Corrosion in Titanium Dental Implants/Prostheses - A Review


Rahul Bhola*, Shaily M. Bhola, Brajendra Mishra and David L. Olson
Dept of Metallurgical & Materials Engineering, 1500 Illinois Street, Golden, Colorado 80401, USA *Corresponding author (rbhola@mines.edu) - Rahul Bhola Received 31 July 2010; Accepted 3 August 2010; Available online 1 March 2011 The corrosion of biomaterials primarily dental implants/prostheses has a significant clinical relevance. In spite of several recent advances in metallurgical and materials science and remarkable improvements in the design and development of surgical and prosthetic materials, failures do occur sometimes. The current paper focuses the problem of corrosion in titanium based implants and prostheses with its clinical significance and emphasis on galvanism induced in-vitro and in-vivo.

Introduction The general consensus about the most corrosion resistant biocompatible metallic biomaterials are the special metals namely-titanium, niobium, tantalum and their alloys, followed by cobalt based alloys and finally the stainless steel grades [1-3]. The most commonly used implant/ prostheses material used today has been summarized below in Table 1 with their common names, UNS, ASTM, ISO and alloy designations [4-6]. Commercially pure titanium and its alloys are known for their use in medical application owing to their good corrosion resistance, biocompatibility and bioactivity in the human body [7].Titanium and its alloys have been used as prosthetic material in several reconstructive and resective procedures in human body since many years [8]. The use of dental implants in the partial and complete edentulism has become the primary treatment regimen in the modern dentistry [9]. These implants appeared as early as late 1920s and gained widespread usage during the last 2-3 decades. Although these biomaterials have good mechanical and biological properties there corrosion resistance is still critical for the overall success of the treatment procedure. It has been long recognized that the corrosion products formed as a result of metal-environment interactions have a significant bearing on the biocompatibility and long term stability of the prostheses/implant. The material used must not cause any biological adverse reaction and must retain its form and properties [10-12] during function. Human stomatognathus is subjected to varying changes in pH and temperature owing to differences in local, systemic, environmental, economic and social conditions for each individual. Corrosion can result from the presence of a number of corrosive species like hydrogen ion (H+), sulfide compounds (S2-), dissolved oxygen, free radicals (O2, O-), and chloride ion (Cl-) resulting in the metal surface breakdown and a consequent adverse tissue reactions [13].

Figure 1: Prerequisites for an implant biomaterial [17]

Corrosion in Titanium Dental Implants/Prostheses - A Review


Table 1: Metallic biomaterials for surgical implant and prostheses. [4-6]
Metallic Biomaterials for Implants/Prostheses Material CommonName/ Designation Trade Name Speciality Metallic Biomaterials Ta, Unalloyed, cast Zr2.5 Nb Ni45Ti Titanium Base Biomaterials Ti CP1 Ti CP2 Ti CP3 Ti CP4 Ti3Al2.5V Ti5Al2.5Fe Ti6Al4V Ti6Al4V cast Ti6Al4V ELI Ti6Al7Nb Ti15Mo Ti12Mo6Zr2Fe Ti11.5Mo6Zr4.5Sn Ti13Nb13Zr Ti45Nb Ti35Nb7Zr5Ta Cobalt Base Biomaterials Co28Cr6Mo casting alloy Co28Cr6Mo wrought alloy #1 Co28Cr6Mo wrought alloy #2 Co28Cr6Mo wrought alloy #3 Co20Cr15W10Ni1.5Mn Co20Cr20Ni5Fe3.5Mo3.5W2Ti Co19Cr17Ni14Fe7Mo1.5Mn Co20Cr15Ni15Fe7Mo2Mn Co35Ni20Cr10Mo Steel Base Biomaterials Fe18Cr14Ni2.5Mo Fe18Cr12.5Ni2.5Mo cast Fe21Cr10Ni3.5Mn2.5Mo Fe22Cr12.5Ni5Mn2.5Mo Fe23Mn21Cr1Mo1N Unalloyed Tantalum () Zircadyne Nitinol CP1 () CP2 () CP3 () CP4 () Ti3Al2.5V (/ ) Tikrutan (/ ) Ti6Al4V (/ ) Ti6Al4V (/ ) Ti6Al4V ELI (/ ) Ti6Al7Nb (/ ) Ti15Mo (Metastable ) TMZF (Metastable ) Beta 3 (Metastable ) Ti13Nb13Zr (Metastable ) Ti45Nb (Metastable ) TiOsteum (Metastable ) Cast CoCrMo Alloy 1 Alloy 2 GADS L605 Syncoben Grade 2 Phynox Grade 1 Elgiloy 35N 316 L Stainless Steel 316 L Stainless Steel REX 734 XM 19 108 UNS Design ation R05200 R60705 N01555 R50250 R50400 R50550 R50700 R56320 Unassigned R56400 R56406 R56401 R56700 R58150 R58120 R58030 R58130 R58450 R58350 R30075 R31537 R31538 R31539 R30605 R30563 R30008 R30003 R30035 S31673 Unassigned S31675 S20910 Unassigned ASTM Designation F 560 F 04.12.45 F 2063 F F F F F F F F F F F F F F F 67 67 67 67 2146 1472 1108 136 1295 2066 1813 9046 1713 04.12.44 04.12.23 ISO Designation ISO 58322 ISO 58322 ISO 58322 ISO 58322 ISO 583210 ISO 58323 ISO 58323 ISO 583211 ISO 58324 ISO 5832 12 ISO 583212 ISO 58325 ISO 58328 ISO 58327 ISO 58327 ISO 58326 ISO 58321 ISO 58329 Alloy #

35

0.0 2.5 55.0 0.1 0.2 0.3 0.4 5.5 7.5 10.0 10.1 10.2 13.0 15.0 20.0 22.0 26.0 45.0 47.0 34.0 34.1 34.2 34.3 46.5 49.0 58.5 59.0 65.0 34.5 34.6 37.0 42.0 46.0

F 75 F1537 F 1537 F 1537 F 90 F 563 F 1058 F 1058 F 562 F 138 F745 F 1586 F 1314 F 04.12.35

Nowadays, osteo-integrated implants/prostheses can be regarded as well proven safe alternatives on a large scale to functionally rehabilitate both partially and completely edentulous patients. One of the chief success factors related to modern titanium implant system is the achievement of a fast and tight interconnection between the implant surface and the bone tissue owing to its low thermal conductivity, low density, lower elastic modulus mismatch compared to bone, high hardness, outstanding biocompatibility and remarkable corrosion resistance [1416]. The prerequisites of a successful implant system have been summarized in Figure.1. Physical metallurgy of titanium alloys Titanium is a transition metal with an incomplete d-shell in its electronic structure that enables it to form solid solutions with most substitutional elements having a size factor within 20% (Hume-Rotherys principles for substitutional and interstitial solutions). In its elemental form titanium has a high melting point (1678C), exhibiting a hexagonal close packed crystal structure (a/alpha) up to the beta transus temperature (882.5C), transforming to a body centered cubic structure (b/beta) above it[18]. Titanium alloys may be classified accordingly either a , near a or a + b, metastable b or stable b depending upon their room temperature microstriucture.[18-19]. In this regard substitutional alloying elements for titanium fall into three basic categories 1. Alpha stabilizers like Al, O, N, C 2. Beta stabilizers like Mo, V, Nb, Ta (isomorphous), Fe, W, Cr, Si, Ni, Co, Mn, H (eutectoid) 3. Neutral stabilizing elements like Zr. Alpha and near alpha titanium alloys exhibit superior corrosion resistance but lower strength. On the other hand a + b alloys exhibit higher strength due to presence of both phases in the microstructure.Their properties depend upon composition, the relative proportions of each phase present with their prior thermal treatment and thermo mechanical processing conditions. Beta alloys in contrast have high strength, good formability and hardenability. Beta alloys also offer lower elastic modulus and higher corrosion properties compared to other alpha and mixed alloys there by suggesting lower coefficient of elastic

