Hote
Hote
Foreword
This booklet was developed as a guide to biosafety cabinets that provides basic
knowledge of biosafety and biosafety cabinets. The information presented
is unbiased and generic in nature compiled with help from experienced
microbiologists, engineers and safety enclosure users.
Biosafety Level 1
Practices, safety equipment and facilities appropriate for work with defined
and characterized strains of viable micro organisms not known to cause
disease in healthy adult humans. The laboratory is not necessarily separated
from the general traffic patterns in the building. Work is generally conducted
on open bench tops using standard microbiological practices. Special
containment equipment or facility design is neither required nor generally
used. Laboratory personnel have specific training in the procedures conducted
in the laboratory and are supervised by a scientist with general training in
microbiology or a related science. A biohazard safety cabinet is generally not
required for work involving these agents.
Biosafety Level 2
Practices, safety equipment and facilities appropriate for work done with
a broad spectrum of indigenous moderate-risk agents present in the
community and associated with human disease in varying severity. It differs
from biosafety level 1 in that:
a) laboratory personnel have specific training in handling pathogenic agents
and are directed by competent scientists;
b) access to the laboratory is limited when work is being conducted;
c) extreme precautions are taken with contaminated sharp items; and
d) certain procedures in which infectious aerosols or splashes may be
created are conducted in biological safety cabinets or other physical
containment equipment. A Class I or Class II biohazard safety cabinet is
highly recommended for work involving these agents.
Biosafety Level 3
Practices, safety equipment and facilities appropriate for work done with
indigenous or exotic agents with a potential for respiratory transmission
2
A Guide To Biosafety And Biosafety Cabinets
which may cause serious and potentially lethal infection. More emphasis
is placed on primary and secondary barriers to protect personnel in the
contagious area, the community, and the environment from exposure to
potentially infectious aerosols. A Class I or Class II biohazard safety cabinet
is required for work involving these agents.
Biosafety Level 4
Practices, safety equipment and facilities appropriate for work done with
dangerous and exotic agents which pose a high risk of life threatening
disease. May be transmitted via the aerosol route, and for which there is no
available vaccine or therapy. Members of the laboratory staff have specific
and thorough training in handling extremely hazardous infectious agents
and they understand the primary and secondary containment functions
of the standard and special practices, the containment equipment, and
the laboratory design characteristics. They are supervised by competent
scientists who are trained and experienced in working with these agents.
Access to the laboratory is strictly controlled by the laboratory director. The
facility is either in a separate building or in a controlled area within a building,
which is completely isolated from all other areas of the building. A specific
facility operations manual is prepared or adopted. A Class III biohazard
safety cabinet is required for work involving these agents.
3
Table 1. Summary of Biosafety Levels and Infections Agents
Bacillus subtilis,
Not known to
Naegleria gruberi, Standard
consistently cause Open benchtop
1 Infectious canine, Microbiological None Required
disease in healthy sink required.
hepatitis virus, practices
adults
[Link]
required needed
Level 2 plus:
4
Primary barriers:
Indigenous or Level 2 plus: • Physical separation
M. Tuberculosis, Class I or II BSCs or other physical
exotic agents with • Controlled access from access corridors
St. louis containment devices used for all open
potential for aerosol • Decontamination of waste • Self-closing, double
encephalitis virus, manipulations of agents.
3 transmission, • Decontamination of lab door access
Coxiella Burnetii,
disease may have clothing • Exhaust air not
Bacillus anthracis Personal protective equipment:
serious or lethal • Baseline serum samples of recirculated
(production level) lab clothing, gloves, respiratory
consequences lab personnel • Negative airflow into
protection as needed
laboratory
Level 3 plus:
• Separate building or
Dangerous / exotic
isolated zone
agents that pose
Level 3 plus: • Dedicated supply/
a high risk of life Primary barriers:
• Clothing change before entry exhaust vacuum,
threatening disease. Ebola zaire, Sin All procedures conducted in Class III
into lab and decontamination
4 Aerosol-transmitted Nombre Virus, Rift or I or II BSCs in combination with
• Shower on exit systems
lab infections or Valley Fever full-body, air-supplied, positive-
• All material decontaminated • Additional
related agents with pressure personnel suit
on exit from lab requirements as
uknown risk of
outlined in Biosafety
transmission
in Microbiological and
Biomedical Laboratories
Source: Biosafety in Microbiology and Biomedical Laboratories,U.S. Department of Health and Human Services, HHS publication (CDC) 99-8395, 4th ed. p 52-53.
A Guide To Biosafety And Biosafety Cabinets
The terms biological safety cabinet and biosafety cabinet have been widely
used to describe a variety of containment devices equipped with HEPA filter(s),
designed to provide personnel or both personnel and product protection from
biohazardous materials. The terms should only be applied to those devices
that meet the requirements of Class I, II, or III specifications, based on their
construction, airflow velocities and patterns, and their exhaust systems.
5
A Guide To Biosafety And Biosafety Cabinets
The Class I cabinet has the most basic and rudimentary design of all biohazard
safety cabinetry available today. A stream of inward air moving into the
cabinet contains aerosols generated during microbiological manipulations.
It then passes through a filtration system that traps all airborne particles
and contaminants. Finally, clean, decontaminated air is exhausted from the
cabinet. The filtration system usually consists of a pre-filter and a HEPA
(High Efficiency Particulate Air) filter.
Although the Class I cabinet protects the operator and the environment from
exposure to biohazards, it does not prevent samples being handled in the
cabinet from coming into contact with airborne contaminants that may be
present in room air. Naturally, there is a possibility of cross-contamination that
may affect experimental consistency. Consequently the scope and application
of Class I cabinets is limited and it is largely considered obsolete.
All Class I biohazard safety cabinets are suitable for work with microbiological
agents assigned to biological safety levels 1, 2 and 3.
Like Class I safety cabinets, Class II cabinets have a stream of inward air
moving into the cabinet. This is known as the inflow and it prevents the
aerosol generated during microbiological manipulations to escape through
the front opening. However, unlike Class I cabinets, the inflow on Class II
cabinets flows through the front inlet grille, near the operator. None of the
unfiltered inflow air enters the work zone of the cabinet, so the product
inside the work zone is not contaminated by the outside air.
The differences between the various Class II cabinets available lie primarily
with the percentage of air exhausted to that of air re-circulated from the
common air plenum. In addition, different Class II cabinets have different
means of cabinet exhaust. Some cabinets may exhaust air directly back to
the laboratory, while others may exhaust air through a dedicated ductwork
system to the external environment.
Despite these differences, all Class II cabinets, like Class I cabinets, protect
both the operator and environment from exposure to biohazards. In addition,
Class II cabinets also protect product samples from contamination during
microbiological manipulations within the cabinet interior and are all suitable
for work with agents assigned to biological safety levels 1, 2 and 3.
6
A Guide To Biosafety And Biosafety Cabinets
The Class II Type A biohazard safety cabinet is the most common Class II
cabinet. It is also the most common safety cabinet of all the different types
available. It has a common plenum from which 30% of air is exhausted, and
70% re-circulated to the work area as the downflow.
Type A cabinets exhaust air directly back to the laboratory, and they may
contain positive pressure contaminated plenums. When toxic chemicals must
be employed as an adjunct to microbiological processes, these cabinets
should not be used. Exhaust HEPA filtration only removes airborne aerosols
including biohazards, and not chemical fumes.
The main difference between Type A and Type B cabinet is: Type B cabinets
must be operated with an external blower and it exhausts air to the external
environment via a dedicated ductwork system. Without the external blower,
the cabinets internal blower will blow the air (and microbiological agents)
inside the work zone through the front operator, towards the operators face,
creating a dangerous situation. This cabinet is not self-balancing in the sense
that its own blower can only create downflow and the cabinet relies on the
external blower to create inflow.
7
A Guide To Biosafety And Biosafety Cabinets
Although Type B cabinets are commonly used when chemicals are involved in
your work processes, they theoretically provide an increased level of safety
as compared to other Type A cabinets. By exhausting air directly to the
external environment, they provide an additional “fail-safe” in the event that
the regular exhaust HEPA filtration ceases to function.
The Class II Type B1 biohazard safety cabinet was originally specified by the
American National Cancer Institute. It has a common plenum from which
70% of air is exhausted, and 30% re-circulated to the work area as the
downflow.
In the Class II Type B2 cabinet all inflow and downflow air is exhausted after
HEPA filtration to the external environment without recirculation within the
cabinet. Type B2 cabinets are suitable for work with toxic chemicals employed
as an adjunct to microbiological processes under all circumstances since no
re-circulation occurs. In theory, Type B2 cabinets may be considered to be
the safest of all Class II biohazard safety cabinets since the total exhaust
feature acts as a fail-safe in the event that the downflow and / or exhaust
HEPA filtration systems cease to function normally.
8
A Guide To Biosafety And Biosafety Cabinets
The Class III biohazard safety cabinet provides an absolute level of safety,
which cannot be attained with Class I and Class II cabinets. All Class III
cabinets are usually of welded metal construction and are designed to
be gastight. Work is performed through glove ports in the front of the
cabinet. During routine operation, negative pressure relative to the ambient
environment is maintained within the cabinet. This provides an additional
fail-safe mechanism in case physical containment is compromised.
On all Class III cabinets, a supply of HEPA filtered air provides product
protection and prevents cross contamination of samples. Exhaust air is
usually HEPA filtered and incinerated. Alternatively, double HEPA filtration
with two filters in series may be utilized. Materials are transferred into the
cabinet using a pass-through unit installed at the side of the work area. Class
III cabinets usually exhaust air back to the laboratory; however, air may also
be exhausted via a dedicated ductwork system to the external environment.
When a dedicated ductwork system is employed, they are also suitable for
work employing toxic chemicals as an adjunct to microbiological processes.
All Class III biohazard safety cabinets are suitable for work with microbiological
agents assigned to biological safety levels 1, 2, 3 and 4. They are frequently
specified for work involving the most lethal biological hazards.
9
A Guide To Biosafety And Biosafety Cabinets
The HEPA filter is the heart of the biosafety cabinet. The HEPA filter is a
disposable dry-type filter, constructed of boron silicate microfibers cast into
a thin sheet, much like a piece of paper
Fig 5. Pack is glued into a frame - Frame is gasketed to form final assembly
10
A Guide To Biosafety And Biosafety Cabinets
Volatile radionuclide / chemical protection, Class IIB1, Class IIA2 vented to the
minute amounts outside
Volatile or toxic chemicals should not be used in BSCs that recirculate exhaust
air to the room, i.e. Class I that are not ducted to building exhaust systems,
or Class IIA1 or Class IIA2 cabinets. Class II B1 cabinets are acceptable
for work with minute amounts of volatile chemicals and radionuclides. A
Class IIB2 cabinet, also called a total exhaust cabinet, is necessary when
significant amounts of radionuclides and volatile chemicals are expected to
be used.
11
A Guide To Biosafety And Biosafety Cabinets
1. Location
The velocity of air flowing through the front opening into a BSC is about 0.45
m/s. At this velocity the integrity of the directional air inflow is fragile and
can be easily disrupted by air currents generated by people walking close
to the BSC, open windows, air supply registers, and opening and shutting
doors.
2. Operators
If BSCs are not used properly, their protective benefits may be greatly
diminished. Operators need to be careful to maintain the integrity of the front
opening air inflow when moving their arms into and out of cabinets. Arms
should be moved in and out slowly, perpendicular to the front opening.
BSC
BSC Biological Safety Cabinet
Personnel Traffic
Barrier Airflow
BSC BSC
Air Inlet
�
BSC
BSC
12
A Guide To Biosafety And Biosafety Cabinets
3. Material Placement
The front intake grill of Class II BSCs must not be blocked with paper,
equipment or other items. Materials to be placed inside the cabinet should
be surface-decontaminated with 70% alcohol. Work may be performed on
disinfectant-soaked absorbent towels to capture splatters and splashes.
All materials should be placed as far back in the cabinet, towards the rear
edge of the work surface, as practical without blocking the rear grill. Aerosol-
generating equipment (e.g. mixers, centrifuges, etc.) should be placed
towards the rear of the cabinet. Bulky items, such as biohazard bags, discard
pipette trays and suction collection flasks should be placed to one side of the
interior of the cabinet. Active work should flow from clean to contaminated
areas across the work surface.
The autoclavable biohazard collection bag and pipette collection tray should
not be placed outside the cabinet. The frequent in-and-out movement
needed to use these containers is disruptive to the integrity of the cabinet’s
air barrier, and can compromise both personnel and product protection.
Cabinets should be turned on at least 5 min before beginning work and after
completion of work to allow the cabinet to “purge”, i.e. to allow time for
contaminated air to be removed from the cabinet environment.
5. Ultraviolet Lights
Ultraviolet lights are not required in BSCs. If they are used, they must be
cleaned weekly to remove any dust and dirt that may block the germicidal
effectiveness of the light. Ultraviolet light intensity should be checked
when the cabinet is recertified to ensure that light emission is appropriate.
Ultraviolet lights must be turned off while the room is occupied, to protect
eyes and skin from inadvertent exposure.
6. Open Flames
Open flames should be avoided in the near microbe-free environment created
inside the BSC. They disrupt the air flow patterns and can be dangerous when
volatile, flammable substances are also used. To sterilize bacteriological
13
A Guide To Biosafety And Biosafety Cabinets
7. Spills
A copy of the laboratory’s protocol for handling spills should be posted,
read and understood by everyone who uses the laboratory. When a spill of
biohazardous material occurs within a BSC, cleanup should begin immediately,
while the cabinet continues to operate. An effective disinfectant should be
used and applied in a manner that minimizes the generation of aerosols. All
materials that come into contact with the spilled agent should be disinfected
and / or autoclaved.
8. Annual Certification
The functional operation and integrity of each BSC should be certified to
national or international performance standards at the time of installation and
regularly thereafter by qualified technicians, according to the manufacturer’s
instructions.
Evaluation of the effectiveness of cabinet containment should include tests
for cabinet integrity, HEPA filter leaks, downflow velocity profile, face velocity,
negative pressure/ventilation rate, air flow smoke pattern, and alarms and
interlocks.
Optional tests for electrical leaks, lighting intensity, ultraviolet light intensity,
noise level and vibration may also be conducted. Special training, skills and
equipment are required to perform these tests and it is highly recommended
that they are undertaken by a qualified professional.
10. Decontamination
BSCs must be decontaminated before filter changes and before being
moved. The most common decontamination method is by fumigation with
formaldehyde gas. BSC decontamination should be performed by a qualified
professional.
14
A Guide To Biosafety And Biosafety Cabinets
12. Alarms
BSCs can be equipped with one of two kinds of alarm. Sash alarms are found
only on cabinets with sliding sashes. The alarm signifies that the operator
has moved the sash to an improper position. Corrective action for this type
of alarm is returning the sash to the proper position. Air-flow alarms indicate
a disruption in the cabinet’s normal air-flow pattern. This represents an
immediate danger to the operator or product. When an air-flow alarm sounds,
work should cease immediately and the laboratory supervisor should be
notified. Manufacturers’ instruction manuals should provide further details.
Training in the use of BSCs should cover this aspect.
15
A Guide To Biosafety And Biosafety Cabinets
3. Bunsen Burner
Burners can cause an outflow of air from inside the cabinet that can endanger
the operator. On Class II cabinets, protection against cross contamination
may also be compromised. There have been incidents where the sash of a
recirculating Class II safety cabinet was closed while the burner was still on,
causing heat build up in the cabinet that damaged the filters.
16
A Guide To Biosafety And Biosafety Cabinets
3. The burner should be placed as far into the cabinet working space as
possible to minimize its effect on cabinet performance.
4. A solenoid valve can be interlocked with the gas supply to automatically
shut off the burner
5. User maintenance
Pre-filter replacements once every 3 months
UV lamp replacement once a year
Operational verification if appropriate (e.g. Alarm checks, basic airflow
measurements etc.)
