THE RF VECTOR VOLTMETER-AN IMPORTANT
NEW INSTRUMENT FOR AMPLITUDE AND
PHASE MEASUREMENTS FROM 1 MHz TO 1000 MHz
A broadband two-channel milliveltmeter and phasemeter
simplifies many measurements heretofore often
neglected. Included are device gain and loss, impedance
and admittance, length inequalities in transmission paths, and
precision frequency comparisons.
Fig. 1, ~hp- Model 8405A Vector Voltmeter measures amplitudes end
‘Phase simulianeoualy. Inatrument has frequency range of 1 MHz to]
Gla, sensitivity of 100 nV fullscale, dynamic range of 5 AB, phase
‘resolution of 0.1", and is simple to operate. Thus it makes feasible many
imeasurements which formerly were difficult or impossible.
n where inforn
nly difficult o obtain. ‘ning feedback ampli-
ul phase are so fundamen. ng the electrical length of
ical engineering that the cables, measuring group delay, and.
tor Voltmeter hasan extraordi- many others.
‘Vector Voltmeter (Fig. 1) isa wochan: nary number of can, Although adequate voltmeters for
nel millivoltmeter and phasemeter: it for example, measure complex or vec- measuring amplitudes over a wide fre-
nd tor parameters such as impedance yy range have been available for
and phase no equally
shift, complex insertion loss or gain, convenient means for measuring phase,
nents of the signals in ch nnd complex reflection coefficient, ¢wo-port Consequently, simultaneous measure.
B; it may then be switched! to network parameters, and filter transfer ments of voltage and phase have not
the voltage in channel B so that gain functions. It always been easy to make, Most systems
for loss may be determined. It makes lective receiver and as a design tool: which areable 10 measure phase angles
these measurements over a broad fre- possible require several control adjustments for
quency range (Ito 1000 MHL) ina part leak ‘each measurement, and many of them
NeW INstREMENT, of the spectr
in to become one of often pecul
ajor electronic measuring instr
recently been developed by
the -hp- Microwave Division. The RF
fiers, meas
applications,
the voltage in channel A,
simultaneously measures the phase an- admittanee, amplifier
gle between1 1000 Me
YW —,
Sampling
‘robe
Fig. 2. Block diagram of hp *
Modet 84054 Vector Voltmeter
‘Automatic phase control (APC)
tuses phase-locked Toop to tune
‘and phase-lock meter to channel vio [+4 ape vourmerer lonasemereR|
A signal. APC adjuste frequency | Ye
Of voltage-luned local’ onlletor Sampling
(VTO) whick triggers sampling: Probe
type mixers in probes, RE signals e
are reconstructed from samples
At intermediate frequency of 20
Hz, where voltage and phase are | 1g|
measured.
20KHe
are Himited in freq)
tivity, and dynamic ra
yrange, sensi:
we
he new Vector Voltmeter (VVM),
‘on the other hand, operates over a fre
quency range of 1 MHz to 1 GHz, It
has high sensitivity and wide dynamic
range. Its phase resolution is 01°
any phase angle at all frequencies, and
it operates with the simplicity of «vol
meter rely selects ap
propriate meter ranges, touches two
probes to the points of interest, and
Teads voltage and phase on two meter.
cr, the WM has nine
voltage ranges, which have fullscale
sensitivities of 100 gV to. V rims. Its
dynamic range is 99 AB, which
that it can measure gains or losses of
(0 95 dB, The 10:1 voltage divide
suppl sith the ins
tw measure voltages up ta 10 V
As a phasemeter, the VWM will
measure phase angles between +180"
and —180°. 1, has four ranges: + 180°,
160°, 18°, and 6°, The phase
rieter can be offset up to.“ 180° in 10°
steps so that any phase angle may b
read on the =+6 range, whch fas 0.1°
resolution, For example, a phase angle
of 115° can be measured with 0.1
revolution by selecting a phaemeter
offet of -+140" oF 150° and asing
the 6° range. Phase readings ate in
dlepensdeni of the voltage levels in the
two channels,
As a volun
iment enable it
‘The reference signal for the phase
measurement is channel A, An auto
matic phase control eireuit (APC
tunes and phase-locks the instrument
tothe channel A signal. The frequency
range of the APC is selected by means
of a frontpanel control; there are 21
overlapping ranges, each more than aa
octave in width, In making 2 measure
ment, the operator selects a frequency
range which includes the frequency of
the signal whieh is driving the circuit
tunder test, ‘The APC then tunes the
um tically and esen-
tially instantaneously (10 milliseconds).
and Keeps it tuned! even if the input
Frequency drifts or sweeps at moderate
p 10 15 MHz/second),
In the input probes of the VWM are
ype mixers which convert
the RF signals to a 20kHy interme:
diate frequency, where the voltage and
phase measurements ate made
back stabilization of the mixers keeps
the voltage conversion loss at 0 dB de.
spite H influences, and
common local-oscillator drive for both
samplers keeps the pl
tuccen the IF signals equal to the phase
difference between the RF signals,
Feed:
we diflerence be:
‘The RF waveformsare reconstructed
at the intermediate frequency: the
1 components of the RV
converted 10 90 LH, the
‘
wavefon
seconel harmonies to 40 KH, the thi
13s
monies to 60 KHz, and so on, up to
highest harmonic of the inpue sig:
nal which falls within the lKGHz band:
width of the samplers, Outputs ate
provided directly from the sampling
mixers in both channels. Since the in
put waveforms are preserved in the IF
Is. the VWM ean be used to con:
ny low-frequency oscilloscopes,
wave analyzers, and distortion analyz:
cers to high-frequency sampling instru:
ments for signals of moderate harmonic
content. A similar sampling prineiple
was originally employed by -bp- in
sampling.type oscilloscopes.”
