Instrument Test Specifications: 0.25 Ms Sample Rate
Instrument Test Specifications: 0.25 Ms Sample Rate
FDU-428
0.25 ms Sample Rate
Max. Distortion: -103 dB. Min. Common-Mode Rejection: 100 dB. Max. Gain error: 3.0%. Max. Phase error: 30 s. Max. Noise (0 dB gain, 1600 mV scale): 18.0 V. Max. Noise (12 dB gain, 400 mV scale): 4.5 V. Min Crosstalk rejection: 110 dB.
296
1 ms Sample Rate
Max. Distortion: -103 dB. Min. Common-Mode Rejection: 100 dB. Max. Gain error: 1.0%. Max. Phase error: 20 s. Max. Noise (0 dB gain, 1600 mV scale): 1.4 V. Max. Noise (12 dB gain, 400 mV scale): 0.35 V. Min Crosstalk rejection: 110 dB.
2 ms Sample Rate
Max. Distortion: -103 dB. Min. Common-Mode Rejection: 100 dB. Max. Gain error: 1.0%. Max. Phase error: 20 s. Max. Noise (0 dB gain, 1600 mV scale): 1.0 V. Max. Noise (12 dB gain, 400 mV scale): 0.25 V. Min Crosstalk rejection: 110 dB.
4 ms Sample Rate
Max. Distortion: -103 dB. Min. Common-Mode Rejection: 100 dB. Max. Gain error: 1.0%. Max. Phase error: 20 s. Max. Noise (0 dB gain, 1600 mV scale): 0.7 V. Max. Noise (12 dB gain, 400 mV scale): 0.18 V. Min Crosstalk rejection: 110 dB.
297