Practical Leakage Testing
of Compressor Valves a
1. Introduction 2.1 Flow Meter Testing
Compressor valves should be flow- The leakage is determined by and the escaping air will blow out
efficient and durable in service. subjecting the closed valve to a any small particles that might be
Leakageis onecriteria thatimpacts “set air pressure of e.g. Shar (70 trapped between the sealing
‘on both of the above. Excessive PSI) and measuring the leakage surfaces and cause excessive
leakage causes backflow of gas through the valve hy means of a leakage.
when the valveis closed, raises the flow meter (fig. 1) This should be done with an
temperatureofthe gas.andreduces Since small contaminants at times appropriate tool flat or sof-tipped)
efficiency. Elevated temperatures can be caught between the valve so no damage is done to the valve
also adversely affect valve lite. plate and seat lands, it is common plate or sealing element. Particular
practice to "tap" the valve plate in cares called for when non-metallic
For this reason, HOERBIGER's —_ several areas repeatedly. This will plates are used so no damage is
quality assurance program dictates _brieflylitthe plate offthe seatlands done to these components.
that all valves are subjected to a
suitable leakage test. This policy
covers newly made valves in our
production plants and repaired
valves in all our service centers.
Ditterent geometry valves and |
sealing elements made from steel
‘ornor-metallic material all undergo
such rigorous testing,
@ oe
Test methods may vary but prefer- Poem
cence ie given to those procedures
that simulate the conditions as they
" rt
2. Valve Testing Methods
Soman
There are 8 customary leakage test | ot
procedures, and HOERBIGER | =
recommends 2 of the testing
methods.
Lig
HOERBIGER CORPORATION OF AMERICA, INC.
1981 S. W. 0th Avenue, Pompano each, Florida 33069
Phone (954) 974-5700 or Toll-Free (800) 327-8961+Fax (964) 974-0964 a
VO15A. ‘AOVO1SA120X0002.2 Measuring Pressure Drop
In this test method, leakage is
‘measured in pressure drop over
time, using a fixed volume "V" (air
supply is shut off) and measuring
the time it takes for the pressure to
drop trom pt to p2 (fig. 2).
The time for the pressure drop to
occur and parameter for acceptable
leakage are dependenton the valve
size, volume of the vessel and the
selected pressures of p1 and p2
‘The prior “tapping” of the vaive to
blow out any impurities should be
standard practice in this test
procedure as well
2.3 Liquid Testing
Test procedures as discussed
under 21 and 22 are compatible
and represent conditions as they
prevail in the compressor cylinder
The sealing element is “pressure
loaded” much like it is in its
operating environment when
installed in a cylinder (the suction
valve is kept closed by the cylinder
pressure while the discharge vaive
is under the influence of the line
pressure).
Frequently field people donothave
access 10 leakage testers and
resort to a simple "liquid testing* of
the valve.
Filling the seat ports of a valve with
a fluid and observing the leakage
rato of the fluid through the valve is
‘a common practice. Kerosene,
Varsol and other light viscosity
fluids (even water) are employed in
such crude leakage tests.
‘Although the procedure is widely
used, itis not withoutcontroversy as
tothe viability ofits results. Contrary
to the previously described
methods of air pressure testing, the
liquid test does not simulate
conditions as found in the cylinder.
The fluid pre-loads the sealing
element towardsits opening ift, and
test results can be distorted,
especiallyifvalves use non-metallic
sealing elements andif the pre-load
of the springs used in the valve 1s
very light.
Certain valves, due to their design,
cannot be tested at ll with the fluid
method, and others will create the
9
—
ee
Raper
ch,
appearance of “leakers" when Uir
sealing characteristic under air
roseure ie perfectly acceptable.
This test method, therefore, is
controversial and should be avoided
whenever possible. The cleaning
process described in air testing
cannot be applied in liquid tests, and
any impurties in the sealing area will
show as excessive leakage.
Frequently, such liquid tests lead to
unnecessary labor and repair of the
valve which would possibly be termed
fine under air tests.
Itis recognized that proper testing of
valves with air pressure is not always
possible, but it remains the superior
prodedure and should be adopted
wherever possible.