Class 2 - 3 Medical Device Cleanrooms
This post classifies cleanroom design parameters during medical device
manufacturing. Establishing criteria is a challenge. The FDA does not set forth any
cleanroom classification or particulate level based on application.
Essential Medical Devices Require Cleanroom Production
All invasive procedures involve contact between a medical device or surgical instrument
and a patient’s sterile tissue or mucous membranes. A major risk of all such procedures
is the introduction of pathogenic microbes that could lead to infection. Failure to
properly disinfect or sterilize new or reusable medical equipment is a perilous risk in the
case of a breached sterile barrier.
Environmental controls are dependent on the intended use of an object: items that
contact sterile tissues are considered critical items such as surgical instruments or
implantable devices. Semicritical items which contact mucous membranes including
endoscopes in medical fields or even drill guides for dental work. Items that only
contact skin are considered noncritical items, such as stethoscopes. Each of these
categories designates a different threshold of sterilization and disinfection which
include high-level disinfection and low-level disinfection.
ISO 14644 - FS-209E Medical Device Cleanrooms
Authoritative documents such as ISO 14644 and FS-209E provide no specific
instructions regarding activities carried out in medical device environments. Broadly,
medical device manufacturing is conducted in an ISO 5 – 8 cleanroom (Class 100 –
100,000). Medical device packaging is conducted in an ISO Class 7- 8 cleanroom. An
ISO 8 environment may include a dedicated gowning room and sometimes softwall
environments in packaging or preparation areas. Classification of surrounding
environments, isolators, and sterile barriers all require very specific performance
criteria.
Related Post: USP 797 Sterile Compounding
The above standards do not apply expectations of the state board or FDA mandates.
ISO classifications simply standardize a benchmark of environmental quality.
Recently expanded guidelines, ISO 13485 (medical devices) and USP
800 (pharmaceutical compounding), emphasize an evolving landscape of risk
management solutions for medical sciences. Harmonizing medical device cleanroom
standards requires an evaluation of not only cleanroom design, but also processes
equipment, materials, consumables, and assemblies.
FDA Medical Devices in Plain Language (Summarized)
Medical devices are defined in Section 201(h) of the Food, Drug, and Cosmetic Act as
any healthcare product that does not require chemical action or metabolization.
A medical device is an instrument, apparatus, implement, machine, contrivance implant,
in-vitro reagent, or other similar or related article including any component, part or
accessory. The FDA classifies medical devices in over 1700 generic device categories
within 16 medical specialties. Each classification outlines a unique compliance based
on risk estimates.
(1) A medical device is recognized in the official National Formulary, or the US
Pharmacopeia, or Formulary, or the US Pharmacopeia, or any supplement to them.
(2) A medical device is intended for use in the diagnosis or cure of disease or other
conditions including the cure, mitigation, treatment, or prevention of mitigation, or
prevention of disease, in man or other animals in man or other animals, or…
3) A medical device does not achieve its primary intended purposes through chemical
action or metabolization within or on the body of a man or other animals.
Medical Devices Classes and Quality Management
Class 1 devices are perceived to be the lowest risk devices. A formal review (i.e., 510(k) or
PMA) is not typically required. General Controls are usually sufficient. Approximately ¾ of
all devices listed with the FDA each year are Class 1. Examples include medical tape and
bandages.
Class 2 devices are perceived to be of greater risk to the patient or user. General controls
are not sufficient for safety or efficacy. Most of these devices require a 510(k) application.
Approximately 3,000 Class 2 devices are cleared by the FDA each year. Most coronary
catheters fall within this classification.
Class 3 devices are perceived to represent the highest risk to patient or user. These
devices support or sustain human life, may prevent or impair human health, or present
potential risk of illness or injury. Most Class 3 devices require a PMA application.
Approximately 40 Class 3 devices are approved by the FDA each year. Examples include
heart valves and pacemakers.
In regards to medical device cleanroom design, Class II and Class III devices require a
Quality Management System established by ISO 13485. A cleanroom is a critical
component for ensuring traceability, lot-to-lot tracking, establishing an aseptic workflow,
and for monitoring air quality to identify and eliminate sources of contamination.
