OFFICE OF THE MEDICAL SUPERINTENDENT
CHANDKA MEDICAL COLLEGE HOSPITAL
LARKANA
e
Tender Document
Purchase of Machinery, Equipment /
Instruments & Furniture / Fixture for
various Departments / Wards, CMC
Hospital, Larkana for the financial
* year 2021-22 "
(Single Stage Two Envelope system)Instruction to Bidders
2 Salient Feature / Terms & Conditions :
3 ‘Terms and Conditions of Tender
Documents Cheeklist
Evaluation Criteria
Bid Letter Form
Manufacturer Authorization Farm
Contract Agreement Form
Bank Guarantee Form
Integrity Pact .
Bid Form & Price Schedule
Schedule of Requirements
Invitation for Bids
Technical Specifications of Machinery ‘Medical Equipment
Se :
Sa .
(fo
ovarinieu wit ©
2of 146
seew
10,
Bidders are required to check that Tender Documents issued to them are complete in all respects
INSTRUCTIONS TO BIDDERS
Office of the Medical Superintendent Chandka Medical College Hospital Larkana invites
sealed bids on single stage wo envelope basis as per Sindh Publie Progurement Rules, 2010
(Amended 2019 up to date) from Manufacturers within the Pakistan and aboard, and their
Authorized Agent / Importers! Bidders / Distributor / Contractors for the Purchase of Machinery
Equipment Instruments & Furniture Fixture for various Departments ¢ Wards, CMC
Hospital Larkana for the financial yenr 2021-22 {
as per Summary of contents related.to instructions to Bidders, Salient Features, Evaluation Criteria,
Form of Tender, Form of Agreement, Specifications (Technical), Scope of Work, Drawings etc.
‘The rates to be inserted in the Bill of Quantity shall be deemed to include all casts of whatsoever
description and expenses necessary for the Supply, Installation, Testing, Commissioning and
Maintenance of the Equipment / Machinery together with all risks, taxes, liabilities and
obligations, specific or implied, in the Tender Documents, Arithmetical errors, if any shall be
Sorrected and Tender price amended secordingly
‘The original bid shall be typed or written in indelible ink by the bidder or person duly authorized.
‘The person or persons signing the bid shall initial all pages of the bid. The name and designation
of each person signing must be mentioned below the signature
No bidder shal] be allowed to alter or modify his bid after the bids have been opened. However,
tbe proatei agency may seck and accept clarification to the bids that do not change substances
of the bids, F
No unauthorized alteration may be made in the Tender documents, [any such alteration is made,
tender may be liable for rejection.
‘The vendor has to quote only one rate for each work as per tender specifications, Hand written
‘tenders or any over writing, cutting, should be signed.
‘The Procuring Agency may reject all bids or proposal st any time prior to the acceptance ofa bid ,
or proposal. The Procuring Agency upon request communicate to any supplier or contractor who
submitted a bid or proposal, the grounds for its ejection of all bids or proposal, but is not required
to justify those grounds,
Any clarification, revision, addition or deletion, in the Tender dacuments may be made by the
Procuring Ageacy before the submission and opening of Tender in the form of Addendum |
Corrigendum. This will be made only be formal Addendum / Corrigendum issued by the Procuring
Agency and will become part of the Contract Documents. Each addendum shall be signed by the
Tenderer and returned with other Tender documents.
The entire Tender Documents, listed duly priced, signed & stamped on cach page and completed.
must reach at designated place in due time and dates as defined in the Bidding Data of the Tender.
<
ee
Er Bef 146
ocanmead eur cae
Zi11. The contractors whose Tender is accepted, by the Procurement Committee, will be required to
enter in to 8 Contract Agreement as defined in the Form af Agreement.
12. The contract resulting from this invitation shall be governed by the Sindis Public Procurement =|
Rules, 2010 (Amended 2019 up to date), i
13. All manufaetured and ether items should be used in the work in accordance with the instructions,
Specification in the Tender Document and also in accordance with generally accepted norms of
good workmanship.
14. The Tender documents referred to above constitute a part of this Tender and comprise of the
following decuments:
4) Salient Features / Terms & Condition of the Tender Questionnaire /Proforma.
6) Evaluation Criteria, *
) Form of Agreement.
) Scope of work / Bill of Quantities / Specification (Technical).
@) The Addendum (if any).
f) ‘The Drawings (if any).
1S. Bidding Procedure: Single Stage - Two Envelope Procedure
8) Bid shall comprise a single package containing two separaie envelopes, Each envelope shall
contain separately the financial proposal and the technical proposal;
') Envelopes shall be marked as FINANCIAL PROPOSAL and TECHNICAL PROPOSAL
in bold and legible letters to avaid confusion;
©) Initially, only the envelope marked —TECHNICAL PROPOSAL shall be opened;
d) Envelope marked as FINANCIAL PROPOSAL shall be retained in the eustody of the
procuring agency without being opened;
¢) Procuring ageney shall evaluate the technica proposal in a manner prescribed in advance,
without reference to the price and reject any proposal which does not ebnform to the
specified requirements;
f) No amendments in the technical proposal shall be permitted during the technical evaluation;
2) Financial proposals of technically qualified bids shall be opened publicly at atime, date and
venue announced and communicated to the bidders in advance;
h) Financial proposal of bids found technically non-responsive shall be returned un-opened to
the respective bidders;
i) Bid found to be the lowest evaluated or best evaluated bid shall be acogpted.
16, In ease of diserepan between the Invitation for Bids (IFB) ‘ Tender Notice and the Bidding.
Documents, the Bidding Documents shall take precedence,
17, The bidders are expected to examine all instructions, forms, terms, and specifications in the
bidding documents. Failure to furnish complete information required in the bidding documents or
to submit a bid not oo responsive to the bidding documents may result in rejection,
- ocanneu1, | Name of Work & Address
“Supply and lasiallation of Machinery /Equipment
“Instruments & Furniture Fixture for various |
Depurtments / Wards, CMC Hospital Larkana for |
the financial year 2021-22,
2. | Time & Date of issuc of Tenders
From the date of publishing to one day before opening
of Technical Proposal, during offiec hours. |
Itshould be (As mentioned in Notiee inviting tender).
3. [Place of Issuance of Tender
Oilice of the Medical Superintendent Chandka |,
Medical College Hospital Larkana
4 | Date of prebid mecting
5, | Method of opening of Tender
Tn case requested by bidders or infrmed by Procuring
Agency.
Mentioned pre-bid mecting date if any application
received considered for discuss same mentioned
date
Single Stage — Two Envelope; os per SPPRA-Rules
2010 (Amended 2019 up to da
6. | Date & Time of Submission of Tenders
‘As published in newspapers
Ii should be (As mentioned in Notice inviting tender).
7, | Date & Time of Opening of Tenders /
Technical Proposals
“As published in newspapers
t should ke (As mentioned in Notice inviting tender)
8. | Date of Opening of Financial Proposals
‘As informed by the Chairihlan of Procurement |*
‘Committee to all participants
9, | Venue of submission & opening of
Tenders
10.) Validity of Tenders
Difice of the Chairman of Central Procurement
Committee, Vice Chancellor Office at Dow University
of Health Seiences Ojha Campus
‘90 days as per Sindh Public Procurement Rules, 2010
(Amended 2019 up to date).
11,| Amount of Earnest Money/Bid Security
1% of the Contract value total quoted amount in shape |”
of Pay Order / Call Deposit / Bank Guarantee from
schedule Bank in favor of office of the Medical
‘Superintendent Chanuika Medical College Hospital
Larkana to be submitted slong with Financial
Proposal and also allached photo copy without
showingamount of that pay order else the offer will be
rejected. i
12,|Supply of Equipment
| siation
Within up to 10 weeks. However, this period is
extendable as per mutual agreement/demand of the
Ss
4
ocanney WIL LaIns
a1
13.| Installation Period Ti will start afler receiving of equipment at site.
‘Twill start aftcr satisfactory installation of equipment
atsite.
Tt Will start aficr the satisfactory installation of the
equipment and issuance ‘of Installation Certificate
from the concemed official. This period will remain
i functional sill two yeurs with replacement of parts.
15.
‘Warranty Period with replacement of “The waranty period of parts with Free of ‘cost
bbe for two years. Along with two
14) Maintenance Period
Parts and services replacement
years free service & maintenance from the date of
Installation.
16.) (a) Contract Agreement The Contractor shall enter & execule a formal y
Agreement as per the “Form” annexed with such
Stamp Paper requirement for Ageeement| modification as may be necessary.
= Rs, 0.35% of the Contract Value ar as prescribed by
Government Laws. .
{hy Opening of LC (eller of Credit) | Letter of Credit (L.C.) will be opened by Procuring
‘Agency 2s consignee and as per the prevailing rule of |
same, Performa invoice of the quoted amount on C&P
| CAF, basis along with insurance cover note if
so be submitted by the contractor(s).
establishment of Letter of Credit
in currency if any and other
miscellaneous chasges on logistics, transportation
clearing from sea port airport will also be paid/bear
by the Procuring Agency.
17,| Release of Earnest Money/Bid Security | To un-successful bidders, after financial bid is opened
or technical bid is rejected,
Earnest Money will be released to successful Bidder
after purchase order is released, but efter 3% Security
Deposit submission. "
18.) Security Deposit 3% Security Deposit in the shape Pay Order/Banke
Guaranty will be submitted by Contractor with
Procuring Agency which will be released as per item
Bank Guarantee of equal amount will also be
acceptable.
Security Deposit is deposited as per item ~ 19. He
‘After completion of installation and maintenance of At
Defect Liability period.
70, Variation in Contract Price ‘No variation in price shall be allowed on any ground
including currency fluctuation/variation or
. whatsoever.
==
19,| Release of Security Deposit
Gof 146
ae
ocanneu withl Gansta21] Discrepancy
Tender /Bis
mentioned elsewhere, Salient Features / Terms &
Conditions and Additional Terms & Conditions of |,
Tender will govern.
| Taxes
“All taxes will be deducted as per prevalent laws of
=
Approved makes
.| Special Note Regarding Equipment
Technical offers / Commercial, offers falling 10
demonstrate below details would be rejected:
8. Quoted system must be of advanced & latest
version. :
b, Tender must cover complete equipment.
€. Tender naust provide ail technical details up to the
satisfaction of the end user.
d. lems should be quality approved from the
concerned international body of the respective
indus
25.| Training
Supe!
the personnel working amv/operating the said
Equipment / Machine as long as the need prevai
26.| Maintenance.
‘Maliitenanee cost for all items for Two years from the
date of successful Installation shall be undertaken by
the Contractor (Maintenance includes all Parts &
Labor, ete, with sufficient staff, during maintenance
period).
27] Default in Preventive Maintenance,
Breakdown and Emergeney Calls.
In case of default by the Contractor with respect t0
‘maintenance, break down and emergency calls, the
same will be carried out within 2 hours by the
Procuring Agency and the cost so incurred will be
paid from the Retention Money.
+ Moreover, an additional 10% of the amount spent
would be charged from the, concemed contractor
being defaulter. be
78. Cost to be quoted in BOQ.
‘© The Contractors shall quote DDP price of the
Equipment’s including Installation = &
Commissioning cost,
It should also include cost during maintenance
period including the parts and maintenance staff.
will provide onsite suncessful training to al |,
ae Exemption.
