Calibration Certificate
Certificate No. 5019348 Sold to: Green Group
Product Defender 520 High Flow ‘Ay, San Luis 2687 - Of. 202
Serial No. 118988 ‘San Borja ~ Lima, Peru
Cal. Date ‘9/28/2012
{calibrations are performed in accordance with 'SO 17025 at Bios International, a division of Mesa Laboratoire, Inc, 10 Park Place, Butler, NU,
(07405, 800-653-4977, an SO 17025:2005 - accredited laboratory through NVLAP. This report shall not be reproduced except in full without the
written approval of the laboratory. Results only relate to the items calibrated. This report must not be used to claim product certification,
‘approval, or endorsement by NVLAP, NIST, ar any agency ofthe Federal Government.
‘Al units tested in accordance with our test number PR17-13 using high-purity nitrogen or filtered laboratory a
As Received Calibration Data
Lab. Pressure 755 mmHg
Technician Sonia Otero Lab. Temperature 22.9 °C
Instrument Reading Tab Standard Deviation ‘Allowable as
Reading Deviation Received
50630 cer 500,905 ccm 0.68% 1.00% __|____In Tolerance
5019.8 ccm 5006.3 ccm 0.27% 7.00%. In Tolerance
30027 ccm '30078.5 ccm: 0.17%. 7.00%, In Tolerance
232°C. 229°C. 03°C. 308°C In Tolerance
754 mmbig 755 mmHg, =immig Z5mmiig In Tolerance
Bios International Standards Used
Description ‘Standard Serial Number Calibration Date Calibration Due Date
ML 500-46 113762 3/21/2012 9/21/2013,
Precision Thermometer 300907 ‘4/26/2012 ‘4/26/2013
Precision Barometer 2981392 f4/2012 6/4/2013
Bios International, a division of Mesa Laboratories Inc. 10 Park Place Butler, NJ 07405 USA
(973) 492-8400 FAX (973) 492-8270 www. com wwww.mesalabs.com Symbol "MLAB" on the NASDAQ
Page 1 of 2 ‘CALO2-50 Rev D02| :
MesaLabs nya
As Shipped Calibration Data
Certificate No. 5019348 Lab. Pressure 752 mmHg
Technician Brian Roberts Lab. Temperature 225°C
Instrument Reading Lab Standard Deviation ‘Allowable is
Reading Deviation Shipped
502,83 com 500.66 ccm: 3% 1.00%. Tn Tolerance
5010.1 com. 5004.2 ccm. 0.12% 1.00%. In Tolerance
29916 ccm ‘30062 cem 0.49% 1.00%. Inolerance
221°C 221°C : 30.8°C. In Tolerance
752 mmHg 752 mmiig : =3.5mmHig In Tolerance
Bios International Standards Used
Description Standard Serial Calibration Date Calibration Due Date
Number
ML-500-46 113762 3721/2072, 3721/2013,
Precision Thermometer. 305460. 8/20/2012. 8/20/2013,
Precision Ssrometer 2981390 42012 t of6/2013 1
Calibration Notes
Bios isan 1SO 17025-accredited metrology laboratory. Each Bios primary gas flow standard is dynamically verified by comparing it to one of our
laboratory standards, which isa Proven DryCal@® Technology volumetric piston prover of much higher accuracy (0.25 or better] but of similar
‘operating principles. For this purpose, a flow generator of 20.10 % of less stabilly is used. ur laboratory standards are qualified by direct
‘measurement of their dimensions (diameter, length an time] using NIST-traceable precision gauges and instrument, such as depth micrometers
{and laser micrometers, Calibration Certificates for these gauges and instruments are available upon request. Rigorous analyses of our laboratory
standards’ uncertainties have been performed, in accordance with The Guide tothe Expression of Uncertainty in Measurement ithe GUM assuring
their traceable accuracy.
Technician Notes:
Dwwrn
David W. Wilson, Chief Metrologist
Bios International, a division of Mesa Laboratories Inc. 10 Park Place Butler, NJ_07405 USA
(873) 492-8400 FAX (973) 492-8270 wwww.biosint.com www.mesalabs.com Symbol "MLAB" on the NASDAQ
xis Page 2 of 2 (CALO2-50 Rev D02