DICOM
Conformance Statement
105952
Revision 0
Copyright 2000, 2001, 2002, 2003, 2004 by GE Medical Systems Kretztechnik GmbH & Co OHG
GE Medical Systems
Kretz Ultrasound
TIEFENBACH 15 A-4871 ZIPF AUSTRIA Telefon: Fax: E-Mail: Internet: +43 7682 / 3800 0 +43 7682 / 3800 47 info@kretztechnik.com http://www.kretztechnik.com
Table of Contents
0 Introduction ................................................................................................................................................................ 5 0.1 DICOM Background ........................................................................................................................................... 5 1 Implementation Model ............................................................................................................................................... 6 1.1 Application Data Flow Diagram.......................................................................................................................... 6 1.2 Functional Definitions of AEs............................................................................................................................ 7 1.3 Sequencing of Real-world Activities ................................................................................................................... 7 2 AE Specifications ....................................................................................................................................................... 8 2.1 Printing AE - Specification.................................................................................................................................. 8 Association Establishment Policies ....................................................................................................................... 8 2.1.1.1 General 2.1.1.2 Asynchronous Nature 2.1.1.3 Implementation Identifying Information 2.1.1.4 Association Initiation by: Print in SonoView 8 8 8 8
Association Initiation by Real-world Activity....................................................................................................... 8 2.2 Storing AE - Specification................................................................................................................................. 14 Association Establishment Policies ..................................................................................................................... 14 2.2.1.1 General 2.2.1.2 Number of Associations 2.2.1.3 Asynchronous Nature 2.2.1.4 Implementation Identifying Information 2.2.1.5 Association Initiation by: Send in SonoView or by Direct Send 14 14 14 14 14
Association Initiation by Real-world Activity..................................................................................................... 14 2.3 Worklist AE - Specification............................................................................................................................... 21 Association Establishment Policies ..................................................................................................................... 21 2.3.1.1 General 2.3.1.2 Number of Associations 2.3.1.3 Asynchronous Nature 2.3.1.4 Implementation Identifying Information 2.3.1.5 Association Initiation by: Search 21 21 21 22 22
Association Initiation by Real-world Activity..................................................................................................... 22 3 Communication Profiles ........................................................................................................................................... 23 3.1 TCP/IP Stack Supported .................................................................................................................................... 23 Physical Media Supported ................................................................................................................................... 23 4 Extensions/Specializations/Privatizations................................................................................................................. 23 4.1 Standard Extended/Specialized/Private SOPs ................................................................................................... 23 4.2 Private Transfer Syntaxes .................................................................................................................................. 24 5 Configuration............................................................................................................................................................ 24 5.1 AE Title/Presentation Address Mapping ........................................................................................................... 24 5.2 Configurable Parameters.................................................................................................................................... 24 6 Support of Extended Character Sets ......................................................................................................................... 24 Voluson 730Expert/Pro DICOM Conformance Statement 105952 Rev. 0
0 Introduction
This document describes the Kretztechnik Voluson 730 Expert/Pro Ultrasound Systems conformance to the ACRNEMA DICOM (Digital Imaging and Communications in Medicine) standard and satisfies the DICOM requirement for a vendor conformance specification. The Voluson 730 Expert/Pro system is an ultrasound imaging device. The SonoView option of the Voluson 730 Expert/Pro system provides a means to select images and send them via DICOM to storage servers and printers.
1 Implementation Model
The Voluson 730 SonoView feature incorporates the DICOM 3.0 standard for networked image printing and image store functions. Images are transferred from the Voluson 730 ultrasound system using standard network connections to be processed on a centralized printer or stored on a DICOM compliant file server.
