Using MRI Safely - Practical Rules For Employees
Using MRI Safely - Practical Rules For Employees
PRACTICAL RULES
FOR EMPLOYEES
Holland HealthTech
Using MRI safely – practical rules for employees
MRI Working Group
2
The Netherlands has played an im- chairmanship of this working group.
portant positive role in the postpo- Mathieu Pruppers of the RIVM has
nement process. Those responsible supported me expertly as the secre-
within SZW [Ministry of Social Affairs tary. The Working Group has defined
and Employment] acknowledged the rules of conduct described in this
the negative effect of the Directive document in an exceptionally con-
on the use of MRI in patient care at structive and progressive way; the
an early stage. Based on their recom- Working Group believes that these
mendation, a working group was set will ensure that MRI can be used
up comprising representatives from safely in practice.
employers’ and employees’ organi-
sations plus a number of additional
experts, with the aim of developing
practical rules defining safe ways for
employees to work with MRI. These
practical rules could be the basis for Rotterdam, juli 2008.
drawing up an arbocatalogus [wor-
king conditions catalogue]. Because Gabriel P. Krestin
I was closely involved in the discus- Hoofd Radiologie Erasmus MC
sions about the Directive both in Past-president ESMRMB.
the Netherlands and at the Europe-
an level, I was asked to take on the
3
MRI Working Group
Index
1 Introduction 5
1.1 Background and Objectives 5
1.2 Concepts and definitions 6
1.3 Legislation and regulations in the Netherlands 7
2 Practical risks 10
2.1 Exposure to electromagnetic fields 11
2.2 Exposure to noise 12
2.3 Risks due to ferromagnetic projectiles 12
2.4 Risks from cryogenic liquids 12
3 Defined working situations 13
4 Types of effects 15
5 Precautions and rules of conduct 18
5.1 General precautions 18
5.2 Rules of conduct for working situation I 19
5.3 Rules of conduct for working situation II 21
5.4 Rules of conduct for working situation III 22
5.5 Test subjects – the voluntary basis 22
5.6 Education, training and instruction 23
4
1 Introduction 1.1 Background and
Objectives
MRI, magnetic resonance imaging or
nuclear magnetic resonance tomog- To be able to work safely with MRI,
raphy, is a powerful diagnostic tool it is important that staff are aware of
that has been in use for 25 years now the safety aspects and handle them
in healthcare and for scientific stud- appropriately. Organisational and
ies. The use of MRI has major benefits technical measures are needed in or-
for the patients. It has now become der to reduce the potential hazards.
an essential part of the diagnosis and This document describes the various
treatment of illnesses and in medical safety aspects and the corresponding
research. Using magnetic and elec- practical rules. It has been drawn up
tromagnetic fields, MRI makes it pos- by the MRI Working Group, an ad hoc
sible to produce detailed images of working group of institutions with
the interior of the human body. MRI interests affected by the application
therefore plays a key role in the rou- of MRI in the Netherlands (please re-
tine treatment of numerous diseases, fer to Appendix 1 for an overview of
such as cancer, cardiovascular disease the members of the Working Group).
and neurological conditions. Further The purpose of this document is to
scientific investigations into the ap- protect all employees1 who are in-
plication of MRI to other illnesses and volved with MRI and to do so at the
into functional and interventional most practical level possible. This
applications are still forging ahead. document will be brought to the at-
MRI provides a much higher contrast tention of those who are involved
between soft tissues than CT (com- in the discussion about changes to
puter tomography) and, unlike CT, the EU Directive for the protection
does not use ionising radiation. of employees from the dangers of
exposure to electromagnetic fields.
This document is also the framework
for drawing up a working conditions
catalogue for handling MRI.
