To calibrate the detector, use an appropriate span gas cylinder, constant flow regulator and the Sensepoint Toxic Gassing Cap (refer to Sensepoint Technical Manual 2106M0502 for details). The flow rate is used of approximately 1 to 1.5 litres per minute for calibration. Apply the gas at the flow rate for the recommended application concentration and time please see Table 6.
Range 0 to 20ppm 0 to 50ppm 0 to 100ppm 0 to 100ppm 0 to 200ppm 0 to 500ppm 0 to 5ppm 0 to 15ppm 0 to 25% V/V 0 to 50ppm 0 to 1000ppm to 1000ppm to 10000ppm 0 to 15ppm 0 to 50ppm 0 to 100ppm 0 to 10ppm 0 to 50ppm
Recommended Test Concentration 10ppm 20ppm 50ppm 50ppm 100ppm 250ppm 3ppm 10ppm 19% V/V 25ppm 500ppm 500ppm 3000ppm 10ppm 20ppm 50ppm 5ppm 20ppm
Table 6 33
Application Operating Temp. Time Minutes Min. Max. 3 mins 3 mins 3 mins 3 mins 3 mins 3 mins 10 mins 10 mins 1 min 10 mins 10 mins 3 mins 3 mins 5 mins 5 mins 5 mins 5 mins 5 mins -20C +50C -20C +50C -20C +50C -20C +50C -20C +50C -20C +50C -20C +50C -20C +50C -15C +40C -20C +40C -20C +40C -15C +40C -15C +40C -15C +40C -15C +40C -15C +40C -15C +40C -15C +40C