EmbeddedBlue Transceiver AppMod Motorola V600 Labor Total Quantity Box of 50 1 1 3 1 4 2 1 1 4 1 1 1 1 288 hrs Unit Price $0.90 $0.11 $0.08 $0.05 $0.05 $0.05 $0.05 $0.05 $0.05 $0.68 $45.08 $239.00 $99.00 $369.00 $40/hr Subtotal $45.00 $0.11 $0.08 $0.15 $0.05 $0.20 $0.10 $0.05 $0.05 $2.72 $45.08 $239.00 $99.00 $369.00 $11,520.00 $12,320.52 Status Pending Obtained Obtained Obtained Obtained Obtained Obtained Obtained Obtained Obtained Obtained Pending Pending Pending -

Schedule Week Sep 12 Sep 18 Description of Activities Submit Proposal. Diane Design analog circuit and research ECG waveforms. Kuangwei Gain familiarity with BASIC Stamp. Lianhan Gain familiarity with V600 phone platform. Procure required hardware. Begin implementation. Diane Analog circuit and ECG waveform analysis. Kuangwei Microcontroller, algorithm design, Bluetooth. Lianhan V600, algorithm design, Bluetooth. Design Review. Continue implementation. Diane Search for ECG data files. Tune circuit. Kuangwei Code algorithm in Java. Lianhan Background work on V600 J2ME environment. Continue implementation. Diane Look into constructing PCB for circuit. Kuangwei Code ADC and Bluetooth interfaces. Lianhan Code SMS and Bluetooth links. Finish implementation. Diane Fabricate and populate PCB. Kuangwei Integrate modules. Lianhan Integrate modules. Integration and testing. All. Integration and testing. All. Mock-up Demo. All. Debugging. All. Debugging. All. Thanksgiving Break LAZINESS Demonstration and presentation. All. Turn in final paper and lab notebooks. All.

Sep 19 Sep 25

Sep 26 Oct 2

Oct 3 Oct 9

Oct 10 Oct 16

Oct 17 Oct 23 Oct 24 Oct 30 Oct 31 Nov 6 Nov 7 Nov 13 Nov 14 Nov 20 Nov 21 Nov 27 Nov 28 Dec 4 Dec 5 Dec 11

You might also like