1 This document primarily considers occupational working conditions for employees. Safe use of MRI
has consequences for the way patients are handled; these are not covered in this document (e.g.
instructions for the patient, the use of upper and lower SAR limits, positioning of the patients, checking
for contra-indications, etc.). The description of good practice is limited to the ‘imaging of people and
animals’. 5
MRI Working Group
in which employees can be exposed the static magnetic field may not ex-
to magnetic fields due to the applica- ceed 0.5 mT (the field strength above
tion of MRI. The possible effects that which hazards can occur for medi-
could arise when working with MRI cal implants such as pacemakers). In
scanners are then listed. Finally, pre- clinical practice, the controlled access
cautions are recommended for each area is often the same as the scanner
of these effects, grouped according room. It is possible that the magnetic
to the working situations. field may be greater than 0.5 mT in
auxiliary areas below or above the
The Working Group is convinced that scanner area, and additional mea-
if the rules in this document are ob- sures are required in such cases.
served, the safety of staff during cur-
rent working practices and accord- Static magnetic field
ing to the current state of knowledge The static magnetic field is one of
is sufficiently guaranteed. Exposure the elements required to obtain an
limits have not been discussed in MRI signal. In an MRI scanner, this
this document, partly because of the field is generated with a solenoid
ongoing international discussions coil that is often made of a supercon-
about how high these limits should ducting material. When the magni-
be. tude and direction of the magnetic
field do not vary over time, then
this is referred to as a static mag-
1.2 Concepts and definitions netic field. In almost all MRI systems,
this static magnetic field is continu-
Scanner room ously present, i.e. 24 hours a day.
The room in which the MRI scanner is The magnetic field strength of MRI
located and in which usually only the systems varies from 0.2 to 1.0 T for
patient is present during scanning. open systems with a vertical static
magnetic field, and from 0.5 to 3.0 T
Controlled access area (around for the classical cylindrical systems
the scanner) with a horizontal field. Systems us-
According to NEN-EN-IEC standard ing 7 T are also in operation in the
60601-2-33, the controlled access Netherlands for scientific research.
area is the “area to which access is For comparison: the strength of the
controlled for safety reasons”. Outside earth’s (static) magnetic field is about
the controlled access area, this stan- 50 microtesla in the Netherlands,
dard states that the stray field from whereas the order of magnitude for
6
the strength of a refrigerator magnet ing coil. This can only be done at very
is a few millitesla. low temperatures, cooling the coil
with liquid helium (sometimes with a
Switched gradient fields secondary layer of cooling using liq-
As well as a static magnetic field, uid nitrogen).
switched magnetic fields are needed
in order to localise the MRI signal and
thereby generate an image. These 1.3 Legislation and
switched magnetic fields generate a regulations in the
linear field gradient and are therefore Netherlands
also referred to as “gradient fields”.
The gradient fields are switched on
and off in timeframes in the millisec- Under the Arbowet [Working Condi-
ond range, thereby producing a mag- tions Act], the employer will ensure
netic field in the kilohertz range. (with the cooperation of the em-
ployee) that the work does not have
Radio-frequency fields any negative effects on the health
In order to obtain the MRI signal, a and safety of the employee. The Ar-
radio-frequency magnetic field is re- bobesluit [Working Conditions De-
quired in addition to the static mag- cree] discusses the risks in greater
netic field and the switched gradient depth.
fields. The frequency of this magnetic
field depends on the strength of the The Working Conditions Decree
static magnetic field and the atoms states among other things that de-
used. For hydrogen under a field of vices that produce electromagnetic
1.5 T, this frequency is 63 MHz. Mag- fields must be properly manufac-
netic fields with these frequencies tured, must be in good working
are also known as radio-frequency order and must be located and as-
fields, or RF fields for short. sembled in such a way or screened
in such a way that damage to health
Cryogenic liquids as a result of electromagnetic fields is
The majority of MRI scanners have avoided as far as possible. In order to
a superconducting solenoid coil. To prevent or restrict exposure to noise,
generate a strong magnetic field, a it also states that appropriate techni-
high current must be passed through cal or organisational measures must
the solenoid coil. This is achieved be taken so that the risks of expo-
technically by using a superconduct- sure are eliminated at the source or
7
MRI Working Group
2 NTA 8050:2007 nl. (Nederlandse Technische Afspraak = Netherlands technical agreement) Guideline
for producing working conditions catalogues. Publication date: 01-Jun-2007.
