ReferenceDOC1017forcompleteoperatinginstructions forthe6101Integrator.

After the integrator/Belt Scale has been initially setup and calibrated, the following
procedurewillprovideinstructionsforperformingzeroandspancalibrations:

RUNBELTEMPTY
ZEROCALIBRATION:
1.
2.
3.

4.

PresssoftkeyZEROCAL
PresssoftkeyAUTOZERO
PresssoftkeyBGINZERO
Message Auto Zero in progress, please wait is displayed and a timer will
begincountingdown.Whenthecounterreaches0,AutoZeroCompleteis
displayedandthezeroerrorin%isdisplayed.
PresssoftkeySUREZEROtoacceptcalibration.Ifforsomereasonyouknow
thezerotestwasbad,presssoftkeyEXITZEROtoendprocedureorBGIN
ZEROtostartanotherzerocalibration.

PRESSRUNWHENFINISHEDORPROCEEDTOSPANCALIBRATION

SPANCALIBRATION:
1.

2.
3.

4.

5.

To verify or change the span calibration method (Electric, Weigh or Chain),


press MENU key, and then press 3 to select 3 SYSTEM DATA. PAGE
DOWN to CALIBRATION DATA 1 menu. The current method will be
displayednexttoSIMULATED[].Ifcorrectmethodisdisplayed,skipto
step2below.Tochangemethod,presskeypadENTkey,thenpresssoftkey
ENTR to scroll through options. When your desired option is displayed,
presssoftkeySUREtoselect.
If using static weights [Weigh] or Calibration Chain [Chain], apply the
appropriatemethodtothebeltscale.
PressMENUkey,andthenpress2toselect2CALIBRATION.Pressscroll
down arrow to 2.2 SPAN CALIBRATION, then press keypad ENT. The
methodselectedinstep1willbedisplayedonthescreen.ieCalibration,
StartBelt
PresssoftkeyAUTOCAL,andthenpresssoftkeyBGINCALtostartspan
calibration. Message Auto Span in progress, wait is displayed and a
timerwillbegincountingdown.Whenthecounterreaches0,ElectricAuto
SpanComplete
PresssoftkeySUREspantoacceptcalibration.Ifforsomereasonyouknow
thespantestwasbad,presssoftkeyEXITCALtoendprocedureorBGIN
CALtostartanotherzerocalibration.

PRESSRUNWHENFINISHED