Related Post: ISO 13485 Quality Management Systems
The requirements of ISO 13485 apply to any medical device organization regardless of
size and activity as a basis for demonstrating and supporting compliance with
regulatory requirements.
Medical Device Cleanroom Design
The principles of cleanroom design for medical devices must support three objectives:
Suitable size, construction, and location to facilitate cleaning, maintenance, and
proper operations
Plan adequate space for orderly placement of equipment and materials to prevent
mix-ups and contamination
Design the adequate flow of materials & persons to prevent contamination.
FDA 21 CFR 211.42 addresses design and construction features.
According to the regulation:
The facility must be of suitable size, construction, and location so that cleaning,
maintenance, and operations can be properly performed.
The facility must have adequate space and flow design to prevent mix-ups and cross-
contamination.
Surfaces must be easy to clean.
Temperature and humidity must be controlled as appropriate.
Air must be positive pressure and HEPA filtered. (Non-hazardous)
Environmental conditions must be monitored.
Room and equipment must be cleaned and disinfected properly.
Equipment must remain in aseptic conditions.
Medical Device Cleanroom Requirements?
What Cleanroom Classification is Required for Class I, Class II,
and Class III Devices
Our cleanroom teams often field inquiries from large and small operations alike. Many
of our customers want to know what ISO Classification is required for a Class I or Class
II device, or for a particular technology. Some organizations don’t require a standardized
compliance and only intend to make a particular area cleaner, or create manufacturing
cells that isolate cleaner tasks from dirtier ones.
ISO 7 – 8 cleanrooms are common for Class I and Class II medical devices. Because a
Class 1 device does not require a Quality Management System, a softwall cleanroom is
common instead of a dedicated installation. Class III devices carry greater risk and a
higher degree of regulatory control, therefore justify the most stringent manufacturing
conditions. As device complexity grows and part sizes shrink, particles become more
problematic. This leads many facilities towards a cleaner classification than
immediately necessary in order to future-proof the build for evolving process
requirements.
Broadly, most injection molded medical devices are produced in ISO Class 8 (Class
100,000) cleanrooms. Sterile devices may require an ISO Class 5 cleanroom (Class
100), while medical device packaging is conducted in an ISO Class 7-8 cleanroom.
Classification of surrounding environments, isolators, and sterilization processes all
require very specific performance criteria.
Plastic injection molding and packaging may extend to ISO Class 8 (Class 100,000)
which requires an average airflow velocity of 8 ft/min with 5 – 48 air changes per hour.
The space will require HEPA filtration fan filter units with 5% – 20% ceiling surface
coverage depending on the process.
Want More Info on a Cleanroom for Your Specific
Application?
We help facilities produce yields faster with better process control. Get help
identifying the proper equipment and cleanroom design for your facility.
Free Cleanroom Consultation
Related Posts
Medical Device Cleanroom Construction Part 1: Walls and
Installation
How is a medical device cleanroom built? Here's a component-by-
component guide to ISO Class 7 cleanroom construction.
Medical Device Cleanroom Construction Part 2: Doors, Electrical, and Layout
In medical device cleanrooms, a common configuration is an ISO 8
gowning room and ISO 7 primary production area. A cascading air
flow design and positive pressure HVAC calibration move the
cleanest air at the…
Medical Device Cleanroom Construction Part 3: Fan Filter Units -
Ceiling Grids
In this post, we overview technical considerations for clean room
fan filter unit selection, installation, and features. You'll learn how
PAC approaches design and configuration of cleanroom HEPA filter
units and fan filter housing specifications.…
Medical Device Cleanroom Construction Part 4 - VCT Flooring
Design
There are a few final preparations required before the final
cleanroom flooring installation. Here's what to expect and consider
for cleanroom tile installations and flooring installation.