3 Corrigendum to Directive 2004/40/EC of the European Parliament and the Council of 29 April 2004 on
the minimum health and safety requirements regarding the exposure of workers to the risks arising
from physical agents (electromagnetic fields). OJ 2004, L184.
4 Health Council of the Netherlands. Comments concerning possible MRI restrictions due to
implementation of an EU directive. The Hague: Gezondheidsraad, 2007; publication number
2007/17.
5 Directive 2008/46/EC of the European Parliament and of the Council of 23 April 2008 amending
Directive 2004/40/EC on minimum health and safety requirements regarding the exposure of workers
to the risks arising from physical agents (electromagnetic fields) (18th individual Directive within the
meaning of Article 16(1) of Directive 89/391/EEC). OJ 2008, L114/88-89.
8
mendations for static magnetic fields - risks must be excluded or
and (extremely) low-frequency elec- restricted as far as possible
tric and magnetic fields. (also meaning a safe design
and safe construction)
MRI equipment must comply with - in relevant cases, appropriate
the essential requirements that are measures will be taken,
defined in European Directive 93/42/ including alarm features if
EEC of 14 June 1993 (Medical De- necessary, to protect against
vices)8. MRI equipment must be de- those risks that cannot be
signed and manufactured in such a excluded, and
way that its use does not create any - the users must be informed
hazards for the clinical condition about any risks that may still
or the safety of the patients, nor for be present as a result of any
the safety and health of the users or, shortcomings in the protective
should such cases arise, any other measures
persons. The equipment must also
be used in the prescribed manner The IEC (International Electrotechni-
and for the purposes for which it is in- cal Committee) has issued a standard
tended. It should also be understood that defines safety procedures and
that any risks that may be associated threshold values for patients and
with its use must be acceptable, tak- staff for exposure to electromagnetic
ing account of the usefulness of the fields during MRI procedures. This
equipment for the patient, and com- standard, which is primarily intend-
patible with a high degree of pro- ed for the equipment manufacturers,
tection for health and safety. When has been published by the European
choosing the most suitable solutions and Dutch governments as a har-
for the design and construction of monised standard under the Medical
the equipment, the manufacturer Devices Directive9. According to this
must adopt the following principles standard, the static magnetic field
in the sequence given below: is limited in principle to 4 T. Above
6 ICNIRP. Guidelines on limits of exposure to static magnetic fields. Health Phys, 1994, 66(1) 100-106.
7 ICNIRP. Guidelines for limiting exposure to time-varying electric, magnetic, and electromagnetic fields
(up to 300 GHz). Health Phys, 1998, 74(4) 494-522.
8 Council Directive 93/42/EEC of 14 June 1993 concerning medical devices. OJ 1993, L169 of 12-Jul-
1993 pp. 0001-0043.
9 NEN-EN-IEC 60601-2-33:2002 en;fr. Medical electrical equipment - Part 2-33: Particular requirements for
the safety of magnetic resonance equipment for medical diagnosisMedical electrical equipment, Part
2-33: Particular requirements for the safety of magnetic resonance equipment for medical diagnosis.
Publication date 01-Nov-2002. This standard has now been modified by two amendments in 2005
and 2008 (NEN-EN-IEC 60601-2-33:2002/A1:2005 and NEN-EN-IEC 60601-2-33:2002/A2:2008); please
refer also to Commission communication (2001/C 319/05) and “New NEN norms” in the Staatscourant
[Dutch Government Gazette] (13 December 2002, no. 241 / page 47); the second amendment had not
yet been published as a harmonised standard under the Medical Devices Directive on the date upon 9
which this document from the MRI Working Group was finalised (see page 1).
