You are on page 1of 1

Method: Curva_Cu_25.07.

14

Page

Date: 8/22/2014 10:09:41 AM

====================================================================================================
Analysis Begun
Logged In Analyst: Dieq
Spectrometer Model: AA700, S/N 700S10080202

Technique: AA Flame
Autosampler Model: S10

Sample Information File: C:\data-AA\Dieq\Sample Information\Pedro_Cu_22-08-2014.sif


Batch ID:
Results Data Set: Pedro_Cu_22-08-2014_r
Results Library: C:\data-AA\Dieq\Results\Results.mdb
====================================================================================================
Method Loaded
Method Name: Curva_Cu_25.07.14
Method Last Saved: 5/3/2003 4:07:11 AM
Method Description: Curva_Cu_25.07.14
====================================================================================================
Sequence No.: 1
Autosampler Location: 1
Sample ID: h2o
Date Collected: 8/22/2014 10:06:26 AM
Analyst:
Data Type: Original
---------------------------------------------------------------------------------------------------Replicate Data: h2o
Repl
SampleConc StndConc BlnkCorr
Time
Signal
#
mg/L
mg/L
Signal
Stored
1
[0.00]
0.505
10:06:47
Yes
2
[0.00]
0.505
10:06:51
Yes
3
[0.00]
0.505
10:06:56
Yes
Mean:
[0.00]
0.505
SD:
0.00
0.0001
%RSD:
0.00
0.02
Auto-zero performed.
====================================================================================================
Sequence No.: 2
Autosampler Location: 2
Sample ID: padrao1
Date Collected: 8/22/2014 10:07:21 AM
Analyst:
Data Type: Original
---------------------------------------------------------------------------------------------------Replicate Data: padrao1
Repl
SampleConc StndConc BlnkCorr
Time
Signal
#
mg/L
mg/L
Signal
Stored
1
[2]
0.106
10:07:43
Yes
2
[2]
0.105
10:07:47
Yes
3
[2]
0.106
10:07:52
Yes
Mean:
[2]
0.106
SD:
0
0.0004
%RSD:
0
0.38
Standard number 1 applied. [2]
Correlation Coef.: 1.000000
Slope: 0.05285
Intercept: 0.00000
====================================================================================================
Sequence No.: 3
Autosampler Location: 3
Sample ID: padrao2
Date Collected: 8/22/2014 10:08:18 AM
Analyst:
Data Type: Original
---------------------------------------------------------------------------------------------------Replicate Data: padrao2
Repl
SampleConc StndConc BlnkCorr
Time
Signal
#
mg/L
mg/L
Signal
Stored
1
[4]
0.200
10:08:39
Yes
2
[4]
0.199
10:08:44
Yes
3
[4]
0.199
10:08:48
Yes
Mean:
[4]
0.199
SD:
0
0.0003
%RSD:
0
0.15
Standard number 2 applied. [4]
Correlation Coef.: 0.997081
Slope: 0.05045
Intercept: 0.00000
====================================================================================================
Sequence No.: 4
Autosampler Location: 4
Sample ID: padrao3
Date Collected: 8/22/2014 10:09:14 AM
Analyst:
Data Type: Original
---------------------------------------------------------------------------------------------------Replicate Data: padrao3
Repl
SampleConc StndConc BlnkCorr
Time
Signal
#
mg/L
mg/L
Signal
Stored

You might also like