Get Worklist
Voluson 730 AE
Worklist Server
Print Image
Print Server
2 AE Specifications
2.1 Printing AE - Specification
The Printing AE provides conformance to the following DICOM SOP Classes as an SCU: SOP Class Name Basic Grayscale Print Management Meta SOP Class Basic Color Print Management Meta SOP Class Basic Film Session SOP Class Basic Film Box SOP Class Basic Gray Image Box SOP Class Basic Color Image Box SOP Class Printer SOP Class SOP Class UID 1.2.840.10008.5.1.1.9 1.2.840.10008.5.1.1.18 1.2.840.10008.5.1.1.1 1.2.840.10008.5.1.1.2 1.2.840.10008.5.1.1.4 1.2.840.10008.5.1.1.4.1 1.2.840.10008.5.1.1.16 Conformance Level Standard Standard Standard Standard Standard Standard Standard
2.1.1.1 General
Maximum PDU size offered: 28,672 bytes Minimum PDU size accepted: 1,024 bytes Number of Associations The maximum number of simultaneous associations for the Printing AE is 2.
Printing AE Proposed Presentation Contexts to a Gray Print Server Presentation Context Table Abstract Syntax Transfer Syntax Name UID Name List UID List Basic Gray Print 1.2.840.10008.5.1.1.9 DICOM Implicit 1.2.840.10008.1.2 Managment VR Little Endian Meta SOP Class Transfer Syntax Role SCU Extended Negotiation None
2.1.1.4.1.2 SOP Specific Conformance to Basic Gray Print Management Meta SOP Class
The Printing AE provides Standard Conformance to the Basic Gray Print Management Meta SOP Class as an SCU. This implies standard conformance for the Basic Film Session SOP Class, Basic Film Box SOP Class, Basic Grayscale Image Box SOP Class, Printer SOP Class. Each of these SOP classes is described in the paragraphs to follow.
Supported SOP Class Elements Name Image Display Format Usage M Range Standard \1,1 Standard \2,2 Standard \3.3 Standard \3,5 Standard \4,5 Standard \1,2 Standard \2,3 Standard \3,4 Standard \4,4 Standard \4,6 Description Range may be further restricted by printer.
Referenced Film Session Sequence Referenced SOP Class UID Referenced SOP Instance UID Film Orientation Film Size ID
Used Film Session SOP Class UID Referenced Film Session SOP Range may be further restricted by printer. Range may be further restricted by printer.
Magnification Type Max Density Configuration Information Annotation Display Format Id Smoothing Type Border Density Empty Image Density Min Density Trim
U U U U U U U U U
2.1.1.4.1.2.3 SOP Specific Conformance to Basic Grayscale Image Box SOP Class
Supported DIMSE Services Name N-Set Usage M Description An image box instance is created by the SCP for each potential image of the film box. Only the instances that will actually contain images will be updated with the N_SET message.
Supported SOP Class Elements Voluson 730Expert/Pro DICOM Conformance Statement 105952 Rev. 0
10
Name Image Position Pre-formatted Grayscale Image Sequence Samples/pixel Photometric Interpretation Rows Columns Pixel Aspect Ratio Bits Allocated Bits Stored High bit Pixel Representation Pixel Data Polarity Referenced Overlay Sequence >SOP Class UID >SOP Instance UID Magnification Type Smoothing Type Requested Image Size
Usage M M M M M M M M M M M M U U U U U U U
Description
Used 0 = Black, 255 = White Pixels Pixels Not used 8 bits per sample Used Bit 7 is MSB Unsigned pixel values Gray pixel data Not used Not used Not used Not used Used Not used Not used
11
Supported SOP Class Elements Note: These attributes are not set by this device. The attribute description here indicates which attributes are used by this device when they are returned by the printer. Name Printer Status Print Status Info Printer Name Manufacturer Model Name Serial Number Software Version Calibration Date Calibration Time Usage U U U U U U U U U Range NORMAL WARNING FAILURE Description Warning and Failure are reported to user. Reported to user. Ignored Ignored Not used Not used Not used Not used Not used
2.1.1.4.2.2 SOP Specific Conformance to Basic Color Print Management Meta SOP Class
The Printing AE provides Standard Conformance to the Basic Color Print Management Meta SOP Class as an SCU. This implies standard conformance for the following SOP classes: Basic Film Session SOP Class Basic Film Box SOP Class Basic Color Image Box SOP Class Printer SOP Class
12
Similarly, the Basic Grayscale Print Management Met SOP Class uses: Basic Film Session SOP Class Basic Film Box SOP Class Basic Grayscale Image Box SOP Class Printer SOP Class
2.1.1.4.2.3 SOP Specific Conformance to Basic Color Image Box SOP Class
Supported DIMSE Services Name N-Set Usage M Description An image box instance is created by the SCP for each potential image of the film box. Only the instances which will actually contain images will be updated with the N_SET message.
