Calibration A Dew Point Measurement System
Calibration A Dew Point Measurement System
Calibrationadewpointmeasurementsystem
Calibrationadewpointmeasurementsystem
Howtocalibrateadewpointmeasurementsystem?Howcouldwemakesurethatthereadingis
correct?
ByVahidon29June,201011:19am
WehaveaGEdewpointmeasurementequipmentinstalledontheoutletgaspipelineofourplant,I
wouldneedtohavemoreinformationaboutthewaysthatwecouldmakesurethatthereadingis
correct.
Idoappreciateyourconcernandadvice.
ByCSA
on29June,20107:31pm
Doyouhavetheinstructionmanualforthedevice?
Arethereanyinstructionsinthemanualaboutverifyingcalibration/output?
Theseareusuallyverydifficulttocalibratewithoutalabsetupwhichcanflowairofaknown
humidityoverthesensor.
Ibelievethebestwayistosendthedevicetoapropercalibrationlab,unlessyoursitehasavery
goodinstrumentationshop/lab.
BySam on30June,20108:21am
Hithere,
NotmuchyoucandotoverifytheiraccuracyinthefieldI'mafraid.Isuggestyoutaketheoriginal
datasheetsuppliedwiththeanalyzer/transmitterandseeifthereissomecalibrationconstantsthat
youcancheck.Consultthemanualsincetheymighthaveacalibrationvalidationprocedurein
there.Itwillbesomeconstantsvaluesyoucancheckonthedisplayorsomevoltagesyoucan
measureonthePCB.
SomethingyoucancheckisthemAoutputsaccordingtothefactorycalibratedzeroandspan
values.Ifitisdewpointyouhaveontheoutputitshouldhaveafacilityonthematrixofthe
transmittertosimulatethe0to150degreesCorwhateveryourfactorycalibrationshouldbe
accordingtothedatasheet.
Youcanthenmakeadjustmentsonthetransmitter'soutputtomakesurethecorrectZ/Sdewpoint
values(asperdatasheet)giveaaccurate4to20mAoutput.
Iknowthemanufacturercanalsosupplyyouwithvarioussaturatedsaltsolutionsthatyoucanuse
inalaborw/shopenvironmenttoverifytheaccuracyofthetransmittersogetholdofthemandask
themaboutthat.
Anotherpossibilityistobuyahotfactorycalibratedstandbyidenticaltotheinstalledunitandjust
http://control.com/thread/1277824746
1/2
5/11/2015
Calibrationadewpointmeasurementsystem
changeitoutwiththisspareonewhileyousendtheotheroneforthecalibrationverificationtothe
factoryoneayearoronceeverytwoyears.Theremightbealimitedsensorshelfliveorspecial
sensorstoragerequirementsinvolved,sospeaktothesupplierandseeifthismightbepossible
option.
YouruseofthissiteissubjecttothetermsandconditionssetforthunderLegalNoticesandthe
PrivacyPolicy.Pleasereadthosetermsandconditionscarefully.Subjecttotherightsexpressly