17
A Guide To Biosafety And Biosafety Cabinets
18
A Guide To Biosafety And Biosafety Cabinets
V. References:
19
PRODUCTS AVAILABLE FROM ESCO BIOTECH:
Cytotoxic Cabinets
IVF Cabinets
Weighing Cabinets
Animal Handling Workstations
1.5 meters (5') AC2-5E_ AC2-5S_ NIL Note: The last digit of the Model Number _ indicates desired
electrical service (60Hz, 50Hz). See the Specification Chart for
1.8 meters (6') AC2-6E_ AC2-6S_ AC2-6D_
details. Specify voltage/Hz when ordering.
Biological Safety Cabinet • Class II Microbiological Safety Cabinets (S-Series, E-Series and Airstream Duo)
Improved lighting is brighter, more Esco antimicrobial surface
uniform, reduces glare. on all painted surfaces minimizes
contamination.
Optional UV lamp is located behind
control panel away from line of sight; Integrated RFI and electrical noise filters
lamp operates on programmable timer. eliminate interference with and from
adjacent equipment.
Biological Safety Cabinet • Class II Microbiological Safety Cabinets (E-Series, S-Series and D-Series)
Touchpad data entry buttons permit Programmable automatic UV light A graphical interface indicates cabinet performance.
control settings and access to diagnostics, timer simplifies operation, enhances
default settings and hierarchical menus. contamination control, extends UV Digital read-out with alpha-numeric display indicates
lamp life and saves energy. all input, status and alarm functions.
Color coded indicator lamps display green
for primary function (fan operation); blue
All functions can be user activated through touch-pad
for secondary function (fluorescent lights
programming access; see Operations Manual.
and electrical outlet); and orange for
caution (UV lamp ON).
Sentinel Microprocessor Control System, Programmable When programmed ON • the start-up sequence confirms status with Air Safe and local time display.
• the Personal Identification Number (PIN) access restricts unauthorized adjustments.
• an airflow alarm warns of deviations from normal velocities.
• Optional service fittings openings are Electrical Safety and Certification Support Stands
offset for easier access. All components meet or exceed applicable • Fixed height, available 737 mm (29")
safety requirements. or 838 mm (33"), ±38.1 mm (1.5")
Comfortable Ergonomic Design - With leveling feet
The cabinet is engineered for comfort, • Each cabinet is individually factory - With casters
utility value and safety. tested for electrical safety. • Adjustable height, hydraulic range
• The angled viewing window and narrow • Documentation specific to each cabinet 737 mm (29") to 838 mm (33")
profile front grille improves reach into serial number is maintained on file. - With casters
the work area. • Certified to major world standards • Telescoping height stand, nominal
• The instant-start 5000k fluorescent for microbiological safety cabinets, range 737 mm (29") or 838 mm (33")
lamp operates on an electronic ballast including the EN12469 (refer to technical - Adjustable in 25.4 mm
to reduce heat, improve comfort and specifications for the full list). (1") increments
conserve energy. • Contact Esco or your Sales Representative
• Infinitely adjustable cradle stand,
• The lamp delivers uniform lighting to for site preparation information; see
with casters
the work surface for greater comfort, Electrical Specifications. - Elevates to seating or standing work
reduced glare and improved productivity; surface height
Warranty
see Specifications. - When lowered permits movement
Airstream AC2 Series cabinets are war-
• The front armrest is raised above the through standard doorway
ranted for 3 years excluding consumable Note: Increases exterior dimensions
work zone to improve comfort and to
parts and accessories.
minimize blockage of forward airflow Electrical Outlets and Utility Fixtures
perforations. • Each cabinet is shipped with a com-
• Electrical outlet, ground fault,
prehensive User’s Manual complete
• The optional adjustable support stand North America
with a report documenting all test
provides work surface height control. • Electrical outlet, Europe / Worldwide
procedures.
• The frameless sash eliminates operator’s • Petcock (air, gas, vacuum)
• Additional IQ/OQ/PQ documentation
line of sight blockage. - North America (American) style
is available upon request.
• A generous sash opening allows for - Europe / Worldwide style DIN 12898,
• Contact your local Sales Representative
easier access into the work zone, pro- DIN 12919, DIN 3537
for specific warranty details or docu-
vides ample room for transferring of Cabinet Accessories
mentation requests.
small equipment; see Specifications. • Germicidal UV lamp
• The sliding window can be fully Accessories and Options - Controlled by automatic UV lamp
TM
opened to insert and remove larger timer through Sentinel micro-
Esco offers a variety of options and
instrumentation and equipment. processor control panel
accessories to meet local applications.
- Emission of 253.7 nanometers for
Contact Esco or your local Sales Represen-
most efficient decontamination
tative for ordering information.
Cabinet Filtration System
Stainless Steel Side Walls/ Tempered Glass
1
8 2
1360 mm (52.9")
11 9
4
10
12
7
560 mm (22")
1035 / 1340 / 1645 / 1950 mm 740 mm (29.1")
(40.7" / 52.7" / 64.7" / 76.7")
1. Exhaust ULPA filter 5. Plugged service fixture provisions 7. Electrical outlet Retrofit Kit provision 10. IV bar Retrofit Kit provision
2. Blower (2 on each side) (2 & 3 ft model - only 1 single outlet 11. Esco Sentinel microprocessor
6. Stainless steel work zone in work zone left side wall) control system
3. Downflow ULPA filter
8. Electrical panel 12. Sash window
4. Standard UV light Retrofit Kit
Provision 9. Fluorescent lamp
Esco Airstream cabinets use a combination of high performance scroll blowers (supply) and German made
ebm-papst® permanently lubricated, centrifugal motor/blowers with external rotor designs (exhaust).
Selected for energy efficiency, compact design, and flat profile, the completely integrated exhaust blower
assembly optimizes motor cooling, with unified rotating parts and overall component balance for smooth,
quiet, vibration-free operation.
Weight is equally distributed to all bearings to extend bearing life, transfer heat and maximize speed control.
Biological Safety Cabinet • Class II Microbiological Safety Cabinets (E-Series, S-Series and D-Series)
Airstream Model AC2 (S-Series) Biological Safety Cabinet Technical Specifications,
Stainless Steel Side Walls
2
8
1360 mm (52.9")
11 9
4
12 10
1. Exhaust ULPA filter 5. Plugged service fixture provisions 7. Electrical outlet Retrofit Kit provision 10. IV bar Retrofit Kit provision
2. Blower (2 on each side) (2 & 3 ft model - only 1 single outlet 11. Esco Sentinel microprocessor
6. Stainless steel work zone in work zone left side wall) control system
3. Downflow ULPA filter
8. Electrical panel 12. Sash window
4. Standard UV light Retrofit Kit
Provision 9. Fluorescent lamp
Antimicrobial Powder-Coating
Ambient air is pulled through the The uniform, non-turbulent air stream
perforations located towards the work protects against cross contamination
zone front to prevent contamination of within and throughout the work area.
the work surface and work product. The
Near the work surface, the downflow
inflow does not mix with the clean air
air stream splits with a portion moving
within the cabinet work zone. Inflow air
toward the front air grille, and the
travels through a return path toward the
remainder moving to the rear air grille.
common air plenum (blower plenum) at
A small portion of the ULPA filtered
the top of the cabinet.
downflow enters the intake perforations
Dual blowers and dual exhaust filters at the side capture zones at a higher
provide an added measure of protection. velocity (small blue arrows).
If the primary blower fails, the secondary
A combination of inflow and downflow
blower still pushes the air across the
air streams forms an air barrier that
exhaust filters to maintain inflow and
prevents contaminated room air from
containment.
entering the work zone, and prevents
Approximately 35% of the air in the work surface emissions from escaping
common plenum is exhausted through the work zone.
the ULPA filter to the room. The
Air returns to the common air plenum
remaining 65% of the air is passed
where the 35% exhaust and 65%
through the downflow ULPA filter and
ULPA-filtered air recirculation process is continued.
into the work area as a vertical laminar
Unfiltered / potentially contaminated air flow air stream bathing the work surface
Room air / Inflow air in clean air.
Airstream Duo Model AC2 (D-Series) Biological Safety Cabinet Technical Specifications,
Stainless Steel Side Walls, Dual Blowers
9 1
3
1400 mm (55.1")
1475 mm (58")
10
11 4
5
670 mm (26.4")
12
6
1. Exhaust ULPA filter 5. IV bar Retrofit Kit provision 8. Electrical outlet Retrofit Kit provision 10. Fluorescent lamp
2. Blower 6. Plugged service fixture provisions (2 & 3 ft model - only 1 single outlet 11. Esco Sentinel microprocessor
(2 on each side) in work zone left side wall) control system
3. Downflow ULPA filter
7. Stainless steel work zone 9. Electrical panel 12. Tempered glass sash window
4. Standard UV light Retrofit Kit
Provision
Biological Safety Cabinet • Class II Microbiological Safety Cabinets (E-Series, S-Series and D-Series)
General Specifications, Airstream Class II, Biological Safety Cabinets (E-Series)
Note to customer: Insert electrical voltage number into last model number digits _ when ordering
Model AC2-2E_ AC2-3E_ AC2-4E__ AC2-5E__ AC2-6E__
Nominal Size 0.6 meters (2') 0.9 meters (3') 1.2 meters (4') 1.5 meters (5') 1.8 meters (6')
External Dimensions 730 x 740 x 1435 mm 1035 x 740 x 1360 mm 1340 x 740 x 1360 mm 1645 x 740 x 1360 mm 1950 x 740 x 1360 mm
(W x D x H) 28.7 " x 29.1" x 56.5" 40.7" x 29.1" x 52.9" 52.7" x 29.1" x 52.9" 64.7" x 29.1" x 52.9" 76.7" x 29.1" x 52.9"
Internal Work Area, Dimensions 660 x 560 x 670 mm 965 x 560 x 670 mm 1270 x 560 x 670 mm 1575 x 560 x 670 mm 1880 x 560 x 670 mm
(W x D x H) 26.0" x 22.0" x 26.4" 38.0" x 22.0" x 26.4" 50.0" x 22.0" x 26.4" 62.0" x 22.0" x 26.4" 74.0" x 22.0" x 26.4"
Internal Work Area, Space 0.29 m2 (3.1 [Link]) 0.43 m2 (4.6 [Link]) 0.58 m2 (6.2 [Link]) 0.73 m2 (7.7 [Link]) 0.87 m2 (9.3 [Link])
Average Inflow 0.45 m/s (90 fpm) at initial setpoint, audible/visual alarm will activate at 0.40 m/s (80 fpm)
Airflow
Velocity Downflow 0.30 m/s (60 fpm) at initial setpoint with uniformity of better than +/- 20%
Inflow 185 m3 /h (109 cfm) 270 m3 /h (159 cfm) 356 m3 /h ( 209 cfm) 441 m3 /h (260 cfm) 527 m3 /h (310 cfm)
Airflow
Downflow 423 m /h (249 cfm)
3
563 m /h (331 cfm)
3
741 m /h (436 cfm)
3
919 m /h (591 cfm)
3
1096 m3 /h (645 cfm)
Volume
Exhaust 185 m3 /h (109 cfm) 270 m3 /h (159 cfm) 356 m3 /h (209 cfm) 441 m3 /h (260 cfm) 527 m3 /h (310 cfm)
ULPA Filter Downflow >99.999% at 0.1 to 0.3 microns as per IEST-RP-CC001.3 USA
Typical
Efficiency Exhaust >99.995% at MPPS as per EN 1822 (H-14) EU
Sound NSF 49 <65 dBA <64 dBA <65 dBA <65 dBA <66 dBA
Emission
(Typical) EN 12469 <62 dBA <61 dBA <62 dBA <62 dBA <63 dBA
Fluorescent Light Intensity At Zero >900 Lux >1130 Lux >1280 Lux >1050 Lux >1220 Lux
Ambient (> 84 foot candles) (>105 foot candles) (>119 foot candles) (>97 foot candles) (>113 foot candles)
1.2 mm (0.06") 16 gauge electrogalvanized steel with white oven-baked epoxy Isocide
Main Body
antimicrobial powder coated finish
Cabinet
Construction Work Zone 1.5 mm (0.06") 16 gauge stainless steel, type 304, with 4B finish
for 48 hours and the Colony Forming • Target slit air samplers and impingers • Target agar plates are placed through-
Units (CFU) are counted to determine the are placed outside the work zone to out the entire work surface.
testing results. Increased microbiological capture possibly escaping [Link] • Acceptance: The number of Bacillus
challenge tests with objects inside the spores, then the sample is incubated. atrophaeus CFU recovered from the
cabinet work zone, Bunsen burner, • Acceptance: The number of Bacillus agar plates shall not exceed 5 CFU per
external airflow disturbance, and Human- atrophaeus CFU recovered from the test.
As-Mannequin test adapted from Fume agar plates shall not exceed 10 CFU
Hood development were performed to per test. Cross Contamination Test
simulate real-world conditions.
The test objective is to evaluate cabinet
Personnel Protection Test Product Protection Test protection from cross contamination of
The test objective is to determine cabinet samples placed simultaneously inside the
The test objective is to evaluate the safety
protection to the product/samples inside work zone.
of the cabinet for the personnel operating
the cabinet work zone from environmental • A nebulizer containing 55 mL of spores
on potentially hazardous samples in the
contaminants. (5 to 8 x 104/mL) is placed against one
cabinet work zone.
• A nebulizer containing 55 mL of 5 to 8 of the work zone sidewalls.
• A nebulizer containing 55 mL of 5 to
x 106 spores/mL [Link] is placed • Target agar plates are placed 360 mm
8 x 108 spores/mL [Link] spores
at 10 cm (4") in front of sash window. (14") away from the same side wall
is placed inside the work zone, 10 cm
(4") behind the front opening sash.
General Specifications, Airstream Class II, Biological Safety Cabinets (S-Series)
Note to customer: Insert electrical voltage number into last model number digits _ when ordering
Model AC2-3S_ AC2-4S_ AC2-5S_ AC2-6S_
Nominal Size 0.9 meters (3') 1.2 meters (4') 1.5 meters (5') 1.8 meters (6')
External Dimensions 1035 x 740 x 1360 mm 1340 x 740 x 1360 mm 1645 x 740 x 1360 mm 1950 x 740 x 1360 mm
(W x D x H) 40.7" x 29.1" x 52.9" 52.7" x 29.1" x 52.9" 64.7" x 29.1" x 52.9" 76.7" x 29.1" x 52.9"
Internal Work Area, Dimensions 965 x 560 x 670 mm 1270 x 560 x 670 mm 1575 x 560 x 670 mm 1880 x 560 x 670 mm
(W x D x H) 38.0" x 22.0" x 26.4" 50.0" x 22.0" x 26.4" 62.0" x 22.0" x 26.4" 74.0" x 22.0" x 26.4"
Internal Work Area, Space 0.43 m2 (4.7 [Link]) 0.58 m2 (6.2 [Link]) 0.73 m2 (7.7 [Link]) 0.87 m2 (9.3 [Link])
Average Airflow Inflow 0.45 m/s (90 fpm) at initial setpoint, audible/visual alarm will activate at 0.40 m/s (80 fpm)
Velocity Downflow 0.30 m/s (60 fpm) at initial setpoint with uniformity of better than +/- 20%
Inflow 270 m3 /h (159 cfm) 356 m3 /h ( 209 cfm) 441 m3 /h (260 cfm) 527 m3 /h (310 cfm)
Airflow Volume Downflow 563 m3 /h (331 cfm) 741 m3 /h (436 cfm) 919 m3 /h (591 cfm) 1096 m3 /h (645 cfm)
Exhaust 270 m /h (159 cfm)
3
356 m /h (209 cfm)
3
441 m /h (260 cfm)
3
527 m3 /h (310 cfm)
ULPA Filter Downflow >99.999% at 0.1 to 0.3 microns as per IEST-RP-CC001.3 USA
Typical
>99.995% at MPPS as per EN 1822 (H-14) EU
Efficiency Exhaust
Sound Emission NSF / ANSI 49 <64 dBA <65 dBA <65 dBA <66 dBA
(Typical) EN 12469 <61 dBA <62 dBA <62 dBA <63 dBA
Fluorescent Light Intensity At Zero >1040 Lux >1190 Lux >920 Lux >1020 Lux
Ambient (>97 foot candles) (>111 foot candles) (>85 foot candles) (>95 foot candles)
1.2 mm (0.06") 16 gauge electrogalvanized steel with white oven-baked epoxy Isocide
Main Body
antimicrobial powder coated finish
Cabinet
Construction Work Zone 1.5 mm (0.06") 16 gauge stainless steel, type 304, with 4B finish
• Acceptance: The number of Bacillus Discus test for our customers. The KI and is useful for validating the actual
atrophaeus CFU recovered on agar plates Discus test is defined in the European containment performance of the
shall not exceed 2 CFU per test. Standard for microbiological safety cabinet on-site.
cabinets, EN12469:2000, as a test method • The KI-Discus takes only 45 minutes as
KI Discus Containment Test for validating the operator/personnel opposed to 2 days for microbiological
According to EN 12469:2000 protection capabilities of the cabinet. testing.