Supported SOP Class Elements Name Image Position Pre-formatted Color Image Sequence Samples/pixel Photometric Interpretation Planar Configuration Rows Columns Pixel Aspect Ratio Bits Allocated Bits Stored High bit Pixel Representation Pixel Data Polarity Referenced Overlay Sequence >SOP Class UID >SOP Instance UID Magnification Type Smoothing Type Requested Image Size Usage M M M M M M M M M M M M U U U U U U U Range 1-n 3 RGB 1 600 800 8 8 7 0 Used Used Used Used Planar - red plane first, then green, and blue. Pixels Pixels Not used 8 bits per sample Used Bit 7 is MSB Unsigned pixel values Color pixel planes data Not used Not used Not used Not used Used Not used Not used Description
13
2.2.1.1 General
Maximum PDU size offered: Minimum PDU size accepted: 28,672 bytes 1,024 bytes
14
If there are queued Direct Send transfers pending an association will be initiated as well. In this case the transfers will be done in the background.
1.2.840.10008.5.1.4.1.1.3.1
JPEG Lossless Implicit or Explicit VR Little Endian Implicit or Explicit VR Little Endian JPEG Baseline JPEG Lossless
SCU
NOTE: For each storage server it is possible to enable or disable the usage of JPEG transfer syntax independently. If JPEG is enabled the compression factor can be set to a value between 80% and 100% . If the compression factor is set to 100% then the JPEG Lossless transfer syntax will be used.
2.2.1.5.1.1 SOP Specific Conformance to Verification SOP Class
The Storing AE dose not use the Verification SOP Class as an SCU.
2.2.1.5.1.2 SOP Specific Conformance Statement to the Ultrasound , the US Multiframe and the Secondary Capture Storage SOP Class.
The Ultrasound Image Storage SOP uses the Ultrasound Image IOD Modules as follows: Ultrasound Image Storage Modules Used Module Patient General Study Patient Study General Series Frame of Reference US Frame of Reference General Equipment 105952 Rev. 0 Usage M M U M U C M Description Used Used Used Used Not used Not used Used
15
General Image Image Pixel Contrast/bolus US Region Calibration US Image Overlay Plane VOI LUT SOP Common Curve Identification Curve Audio Curve SOP Common
M M C U M U U M M M U M
Used Used Not used Used Used Not used Not used Used Not used since the Curve IE is mutually exclusive with the Image IE. Not used since the Curve IE is mutually exclusive with the Image IE. Not used since the Curve IE is mutually exclusive with the Image IE. Not used since the Curve IE is mutually exclusive with the Image IE.
The Ultrasound Multi Frame Image Storage SOP uses the Ultrasound Multi Frame Image IOD Modules as follows: Ultrasound Multi Frame Image Storage Modules Used Module Patient General Study Patient Study General Series Frame of Reference US Frame of Reference General Equipment General Image Image Pixel Contrast/bolus Cine Multi-Frame Palette Color LUT US Region Calibration US Image VOI LUT SOP Common Curve Identification Curve Audio Curve SOP Common Usage M M U M U C M M M C M M C U M U M M M U M Description Used Used Used Used Not used Not used Used Used Used Not used Used Used Not used Not used Used Not used Used Not used since the Curve IE is mutually exclusive with the Image IE. Not used since the Curve IE is mutually exclusive with the Image IE. Not used since the Curve IE is mutually exclusive with the Image IE. Not used since the Curve IE is mutually exclusive with the Image IE.