36

R. Bhola, S.M. Bhola, B. Mishra and D.L. Olson DE = DE - lnK electrochemical reaction Nernst Equation for the

Where DG, DE are the energy and potential change, DG and DE are the standard free energy and potential change, R is universal gas constant, T is the temperature under study, n is the number of valence electrons involved in the reaction and K is the equilibrium constant of the given reaction at a specified temperature. 2. Corrosion Kinetics The electrochemical kinetics include factors that physically, obstruct corrosion from taking place [24]. These factors tend to hinder the various electrochemical reactions going in the system either through concentration, temperature or velocity kinetics in the system. It is represented by log current density axis on Evans diagram and is majorly driven by the overpotential equation hconc = log[1 Where, h conc is the concentration overpotential, R is universal gas constant, T is the temperature under study, n is the number of valence electrons involved in the electrochemical reaction, ic is the cathodic current density and iL is the limiting current density. Types of corrosion Uniform corrosion (Generalized corrosion) A uniform regular removal of the metal from the surface is usually the most common mode of corrosion. The corrosive environment must have same access to all parts of the metal surface and the metal itself must be metallurgically and compositionally uniform. At times general corrosion in aqueous body fluids like phosphate buffer saline (PBS), ringers lactate (RL), normal saline (NS) etc may take the mottled form, severely roughened metal surface that resembles localized attack. This uneven localized attack results from variations in the corrosion rate of localized surface patches due to localized masking of metal surfaces by process scales, corrosion products, food lodgment and surrounding and adjacent super structures. When titanium is in the fully passive condition, corrosion rates are typically less than 0.02 mm/yr (0.8 mil/yr) and well below the 0.13 mm/yr (5 mils/yr) maximum corrosion rate commonly accepted for biomaterial design and application. This minimal acceptable corrosion rate is primarily due to the finite +4 oxidation of titanium alloys owing to the formation of adherent TiO 2 film although the surface oxide is more complex than a single TiO 2 oxide over their surface. General corrosion becomes a concern at high temperatures in highly acidic environments owing to consumption of hot, spicy and sticky foods. In strong and/ or hot reducing acids (plaque deposits) the oxide film of titanium can deteriorate and dissolve, and the unprotected metal is oxidized to the violet colored soluble trivalent ion (Ti3+) in acid solutions which is further converted to pale yellow Ti4+ ion in presence of oxidizing species which on further hydrolysis may form insoluble TiO 2 precipitates/ scales and inhibiting subsequent corrosion. Uniform corrosion for titanium implants can be determined from weight loss data (increase or decrease in weight

Figure 2: Pseudo binary phase diagram for titanium beta stabilizer

mismatch with the bone and favorable stress distribution and outcome.[20]. A beta alloy is fundamentally defined as an alloy whose chemical composition lies above bc as shown in Figure 2 that is, it contains sufficient amount of beta stabilizing agents to retain 100% body centered cubic microstructure upon quenching from above the beta transus temperature [20-21]. Alloys lying above this critical minimum level of beta stabilizing content may still lie within the two-phase region, making the as-quenched beta phase being metastable with the vulnerability of precipitating a second phase upon aging. Titanium alloys with increasing alloying content, exceeding a critical beta value are considered stable beta alloys where no precipitation of the second phase takes place during practical long-time thermal exposure. Process variations are traditionally used to control the alloy microstructure and therefore to optimize titanium alloy properties such as ductility, strength, fatigue resistance and fracture toughness. The effects of various microstructures are then correlated with engineering properties and the most common microstructural features studied are the metastable beta alloys, beta grain size, size and age distribution of aged alpha phase. Apart from the alpha phase precipitation transient b or w phase or several other intermetallic compounds may also be observed in the metastable beta region depending upon alloy composition, heat treatment, processing history and service conditions [22-23]. General Concepts related to corrosion There are two essential features determining corrosion of an implant 1. Corrosion Thermodynamics The application of chemical thermodynamics are primarily related to the oxidation-reduction reactions for the given electrochemical system. The system tends to attain the lowest energy by releasing energy during the reaction [24]. It is represented by the potential (voltage) axis on the Evans diagram and is given by equation DG = DG + RT lnK for the chemical reaction

Corrosion in Titanium Dental Implants/Prostheses - A Review depending upon the environment and by products in accordance to ASTM G1 & G31), dimensional changes (shape, size, appearance and texture) and electrochemical methods (anodic and cathodic polarization, cyclic voltammetry and electrochemical impedance measurements). Corrosion rates in millimeters per year for titanium alloys can be calculated from weight loss data as under: Corrosion rate (mm/yr) = where d is the titanium implant alloy density (in grams per cubic centimeter which is approximately 4.51g/cc for c.p.Ti), A is the sample surface area (in square centimeters), t is the exposure time (in hours), and W is the weight change (in grams). Corrosion rates in millimeters per year can be calculated from electrochemical measurements on the other hand using the equation: Corrosion rate (mm/yr) = where icorr is the measured corrosion current density (in milliamps per square centimeter), d is titanium alloy density (in grams per cubic centimeter), and EW is the equivalent weight for titanium. The equivalent weight for titanium is approximately 16 under reducing acid conditions and 12 under oxidizing conditions depending upon the number of valence electrons involved. The value of i corr is typically determined from Tafel slope extrapolation or linear polarization methods [25-26] Pitting corrosion Localized corrosion attack in an otherwise resistant surface produces pitting corrosion. It usually occurs on base metals which are protected by a naturally forming thin film of an oxide (for instance the firmly adherent TiO 2 over the Ti surface) when the potential of the film exceeds the breakdown potential of the oxide in a given environment. In the presence of certain ions like chlorides and sulphides the film locally break downs and rapid dissolution of underlying metal occurs in the form of pits. When the anodic breakdown (pitting) potential of the metal is equal to or less than the corrosion potential under a given set of conditions, spontaneous pitting can be expected. Because of the protective oxide films the titanium implant surface exhibits anodic pitting potential that are very high (1V) compared to other biomaterials used (iron, steel, cobalt-chromium alloys etc). Thus pitting corrosion is not of much concern in the oral environment for titanium alloys. For example pitting potentials exceed +5 to +10 V versus the saturated calomel electrode (SCE) in body fluids like chlorides (NS) and lactates(RL) and typically +60 to +80 V in sulfates and phosphates(PBS).Although pitting potential for titanium dental implants is not of much clinical significance but this property of protection potential or repassivation can still be used to define the minimum potential at which pitting can be maintained [27]. Guiding principles under the American Standards for Testing of Materials G3 & G5, electrochemical techniques involving potentiostatic and potentiodynamic testing are