Tn case of purchase on C&F/ C.F basis, the required |,
certificate will be issued by this office that the import
hhas been made for this hospital, so as to avail the
facility of exemption of dutlesftaxes, 9s per
Goverment Rules/Policy according to the Sindh
Public Procurement Rules, 2010 (Amended 2019).
Scanneu WILT &TERMS & CONDITIONS OF TENDER
lote: No t Ibe accepted 9 ing of the Tender box, whatsoever reason may be,
+ The Contractors responsible for Supply, Installation, Testing and Commissioning of Medical
Equipment, Machinery Instrument while Installing Medical Equipment at no extra-cost. So, the
‘Contractors should take into- account all such expenditures while quoting the bid, af
E
An Technical Bid, the bidder must provide original data sheet, technical brochure, and all other
felevant documents, along With copy of Pay Order / Bank Draft / Bank Guarantee without showing
the amount figure otherwise, the bid will be ignored/rejected, The technical eva
‘on the basis of criteria given in tendet document.
we
3 They endor has to fill up compliance sheet with YES/NO or put value as desired in the column ageinst
Provided specification, These specifications must be highlighted in the brochure attached; else the
offer will be rejected,
4. In Financial Bid, the vendor should mention financial offer along with Pay Order/ Bank Draft! Bank
Guarantee 1% value of quoted items in favor of, Office of the Medical Superintendent, Medical
‘Superintendent Chandks Medical College Hospital Larkana, In ease of purchase on C&P / C.LF
Basis, the procuring agency will invest open Letier of Credit after getting proforma invoice and
insurance cover note ftom the bidder within thirty days after receiving the supply order. Bidder will ae
bear all the expenses such as bank cherges, freight charges etc. The required certificate will be issued varmeuthe administration of Medical Superintendent Chandka Medical College Hospital Larkana, The
warranty period will commence fram the date of installation of the Equipment at Office of the Medical
Superintendent Chandka Medical College Hospitn! Larkan
‘of their financial standing, listing of similar items supplied in
iems and details of office with number of personals.
1.Finn must provide complete deta
Pakistan, available local stock 0
12. The vendor has to quote only one rate for each item as per tender specifications. Hand written tenders
or any over writing, cutting, should be signed.
13. Vendors should quote the items they intend to supply,
14. The bidder shall submit with the bid, an Undertaking on Non-Judicial stamp paper of Rs.S0/- duly
attested by Oath Cortimissioner, stating that the medical equipment will be supplied brand new, free
from defects as per the approved specification, also provide two years warranty for service and
maintenance and will bind himself for availability of spares for a period ‘of 10 years from the date of
supply. In case of winding up of his business, the supplier will be bound to make an alternate
arrangement of spares and maintenance af medical equipment Incase of default, depreciated cost of
medical equipment will be recoverable fiom vendor. :
IS IF it has been found that the information submitted by the bidder regarding his qualification and
professional, technical, financial, legal or managerial competence as supplier / contractor is /was false
and materially inaecurate or incomplete at any stage, then the bidder will be disqualified
free of cast and also
16 Suppliers / Contractors will be responsible for supply at consignee end,
responsible for free installation work, if required. ‘
¢ Store Keeper Ghulam Medical Superintendent Chandka,
17.Equipment will be handed over to th
be allowed,
‘Medical College Hospital Larkana, Price escalation will not
18.1f vendor doesn't supply medical equipment in accordance with the supply order or fails ta supply
according to terms and conditions or in case of any other defoutt, the Performance Security and
Security Deposit would be forfeited and further action as per SPPRA rule would be taken against such
vendor.
19.1f the supplier fails to give sepply and install within the stipulated period, Liguidated charges will be
imposed.
‘4. The decision ance taken will be final and will not be challenged in any Court of Law.
2LA copy of Income Tax.and GST. Regisiration Certificate should be attached along, with Technical Bid.
2 The bidding firm must be registered with PNRA
items).
23 The bidder shall furnish copy of ATL (Active Tax Payer List certificate).
cate, Income Tax Certificate and GST
ocdi ine wi
department for minimum past three years (for x= ray35, The bidder shall fumish three years Income Tax Retum Certificate,
6 Conditional Tender(s) against the Goverment rules and non-compliance of instructions mentioned in |
‘Tender Form & Hand written tender(s) will not be enieriained,
21. The vendor should submit a Data Sheet according to the Tender Specifications. Form as mentioned in
the tender and highlight the tender specifications in their Technical Data Sheet. Additional Features
can be mentioned separately,
28 In case the information provided by bidder in Data Sheet as per Clause No. 30 is incomplete or
incorreet, the quoted item will be liable for rejection,
‘3 Supplier must have to provide Manual book with circuit diagram where applicable.
30 Supplier must have to arrange training on site for concerned technical staff for the supply of Machinery
and Medical Equipment
31. Procurement Committee, Medical Superintendent Chandka Medical College Hospital Larkana
Health Department, Govt, of Sindh reserves the right to reject or accept any’/ all tender(s) as per SPPR-
2010 (Amended 2019),
a
% The system will have the minimum uptime 95% in case of down time beyond this period the following
penalty will be applied:
a) 95% to 100% No penalty.
b) 90%6t095%the warranty period will be extended by the number of days system in down.
©) 85% to 90% warranty period will be extended 1.5 time the down time period
d) 80% to 859% warranty period will be exteaded two time for the down ns and suppl
supposed to give justification of delay.
Uptime less than 80% hospital will complain to the principal to take action against the supplier
and warrant\’ will be the three time for the dawn time periad.
{) However, PPM (Periodic Preventive Maintenance) will not be considered in the downtime and
PPM of the system will be done as per manufacturer recommendation.
33 Payment of DDP basis, the vendor shall submit Bill“Invoice for the purchase order value after delivery,
installation and issuance of satisfactory report by the end user. . it
®)
34 Bids shall remain valid for 90 days after the date of bid opening and same may be extended in terins
of Rule 38 (2) (3) (4) of SPPRA Rules,
38 Bid should be inclusive of all Government taxes (if applicable) and the same will be paid by the
Contractor except withholding tax and 0.35% Stamp Duty which will be deducted at source in office
of the Accountant General Sindh Karachi,
34 The bid security will be reas ti og Covert bidder thd bis ba tet openingand ,
before the expiry of riod or fails to sign the contract in splat time i i
ge
Ocarimeu Witaccepted,
37 The Procuring Agency may seject all or any bid at any time prior to the acceptance of a bid or
Proposals, subject 10 the relevant provision of SPP Rules, (Amended 2019 upte date). Payment
Will be made within four weeks afler receipt of bill / invoice duly fills in all respects.
38 Prices quoted shall remain valid up to 90 Days.
3.Bids submitted late due to any reason what so ever, shall not be considered and retumed unopened to
the bidder or his authorized representative,
4. Registration from Sindh Revenue Board is not required in procurement of Goads.
1/ We agree to above menitlaned terms & conditions:
Name of Contractor Signature
(CNIC NO (Copy must be attached.)
Full Address
Rubber Stamp
ocadiineuDOCUMENT CHECKL
Please review the following list of all possible documents (o be enclosed with the Technical
Propos
Tender Purchase Receipt (Original)
‘Bid Security (Copy with value hidden in Technical Proposal; Original
in Financial
Bidding Documents (Duly filled, Signed & Stamped by Bidder)
‘Technical Proposal on Bidder’s Letterhead
Compliance sheet on letter head
Undertaking (as per sample form)
Certificate (as per sample form)
Inconte Tax & GST Registration Certificate
Professional Tax Certificate (Sindh)
SECP Incorporation Certificate
PNRA Registration Certificate
PEC Registration Cerificate of Bidder Code ME-06 for Medical
Equipment (for Electro Medical Equipment only)
DRAP license of Bidder showing importer of medical devices
FORM-™4 (for Electro Medical Equipment only)
Company Profile
‘Bank Account Maintenance Certificate
Income Tax Returns (last 3 years)
‘Audited Financial Statement (last 3year)
"Workshop for After Sales Services
Technical Team Deiails (Workshop details & Technical Staff Bio-
data with Training Certificates}
20
‘Manufacturer's Authorization
21
‘Catalogue / Brochures (original)
2
Product Technical Data Sheet
Manufacturer [50 9001
a
4
Manufacturer [SO 13485
Manufacturer 150 14001
= |
‘Manufacturer CE/ FDA [JIS J MHLW
26
All pages of the bide
‘Mandatory documents are mentioned in “Instructions to Bidders’. i
inted literature, shall be initialed by the Bidder, ~ ie
slg :
for un-amend
ocainleu Wi
: 4h
120fBIDDER'S DETAILS FOR NOTICE PURFOSES
ocdiineuSECTION A
| Supply, Installation, Testing and Commissioning of Medical Equipment and
Machinery, Instrument
‘ VALUATION C) TA "
Financial proposal must be submitted on company’ letter head duly signed and stamped. Bidder is required
10 type their offer in figure and as well as in words of the total amount; elsethe offer would be rejected.
EVALUATION / QUALIFICATION
11. The procuring agency reserves the right to evaluate and compare the bids oti itemized basis
OR on the basis of a group of similar nature goods OR. goods compatible with each other.
1.2 The following merit point syste for weighing evaluation factors/eriteria will be applicd for iH li
technical proposals a
13 Bidders achieving minimum 70 marks will be considered only for further process,
« Documentary evidence must be attached in support of each parameter,
14 Any Bid not meeting the mandatory requirements of evaluation criteria will be disqualified /
rejected straight away and will not be considered for further evaluation.
A___ PRODUCT EVALUATION
12 | Compliant with minar deviation (up to 10% subjeet to main function is not
affected)
1.3 | Non- Max, Output :( Generator) 4 KW or more
"| Max, Inverter Frequeney: 20kHz or more ‘
Radiographic kV Renge:40 ~ 150kV at TY steps
Radiographic mA Range: 10-S00mA.or mare
Radiographic mAs Range : 0. ~S00mAs or mare
+ Analytical Programs: Maximum 800 Programs $00
Programs or more
+ Indicator: Color Coded LED Indicator
‘Configuration:
4 High Veltage Gencesior 40kW or more
X-ray Tube: 200 KE or more
Focus Spot: Dual Focus 0.6 / 1.2. mm or beter
oray rt Snesification
+ Vertical X-ray Tube Stroke: 580-1800mm or more
+ Maximum SID with Table: 1100-2000mm oF more
Tube Rotation on Horizontal Axis; «180 r0-+180 deg, or
= more
+ Tube Rotation on Vertical Axis: 180 deg. (Mechanical ‘ tii!
Lock) or better ane}
Radiography Table s
+ Loading Capacity: 200kg_ or more
Collimator: LED Callimator Lamp
Consol Digital Console
Side Hand Grip: Stainless Steel
Hand Switel Hand Exposure
Flat Panel Detector (17"x17") iis
Scintillator C31 (Cesium lodide) by Direct Deposition * ak
Pixel Pitch 145 jum or better.
