Detection 5
Detection 5
Abstract
Several non-invasive and novel aids for the detection of (and in some cases monitoring of ) caries
lesions have been introduced in the field of ‘caries diagnostics’ over the last 15 years. This chapter
focusses on those available to dentists at the time of writing; continuing research is bound to lead to
further developments in the coming years. Laser fluorescence is based on measurements of back-
scattered fluorescence of a 655-nm light source. It enhances occlusal and (potentially) approximal
lesion detection and enables semi-quantitative caries monitoring. Systematic reviews have identi-
fied false-positive results as a limitation. Quantitative light-induced fluorescence is another sensitive
method to quantitatively detect and measure mineral loss both in enamel and some dentine lesions;
again, the trade-offs with lower specificity when compared with clinical visual detection must be
considered. Subtraction radiography is based on the principle of digitally superimposing two radio-
graphs with exactly the same projection geometry. This method is applicable for approximal sur-
faces and occlusal caries involving dentine but is not yet widely available. Electrical caries
measurements gather either site-specific or surface-specific information of teeth and tooth struc-
ture. Fixed-frequency devices perform best for occlusal dentine caries but the method has also
shown promise for lesions in enamel and other tooth surfaces with multi-frequency approaches. All
methods require further research and further validation in well-designed clinical trials. In the future,
they could have useful applications in clinical practice as part of a personalized, comprehensive car-
ies management system.
Copyright © 2009 S. Karger AG, Basel
Laser Fluorescence
Laser fluorescence (LF) has increasingly been adopted in the caries detection proce-
dure for the last decade. In the Diagnodent device (Kavo, Biberach, Germany), a 655-
nm monochromatic light is emitted from an optical tip/sensor, which can also detect
the amount of back-scattered fluorescence [1]. During the caries detection proce-
dure, the light beam enters the enamel and will either go unhindered to the dentine
or be partially scattered. A regular crystalline structure like mature enamel is more
198.143.38.1 - 1/27/2016 7:17:48 AM
Downloaded by:
UCONN Storrs
a b
Fig. 1. a In LF measurements, the correct angulation of the probe is crucial. b In this case, a wrong
angulation of the probe leads to false-negative readings.
transparent; thus, light can pass through the enamel layer with only little deflection.
If the enamel layer is more inhomogeneous, more light will be diffracted and scat-
tered. The scattered portion of the light then excites either the dental hard substance
itself (and renders a measurement of the tooth’s autofluorescence) or excites so-called
fluorophores within the lesion. Fluorophores are particles with the property of fluo-
rescing when excited by a specific wavelength of electromagnetic energy. In the case
of 655 nm, the fluorescing particles have been identified as bacterial protoporphy-
rins [2, 3]. Thus, when a critically increased pore volume is exceeded, the amount of
back-scattered fluorescence is – theoretically – proportional to the amount of bacte-
rial infection, pore volume and lesion depth.
The measurements on the Diagnodent display render values between 0 (minimum
fluorescence) and 99 (maximum fluorescence), thus theoretically making quantitative
caries monitoring possible (fig. 1a, b). The manufacturer’s instructions make clear, how-
ever, that only certain limited ranges can be used for assessing lesion extent, that there is
overlap between some of these bands and that values over 45 may not be diagnostically
useful. A more recently developed LF device (Diagnodent pen) provides 2 tips for both
occlusal and approximal caries assessment (fig. 2, 3), although data supporting approxi-
mal detection on contacting teeth in vitro and in vivo is very limited at present.
198.143.38.1 - 1/27/2016 7:17:48 AM
a b
To date, many in vitro studies as well as an increasing number of in vivo clinical tri-
als have been undertaken to validate the aforementioned general principle of detect-
ing caries with LF. Generally, in vitro studies using teeth that were not frozen may be
considered to be biased, because of a significant drop of the fluorescence signals due to
degradation of porphyrins [4]. Narrative and systematic reviews have shown that LF
is more sensitive on the occlusal aspect of posterior teeth than traditional diagnostic
198.143.38.1 - 1/27/2016 7:17:48 AM
Baseline image
Fig. 4. Subtraction radiography requires 2 radiographs with exactly the same angulation.
also been tested in vivo in 44 eight-year-old children and was calculated to be able to
assess lesion activity [38]. In a single-blind cluster randomized trial with 296 school-
children, QLF was used to calculate the remineralization capabilities of different flu-
oride-containing toothpastes on initial caries [39]. QLF is suitable for early lesion
detection on smooth surfaces, although its validated performance in terms of sensi-
tivity and specificity across the full range of caries predilection sites and the impact of
surface staining are not yet established independently. However, its time-consuming
image processing and analysis and its costs are the biggest obstacles for wide use in
private dental practice. Concerning a wider use within this target group, a recently
developed QLF-based tool for rapid, but not quantitative, caries detection might be
promising [40].