37

used similar to uniform corrosion measurements to study pitting behavior of titanium dental implants [28-29]. Anodic pitting potential requires slow scan rates (d0.5 mV/s) and surface condition of the implant surface under test. For example, abraded or sandblasted implants will exhibit significantly lower pitting potentials compared to acid pickled implant surfaces. Repassivation potentials on the other hand can be measured using the galvanostatic technique [27,30] by impressing an anodic current density of approximately 200 mA/cm2 on the specimen for at least several minutes before measuring the repassivation potential of the sample. Weber and his coworkers [31] studied the aspects of dental corrosion on titanium system using various electrochemical techniques on titanium and its alloys with iron as an important constituent in dental media. The susceptibility to localized pitting corrosion of titanium and its alloys were evaluated by the breakdown potential, pitting potential, corrosion current density and the corresponding anodic polarization curves and tafel slopes. Crevice corrosion Localized crevice corrosion occurs from the geometry of the implant/prostheses assembly. Corrosion of an alloy is greater in the small sheltered volume of the crevice created by contact with another material. The other metal could be part of the fastener of the same or different alloy, a sheltered crown, cement packing or implant prostheses joint [32-33]. Crevice corrosion seems to prefer (metalmetal) Ti implant superstructure and base metal crown crevice with constricted space and oxygen gradient. Titanium alloy implants may be subjected to localized crevice attack exposed to short time periods of hot (>70 C, or 160 F) chloride, bromide, iodide, fluoride or sulfate containing solutions during electrosurgery, electrocautery or thermocautery procedures. The reduction in pH and increase in crevicular chloride ion concentration are the essential factors in the initiation and propagation of the pits. When the acidity of the micro-environment around crevice increases with time it dissolves the passive oxide layer thereby causing localized destruction and crevice corrosion [34]. The mechanism for crevice corrosion in titanium is very similar to that seen in stainless steel, in which oxygendepleted reducing acid conditions develop within tight crevices [32]. The model for crevice corrosion is illustrated in Figure 3. Dissolved oxygen or other oxidizing species in the bulk solution are depleted in the restricted volume of solution in the crevice. Finite surface oxidation in crevices consumes these species faster than diffusion from the bulk solution can replenish them as a result metal potentials within crevices become more active compared to metal surface exposed to the bulk body tissues. This situation creates a micro electrochemical cell in the bony socket where the shielded crevices become anode and corrode, and the surrounding more noble metal surface becomes cathode and is protected. Titanium chlorides formed within the crevice are unstable and tend to hydrolyze, forming hydrochloric acid (HCl) and titanium oxide/hydroxide corrosion products. Because of the small, restricted volumes of solution within tight implant-superstructure crevice, low crevice pH levels (0-

38

R. Bhola, S.M. Bhola, B. Mishra and D.L. Olson

Figure 3: Illustration showing the mechanism of crevice corrosion for titanium in aqueous chloride media (NS)

Figure 4: Micrograph of hydrogen embrittled unalloyed titanium at a magnification of 200

1) with high H+ ions concentration can develop. These local reducing acidic conditions can result in severe and rapid localized active corrosion within crevices, depending on alloy resistance, temperature, and redox potential of the surrounding crevicular fluid. Several oxidizing species such as oxygen, chlorine, ferric ion (Fe3+), and cupric ion (Cu2+)from food and saliva, tend to effectively inhibit the general corrosion of exposed titanium implant surfaces, but most of these tend to aggravate the onset and propagation of titanium alloy crevice corrosion. These ionic species are excellent cathodic depolarizers and rate controlling, thus accelerate cathodic reduction kinetics involved in the electrochemical process. These cationic oxidizing species will not diffuse into the active crevice to inhibit attack compared to certain anionic oxidizing species such as NO3-, ClO3-, OCl-, CrO32, ClO4-, and MnO4-, which can migrate into the implantsuperstructure crevice inhibiting corrosion under halide solutions. Crevice corrosion testing of titanium implants in function is insidious and very rapid, and may leach several ions into the crevicular space activating the host complement response and causing an adverse reaction that may or maynot be tolerated. Several crevice test assemblies are employed including the multiple-crevice serrated washer and the sandwichtype crevice test assembly (consisting of 25 to 38 mm square flat sheet or plate specimen, with thin polytetrafloroethylene gaskets interspersed to provide the desired number of metal-to-metal and metal-to-gasket crevices) for the laboratory testing of crevice corrosion in titanium implants over varying time duration. In addition to visual and weight change measurements, monitoring of creviced implant potential and current has been used to a limited extent to identify the initiation of titanium alloy crevice corrosion.[35] Environmentally induced cracking Titanium alloy implants/prostheses processed under hydrogen-containing environments and under conditions

in which galvanic couples or cathodic charging (impressed current) causes hydrogen to be evolved on metal surfaces may cause hydrogen induced cracking (HIC) [36-37] and the impairment of ductility of implant/prostheses permanently and irreversibly. The tenacious TiO 2 surface film over the implant surface is a highly effective barrier to hydrogen penetration. Small amounts of moisture or oxygen in hydrogen-gas-containing environments effectively maintain this protective film, thus avoiding or limiting hydrogen uptake compared to anhydrous hydrogen atmospheres at higher temperatures and pressures causing increased uptake [38-39]. Hydrogen attack is the reaction of hydrogen with carbides in titanium to form methane, resulting in decarburization, surface voids and blisters. These voids are formed when the atomic hydrogen migrates from the surface to internal defects and inclusions, where molecular hydrogen gas can nucleate, generating internal stresses to deform and rupture the metal locally. In and - alloys, excessive hydrogen uptake can induce the precipitation of titanium hydride in the phase. These acicular-appearing hydride platelets, as shown in Figure 4 are brittle and have been well characterized in literature in detail [38]. Biaxial and triaxial mechanical properties such as ductility, cold-drawing/formability and impact toughness in and near- alloys are very sensitive to hydrogen levels [4042]. Hydrogen contents above critical levels can result in sustained load cracking which drastically reduce the service life of cracked/notched or etched surfaces of implants under functional load of mastication and occlusion [40-42].Beta titanium alloys on the other hand have high solubility for hydrogen, therefore embrittlement is generally not associated with hydride precipitation. Severe loss in ductility and formability may not occur below several thousand parts per million of hydrogen [43]. This enhanced tolerance of beta alloys to hydrogen decreases with aging (tempering/ normalizing) as more of the phase is precipitated. This increased alloy tolerance must be weighed against higher hydrogen uptake rates that result from the higher hydrogen diffusion coefficient in titanium alloys [43]. Covington et al. [39] have characterized the three conditions that must coexist to cause embrittlement