MRI Working Group
that, according to the IEC, use on pa- through the day and on the given
tients or test subjects is permissible peak acoustic pressure. The average
as long as the consent of the local level to which the employee is sub-
Medical Ethics Committee has been jected during the day is hence an
obtained. important parameter. For example,
if the employee works for four hours
Noise a day at 90 dB(A) and four hours at
In 2006, the Working Conditions 75 dB(A), then the daily exposure is
Decree was modified regarding the above 80 dB(A).
exposure of employees to the risks
of noise10. This amendment was im- Cryogenic liquids
plemented as a result of European Policy rule 4.4-5 is applicable to the
Directive 2003/10/EC dated 6 Febru- use of cryogenic liquids, in order to
ary 200311. Noise in the workplace is prevent undesirable incidents when
harmful if the daily average exposure working with dangerous materials.
of an employee on a nominal work- This is based on article 4.4 of the
ing day of 8 hours or more is greater Working Conditions Decree, in which
than 80 dB(A)12 [or a peak acoustic the specific circumstances contribut-
pressure of 112 Pa13]. If this is the ing to elevated risk levels from these
case, the employer must provide ear materials are named, such as the ex-
protection and arrange for informa- tremely low temperature that can
tion and training. If the noise level is cause severe frostbite and the ex-
higher on a daily basis than 85 dB(A) pansion during evaporation, which
[or a peak acoustic pressure greater can displace oxygen and cause suf-
than 140 Pa], the employee must focation.
wear ear protection and the employ-
er must take protective measures.
Whether or not sound causes harm
depends on how long the employee
is exposed to a particular sound level
10 Decree dated 25 January 2006 amending the Working Conditions Decree, containing rules relating
to the exposure of employees to the risks of noise. Staatsblad [Dutch Bulletin of Acts, Orders and
Decrees] 2006, number 56, 1 25.
11 Directive 2003/10/EC of the European Parliament and of the Council of 6 February 2003 on the
minimum health and safety requirements regarding the exposure of workers to the risks arising
from physical agents (noise) (Seventeenth individual Directive within the meaning of Article 16(1) of
Directive 89/391/EEC). OJ 2003, L42, 38.
12 The daily exposure is expressed in dB (decibels). For filtering, the A filter is generally used, which
corresponds to the sensitivity of the human ear; this is therefore referred to as dB(A).
13 The peak acoustic pressure is taken to mean the unweighted acoustic pressure that is measured
with a noise level meter in the peak hold position, expressed in Pa. This uses a frequency weighting
according to the C filtering. This measurement allows the maximum value (peak acoustic level) of brief
10 or momentary noise signals to be measured.
2 Practical risks immediately outside the ends of
the MRI scanner’s magnet core than
In the application, development, or the fields in the centre of the scan-
maintenance of MRI systems, em- ner, and it decreases rapidly with
ployees may be exposed to various distance from the system. The stray
types of risks, namely those result- field from the static magnetic field
ing from exposure to magnetic fields is always present, whereas the stray
and to noise, and as the result of fer- fields from the switched gradient
romagnetic projectiles, and risks de- fields and the RF fields are only pres-
riving from cryogenic liquids. ent during scanning.
11
MRI Working Group
The situation as regards the switched pends on the location and can reach
gradient fields needs examining levels of greater than 80 dB(A) in the
with greater care. These stray fields majority of systems. NEN3418 is gen-
are also considerably lower immedi- erally adopted as the measurement
ately outside the system than are the protocol.
fields in the centre of the system, but
it is nevertheless possible that these
stray fields might cause effects. This 2.3 Risks due to
depends on the scanning technique ferromagnetic
being used, because the strength projectiles
and the switching speed of the gra-
dient fields can vary.
In the static magnetic field of an MRI
Finally, a new technique is being de- scanner, there is the risk of ferromag-
veloped: MRI-guided interventions. netic items flying around. Employ-
This is a technique in which e.g. a ees could be injured as a result of an
physician or other professional prac- item turning into a projectile, or get
titioner carries out interventions trapped between the magnet and
with the help of MRI imaging, such the ferromagnetic material. This can
as taking biopsies, introducing cath- only happen in the scanner room it-
eters, and so forth. This may mean self.
that it is necessary for the employee
to be close to the magnet core dur-
ing scanning, with hands or head in 2.4 Risks from cryogenic
the switched gradient field and the liquids
RF field.