Effective Imaging Area 430 x 430mm
Image Matrix Size 3072 x 3072 pixels
AMID Conversion 14 bits or better
Image Acquisition and 7
DICOM 3.0 Com
Page 24 of 146
at
aoe
ocaiieu with GarnstaAcquisition Workstation nnd Soitware Latest; Image quality contral
‘ools with piel fixing tools exible configuration including,
‘angsiige localization Up to three image marks aind ene ear
supported jn one view (ean be done aulomatically according to
sonfigoration) Image cropping (can be dene automatically necording
to configuration)
09001, 180 13488,
FDA S10(K) approved / CE marked,
‘Manufaeturers frov other couniries (other than USA, Europe
oF Japan) will be nccejted if equivalent has FDA SI0(K) & CE
JIS.
i
Nem No.2] Wigh-Ead Color Doppler Ultrasound machine With mel
oe
7
grade wide screen monitor and touch pane! control sereen,
Convex, Linear and TVS Probes included with all application
And measurement packages with Qaline UPS and BAW
‘Thermal Printer,
(each for CMCH City Block & Teaching Hospital Black)
Digital Bean Color Doppler Ultrasound two machine for General
and two machines with Echocardiography facilities.
‘+ The system shall wilize a new generation all-digital beam
former technology which enables parallel Quad beam
rovessing of the RF signal data inthe time and amplitude
domains
‘= The system shall provide at least 512 lines of density or better,
= The sysiem shall provide up to 200 000-channcls ot more
* The system shall have system imaging frequency range from 1.610
IBM Tze more PST maine eaency fanee
«+ The systems shall provide-a dynamié range of at least 300 dB
‘or more
‘The system shall have touch sereen control with
minimum 12 inch of more and 1024x768 resolution
* Flat Panc! Display (FFD) 23 inch LCD or more
‘widescreen, color, high resolution and progressive scan
(non-interlazed)
= The system shall have minimum 3 active transducer parts
* Themes of echocardsoprapty shal have system shall display
hicast rate, integrated 3 lead ECG imertace, input for ECG and
other physiological signal
© The system shat have minimum of 2000 fps ot mace.
*# The system shall have minimum of7 selectable
transmit frequencies or mare
# The system shall have LAR flip and U/D flip forall formats
«© The system shall have a minimum of 45 em depth or more
+ The system shall have read/write 2oom by factor up to 10 x3 0°
* B-mode, Anstomical M- mede in live und cine, Colar
Dogipler, Power Doppler, PW / CW Spectral Doppler,
Mixed modes ete.
+ The system shall have one-button, automatic image
ization Feature that optimizes imaging Parameters based
‘on the actual image data Thesystem shall have technology to
optimize hyper choke signals in real time The system shall
Page 25 of 146 “S
DCANINEU WILT Larnoc(8
[provide s method for speckloreduction and edge enhanceinent
‘ar improved contrast
+ resolution selectable of minivan 3 levels
* The system shall have color DT1 and spectral DTI
functionality. and including quaalifcation package.
+ The system shall also have DTE (Doppler Tissue Energy)
* The system shall have assisted boundary detection technique
for rapid assessment of left ventricle volume
* and Bjection Fraction
+» The system shall have dynamic 20 method to visualize,
measure and display plohal and regional
* The system shall support real time quantitative imaging
method that caloulotes and displays the relative
« Stiffiess ofthe tissue. (fo be quoted separately as an option).
Transducers.
Two units should have convex transducer frequency range
of 2— 3 MHz or better. Linear transducer frequency’ range of
4 — 12 MHz or better Endocavitory transducer frequency
range of 6 = 11 MHz or better.
= Two units should have Phased Array Probe 2-6 MHz for
adults and Phased Array Probe 4-8 MHz for pediatrics and
Linear transducer frequency range of 4~ 12 Miz
‘© The cost of options to have contrast imaging and #0 imaging shoud
be quoted separately:
+ Cost of Bi-plane Endo cavity Transducer frequency range
= LO MH zor better may be quoted separately.
‘Standards: ISO 9001, 180 13485,
© Certifications: FDA $10(k)/CE/MHLW
« Manufacturers from other countries (other than USA,
Europe or Japan) will be accepted if equivalent has FDA
510(k) AND CE, JIS, MHLW or JQAO certifications.
Tem No. 3
var
‘Queue Management System with Software & Hardware
t
(01 for CMCH City block, 01 for Teaching Hospital block 01 for
opp)
cation:
Project Overview Queue Management System is next generation
advanced software-based queuing system must be integrating with
Existing Radiology Department Software. It aulomates not only
Complete visitor workilow at servicing area but also calculate the
total jsumey of Patient from arrival to-complete processed. It
provides all comfort your patient deserve without any hassles, can
provide vital MIS reperting of your patient Row and Inerease patient
Tatisfaction by eutting down waiting times,
Core Funetionality ;
Following are the core functionalities for the queue management
system, Waiting. arco, Display called tickets and display estimated
serve time (optional) Management and Anaiytic app, Smart
‘analytics with monitoring and reporting of siaff performance and
patient satisetign-and ealculate the total journey from start to end,
‘lea includes 5 ~
Page 26 of 146
OCaNeu WIth U-Queue management system is & fully automaled but it still rely on
the personal human toiich, by using app (Tablet, Web & Chrome)
which your representatives can use to call customers and offer them
a more personatized experience,
Scheduling & Appointment
‘An interface will be provided for scheduling the appointment haurs,
So the appointment based on branch wise schedule.
Monitoring,
‘You can monitor staff performance in realtime and receive Patient
feedback. Monitoring consists of Performance Reports Station
serving tine Agent Performance Chhnrts patient counts Station wise
Patient count Core Benefits For Visitors A QMS takes care of that
by: Reducing actual and perceived wait time, Eliminating the
pesibility of joining the wrong line. Informing patient of their real-
¢ status in a quenie. Allowing patient to spend their queuing time
more productively. For Managers Frontline slerks and managers
care about preventing the chaos of queuing. Or Equivalent
Standards: ISO 9001, ISO 13488,
Certifications: FDA $10(k) approved /CE marked.
Manufacturers from other countries (other than USA,
Europe or Japan) will be accepted if equivalent has FDA
510(k), CE, JIS , MHLW or JQAO certifications,
©.T Table Electra-hydraulic remote controlled, with
battery hackup and complete accessories
(for Teaching Hospital block 2 for Urology | for Gyne 2 for
Ortho),
Specification:
/+ Automatic Control System Satisfying Various Demands of All
Surgeries.
‘Extreme Weight Load 250kg-280kg with Highest Safery &
Stability at static position.
‘© Memory Pad with Antistatic Waterproof design.
«* Standard Override Panel Control Makes the control Safer,
«Easy Click Leg Plate Controlled by Manual Control.
jouble Contral System.
ICA CATION:
Dimension 2060°520mm_
530mm to 970mm
and Reverse Trendefenberg 30°30"
25°-25°
Head Plate up 45*-down 90°
Back Plate up 90°-down 45°
Leg Plate up 20°-down 94°
Longitudinal Sliding 300mm
Kidney Bridge Lifting 120mm
Battery Standby 120 minutes
Patient Weight Capacity 250kg -80kg
Accessories: . _
AmRet pes c
LegCmiches pes ~
Shoulder Support 2 pes
Belt Ipe
Screen Frame Ipe
Body Supporter 2pes
Page 27 of 146
ocdiineu WiHand Remote Control pe
Power Cable Ipe
Foot Suppor’ |pe
Mattress Ipe
Long Fixing Clamp tpe
Square Clarnp 1 pes
Square Clamp U1 2pes
Standards: 180 9001, ISO 13485.
Certifications: FDA 310(k) approved / CE marked,
Manufacturers from other countries (other than USA,
Europe or Japan) will be accepted if equivatent has FDA
$10(k), CE, JIS , MHLW or JQAO certifications.
Tem Sook
OT Light LED Double Domel60,000+ 90,000 Lux Ceiling
with Ontional Camera
ee ‘Teaching Hospital black 2 for Ortho 2 for urology 1 for Gyne |
‘60000-120000/80000-160000
Color Temp 350k, 3750, 4050K, 350k, 4650k, $000k
Spot Diameter (mm) " 150-350
Led bull power (W) German OSRAM: I'W each
‘Total Power (w) 4gwe6sW
Bulb service life ‘more than 50000 hours
Diameter of light head SO0mm+700mm. |
Focus manual
Sterilize Handle Foous Function Yes
Three Mode Yes
LED touch sereen Yes
Bulb number 48+78pes
‘HD Camera system
include:2mega pixel SONY / Equivalent integrated camera, control
system,22' monitor and display, accordingly rotating arm and spring
arm, accordingly video recorder, decorer, 1T hard disk, cable lines
Note : CAMERA SYSTEM SHOULD BE QUOTED
SEPARATELY AS OPTIONAL
Standards: ISO 9001, ISO 13485.
Certifications: FDA $10(k) approved /CE marked,
Manufacturers from other countries (other than USA,
) will be accepted if equivalent has FDA
510(k), CE, JIS , MHLW or JQAO certifications,
‘Anesthesia Machine with built in Ventilator
G for Teaching Hospital block 2 for Ortho 2 for Urology)
TECHNICAL SPECIFICATIONS
chine to administer anesthetic agents in precise
control and flaw manner.
# The machine should equip to monitor anesthetic agents during
‘operation,
= 3egases O2/NZOIAIR.
+ Provision of communication port for sharing and transfer of data,
Unit shall comprise of the following component
« Electronically/digitally control, mixing and monitoring of anesthetic
gases (02, AIR, and N20) bot by digits as well as virtual tubes.
+ Built-in illarina —s)
+ Non- interchangeable pipe
Page 28 of 146
— Scannec win Laitan
* Pipeline & cylinder gauges for O2, NIO and Alle
© Central gas! electronically driven wit,
ler yokes for Oxygen & N2O (One each), as
ylinders will be provided with the unit (2x02 and
2NN2O; BS standard)
* Gas ontlet and O2 flush contral
+ Lausiliary ©2 outlet
+ Two Lockable castors
* Fiber work surface
# Absorber bag support arm
« Integrated heated breathing system,
«= Thrce gas electronic digital flow meters for precise control and
monitoring of gases
+ Drawer unit 5-6" high. Mintmum three drawers,
*# Calculation of gas consumption during surgery.
*# Power outlet with 3/4 socket outlets to connect the auniliary
equipment
# COZ absorber 1,500 em or better with changeableduring the
surgery.
+ Complete with vaive for bap/veatilater, manometer
‘© Autoclavable Breathing tube (adult and paeds).
‘Additional breathing hose and connector (adult and paeds),
© Articulmed Arm
* Scavenging system active type.
+ Built in Suction system,
1 Sx water taps for agent monitoring
++ 5x Sampling line for agent and EtCO2 monitoring
ANESTHESIA VENTILATOR:
« Anesthesia Ventilator with minimum 15" or more LCD /TFT
touch Sereen.