Corrosion in Titanium Dental Implants/Prostheses - A Review of commercially pure titanium implant alloys for dental applications in aqueous media. 1. A mechanism for generating nascent (atomic) hydrogen on a titanium surface. This hydrogen generation may be from a galvanic couple, an impressed cathodic current, corrosion of titanium, or severe continuous abrasion of the titanium surface in an aqueous medium. 2. Metal temperature above approximately 80 C (175 F), above which the diffusion rate of hydrogen into titanium becomes significant [44] 3. Solution pH less than 3 or greater than 12, or impressed potentials more negative than -0.75 V (versus Ag/AgCl reference electrode) [45-47] No hydrogen uptake and embrittlement problems occur when titanium implant is galvanically coupled to fully passive materials like titanium alloys, resistant (passive) stainless steels, copper alloys, and nickel-base alloys, depending on surrounding environments. Very high cathodic charging of dental implant titanium alloys may cause enhanced hydride film formation and penetration over the implant surface and HIC at room temperature. Hot alkaline (>80 C, and pHe12), conditions may also result in increase hydrogen uptake and embrittlement of titanium alloys. Hydrogen analysis for the titanium implants can be done using the electrochemical impedance spectroscopy and measuring the relative impedance of the retained hydrogen as a function of impedance or by the hot vacuum extraction method, where the implant samples are heated to 1100 to 1400 C for a stipulated time to reversibly release the absorbed hydrogen, followed by evolved gas measurements. Stress induced cracking Stress-corrosion cracking (SCC) is a fracture phenomenon caused by the combined factors of tensile stress, a susceptible alloy, and a corrosive environment. The metal normally shows no evidence of general corrosion attack, although slight localized attack in the form of pitting may be visible in certain metals. Usually specific combinations of metallurgical and environmental conditions cause SCC. This combination of conditions is important because it is often possible to eliminate or reduce SCC susceptibility by modifying either the metallurgical characteristics of the metal and/or the environment. Another important aspect of SCC is the requirement for the tensile stress to be present, such as those developing from cold work, residual stresses during fabrication/machining, and externally applied functional/occlusal loads. Different surfaces of a metallic restoration (implant or crown structure) may have small pits and crevices and may be differentially exposed to different stresses consequently leading to stress corrosion cracking [36-37].The primary idea behind, titanium alloy SCC is the observation that no apparent corrosion, either uniform or localized, usually occurs before the cracking process [48-49] as a result it is difficult to translate the real oral situation into a laboratory experiment [49].

39

Over decades several models [49-54] have been proposed to explain the mechanism behind SCC in titanium alloys and broadly fall under two large categories. 1. Anodic-assisted cracking may begin where localized corrosion has occurred in the presence of a tensile stress. If corrosion is not rapid to impede the advancing crack tip, the crack will continue to advance into the metal and eventually lead to failure. Once a crack initiates, the balance among the crack tip corrosion rate, the crack tip environment, and the crack tip stress state is critical to crack propagation. 2. Hydrogen-assisted cracking (HIC) is said to occur by absorption of hydrogen near the crack tip. Hydrogen absorption leads to embrittlement of the metal ahead of the crack tip and promotes crack formation. The source of hydrogen is normally associated with anodic dissolution (that is, from the concurrent cathodic hydrogen-reduction reaction) at freshly exposed metal at the crack tip. As a result, anodic dissolution in the vicinity of the crack tip is normally required for this mechanism to operate. ASTM has formulated the standard practice for testing SCC of titanium alloys in three basic categories [49]: 1. Use of smooth and statically loaded specimens such as U-bend, C-ring, bent beam, and dead-loaded tensile specimens. 2. Use of notched and pre-cracked specimens that are statically or dynamically loaded such as cantilever beam bend specimen, compact tension specimens and doublecantilever beam specimens. 3. Use of smooth or notched tensile specimens that are dynamically loaded at relatively low strain rates. Fretting Corrosion / erosion corrosion The combination of corrosive fluid (saliva with several enzymes and food particles) and high velocity in the oral environment results in erosion-corrosion or fretting. It is responsible for most of the metal release in tissue. Conjoint action of chemical (enzymes and proteins) and mechanical wear (mastication) during function further aggravates the attack [36-37,55-56].In general during the passive environments, the hard and tenacious TiO 2 surface film over the metal surface provides a superb barrier to erosion-corrosion. For this reason titanium alloys can withstand flowing water velocity as high as 30 m/s with little or no metal loss. The ability of the oxide film to repair itself when damaged and the intrinsic hardness of titanium alloys both contribute to their excellent resistance to erosion-corrosion. Titanium alloys exhibit relatively high resistance to fluids containing suspended solids. Critical velocities for excessive metal removal depend upon the concentration, shape, size, hardness of the suspended particles, fluid impingement angle,[57] local turbulence, and titanium alloy properties. The typically low concentrations of organic material in oral cavity is of little importance but continuous exposures to local changes around the implant during function can lead to finite removal of the metal as well as the cementing

40

R. Bhola, S.M. Bhola, B. Mishra and D.L. Olson electrolyte, a potential is generated. The net result is a chemical reaction with oxidation occurring at the anode and reduction occurring at the cathode, the electronic exchange occurs through the contact and ionic exchange occurs through the electrolyte. The electrochemical cell reactions occurring at the different electrodes, depending upon the pH and aeration conditions as well the addition of oxidizers are [37]: 1) Anode (Oxidation) M Mn+ + ne2) Cathode (Reduction) 2H+ + 2e- H2 4H+ + 4e- + O2 2H2O 2H2O + 2e- H2 +2OH2H2O + 4e- + O2 4OHMn+ + e- M(n-1)+ Thus if a base metal alloy superstructure is provided over the titanium implant, the less noble metal alloy forms the anode and the more noble titanium alloy is protected as a cathode. The electronic exchange occurs through the metallic contact and ionic contact occurs through saliva, mucosa, tissue and bone fluids. Though the hydrogen evolution at the implant surface may further complicate the situation by pressure under the prosthesis and hydrogen embrittlement of the implant surface. Mechanisms of Corrosion resistance The excellent corrosion resistance of titanium and its alloys used for implants is due the formation of a thermodynamically stable, continuous, highly adherent, and protective surface oxide film. Since titanium metal itself is highly reactive and has an extremely high affinity for oxygen, this beneficial surface oxide film is formed spontaneously and instantly when fresh metal surface is exposed to air and/or moisture. In fact, a damaged oxide film can generally reheal itself instantaneously if at least traces (that is, a few parts per million) of oxygen or water are present in the environment. Within a millisecond of exposure to air, a 10 oxide layer will be formed on the cut surface of the exposed pure Ti which will grow to about 100 thick within a minute. The nature, composition, and thickness of the protective surface oxides that form on titanium alloys depend on environmental conditions. In most oral environments the oxide is typically TiO2 but may consist of mixtures of other titanium oxides as well including TiO2, Ti2O3, and TiO[61]. High-temperature oxidation promotes the formation of denser, more chemically resistant form of TiO 2 known as rutile, whereas lower temperatures often generate a less crystalline and protective form of TiO 2, anatase, or a mixture of rutile and anatase [61]. In dilute reducing acids, a 20 to 100 multiplex film consisting of a hydrated titanium sesquioxide (Ti 2O3) inner layer and a TiO2 outer layer has been shown to form [62]. Increasing redox potential favors TiO 2 formation, and