14 The term ‘quench’ refers to the process in which the superconductive coils of the magnet become
resistive. This converts the energy of the static field into heat, which causes part of the liquid helium
(and the liquid nitrogen, if present) to be vaporised. This large amount of gas is vented outside
through a special pipe, known as the quench pipe. There is a quench button to activate this process in
12 emergency situations.
tional circumstances, then parts of constructed in such a way that no
the sides of the magnet can become condensed air can drip onto the pa-
cold. This can clearly be seen from tient or employees in the event of a
the ice deposition and the formation quench.
of a “cloud”. The evaporating cryo-
genic liquids are vented outside via All the working situations mentioned
the quench pipe. Gaseous helium is here are the cases in which employ-
only released into the scanner room ees are present within the controlled
(with the concomitant danger of suf- access area and, in the majority of
focation) in those quenches where cases, in the scanner room. No spe-
this pipe leaks or is blocked. cial measures are applicable out-
side the controlled access area. The
working situation in which the con-
3 Defined working trolled access area extends outside
situations the scanner room is only discussed
tangentially here. This then refers to
adjacent work areas that are directly
The scanner room does not restrict related to the MRI application and to
the stray field from the switched which access is restricted for safety
gradient fields, but the strength of reasons, such as rooms for access,
this field is so low outside the scan- administration, operating the equip-
ner room that it can no longer have ment, and waiting. These rooms are
any effect. The stray field from the RF only accessible after controls (in-
fields is shielded by the ‘Faraday cage’ cluding screening and instruction)
that is constructed around the scan- and have the necessary warnings on
ner room and sometimes around the view.
scanner itself. The noise levels outside
the scanner room will in general be The relevant distinction between the
lower than the legal limits. If not, ad- various situations is the question of
ditional measures will be necessary whether or not scanning is being
for the adjacent rooms. The risks due performed, and where the employ-
to cryogenic liquids are also present ee is at that time. This means that
outside the scanner room. For this, three working situations can be dis-
the only checks required are that the tinguished for staff whose activities
quench pipe does not vent into areas have to be carried out in the vicinity
where staff could come into contact of MRI equipment.
with it and that the quench pipe is
13
MRI Working Group
III The situation in which a staff Areas for working situation II are only
member is partly or entirely accessible to staff who have been
inside the scanner during screened (for metal objects, metals
scanning. implants, active medical implants,
In this situation, the employee etc.), who have been given instruc-
can be exposed to risks as a result tion about the potential risks, and
of: who are strictly required in the scan-
15 A small bore system is an MRI scanner with a magnetic core of less than approximately 30 cm in
diameter. Such systems are mostly used for scanning small experimental animals and for scanning the
14 limbs of human beings.
ner room during scanning, due to Working situation III is applicable
the activities to be carried out. Work- when an employee:
ing situation II is also applicable to - is entirely or partially within
staff members who have to be close the scanner for intervention
to the magnet of a small bore system work17
during scanning, for example in or- - is in the scanner together
der to start an infusion. with a patient as part of
their work, for example
For each type of employer (hospital, when accompanying a child,
scientific research institute or MRI or
manufacturer), consideration must - has to be inside the scanner
be given to staff with the following e.g. to track down a fault
occupations and/or duties:
A Hospital A member of staff will be treated as
- radiological technologist a pregnant employee as soon as she
(radiographer)16 has reported to the employer that
- physician she is pregnant. Consideration may
- anaesthesiology staff be given to excluding a pregnant
member employee from working situations II
- nurse and III as a precautionary measure.
- technical or medical
researcher
- maintenance staff 4 Types of effects
- patient supervisor
B Scientific institute The term ‘effects’ is interpreted in a
- operator / biotechnician wide sense in this document, i.e. cov-
- researcher ering everything that could be asso-
- maintenance staff ciated with exposure to all the iden-
C MRI manufacturer tified risks. It is moreover important
- applications specialist that not all physical and biological
- researcher effects lead to transitory or perma-
- developer nent effects on health.