Medical Device Cleanroom Construction Part 5: HVAC Balancing &
Calibration
Before cleanroom validation, a cleanroom requires mechanical
calibration. HVAC calibration first ensures that the room meets the
performance parameters which would deliver the expected results
during final ISO validation, such as differential pressure, air
change…
Medical Device Cleanroom Construction Part 6: Pressure
Differentials, Humidity and Temperature Calibration
The central controller integrates the data from both the
thermostats, fan filter processors, and room-side environmental
sensors. Fan filter speed is then adjust by each group to achieve
the specified room pressure while at ideal…
ISO 13485:2016 Medical Devices
ISO 13485 is designed for use throughout the life cycle of a medical device. It supports
each stage of medical device development and operation from initial concept to
production and disposal. The standard helps internal…
Cleanroom Classification Comparison Specifications Defined
A comparison of cleanroom classifications and specifications.
CleanPro® Softwall Cleanroom Enclosure
This customer needed to enclose a piece of machinery, and
CleanPro® was able to provide a solution.
CleanPro® Turnkey Cleanroom Install
CleanPro’s turn-key cleanroom solution provided a one-stop, one-
contact result for the initial delivery and on-site installation of
walls, ceiling grids, electrical systems, flooring, filters, HVAC, and
more.
Cleanroom Construction FAQ
Definitive Cleanroom Construction FAQ: How much does a
cleanroom cost? How does the level of ISO classification impact
build and operating costs? How much supporting space will I need
for a cleanroom addition or ISO…
CleanPro® Chemotherapy & CSP Cleanroom Installation
This modular, sterile compounding cleanroom is designed for USP
797 and USP 800 compliance, particularly for compounding
chemotherapy drugs. Safe handling of sterile compounds requires
special considerations: heat-welded floors, anterooms and buffer
areas.
Cleanroom Construction FAQ
Definitive Cleanroom Construction FAQ: How much does a
cleanroom cost? How does the level of ISO classification impact
build and operating costs? How much supporting space will I need
for a cleanroom addition or ISO…
Share this Article:
Facebook Twitter LinkedIn Reddit Email Print
Mitch Walleser
Mitch is a contributing writer for Production Automation Corporation. PAC is a factory-
direct distributor of products and environmental solutions for industrial and critical
requirements within electronics, medical device, life science, pharmaceutical, and general
manufacturing industries. Mitch has worked with manufacturing engineers, in-house
specialists, and factory experts to highlight and uncover manufacturing solutions. His
background includes 3D printing, electronics, and cleanroom manufacturing.
Was this post helpful?
Yes No
PREVIOUS
New: Low Particulate Cleanroom Tape
NEXT
Ethanol Purity, Grades, and Extraction Techniques
Recent Posts
PAC Designs a Crucial Cleanroom Facility for the Webb Space Telescope
PFAS Chemicals: What are They and How Can You Replace Them?
3M™ Novec™ Chemical To Exit Market: Here’s The Replacement You
Need
Search Posts
Search …
Search
About Production Automation
Production Automation Corporation (that’s us) is an essential business
supplier for medical device, cleanroom, and general manufacturing
industries.
All products featured on the PAC blog are available through the
www.gotopac.com webstore. Online orders are shipped and fulfilled
Monday – Friday from PAC warehouses in California, Minnesota, and
Texas.
Production Automation Corporation (PAC) stands 100% behind a product
catalog of over 40,000 products, which includes support for factory
warranties, value added services, and returns. Get help from a PAC sales
rep during business hours via email, phone, or live chat.
Shop Online: www.gotopac.com
Toll Free: (888) 903-0333
Email: blog@gotopac.com
Office Hours: 8am - 5pm CST
Sign Up to Our Newsletter
Email address:
Your email address
Sign up
PAC Logo
Production Automation Corporation
121 Cheshire Lane, Suite 400
Minnetonka, MN 55305
(888) 903-0333
Production Automation Corporation (PAC) is a distributor of supplies, equipment, and
controlled environments for life science, aerospace, electronics, and industrial
manufacturing industries. Learn more at www.gotopac.com
Subscribe To Our Newsletter
Email Address
View previous campaigns.
Subscribe
Trustpilot
Production Automation Corporation is ISO 9001:2015 Certified
© 2023 Production Automation Corporation. All Rights Reserved.