material between the implant and superstructure there by not only promoting erosion corrosion but crevice and galvanic corrosion as well. Intergranular corrosion/cracking For intergranular cracking to occur reactive impurities may segregate, or passivating elements may deplete at grain boundaries. This results in grain boundaries as a preferential region for corrosion owing to its high susceptibility and so the grains might fall out of the surface leading to material cracking [36-37]. This type of corrosion is not usually observed in titanium oral implants. Galvanic corrosion of dental implants The most common form of corrosion occurring in titanium implants is the galvanic corrosion or dissimilar corrosion. Titanium has been chosen as the material of choice for several trans-osseous and end-osseous implantations. Long term studies and clinical observations have established the fact that titanium is noble (due to the presence of adherent oxide) and does not corrode in human tissues however galvanic coupling of implant to several other metallic restorations may induce one of the several forms of corrosion. Thus coupling remains a great concern for the metallic superstructures covering the implant body. Gold alloys have widely been used as a super structure owing to their excellent biocompatibility, corrosion resistance and mechanical properties. Owing to higher cost of the precious metal alloys (noble alloys)used in prosthodontics, it has led to the development of cost effective semi-precious metallic alternatives(non-noble alloys) [58-59] like gold-palladium, nickel-chromium, cobalt-chromium, nickel-titanium and several other titanium alloys. Galvanic coupling occurs when two dissimilar metals are placed in direct contact within the oral cavity (adjacent or contralateral) or within the tissues. The complex corrosion process occurring involves the electrochemical reactions occurring at the dissimilar metals interface in presence of electrolyte (saliva or oral fluids or body fluids) resulting in the flow of electric current between them [37,60]. An invivo (in the living tissue) electrochemical cell is formed and galvanic current causes the corrosion of active metal and the noble metal is protected. The current also passes through the cellular junctions and tissues (desmosomes, hemi-desmosomes and cellular attachments) thereby activating the proprioceptors causing pain. As already discussed in the section on hydrogen damage above attention should be given to possible excessive hydrogen uptake by titanium implant when it is galvanically coupled to active metal superstructures. This situation is of great concern in titanium alloys when temperatures exceed 80C in aqueous electrolytes during implant processing especially when hydrogen recombination poisons, such as sulfides, arsenides and cyanides are present. Phenomena of galvanic corrosion When two dissimilar metals with different electrode potentials come in contact in the presence of a corrosive

Corrosion in Titanium Dental Implants/Prostheses - A Review

41

The Pourbaix (potential-pH) diagram [64] for the titaniumwater system at 25C has been shown in Figure 5 and depicts the broad range around which the passive TiO 2 film is stable, based on thermodynamic (free-energy) considerations. Oxide stability over the full pH scale is indicated over a wide range of highly oxidizing to mildly reducing potentials. Oxide film breaks and the corrosion of titanium implant surface occurs under reducing acidic conditions. Under strongly reducing (cathodic) conditions, titanium hydride formation is predicted. This range of oxide film stability and passivation is relatively insensitive to the presence of chlorides, explaining the high innate resistance of titanium to aqueous chloride environments [65]. The nature of the oxide film on titanium alloys remains unaltered in the presence of minor alloying constituents. Small additions (<2 to 3%) of most commercially used alloying elements or variations in interstitial impurities have little effect on the corrosion resistance of titanium in normally passive environments. Increasing the alloying iron and sulfur content may increases corrosion rates exceeding ~0.10 mm/yr (3.9 mils/yr) [66]. Thus minor variations in alloy chemistry may be of concern only under conditions in which the passivity of titanium is borderline or when the metal is actively corroding. Clinical significance of corrosion Although titanium alloys have better corrosion properties compared to Co-Cr and stainless steel (other implant materials) their corrosion leads to dissolution of titanium and other alloying elements like aluminum, vanadium, niobium, molybdenum etc [67] causing localized to generalized host response as illustrated in Figure 6. The leached ions may induce potentially osteolytic cytokines into tissues leading to implant loosening [67] and may even cause severe allergic reactions or hypersensitivity [67]. Incidence of tumors and malignancies has also been reported in the literature but are few. Fracture of Dental Implant Fracture of dental implant/prostheses is a very rare phenomenon more often associated with mechanical function and previously use of screw preload systems to

Figure 5: Pourbaix (potential-pH) diagram for the titaniumwater system at 25 C

increasing exposure temperature and/or time motivate conversion to the more stable crystalline form of TiO 2. These naturally formed films are typically less than 10 nm thick[63] and are not visible to the human eye. TiO 2 oxide is highly resistant chemically and is attacked by very few chemicals like hot concentrated HCl, H2SO4, NaOH, and HF. This thin surface oxide is also a highly effective barrier to hydrogen penetration of the alloy, as is discussed in a later section of this article. Furthermore, the TiO 2 film being an n -type semiconductor exhibits increasing electronic conductivity with increasing temperature. As a cathode, titanium readily passes current and permits electrochemical reduction of ions in an aqueous electrolyte. On the other hand, very high resistance to anodic current flow (anodic polarization) across the passive oxide film can be seen in most aqueous solutions. Because the passivity of titanium occurs by the formation of a stable oxide film the corrosion behavior of implant metal interface can be understood by recognizing the conditions under which this oxide is thermodynamically stable.

Figure 6: Host tissue response to an implant material

42

R. Bhola, S.M. Bhola, B. Mishra and D.L. Olson Several studies have further shown that the cellular uptake of hexavalent chromium is many folds greater than the trivalent chromium ion and its increased uptake causes reduction of the alkaline phosphatase activity of the osteoblastic cells [77-79]. In-Vitro Studies Several clinical cases of galvanic corrosion have been reported in the literature. Ravnholt et al.[ 80] studied the galvanic corrosion of titanium implants in contact with amalgam and cast metal alloys. As a continuation of their work Ravnholt and Jensen et al. [ 80-81] reported no changes in current or pH when titanium superstructures were in metallic contact with gold, cobalt-chromium, silver palladium and composite carbon alloys. Geis-Gerstorfer et al.[31] stated the importance of galvanic corrosion in implant system in two ways; first the possibility of biological effects like discoloration, pigmentation, localized reaction occurring from alloy components and secondly the osteolysis occurring as a mediation to phagocyte cascade activation. Recalru and Meyer [60] studied the potential alloys that could be used as implant superstructure and concluded that an alloy that is potentially usable for superstructures in galvanic coupling to titanium must fill the following requisites 1. During coupling the titanium should have weak anodic polarization 2. The current generated during galvanism should be weak 3. The crevice potential must be much higher in comparison to common potential Recently, Venugopalan and Lucas [82] defined the parameters for an acceptable couple combination to fulfill the following 1. The difference in the OCP (open circuit potential) and the corrosion current density (coupled materials) should be as small as possible 2. The breakdown potential of the anodic component should be significantly noble than the coupled corrosion density of the combination

clamp flat to flat abutment implant junctions. They can have serious clinical complications. Corrosion can severely limit the fatigue strength and ultimate tensile strength of the material leading to its mechanical failure. As reported by Green[68], the end-osseous implant superstructures, leached metal ions into the surrounding tissues as an event to corrosion, leading to fatigue fracture, following four years of functional loading into the oral cavity. Yokoyama et al.[ 69] on the other hand investigated the delayed fracture of titanium implant into the oral environment owing to hydrogen embritlement and environmentally induced cracking (EIC). Bone loss and osteolysis Olmedo, Fernandez and Guglidomotti [70] have speculated the corrosion induced ionic leaching to be responsible for peri-implantitis and treatment failure. As already explained in Figure 6. the particles that are leached as a result of corrosion process are phagocytosized by macrophages (under the influence of host response) and release mediators of inflammation in the form of cytokines (host defense) [67] through the inflammatory cascade which inhibit the production of osteoblasts causing increased activity of the bone destroying cells leading to peripheral osteolysis and loosening of the implant. Localized pain and inflammation (swelling) Watterhehn et al .[71] have studied the effect of ionic release as a result of corrosion, causing localized pain and swelling with or without infection in the region of implant insertion. Figure 7 shows the localized inflammatory response to a freshly inserted implant in the right lateral incisor region (12 region). [71] Cytotoxic tissue response Corrosion release of the several substitutional alloying elements from various titanium alloys used in dentistry have been widely known in literature. Watterhehn et al. [71-76] have reported these metal ion release to be associated with carcinogenic and mutagenic activity of the oral cavity as shown in Figure 8.