‘© The ventilator shall be eapable of ventilating Neonates /pediatrié
ppatienis/Adult Patients)
‘The ventilator shall have following features as a minimum
requirement:
« Volume Preset Time Cyeled Ventilator (IPP Mode/ CMV)
‘* Manual, spontaneous; Volume Mode (PPV) (CMV
‘= Assisted Control Ventilation
‘= Pressure Mode (PCV)
«© Assisted Pressure Controt ventilation
+ Proportional control of minute venti
mades
« Synchronized volume controlled ventilation (SIMV) with PS
«= PS with apnes back up
® Breathing Mode Selection (Standby / Volume/ Spontaneous and
Pressure)
«= Byilt in Oxygen Monitor
# Inverse LE ratio Capability
«# Gas Specific Input Connectors (Air or Oxygen ISO or ANSI
Standards) (
+ Tidal Volume fio Sm! to 1500ml,
# Rate oF Frequency 4 to 90 bpm
« PEEP 2 to $0.cm of H20.
tion in all pressure eonteol
Page 29 of 146
OcCaneu WIth U‘+ Inspiratory Pressure Limit
* Pressure and Volume (Spirometry}) Loops / Curve.
* Oxygen / Electronically Driven
* Battery Backup minimum 100 minutes
+ Low (High FiO2 Alarm
«Incorrect Rate or Ratio alarm
* Mains Failure alarm
+ Low battery alarm.
‘+ Oxygen Senor: Paramagnetic
4 Lynne Device
se ventilater shall be supplied With complete drive hose and power
7M pp plete drive hose and po
Optional Items: (Mandatory to Quote)
* Two pre calibrated Vaporizers of Isoflurane & Sevoflurane
vaporizer (or by choice), temperature and flow compensated,
* Cardiac bypass mode / HLM / Spontaneous Mode.
* Imported Online UPS with backup of 30 minutes for complete unit
* Modular Vital sign monitor,
Size of minimum 12" touch screen or more for display of vital sign
Parameters of neonates, infants and adults.
Measurement of ECG, NIBP,SpO2,Temperature, Respiration
Provision of communication port for sharing and transfer of data.
220¥, 50 Hz operated. Battery backup of at least 9 minutes
ACCESSORIES:
2. NIBP Cuff for neonates, child and adults for each
2 Spot probe for neonates, child and adults for each
‘2 temperature skin probes
‘2temperature Nasal probes
2ECG Leads
Standards: ISO 9001, ISO 13485.
Certifications: FDA $10(k) approved / CE marked.
‘Manufacturers from other countries (other than USA,
Europe or Japan) will be accepted if equivalent has FDA
510(k), CE, JIS , MHLW or JQAO certifications,
Tew Na
Ie
Surgical Di thermy 300 with vesscl sealing facility
mn ith accessories
(4 for Teaching Hospital block 1 for Ortho 3 for Urology | for
Gyne)
Specification:
‘The energy Platform delivers both enhanced Maaopolar & Bipolar energy
wre eeeat geoerten af vessel sealing And Cerue Sion and taut
Sensing technology all in one unit compatible with Argan Beam Congulatar
System, Systems also have Bipolar Resection capabilities. The system
cludes 02 ladependent Monopolar outputs, O1 separate Menopolar fact
switching port, 01 bipolar part Awto-bipolar with Current meter (ammeter)
Sand Of esse! scaling ports with fost-svtching or handeswitching option.
. ‘Output Characteristics
Rated Lond Rated Ovtpat
Monapolar CUT
PURE 300 00 a0
BLEND: x 200 1100
yn te
Page 30 of 146
s
OcCaINeu WIL U-Monopolar COAG
FULGURATE $00 120 313s * ibe
SPRAY 500 0 3578 ay
‘SOFT 100 120 240 Ta
Bipolar
LOW (1-15 Wp 100 B a
MEDIUM (16-40 W) 100 aD a3,
| igi 395 W) 109 95 310
‘Vensel Sealing 20 30 163
Bipolar Resection
cUuT 300 35 495 :
COAG 109 198 212
Main System with following accessories:
Description
Generar
Monopole footswitch
Bipolar footswitch
PolsHlesive REM Cordless
Maryland Jaw 23cm
Mayland tow 3¥om open shew
Small Jaw festrument
m Clamp/Cord Assembly
‘Safety Standards
(Class F-Equipment (LEC 60601-1)
‘Type CF Equipment (IEC 60601-1)/Defibrillator Proof
Liquid Spilage (IEC 6601-22)
lectromagnetic Compatibility (IEC 60601
Few Ralienent
a. Line Voltage: 220-240 VAC ‘
tb, Line Frequency 50/60 Hz
‘Standards: ISO 9001, 180 13485,
Certifications: FDA $10(k) anproved / CE marked,
Manufacturers from other countries (other thas USA, Europe or
Japan) will be accepted if equivalent has FDA 510(k), CE, JIS »
2 and IEC 0601-22)
Trem No.8 Fy
(4 for Teaching Hospital block 5 for Ortho for Medical Unit Ill
6 Peads Medicine 3 for Urology 3 for Neurology 2 for neuro
surgery)
Specification:
vi 15" Touch Colour LCD Screen. Display 1S" Color TFT- Resolution
fr 1024X 768 pixels w Intelligent audio, three level, and Visual, Red
Led, comprehensive alarm Multischanne! waveform display.
Portable operation, multi-prongod. Full-screen touch, fingertip
operation, Fast button backlit design for easy night operation.
Ergonomic knob operation, elassie and flexible. Plugin & use, '
powerful 3 module slots, easy to upgrade, » Flexible combination,
Fesource sharing, Global gold standard Niassimo SpO2, must ensure
ow perfusion measwrement, Sun tech NIBP blood pressure, for
accurate blood pressure measurement.
‘Thermal Printer
Method: Thermal det array with 3 tracks » Paper width: $0 mm with
Paper Speed of 12.5/25/50 (mmisee}
System output '
Ethernet Network standard RdS socket; Defibrillation Out. BNC :
‘connector USB 2 sockets; Video Output 1VGA (Option); Nurse call
1 Rj. connector =
Page 31 of 146 ie <
é SS
/ i
oOcdiineu WiData Storage & Review
‘Trend: Land trend: 168h, min, resolution is mia (store when power
goes off J ai
NIBP meassirement reviewing: 1000 groups i
ARR event: 128 groups of ARR event and the assccieted waveform i
‘Alarm events: 128 groups of parameter alarm events and associated
parameter waveform at the alariy moment
Full Disclosure waveform: 96 hours for 3 waveforms {with 4 SD
ant)
ECG .
, Hl, aVR, aVL, aVF, V1-V6 (option) ‘
Gain selection; AUTO, 0.25x, 0-50x, 1.0x, 2.0%, 4,0x |
Sweep speed: 6,25mmis,L2,smnw,25muns 50mm
Heart Rate:15-300 bpm —HR accuracy: + | %
ST segment deviation analysis: -0.8-+0.8mY ST accuracy: 40.02mv
NIBP (Non-invasive blood pressure)
Method: Automatic Oscillometry
Displayed parameters: Systolic, Diastlie-and sean arterial pressure :
Measurement type: . ages
‘adult, pediatric, neonatal i
Operation Modes; Manual, Auto and Continuous .
Meastirement range:
Systolic 4.0 - 36.0 kPa
Diastolic 1.3 - 33.0 kPa » Mean 2,635.0 kPa
Accuracy: + 0.4: kPa or 5 % Resolution: 0.1 kPa
Protection: Hardwarc and SW overpressure protection
SpO2, Masimo SpO2 (Optional)
Display type: waveform data
Measurement range (0 to 100) % t
Measurement precision: (70 to 100) %: +2 %
SpO2 resolution 1%
‘Temperature (Rectal & Surface)
Method: Thermal resistance
Number of channels: 6
Displayed parameters: TI, T2 and AT
‘Measurement range: 0,0°C w 50.0°C, Resolution: #0.1 °C '
ETCO2 Side stream, Mainstream, ETCO2 Micro stream "
(Optional)
Measure method: infrared spectrune
‘Warm up lime: Caprogram displayed in bess than 15 seconds
Measurement Range: 0 - 19.7%(U-LS0mmitg)
Resolution: 1 mil fg, Unit: % mmHg, kPa
C2 Accuracy; 0—AOmmbg,2296,101-150 mmHg, + 10%
IBP (Invasive blood pressure) z
Method: Direclly invasive pressure measurement :
Number of channels; 6
Pressure readings: Systolic pressure, diastolic pressure and mean
pressure .
Pressure labels: ART, PA, LAP, RAP, RAP, C¥P, ICP, P1/P2
Measurement range:(-i0 to 350) mntlg
Dynamic Accuracy: £ 4rumHg or 4%, whichever is greater
Cardiae Output Method:
‘Thermal dilution
‘Measurement paramfters: C.O., TB, Tl, C:———
Page 32 of 146 ‘
OCaneu Witfin
U2
TT Adee aoe: 28.0806 Who i Hl ito
ceontinuonsty
Resolution:
Calibrations
adapter (<$ see)
‘Warm-up tiie: 10, tll neetraey wil [vi
fare measurement tare: 18th
RESP
Method: Thorneie impedance
Lead Selected fiom: | (RAsLA) neh (KALL ) Hetil:
Gan: 0.25, x1 x2 xelothmaloia: (25 HM am th (ll)
Sweep spocdt 6.25unmi/s, 12,51
Measurement Range: (e150 1p
Apnea Alam: 10s, 138,255,
Standard configuration:
15" Touch seteen digplny, 1 Stamnlon
ECG, HR, KI
socket, , 1 VGA port, 21
FSO 9001, 180 Lads
‘Certifications: FIA 510(k) approv
‘Manufacturers from other eountrtes (0
Europe or Japan) will be neceted if ey
510(k), CE, JIS, MITLW.
6 Haun aly cl tf ese efoto alain
HeLa
TE
Tem Nao
Surgical Ceiling Pendant Single Arm
Specification:
Aluminum Extrusion Profile
‘Arm length 800mm !8(00
Load Capacity |20kg or more
Brake System
‘Column Length (OEM) 600m
Gas outlet — 6 (02, N20, Air
Socket a
Teeminal box 400mm
Shelf dimensions (WD) 5!
Shelf load capacity — 70ky,
Rotate 340°
Drawerlight Yes
Background light Yew
20000
Air’? & Vawinunny
IV Pole capacity Sky.
Standard configuration 4-gas vutle ¢
IV pole, | basket
Standards: ISO 9001, 180 LMHS,
Certifications: FDA 510(K) approved f
Manufacturers from other countries (uth
Europe or Japan) will he wevepted if
510(K), CK, JIS, MUL or QAO certiflent lone
INA,
Page 33 of 146
V0 27.04 Meenilinty U0. EE
i ane 13 valves: CO2 NAGA AKL HAT INO), HIVL AVNE FAM)
A tin HEYA
whe, | obo I eins | ! ill
il
‘
OCaiINeu WitTrem Nox
ottles
for Teaching Hospital block 2 for Urology)
Specification:
antistatic wheel with 2 brackes option
2 jars of 2000 liters of each
‘+ Motor Oiless and maintenancesfree piston pump
Power Feeding 230V-50 Hz
180 10079-1Classification HIGH VACUUM / HIGH FLOW
‘Max free air flow rate 60 Vinin
‘Max Vacuum (adjustable) -0.90 Bar -00 kPa -675 mmHg
‘Noise level 51,7 B
‘Power consumption 230 VA
‘Operating Theatre suetion units
Fuse 1x F4.A 250 V
* Duty cyele Non-stop operation
Standards; ISO 9001, ISO 13485.