(Potassium Iodide) • The KI Discus test shows excellent • Thus, each Esco Airstream FC2 model
Esco is currently one of the few companies correlation with the microbiological is type tested using the KI Discus
in the world equipped to perform the KI test method for operator protection, method for operator safety.
For Microbiological
For Air Quality For Filtration For Electrical Safety
Safety Cabinets*
* For details on most recent certifications contact Esco or your Esco Sales Representative.
Biological Safety Cabinet • Class II Microbiological Safety Cabinets (E-Series, S-Series and D-Series)
General Specifications, Airstream Class II, Biological Safety Cabinets (D-Series)
Note to customer: Insert electrical voltage number into last model number digits _ when ordering
Model AC2-4D_ AC2-6D_
Average Airflow Inflow 0.45 m/s (90 fpm) at initial setpoint, audible/visual alarm will activate at 0.40 m/s (80 fpm)
Velocity Downflow 0.30 m/s (60 fpm) at initial setpoint with uniformity of better than +/- 20%
Inflow 270 m3 /h (159 cfm) 527 m3 /h (310 cfm)
Airflow Volume Downflow 563 m3 /h (331 cfm) 1096 m3 /h (645 cfm)
Exhaust 270 m /h (159 cfm)
3
527 m3 /h (310 cfm)
ULPA Filter Downflow >99.999% at 0.1 to 0.3 microns as per IEST-RP-CC001.3 USA
Typical
>99.995% at MPPS as per EN 1822 (H-14) EU
Efficiency Exhaust
MAIN FILTRATION AGENTS cycle is also enabled at shutdown, ensuring operating costs; extremely low noise and
all residue contaminants are purged out of vibration levels (less than 60dBA at working
ISO Class 3 air cleanliness within the workzone before deactivation. position) due to proprietary construction
workzone as per ISO 14644.1 (equivalent and mounting technology.
to Class 1 as per the US Federal Standard An admin PIN can be set to restrict
209E, 100 times cleaner than the usual unauthorized access to all menu functions. The cabinet’s airflow is self-
Class 100 classification on cabinets offered regulating: the motor / blower system
by the competition). ERGONOMIC FEATURES is able to compensate automatically to
maintain airflow as the filter is loaded with
Minipleat separatorless ULPA filter Ergonomic sloped front design allows particulates.
technology, operating at the typical for maximum visibility into the workzone.
efficiency of 99.9999% at MPPS, 0.3 and Built-in solid state variable speed
0.12 microns provides better operator, The frameless tempered glass controller(s) (infinitely adjustable from zero
product and environmental protection than sliding front sash and tempered glass to the maximum setting) with built-in RFI
conventional HEPA filters. This cabinet sidewalls provide a greater visibility into and noise filters is superior to conventional
employs 2 ULPA filters, 1 for downflow and the workzone. “step” controllers.
1 for exhaust airflow filtration.
Built-in 5000k fluorescent lighting Superior electrical safety for the
CABINET CONTROL SYSTEM offers excellent illumination throughout operator; all electrical components are
the workzone. Electronically-ballasted UL listed or UL recognized; all cabinets are
Esco Sentinel™ Microprocessor lighting system is instant-start, non- factory tested for electrical safety after
Cabinet Control System allows the user flickering and energy-efficient. production at manufacturing site.
to easily access all cabinet functions. The
control system monitors and displays Cabinet armrest is raised above the Designed to meet the general safety
cabinet airflow constantly on the control workzone to ensure that the operator’s requirements of the IEC 61010-1 / EN
LCD screen, prompting the user with arms do not block the front inflow 61010-1 / UL 61010A-1 / CSA C22.2 No.
audible and visual alarms in case of any perforations. (Optional PVC soft armrest- 1010.1-92.
unsafe condition. padding available upon request. Refer to
our Accessories catalogue.)* OPTIONAL ACCESSORIES
Fail-safe system ensures that in case of
airflow failure, the cabinet automatically CABINET BODY CONSTRUCTION Support Stand with
shuts down to ensure safety to the user castor wheels or
and the environment. All components designed for maximum levelling [Link]
chemical resistance for long service life construction allows for
True airflow velocity (both for downflow and durability. All cabinet components are the cabinet to be lifted
and inflow velocity) sensing technology, directly on to our optional
cleanroom compatible.
with temperature compensation for im- support stand (available
proved sensor accuracy. with castor wheels or
Industrial-grade main body constructed
levelling feet).
of electrogalvanised steel: with an abrasion-
Audible and visual alarms for low airflow, resistant white oven-baked powder-coated
unsafe sash positions. Exhaust collar for non-
finish. Attractive two-piece stainless steel
airtight thimble ducting
work trays are easy to remove for purposes
Automatic warm-up cycle is enabled of the cabinet.
of cleaning and auto-claving.
upon turning on the cabinet, ensuring all
contaminants are purged from the cabinet For other optional and
Permanently lubricated direct drive
workzone before the operator can use the ergonomic accessories,
centrifugal blower(s); energy efficient
cabinet. All menu functions are inactive check our catalogue:
external rotor motor type design reduces
during this period. Automatic post-purge [Link]
Esco Catalogue Revision No: AC2.2ft.v1.12.2005
Specifications subject to change without notice.
Technical Specifications and Engineering Diagram ©2005 Esco Micro Pte. Ltd.
173 mm / 6.8”
15
2
7
1435 mm / 56.5”
8
12 4
5
13
670 mm / 26.4”
9 6 Work zone heignt
14
10
11
660 mm / 26.0”
732 mm / 28.8”
730 mm / 28.7”
1. Exhaust ULPA Filter 2. Blower 3. Downflow ULPA Filter 4. UV Light Retrofit Kit™ Provision 5. IV Bar Retrofit Kit™ Provision
6. Service Fixture Retrofit Kit™ Provisions (2 One Each Side) 7. Electrical & Electronics Panel 8. Fluorescent Lamps
9. Tempered Glass Sidewalls 10. Stainless Steel Divided Work Tray (2pcs) 11. Stainless Steel Armrest
12. Esco Sentinel™ Microprocessor Control System 13. Tempered Glass Sliding Sash
14. Electrical Outlet Retrofit Kit™ Provision 15. Optional Exhaust Collar
21 Changi South Street 1 · Singapore 486 777 · Tel: +65 6542 0833 · Fax: +65 6542 6920 · Email: biotech@[Link] · [Link]
NSF 49, UL 61010, JIS K3800, SFDA YY-0569, EN 12469, SANS 12469
Ambient air is pulled through the Near the work surface, the downflow
perforations located towards the work air stream splits with a portion moving
zone front to prevent contamination of toward the front air grille, and the
the work surface and work product. The remainder moving to the rear air grille.
inflow does not mix with the clean air A small portion of the ULPA filtered
within the cabinet work zone. Inflow air downflow enters the intake perforations
travels through a return path toward the at the side capture zones at a higher
common air plenum (blower plenum) at velocity (small blue arrows).
the top of the cabinet.
A combination of inflow and downflow
Approximately 35% of the air in the air streams forms an air barrier that
common plenum is exhausted through prevents contaminated room air from
the ULPA filter to the room. The entering the work zone, and prevents
remaining 65% of the air is passed work surface emissions from escaping
through the downflow ULPA filter and the work zone.
ULPA-filtered air into the work area as a vertical laminar
Air returns to the common air plenum
flow air stream bathing the work surface
Unfiltered / potentially contaminated air where the 35% exhaust and 65%
in clean air.
Room air / Inflow air
recirculation process is continued.
The uniform, non-turbulent air stream
protects against cross contamination
within and throughout the work area.
• The counterbalance locking mechanism • The blower/motor can maintain airflow Sentinel Microprocessor Control,
is inherently safe; it locks the counter- for a 150% increase in pressure loss Alarm, Monitoring System
balance in place if either of the 2 cables over clean filters without manual The Esco Sentinel™ microprocessor-based
is detached. The sash cable and cable speed control adjustment, or 210% control system supervises operation of all
clip have rated strength of more than increase in pressure drop across the filter cabinet functions.
6 times the weight of the sash window with manual speed control adjustment.
• The control panel is located on the
of the largest Labculture Class II cabinet (Applies to 1.2 meter / 4 ft. model; contact
center of the cabinet, and angled down
(1.8 meter / 6 ft. model). Esco for information on other models.)
for easy access by the operator
• The safety glass maintains containment • The permanently lubricated direct-drive
• Continuous monitoring of cabinet
if the sash is accidentally broken during external rotor motor/blower reduces
airflow is displayed on a bright, easy-to-
cabinet operation. operating costs.
read LCD panel.
• Built-in RFI and electrical noise filters
Blower Efficiency • An integrated, temperature-compen-
eliminate interference with adjacent
The Labculture blower system is designed sated, true airflow velocity sensor
instrumentation.
for high performance operation, provides an accurate airflow reading
• An integral blower hour meter tracks
maximum energy efficiency and minimal despite room temperature fluctuation.
operating life for predictive maintenance
maintenance. • All electronic parts are contained inside
planning.
• The external rotor motor design allows a plug-and-play module that permits
• To prevent fan damage, a paper-catch
for optimum cooling of the motor easy exchange if required.
grille traps papers or towels that may
during extended operations and extends • Microprocessor software updates are
drop down on the drain pan, preventing
the motor bearing life. available from Esco for download via
them from being pulled into the column
the Internet.
by fan suction.
Sentinel functions are factory set to default
to ON or OFF, depending on worldwide
destination and local preferences. Default
Esco Centrifugal Fan with External Rotor Motor (left) vs.
settings can be user activated through the
Conventional Fan with Standard Motor (right)
touchpad data entry access.
Esco cabinets use German made ebm-papst®
permanently lubricated, centrifugalmotor/
• Automatic start-up sequence will
blowers with external rotor designs. prepare the cabinet for normal operation
Integrated blades narrow the profile and and advise when safe conditions are
eliminate need for a motor shaft.
established.
Motors are selected for energy efficiency,
compact design, and flat profile. The completely • An administrator controlled PIN
integrated assembly optimizes motor cooling.
(Personal Identification Number) can be
All rotating parts are unitized and balanced
for smooth, quiet, vibration-free operation. set to restrict access to main menu.
Sentinel Microprocessor Control System, Programmable When programmed ON • the start-up sequence confirms status with Air Safe and local time display.
• the Personal Identification Number (PIN) access restricts unauthorized adjustments.
• an airflow alarm warns of deviations from normal velocities.
• The airflow alarm can be activated • The interior work area is formed from a walls to eliminate concealed or hard-
or deactivated depending on user single piece of stainless-steel with large to-clean spaces.
preference and nature of the work. radius corners to simplify cleaning. • There are no screws in on the front or
Consult your Esco Operating Manual • The cabinet work zone has no welded sides to trap contaminants or complicate
or contact your Sales Representative joints to collect contaminants or rust. cleaning.
for information on user-preference • All stainless steel work surfaces are • Optional service fittings are offset for
programming capabilities built into the accessible for cleaning. easier access.
Sentinel microprocessor platform.
• Tray components lift and remove to • External plumbing is concealed behind
provide easy access and encourage trim panels to maintain aesthetics.
Cabinet Construction
surface decontamination. • External surfaces are coated with Esco
Robust construction and enhanced safety
• A recessed central area and stainless Isocide™ antimicrobial coating to protect
features qualify the cabinet for the most
steel drain pan channels spills and against surface contamination and
demanding laboratory applications. The
prevent liquids from entering the lower inhibit bacterial growth. Isocide
cabinet is fully assembled and ready to install
filtration and blower systems. eliminates 99.9% of surface bacteria
and operate when shipped.
• The drain pan is flush with the side within 24 hours of exposure.
Service fixtures are offset for easier reach. Standard cabinets include two fixture provisions on each sidewall
(one provision on each sidewall for 0.9 meter/3 ft. cabinet). Electrical outlets are mounted below service
fixtures to minimize obstructions.
Helpful for certifiers, the hinged maintenance assembly opens to a fixed position on integrated, gas spring
struts providing front service access.
All key components, with the exception of the blower/motor assembly, are mounted outside the air stream
and away from contaminated air to permit service without decontamination. These include fluorescent lamps,
UV lamp, electrical harnesses, electronic boards and microprocessor control.
The telescoping Dynamic Chamber™ plenum minimizes physical lifting and accelerates filter change when required.
Work area containment is maintained even when removable components are lifted out for cleaning.
The one piece stainless steel work tray edges are radiused and easy to clean without crevices or joints.
The lower drain trough is a single-piece fabrication with wide open angles and a channel to direct spills to
the drain.
The closed sidewall contains no perforations, air return slots or other hidden areas where contaminants
can accumulate.
Biosafety Cabinets Air Quality Filtration Electrical Safety
Standards NSF / ANSI 49, USA* ISO 14644.1, Class 3, Worldwide EN-1822 (H14), Europe UL-C-61010A-1, USA
Compliance EN 12469, Europe JIS B9920, Class 3, Japan IEST-RP-CC001.3, USA CSA22.2, No.1010-192, Canada
JIS K 3800, Japan JIS BS5295, Class 3, Japan IEST-RP-CC007, USA EN-61010-1, Europe
SFDA YY-0569, China US Fed Std 209E, Class 1 USA IEST-RP-CC034.1, USA IEC61010-1, Worldwide
* Please refer to the specifications table on page 8 for the model listing.
Dynamic Chamber™ Plenum Design • The front armrest is raised above the Warranty
workzone to improve comfort and to The Labculture cabinet is warranted for
ensure that the operator’s arms do not 3 years excluding consumable parts and
block the forward airflow perforations. accessories.
• The optional adjustable support stand
• Each cabinet is shipped with a compre-
provides work surface height control.
hensive user’s manual complete with a
• The frameless sash eliminates operator's report documenting all test procedures.
line of sight blockage
• Additional IQ/OQ/PQ documentation
• A generous sash opening allows for is available upon request.
easier access into the work zone,
• Contact your local Sales Representative
provides ample room for transferring of
for specific warranty details or docu-
small equipment.
mentation requests.
• The sliding window can be fully
opened to insert and remove larger Accessories and Options
Negative pressure Positive pressure
instrumentation and equipment. Esco offers a variety of options and
The Esco triple-wall design creates a Dynamic Chamber
plenum which surrounds contaminated areas with accessories to meet local applications.
negative pressure, preventing the possibility of Electrical Safety and Certification Contact Esco or your local Sales Represen-
contamination from leaks in filter seal, gasket or
cabinet structure. The third wall conceals utilities. All components meet or exceed applicable tative for ordering information.
safety requirements.
Comfortable Ergonomic Design Support Stands
The LA2 cabinet is engineered for comfort, • Each cabinet is individually tested for • Fixed height, available 711 mm (28")
utility value and safety. electrical safety at factory. or 864 mm (34"), ±38.1 mm (1.5")
• Documentation specific to each cabinet - With leveling feet
• The 10° angled viewing window and
serial number is maintained on file. - With casters
narrow profile front grille improves
reach into the work area. • UL Listed for USA and Canada. • Adjustable height, hydraulic range
• Certified to all major world standards 711 mm to 864 mm (28" to 34")
• The instant-start 5000k fluorescent
for biological safety cabinets, including - With leveling feet
lamp operates on an electronic ballast
the NSF / ANSI 49 and EN 12469 (refer - With casters
to reduce heat, improve comfort and
conserve energy. to Standards Compliance table on this • Telescoping height stand for leveling
page for the full list). feet, nominal range 660 mm or
• The lamp delivers uniform lighting to
• Contact Esco or your Sales Representative 960 mm (26" or 37.8")
the work surface for greater comfort,
reduced glare and improved productivity; for site preparation information; see • Telescoping height stand for casters,
see Technical Specifications. Electrical Specifications. nominal range 660 mm or 880 mm
(26" or 34.6")
- Adjustable in 25.4 mm (1") increments
Antimicrobial Powder-Coating
(Exterior height for 110-130 V, 60 Hz models of 1.2 m and 1.8 m / 4' and 6' is 1640 mm / 64.6")
16 1
76
27
8.