Figure 7: Localized inflammatory response in a 17 year old adult as a result of ionic leaching from an implant in 12 region.

Figure 8: Malignancy formation as a response to adjacent implant leach (corrosion)

Corrosion in Titanium Dental Implants/Prostheses - A Review 3. The repassivation potential of the anodic component of the couple should be as similar as possible with minimal hysteresis Johanson et al. [83] studied the effect of fluoride on corrosion behavior of machined and cast titanium with varying surface treatments and exposed area in contact with conventional and high copper amalgams. He observed higher current density for conventional amalgam in contact with titanium and concluded that surface preparations and anionic inclusions affect corrosion behavior of titanium implants. In-Vivo Studies Despite the use of titanium and its alloys as implants in human body increasing evidence is found that titanium and various substitutional alloying elements leach into the crevicular space around the implant [11-12,84-86]. Titanium like the other active-passive metals is covered by a protective air formed oxide of TiO2 which is adherent and stable. Although the oxide coating is thermodynamically stable, any form of chemical, electrochemical and mechanical trauma can lead to release of alloying elements through passive dissolution process. Ferguson and coworkers [87] first documented the elevated titanium levels locally around the implant periphery. Bumgardner et al. [ 88] reported the increased release of metal ions through a gallium titanium galvanic couple causing increased implant and tissue deterioration.Cortada et al.[89] used the ICP-AES (inductively coupled plasma- atomic emission spectroscopy ) technique to confirm the release of metallic ions during titanium implant galvanism. Nakagawa et al. [90] studied the effect of fluoridation during prophylactic therapy on the corrosion behavior of an implant and reported higher corrosion current density with increasing anionic concentrations. Extended work has been done by Bhola et al. [ 91] to study the effect of Knutson fluoridation on the corrosion behavior of titaniumniobium dental implant alloys in normal saline. Kirkpatrick et al. [ 92] highlighted the patho-physiology and patho-mechanics behind impaired wound and tissue healing under metallic ions release during corrosion. Dental corrosion Mitigation With the development of corrosion science and advances in material engineering and design several strategies have been employed today to prevent in-use (within the oral cavity) and out-use (outside oral cavity) corrosion of titanium implants and prostheses. 1. Alloying titanium with noble metals (platinum-group metals) which facilitates cathodic depolarization by catalyzing the hydrogen reduction step in acid solutions used during active processing of dental implants in the industry for out-use prevention. 2. Alloying titanium with more thermodynamically stable acid-resistant elements such as molybdenum, zirconium, tantalum, niobium used during processing of dental

43

implants, also makes the implant surface more biocompatible and cell adherent for in use prevention. 3. Addition or presence of soluble oxidizing ions or compounds (cathodic depolarizers) in the oral media (e.g., Fe3+, O2, Cl2, NO3-), can be used during implant insertion and placement using NS, RL, PBS etc, for in-use prevention. 4. Noble metal surface treatments (platinum-group metal coatings or enriched/modified surfaces) can be used during implant surface processing for in-use corrosion prevention. 5. Anodic protection via impressed positive potentials from an external power source or from direct galvanic coupling with a noble metal (platinum-group metal) can be used during processing for out-use prevention. 6. Thermal oxidation or nitriding of titanium surfaces can be used during implant manufacturing and processing to facilitate in-use prevention. To summarize several techniques do exist to effectively reduce the corrosion in dental implants, but we can mainly rely on material selection and coating the surface for an effective in-use clinical performance into a patients oral cavity.

Conclusion
The metallic titanium dental implants/prostheses used in dentistry today derive their biocompatibility from the alloying elements responsible for the formation of a continuous stable TiO2 passive film on its surface. There is a significantly small release of alloying ions even under the ideal conditions of passivity and with no damage to the implant surface. Corrosion of these implants may occur when the oral conditions are unfavorable as under mechanical trauma to the implant surface (during placement, subject induced, and trauma to assault) or the use of inappropriate metal combination as auxiliary prostheses (galvanism). The potential adverse effects of metal ion release into living tissues can be proposed based on information from literature and various clinical, preclinical and animal trial studies in-vivo and in-vitro. The results of in-vivo and invitro testing do not necessarily take into account all of the protection mechanisms and physiological host response characteristics of the actual implant environment in oral cavity. It is clear that corrosion is bound to occur and its consequences can be quite severe. Our current research standards, regulations for biocompatibility testing, material and design understanding and composition of metals and its alloys suitable to a special application have drastically reduced the incidence of implant failures in oral physiological environment under corrosion and its adverse events.

