Ali Mohammed Jobayer; Shishir Bordhan; Md, Towhidul Islam; Nipu kumar Das
Department of Electrical and Electronic Engineering
Chittagong University of Engineering and Technology
Chittagong, Bangladesh
jobayer.ieee@yahoo.com; bordhanshishir@gmail.com; towhid.pappu09@gmail.com; nipu16@gmail.com
Abstract-The number of decease due to Cardiovascular Disease heart muscles such that heart pumps enough blood to the
(CVD) in low and middle-income countries has become a major whole body. Electrocardiogram is the widely used diagnostic
concern. From a clinical standpoint, electrocardiogram is an tool to test periodically the electrical and muscular functions
useful mechanism to diagnose cardiac diseases. Therefore,
of the heart. Although cardiac signal is a low amplitude signal
electrocardiogram based machine design has taken into
distorted by many artefacts. Power line interference is the
consideration to create a framework for building a simple, low
main problem and that is why low noise cable gets privileges
cost microcontroller based ECG monitoring system. However,
most of the machines in current research has high price with in designing process[II ]. Besides instrumentation amplifier
bulky shape which deprives the people of remote area to get with high common mode rejection ratio, RFI protection
benefit from it. Therefore, the paper tries to design a smart circuit, right leg drive circuit and analogue filter design can
economical diagnosis system which can maintain reliable provide better extraction. To eradicate the noise, digital low
diagnosis with better portability. This system has been pass and band stop filter have been used. Java based
implemented by instrumentation amplifier with high common Processing IDE has been used in this system to display the
mode rejection ration and right leg circuit to remove 50 Hz noise
cardiac signal graphical view on the monitor. The following
of ac main. Digital and analog filtering of the system has
fig.I shows the normal electrocardiogram with waves,
optimized the signal by removing baseline wandering and power
intervals, segments.
frequency noise. The communication between the hardware and
software is interfaced through USB port using microcontroller.
The implementation of this system shows greater benefits in
domestic health monitoring and rural area's medical camp.
I. INTRODUCTION
In strumentation
Amplifi er A D 6 2 0 . B and Pass non-inverting L evel Arduino Computer
with rig ht leg drive Filter Amplifi er S hifter uno Interface
circ u it
49.4Kn nd
The cut- off Frequency of 2 order low pass filter is :
RG =
G -1
1
For G = 120, RG = 440n. For G = 1, the RG pins of cut. off Frequency, f =
(2xnx1.8kx 4.7f.t)
= 33.73Hz
fd
So the Overall cut-off Frequency of 3 order low pass filter is
approximately 25-30 Hz. The following fig. 6 shows the
simulated value of the active low pass filter.
Fig. 3. First order high pass filter.
1
cut. off Frequency , f
2 π R C
= π
= 0.05Hz
μ
Rl1 II RIO
VI = O.5 V
(Rl1 II RIO) + RI4
Voltage at noninverting pin due to V/N
Fig. 7. Noninverting amplifier.
RI4 II RiO
V2 =
V/N
(R14 II RiO) + R11
The close loop gain (ACL) of the noninverting amplifier can be
calculated by the following formula. 0.5 * V/N
R13
- VOUT _ 1 RI6 So the output Voltage, Vout= (VI + V2)
A CL - - + R12
V/N RIS
[f the R16= 560k and R15 = [Ok, then the close loop gain is,
ACL = 57. Close loop gain is responsible to amplify the signal.
Fig. 8. Amplification of Noninverting Amplifier. Fig. 10 shows the simulation of level shifter. Here, level
shifter shifts the signal by 1 volt
Fig. 8 shows the simulation of amplification of noninverting
amplifier. Here, noninverting amplifier amplifies signal by 5 7.
-40
VI. SYSTEM OUTPUT
-10 Bandstop Bu
utterw orth
Magnitude (dB)
-20
-30
-40
0 50 100 150
Frequency (Hz)
Fig. 12. Frequency Response of bandstoop filter. Fig. 14. Measured ECG Signal.
Fig. 11 and 12 shows the frequency response of low pass filter VII. CONCLUSION
and bandstop filter respectively. Fdatool, which comes with Thus this ECG machine, provides user friendly operation
Matlab, facilitates to determine IIR filter order, section, and on windows based operating system and can be connected to
coefficient for implementation in filter coding. any PC easily with automatic and instantaneous interpretation.
[t includes noise free ECG acquisition and works as storage of
V. INTERFAClNG AND SOFTWARE DESIGN ECG signal.This setup can be useful to doctor as well as mass
people. However, ECG signals are to be prepared at actual
Arduino Uno communicates with PC through Serial
state after some time delay because of the processing of digital
interface using RS232 protocol. In firmware, data rate is
filter which is necessary to improve more. This machine can
28800bps[7]. It gives baud rate 320Bd, and is enough for ECG
provide Lead I, Lead II, Lead III electrode placement
bandwidth. For displaying ECG signal, GUI window is
facility[II ]. Use of both analog and digital filter enhances
designed using processing software depicted in fig. 13 [8]. It is
noise reduction. Using laptop/computer monitor as a
an open source programming language and integrated
displaying system reduces cost of this device.
development environment for visual design and builds on the
java language. The software has some capabilities such as
zoom in, zoom out, hold on, data saving in.txt format. As
every analog and digital filter has unity gain, calibration
provides approximate O.75mV per box after calculating
amplifier gain and instrumentation amplifier gain.
REFERENCES ported-files/tutorials/MT-070.pdf
10 World Health Rankings Live Longer Better,[Online].
Available: http: //www.worldlifeexpectancy.comlbangla
David A. Bell."Operational Amplifiers and Linear Ics."
desh-coronary-heart-disease.
2nd edition, Prentice Hall of India Private Limited '
New Delhi - 110001,2005, pp. 120-123.
II Abdullah AI-Amin, A K M Bodiuzzaman, Ahmed
2 Charles Kitchin and Lew."A Designer's Guide to
Imtiaz Khan, K Siddique-e Rabbani. "Design and
Instrumentation Amplifiers." 2nd edition, Printed In
Development of a PC based ECG Equipment",
U.S.A, 2005, pp. I 23-130.
Bangladesh Journal of Medical Physics, Vol. 6, No.1,
3 Albert Malvino, David J Bates."Electronic principles."
2013.
7th edition, Tata McGraw-Hili Publishing Company
12 EI Mimouni EI Hassan and Karim Mohammed.
Limited, New Delhi, 2010-2011, pp. 203-223.
"Design andImplementationof an Embedded
4 Sallen-Key Low-pass Filter Design Tool, [Online].
Systemfor Ambulatory Cardiac Monitoring."
Available:
University Sidi Mohammed Ben Abdellah ' F'es,
http://sim.okawadenshijp/en/OPseikiLowkeisan.htm
Morocco.
5 Non-Inverting Op-Amp Level Shifter. [Online].
Available: http: //www.daycounter.cOlnJCircuits/OpAmp
-Level-Shifter/OpAmp-Level- Shifter.html
6 John G. Prokis."Digital Signal Processing Principles,
Algorithms And Applications." 4th edition, Ed. Pearson
Educationlnc., 2007, pp-214-233.
7 Arduino, [Online]. Available: http://www.arduino.cc