Subtraction Radiography
Subtraction radiography is based on the principle that 2 radiographs taken with the
same projection geometry are spatially and densitometrically aligned. They are then
digitally subtracted to remove structural noise in the image. Following subtraction
the contrast is stretched in the image. This process enhances any changes in the lesion
between 2 points in time. This technique has been made a reality in the last decade,
whereas earlier attempts had proved to be unsuccessful. It is clear that this subtraction
process will not necessarily improve the detection of a lesion, but will only provide
198.143.38.1 - 1/27/2016 7:17:48 AM
Over the last 6 decades, the relationship between the extent of caries in teeth and elec-
trical resistance has been investigated. These investigations have focussed on the vari-
ous parameters affecting electrical measurements of teeth, including porosity, surface
area of the contact ‘electrode’, the thickness of enamel (and dentine) tissue, hydration
of the enamel, temperature, ionic content (concentration) of the dental tissue fluids,
as well as the maturation time of the tooth in the oral environment.
The validity of electrical caries detection methods was reviewed by Huysmans
[42], as well as Longbottom and Huysmans [43], and subject to a systematic review by
Bader et al. [44]. These mainly relate to the commercially available fixed-frequency
ECM device (electronic caries monitor). The relatively few papers on this ECM pub-
lished since those reviews indicate that only slight modifications are required to their
conclusions.
The ECM device can be used in a ‘site-specific’ or ‘surface-specific’ mode, the latter
involving a conducting medium used to cover the whole occlusal surface of a tooth –
see table 1 for sensitivity and specificity values.
In vitro studies by Ellwood and Cortes [45] and Cortes et al. [46] indicated that the
presence of stain is a confounder for ECM measurements and that different ECM cut-
off values for enamel and dentine caries may be required for stained teeth.
The reproducibility of the device has been assessed and ranges from 0.53 to 0.92
for the site-specific and from 0.55 to 0.89 for the surface-specific method, but due to
198.143.38.1 - 1/27/2016 7:17:48 AM
Site-specific Surface-specific
the wide variations, Huysmans [42] cast doubt on its suitability for the longitudinal
monitoring of lesion behaviour.
More recently, analysis by Huysmans et al. [47] has indicated that there may be a
systematic error in the fixed-frequency ECM measurement technique, possibly attrib-
utable to inconsistent probe contact with the tooth surface. The authors suggested that
a modification of the probe design may be necessary and a newer prototype device
with such a modified probe design appears to reduce this systematic error [48].
Root Caries
A small number of studies have looked at the use of the fixed-frequency ECM
for root caries detection. Wicht et al. [49] found a moderate relationship between
ECM measurements and root caries lesion depth in an in vitro study. Baysan et al.
[50], in vitro, and Petersson et al. [51], in vivo over 12 months, found that ECM
measurements appeared capable of distinguishing between different severities of
primary root caries lesions. Electrical measurements for caries detection have been
investigated for several decades. The literature indicates that there is value in such
measurements using fixed-frequency devices for dentine caries detection at occlusal
sites. However, at present, there are limitations to these measurements, in particu-
lar in relation to the number of false-positive readings. This type of ECM may be
used as an adjunct to clinical information and other detection methods. The use of
multiple-frequency electrical caries measurement, using AC impedance spectros-
copy, was reported with very encouraging in vitro results some time ago [52, 53].
The detection of caries and its extent, in terms of the order of magnitude differ-
ences between sound approximal surfaces, surfaces with initial lesions and surfaces
with cavitation, was shown to be feasible [52, 53]. Although further development
of a clinical device which can be used on other tooth surfaces has been reported
in abstracts and a device is apparently now available to dentists in the UK, fur-
ther published studies are required to assess the implementation of this technology
further.
