Professional Documents
Culture Documents
Metrology Practical Question 5 & 9
Metrology Practical Question 5 & 9
Apparatus needed:
1. Gauge blocks
2. Calibrated micrometer
3. Reference surface (e.g. Granite plate )
Calibration procedure:
Clean the Vernier Calipers measuring surfaces, the granite surface, and the
gauge blocks to be used.
Zero the Vernier Caliper at the start and adjust as required by the
manufacturers specifications. If cannot be zeroed, it is then mark as a fail.
Make a visual and touch inspection. Inspect the inside and Outside Jaws,
Depth Gauge and sliding jaw for smooth movement. There should not be any
damage, nicks, or burrs. It should have straight and parallel faces with no
free play over the whole length.
Question 9
National Standard
Secondary Standards
It is made nearly as close as possible to the primary standard with which they are
compared at intervals and records of their derivation from it. It used occasionally for
comparison with the tertiary standard.
Working Standards
These standards are used in laboratories and similar institution. They are derived
from the fundamental standards. They are referred to as the reference standard
and calibration standards.