‘Certifications: FDA 510(k) approved / CE marked.
Manufacturers from other countries (other than USA,
Europe or Japan) will be accepted if equivalent has FDA
510(k), CE, JIS, MHLW or JQAO certifications
tiem Nat
‘Suction Machine
G for Ortho 2 for Neurology 4 for neuro surgery & 1 for Peads)
Specification:
‘Vacuum 0 100.85 bar /0 to 650 mmifg
Flow rate 40 Wmin
Collection Jars 2 x 2 litre
Noise Level S0dB
Voltage 230v +/- 102%~ 50/6012
110y +/- 5%~ SO/60Hz
Moveablé Gn five wheels
Or Equivalent
Standards: ISO 9001, ISO 13485.
Certifications: FDA 510(k) approved / CE marked.
‘Manufacturers from ather countries (other than USA,
Europe or Japan) will be accepted if equivalent has FDA
§10(k), CE, JIS, MHLW or JQAO cet zations
10
‘
(
ome
joor mounted air cleaner with
‘G for Teaching Block }
‘Specification:
Tntegrated 6,8" Touch Soreen os better
Multi-functioas contral panel with LCD touch seréen.
Plug & Play system
Bactericidal, veridical, fungicidal (including spores) actions
and molecular decontamination
* Decontamination kinetics CP1O(particles 0.54)
Bacteriological elass MS/BS
Particular class ISO 8 / 1SO-7/ 180 6
microbiological reduction: up to 99.99% ina single pass
very low sound level: 42 dBA at 900 m3h
Device capable of running 24/24hrs, 7/7 days
‘Air flow speed adjustable up to 1200 m3/h
Mobile ss
Soosiopernon_———
a” Easy to move by a
Page 34 0f 146
OcCaiIneu WitTECHNICAL CHARATISTICS
Ar Row 300 -1200 m3¢h (with
‘constant ar low regulation)
ir supply ‘Vin plenum
Mobility 4 wheel
+ Costco! panel Multi-function touch soreen
Pcae cot 740x500 x1550 mm
ie inake filtration G4 + £7 (low pressure dro
filter made of polypropylene) =
* Air supply filtration 114 (low pressure drop filer made
‘of polypropylene) ~ single ur double stage
Photo catalysis module Photo catalysis lamp
Probe VOC, temperature, humidity Probe E4000 at air
intake
Particular ProbeProbe P4000 at air supply:
Pressure probe ir intake, air supply , Fan
Remote control touch pad 7"WiFT (optional)
Internal structure “double skin" galvanized stee]
Panels
External siructure —_Thermoformed panels
Electrical power 120-230 V/ 50-60 He,
= Interfoee language French /English / Spanish /
German / Chinese
Power consumption 450 W
‘Air flow (day sajght! auto/ manual) in m3/h
Humidity
Temperature
Level of VOCs
Particulate concentration
« Alarms on all points (probes, fiers, fan...)
‘Maintenance menu (date of filters changing. of other type
of maintenance operation)
* Secured information with access code
Standards: ISO 9001, ISO 13485.
Certifications: FDA 510(k) approved J CE marked,
‘Manufacturers from other countries (other than USA,
es
fe Europe or Jap2n) will be accepted if equivalent has FDA
510(k), CE, JIS, MHLW or JQAO certifications
Tem No. 13 ical gas pipeli i i
‘Medical gas pipeline system installation in 2 x Operation Theatres
‘and up-gradation of existing system:
‘The pressure of gas in collectors (150 bar —MPa—2176 psi i
reduced to 4.7 bar that is appropriate pressure to provide for patents
‘and medical devives through pi
Technical Specifications
Input Pressure: Max. 220 Bar
(Output Pressure: 4—7 Bar
Capacity; 150m*/h
Input Pos (Main ramp, Backup Ramp, External Supply,
Emergency Case Supply Unit
Operating Temp.: (+10) - (#45) C*
Power S
Page 35 of 146
ocdiineu Wi‘Alani System: Digital LCD, Alon, Sound i
MA IFOLD FOR OXIDE: (ar
‘The pressure of gis in collectors (150 bar —MVa—2176 psi) is
reduced to 4-7 bar that is appropriate pressurs to provide for patients
and medical devices through pipetine systems,
‘Technical Specifications
4 (Main rainp, Backup Ran
Emergency Case Supply Unit
Operating Temp. (#10) - (48) C*
Power Supply: 220°V $0 Hz 100 Watt
Alam System: Digital LCD, Alarm, Sound
Medical DUPLEX COMPRESSED AIR SYSTEM (02 Separate.
sompressors)
Technical Specifications:
‘Number of Compressor: Min. 2 i
‘Compressor Capacity: 40 m3
Control Panel: Operating (Compe) hour equalization by PLC
Tank Capacity: SOOLt
Operating Temperature: (+19) - (+45) C°
Power Supply: 30, 380VAC, 50 Hz
Dryer: 2 Pes.
Filter Group; | Pes with By-Pass System
Regulation Group: 4-7 Bar Regulators :
GAS OUTLETS (BS SYSTEM)
Gas outlets, complete in Bondeasing gas Specific, Self-seating chock
valve, indexed to eliminate interchangeability (erroneous tapping of
‘gas, Filter and cover plate as pee Standard Design
requirements of the bidder with following colour eading:
© Oxygen White (50-n05,)
+” Medical Air Black (50 Nos.)
+ Nitegus Oxide ‘Blue (12 Nos) 4
Colour coding. BS type. Internal size for pipe connection will not be |
Jess than 10mm
VENTURE SUCTION REGULATOR (30 Nos) |
‘System with Gauge and injector for connestion to the-wall source of
‘Air pressure gas using direct probe with suction trap to prevent
overflow of waste fluid.
‘Vacuum capacity 0 to 760 mun/fig adjustable. ‘1
Adjustable knob to control the vacuum directly . Hy it
Collection jar? Liter (from Same Manufacturer) graduated, Li
unbreakable with overflow safety
ox FLO
Flow meter for O2 Therapy with regulator.
Adjustable regulated flow from 0-15 Limin.
Bubble humidifier with autoclave ablet unbreaksble bottle |
(Complete set),
Safety valve in case of pressitre increase |
Standard BS to be attached
Locking mechanism for safety.
GAS AL,
Page 36 of 146
ocanned WIT GdlTiocaVisual and audible
Individeat display and sensor module,
mitestie ® eror message dipay readable free of
nance psi, KPa or bar read out, Test d Mut ut
alarm condition. LCD serven, seh
ARM FOR WARDIOT (4 no
sta nd ici mourned on wal inside he War,
splay of pressure High/Low pressure 00 hi
Test & Mute ten, LED scene on 2 ih Bow.
VALVI x FOR TI
VALVE BC X FOR TWO. GASES (4 nos.
OPPER PIPING: te per fe ig
Supply and instalation of seamless medical grade copper pipe,
eoxidized and degreased along with required fitting ec, various
sizes (diameter as per Drawing with matching color indications of
‘ut lets. The sizes will be 2mm, {Sim 22mim depending upon the
design.
12mm fet
‘1Smm 0] feet
22mm O1 feet
n i it
Bead Head Unit:
IZANT. IVE CARE UNTT (Aluminum Rai
‘anel di
> Oxygen Poinis 02 Nos
> Aid Points 02 Nos
> Electrical Socket 08 Nos
> Barthing Nodes; 04 Nos
IGH PRE: ICAI HANDLE v,
Sizes in mm 1215.2
Imported with complete capper/brass fittings
Standards: ISO 9001, ISO 13485,
Certifications: FDA 510(k) approved / CE marked,
Country of Origin: UK, USA, Japan, Turkey,
OR Equivalent
Tem Noid
CSSD (OT)
(for Teaching Block }
rac
CSSD Machinery, Equipment, Furniture & Instruments
a) Steam Sterilizer 6STU Double Door 485L, or more (2 units)
« Autoclave should be fully sutomatic, programmable,
microprocessor type.
© Chamber capacity of 485L Or more rectangular shape
# Minimura 8 sterilization cycles including Bowie & dick test and
‘vacuum test
* High pressure Stcam Sterilizer with extemal steam supply
connection facility
/# Integrated steam generator as backup.
* Closing, locking and opening of door should be completely
automati¢ and through touch sereen
}* Automatic ie double door pass through system, Chamber
and door valves should be pneueiatie:
-* Built in Printer, Réal-t
Page 37 of 146
ocailineu‘Time cycled, working pressure 32 pas
Autoclave should be equipped with door safety locks and safety
valve for Jacket, chamber, and steam generator
‘* Built in touch séreen of 7.0" or mare which display current cyele u
Phase, temperature and pressure, Program/Cycle selection.
** Monometers for Jacket, chamber, and steam generator
** Emergency stop push button
* Complete with standard accessories and removable shelves,
* Capable of taking both packets and containers of all standard sires
© Chamber, Jacket and doors made of AISI 316 UT, The system
‘complete with built-in water saving system
* Chamber electrolytic polishing; z
* Automatic heat exchanger
© 3 level water indicator
© UPS of suitable Sapacity with minimum 13 minutes for Controller
Sum display for monitoring and controlling of parameters during
Power shedding provided! installed by the manufacturer.
Accessories;
# Removable shelve: 02
+ Loading Cart: 01 ‘
+ Loading/Unioading Trolley: 02 '
b) Washer Disinfector 10DIN 300L or more (2 waits)
Fully automate, microprocessor controlled, washer disinfector.
‘Processing cycles should be faciory programmed with recommended
‘reatment parameters for pre-rinsing cleaning, post-rinsing, thermal
disinfection, final rinsing and drying.
Features
= Rotary spray arms should be fixed at the top and bottom of the 2
chamber.
«washing up to 95 °C - antibacterial thermal
# capacity of 12 DIN baskeis
|* programs: 20 default + 20 custom
+ multislanguage TFT colour towch screen on both sides
= functioning protected by different levels password
© integrated built-in panel printer
‘© USB port for downloading washing cycles log .
«2 sliding dew fill glass doors
+ doors interlocked for complete separation dicyiclean side
+ spray arms, one an the tap and on the bottom of the chamber
« high pressure hot air drying system with absolute 99.9999 HEPA
‘ilters
+ high efficiency Eeuslim stean condenser
+ peristaltic pump for liquid detergent and level sensor
+ peristaltic pump for liquid ncurralizer and Jevel sensor
« possibility of installing up 2 additional peristaltic pumps z
« detergent cabinet with removable tray for storing up ta 3 x $ liters
cans + 1x 1 fiter can
+ washing chamber made of AISI 316L stainless steel
+ external panels made of AISI 304 stainless steel
* cold water, hot water and demineralized water connections
* floordrain connection:
* drain pump (optionaf) _ :
+= clectrical connection: 3/QVPE 400 V =50 N= 1 kW thax. ~
7
infection
Page 38 oF 146
ovarieu Wi+ 60 He version available without racks |
* Compete with $ level wash cart, transfer tralley ad 10 DIN trays
6) Bench top Steam Steritizer 231, (1 anit)
# Poetable Horizonlal Steam Sterilizer
* Water feeding system with iter tank sensor to eicck the waler
quality
* Stainless Stee} ‘Chamber of 304/316L,
* Tenipersiure Range pte 134° C at least
+ Built-in thental printer
+ LCD display
+ Fully automati, programmable microprocessor type
> Capacity 23 liters
+45 pisce of stainless steel/Aluminum trays compatible with the
‘steam sterilization,
* One piece of holder for unlonding of trays.