- system tester
- production staff member The ICNIRP distinguishes three types
- maintenance staff of effects that electromagnetic fields
can have on health (their terminol-
16 As of 17 April 2008, the new professional term for medical nuclear science workers, radio-diagnostic and
radio-therapeutic technicians in Dutch is MBB’er (Medisch Beeldvormings- en Bestralingsdeskundige
= medical imaging and radiation expert).
17 It is possible that people will in future carry out MRI interventions on experimental animals. In terms
of the exposure of the employees, these situations are comparable with working situations II and III 15
respectively.
MRI Working Group
18 ICNIRP. General approach to protection against non-ionizing radiation. Health Phys, 2002, 82(4):540-
16 548.
when the staff member tickling. The threshold at
moves within the static which this occurs varies from
magnetic field, with the one person to another.
speed and direction also b objective, transitory: muscular
being important. The fact contractions
that the static magnetic As a result of nerve
field outside the magnet stimulation, muscular
is not homogenous is also contractions can arise which
important. There is also may be painful (subjective).
variability in individual c objective, possibly
sensitivities to these effects. permanent harm: disruption
The stronger the magnetic of the functioning of active
field is, the more often such implanted medical devices
effects are reported. (AIMDs) such as pacemakers
d subjective, transitory: or infusion pumps
cognitive effects
Examples are a possible 3. RF fields
worsening of hand-eye a objective, possibly permanent
coordination and memory harm: direct heating of tissues
effects. Unlike other well- (although the effect itself is
documented effects, these transitory)
are only based on a few Tissue heating can cause
references in the literature. It burns and therefore give rise
is thought that these effects to permanent injury.
are caused by the above- b objective, possibly permanent
mentioned effects, such as harm: injuries as a result of
dizziness and nausea. There heating (although the effect
is also variability in individual itself is transitory) of metals or
sensitivities to these effects. implants containing metals,
wires, piercings, tattoos,
2. Switched gradient fields permanent makeup, etc.
a subjective, transitory: nerve This heating can cause burns
stimulation and therefore give rise to
Stimulation of the peripheral permanent injury.
nervous system in the limbs, c objective, possibly permanent
head and torso. The first harm: disruption of active
sensation observed is a slight implanted medical devices
17
MRI Working Group
18
as far away from the scanner as pos- fore advisable to automate infusions
sible. The greater the distance, the of anaesthetics, medications or con-
lower the exposure to the electro- trast liquids during scanning as fully
magnetic fields. As a precautionary as possible and to place operating
measure, it can also be recommend- consoles, anaesthesia equipment
ed that people do not remain in the and so forth as far away from the
scanner room longer than strictly magnet core as possible. It should
necessary, that additional measures also be noted here that the mini-
are taken to minimise exposure, and mum distance that should be kept
that awareness of the possible ef- depends strongly on the strength of
fects is encouraged. the RF field and the switched gradi-
ent field that the system can gener-
For exposure to the static magnetic ate. In practice, a distance of more
field in particular, in all the working than approximately 1m is sufficient
situations, an extra control measure to prevent the switched gradient
is that the changes in magnetic flux fields from having any effect. There is
density can be minimised by moving in practice no effective personal pro-
slowly through those areas where the tection equipment available yet for
static magnetic field gradient (dB0/ shielding against the switched gradi-
dx) is greatest, so that the product of ent fields. Marking out the 1m limit
the speed (dx/dt) and the field gradi- on the floor of the scanner room can
ent (dB0/dx) produces the smallest help make people aware of the risk.
possible value for the flux change
(dB0/dt). Within 1m of the aperture The following sections contain addi-
of the magnet core, the gradient of tional rules of conduct for the three
the static field is at a maximum and different working situations. Appen-
the employee must therefore move dix 2 lists the measures to be taken,
slowly. The gradient of the field is grouped by the type of area, working
much less at the sides of the mag- situation and type of activity.
net.