462 mm (1
4a9
13 4b9
87
670 mm (26.4")
9
14
109
ING
ACC S
15
OPENES
16 11
45 mm
(1.8")
955 mm / 1260 mm / 1565 mm / 1870 mm 180 mm 12
(37.6"/ 49.6"/ 61.6"/ 73.6") (7.1") 139 mm
(5.5")
1115 mm / 1420 mm / 1725 mm / 2030 mm 815 mm (32.1")
(43.9"/ 55.9"/ 67.9"/ 79.9")
*Cabinet Size 0.9 meter (3') 1.2 meter (4') 1.5 meter (5') 1.8 meter (6')
Testing Opening Height 228.6 mm (9") 228.6 mm (9") 228.6 mm (9") 203.2 mm (8")
Working Access Height 274 mm (10.8") 274 mm (10.8") 274 mm (10.8") 218 mm (8.6")
1. Exhaust ULPA filter 6. Electrical outlet Retrofit Kit provision 10. Stainless steel single-piece work tray 14. Safety glass sliding sash window
2. Blower (0.9 meter /3' model - one single 11. Stainless steel armrest 15. Single-piece stainless steel back wall
outlet in workzone) (1.2, 1.5 and and side walls
3. Downflow ULPA filter, angled 12. Drain valve Retrofit Kit provision
1.8 meter / 4',5' & 6' models - two
4a. UV light Retrofit Kit provision for 0.9, 13. Esco Sentinel microprocessor control 16. Removable side panel for plumbing
single outlets in workzone)
1.2 and 1.5 meter (3',4' & 5') models. system access
7. Electrical / Electronics panel
4b. UV light Retrofit Kit provision for 1.8
meter (6') models. 8. Fluorescent lamp
5. IV-Bar Retrofit Kit provision 9. Plugged service fixture provisions
(2 on each side)
333 mm 610 mm
(13.1") (24")
269 mm
(10.6")
269 mm
269 mm
(10.6")
(10.6")
249 mm
(ø9.8")
129 mm
5.1"
350 mm 691 mm
(13.8") (27.2")
269 mm
269 mm
(10.6")
(10.6")
Side View - All Sizes Top View - LA2-4A_ Top View - LA2-6A_
General Specifications
North America Models LA2-3A2 LA2-4A2 LA2-5A2 LA2-6A2
(USA, Canada, Mexico / 115 V, 60 Hz)
Nominal Size 0.9 meters ( 3') 1.2 meters ( 4') 1.5 meters ( 5') 1.8 meters ( 6')
External Dimensions 1115 x 810 x 1540 mm 1420 x 815 x 1640 mm 1725 x 815 x 1540 mm 2030 x 815 x 1640 mm
(W x D x H) 43.9" x 31.9" x 60.6" 55.9" x 32.1" x 64.6" 67.9" x 32.1" x 60.6" 79.9" x 32.1" x 64.6"
Gross Internal Dimensions 955 x 623 x 670 mm 1260 x 623 x 715 mm 1565 x 623 x 670 mm 1870 x 623 x 715 mm
(W x D x H) 37.6" x 24.5" x 26.4" 49.6" x 24.5" x 28.1" 61.6" x 24.5" x 26.4" 73.6" x 24.5" x 28.1"
Usable Work Area 0.45 m2 (4.8 [Link].) 0.6 m2 (6.5 [Link].) 0.75 m2 (8.1 [Link].) 0.9 m2 (9.7 [Link].)
Tested Opening 228.6 mm (9") 228.6 mm (9") 228.6 mm (9") 203.2 mm (8")
Working Opening 274 mm (10.8") 274 mm (10.8") 274 mm (10.8") 248 mm (9.8")
Downflow, 63% 702 m3 / h (413 cfm) 924 m3 / h (544 cfm) 1147 m3 / h (675 cfm) 1289 m3 / h (758 cfm)
ULPA Filter Typical Efficiency >99.999% for particle size between 0.1 to 0.3 microns per IEST-RP-CC001.3
NSF / ANSI 49 <66 dBA <64.5 dBA <66 dBA <67 dBA
Sound Emission
EN 12469 <63 dBA <61.5 dBA <63 dBA <64 dBA
> 1230 Lux > 1400 Lux > 1070 Lux > 1225 Lux
Fluorescent Lamp Intensity
(> 114 foot-candles) (> 130 foot-candles) (> 100 foot-candles) (> 114 foot-candles)
Net Weight ** 243 kg / 536 lbs 336 kg / 741 lbs 317 kg / 698 lbs 434 kg / 957 lbs
Shipping Weight ** 292 kg / 644 lbs 400 kg / 880 lbs 410 kg / 904 lbs 523 kg / 1153 lbs
Shipping Dimensions, 1230 x 940 x 1900 mm 1530 x 940 x 1900 mm 1910 x 940 x 1900 mm 2150 x 940 x 1980 mm
Maximum (W x D x H) ** 48.4" x 37" x 74.8" 60.2" x 37" x 74.8" 75.2" x 37" x 74.8" 84.6" x 37" x 77.9"
Shipping Volume, Maximum ** 2.20 m3 (78 [Link].) 2.73 m3 (97 [Link].) 3.41 m3 (121 [Link].) 4.00 m3 (141 [Link].)
Nominal Size 0.9 meters ( 3') 1.2 meters ( 4') 1.5 meters ( 5') 1.8 meters ( 6')
External Dimensions 1115 x 810 x 1540 mm 1420 x 815 x 1540 mm 1725 x 815 x 1540 mm 2030 x 815 x 1540 mm
(W x D x H) 43.9" x 31.9" x 60.6" 55.9" x 32.1" x 60.6" 67.9" x 32.1" x 60.6" 79.9" x 32.1" x 60.6"
Gross Internal Dimensions 955 x 623 x 670 mm 1260 x 623 x 670 mm 1565 x 623 x 670 mm 1870 x 623 x 670 mm
(W x D x H) 37.6" x 24.5" x 26.4" 49.6" x 24.5" x 26.4" 61.6" x 24.5" x 26.4" 73.6" x 24.5" x 26.4"
Usable Work Area 0.45 m2 (4.8 [Link].) 0.6 m2 (6.5 [Link].) 0.75 m2 (8.1 [Link].) 0.9 m2 (9.7 [Link].)
Tested Opening 228.6 mm (9") 228.6 mm (9") 228.6 mm (9") 203.2 mm (8")
Working Opening 274 mm (10.8") 274 mm (10.8") 274 mm (10.8") 248 mm (9.8")
Downflow, 63% 702 m3 / h (413 cfm) 924 m3 / h (544 cfm) 1147 m3 / h (675 cfm) 1289 m3 / h (758 cfm)
ULPA Filter Typical Efficiency >99.999% at particle size between 0.1 to 0.3 microns
NSF / ANSI 49 <66 dBA <63.5 dBA <66 dBA <67 dBA
Sound Emission
EN 12469 <63 dBA <60.5 dBA <63 dBA <64 dBA
> 1230 Lux > 1400 Lux > 1070 Lux > 1225 Lux
Fluorescent Lamp Intensity
(> 114 foot-candles) (> 130 foot-candles) (> 100 foot-candles) (> 114 foot-candles)
220-240V, AC,
LA2-3A1 LA2-4A1 LA2-5A1 LA2-6A1
50Hz, 1ø
220-240V, AC,
Electrical * LA2-3A3 LA2-4A3 LA2-5A3 LA2-6A3
60Hz, 1ø
100V, AC,
LA2-3J5 LA2-4J5 LA2-5J5 LA2-6J5
50/ 60Hz, 1ø
Net Weight ** 243 kg / 536 lbs 283 kg / 624 lbs 317 kg / 698 lbs 350 kg / 772 lbs
Shipping Weight ** 292 kg / 644 lbs 345 kg / 761 lbs 410 kg / 904 lbs 486 kg / 1072 lbs
Shipping Dimensions, 1230 x 940 x 1900 mm 1530 x 940 x 1900 mm 1910 x 940 x 1900 mm 2150 x 940 x 1980 mm
Maximum (W x D x H) ** 48.4" x 37" x 74.8" 60.2" x 37" x 74.8" 75.2" x 37" x 74.8" 84.6" x 37" x 77.9"
Shipping Volume, Maximum ** 2.20 m3 (78 [Link].) 2.73 m3 (97 [Link].) 3.41 m3 (121 [Link].) 4.00 m3 (141 [Link].)
The Performance Envelope Concept • HPV requires approximately 4-7 hours for
set-up, decontamination, and tear-down,
0.65 Esco cabinets are designed to operate compared to a total of 12-15 hours
m/s
within a performance envelope to maintain needed to complete a formaldehyde
protection for personnel, product and the
High Inflow,
High Downflow environment. decontamination process.
0.60 (0.61, 0.43)
High Inflow,
Low Downflow
Airflow parameters used to frame the • HPV decontamination effectiveness
(0.58, 0.27) performance envelope include both inflow
velocity and downflow velocity.
is independent of temperature and
0.55
The graph illustrates the boundaries of humidity. Formaldehyde requires
Nominal
(0.53, 0.35) the performance envelope (based on a temperature above 20°C and relative
Inflow Velocity
Key Features
2
7
1436 mm / 56.5”
4
8
12
5
670 mm / 26.4”
9
13 10
/ 8.0”
14
203m m
6
15 11
Engineering Details
1. Exhaust ULPA Filter 2. Blower 3. Downflow ULPA filter 4. Standard UV Light Retrofit™ Kit Provision 5. IV Bar Retrofit Kit™ Provision
6. Electrical Outlet Retrofit Kit™ Provision (2 single outlets) 7. Electrical / Electronics Panel 8. Fluorescent Light
9. Plugged Service Fixture Provisions (2 on each side wall) for Gas/Vacuum/Water. (Solenoid Gas Valve Option Available)
10. Stainless Steel Multi-piece Work tray 11. Stainless Steel Armrest
12. Esco Sentinel™ Microprocessor Control System 13. Laminated Shatter-Proof Sliding Sash Window
14. Single-piece Stainless Steel Back Wall and Side Wall 15. Removable Side Panel for Plumbing Access
Main Features
The air enters the cabinet through the front air Extremely low noise level for quiet working
grille, and travels through the KI-impregnated environment
carbon filter on which radioactive iodine is Additional options:
adsorbed. Air then travels to the blower plenum Lead-shielded back panel
where it is filtered by high-efficiency ULPA filters Increased thickness of lead shielding
before being exhausted. (standard is 3mm)
Technical Specifications and Engineering Diagram Esco Catalogue Revision LS2.v2.7.2006
Specifications subject to change without notice
©2006 Esco Micro Pte Ltd
1
7 2
/18.7”
ING
MAX. OPEN
4
475 mm
13 8
2190 mm / 86.2”
9 5
670 mm /26.43”
14
10
/9.8”
6
OPENING
ACCESS
15
248 mm
17
16 11
760 mm / 30.0”
12
750 mm / 29.5”
1420/2030 mm
60”/ 80” 780 mm / 30.7”
1. Exhaust ULPA Filter 2. Blower 3. Downflow ULPA Filter 4. Standard UV Light Retrofit™ Kit Provision
5. Standard IV-Bar Retrofit™ Kit Provision 6. Electrical Outlet Retrofit™ Kit Provision (Two Single Outlets in Workzone)
7. Electrical and Electronics Panel 8. Fluorescent Light 9. Plugged Service Fixture Provisions (2 on each side)
10. Stainless Steel Multi-piece Work Tray 11. Stainless Steel Armrest 12. KI-Impregnated Carbon Filter
13. Esco Sentinel™ Microprocessor Control System 14. Lead-Shielded Glass Sliding Sash Window
15. Stainless Steel Back Wall 16. Lead-Shielded Side Walls 17. Horizontal Lead-Glass Sliding Door
..
Tested to EN 12469 and TUV-GS certified
for safety and performance.
Advanced Engineering Containment and Protection • Auto-purge holes located at the front
The Esco Infinity microbiological safety The Esco Infinity Class II Microbiological side walls eliminate eddy currents and
cabinet, includes a number of design and safety cabinet (FC2) provides operator, dead air pockets in the critical area
performance features not found on our product and environmental protection behind the sash window.
popular Labculture® series cabinets. These against Biosafety Levels 1, 2 and 3. This • The downflow (supply) filter is tilted
include: cabinet can be used for handling Biosafety proportional to the cabinet front angle
Level 4, provided that the operator wears to direct more air forward to the front
• An aerosol tight window for additional
positive pressure suit. air grille.
safety while the cabinet is inoperative.
• The airflow ratio of 65% recirculation to • The inflow velocity, downflow velocity,
• Double fan design guarantees safety in
35% exhaust increases operator pro- and airflow path, and intake geometry
the event of the failure of one fan.
tection beyond the 70% / 30% ratio are precision tuned and tested to create
• Motorized front sash for one-hand an optimum air curtain on the front
of conventional microbiological safety
operation. aperture; this curtain maintains per-
cabinets.
• Larger LCD display for easy monitoring sonal and product protection even in
• Inflow of room air enters the front air
of operational parameters. the unlikely event of a severe inflow
grille to establish operator protection;
• Reduced height permits cabinet instal- or downflow imbalance that would
room air does not enter the work zone,
lation for seated operation in standard compromise protection in a conven-
preventing product contamination.
2.4 meter (nominal, <8') laboratory ceil- tional cabinet.
• Raised armrest prevents the likelihood
ing height. Cabinet shown on optional
of inflow grille blocking by operator's Integrated Filtration System
mounting stand with casters, seated
arms. Independent supply and exhaust filters
height work surface 711 mm (28").
provide 99.999% typical efficiency for
particle sizes of 0.1 to 0.3 microns. Infinity
Series filters meet the IEST-RP-CC001.3
recommended practice for ULPA per-
formance (USA), and EN 1822 for H14
Esco Filter Efficiency
performance (EU).
Independent supply and exhaust
filters provide 99.999% typical efficiency • ULPA filters (per IEST-RP-CC001.3),
for particle sizes of 0.1 to 0.3 microns.
Infinity Series filters meet the IEST-RP- are tested to a typical efficiency of
CC001.3 recommended practice for >99.999% for 0.1 to 0.3 micron par-
ULPA performance (USA), and EN 1822
for HEPA performance (EU). ticles; these provide better filtration
capability than conventional H13 HEPA
Typical Penetration filters that have a typical efficiency of
> 99.99% for 0.3 micron particles.
Enlarged touchpad data entry buttons with tactile feedback permit control settings and access
to diagnostics, default settings and hierarchical menus.
Mini-pleat Separatorless Filter (left) vs. Sentinel Microprocessor Control, • A pressure sensor monitors pressure
Conventional Aluminium Separator Filter (right)
Alarm, Monitoring System drop across the filter plenum and com-
The Esco Sentinel microprocessor-based putes estimated remaining filter life.
control system supervises operation of all • A one-touch half-speed setting can be
cabinet functions. activated at the end of the work day to
reduce energy consumption, prolong
• Continuous monitoring of cabinet air-
filter life and maintain interior work
flow is displayed on a bright, easy-to-read
zone cleanliness for the next day's
LCD panel. The large display monitors
work.
operational parameters.
Esco cabinets use Swedish Camfil Farr® mini-pleat • All electronic parts are contained inside
filters without aluminum separators to increase filter • The control panel is located on the center
Input Voltage
efficiency, minimize the chance of leakage, and to a plug-and-play module that permits
of the cabinet, and angled down for easy
prolong filter life. Filters include a lightweight alumi- easy exchange if required.
num frame for structural stability and elimination of access by the operator. Blower S
swelling common to conventional wood frames. • Microprocessor software updates are
• A back-up battery maintains alarm
available from Esco for download via
function for airflow, power and pressure
• Filter assembly is constructed in accor- the Internet.
alarm in the event of a power failure.
dance with EN1822 requirements. • Sentinel functions are factory set to
• Two integrated, temperature-com-
• The supply filter provides ISO Class 3 default to ON or OFF, depending on
pensated true airflow velocity sensors
(per ISO14644.1) clean air to the work worldwide destination and local pref-
provide independent measurement of
surface in a gentle vertical laminar flow erences. Default settings can be user
inflow and downflow velocities despite
for product protection. activated through the touchpad data
room temperature fluctuation.
