BD CSampler Software User Guide
3 VALIDATING THE PERFORMANCE OF THE BD
ACCURI C6 FLOW CYTOMETER
Perform a validation of the system at least once each day of use. This ensures that
the flow cytometer is working properly before running experimental samples.
Using the same BD CSampler Software file each day allows identification of trends
over time. When collecting validation data, BD CSampler Software advances to the.
next empty well in row A-B (for 8-peak beads) or C-D (for 6-peak beads). Include the
date in the Sample Naming Field for each day of validation. Start a new validation
bead file when all data wells are full.
Reagents required:
‘Spherotech 8-Peak Validation Beads (PN 653144, supplied with the initial flow cytometer
‘Spherotech 6-Peak Validation Beads (PN 653145, supplied with the initial flow cytometer
+ Sheath fluid: Deionized, filtered water (0.2 ym filter) plus Bacteriostatic
Concentration Solution (PN 653156, supplied with the initial flow cytometer
shipment)
Running Validation Beads (Daily Start-Up)
Setup:
* Verify that the Bead Template has been copied to the BD CSampler Software
computer. The file is located on the BD CSampler Software CD or flash drive and
is on the BD Accuri website
(ntto/www.bdbiosciences.com/support/resources/accuri/index.isp/).
NOTE: The template is set up for the 24-tube rack only.
* Open BD CSampler Software.
* Select File > Open Workspace or Template.
* Inthe Open dialog box, browse to the location of the bead template file and open
the file. If adding to an established BD CSampler Software bead file, browse to
the location of the file.
7820099-01 Rev-2 "7