The Secondary Capture Storage SOP uses the SC Image IOD Modules as follows: Secondary Capture Storage Modules Used Module Patient General Study Patient Study General Series General Equipment SC Equipment General Image Image Pixel SC Image Overlay Plane Modality LUT VOI LUT SOP Common Usage M M U M M M M M M U U U M Description Used Used Used Used Used Used Used Used Used Not used Not used Not used Used
16
Each module which is used by the Storing AE has a table below which indicates the elements supported. Patient Module Elements Name Patient's Name Patient ID Birth Date Patient Sex Referenced Patient Sequence Patients Birth Time Other Patient ID Other Patient Names Ethnic Group Patient Comments Use 2 2 2 2 3 3 3 3 3 3 Tag 0010, 0010 0010, 0020 0010, 0030 0010, 0040 Range xx xx xx xx Description Patient name with ^ delimiters 64 char max Used Used Not used Not used Not used Not used Not used Not used
General Study Module Elements Name Study Instance UID Study Date Study Time Referring Physician Name Study ID Accession Number Study Description Name of Reading Physician(s) Referenced Study Sequence Use 1 2 2 2 2 2 3 3 3 Tag 0020, 000D 0008, 0020 0008, 0030 0008, 0090 0020, 0010 0008, 0050 0008,1060 Range xx yyyymmdd hhmmss xx xx Description Used Exam date Exam time Used Zero Length Used Not used Used Not used
General Series Module Elements Name Modality Series Instance UID Series Number Laterality Series Date Series Time Performing Physicians Name Protocol Name Series Description Operators Name Referenced Study Component Seq. Body Part Examined Patient Position Smallest Pixel Value in Series Largest Pixel Value in Series Use 1 1 2 2C 3 3 3 3 3 3 3 3 2C 3 3 Tag 0008, 0060 0020, 000E 0020, 0011 0020, 0060 0008, 0021 0008, 0031 0008, 1050 0018, 1030 0008,103E 0008, 1070 0008, 1111 0018, 0015 0018, 5100 0028, 0108 0028, 0109 Range US xx xx yyyymmdd hhmmss Description Always US for ultrasound. Used Series number in exam Not used Used Used Used Not used Not used Used Not used Not used Not used Not used Not used
17
General Equipment Module Elements Name Manufacturer Institution Name Institution Address Station Name Institutional Department Name Manufacturers Model Name Device Serial Number Software Version Spatial Resolution Date of Last Calibration Time of Last Calibration Pixel Padding Value Use 2 3 3 3 3 3 3 3 3 3 3 3 Tag 0008, 0070 0008,0080 0008,1010 0008, 1090 0018,1000 VOLUSON 730 0 Range Kretztechnik Description Used Used Not used Used Not used Used Used Not used Not used Not used Not used Not used
Secondary Capture Equipment Module Elements Name Conversion Type Modality Secondary Capture Device ID Secondary Capture Device Manufacturer Institutional Department Name Secondary Capture Device Manufacturers Model Name Secondary Capture Device Software Version Video Image Format Acquired Digital Image Format Acquired Use 1 3 3 3 3 3 3 3 3 Tag 0008,0064 0008,0060 0018,1010 0008,1016 0008, 1018 0018,1019 0018,1022 Range WSD US Description Used Used Not used Not used Not used Not used Not used Not used Not used
General Image Module Elements Name Image Number Patient Orientation Image Date Image Time Image Type Acquisition Number Acquisition Date Acquisition Time Referenced Image Sequence Derivation Description Source Image Sequence Images in Acquisition Image Comments Lossy Image Compression Image Pixel Module Elements Name Samples Per Pixel Photometric Interpretation Use 1 1 Tag 0028, 0002 0028, 0004 Range 1 or 3 RGB, YBR FULL 422 Description RGB: 3 YBR_FULL_422: 3 MONOCHROME2: 1 Color: RGB: uncompressed, JPEG Use 2 2C 2C 2C 3 3 3 3 3 3 3 3 3 3 Tag 0020,0013 0020,0020 Range 1-n Description Image number in exam Zero length Not used Not used Used Not used Not used Not used Not used Not used Not used Not used Used for lossy compressed image