• Modern separatorless mini-pleat filter entry access.
construction maximizes the filter surface
area to extend filter life and eliminate
Esco Accuflow Microprocessor Speed Controller
possible filter media damage by thin
and sharp aluminum separators used in
conventional HEPA filter construction.
Input Voltage
Ambient air is pulled through the perfora- Near the work surface, the downflow
tions located towards the work zone front air stream splits with a portion moving
to prevent contamination of the work sur- toward the front air grille, and the remain-
face and work product. The inflow does der moving to the rear air grille. A small
not mix with the clean air within the cabi- portion of the ULPA filtered downflow
net work zone. Inflow air travels through enters the intake perforations at the side
a return path toward the common air capture zones at a higher velocity (small
plenum (blower plenum) at the top of the blue arrows).
cabinet.
A combination of inflow and downflow
Approximately 35% of the air in the air streams forms an air barrier that pre-
common plenum is exhausted through vents contaminated room air from enter-
the ULPA filter to the room. The remain- ing the work zone, and prevents work
ing 65% of the air is passed through the surface emissions from escaping the work
downflow ULPA filter and into the work zone.
ULPA-filtered air area as a vertical laminar flow air stream
Air returns to the common air plenum
Unfiltered / potentially contaminated air bathing the work surface in clean air.
where the 35% exhaust and 65% recircu-
Room air / Inflow air The uniform, non-turbulent air stream lation process is continued.
protects against cross contamination
within and throughout the work area.
• Automatic start-up sequence will pre- • Dual permanently lubricated direct-drive operating life and aids in predictive main-
pare the cabinet for normal operation external rotor motor/blowers safety in tenance planning.
and advise when safe conditions are the event of a motor failure. • To prevent fan damage, a paper-catch
established. • The external rotor motor design allows grille traps papers or towels that may
• An administrator controlled PIN (Per- for optimum cooling of the motor drop down on the drain pan, preventing
sonal Identification Number) can be set during extended operations and ex- them from being pulled into the column
to restrict access to main menu. tends the motor bearing life. by fan suction.
• The airflow alarm can be activated or • The inflow and downflow balance is
Cabinet Construction
deactivated depending on user prefer- precisely established by two indepen-
ence and nature of the work. dent Accuflow fan speed controls. Robust construction and enhanced safety
features qualify the cabinet for the most
Consult your Esco Operating Manual or • The Accuflow microprocessor based
demanding laboratory applications. The
contact Esco or your Sales Representative speed controller maintains constant,
cabinet is fully assembled and ready to
for information on user-preference pro- stable airflow despite building supply
install and operate when shipped.
gramming capabilities built into the Sentinel voltage fluctuations.
microprocessor platform. • Speed can be adjusted electronically • The interior sides and back wall are
without mechanical adjustment. formed from a single piece of stainless-
Redundant Blower System steel with large radius corners to sim-
• Built-in RFI and electrical noise filters
The Infinity blower system is designed for plify interior cleaning.
eliminate interference with adjacent
high performance operation, redundancy, instrumentation. • The cabinet work zone has no welded
maximum energy efficiency and minimal joints to collect contaminants or rust.
maintenance. • All stainless steel work surfaces are
• An integral blower hour meter tracks
accessible for cleaning.
Esco Centrifugal Fan with External Rotor Motor (left) vs. • Multi-piece tray components are easily
Conventional Fan with Standard Motor (right) lifted and removed to encourage sur-
face decontamination.
Esco Infinity cabinets use a combination of high
performance scroll blowers (supply) and German • A recessed central area and stainless
made ebm-papst® permanently lubricated, centrifugal steel drain pan channels spills and
motor/blowers with external rotor designs (exhaust).
Selected for energy efficiency, compact design, prevent liquids from entering the lower
and flat profile, the completely integrated exhaust filtration and blower systems.
blower assembly optimizes motor cooling, with
unified rotating parts and overall component balance
for smooth, quiet, vibration-free operation. Weight
is equally distributed to all bearings to extend bearing
life, transfer heat and maximize speed control.
AIRAIR
SAFE
AIR
SAFESAFE SUPPLY FAIL
SUPPLY
SUPPLY FAILFAIL EXHAUST FAIL
EXHAUST
EXHAUST FAILFAIL
Downflow
DownflowFail
DownflowFailFail Inflow Fail
Inflow
Inflow FailFail
1A: Under normal operation with both blowers 1B: If the supply blower fails (1b), downflow to 1C: If the exhaust blower fails (1c), the supply
operating (1a) the supply blower creates a negative the work area is suspended. The exhaust blower blower continues to provide inflow to the cabinet
pressure surrounding the contaminated positive ramps up to maximum power to provide inflow and downflow to the work area. The control
pressure plenum and pushes air across the supply to the cabinet to help maintain containment. The panel warns of inflow failure. With the exhaust
and exhaust filters. The exhaust blower boosts the control panel warns of downflow failure. With blower offline, the cabinet maintains protection
air pressure through the exhaust filter to create the supply blower offline, the cabinet still provides by maintaining inflow above 0.40 m/s as required
better inflow and operator protection. Supply protection by maintaining inflow above 0.40 m/s by EN 12469.
and exhaust blowers automatically operating at as required by EN 12469.
reduced speeds extend blower life.
Dynamic Chamber™ Plenum Design • The External plumbing is concealed • The sliding window can be fully opened
behind trim panels to preserve cabinet to insert and remove larger instrumen-
aesthetics. tation and equipment.
• A normally closed gas solenoid valve
Electrical Safety and Certification
automatically shuts off gas flow in
the event of a cabinet alarm or unsafe All components meet or exceed applicable
condition. safety requirements.
• Each cabinet is individually factory
Comfortable Ergonomic Design tested for electrical safety.
The cabinet is engineered for comfort,
• Documentation specific to each cabinet
utility value and safety.
serial number is maintained on file.
• The angled viewing window and narrow • Certified to the renowned world stan-
profile front grille improves reach into dard for microbiological safety cabinets;
the work area. the TUV GS mark for compliance, to
Negative pressure Positive pressure
The Esco triple-wall design creates a Dynamic Cham- • The instant-start 5000k fluorescent EN 12469 (May, 2007).
ber plenum which surrounds contaminated areas lamp operates on an electronic ballast
with negative pressure, preventing the possibility • Contact Esco or your Sales Representative
of contamination from leaks in filter seal, gasket or to reduce heat, improve comfort and for site preparation information; see
cabinet structure. The third wall conceals utilities. conserve energy. Electrical Specifications.
• External surfaces are coated with Esco • The lamp delivers uniform lighting to
the work surface for greater comfort, Warranty
Isocide antimicrobial coating to protect
reduced glare and improved productivity; The Infinity is warranted for 3 years exclud-
against surface contamination and in-
see Specifications. ing consumable parts and accessories.
hibit bacterial growth. Isocide eliminates
99.9% of surface bacteria within 24 • The front armrest is raised above the • Each cabinet is shipped with a compre-
hours of exposure. work zone to improve comfort and to hensive User's Manual complete with a
• There are no screws in the front or sides minimize blockage of forward airflow report documenting all test procedures.
to trap contaminants or complicate perforations.
• Additional IQ/OQ/PQ documentation is
cleaning. • The optional adjustable support stand available upon request.
provides work surface height control.
• Contact your local Sales Representative
Service Fitting Access • The frameless sash eliminates operator's for specific warranty details or docu-
The cabinet is pre-plumbed for easy instal- line of sight blockage. mentation requests.
lation optional gas and vacuum fittings;
• A generous sash opening allows for
see Accessories.
easier access into the work zone, provides
• The Service fitting openings are offset for ample room for transferring of small equip-
easier access. ment; see Specifications.
Microbiological
Air Quality Filtration Electrical Safety
Safety Cabinets*
ISO 14644.1 Class 3, Worldwide
Standards AS 1386 Class 1.5, Australia IEC 61010-1, Worldwide
IEST-RP-CC034.1, USA EN 61010-1, Europe
Compliance Certified, TUV-GS JIS B9920 Class 3, Japan
IEST-RP-CC007.1, USA UL 61010A-1, USA
to EN 12469:2000 BS 5295 Class C, UK
IEST-RP-CC001.3, USA CSA C22.2 No. 1010.1-92, Canada
(May, 2007) KS27030.1, Class M10,000, Korea
EN 1822, Europe
IEST-G-CC1001, USA
IEST-G-CC1002, USA
*Models FC2-4A1 and FC2-6A1. Certification for additional models is pending or in process.
For details on most recent certifications contact Esco or your Sales Representative.
Helpful for certifiers, the hinged maintenance assembly opens to a fixed position on integrated,
gas spring struts providing front service access.
All key components with the exception of the motor/blower assembly are mounted outside the air stream and
away from contaminated air to permit service without decontamination. These include fluorescent lamps, UV
lamp, electrical harnesses, electronic boards and microprocessor control.
Panels enclosing potentially hazardous areas or components such as microbiological contamination
or electrical shock are color-coded red to warn service technician.
The telescoping Dynamic Chamber™ plenum minimizes physical lifting and accelerates filter change when
required.
Work area containment is maintained even when removable components are lifted out for cleaning.
The lower drain trough is a single-piece fabrication with wide open angles and a channel
to direct spills to the drain.
1
76
1 6
87 27
1540 mm (60.6")
49
15
109
14 5
670 mm (26.4")
11
16
12
ING
ACC S
OPENES
17
18 13
1015 mm
(40.0")
Top View - FC2-4A_
262 mm
(10.3")
Nominal Size 0.9 meters ( 3') 1.2 meters ( 4') 1.5 meters ( 5') 1.8 meters ( 6')
External Dimensions 1115 x 805 x 1535 mm 1420 x 805 x 1535 mm 1725 x 805 x 1535 mm 2030 x 805 x 1535 mm
(W x D x H) 43.9" x 31.7" x 60.4" 56.0" x 31.7" x 60.4" 67.9" x 31.7" x 60.4" 80.0" x 31.7" x 60.4"
Internal Work Area Dimensions 955 x 610 x 670 mm 1260 x 610 x 670 mm 1565 x 610 x 670 mm 1870 x 610 x 670 mm
(W x D x H) 37.6" x 24.0" x 26.4" 49.6" x 24.0" x 26.4" 61.6" x 24.0" x 26.4" 73.6" x 24.0" x 26.4"
Internal Work Space Area 0.47 m2 (5.06 [Link].) 0.62 m2 (6.67 [Link].) 0.77 m2 (8.29 [Link].) 0.92 m2 (9.90 [Link].)
Tested Opening 200 mm (7.9") 200 mm (7.9") 200 mm (7.9") 200 mm (7.9")
Working Opening 210 mm (8.3") 210 mm (8.3") 210 mm (8.3") 210 mm (8.3")
Power Rating
Code1
1, 3 332 m3 / h (195 cfm) 437 m3 / h (257 cfm) 593 m3 / h (319 cfm) 646 m3 / h (380 cfm)
Inflow
2 336 m3 / h (216 cfm) 483 m3 / h (284 cfm) 599 m3 / h (353 cfm) 714 m3 / h (420 cfm)
Airflow Volume Downflow 1, 2, 3 702 m3 / h (413 cfm) 929 m3 / h (544 cfm) 1147 m3 / h (675 cfm) 1367 m3 / h (804 cfm)
1, 3 332 m3 / h (195 cfm) 437 m3 / h (257 cfm) 543 m3 / h (319 cfm) 646 m3 / h (380 cfm)
Exhaust
2 336 m3 / h (216 cfm) 483 m3 / h (284 cfm) 599 m3 / h (353 cfm) 714 m3 / h (420 cfm)
ULPA Filter Downflow >99.999% at 0.1 to 0.3 microns as per IEST-RP-CC001.3 USA
Typical Efficiency Exhaust >99.995% at MPPS as per EN 1822 (H-14) EU
Net Weight ** 228 kg / 503 lbs 298 kg / 657 lbs 372 kg / 820 lbs 447 kg / 985 lbs
Shipping Weight ** 280 kg / 617 lbs 354 kg / 780 lbs 443 kg / 977 lbs 522 kg / 1150 lbs
Shipping Dimensions, 1200 x 900 x 1870 mm 1530 x 900 x 1870 mm 1910 x 900 x 1870 mm 2150 x 900 x 1870 mm
Maximum (W x D x H) ** 47.2" x 35.4" x 73.6" 60.2" x 35.4" x 73.6" 75.2" x 35.4" x 73.6" 84.6" x 35.4" x 73.6"
Shipping Volume, Maximum ** 2.02 m3 (71 [Link].) 2.58 m3 (91 [Link].) 3.22 m3 (114 [Link].) 3.62 m3 (128 [Link].)
• If there is a gap on the cabinet sealing, for operator protection, and is useful for
spores/mL [Link] spores is placed
escaping HPV to the lab will decompose to validating the actual containment perfor-
inside the work zone, 10 cm (4") behind the
become oxygen and water. Escaping form- mance of the cabinet on-site.
front opening sash.
aldehyde, however, is harmful to people in • The KI-Discus takes only 45 minutes as
• Target slit air samplers and impingers are
the lab. Therefore HPV decontamination opposed to 2 days for microbiological
placed outside the work zone to capture 11
can be performed while people are work- testing.
possibly escaping [Link] spores, then
ing inside the lab, while formaldehyde • Thus, each Esco Infinity FC2 model is fac-
the sample is incubated.
decontamination must be performed with tory tested using the KI-Discus method for
• Acceptance: The number of Bacillus atro- no one present in the lab. The HPV method operator safety.
phaeus CFU recovered from the agar plates improves safety, productivity, and reduces
shall not exceed 10 CFU per test. the time to seal the cabinet.
• HPV biological efficacy is independent of
Product Protection Test
environmental variables, whereas form-
The test objective is to determine cabinet pro- aldehyde efficacy is dependent on such
tection to the product/samples inside the cabinet variables.
work zone from environmental contaminants.
• HPV has a better penetration capacity,
• A nebulizer containing 55 mL of 5 to 8 x 106 resulting in a full decontamination of the
spores/mL [Link] is placed at 10 cm cabinet. The formaldehyde method is known
(4") in front of sash window. to result in incomplete decontamination.
• Target agar plates are placed throughout • HPV is more effective and rapid against bio-
the entire work surface. logical organisms compared to formaldehyde.
• Acceptance: The number of Bacillus atro- Esco builds quality from the inside out.
phaeus CFU recovered from the agar plates
shall not exceed 5 CFU per test.
Biological Safety Cabinet • Class II, Type B2 (Total Exhaust) Biological Safety Cabinets
spills in the work zone from entering • Modern separator-less mini-pleat filter
the side air column. construction maximizes the filter surface
• The ULPA downflow (supply) filter is area to extend filter life and eliminate
tilted proportional to the cabinet front possible filter media damage by thin
angle to direct more air forward to the and sharp aluminum separators used in
front air grille. conventional HEPA filter construction.
• The inflow velocity, downflow velocity, • The filter frame and media is constructed
and air flow path, and intake geometry in accordance with EN1822 requirements
are precision tuned and tested to for fire retardant properties
create an optimum air curtain on the • The supply ULPA filter provides ISO Class
front aperture; this curtain maintains 3 (per ISO14644.1) clean air to the work
personal and product protection surface in a gentle vertical laminar flow
even in the unlikely event of a severe for product protection.
inflow or downflow imbalance that • The exhaust HEPA filter traps biohazard
would compromise protection in a particles acquired from the work surface
conventional cabinet. before air is exhausted via the ducting
system to the external environment. The
Integrated Filtration System
Model LB2 is provided with an integral
A combination of a supply ULPA filter and exhaust collar for connecting the cabinet
an exhaust HEPA filter give the Labculture exhaust to an airtight ducting system.
