JURUSAN ANALIS KESEHATAN POLTEKKES DEPKES PONTIANAK LINGKUP KERJA INSTRUKSI KERJA
: :
Tanggal Berlaku:
Halaman :
Laboratorium Jurusan Analis Kesehatan
Daily Probe Cleaning
Supplies :Daily Probe cleansing Kit
Fill the probe cleansing wedge with 2 ml probe cleaning
solution Place the Probe Cleansing wedge into position 1 of the reagent carousel. Select DIAGNOSTICS FROM THE Start-up screen. Select PRBCLEAN from the common Diagnostics screen,and then RUN. Select GO on the Display Panel. When the probe Cleaning is finished,select GO on the Display Panel to run the PROBE ANGLE test,and observe the liquid dispersed from the probe. Simultaneously press ALARM MUTE and GO on the display Panel yo stopthe diagnostic procedure. Remove the Probe Cleaning Wedge from the Reagent Carousel. Select CLOSE to return to the Start-up screen.
Daily Startup
Select RUN IMMULITE or RUN TURBO from the Startup screen.
Enter the User ID and select OK. Wait for program files to download to the instrument. Select GO on the Display Panel,and follow the prompts. Review the Reagent Status screen after loading reagents,samples,and Test Units. Select CLOSE 1 to return to the Home Screen.
Daily Shutdown
Select LOG OFF from the menu bar.Logging off
automically initiates a system backup, and return you to the Start-up screen. Select SHUTDOWN from the Strart-up screen and then YES. Select Start from the task bar, and then Shut Down. Select Restart from the drop-down menu, and then OK. The computer will restart,and Log On to Windows box displays. Select OK.(No password is required) Double-click the IMMULITE 1000. Icon on the desktopto redisplay the Start-up screen.
Activating Remote Diagnostics
From the running software,press[Alt] and [Tab]
simultaneously on the keyboard to display the task bar and select Start. Select pcAnywhereHost Mode option. The pcAnywhere Managerwindow will open. Double click the icon. A computer monitor icon appears on the bottom right side of the screen conforming the connection.