0008,0008
Xxx
0028,2111
0020,4000 0028,2110
Xxx 01
18
or MONOCHROME2
lossless or YBR_FULL_422: JPEG Baseline Grayscale: MONOCHROME2 V730 Expert: configurable per DICOM destination (800x600 or 640x480) V730 Pro: SC Images: configurable per DICOM destination US Images: always 640x480 V730 Expert: configurable per DICOM destination (800x600 or 640x480) V730 Pro: SC Images: configurable per DICOM destination US Images: always 640x480 Used Used Used Unsigned int Used Used unless MONOCHROME2 Not used Not used Not used Used if MONOCHROME2 Used if MONOCHROME2
Rows
0028, 0010
Columns
0028, 0011
Bits Allocated Bits Stored High Bit Pixel Representation Pixel Data Planar Configuration Aspect Ratio Smallest Image Pixel Value Largest Image Pixel Value Window Center Window Width Cine Module Elements Name Preferred Playback Sequencing Frame Time Frame Time Vector Start Trim Stop Trim Rec. Display Frame Rate Cine Rate Frame Delay Effective Duration Actual Frame Duration
1 1 1 1 1 1C 1C 3 3 3 1C
0028, 0100 0028, 0101 0028, 0102 0028, 0103 7FE0, 0010 0028, 0006 0028, 0034 0028, 0106 0028, 0107 0028, 1050 0028, 1051
8 8 7 0 0
127 256
Use 3 1C 1C 3 3 3 3 3 3 3
Tag 0018,1244 0018,1063 0018,1065 0008,2142 0008,2143 0008,2144 0018,0040 0018,1066 0018,1072 0018,1242
Range
Description Not used Set to interframe time Not Used Not used Not used Not used Not used Not used Not used Not used
Multi Frame Module Elements Name Number of Frames Frame Increment Pointer Use 1 1 Tag 0028,0008 0028,0009 Range Set to number of frames in image. Is set to Frame Time (0018,0063) Description Used Used
US Image Module Elements Name 105952 Rev. 0 Use Tag Range Description Voluson 730Expert/Pro DICOM Conformance Statement
19
Photometric Interpretation
0028, 0004
Color: RGB: uncompressed, JPEG lossless or YBR_FULL_422: JPEG Baseline Grayscale: MONOCHROME2 Unsigned int Not used for lossy compressed image Not used Not used Not used Not used Not used Not used Not used Not used Not used Not used Not used Not used Not used Not used Not used Not used Not used Not used Not used Not used Not used Not used Not used Not used Not used Not used Not used Not used Not used Not used Not used Not used Not used Not used
Pixel Representation Frame Increment Pointer Lossy Image Compression Number Stages Number Views in Stage Referenced Overlay Sequence Referenced Curve Sequence Stage Name Stage Number View Number Number of Event Timers Event Elapsed Times Event Timer Name Transducer Position Transducer Orientation Anatomic Structure Trigger Time Nominal Interval Beat Rejection Flag Low R-R Value High R-R Value Heart Rate Output Power Transducer Data Transducer Type Focus Depth Preprocessing Function Mechanical Index Bone Thermal Index Cranial Thermal Index Soft Tissue Thermal Index Soft Tissue-focus Thermal Index Soft Tissue-surface Thermal Index Depth of Scan Field Image Transformation Matrix Image Translation Vector Ultrasound color data present
1 1C 1C 2C 2C 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3 3
0 0,1
Remarks: Internally the V730 Expert/Pro stores all images in color (RGB or YBR_FULL_422). Whenever a destination is configured for MONOCHROME2 the images are converted on the fly and then transmitted. In this case the SOP instance UID of the transmitted DICOM object is changed.