‘lectrically heated steam generator.
* Connection te external data logger
* Standard programs should include:
+ Test programs: HELIX TEST - B&D TEST, Vacuum Test
e + 121°C Hollow Wrapped
* 134°C Hollow Wrapped
121° € Hollow Open
+ 134° € Hollow Open
© 134°C Prion
© 134° C Porous
d) Ultrasonic Cleaner 25L (I unit)
* Ultrasonic leaner integrated type,
* Cleaning chamber constructed of corrosive resistant 304 8S,
« Heating thermostatically eontolled, :
+ Digital display of time and temperature
city 25Liter.
« dyrtan compl i bask LD and other sandr cessor.
* Operation 220¥, 50 Hz.
scaler (1 unit) .
aL ier isnt ROR
of different materials.
= Automatic start & stop by phatocell
> eereating automate 309 (250°
J» Pro-programmied temperatures:
'# Temperature range: 20-230C
= Sealing width: [5mm appr
= pressure range adjustable: 40 -100 N
'= Standby time (adjustable): 1-60 min
'» Heat-up time to; 180°C 180s
= Speed range (adjustable): 120-170 man/s
ae deste Appin 20H £70820 mn (WDD
ae eo Sule Stel quran TbleTop 4050-0
tesa, ‘made of Ni-chrome stcel/SS sheet. 02 No's Sinks each
approx. 500%500%250 mim made af stainless steel wis es meriien
Shes Op Sa secol selve Catpe i rechrame
sme amit rilegrated Spa for
Page 39 of 146
ocdiineuematerialized waiee? air with 08 cleaning nova
throegh Space to hase he Olultnsoieeteners ee
8) Sterile Materia! Transport ‘Trolley 8S (1 unit)
Senses
~ Storie Materia! Handlin Sontaines Capacity
alters pad Tale ntainer Capacity
‘) Linen Inspect ote SS (1 unit)
ding Table Wilh Light
~ 140870985 em Approx,
i) Teck Table SS (1 unit)
+ Packing Table With Drawer, Roll Han "
= TUaI4D90160 em Appron LMeR aM 2 sets
3) Paper Storage Trolley $$ (1 unit)
~ Sthitiless steel
> 408120390 cm Approx.
‘ky Paper Holder with Cutter SS (1 unit) an
+ Rotary Roll Cutter, Double, Manvel 4 a
1) Single Face 6 Baskets Transport Trolley SS (1 unit) oe
+ Stainless steel
~ 56 60 x 170 cm Approx.
m) Transport trolley SS 3 Shelves
+ Container Transport tralley, 3 Shelves
= 62x66v90 em Approx,
2) Inspection Lamp with Magnifying Leas(1 wnit)
6) Tape Dispenser (1 unit) §
~ This dispenser for sterilizer tape holds two reels of tape
= Made of high quality coated stee! for long use
p) Storage Rack MS (2 units)
~ System, Modular construction, adjustable shelves
~ Solid Shelf System
4) RO Plant 1500 GPD (4 unit)
+ Locally assembled with imported membranes
~ Tank capacity 1000L or better 4
= Automatic operation
+ TDS level digital display
+) Air Compressor (2 unit)
+ Fully automatic oil-free compressor
© Single stage type, two-cylinder compressor powered by an
electric motor mounted on air receiver
= Capacityfpeessute 196 L/min at 7 bar or better
= Compresser speed 1400 ¢pm or better sand
Supply of goods complete with installation including electri
plumbing works to install machinery and equipment within the
ssp
OPTIONAL 15!
1- Low Temp Plasma Sterilizer 110L (1 unit)
Display: 7* touch screen
Chamber Volume: 110 L
> Sterilization temperature Average: up to 60°C
Structure and panels made of stainless steel .
= Recognition of recharge bottles by RFID
= Sterilization cell: Rectangular, 3167304 L.
=
= All tubes, wins angel
Indication of the sterilizis
Page 40 of 146,
OCaineu WILT~ Application fer the
cycles! phases om hi
sereen
Mobite/Portable
~ Automatically operated single sting door.
+ Safety Syste avoiding pinching during dave closing
Operator ofthe Sterilizer to see in real time the “T i;
is mobile phone, as a duplicate of the sterfizer
~ A Standard 1 thermal printer mounted inte the fascia en the
Control Side
~ Chantber heating through electric heaters .
+ PC cannestion port for direct nscess to PC * ae
= 2 loading shelves bie
~ As intake Through medical-grade HEPA filter effisioncy’ upto .
O.2j0m
~ 202 as stetlizing agent
~ Independent temperature control system for each element with
probe
Alans: temperature and pressure sensor fhilure, excess moisture
alarm, excess cellulose material alarm, Injector malfunction,
concentration preparation alarm, s
~ In-line catalytie converter to avoid leakage of residual H202
~ An efficient catalytic converter to ensure of no residual H202
leakage from the ster
‘The standard program to program to sterilize general surgical
instruments with single ehsnnel flexible and moderate-length
figid lumens, The fast program should provide shortened
sterilization time for surface sterilization loads, The advanced | ip
[program to sterilize surgical instruments with long rigid lumens | ei
2 Drying Cabinet (1 unit K
A drying cabinet for anesthetic equipment and instruments in
standard trays, Should also possible to dry laboratory equipment
and other wlensils
‘The hoses should hung cassettes during the drying process,
enabling water to drip aff easily.
= Should process up to 36 hoses in the same drying cycle. 4,
= The drying time ean be selected ina range from 0 10 90 minutes ‘ et
= Capacity up to 18 DIN tray OR 6 AN-casscttes with 36 anesthetic i
hoses OR 2 AN-casscttes with 12 anesthetic hoses and up to 18
dards: 0 9001, $0 13485
itandards: ISO 9001, .
Certieations FDA 510(k) approved / CE marked.
Manufacturers from other countries (ather than USA,
Europe or Japan) will be accepted if equivalent has FDA :
510(K), CE, JIS , MHLW or JQAO certifications
Trem No
eP Raa (00 mA baie T
{for Orbo Deparimen!)
‘Technical Specifications: .
ee oes Battery Operated Mobite X-ray system For all
General-Purpose X-Rays.
* High Frequency inverter High Voltage Generator
Maximum Inverter Frequency:20KHz or more
Input Voltage: 220 volts single Phase «
Tube Voltage Range: $0-125kV at 1 KV step
mAs Range: 00)
300 ,As oF more
Maximum Power: ore geen
Anatomical Programs:_35 oF mi
Page 41 of 148
ocadiineu* Main Mobile Cari with Telescopl
Power assisted Optimization system
Support Column
‘Arm Rotation : “tie 270 deg
Collimator Rotation: He 110 deg
Narrow Widih: Stem
Arm Horizontal Travel : 120emn
* Xray Tube Unit
Anivle Heat Capacity; 300KHU
Focal Sp 6.Tonm
* Collimator LED collimater lamp
+ Hand Swviteh for Exposure
* Cassettes Contpartinent (05 cassettes 14"'x17" or mare)
Standards: ISO 9001, ISO 13488,
Certifications: FDA 510(k) approved / CE marked,
Manufacturers from ather countries (other than USA,
Europe or Japan) will be accepted if equivalent has FDA
$10() AND CE certifications
ters Nowe
Page 42.0f 146
‘Computed Radiography (CR)
(for Oho Department)
Specification:
> READER (INoy
* Diroet Digitire
Should be capable 10 read X-Ray exposed Imaging Plates (IP)
‘of all standard sizes in incheslem.The productivity of
reading/digitizing should be minimum 60 IPs/Hr of 14x17"
amy itch:
12 Types: 87.5/175 micro meter or better
* Digital Gradation Level
4096 Levels (12-Bit) or better
* Processing Capability
_, Ae SE Pikes Aa aT a tae
ass ‘Tim
" Rppron: 45 seconds 4" x 1271 7Sar or ss
} IMAGE SOFTWARES & CR CONSOLE:
© Software should support all the standard connectivity from a
high quality console, such a DICOM modality work list,
DICOM Print, DICOM Storage which gives fill control over
image quality and allow each exam to be Tully customized,
«# Body part selection screen.
+ Easy list switching by tab (Study/Exam.),
sy viewable status icon of study
“Te Image Processing:
Zoom infout
be Seumodify ROI
cc SiG value
d- Rotate/Fl
ce Masking
f Cropping
a Markers: Pre-defined text (e.g. LR, AP, Lat,)
b+ Stamps: Dynapfic text elements (e.g. Patient Name
Annotations:
ocarleu
fe
win varniscaqe
Multiple
i
ased images con be laid oat ws film
nul enoy
by moving berilers
& Print Composer:
Print composer and nnalti-sily cam fe used! at the sine thine,
Images from different stidics of the same. patient can be
pritited on one film,
S Study Sharing
‘Shauld have capability that Multiple Consoles should be able
to:
a Share Examination Information
be Send/Receive
© Perform Exposures
ad VerifyiCheck images
6 Tube/Gauze Enhancement:
‘Tube/gaure enhancement to be applied to the following exam
tags:
a Chest/Abdomen APYLAT (infant and adult)
‘be Skull APYLAT/OBL
7 Settings:
a Institution Information
b+ Screen Layout
a. Paticnt Registration Sercen
‘b. List Sereen
t. Quick Tools
c= Prodefine Overlays'Annotations
a Institution Information
e Seren Layout:
1, Potient Registration Screen
List Seren
©. Quick Tools
£Predefine Qverlays/Annotations
gr Exam Tags:
a RIS Code Mapping
be Exposure Conditions
Properties
4 Body Part Selection Seren
a+ Hospital Information: E.g. Define User logins
to CS-7
he Backup/Restore: ae
Backup and Restore device settings, patient information
and logs
i. Analysis: _
a QA/QC tool for evaluation and managing image
quality
bbe Statistics Analys
c Eg. Exposure St
& Ooi Ma gement: Raw data output
8 LCD/TET Monitor 21" (10 be supplied locally).
LASER IMAGER CINa.)
Laser Source! jemicondustor Laser
Film Size; 14°17" 5843 em),
14" (35
It “Ganisemt
Page 43 of 146
ovamieuTONITE G5xi0 em),
S°XIO" (20x25 cm) selectable,
Fil Dry Image Recording Filey
Image Memory: Print Memory 2 GI or more
Pixel 7A jim ov better
Image Data Inuit: 8-Bits/ 12-Biee
Meoeessing Capab IS-sheets (hour or mare
Provecol: DICOM Print Management
Multi range formats 1.35 or more
Input Interface Ethemet 10/1 00/1000 Base-T. '
Integrated DICOM interface supports printing from DICOM print
thedalities Should have an option to connect non-DICOM
modalities using Print Link 1V.ID
‘+Casscttes & Imaging Plates:
1417" G5 X Adem) (4Nes)
10 12" (24 x 30cm) (4Nos.)
