You are on page 1of 5

UN-III oximeter module users manual

Page 1 of 5
Address: Rm1,10,14F,Wan'anguojiqiyedaodao,Gexin Road #568,
Wujiashanjingjikaifaqu, Dongxihu District,Wuhan,China
Postcode: 430040
Phone: (86) 027-83083133
Fax: (86)027-83080833
Email: info@unmedtech.com
Website:www.unmedteh.com
UN-III oximeter module users manual
Page 2 of 5
Advanced DSP algorithm minishes the influence of motion
artifact
All digital-signal processing improves accuracy of low
perfusion
Small size, easy to be assembled
UN-III oximeter module users manual
Page 3 of 5
Range and accuracy
Range
Pulse rate: 30 250 bpm
SpO2: 0 100%
Resolution1SD
Pulse rate 30250BPM
2bpm Normal
3bpm Motion or low perfusion
SpO2 70100%
2% Normal
3% Motion or low perfusion
Below 70% Not specified
Environmental Requirements
Operating temperature: 045 C
Storage temperature: -2060 C
Operating Humidity3095%
Storage Humidity10100%
Operating Altitude-5005000m
Storage Altitude-50010000m
Power Specifications
Par. Min. Typ. Max.
5V 3.6V 5V 5.25V
Ripple --- --- 100mVp-p
Communication Protocol
QSZ-MN-III Module communicates with the host via a asynchronous serial port at
TTL levels, the data format is 4800baud, 8 data bits, odd parity, 1stop bit 4800
8-Odd-1.
UN-III oximeter module users manual
Page 4 of 5
Byte Bit Description Byte Bit Description
0
0 Signal strength 0
3
0 Pulse rate 0
1 Signal strength 1 1 Pulse rate 1
2 Signal strength 2 2 Pulse rate 2
3 Signal strength 3 3 Pulse rate 3
4 1 = Searching too long 4 Pulse rate 4
5 1 = Sensor unplugged 5 Pulse rate 5
6 1 = Pulse beep 6 Pulse rate 6
7 1 (Syn. bit) 7 0 (Syn. bit)
1
0 Plethysmogram 0
4
0 SpO2 0
1 Plethysmogram 1 1 SpO2 1
2 Plethysmogram 2 2 SpO2 2
3 Plethysmogram 3 3 SpO2 3
4 Plethysmogram 4 4 SpO2 4
5 Plethysmogram 5 5 SpO2 5
6 Plethysmogram 6 6 SpO2 6
7 0 (Syn. bit) 7 0 (Syn. bit)
2
0 Bargraph 0
1 Bargraph 1
2 Bargraph 2
3 Bargraph 3
4 1 = Sensor error
5 1 = Searching for pulse
6 Pulse rate 7
7 0 (Syn. bit)
UN-III oximeter module users manual
Page 5 of 5
Connections
P1-Probe connector
P1 Pin Description
1 Probe output
2 3.3V
3 RED+
4 RED-
5 GND
6 NC
P2-Host connector
P2 Pin Description
1 TX Output
2 RX Input
3 GND
4 3.6V-5V
Size
Unitmm
Probe
Only digital probes can be used for this module.