Operator, Product and Environmental Protection
cabinet a fully integrated performance NB: Ductworks are not provided with
The Esco Labculture Class II, Type B2 Biosafety Cabinet
provides operator, product and environmental protection
envelope for product, operator and standard product.
against Biosafety Levels 1, 2 and 3. This cabinet can be environmental protection.
used for handling Biohazard Level 4, provided that the
operator wears positive pressure suit. • ULPA supply filter (per IEST-RP-
Front Sash Assembly
CC001.3), is tested to a typical efficiency • Integrated sash proximity contacts sense
Containment and Protection of >99.999% for 0.1 to 0.3 micron sash position, serve as an interlock for
• The total exhaust design of the LB2 makes particles; providing better filtration the UV lamp, and activate an alarm if
the cabinet suitable for microbiological capability than conventional HEPA the sash is improperly positioned.
work involving trace amounts of toxic filters that have a typical efficiency of • The magnetic switch eliminates the
chemicals. > 99.99% for 0.3 micron particles. chance of mechanical wear and tear
• The non-recirculating design guarantees • HEPA exhaust filter (per IEST-RP- typical of a mechanical switch.
safety in cases where exhaust HEPA- CC001.3), is tested to a typical efficiency • The back of the sash can be easily
filtration fails to ensure containment. of >99.99% for 0.3 micron particles. cleaned by removing the sash track
• Inflow of room air enters the front air cover and swing up the sash glass.
Mini-pleat Separatorless Filter (left) vs.
grille to establish operator protection; Conventional Aluminium Separator Filter (right) • The sash is counterbalanced for smooth
room air does not enter the work zone, and light sash moving operation.
preventing product contamination. • The counterbalance locking mechanism
• The front grille has proportionally larger is inherently safe; it locks the counter-
perforation on the extreme left and balance in place if either of the 2 cables
right side. Combined with the side air is detached. The sash cable and cable
gap between the work tray and the side clip have rated strength of more than
wall, the stronger side capture zones 6x the weight of the sash window of
increase protection in this critical area Esco cabinets use Swedish Camfil Farr® mini-pleat
the largest Labculture Class II cabinet
where contaminants tend to bleed out filters without aluminum separators to increase filter (1.8 meter/6 ft model)
efficiency, minimize the chance of leakage, and to
in conventional cabinet. prolong filter life. Filters include a lightweight • The safety glass maintains containment
• The inflow grille has a convex shape aluminum frame for structural stability and if the sash is accidentally broken during
elimination of swelling common to conventional
to maintain operator protection by wood frames. cabinet operation.
discouraging users from placing objects
on it.
• Raised armrest prevents the likelihood of Esco ULPA Filter Efficiency
inflow grille blocking by operator’s arms.
• Auto-purge holes located at the front Esco cabinets use ULPA filters (per
side walls eliminate eddy currents and IEST-RP-CC001.3) instead of conventional
HEPA filters commonly found in
dead air pockets in the critical area biological safety cabinets. While HEPA
filters offer 99.99% typical efficiency
behind the sash window. Per NSF / ANSI Typical Penetration
at 0.3 micron level, ULPA filters provide
49 requirement, these side perforations 99.999% typical efficiency for particle
sizes of 0.1 to 0.3 micron level.
do not extend completely to the work
surface to prevent accidental liquid
Cabinet Filtration System
Side capture zones
Dynamic air barrier, inflow and
forward-directed downflow air converge
Ambient air is pulled through the front The uniform, non-turbulent air stream
grille to prevent contamination of the protects against cross contamination
work surface and work product. The within and throughout the work area.
inflow does not mix with the clean air
Near the work surface, the downflow
within the cabinet work zone. Inflow air
air stream splits with a portion moving
travels through a return path toward the
toward the front air grille, and the
common air plenum (blower plenum) at
remainder moving to the rear air grille.
the top of the cabinet.
A small portion of the ULPA filtered
Ambient air is taken in through a pre- downflow enters the intake perforations
filter at the top of the cabinet, and at the side capture zones at a higher
passes through the downflow ULPA filter, velocity (small blue arrows).
entering the work zone as laminar flow.
A combination of inflow and downflow air
The uniform, non-turbulent air stream
streams forms an air barrier that prevents
protects against cross contamination
contaminated room air from entering the
within and throughout the work area.
work zone, and prevents work surface
Near the work surface, the downflow air emissions from escaping the work zone.
stream splits with a portion moving toward The downflow combined with the inflow
ULPA-filtered air the front air grille, and the remainder air enters the common air plenum.
Unfiltered / potentially contaminated air moving to the rear air grille. A small
All air in the common plenum is HEPA-filtered
Room air / Inflow air portion of the ULPA filtered downflow
and exhausted via a dedicated ducting
enters the intake perforations at the
system to the external environment.
side capture zones (small blue arrows).
Blower Efficiency • Built-in RFI and electrical noise filters • The control panel is located on the
The Labculture blower system is designed eliminate interference with adjacent center of the cabinet, and angled down
for high performance operation, instrumentation. for easy access by the operator.
maximum energy efficiency and minimal • An integral blower hour meter tracks • Continuous monitoring of cabinet
maintenance. operating life and aids in predictive airflow is displayed on a bright, easy-
maintenance planning. to-read LCD panel.
• The blower/motor can maintain airflow
velocity within 10% of original setpoint • To prevent fan damage, a paper-catch • An integrated, temperature-compen-
for a 50% increase in pressure loss over grille traps papers or towels that may sated true airflow velocity sensor
clean filters without manual speed drop down on the drain pan, preventing provides an accurate airflow reading
control adjustment. them from being pulled into the column despite room temperature fluctuation.
by fan suction. • All electronic parts are contained inside
• The permanently lubricated direct-drive
external rotor motor/blower reduces a plug-and-play module that permits
Sentinel Microprocessor Control,
operating costs. easy exchange if required.
Alarm, Monitoring System
• The external rotor motor design allows • Microprocessor software updates are
The Esco Sentinel microprocessor-based
for optimum cooling of the motor available from Esco for download via
control system supervises operation of all
during extended operations and extends the Internet.
cabinet functions.
the motor bearing life. Sentinel functions are factory set to default
to ON or OFF, depending on worldwide
destination and local preferences. Default
settings can be user activated through the
touchpad data entry access.
Esco Centrifugal Fan with External Rotor Motor (left) vs.
Conventional Fan with Standard Motor (right) • Automatic start-up sequence will
prepare the cabinet for normal operation
and advise when safe conditions are
Esco cabinets use German made ebm-papst® established.
permanently lubricated, centrifugal motor/
blowers with external rotor designs. • An administrator controlled PIN
Integrated blades norrow the profile and (Personal Identification Number) can
eliminate the need for a motor shaft.
be set to restrict access to main menu.
Motors are selected for energy efficiency,
compact design, and flat profile. The completely • The airflow alarm can be activated
integrated assembly optimizes motor cooling. or deactivated depending on user
All rotating parts are unitized and balanced for
smooth, quiet, vibration-free operation.
preference and nature of the work.
Biological Safety Cabinet • Class II, Type B2 (Total Exhaust) Biological Safety Cabinets
Touchpad data entry buttons permit control Programmable automatic UV light A graphical interface indicates cabinet performance.
settings and access to diagnostics, default timer simplifies operation, enhances
settings and hierarchical menus. contamination control, extends UV Digital read-out with alpha-numeric display indicates all
lamp life and saves energy. input, status and alarm functions.
Color coded indicator lamps display green
for primary function (fan operation); blue All functions can be user activated through touch-pad
for secondary function (fluorescent lights programming access; see Operations Manual.
and electrical outlet); and orange for
caution (UV lamp ON).
Sentinel Microprocessor Control System, Programmable When programmed ON • the start-up sequence confirms status with Air Safe and local time display.
• the Personal Identification Number (PIN) access restricts unauthorized adjustments.
• an airflow alarm warns of deviations from normal velocities.
Consult your Esco Operating Manual • The cabinet work zone has no welded • There are no screws in on the front or
or contact your Sales Representative joints to collect contaminants or rust. sides to trap contaminants or complicate
for information on user-preference • All stainless steel work surfaces are cleaning.
programming capabilities built into the accessible for cleaning. • Optional service fittings are offset for
Sentinel microprocessor platform. easier access.
• Tray components lift and remove to
provide easy access and encourage • External plumbing is concealed behind
Cabinet Construction
surface decontamination. trim panels to maintain aesthetics.
Robust construction and enhanced safety
• A recessed central area and stainless • External surfaces are coated with
features qualify the cabinet for the most
steel drain pan channels spills and Esco Isocide antimicrobial coating to
demanding laboratory applications. The
prevent liquids from entering the lower protect against surface contamination
cabinet is fully assembled and ready to install
filtration and blower systems. and inhibit bacterial growth. Isocide
and operate when shipped.
• The drain pan is flush with the side walls eliminates 99.9% of surface bacteria
• The interior work area is formed from a to eliminate concealed or hard-to-clean within 24 hours of exposure.
single piece of stainless-steel with large spaces.
radius corners to simplify cleaning.
Service fixtures are offset for easier reach. Standard cabinets include two fixture provisions on each sidewall
(one provision on each sidewall for 0.9 meter/3 ft. cabinet). Electrical outlets are mounted below service
fixtures to minimize obstructions.
Helpful for certifiers, the hinged maintenance assembly opens to a fixed position on integrated, gas spring
struts providing front service access.
All key components, with the exception of the blower/motor assembly, are mounted outside the air stream
and away from contaminated air to permit service without decontamination. These include fluorescent lamps,
UV lamp, electrical harnesses, electronic boards and microprocessor control.
Panels enclosing potentially hazardous areas or components such as microbiological contamination or electrical
shock are color-coded red to warn service technician.
The Dynamic Chamber™ plenum minimizes physical lifting and accelerates filter change when required.
Work area containment is maintained even when removable components are lifted out for cleaning.
The one piece stainless steel work tray edges are radiused and easy to clean without crevices or joints.
The lower drain trough is a single-piece fabrication with wide open angles and a channel to direct spills
to the drain.
The closed sidewall contains no perforations, air return slots or other hidden areas where contaminants can
accumulate.
Biosafety Cabinets Air Quality Filtration Electrical Safety
Standards ISO 14644.1, Class 3, Worldwide EN-1822 (H14 & H13), Europe UL-C-61010A-1, USA
NSF / ANSI 49, USA
Compliance JIS B9920, Class 3, Japan IEST-RP-CC001.3, USA CSA22.2, No.1010-192, Canada
(LB2-4B2, LB2-6B2)
JIS BS5295, Class 3, Japan IEST-RP-CC007, USA EN 61010-1, Europe
SFDA YY-0569, China
US Fed Std 209E, Class 1 USA IEST-RP-CC034.1, USA IEC61010-1, Worldwide
Antimicrobial Powder-Coating
Biological Safety Cabinet • Class II, Type B2 (Total Exhaust) Biological Safety Cabinets
Model LB2 Biological Safety Cabinet Technical Specifications
271 mm / 10.7" (LB2-3)
430 mm / 16.9" (LB2-4) 250 mm (9.84")
666 mm / 26.2" (LB2-5)
50 mm
(2.0")
820 mm / 32.3" (LB2-6)
250 mm (9.84")
17
2
7
1568 mm (61.7")
(17.5")
OPENING
4
MAX.
450 mm
13 8
5
715 mm (28.1")
9
14
10
")
*
203mm (8
15
OPENING
ACCESS
16
11
45 mm
180 mm
1.8"
140 mm
955 mm / 1260 mm / 1565 mm / 1870 mm (7.1") 12 5.5"
(37.6" / 49.6" / 61.6" / 73.6")
815 mm (32.1")
1115 mm / 1420 mm / 1725 mm / 2030 mm
(43.9" / 56.0" / 67.9" / 80.0")
*Cabinet Size 0.9 meter (3') 1.2 meter (4') 1.5 meter (5') 1.8 meter (6')
Testing Opening Height 203 mm (8") 203 mm (8") 203 mm (8") 203 mm (8")
Working Access Height 248 mm (9.8") 248 mm (9.8") 248 mm (9.8") 248 mm (9.8")
1. Exhaust ULPA filter 6. Electrical outlet Retrofit Kit provision 9. Plugged service fixture provisions 14. Safety glass sliding sash window
2. Blower (0.9 meter / 3' model - one single (2 on each side) 15. Single-piece stainless steel back wall
outlet in workzone) (1.2, 1.5 and 10. Stainless steel single-piece work tray and side walls
3. Downflow ULPA filter, angled
1.8 meter / 4',5' & 6' models - two
4. UV light Retrofit Kit™ provision 11. Stainless steel armrest 16. Removable side panel for
single outlets in workzone)
12. Drain valve Retrofit Kit provision plumbing access
5. IV-Bar Retrofit Kit provision 7. Electrical / Electronics panel
13. Esco Sentinel microprocessor 17. Integral exhaust collar
8. Fluorescent lamp
control system
Accessories and Options (Cont.) timer through Sentinel micro- - Adjustable 395-490 mm
• Cradle stand, electrical hydraulic, processor control panel (15.6"-19.3")
infinitely adjustable, with casters - Emission of 253.7 nanometers for • Ergonomic foot rest
- Elevates to seating or standing most efficient decontamination. - Angled, helps maintain proper
work surface height - Lamp is positioned away from posture
- When lowered permits movement operator line-of-sight for safety and - Adjustable height
through standard doorway proper exposure to interior surfaces. - Anti-skid coating, chemical
Note: Increases exterior dimensions. Note: UV lamp intensity reduces over time resistant finish
and its effectiveness is subject to factors such
Electrical Outlets and Utility Fixtures • IV bar, with hooks
as relative humidity in the cabinet, ambient
• Electrical outlet, ground fault, - Stainless steel construction
air temperature and microbial species in the
North America - Available for all standard Esco
work zone.
• Electrical outlet, Europe / Worldwide cabinets
• PVC armrest
• Petcock (air, gas, vacuum) - Chemically treated, improves • Microscope viewing device
- North America (American) style operator comfort, easy-to-clean. - Mounting and viewing pouch
- Europe / Worldwide style DIN 12898, 712 mm (28") standard size integrated into sash. Factory installed;
DIN 12919, DIN 3537 specify when ordering
• Ergonomic lab chair
Cabinet Accessories - Laboratory grade construction,
• Germicidal UV lamp meets Class 100 cleanliness; alcohol
- Controlled by automatic UV lamp resistant PVC materials
General Specifications LB2-3B_ LB2-4B_ LB2-5B_ LB2-6B_
Nominal Size 0.9 meters ( 3') 1.2 meters ( 4') 1.5 meters ( 5') 1.8 meters ( 6')
External Dimensions 1115 x 815 x 1568 mm 1420 x 815 x 1568 mm 1725 x 815 x 1568 mm 2030 x 815 x 1568 mm
(W x D x H) 40.7" x 33.3" x 61.7" 52.7" x 33.3" x 61.7" 64.7" x 33.3" x 61.7" 80.0" x 33.3" x 61.7"
Gross Internal Dimesnsions 955 x 603 x 715 mm 1260 x 623 x 715 mm 1565 x 603 x 715 mm 1870 x 603 x 715 mm
(W x D x H) 37.6" x 21.3" x 28.1" 49.6" x 24.5" x 28.1" 61.6" x 21.3" x 28.1" 73.6" x 21.3" x 28.1"
Usable Work Area 0.44 m2 (4.7 [Link].) 0.59 m2 (6.4 [Link].) 0.74 m2 (7.9 [Link].) 0.89 m2 (9.6 [Link].)
Tested Opening 203 mm (8") 203 mm (8") 203 mm (8") 203 mm (8")
Working Opening 248 mm (9.8") 248 mm (9.8") 248 mm (9.8") 248 mm (9.8")
Downflow 702 m3/h (413 cfm) 924 m3/h (544 cfm) 1147 m3/h (675 cfm) 1369 m3/h (806 cfm)
Supply ULPA Filter Typical Efficiency >99.999% for particle size between 0.1 to 0.3 microns.
Exhaust HEPA Filter Typical Efficiency >99.99% at 0.3 microns.