44

R. Bhola, S.M. Bhola, B. Mishra and D.L. Olson

References
1. J. Black, G. Hastings, ibid, p 137 2. M.G. Fontana, Perspectives on corrosion of materials, Metallurgical transactions Vol.1. The 1970 Campbell Memorial Lecture,ASM,p 3251-3266, 1970 3. S.G. Steinemann, Corrosion of titanium and titanium alloys for surgical implants. In: Lutjering G, Zwicker U, Bunk W(eds), Proceedings of the 5th international conference on Titanium(Munich FRG, 10-14 Sep 1984). Deutsche Gesellschaft fur Metallkunde, p 1373-1379, 1987 4. Annual book of ASTM standards 2001, Vol 13-01, ASTM, West Conshohcken PA 2001 5. Metals and Alloys in Unified Numbering System 9 th edition ASTM, West Conshohocken PA 2001 6. Stahlschlussel (Key to Steels) Verlag Stahlschlussel Wegst GmbH, D-71672, Marbach, Germany 7. J. Lemons, R. Venugopalan, L. Lucas, Corrosion and Biodegradation; A. Von Recum, ed., Taylor Francis Inc: New York, 155, 1999 8. X. Cheng, S.G. Roscoe, Corrosion behavior of titanium in the presence of calcium phosphate and serum proteins, Biomaterials, 26, p 7350-7356, 2005 9. R. Adell, U. Lekholm, B. Rockler,P.I. Branemark, A 15 year study of osteointegrated implants in treatment of edentulous jaw, Int J Oral Surg, 10, p 387-416, 1981 10. P. Kovacs, J.A. Davidson, Chemical and electrochemical aspects of the biocompatibility of titanium and its alloys. In: Brown SA, Lemons JE, editors. Medical application of titanium and its alloys: The material and biological issue, ASTM STP 1272. West Conshohocken, PA: ASTM, p163-178, 1996 11. H. Zitter, H. Plenk, The electrochemical behavior of metallic implant materials as an indicator of their biocompatibility, J Biomed Mater Res, 21, p 881-896, 1987 12. K. Nielsen, Br Corrosion J, 22(4), p 272, 1987 13. B. Bozinni, P. Carlino, L. DUrzo, V. Pepe, C. Mele, F. Venturo, An electrochemical impedance investigationof the behavior of anodically oxidized titanium in human plasma and cognate fluids, relavant to dental applications, J Mater Sci: Mater Med, 19, p 3443-3453, 2008 14. S. Kothari, P.V. Halton, J. Douglas, Mater Sci, 6, p 695, 1995 15. T. Sawase, K. Hai, K. Yoshida, K. Baba, R. Hatada, M. Atsuta, J Dent, 26, p 119, 1998 16. Y.T. Sul, C.B. Johansson, S. Petronis, A. Kroger, A. Weenerger, T. Albrektsson, Biomaterials, 23, p 1809, 2002 17. D.W. Hoppner, V. Chandrasekarn, Fretting in orthopaedic implants: a review, Wear, 173, p 189-197, 1994 18. E.W. Collings, The physical metallurgy of titanium alloys, ASM Series in Metal Processing. H.L. Gegel, editor. Cleveland, Metals Park, OH: American Society for Metals, 1984 19. J.J. Polmear, Titanium alloys. In: Light alloys, Chapter 6. London, UK: Edward Arnold Publ, 1981 20. P.J. Bania, Beta titanium alloys and their role in the titanium industry In: D. Eylon, R.R. Boyer, D.A. Koss, editors. Titanium Alloys in the 1990s. Warrendale: The Mineral, Metals & Materials Society, p 3-14, 1993 21. R.W. Schutz, An overview of beta titanium alloy environmental behavior. In: D. Eylon, R.R. Boyer, D.A. Koss, editors. Beta Titanium Alloys in the 1990s. Warrendale, PA: The Mineral, Metals & Materials Society, p 75-91, 1993 22. R.R. Boyer, J.A. Hall, Microstructure-property relationships in titanium alloys (critical review). In: F.H. Froes, I. Caplan, editors. Titanium92 Science and Technology, Warrendale, PA: The Mineral, Metals & Materials Society, p 77-88, 1993 23. A. Vassel, Microstructural instabilities in beta titanium alloys. In: D. Eylon, R.R. Boyer, D.A. Koss, editors. Beta Titanium Alloys in the 1990s. The Mineral, Metals & Materials Society, p 173-185, 1993 24. J.J. Jacobs, J.L. Gilbert, R.M. Urbani, Corrosion of metal orthopaedic implants, J Bone Joint Surg, 80, p 1-2, 1998 25. S.W. Dean, Jr., Electrochemical Methods of Corrosion Testing, Electrochemical Techniques for Corrosion, National Association of Corrosion Engineers, p 5260, 1977 26. E.L. Liening, Electrochemical Corrosion Testing Techniques, Process Industries Corrosion, National Association of Corrosion Engineers, p 85122, 1986 27. L. Szklarska-Smialowska and M. Janik-Czachor, Corros. Sci, 11, p 901914, 1971 28. R. Baboian, Ed., Corrosion Tests and Standards: Application and Interpretation, 2nd ed., ASTM International, p 598612, 2005 29. E.L. Liening, Electrochemical Corrosion Testing Techniques, Process Industries Corrosion, National Association of Corrosion Engineers, p 85122, 1986 30. R.W. Schutz and J.S. Grauman, Compositional Effects on Titanium Alloy Repassivation Potential in Chloride Media, Advances in Localized Corrosion, Proceedings of the Second International Conference on Localized Corrosion, H.S. Isaacs, U. Bertocci, J. Kruger, and S. Smialowska, Ed., National Association of Corrosion Engineers, p 335337, 1990 31. G.J. Geis, J.G. Weber, K.H. Sauer, In vitro substance loss due to galvanic corrosion in titanium implant/ NiCr supraconstruction systems, Int J Oral Maxillofac Imp, 9, p 449-454, 1994 32. R.W. Schutz, Understanding and Preventing Crevice Corrosion of Titanium Alloys Part I, Mater. Perform., p 5762,1992 33. R.W. Schutz, Understanding and Preventing Crevice Corrosion of Titanium AlloysPart II, Mater. Perform., p 5456, 1992 34. N. Sato, Toward a more fundamental understanding of corrosion processes, Corrosion, 45 , p 354-368,1989 35. R. Baboian, Ed., Corrosion Tests and Standards: Application and Interpretation, 2nd ed., ASTM International, p 598612, 2005 36. D.O. Sprowls, Metals handbook vol 13, corrosion 9th edition ASM International, metals park, OH, p 222-225, 1987 37. D.A. Jones, Forms of corrosion recognition and prevention, Dillion CP .ed., NACE Houston, p 19-43, 1982 38. L.C. Covington, Factors Affecting the Hydrogen Embrittlement of Titanium, Paper59, Corrosion/75, National Association of Corrosion Engineers, April 1975 39. L.C. Covington, Corrosion, Vol 35 (No. 8), p 378382, 1979 40. R. Bourcier, D. Koss, Acta Metall., 32(11), p 20912099, 1984 41. G.A. Lenning et al., Effect of Hydrogen on Alpha Titanium Alloys, Trans. AIME, p 1235, 1956 42. C.M. Craighead et al., Hydrogen Embrittlement of Beta-Stabilized Titanium Alloys, Trans. AIME, p 923, 1956 43. R.W. Schutz, Environmental Behavior of Beta Titanium Alloys, JOM, p 2429, 1994 44. R.W. Schutz, Titanium, Process Industries CorrosionThe Theory and Practice, National Association of Corrosion Engineers, p 503, 1986 45. H. Satoh, T. Fukuzuka, K. Shimogori, and H. Tanabe, Hydrogen Pickup by Titanium Held Cathodic in Seawater, Second International Congress on Hydrogen in Metals (Paris), June 1977