DICOM 3 CAPABILITY.
Dicom 3.0 Capability for send, receive, archive, retrieve and print. .
TRUE SINE WAVE ON LINE UPS SKVA FOR CR SYSTEM
NOTE: Vendor quot ing CR system must have exclusive distributor
Ship'sole-ageney for at least § years and should have installed at
feast 50 units all over Pakistan,
Certification : FDA S10k/CE/IS/MHLW
«Table Top CR System,
® Touchscreen operation to ensure-quick and efficient work and
smooth workflow z
Compact desktop unit (0-30 m? footpriit)
Professional image acquisition saftware, the CR system offers
all necessary image processing tools.
+ Multi frequency level analysis Analyses each image on various
frequency levels for ideal sharpness and high subtle contrast,
Automatic noise filter Algorithm for optimal noise reduction.
Can connect with any DICOM printer and paper printer.
‘Sampling Pitch: Types: 50 micrometer or better, k
Processing Capability: Speed up 10°70 plates / houe of more.
Resolution up ta 10 pixels / mm or more,
ICOM Printer ‘
Excellent reliability, minimum maintenance
Direct Digital Imaging technology
Convenient imaging with two media sizes-on-line
Easy doylight loading ; -
Technology: Photo thermo graphic (dry laser) : a
Printing resolution; Geometrical: 320ppi. |
Contrast: 12 bits contrast resolution,
Throughput: Time to first print: 77 seconds
Up to 70 films or more per hour: L4 x 17 in.
Upto 185 fins or mare pe hor 8x 10in,
forage Cay 100 or more sheets.
Fae reine ‘Auto ranging 100 to 240¥; 50'60 Hz
‘Accessories:
© Cassettes, Size 14x17 ingh—
«X-Ray Pils of 125 shed{s of {4 x 17 inch, ——— >
fiee tee eee eh pues ae
Country of Origin:
USA / Europe / Japan.
Page 44 of 146Power Dri
(4 for Ortho Departiient)
Specification:
Elcctric Console 230 Volt, With Universal Cable!
Capable of operate 24 sifferent hid pieces,
2 Hand pieces operate able simultaneously with built-in pump for
ivigation, Both Hand & Foot control.
Use in. Arthroscopy, Large Bone Orthopedics,
Electric Modutor Hand Picce |
For Drilling, Reaming, Wiring, Serewing, Sawing eto,
$ Speed :0-1250 im, Torque 30 in tbs. or better
{ VSJACOBS CHUCK ATTACHMENT 1
* AO SMALL ATTACHMENT1
SAOREAMER ATTACHMENT — |
SDHSREAMER ATTACHMENT — 1
{REAMING ADAPTOR HUDSON |
{HIGH SPEED DRILL ATTACHMENT 1
UNIVERSAL CABLE I2FT/3.6METERS 2
{ Wire Driver Attachment Cannulations 0.7 mn-1.6 mm dia, 1
7 Pin Driver Attachment, Acceptance of Wire Sizes (1,8-4.mm) 1
* ELECTRIC OSCILLATING SAW HANDPIECE 1
~ One-hand collet for easy loading of blades
~Four position rotating collet
- One hand safety switch
BLADES FOR OSCILLATING SAW (DIF. SIZE 3 BXS)
* ELECTRIC RECIPROCATING SAW HANDPIECE 1
~ Maximum Variable Speed 1,000 epm or better
BLADES FOR RECIPROCATING SAW (DIF, SIZE 3 BXS)
STERILIZATION CONTAINER 1
Standards: ISO 9001, ISO 13485,
Certifications: FDA $10(k) approved / CE marked.
Manufacturers from other countries (other than USA,
Europe or Japan) will be accepted if equivalent has FDA
$10(k), CE, HIS , MHLW or JQAG certifications
ay
Tiem Hos
fe
Power Saw
(@-for Ortho Department)
Specification:
+ ELECTRIC OSCILLATING SAW HANDPIECE 1
= Mode of Operation: [PX0 (Ordinary)
= Speed Range: 0-11,000 cpm or beiter
+= Oscillation Range: 4.5° of better
= One-hand collet for easy loading of blades
~ Four position rotating collet
= One hand safety switch
BLADES FOR OSCILLATING SAW (DIF. SIZE 3 BXS)
Electri¢ Console 230 Volt, With Universal Cable
Capable of operate 24 different hand pieces,
‘2 hand pieces onerate able simultaneously with built-in piamp for
irrigation, Both Hand & Poot control,
Use in Arthroscopy, Large Bong Orthopedics,
* 1/4 JACOBS CHUCK ATTACHMENT
* AQ SMALL ATTACHMEI
"AO REAMER ATTACHM!
1
Page 4S of 146°
Ocadilllome
*DHS REAMER ATTACHMENT 1
*REAMING ADAPTOR HUDSON |
*HIGH SPEED DRILL ATTACHMENT — 1
UNIVERSAL CABLE 12FT (3,6 METERS 2
* Wire Driver Attachment Cannulations 0.7 mm- 16mm dia, 1
* Pin Driver Attachinent, Acceptance of Wire Sizes (1.8-4.0mm) 1
STERILIZATION CONTAINER |
Standards: ISO 9001, ISO 13485,
‘Certifications: FDA $10(k) approved / CE. marked,
Manufacturers from other countries (other than USA,
Europe or Japan) will be accepted if equivalent has FDA
510(:), CE, JIS , MHLW or JQAO certifications
Trem Nols
Power Plaster Cutter
@ fer Ortho Department)
Specification:
Plaster saw.
1 Saw blade @ 50 mm.
1 Saw blade @ 6§ mm
4 Pair of Wrenches
‘Technical Specifications:
High oscillating speed 24000 CPM or better
Low oscillating speed 600 CPM or better
Sound level $2-90 dB(A) + SdB depending on the adjusted
‘oscillations
‘Connecting cable 3m
‘Double insulation Safety classification Il Type BF
230 Valt, 50 cycles $00 Watt
Fully automatic toad compensation over the entire oscillating range
Motor protected sgainst dressing-generated dust Double insulation
‘Safety classification I] Type BF
Compliance with 89/336/EEC directive EMC (electromagnetic
iy)
‘The Plaster Be dressing saw is ideal for the removal of plaster-
and synthetie dressings due to its strong motor and quiet motion,
‘The Cuiter saw is equipped with a steplessly adjustable oscillating
speed, thus allowing the spsed of rotation to be adapted to suit
actual requirements. Together with the quiet, low motion this.
especially beneficial in case of anxious patients, Air entry atthe end
the saw reduiees dust aspiration into the saw moter.
Features:
Solid construction; High capacity during permanent use,
Low vibrating; Automatic overload protection;
Maintenace free; Low weight .
Ergenémie handling: Continuously variable adjustment of
oscillations; Motor protected against dust of dressings.”
Dust Extraction Unit Vacuum eleaner 230V
1 Vacuum cleaner (without plaster saw) incl. suction tube, support
device for suction tube and support plate for plaster saw, T suction
nozzle to connect the suetion tube to ‘the plaster saw, 1 HEPA filter,
| perfume stick, 1 paper bag
ron, Tank capacity: 121, Weight; 12 Kg Hei
x, (paper bags, 6 perfurne sticks)
R
INSTRUMENTS FOR PLASTER
Technical as
Stille plaster shears 23em,
L
Page 46
of 146
. " it
Ocdill ein LadltlocdLister bandage selssors Ifem
Lister bandage scissors 20cm.
Lorena bandage seissors 24cm
Brus plaster shears 2.4em serrated
Lister bandage scissors 20em large ring
Hewiig plaster spreader 27em
plasice spreader 23cm Mod USA
Wolff Boller plaster east breaker 27.Sem,
Esmarch plaster knife {em
Engel plaster saw 15cm
Standards: 180 9001, ISO: 13485.
Certifications: FDA 510(k) approved / CE marked.
Manufacturers from other countries (other than USA,
Europe or Japan) will be accepted if equivalent has FDA.
SLO(K), CE, JIS, MHLAY or JQAO certifi
3
Trem Nea
€
duutoclave Electric
2 for Ortho , 1 for Urology , 1 for Neurosurgery
Autoclave Electric
* Autoclave should be fully automatic, programmable,
microprocessor type.
/* Touch sereen colored display with minimum 7.0" with integrated
printer.
+ Preumatically driven Automatic single sliding door system,
Daor should be move vertically or horizontally. Opening and
lasing of door should be from the touch sereen & detect any
obstacles during movement of door.
* Time cycled, working pressure 32 psi,
+ Safety interlock, Temperature & Pressure recorder,
* Chamber pressure indicator on the touch screen.
© Cycle indicator to determine the phase of sterilization cycle.
ProgramiCycle selection.
+ Complete with standerd accessories and remavable shelves,
capable of taking both packets and containers of all standard
sizes.
* Chamber capacity 160L or more, Rectangular shape, Chamber,
Jacket and doors made of AISI 316 LITi.
« Liquid ring vacuum pump for vacuum system Autoclave should
have Bowie dick cycle, vacuum test cycle along with O8 fa
programmed cycles
\« The system should be complete with built-in steam generator in
AISII6 with manual blow down, water saving system and
automatic heat exchanger. RO System Locally provided with
suitable capacity
Standards: [SO 9001, 180 13485,
Certifications: FDA 510(k) approved / CE marked.
Manufacturers from other countries (other than USA, Europe
or Japan) will be accepted if equivalent has FDA $10(k), CE, JIS,
MHLW or SQAO certifications
iH
‘Tem Now2t
(2 for Ortho Department)
Specification: ‘ YS
Dimension 2060°520mm R
Table Lifting _$30mm to 970mm
Page 47 of 146
ocarlleu WI‘Trondelenberg and Reverse Trendalenberg 30°30"
Lateral Tilt 25°-25°
Head Plate up 45*-dow
Back Plate up-90°-down
Leg Plate up 20"down 90°
Longitudinal Stiding 300mm
Kidney Bridge Litting 120mm
Batery Standby 120 minutes
Patient Weight Capacity 250kx - 280k
‘Aru Rest pes
Leg Crutches 2pes
Shoulder Support pes
Body Bind Beh Ipes
Anesthesia Screen Frame pes
Body Supparter2pes
Hand Remote Control pes
Square Clamp 1 Spex
Square Clamp Il
nlvensl Pelvis Res wth Pa
ion Accessory Clamp
Tone enced ‘Counter-traction Support
Abductor Bar (double bar)
Mounting Bracket, Pin type
Patient Transfer Board
Orthopedic Accessory Cart
Multi task Arm tard pair
“Traction Unit
‘Traction Boot, Adult
‘Traction Boot, Peads
Skeletal Traction Bow
Basic Stiseups
Perineal Post with Pad
Straight Traction Extension
Standards: ISO 9001, 10 13485,
Certifications: FDA 510(k) approved / CE marked.