NSF / ANSI 49 <62.5 dBA <62.5 dBA <64 dBA <64.5 dBA
Sound Emission
EN 12469 <59.5 dBA <59.5 dBA <61 dBA <61.5 dBA
> 1335 Lux > 1440 Lux > 1200 Lux > 1230 Lux
Fluorescent Lamp Intensity At Zero Ambient
(> 124 foot-candles) (> 134 foot-candles) (> 112 foot-candles) (> 114 foot-candles)
Cabinet LB2-4B2, LB2-6B2 1.5 mm (0.06") 16 gauge electro galvanized steel with Isocide white oven-baked epoxy polyester powder-coating
Construction All other variants 1.2 mm (0.05") 18 gauge electro galvanized steel with Isocide white oven-baked epoxy polyester powder-coating
Net Weight *** 288 kg / 634 lbs 335 kg / 739 lbs 386 kg / 851 lbs 438 kg / 906 lbs
Shipping Weight *** 336 kg / 741 lbs 391 kg / 862 lbs 450 kg / 991 lbs 508 kg / 1120 lbs
Shipping Dimensions, 1230 x 940 x 1900 mm 1530 x 940 x 1900 mm 1910 x 940 x 1900 mm 2150 x 940 x 1980 mm
Maximum (W x D x H) *** 49.8" x 35.4" x 73.4" 59.1" x 35.4" x 73.4" 74.8" x 35.4" x 74.8" 82.7" x 35.4" x 74.8"
Shipping Volume, Maximum *** 2.20 m3 (71.3 [Link].) 2.73 m3 (89.2 [Link].) 3.41 m3 (114.7 [Link].) 4.00 m3 (126.8 [Link].)
* These figures are the required exhaust fan pressure capacity left at the cabinet's duct connection, plus 50% and 100% extra safety margin, but not including the
presure capacity required to overcome the pressure drop from bends, elbows, valves, etc in the duct work. These required pressure capacity has been claculated to
overcome the increased exhaust HEPA filter pressure drop by +50% and +100% due to particle loading. The minimum recommended pressure capcity is +50%,
but to prolong the "life" of exhaust filter, +100% is a safer choice.
** Additional voltages may be available; contact Esco for ordering information.
*** Cabinet only, excludes optional stand.
Biological Safety Cabinet • Class II, Type B2 (Total Exhaust) Biological Safety Cabinets
Microbiological Testing Cross Contamination Test • If there is a gap on the cabinet sealing,
Esco performs testing in accordance The test objective is to evaluate cabinet escaping HPV to the lab will decompose
with more than 10 of the world’s most protection from cross contamination of to become oxygen and water. Escaping
recognized standards for local, regional samples placed simultaneously inside the formaldehyde, however, is harmful
and international criteria. Testing in our workzone. to people in the lab. Therefore HPV
microbiology laboratory is conducted decontamination can be performed
• A nebulizer containing 55 mL of spores while people still working inside the lab,
according to NSF / ANSI 49, EN 12469, and
(5 to 8 x 104/mL) is placed against one while formaldehyde decontamination
JIS K3800. An NSF-accredited biohazard
of the workzone sidewalls. must be performed with no one present
cabinet field certifier is available in-house
full-time to supervise all tests. • Target agar plates are placed 36 cm in the lab. The HPV method improves
(14”) away from the same side wall as safety, productivity, and reduces the
Harmless Bacillus atrophaeus (formerly the nebulizer. time to seal the cabinet.
Bacillus Subtilis) bacteria is used to
• Acceptance: The number of Bacillus • HPV has a better penetration capacity,
challenge the cabinet, then incubated for
atrophaeus CFU recovered on agar resulting in a full decontamination of
48 hours and the Colony Forming Units
plates shall not exceed 2 CFU per test. the cabinet. The formaldehyde method
(CFU) are counted to determine the testing
is known to result in incomplete
results. Increased microbiological challenge HPV Test Compliant: decontamination,
tests with objects inside the cabinet work Safer Hydrogen Peroxide
zone, Bunsen burner, external airflow Decontamination Compatibility • HPV is more effective and rapid against
disturbance, and Human-As-Mannequin biological organisms compared to
Esco biological safety cabinets are Hydrogen
test adapted from Fume Hood development formaldehyde.
Peroxide Vapor (HPV) compliant and
were performed to simulate real-world decontaminatable cabinets approved by • HPV requires approximately 4-7
conditions. BIOQUELL for this patented process. HPV hours for set-up, decontamination,
is a safer and more efficient alternative and tear-down, compared to a total
Personnel Protection Test to conventional decontamination using of 12-15 hours needed to complete
The test objective is to evaluate the safety formaldehyde (CH20): a formaldehyde decontamination
of the cabinet for the personnel operating process.
• HPV is non-carcinogenic and odorless,
on potentially hazardous samples in the • HPV decontamination effectiveness
while formaldehyde is carcinogenic,
cabinet workzone. is independent of temperature and
toxic and has pungent smell.
humidity. Formaldehyde requires
• A nebulizer containing 55 mL of 5 to 8 • HPV biological efficacy is independent temperature above 20°C and relative
x 108 spores/mL [Link] spores is of environmental variables, whereas humidity above 65%
placed inside the workzone, 10 cm (4 formaldehyde efficacy is dependent on
inches) behind the front opening sash. • For information on the BIOQUELL HPV
such variables.
methodology contact Esco or your Sales
• Target slit air samplers and impingers
Representative.
are placed outside the workzone to
capture [Link] spores that may
escape, then the sample is incubated
• Acceptance: The number of B.
atrophaeus CFU recovered from the The Performance Envelope Concept
agar plates shall not exceed 10 CFU
per test. 0.65 Esco cabinets are designed to operate within a
High Inflow
m/s High Inflow performance envelope to maintain protection
High Downflow
Product Protection Test Low Downflow (0.61, 0.43)
for personnel, product and the environment.
(0.58, 0.30) Airflow parameters used to frame the
The test objective is to determine cabinet 0.60
performance envelope include both inflow
protection to the product/samples inside velocity and downflow velocity.
the cabinet workzone from environmental Nominal Inflow The graph illustrates the boundaries of
0.55 Nominal Downflow the performance envelope, as well as the
contaminants. (0.53, 0.35) nominal performance point at which tests
Inflow Velocity
are conducted.
• A nebulizer containing 55 mL of 5 to 8
0.50 The range between high and low Inflow,
x 106 spores/mL [Link] is placed and high and low downflow, together
at 10 cm (4 inches) in front of sash with the fluid dynamics achieved through
Low Inflow sophisticated cabinet design, proportionally
window. Low Inflow
0.45 High Downflow size capture slots, and uniform laminar
Low Downflow
(0.48, 0.40) airflow, combine to deliver a complete
• Target agar plates are placed throughout (0.48, 0.30)
containment and safety solution expected
the entire work surface. of a professional biological safety cabinet.
0.40 For details on the Esco performance envelope
• Acceptance: The number of Bacillus
0.25 0.30 0.35 0.40 0.45 0.50 contact Esco or your Esco Sales Representative.
atrophaeus CFU recovered from the m/s
Downflow velocity
agar plates shall not exceed 5 CFU
per test.
KI Discus Containment Filtration System 26. Magnetic, not mechanical, proximity sensors shall
work in conjunction with the control system to
Test According to EN 12469 6. The cabinet shall have one supply downflow
indicate proper sash position for containment.
filter and one exhaust filter. Supply filter shall
(Potassium Iodide) 27. Fluorescent lamps shall be mounted behind the
be ULPA-type per IEST-RP-CC001.3 and meet
Esco is currently one of the few companies EN1822 (H14) requirements. Exhaust filter shall control panel module out of the work zone.
be HEPA-type per IEST-RP-CC001.3 and meet EN Electronic ballasts shall be used to eliminate
in the world equipped to perform the KI
1822 (H13) requirements. flicker, extend lamp life and reduce heat output.
Discus test for our customers. The KI Discus 28. The UV lamp, if installed, shall be mounted
7. The filters shall be within an aluminum frame
test is defined in the European Standard with mini-pleat design without aluminum behind the control panel and away from the
for microbiological safety cabinets, EN separators; no wood or fiberboard shall be used user’s line of site for protection. (Note: Does not
in the filter assembly. apply to 1.8 meter/ 6 ft. model.)
12469, as a test method for validating the
8. Typical ULPA filter efficiency shall be >99.9997% at 29. The UV lamp shall operate via an automatic
operator/personnel protection capabilities MPPS and 99.999% for 0.1 to 0.3 microns. Typical timer with automatic shut-off managed by the
of the cabinet. HEPA Filter shall be >99.99% at 0.3 microns. microprocessor controller and shall be interlocked
9. An integral filter guard shall be affixed to with the blower/motor and fluorescent lights for
• The KI Discus test shows excellent prevent damage to the filter media. safety.
correlation with the microbiological 10. The filters shall be (a) individually scan tested by 30. The cabinet shall be designed with a 10° angled
front to optimize user comfort, reduce glare and
test method for operator protection, the manufacturer, (b) individually scan tested after
maximize reach into the work area.
assembly, and (b) easily accessible for scan testing
and is useful for validating the actual in situ by means of a dedicated upstream sampling 31. The front grille shall be raised to prevent airflow
containment performance of the cabinet port accessible from within the cabinet. blockage and improve comfort.
on-site. 11. The supply filter shall be angled and oriented 32. Penetrations for petcocks and service fittings
to the 10° cabinet front angle to maximize shall be provided; penetrations shall be offset to
• The KI-Discus takes only 45 minutes as downflow uniformity over the work surface. improve user access.
opposed to 2 days for microbiological 12. A removable, perforated metal diffuser shall 33. The cabinet shall accommodate an optional
testing. be installed below the supply filter to optimize mounting stand for fixed-height.
airflow uniformity and to protect from damage.
• Thus, each Esco Labculture LB2 model Control and Alarm System
is factory tested using the KI Discus Blower System 34. All cabinet functions shall be managed by a
13. The cabinet shall have a direct drive, permanently programmable microprocessor control system
method for operator safety. capable of software updates via Internet
lubricated centrifugal blower/motor dynamically
balanced in two planes compliant to ISO 2710 downloads.
for low noise, low vibration and long filter life. 35. The microprocessor controller shall be mounted
14. The blower/motor shall have an external rotor on the main control panel facing down toward
design and include an automatic thermal cut-out the user.
to disable the motor in case of overheating. 36. The controller shall include soft-touch keypad
Purchase Specifications
15. The blower/motor shall have an automatic speed controls and backlit LCD displays to permit
LB2 Series Class II, Type B2 (Total Exhaust) control to compensate for filter loading. operation of the blower/motor, light, UV lamp,
10 Biological Safety Cabinet 16. The blower/motor system shall be enclosed electrical outlet(s) and menu.
General Performance and Certifications within a dynamic chamber shaped steel plenum 37. The controller shall be user programmable
and integrated with the removable supply filter in situ to enable or disable functions such as
1. The biological safety cabinet shall comply with
assembly to simplify filter changing. PIN (personal identification number) access
one or more of the following international
17. The cabinet shall be equipped with an integral restriction, cabinet start-up protocol, airflow
standards, and the manufacturer shall provide a
exhaust collar for connection to the laboratory alarm and other microprocessor controlled
certified copy of containment and performance
ducting system. operations outlined in the user manual.
tests equivalent to or greater than specified
in the following independent international 38. When programmed ON, the start-up protocol
standards for biological safety, electrical and Cabinet Design, Construction, Cleaning shall perform an automatic pre-purge and post-
other functional characteristics: Class II, Type 18. The cabinet shall be of triple wall design whereby purge cycle to ensure proper cabinet operation.
B2 per NSF / ANSI 49 (USA). all positive pressure plenums capable of handling 39. The controller shall include a blower/motor hours
2. The cabinet shall protect (a) the operator contaminated air shall be surrounded by negative meter to display aggregate motor running time
and laboratory environment from particulates pressure. No positive pressure areas shall be to assist in predictive maintenance.
generated within the work zone; (b) the product accessible external to the cabinet. The third wall
40. Audible and visual alarms shall be provided for
and process within the work zone from airborne shall conceal utilities.
unsafe conditions such as improper airflow or
contamination from ambient air; (c) and the 19. The cabinet shall maintain containment sash position.
product and process within the work zone from performance even when removable work area
41. Airflow shall be monitored by a temperature
cross contamination. components are removed for cleaning.
compensating, thermistor-based, true air velocity
3. Before shipment each cabinet shall be individually 20. The work tray shall be one-piece, removable, sensor mounted in the cabinet.
tested by the KI-Discus test (European Standard EN stainless steel with radius corners without crevices
42. The airflow display and alarm system shall be
12469) to validate operator/personnel protection. or joints.
individually calibrated before shipment.
The retention efficiency for the front aperture shall 21. The cabinet shall have a stainless steel, one-
be not less than 99.999%. Microbiological testing 43. The main control panel shall exhibit continuous
piece fabricated drain trough with open angles
for cabinet performance shall be performed on a display of air velocity and a 24-hour clock display.
to channel spills to a common drain.
statistical sampling basis.
22. The closed sidewall shall be sealed without Certification, Service and Decontamination
4. Each cabinet shall be listed by Underwriters’ perforations, return air slots or concealed areas 44. The cabinet shall be approved for both hydrogen
Laboratories (UL, CUL) or CE for electrical safety. which can contain contaminants. peroxide vapor (HPV) and formaldehyde
5. Original documentation specific to each cabinet 23. The cabinet shall be free of sharp edges, non- decontamination protocol.
serial number shall be provided with the cabinet functional protrusions, bolts, screws or hardware, 45. All panels leading to potentially contaminated and/
and maintained in the manufacturers’ records. and all metal edges shall be deburred. or hazardous areas shall be color coded red.
Test data verifying all performance criteria shall
be available upon request to include: (a) inflow Ergonomics and Convenience 46. All components with the exception of blower/
velocity through direct inflow measurement motor and ULPA filters shall be located outside
24. The front sash shall be frameless to maximize
method; (b) downflow velocity and uniformity; of contaminated air spaces to facilitate servicing
visibility, and accessible for cleaning front and
(c) filter leak scan with aerosol challenge for without the need to decontaminate the cabinet.
back. Sash glass shall be safety glass.
both filters; (d) light, noise, vibration; (e) and 47. All exterior surfaces shall be painted with a
electrical safety. 25. The sash counterbalance shall be suspended on
permanent antimicrobial inhibitor coating to
two high-strength cables, and the sash shall lock
minimize contamination.
into position in the event one cable becomes
detached.
Biological Safety Cabinet • Class II, Type B2 (Total Exhaust) Biological Safety Cabinets
ESCO Airstream® Class II Type B2 Biosafety Cabinets
Ø305 mm
Ø12.0”
100 mm
3.9”
1
9
2
1460 mm / 57.5”
1360 mm / 53.5”
3
670 mm / 26.4”
7
11
12 8
173 mm
6.8”
13
1270 / 1880 mm
50.0” / 74.0”
710 mm / 28.0”
1340 / 1950 mm
52.8” / 76.8”
safety and other HEPA-filtered cabinets for the laboratory with a history
of quality cabinets since 1978. We are predominantly oriented towards the
international marketplace, with sales in more than 70 countries and 95% of
turnover exported. Our products have been independently tested to stan-
dards such as AS1807.5 and EN12469. Products are manufactured under
an ISO 9001 registered quality system.
21 Changi South Street 1 · Singapore 486 777 · Tel: +65 6542 0833 · Fax: +65 6542 6920 · Email: biotech@[Link] · [Link]
Global Offices: Philadelphia, USA ~ Leiden, Netherlands ~ Santiago, Chile ~ Shanghai, China ~ Kuala Lumpur, Malaysia ~ Mumbai, India ~ Dubai, U.A.E.
ESCO Airstream® Class III Biosafety Cabinet
EN 12469:2000.
MAIN FILTRATION AGENTS in the backwall of the workzone for monitoring Industrial-grade support frame constructed
pressure in the workzone during usage. of electro-galvanised steel with an abrasion-
ISO Class 3 air cleanliness within work resistant oven-baked powder-coated
zone as per ISO 14644.1 (equivalent to *Registered trademark of Dwyer finish.