Corrosion in Titanium Dental Implants/Prostheses - A Review

45

46. S. Sato, K. Nagata, and M. Nagayama, Experiences of Welded Titanium Condenser Tubes in Japan, Technical Research Laboratory, Sumitomo Light Metal Industries Ltd. 47. T. Fukuzuka, K. Shimogori, H. Satoh, and F. Kamikubo, Corrosion Problems and Countermeasures in MSF Desalination Plant Using Titanium Tube, Kobe Steel Ltd., 1985 48. I.R. Lane, Jr., J.L. Cavallaro, and A.G.S. Morton, Stress Corrosion Cracking of Titanium, STP 397, American Society for Testing and Materials, 1966 49. R.W. Schutz, Stress-Corrosion Cracking of Titanium Alloys, Stress-Corrosion Cracking, R.H. Jones, Ed., ASM International, p 265297, 1992 50. T.R. Beck, Electrochemical Aspects of Titanium Stress-Corrosion Cracking, Proceedings of ConferenceFundamental Aspects of Stress-Corrosion Cracking, National Association of Corrosion Engineers, p 605, 1969 51. T.R. Beck and E.A. Grens, An Electrochemical Mass-Transport-Kinetic Model for Stress-Corrosion Cracking of Titanium, J. Electrochem. Soc., 116(2), p 117, 1969 52. D.T. Powell and J.C. Scully, Stress-Corrosion Cracking of Alpha Titanium Alloys at Room Temperature, Corrosion, 24(6), p 151, 1968 53. G. Sanderson, D.T. Powell, and J.C. Scully, The Stress-Corrosion Cracking of Ti Alloys in Aqueous Chloride Solutions at Room Temperature, Corros. Sci., 8, p 473, 1968 54. R.J.H. Wanhill, Aqueous Stress-Corrosion in Titanium Alloys, Br. Corros. J., 10, p 69, 1975 55. P.D.Bianco, P. Ducheyne, J.M. Cuckler, Biomaterials, p 17, 1996 56. P.D. Bianco, P. Ducheyne, J.M.Cuckler, In Proc. 1994 symp, on Medical Applications of Titanium and its alloys: The material and biological issues, 1272, ASTM, Conshohocken, PA, p 346, 1996 57. N. Gat and W. Tabakoff, Effects of Temperature on the Behavior of Metals Under Erosion by Particulate Matter, J. Test. Eval., 8(4), p 177186, 1980 58. K.F. Leinfelder, j.E. Lemons, Clinical restorative materials and techniques, Ler and Febiger, Philadelphia, p 139159, 1998 59. L.C. Lucas, J.E. Lemons, Biodegradation of restorative metallic systems, Adv Dent Res, 6, p 32-37, 1992 60. L. Recalru, J.M. Meyer, Study of corrosion between a titanium implant and dental alloys, J Dent, 22, p 159-168, 1994 61. N.D. Tomashov and P.M. Altovskii, Corrosion and Protection of Titanium, Government Scientific-Technical Publication of MachineBuilding Literature (Russian translation), 1963 62. K. Shimogori et al., Analysis of Passive Film on Titanium Formed in Dilute HCl Solutions at Elevated Temperatures, Titanium and Titanium Alloys: Scientific and Technological Aspects, Vol 2, Plenum Publishing Corp., p 881890, 1982 63. V.V. Andreeva, Behavior and Nature of Thin Oxide Films on Some Metals in Gaseous Media and in Electrolyte Solutions, Corrosion, 20, p 3546, 1964 64. M. Pourbaix, Atlas of Electrochemical Equilibria in Aqueous Solutions, National Association of Corrosion Engineers, p 217, 1974 65. R.W. Schutz, Utilizing Titanium to Successfully Handle Chloride Process Environments, Can. Min. Metall. Bull., 95(1065), p 84 88, 2002 66. L.C. Covington and R.W. Schutz, Effects of Iron on the Corrosion Resistance of Titanium, Industrial Applications of Titanium and Zirconium, STP 728, American Society for Testing and Materials, p 163180, 1981 67. S.D. Rogers, D.W. Howie, S.E. Graves, M.J. Pearcy, D.R. Haynes, In vitro human monocyte response to wear particles of titanium alloys containing niobium and vanadium, J Bone Joint Surg, 79(B), p 311-315, 1997 68. N.T. Green, Fracture of dental implants: Literature review and report of a case, Imp Dent, 137, p 143, 2002 69. K. Yokoyama, T. Ichikawa, H. Murakami, Y. Miyamoto, K. Asaoka, Fracture mechanics of retrived titanium screw thread in dental implant, Biomaterials, 2002, 23, p 2459-2465 70. D. Olmedo, M.M. Farnendez, M.B. Guglidmotti, R.L. Cabrini, Macrophages related to dental implant failure, Imp Dent, 2003, 12, p 75-80 71. K.E. Watterhahn, B. Demple, M.M. Kulesz, E.S. Copeland, Carcinogenesis- a chemical pathology study section workshop, Workshop report from the division of research grants, NIH. Cancer Res 52, p 4058-4063, 1992 72. A. Arillo, F. Melodia, R. Frache, Reduction of hexavalent chromium by mitochondria: methodological implications and possible mechanisms, Ecotoxicol Environ Safety, 14, p 164-177, 1987 73. A. Ryberg, Mechanism of chromium toxicity in mitochondria, Chem Biol Interact, 75, p 141-151, 1990 74. M.D. Cohen, B. Kargacin, C.B. Klein, M. Costa, Mechanism of chromium carcinogenicity and toxicity, Crit Rev Toxicol, 23, p 255281, 1993 75. D.M. Stearns, K.D. Curtne, P.H. Giangrande, L.S. Phieffer, K.E. Wetterhahn, Chromium (VI) reduction by ascorbate: role or reactive internediates in DNA damagein vitro, Environ Health Persp, 102, p 21-25, 1994 76. E.T. Snow, L. Xu, Chromium (III) bound to DNA Templates promotes increased plumerase processivity and decreased fidelity during replication vitro, Biochemistry, 30, p 238-245, 1991 77. K. Merritt, C.D. Fedele, S.A. Brown, Chromium 6 or 3 release during corrosion and in vivo distribution, Biomater Tissue Interf, p 49-53, 1992 78. K. Merritt, S.A. Brown, Release of hexavalent chromium from corrosion of stainless steel and cobalt-chromium alloys, J Biomed Mater Res, 29, p 627-633, 1995 79. P. Debetto, S. Luciani, Toxic effect of chromium on cellular metabolism, Sci Total Environ, 71, p 365-377, 1988 80. G.Ravnholt, Corrosion current, pH rise around titanium implants coupled to dental alloys, Scand J Dent Res, 96, p 466-472, 1998 81. G. Ravnholt, J. Jensen, Corrosion investigation of two material for implant supraconstructions coupled to titanium implant, Scand J Dent Res, 99, p 181-186, 1991 82. R. Venugopalan, C. Linda, L. Lucas, Evaluation of restorative and implant alloys galvanically coupled to titanium, Dent Mater, 14, p 165-172, 1998 83. B.I. Johansson, B. Bergman, Corrosion of titanium and amalgam couples, Effect of fluoride, area size, surface preparation and fabrication procedures, Dent Mater, 1, p, 41-46, 1995 84. J. Black, E.C. Maitin, H.Gelman, D.M. Morris, Serum concentrations of chromium ,cobalt, nickel after total hip replacement: A six month study, Biomaterials, 4, p 160-164, 1983

46

R. Bhola, S.M. Bhola, B. Mishra and D.L. Olson

85. S. Lugowski, D.C. Smith, J.C. Vanloon, The determination of Al,Cr, Co, Fe and Ni in whole blood by electrothermal atomic adsorption spectrophotometery, J Biomed Mater Res, 21, p 657-674, 1987 86. F. Contu, B. Elsener, H. Bohni, Characterisation of implant material in FBS and sodium sulfate by electrochemical impedance spectroscopy, J Biomed Mater Res, 62(3), p 412-421, 2002 87. A.B. Ferguson, P.G. Liang, E.S. Hodge, The ionization of metal implants in living tissues, J Bone Jt Surg, 42, p 76-89, 1960 88. J.D. Bumgardner, B.I. Johansson, Effects of titanium dental restorative alloy galvanic couples on culture cells, J Biomed Mater Res, 43, p 184-191, 1998 89. M. Cortada, L. Giner, S. Costa, F.J. Gil, D. Rodriguez, P.A. Planell, Metallic ion release in artificial saliva of titanium oral implants coupled with different metal superstructures, Biomed Mater Eng, 7, p 213-220, 1997 90. M. Nakagawa, S. Matsuya, T. Shirasishi, M. Ohta, Effect of fluoride concentration and pH on corrosion behavior of titanium for dental use, J Dent Res, 78, p 1568-1572, 1997 91. R. Bhola, S.M. Bhola, B. Mishra, D.L. Olson, Knutson fluoridation effect on corrosion properties of Ti36 in normal saline solution, J Mat Eng & Phy, in press, 2009 92. C.J. Kirkpatric, S. Barta, T. Gerdes, V. Krump-Konvalinhova, K. Peters, Pathomechanisms of impaired wound healing by metallic corrosion products, Mund Kiefer Gesichtschir, 6, p 183-190, 2002