For the voltage and pl
ments, the IF signals from the sav
pling mixers are filtered so that only
their 20kHz fundamentals remain,
nnd the amplitudes of these fun
mentals and the phase angle between
them are measured and displayed on
the two frontpanel ameters (see block
diagram, Fig, 2) Since only the funda.
mentals are measured, the amplitude
nel phase readings are not affected by
the harmonic content of the input sig:
nals, The narrow bandwidth IF filters
(URED) also reduce th
the meter inputs, The de meter signals
for both voltage and phase are avail-
able at the rear panel and ean be sect
tw drive recorders
a.carn ct mama nay, aehp: 6086 hp: 8050
hp. 84054
VECTOR
VOLTMETER,
SIGNAL.
[GENERATOR
‘SLOTTED
LINE
(a)
440
420
400
380
360
PHASE (degrees)
340
320
290 295
300
PROBE POSITION (mm)
(G}
PHASE-MEASURING CAPABILITIES
Figs. 3(b) and 3(¢) demonstrate the
phase-measuring capabilities of the
Vector Voltmeter. They show, first of
all, how the high phase resolution of
the VVM makes possible very precise
measurements of length. Fig. 3(b) also
includes an example of the phase-
measurement accuracy of the instru-
ment.
‘To obtain the data for Figs. 3(b) and
8(¢), a 1-GH> signal was applied first
ton unloaded slotted line and then to
the same slotted line with a 50ohm
load [sce block diagram,
Probe A of the VWM was p
output of the slotted line, and probe B
was attached to the movable slotted-
line probe. Fig. 3(b) 1s a plot of the
phasemeter readings versus. the posi
tion of the slotted-line probe. The
measured curve closely follows the the:
oretical curve fi open
lossless line.
Without the 50-ohm oad, the stand-
ing-wave ratio on the line was 505.
This was determined by measuring the
mum and minimum voltages on
ne with the volumeter of the VVM_
switched to channel B. The phase-vs
position curve fs the stepslike curve of
shows one of the
steep portions of this curve with an
\led horizontal scale. The maxi
jum rate of change of phase can be
determined from Fig. 3(¢) to be 50° per
millimeter, or 0.05° per micrometer
Thus, a change equal to the diameter
sae
100200 ~—«300~—«400,
PROBE POSITION (mem)
(2)
Demonstration of phase-measuring capabilities
of Vector Voltmeter. Phasemeter readings vs. slotted-
line probe positions for setup of (a) are plotted in (b).
Steepest portion of (b) curves is shown expanded in
(c). Maximum rate of change of phase of 50° mm, or
2 per human hair diameter, is easily measured. With
slotted line terminated in 50 ohms, maximum devia-
tion of phase from linear, theoretically 22.8° jor meas-
ured VSWR of 2.26, is measured as 22°. Frequency is
1.003 GHe.
‘of a human hair in the position of the
slotted-line probe was accompanied by
about a 2° phase change, and was easily
resolved by the high-resolution (0.1°)
phasemeter.
With the 50-chm load, the VSWR
was 2.26, Had the VSWR been 1.0, the
phase-vs-position curve would have
CORRECTION
In the article “RFI Measurements Down to
10" kis With. Spectrum i
Vol. 17, No. 7, Mareh, 1966,
high. Optimum LO power is about 8 mi.
cam bo used, bt the thibeen linear, as shown by the dashed
line in Fig. 3¢b). The theoretical maxi
deviation from linear of the
phase curve for a VSWR of 2.26 is
—1
= 228°,
ag=
The measured maximum deviation
shown in Fig. 3(b) is about 22
AMPLIFIER MEASUREMENTS
Fig. 4(b) shows curves of gain, phase,
and group delay versus frequency for
a transistor amplifier stage operating,
1 the 100-12-MIHz range, ‘The curves
were measured with the Vector Volt
meter in the setup of
ed! with previo lable
ods, the time and effort req
take the data were minimal,
ves are show
4(b). With the switch shown
4a) in the closed posit
ig. 4a). Com-
eth
(a)
Tip 608
SIGNAL
GENERATOR
hp S245,
‘COUNTER
the second amplifier stage was reduced
0 zero. The solid curves of Fig. 4(b)
were obtained with the switel open
(second stage gain >1) and the dashed
curves were obtained with the switeh
closed. The difference between the
‘curves shows that the impedance seen
by the first stage has been changed by
the Miller effect of the collectorso.
base capacitance of the second transis:
tor and the gain of the second stag
Besides amplitude and phase curves,
Fig. 4(b) shows group-del
‘which delay distortion produced by the
Miller effect is apparent. A group de
lay curve can be obtained either by
plotting the slope of the phase cu
‘or directly from the phasemeter. By
changing the input frequency in incre
ments of 2.78 KH, or 27.8 kHy, or 278
Kis, ete. the group delay can be read
direetly from the corresponding
changes in the phasemeter readings.
The scale factors will in this ease be
Ips, 100 ns, or 10 ns, etc., per degree,
since 1 ys = 1 degree at 2.78 KHz, and
so on. Group delay information is very
useful in cable testing, where constant
time delay for all frequencies is desir
le.
MEASUREMENTS OF TRANSISTOR AND
NETWORK PARAMETERS
Another important applic:
the new VVM is measuring transistor
gain and other transistor parameters
The wide frequency range of the VVM,
its ability to measure very small
signals, make it wellsuited for tran:
sistor measurements,
Fig. 5(a) shows a test setup which is
being used at -hp- to measure transistor
ng parameters, or «parameters.
The sparameters contain the same in-
ion for
seater
7
Vector
votes
99
7 PO lvoe
Phase / ~
zn S.
Ecol 299 \
S| 229
sol 2
O Babe VN
3 se Jo \\