Class 1 as per US Federal Standard 209E, Instruments, Inc
100 times “cleaner” than the usual Class Cabinet airflow is regulated by double fans
100 classification on cabinets offered by CONSTRUCTION FEATURES (refer to diagrams on page 3), thus ensuring
the competition). All materials used in our a fail-safe system that guarantees absolute
products are cleanroom compatible. State-of-the-art Neoprene™ synthetic safety.
arm-length gloves; flame and abrasion-
Mini-pleat separatorless ULPA filter resistant, cleaner, and more resistant to Permanently lubricated direct drive
technology reduces energy consumption and chemicals and flames than regular latex centrifugal blower(s); energy efficient
delivers increased laminar airflow uniformity gloves; each glove is individually tested external rotor motor type design reduces
for better product and cross contamination for airholes. Glove ports are designed to operating costs.
protection. ULPA filters are double scan- make glove-changing easy and most
tested, at the time of manufacturing, and importantly, absolutely safe. Gloves Built-in solid state variable speed controller
after installation. available in sizes 6 and 7. (infinitely adjustable from zero to the
maximum setting) with built-in RFI and
Exclusive vertical laminar flow design Integrated pass-through with noise filters is superior to conventional
ensures that air within the chamber is interlocking doors at side of main chamber “step” controllers.
decontaminated in a consistent fashion, allows materials to be transferred into the
thus preventing the accumulation of cabinet without the risk of environmental Complete welded construction ensures an
particulate contamination in “dead air contamination. air-tight cabinet carcass for absolute
corners”. Product protection is guaranteed containment and operator/environmental
by the ULPA-filtered vertical laminar air flow A sloped cabinet front ensures an safety.
within the cabinet workzone. ergonomic working posture for increased
operator productivity and reduces glare Exclusive Dynamic Chamber™ design
Exhaust air is double-filtered through from the glass surface. surrounds all contaminated air plenums with
high-quality ULPA filters with a rated negative pressure plenums for additional
efficiency of 99.999% at 0.12 microns, Durable and easy to clean stainless safety. Unique airflow technology maintains
thus ensuring complete product, operator steel work surface will never rust or cause cabinet negative pressure at -274Pa
and environmental protection from airborne contamination. / 1.1inWater within the chamber under
biological hazards; a built-in pre-filter also all circumstances for maximum safety and
extends main filter life. Built-in warm white, electronically containment.
ballasted lighting is comfortable to the eyes
CONTROL FEATURES and offers excellent illumination throughout Designed to meet the safety
the work zone in order to reduce operator requirements of IEC 61010-1 / EN
Microprocessor-based Esco Sentinel™ fatigue. Light tubes are mounted out of the 61010-1 / UL 3101-1 / CSA C22.2 No.
control system provides visual / audible air stream for better airflow uniformity. 1010.1-92. Components are UL listed
alarms for airflow. Easy-to-clean membrane UV-resistant tempered glass window / recognised.
touch control panel is supplied with an LCD with built-in stainless steel glove ports
display. (for the number of gloves ports, refer to Extended warranty period of 3
engineering diagrams provided) is designed years excluding consumable parts and
Magnehelic pressure gauge* is mounted to be 100% leak-free. accessories.
General Features Cont. / Class III Cabinet Airflow Profile
Your safety means the world to us: the following tests are
performed individually at our in-house laboratory on every cabinet
produced:
a. Airflow velocities
b. Cabinet leak test
c. Aerosol challenge test for filter integrity
d. Light intensity / noise / vibration levels
e. Airflow pattern visualisation test
f. Electrical safety tests
Exhaust Air is Esco Airstream® Class III Biohazard Safety Cabinet: Airflow Profile
ULPA-filtered
• Air taken in from above the cabinet • All exhaust air is ULPA-filtered and
Intake Air passes through a pre-filter and an ULPA exhausted directly back to the laboratory
filter. Moving into the main chamber of without re-circulation. Optionally, an
the cabinet in a vertical air stream, it exhaust collar can be fitted to allow air to
provides product protection from airborne be ducted via a dedicated ducting system
particulate matter and cross contamination. to the external environment.
Pre-filter This vertical flow air stream is also known
Removes
Larger as the downflow. • Air velocity through an open glove port will
Particles be a minimum of 0.7m/s (140fpm) when a
• Close to the work surface, the downflow single port is open.
air stream “splits” with a portion entering
grilles towards the front or back of the
cabinet, providing a uniform downflow over Class III Biohazard Safety Cabinets are
the entire working area of the chamber. commonly specified for use with the
world’s deadliest microbes. Drawing on the
• Esco’s unique airflow design ensures that air resources and expertise of our cleanroom
Positive Pressure Plenum
within the cabinet is replaced consistently, engineering division, Esco is able to
Vertical Airflow eliminating excessive air turbulence and the provide a complete turn-key service for
accumulation of particulate contamination the construction of critical environments to
in “dead air corners”. complement applications in which Class III
cabinets may be used.
• Air is moved within the cabinet below
the work table, where is passes through Although a Class III safety cabinet provides
two ULPA filters; one mounted below an unparalleled degree of operator and
the work table, and the other above the environmental protection, safety can
main chamber. This provides a fail-safe still be increased by the construction of
mechanism in case any one of the ULPA negative pressure laboratories and the
filters fails to maintain containment. use of isolation suits. When a Class III
safety cabinet is operated within a negative
• Esco’s unique blower system maintains pressure laboratory, any failure in the
negative pressure within the main work Class III cabinet will still be isolated within
chamber under all circumstances. Unlike the laboratory. Assuming the cabinet
the competition, our Class III biohazard fails, contamination from the cabinet will
safety cabinets do not utilize supply “spill-over” into the laboratory but will not
blowers, which, when improperly balanced, contaminate the general environment,
may result in the working area being since all air within the laboratory is filtered
positively pressurized. before being exhausted to the external
Downflow / Supply ULPA Filter environment.
First Exhaust ULPA Filter
Second Exhaust ULPA Filter
Engineering Diagrams for Available Sizes
GLOVE SIZES
To find out your glove size, measure (in inches) around your hand with a tape measure at the
place indicated on the diagram. This measurement, in inches, is the closest to your glove size,
i.e. 8” is equal to a glove size of 8.
You should use your dominant hand, the right if you are right-handed, and the left if you are left-
handed.
15 15
1 1
ESCO
12 Air Flow 0.3 m/s
ESCO
13 13
2 2
14 15
400 mm 400 mm
15.7” 15.7”
1340 mm / 52.8”
1645 mm / 64.8”
1665 mm / 65.6”
1970 mm / 77.6”
3
10
11
4
1
5
12 Air Flow 0.3 m/s
ESCO
6 290 mm / 11.4”
650 mm / 25.6”
13
340 mm / 13.4”
2250 mm / 88.6”
7
14
400 mm
15.7” 8
900 mm / 35.4”
1950 mm / 76.8”
850 mm / 33.5”
2275 mm / 89.6”
Engineering Details
1. Esco Sentinel™ Microprocessor Control 2. Pass Box 3. Pre-filter 4. Electrical / Electronics Panel
5. ULPA Filter (Downflow) 6. Fluorescent Light 7. Glove Ports (Gloves available in sizes 7,8 & 9)
8. First Exhaust ULPA Filter 9. First Exhaust Blower 10. Second Exhaust ULPA Filter 11. Second Exhaust Blower
12. Optional UV Lamps 13. Pressure Gauge
14. Work Zone Back Wall: Optional Electrical Outlet Retrofit Kit™ Provisions: 2 outlets in work zone one on each side)
15. Optional Exhaust Collar
Esco Catalogue Revision No: AC3.v3.6.2005
Specifications subject to change without notice.
Technical Specifications ©2005 Esco Micro Pte. Ltd.
Esco Airstream® Class The following optional Retrofit Kits™ are available from Esco for
III biosafety cabinets are Airstream Class III biohazard safety cabinets. Esco’s Retrofit
available in full stainless Kit™ system allows the user to field-upgrade their cabinets with
steel on a customized optional accessories post-purchase. NOTE: Service fixtures are
basis, suitable for the not allowed in Class III biosafety cabinets, as Class III biosafety
pharmaceutical industry. cabinet construction is required to be airtight.
Standard
Size
General Specifications AC3-4BX AC3-5BX AC3-6BX
External Dimensions 1340 x 850 x 2250 mm 1645 x 850 x 2250 mm 1950 x 850 x 2250 mm
(Width x Depth x Height) 52.8” x 33.5” x 88.6” 64.8” x 33.5” x 88.6” 76.8” x 33.5” x 88.6”
Internal Work Zone 1240 x 650 x 650 mm 1545 x 650 x 650 mm 1850 x 650 x 650 mm
(Width x Depth x Height) 48.8” x 25.6” x 25.6” 60.8” x 25.6” x 25.6” 72.8” x 25.6” x 25.6”
Number of Glove Ports 2 ports (refer to drawing on page 3) 4 ports (refer to drawing on page 3) 4 ports (refer to drawing on page 3)
Glove Type & Sizes Available Neoprene® polychloroprene synthetic rubber gloves: Available in sizes 7, 8 (standard size) and 9.
Air Volume (At Initial Velocity) 603 cmh / 355 cfm 756 cmh / 445 cfm 902 cmh / 531 cfm
Cabinet negative pressure 274Pa / 1.1inH2O (exceeds EN12469:2000 requirement of 199.3Pa / 0.8inH2O and NSF49 requirement of 124.5Pa / 0.5inH2O
Individually performance tested and certified at factory under controlled conditions for:
General requirements: IEST-RP-CC002.2 and AS1386.5
Standards Compliance Air cleanliness: ISO 14644.1 Class 3, IEST-G-CC1001, IEST-G-CC1002 and other equivalent air cleanliness requirements
Filter performance: IEST-RP-CC034.1, IEST-RP-CC007.1, IEST-RP-CC001.3 and EN1822
Electrical safety: IEC 61010-1 / EN 61010-1 / UL 61010A-1 / CSA C22.2 No. 1010.1-92
Air Cleanliness Within ISO 14644.1 Class 3, US Federal Standard 209E Class 1 / M1.5, AS 1386 Class 1.5, JIS B9920 Class 3, BS5295 Class C,
Working Area Class M10,000 as per KS 27030.1 and other equivalent cleanliness classifications of the VDI 2083 and AFNOR X44101
Main Filter Downflow ULPA filter with integral metal guards and filter frame gaskets; fully compliant with EN1822 and IEST-RP-CC001.3 requirements
Type Exhaust (x2) ULPA filters with integral metal guards and filter frame gaskets; fully compliant with EN1822 and IEST-RP-CC001.3 requirements
Minimum: 99.997% at 0.3µm / 99.988% at 0.12µm / 99.987% at MPPS
Downflow
Typical: 99.99927% at 0.3µm / 99.9977% at 0.12µm / 99.9972% at MPPS
Main Filter
Minimum: 99.99962% at 0.3µm / 99.99965% at 0.12µm / 99.99946% at MPPS
Efficiency 1st Exhaust
Typical: 99.999928% at 0.3µm / 99.99996% at 012µm / 99.999924% at MPPS
Ratings
Minimum: 99.99916% at 0.3µm / 99.9985% at 0.12µm / 99.9982% at MPPS
2nd Exhaust
Typical: 99.99983% at 0.3µm / 99.99979% at 012µm / 99.99971% at MPPS
Pre-Filter Disposable and non-washable polyester fibers with 85% arrestance / EU3 rated
Typically <60 dBA at initial blower speed setting measured as per EN 12469:2000, based on 4 feet cabinet, subject to acoustic
Noise Level
properties of test environment
Light Intensity >1700 lux / >158 foot candles, measured at work surface level (zero background) as per IEST-RP-CC002.2
Main Body Construction 1.5mmt / 0.06” / 16 gauge electro-galvanised steel with white oven-baked epoxy powder-coated finish
Front Window Construction Colourless and transparent UV-absorbing 8mm / 0.31” tempered glass
Work Surface Construction 1.2mmt / 0.05” / 18 gauge stainless steel grade 304
Maximum 220-240VAC
1144W / 4.97A 1164W / 5.06A 1180W / 5.13A
Power / 50Hz 1Ph
Consumption 110-130VAC
/Current 1632W / 13.6A 1652W / 13.77A 1668W / 13.9A
/ 60Hz 1Ph
Net Weight (Approximate) 351.5 kgs / 775 lbs 439.4 kgs / 969 lbs 527.3 kgs / 1163 lbs
Max Shipping Weight 601 kgs / 1325 lbs 751.2 kgs / 1656 lbs 901.5 kgs / 1988 lbs
Max Shipping Dimensions 1995 x 1300 x 2750 mm 2245 x 1300 x 2750 mm 2550 x 1300 x 2750 mm
(W x D x H) 78.5” x 51.2” x 108.3” 88.4” x 51.2” x 108.3” 100.4” x 51.2” x 108.3”
Max Shipping Volume 7.13 cbm / 251.8 cbf 8.03 cbm / 283.6 cbf 9.12 cbm / 322.07 cbf
21 Changi South Street 1 · Singapore 486 777 · Tel: +65 6542 0833 · Fax: +65 6542 6920 · Email: biotech@[Link] · [Link]
ESCO Formalin Vaporizer for Decontamination
Decontamination Cycle Electrical and electronic components Automatic control of the entire
are isolated from the heating chamber to decontamination cycle .
Operator starts process by pressing a prevent exposure to chemical fumes and No manual intervention required for the
button on the touch pad heat which can reduce component life entire process
Pre-delay timer allows decontamination Side mounted handles for easy In case of power failure, the vaporizer
operator to seal up the cabinet before carrying will recover automatically once the power
vaporization of ammonia commences. Delay Generously-sized formalin and ammonia resumes without any intervention
period can be set from 1 to 59 minutes. chambers (800mL maximum capacity in Microprocessor unit has a built-in
Vaporization of formalin solution total) which allows even the largest of watchdog and will self-reset in case of any
commences to release the formaldehyde safety cabinets (up to 6 foot console units) processor failure
gas to decontaminate the cabinet to be decontaminated with the maximum Pre-delay, contact time, neutralisation
LCD display will show reminder for agent volume recommended in EN12469 timers are adjustable using the touch
operator to cycle the cabinet fans to ensure Thermostats for the formalin and ammonia control pad on the front of the unit
even distribution of the gas chambers to prevent overheating Large, backlit, LCD display
Heater automatically cuts off when Backup thermostat to ensure the Audible and visual alarms for in case of
formalin solution has completely boiled temperature of the external casing does overheating during a cycle
off not rise above a safe level Password control to prevent
Contact time timer starts to regulate the Increased diameter liquid chamber inlets decontamination timers from being adjusted
period for which the formaldehyde gas is to allow the formalin and ammonia solutions by unauthorised personnel
exposed to the cabinet internal surfaces for to be poured into the chambers more easily Easy to clean touch control panel without
the decontamination to take effect without spillage any switches or manual relays to corrode
Neutralisation process begins Internal heat insulation to prevent cross
automatically to vaporize the ammonia heating of chambers and damage to the Ordering
solution to neutralise the formaldehyde electrical and electronics components
gas Manual shut-off isolating switch and Esco Formalin Vaporizer is available
LCD display will show reminder for safety fuse with the following electrical configurations.
operator to cycle the cabinet fans to ensure Universal IEC power inlet design Please ensure that the ordering code
even distribution of the gas (appropriate cordset must be specified on corresponds to your electrical configuration
Neutralisation timer starts to regulate order depending on country of use) requirements when ordering.
the period for which the neutralisation
Power Supply
reaction takes place Control System Features Ordering Codes
Options
General Construction Features The first and only microprocessor 220-240VAC,
FV-001
controlled vaporizer unit in the world 50HZ / 60HZ, 1 phase
Durable, heavy-duty full stainless steel specifically designed for safety cabinet 100-130VAC,
FV-002
construction decontamination 50HZ / 60HZ, 1 phase
Esco Catalogue Revision No: FV.v1.10.2004 ~ Specifications subject to change without notice. ~ ©2004 Esco Micro Pte. Ltd.
21 Changi South Street 1 · Singapore 486 777 · Tel: +65 6542 0833 · Fax: +65 6542 6920 · Email: biotech@[Link] · [Link]