Secondary Capture Image Module Elements Name Date of Secondary Capture Time of Secondary Capture Use 3 3 Tag 0018, 1012 0018, 1014 Range Not used Not used Description
20
US Region Calibration Module elements Name Sequence of Ultrasound Regions >Region Spatial Format >Region Data Type >Region Flags >Region Location MinX0 >Region Location MinY0 >Region Location Max X1 >Region Location Max Y1 >Reference Pixel X0 >Reference Pixel Y0 >Physical Units X Direction >Physical Units Y Direction >Physical Delta X >Physical Delta Y SOP Common Module Elements Name SOP Class UID Image Storage. SOP Instance UID Specific Character Set Instance Creation Date Instance Creation Time Instance Creator ID Use Tag 1 0008, 0016 1 0008, 0018 1C 3 3 3 Range 1.2.840.10008.5.1.4.1.1.6.1 1.2.840.10008.5.1.4.1.1.3.1. 1.2.840.10008.5.1.4.1.1.7 xxx Description Used for US image storage Used for US multiframe image storage Used for Secondary Capture storage Same as in Command Set Not used Not used Not used Not used Use 1 1 1 1 1 1 1 1 3 3 1 1 1 1 Tag 0018,6011 0018,6012 0018,6014 0018,6016 0018,6018 0018,601a 0018,601c 0018,601e 0018,6020 0018,6022 0018,6024 0018,6026 0018,602c 0018,602e Range Used 1,2,3 1 0 0-799 0-599 0-799 0-599 0 0-xxx 3,4 3,7 Used Used Used Used Used Used Used Used if Region Spatial Format = 3 Used if Region Spatial Format =3 Used Used Used Used Description
2.3.1.1 General
Maximum PDU size offered: Minimum PDU size accepted: 28,672 bytes 1,024 bytes
21
2.3.1.5.1.1 SOP Specific Conformance Statement to Modality Worklist Service SOP Class
Each module which is used by the Worklist AE has a table below which indicates the elements supported.
Scheduled Procedure Step Module Elements Name Scheduled Procedure Step Sequence >Scheduled Station AE Title >Scheduled Procedure Step Start Date >Scheduled Procedure Step Start Time >Modality >Scheduled Performing Physicians Name >Scheduled Procedure Step Description >Scheduled Station Name >Scheduled Procedure Step ID Tag 0040, 0100 0040, 0001 0040, 0002 0040, 0003 0008, 0060 0040, 0006 0040, 0007 0040,0010 0040, 0009 Return Type 1 1 1 1 1 2 1C 2 1 Mapped into image No No No No Yes Yes No No No Description Used Used Used Used US or zero length Used Used Used if present in the query form Used
22
Requested Procedure Module Elements Name Tag Requested Procedure ID Requested Procedure Description Study Instance UID 0040, 1001 0032, 1060 0020, 000D
Imaging Service Request Module Elements Name Tag Accession Number Referring Physicians Name 0008, 0050 0008, 0090
Patient Identification Module Elements Name Tag Patients Name Patient ID 0010, 0010 0010, 0020
Patient Demographic Module Elements Name Tag Patients Birth Date Patients Sex Patients Size Patients Weight 0010, 0030 0010, 0040 0010,1020 0010,1030
Return Type 2 2 3 2
Tag
0010, 21D0
Description Used
3 Communication Profiles
3.1 TCP/IP Stack Supported
The TCP/IP protocol is used. The port address is configurable as stated elsewhere in the spec DCS.
4 Extensions/Specializations/Privatizations
4.1 Standard Extended/Specialized/Private SOPs
None Voluson 730Expert/Pro DICOM Conformance Statement 105952 Rev. 0
23
5 Configuration
This device obtains configuration information at the time of installation to provide the following. mapping from Application Entity Title to Presentation Address device configuration information
24