Manufacturers from other countries (other than USA,
Europe or Japan) will be accepted if equivalent has FDA
510(k), CE, JIS , MHLW or Jags certifications
Fully Automatic
(4 for OPD)
Specification:
Measurement method Os¢illometric
‘AC power supply unit Switching 100 V=240'V AC 50 Hz~60 Hz
Protection against electrical shock Class 1, Type B
Pressure display range 0 mmig-299 mmttg
‘NIBP measurement range SYS 40 mmHg~270 mmHg, DIA.
20 mmHig-200 mmHg, Pulse 30 bprt~240 ben
Measurement accuracy Pressure: £3 mmHg Pulse: £5 %
‘NIBP clinical test BS EN1060-4:2004, BHS:1993, +/-S mmHg
SD:8 mmilg.
‘Cuff fastening method Torque-Contralied Belt drive Method
(ICBM)
External dimensions 241(W)*324(13)x390(D) mm 9.5(6) *13.0 (H)
15.4(0) inches, .
Page 48 of 146
ocanned wit GarnataWeight ‘Approx 5.5 kg (12 Ib)
Applicable nrm sirsumference 7 inches (18.0 cm) to 13.8 inches
(5.0 cin)Cult size 135x300) men (4.9% inches)
Display type
Systolic (SYS) Jedigit display by LED
Diastolic(DIA)
Paalse mte(PUL)
Counting function Display’ of number of measurements to
date: up to 999,999 tines
‘Applied pressure Automatic inflation by air pump
Pressure detection methed Capacitance type pressure transducet '
‘Operating conditions Temperature: +10 C140 °C lumidity: 30
*ARN-85 RH, no condensing
‘Transpory/Storage conditions Temperature: -20°C=+60°C
‘humidity: 10 %4RH~9$ 94RH, no condensi
Quick release when START/STOP button is pressed
Safety device (Electrical) Quick release when FAST STOP burton is
pressed
Clock function Clock Display, Date & Time (2015 ~2050)
ISO 9001, ISO 13485,
& Certifications; FDA 510(k) approved / CE marked,
Manufacturers from other countries (other than USA,
Europe or Japan) will be accepted if equivalent has FDA.
510(k), CE, JIS , MHLW or JQAO certifications
‘ow NaS | 3-channel ECG Cardiac Monitoring 4
Ger orp)
‘Technical Specification: ‘
3-channels ECG
5. inch high-resolution LCD-
‘Durable and comfortable silicon keys
Rechargeable lithium battery for 2.5 hours operation
Peripheral ports: SD card, USB Up to 20 files internal storage |
Unlimited files with external SD card
‘USB port fur data transmission to PC
Standards: ISO 9001, ISO 13485,
Certifications: FDA $10(8) approved / CE marked,
‘Manufacturers from other countries {other than USA,
€ Europe or Japan) will be accepted if equivalent has FDA
510(K), CE, JIS.
Trem No24 | C Arm Image Intensifier Fluoroscopy,
(for Ortho 1 for Urology 1 for Neurosurgery)
W or JQAO certifications
‘Specification:
| High frequency 40 KHZ microprocessor control mono block x
ray generator
4, System should have mono block $.0 KW or betier X- ray
geiierator
t X ray tube Rotatary of better
Foeus spot 0.6 Or better
‘Aniode heat content: 3O0KHU or better
"Anode heat dissipation 600 w or better
System should have advance heat eontrolémanagement system
X ray tube filter Al and Cu
System should have pre collimatog far Mat panel detector
Continuous fluoroscopy ‘Ky F2ORY or better “y |
a
OCaIINe!Continuous Nuoroscopy mA ranger 1.7-TimA or better
Pulsed fluoroscopy Ky range 40-120Ky ar better
Pulsed fluoroscopy’ mA 02-200 mie or belier
Digital radiography range 40-120 KY or better
ital Radiography mA 24 mA or beiter
Itis type x ray tube eoltimator
Collittator rotation: #80°
Flat panel detector size 2lem x2Ncm or better
EPD should be latest with esi scintillator
Pisel size should be less than 180,m or more
System should have 16 bit er better gray scale values
DQE should be 75% at Slp/mm or better
System should have double laser positioning device integrated on “
both cube and detector housing for patient positioning,
System should have anti-seatter Grid & Auto calibration
System should have Frame rate max.
System should have Detector ihatrin of 2,048 x 2,048 pixels
System should have Mag. mode 1 with 1,536 x 1,536 pixels
System should have Mag. mode 2 with 1,024 x 1024 pixels
User interface should be TFT Touchscreen on Monitor Support
‘Arm stand-alone unit
User interface Resolution should be 64 x 480 pixels or better
‘System should have Motor driven vertical travel 42cm or better
System should have Horizontal travel: 22-25em or better
‘System should have Orbital rotation +45"or better
$225°or better
System should have C-arm dep!
‘System should have steering and braking lever with parallel
movement of the mobile stand in all directions
‘System should have High brightness full HD 27"-30" split
monitor mounted on c-arm
System should have Viewing angle of monitor: 178° or better
[ ‘System should have Native resolution: 1920 X 1080 pixels or
better
t System should have Monitor mounted an C= Arm i
System should have Real-Time Processing Functions
System shiould have Recursive filter Stack filter (LIM) , Edge
enhancement filter up to 5 bevels
system should have special dose control software by using object
detection in Pediatrics
4. System should have Motion detection algorithms Automatic dose
reduction, algorithms Automatic metal correction algorithms
; System should have 2D distance measurement ‘
System should kave Post processing Functions
System should have AEC and ABC eantral
System should have Digital Image rotation and reversal without
radiation
t System should have Gray scal
System should have Image
Page 80 of 146
1
OCaTINNeU |System should hnve Bono, Henri, Abdomen, soft tissue, spine
Procedure , Metal, Repositioning, pediatric,
fp System should have Auto save image option
Siies should have Image and ¢ine Stari, stop de reply rate
contro
. oun should have Inge storage capacity: 100,000 image or
etter
System shuld have Image Archiving options:
t BICOM, TIFF, Multimedia(Av), JPG
Power supply: 240VAC.
SO 9001, ISO 13485,
Certifications: FDA 510(k) approved / CE marked,
Manufacturers from other coutitries (other than USA, Europe
oF Japan) will be accepted if equivalent has FDA S10(I AND
CE! JIS! MHLW or JQAO certifications
‘Shaver machine
(1 for Ortho)
Hand piece (2 Button) with Forward & Reverse Mode (500-12,000
‘Apin) and Oscillate mode Special for Condermalacia (500-4000 cpm)
‘or better Single & Multi Oscillate Light Weight Autoclavable.
Electrie Console 239 Volt, With Universal Cable] Capable of
Speratg 24 different hand pieces, 2 hand pieces operate able
Simultaneously with built-in pump for irrigation. Both Hand & Foot
contral. Use in Arthroscopy, Large Bone Orthopedies,
4.2MM STERLING GREAT WHITE, (B36)
4.2MM FULL RADIUS RESECTOR, (BX/6)
3.5MM STERLING GATOR, (BXi6)
42MM STERLING GATOR, (BX/6)
3.5MM STERLING CUDA, (BX/6)
42MM STERLING CUDA, (BX/6)
Standards: ISO 9001, ISO 13485,
Certifications: FDA 510(k) approved / CE marked.
Manufaeturers from other countries (other than USA,
Europe or Japan) will be accepted if equivalent has FDA.
510(k), CE, JIS , MHLW or JQAO certifications
Tem No 26
Infusion pump
(1 for Ortho)
SPECIFICATION:
Visible handle for easy carrying LCD Display
STRUCTURE: ist, high visibility Quick IV sets installation
Standard [V sets are compatible with the unit
User-defined configuration possible Add/Modity
brand on pump direelly, can copy to other pumps
through PC tool Automatic recognition of IY sets
brand Operating Modes: Rate mode, Time mode,
Body weight mode, Sequential mode Delivery rate
0.10-2000mi/h. VBI: 0. 10-9999), step 0.1 ml
CONSUMABLE Tino: 00;00:01-99:59:59 hh:mm:ss; step Is,
Preset Volue(VTBI} 0.1-9999ml
KVO Rate: 0.1-5,0ml/h adjustable, step 0.Imi/h
Bolus Rate: Automatie/Manual bolus, 0.20-
2000mIn, dei 800mI/
Preset bolus Volume;',10-9999%mn!—
Purge Rate: SOOraP SelEtest system
Page $1 of 146
OCAMMNeU WIL LalTiocaAnti-bolus fimétion. Reduces significnnily bolus
ifler occlusion release Titrmtign fittction:
Availabe (o change the delivery rate during
Infusion at minigun increment of O.1imUh The
‘bolus accurlation value aad bolus rate shall be
slisplayest Drug tibeary with up to 200 cag name,
dl or delete drugs available in seralcfined drug
{ist Howe up 10 1500 history records Have
GENERAL —__mulomatic hols system, with bolus rc and preset
FEATURES Yolume adjustsble Remeinber last infusion
‘configuration wien power off Delivery Accuracy:
45% Dynamic ocelusion pressure displayed on
screen Multifimction interfoce: RS232, DC-input,
Nurse Call Volume collection furietion: ass
infased volute record through automatically
‘olume calculation with four modes: 24 teal,
current total, timefcertain period) timing
volume(tim ing interval) Data transmission is
Available with mol -funetion interface
‘Acoustic and visible alarm 3 levels alam, tow,
‘medium, high Alarm ineluding: occlusion, battery’
empty, VTBI done, syringe empty, syringe
disengaged, KVO finish, system errr, reminder,
battery fow, No battery inserted, syringe near
‘empty, standby time expired, etc. 4 Decluston
alarm level: 15tmmhg, 300mmbg, 325munbg,
Saka 900mmhg-4 Pressure unit selectable: mmf, KPa,
psi, bar Air bubble alarm size:20, 50, 100, 250,
$500, 800 pl Alarm sound I-8 levels adjustable
Visualized alarm with intwtive graphic
information, current status reminder and operation
guidance Yellow and red alarra light with differen
frequency according alarm levet
Screen contrast 1-8 levels adjustable Delivery rate,
DISPLAY ‘current infusion, VTBI, IV set brand, real-time
pressure, batiery capacity, drugs, alarms etc.
POWER SUPPLy: ACI00-240V, 5W/60HZ
Battery type: Rechargeble Lithium battery Battery
BATTERY operating time: more than 4 hours(@25mi"h
Battery charging time: less than 6 hours for 100%
SAFTY ‘Type of shock protection : Class l, Type CF.
SPECIFICATION defibrillation proof Water-Proof Grade : P34
Standards: ISO 9001, ISO 13485,
Ceitifications: FDA $10(&) approved / CE marked.
Manufacturers from other countries (other than USA,
Europe or Japan) will be accepted if equivalent has FDA
$10(k), CE, JIS , MHLW or JQAO certifications
i
4
‘Neat
“Tourniquet machine
(for Ortho) a
Dual Channel Electronic Tourniquel with integral power supply
stand & cuff supply hoses.
which cequires no external —o ‘comes with mobile
‘Technical Specifications
Page 52 of 146
ocaiineu AITIDCa