For both the RF field and the switched 5.2 Rules of conduct for
gradient field, the exposure can only working situation I
be minimised by keeping at a dis-
tance from the MRI scanner and by
avoiding tasks near to the magnet In this working situation, the impor-
core as much as possible. It is there- tant effects are those resulting from
19
MRI Working Group
21
MRI Working Group
22
is because the employment contract issues discussed in this document
or public law appointment does not and the practical rules needed. This
provide a basis for the infringement training must match the local situa-
of personal integrity that is involved tion and can be arranged e.g. by the
in acting as a test subject. Moreover, clinical physicist or MR physicist.
the Working Conditions Act states
that the employer must avoid psy- Clinical MRI scans will generally be
chosocial work stresses. The employ- carried out in hospitals or MRI cen-
ee is clearly in a dependent position tres. This means that the clinical in-
with respect to the employer, and vestigations will be performed by
feeling obliged to act as a test subject qualified radiological technologists16.
in order to maintain a good work- They will have been trained in how
ing relationship could cause stress. to deal with patients and in the inter-
Stress is deemed to be a psychoso- pretation of anatomical images. This
cial work burden. Test subjects must background also means that they
be recruited passively (no compul- will be familiar with the operation of
sion) and must be properly informed equipment such as X-ray and irradia-
about the possible risks. Handling of tion equipment. Since the arrival of
any incidental findings (reported or MRI, the basic principles of MRI have
not) must also be explained clearly also been included in the initial train-
to the test subject beforehand. It is ing of the radiological technologists,
desirable that the ‘agreement’ be- also covering the safety aspects.
tween the employer and the em-
ployee should be recorded in writing One possible way for technnolo-
by means of an informed consent gists working with MRI to obtain the
form. required extra training regarding
safety issues is by following a post-
HBO [higher vocational education]
5.6 Education, training and MRI course. This course emphasizes
instruction the theory and techniques required
to be able to perform the optimum
examinations. This addresses the
Given the central role of the technol- risks and effects of electromagnetic
ogist/operator on the work floor in fields once again, plus the major im-
implementing the working methods portance of screening everyone who
indicated here, such employees must enters the scanner room properly.
be trained regarding the MRI safety
23
MRI Working Group
Staff who do not belong to the group Staff members who do not belong
mentioned above and who carry out to the groups mentioned above and
their duties in working situation I who may have to carry out their du-
must be screened by the qualified ties in working situations II or III must
technologist/operator and informed also have followed a training course
about the potential risks. If there are about safety. Given the importance
practical reasons why this is not pos- of safety, MRI operators must regu-
sible, these employees must follow a larly receive refresher training re-
separate safety training course. This garding safety aspects (for further
training must be tuned to match the information, please refer to the ACR
local situation, the equipment at the Guidance Document 200720).
site, and the activities performed.
20 ACR Blue Ribbon Panel on MR Safety (Kanal et al.). ACR Guidance Document for Safe MR Practices:
24 2007. American Journal of Roentgenology, 188:1–27, June 2007.
Appendix 1
Members of the MRI
Working Group
25
26
type of room Working situation activity risk measures
adjacent working areas Various, no MRI ap- access, longer presence, sMF * < 0.5 mT none
and corridors, outside the plication walking routes
controlled access area
MRI Working Group
technical areas, outside Various, no MRI ap- maintenance release of cryogenic gases protect the quench pipe; warning
the controlled access area plication signals; detection if required
sMF< 0,5 mT none
Appendix 2
MRI staff
controlled access area: Area where access is access sMF> 0,5 mT screening; instruction; access con-
area outside the scanner restricted for safety trol; warning signals;
room reasons
administration; operation sMF> 0,5 mT none
waiting sMF> 0,5 mT none
cleaning sMF> 0,5 mT none
controlled access area: Working situation I: accessing the room to moving in a sMF instruction; move slowly
scanner room situation in which a staff position or fetch the
ferromagnetic projectiles special tools; screening; instruction
Task and risk analysis for
27
Using MRI safely
PRACTICAL RULES
FOR EMPLOYEES
Holland HealthTech