198.143.38.1 - 1/27/2016 7:17:48 AM
Four novel methods for non-invasive caries detection and – to a limited extent – car-
ies monitoring are presented in this chapter. All methods presented have potential for
wider use in the private clinical setting – the balance between enhancing the sensi-
tivity of lesion detection and trading-off lower values for specificity must always be
borne in mind and allowed for when making treatment decisions. All the methods
presented require further research and further validation in well-designed clinical tri-
als. In the future, these types of lesion detection aids could have useful applications
in clinical practice as a complement to clinical visual lesion detection and as part of a
personalized, comprehensive caries management system.
References
1 Lussi A, Imwinkelried S, Pitts N, Longbottom C, 11 Huth KC, Neuhaus KW, Gygax M, Bucher K,
Reich E: Performance and reproducibility of a laser Crispin A, Paschos E, Hickel R, Lussi A: Clinical
fluorescence system for detection of occlusal caries performance of a new laser fluorescence device for
in vitro. Caries Res 1999;33:261–266. detection of occlusal caries lesions in permanent
2 König K, Flemming G, Hibst R: Laser-induced auto- molars. J Dent 2008;36:1033–1040.
fluorescence spectroscopy of dental caries. Cell Mol 12 Bamzahim M, Aljehani A, Shi XQ: Clinical perfor-
Biol (Noisy-le-Grand) 1998;44:1293–3000. mance of Diagnodent in the detection of secondary
3 Buchalla W, Attin T, Niedmann Y, Niedmann PD: carious lesions. Acta Odontol Scand 2005;63:26–30.
Porphyrins are the cause of red fluorescence of cari- 13 Bamzahim M, Shi XQ, Angmar-Mansson B:
ous dentine: verified by gradient reversed-phase Secondary caries detection by Diagnodent and radi-
HPLC (abstract). Caries Res 2008;42:223. ography: a comparative in vitro study. Acta Odontol
4 Francescut P, Zimmerli B, Lussi A: Influence of dif- Scand 2004;62:61–64.
ferent storage methods on laser fluorescence values: 14 Lennon AM, Buchalla W, Switalski L, Stookey GK:
a two-year study. Caries Res 2006;40:181–185. Residual caries detection using visible fluorescence.
5 Ricketts D: The eyes have it: how good is Diagnodent Caries Res 2002;36:315–319.
at detecting caries? Evid Based Dent 2005;6:64–65. 15 Aljehani A, Tranaeus S, Forsberg CM, Angmar-
6 Lussi A, Hibst R, Paulus R: Diagnodent: an optical Mansson B, Shi XQ: In vitro quantification of white
method for caries detection. J Dent Res 2004;83(spec spot enamel lesions adjacent to fixed orthodontic
iss C):80–83. appliances using quantitative light-induced fluores-
7 Bader JD, Shugars DA: A systematic review of the cence and Diagnodent. Acta Odontol Scand 2004;
performance of a laser fluorescence device for 62:313–318.
detecting caries. J Am Dent Assoc 2004;135:1413– 16 Aljehani A, Yousif MA, Angmar-Mansson B, Shi
1426. XQ: Longitudinal quantification of incipient carious
8 Tranaeus S, Lindgren LE, Karlsson L, Angmar- lesions in postorthodontic patients using a fluores-
Mansson B: In vivo validity and reliability of IR flu- cence method. Eur J Oral Sci 2006;114:430–434.
orescence measurements for caries detection and 17 Wicht MJ, Haak R, Stutzer H, Strohe D, Noack MJ:
quantification. Swed Dent J 2004;28:173–182. Intra- and interexaminer variability and validity of
9 Lussi A, Hack A, Hug I, Heckenberger H, Megert B, laser fluorescence and electrical resistance readings
Stich H: Detection of approximal caries with a new on root surface lesions. Caries Res 2002;36:241–
laser fluorescence device. Caries Res 2006;40:97– 248.
103. 18 Takamori K, Hokari N, Okumura Y, Watanabe S:
10 Lussi A, Megert B, Longbottom C, Reich E, Detection of occlusal caries under sealants by use of
Francescut P: Clinical performance of a laser fluo- a laser fluorescence system. J Clin Laser Med Surg
rescence device for detection of occlusal caries 2001;19:267–271.
lesions. Eur J Oral Sci 2001;109:14–19.
198.143.38.1 - 1/27/2016 7:17:48 AM