EASYpure II and EASYpure RODI
Dara DispLavep Upon Startup “
1. Model Identification (see Model Definitions)
2. Allsegments of display turned OFF / ON OFF
3. Software Revision Data
4. Mode of Operation or Error Message (see Mode and Error Message Definitions)
Mover Dermurrons
r-dl___ EASYpure RODI mode!
F Reservoir Fed EASYpure i model
1F-Ut___ Reservoir Fed with UV Lamp EASYpure Il model
LF Line Fed EASYpure if model
uF Line Fed with Uitrafilter EASYpure IT model
ut Line Fed with UV Lamp EASY pure Il model
UtUF Line Fed with UV Lamp and Uluafiiter EASYpure It model
dL Idle Mode_**
3 ‘Standby Mode
TEC Recirculating Di water while in Standby Mode
FLA Flush Mode with number of minutes remaining
‘AFL___ Auto Flush Mode
### Run Mode displays Temperature Compensated Resistivity
Ernor Message DEFINITIONS
Er Purity meter resistivity error
Ea Check feed water supply low/pressure (line fed units only, not on RODI)
‘Add Water level in reservoir tank is low (reservoir fed units only, including ROD)
ES ‘Check UV Lamp
rs Purity meter temperature error
ES ___ Purity meter communication error
Ex6____ RO float stick error
Er? Excessive reservoir tank fill ime
RESETTING THe Cannon Preritrer Tie (EASY pure RODI model only)
Tum unit OFF
ress and bod the Flash key while ring the nit ON
‘When “rESEt 10” begins to scroll across the display release the Flush
When “--- appears:
Press Star/Stop key ~» “yES” appears on display
, Press Starv/Stop key ~+ “000” appears on display
Timer is reset.
Note:
If the Standby key is pressed in step 2. ot b. above, the reset procedure will be terminated, 6
[Fbutton presses between steps are not done within 10 seconds, the reset procedure will be terminated,
Modif Date: 19May04 Page Lof Print Date: 571904
1B Rev